Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K041130

    Validate with FDA (Live)

    Date Cleared
    2004-06-10

    (41 days)

    Product Code
    Regulation Number
    864.8165
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CDS Hematology Calibrator for red cell, white cell and platelet counting is a device that resembles red cells, white cells and platelets in whole blood specimens and is intended to serve as a calibration standard for automated hematology analyzers, including Beckman Coulter: S-Plus IV-VI series, STKR, JS, JR, ST and JT series, STKS, MAXM, MD, Counto, OnyX, T series and A . T and A . T diff series analyzers; the Abbott Cell-Dyn 1100, Cell-Dyn 1400, 1500, 1600 and 1700 analyzers; and the Danam EXCEL 16 and EXCEL 22 and Coulter Ac T diff analyzers used to count red cells, white cells and platelets. It is a stable suspension of particles whose size, shape, concentration and other characteristics have been precisely and accurately determined.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA for the CDS Hematology Calibrator (CDS CAL). It states that the device is substantially equivalent to legally marketed predicate devices, meaning it can be marketed. However, the document does not contain any information about acceptance criteria or a study proving the device meets those criteria, nor does it include details about performance, sample size, ground truth, or expert qualifications.

    Therefore, I cannot provide the requested information based on the provided text. The document focuses on regulatory approval based on "substantial equivalence" rather than detailed performance metrics and studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K024173

    Validate with FDA (Live)

    Date Cleared
    2003-01-14

    (27 days)

    Product Code
    Regulation Number
    864.8625
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CDS 3-PD Hematology Controls are intended to serve as a means of verifying the results obtained from a hematology analyzer system. This material applies to hematology analyzer systems equipped to measure RBC count and size, WBC count and size, PLT count and size, total HGB concentration, and calculated parameters HCT, MCH, MCHC, RDW, WBC percent and number for lymphocyte, mononuclear and granulocyte populations. The products are designed to be sampled directly into the analyzer system as a sample specimen with minimal operator intervention.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a Food and Drug Administration (FDA) 510(k) clearance letter for the "CDS 3-PD Hematology Controls" device. This letter indicates that the FDA has determined the device is "substantially equivalent" to legally marketed predicate devices.

    However, the letter does not contain any information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert involvement. The 510(k) pathway for medical device clearance primarily relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than requiring extensive clinical trials with pre-defined acceptance criteria and detailed performance studies like a PMA (Premarket Approval) application would.

    Therefore, I cannot extract the requested information from the provided text. The document focuses on regulatory clearance based on substantial equivalence, not on a detailed scientific study demonstrating performance against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993375

    Validate with FDA (Live)

    Date Cleared
    1999-12-20

    (74 days)

    Product Code
    Regulation Number
    864.8150
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CDS LTX-Control Latex Particle Suspension is intended to serve as a calibration standard for the d.c. impedance, r.f. impedance and optical light scatter parameters of the VCS module of Coulter® Models STKS, MAXM and Gen.S Hematology Analyzer systems. Models STKS and MAXM analyzers equipped with 5-part differential capability only require aspiration of the CDS LTX-Control Latex Particle Suspension through the secondary mode aspiration inlet with the analyzer set to function F5S. STKS and Gen. S analyzers equipped with 5-part differential and reticulocyte capability require aspiration of the CDS LTX-Control Latex Particle Suspension through the secondary mode aspiration inlet with the analyzer set to function F57.

    Device Description

    The CDS LTX-Control for red cell, white cell, and platelet counting is a device that resembles red cells, white cells, and platelets, and is used to validate the electronic calibration of the volume, conductivity and light scatter parameters on Coulter® MAXM®, STKS® and GEN.S® instruments intended to count red cells, whire cells, and platelets. It is a suspension of uniform particles whose shape, size, and concentration and other characteristics have been precisely and accurately determined.

    CDS LTX-Control is supplied as ready-to-use latex particles in a buffered suspension fluid. The particles pass through the flow cell producing characteristic electrical signals that are measured as volume, conductivity and light scatter for quality monitoring of the previously mentioned COULTER instruments.

    Matrix base - aqueous, buffered, bacteriostatic and fungistatic medium containing a surfactant.

    Constituents - alkali metal salts, detergents, antimicrobial materials and polymer beads of known size.

    AI/ML Overview

    The document describes the CDS LTX-Control, a latex particle suspension used as a calibrator for hematology analyzers (Coulter MAXM, STKS, and GEN.S). It aims to validate the electronic calibration of volume, conductivity, and light scatter parameters.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Stability Testing: Calibration result must be within ±3% of the previous value from the same lot of material.Met. Stability testing was performed at ambient room temperature, 37°C, and 50°C using pilot production material and compared with fresh material and other commercial controls. The product stability-dating interval is revised upward each month upon successful completion of testing, indicating it consistently met the ±3% criterion. Data tables for two lots (503 and 504) showed stability over 90 days from Day 3, with occasional deviations correlated to instrument malfunction, not product instability.
    Analyzer Calibration Validation: Deviation of ±3% for commercial control materials compared to manufacturer's reference value.Met. A Coulter Model STKS analyzer was calibrated using a standard commercial calibration material and validated by assaying at least three levels of commercial control materials for three cycles each. The averaged values were compared to the manufacturer's reference, and a deviation of ±3% was the accepted range. This step ensured the analyzer was properly calibrated before evaluating the LTX-Control.
    CDS LTX-Control vs. Normal Human Blood: Assigned values for CDS LTX-Control should be within ±10% of the mean average of normal human blood sample results.Met. The assigned values for the CDS LTX-Control were compared to results obtained from normal human blood specimens, and the values assigned were within ±10% of the mean average of the normal sample results.
    NIST-traceable Primary Standard Comparison: Statistical comparison with a NIST-traceable particle standard.Met. Performed by assaying the NIST-traceable primary standard 11 times (discarding the first value, taking the mean of the remaining 10). The CDS LTX-Control assay was then compared to this primary standard. The methodology effectively transfers the known size of the NIST-traceable latex standard to the secondary CDS LTX-Control.
    Calibration Parallelism with Competitive Material: Statistical comparison of repetitive analysis of CDS LTX-Control with competitive commercial material (Latron and Latron Primer).Met. Repetitive analysis of commercial competitive material from the instrument system's manufacturer was statistically compared to repetitive analysis of CDS LTX-Control as an indication of calibration parallelism.
    Light Scatter Stability: Light Scatter parameter requires three to five days of aging in the final solution to reach a stable value.Met. Assay values for a given lot number of CDS LTX-Control are assigned five days after manufacture and packaging, indicated by the vertical blue line on Charts 1 and 2, confirming that the product is allowed to stabilize for this parameter before value assignment. Charts 1 and 2 show the trend for the Light Scatter parameter, with upper and lower limits of ±3% of the mean assay value.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: For stability testing and value assignment, 31 cycles are performed repeatedly for the LTX-Control. The first cycle is discarded, and the mean is calculated from the remaining 30 replicates.
      • For the NIST-traceable primary particle standard, it is assayed 11 times, with the first value discarded, and the remaining 10 values used to calculate a mean.
      • For validation of the analyzer system with commercial controls, at least three levels of commercial control materials are assayed for three cycles each.
      • For comparison with normal human blood, "normal human blood specimens" are used, but a specific number / sample size is not stated beyond "mean average of the normal sample results."
      • Data tables (Tables I and II) are presented for two lot numbers of CDS LTX-Control.
      • Table 100 shows the analysis of the NIST-traceable primary particle standard.
    • Data Provenance: The document does not explicitly state the country of origin of the data. Given the address of the manufacturer (Clinical Diagnostic Solutions, Inc. in Plantation, FL, USA) and the FDA 510(k) submission, it is highly likely the data was generated in the United States. The stability studies refer to "pilot production material." The studies appear to be prospective as they involve specific testing protocols for stability, value assignment, and comparison studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • The document does not explicitly mention the use of human experts or radiologists to establish ground truth for the test set.
    • The ground truth for the CDS LTX-Control is established through comparison with:
      • Previous values from the same lot (for stability).
      • Reference values assigned by the manufacturer for commercial control materials.
      • A NIST-traceable primary particle standard.
      • Mean averages derived from normal human blood specimens.
      • This implies an objective, quantitative ground truth based on established measurement standards and existing commercial controls, rather than subjective expert consensus.

    4. Adjudication Method for the Test Set:

    • No human adjudication method (e.g., 2+1, 3+1) is described.
    • The determination of acceptance is based on pre-defined quantitative criteria (e.g., ±3% deviation, ±10% deviation) and statistical calculations (mean, standard deviation, coefficient of variation, t-test). If an instrument malfunction affects results, recalibration to known standards is used to correct values.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This device is a calibrator for automated hematology analyzers, not an AI diagnostic tool used by human readers. Therefore, the concept of human readers improving with or without AI assistance is not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Yes, the studies described are essentially standalone performance evaluations of the CDS LTX-Control itself and its ability to calibrate the specified automated hematology analyzers. The device is a physical standard, and its performance (e.g., consistency, accuracy relative to other standards) is assessed directly through repetitive measurements on the target analyzers. There is no "human-in-the-loop" directly interacting with the calibrator's output in a diagnostic capacity; rather, the calibrator ensures the instrument's algorithm/measurement system is working correctly.

    7. Type of Ground Truth Used:

    • The ground truth used is primarily objective, quantitative, and traceable to established standards:
      • Manufacturer-assigned reference values for commercial control materials.
      • NIST-traceable primary particle standards (for transferring known particle sizes).
      • Statistical averages/means derived from a large number of replicates and normal human blood specimens, serving as a baseline for expected performance.
      • Instrument calibration standards.

    8. Sample Size for the Training Set:

    • The document describes a calibration and quality control product, not a machine learning algorithm. Therefore, the concept of a "training set" in the context of AI/ML is not applicable. The device itself is a "standard" or "control" that is used to calibrate or train other instruments, rather than being "trained" itself.

    9. How the Ground Truth for the Training Set Was Established:

    • As the concept of a "training set" in the AI/ML sense is not applicable, the ground truth for an AI training set was not established.
    • Instead, the "ground truth" for the CDS LTX-Control's assigned values is established through rigorous comparative studies as outlined in point 7, involving traceability to NIST standards, comparisons with commercial controls, and assessment against normal human blood values on properly calibrated instruments.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1