    K Number
    K073000

    Validate with FDA (Live)

    Device Name
    LUMAPROBE
    Manufacturer
    Date Cleared
    2008-10-01

    (343 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LumaProbe is intended to provide light to the body.
    LumaProbe (Red-631) is generally indicated for the treatment of superficial, benign vascular, and pigmented lesions.
    LumaProbe (Blue-415) is generally indicated for the treatment of dermatological conditions and specially indicated for the treatment of mild to moderate inflammatory acne vulgaris.
    LumaProbe (IR-830) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

    Device Description

    The LumaProbe is a device that utilizes Light Emitting Diodes to provide LED light to the body. The LumaProbe Red, IR and Blue are a visible light source of high spectral purity. They provide uniform or "hot-spot" free illumination. The output is pre-tuned to one wavelength with a narrow spectral bandwidth. The output wavelength of Red is 631 +/- 4nm, IR is 830 +/- nm and Blue is 415 +/- 5nm. The LumaProbe unit contains the power supplies and the control unit; it has the capabilities to power and sequence up to 2 attachable probes depending on the product model, that deliver the light to the skin as they are moved over the skin surface. The STOP button directly on the probes allows the user to immediately remove all power to the probes.

    AI/ML Overview

    The provided text is a 510(k) Summary for the LumaProbe device. It does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria. Instead, it aims to demonstrate substantial equivalence to predicate devices already on the market.

    Therefore, many of the requested categories cannot be filled as they pertain to clinical or performance studies that are not detailed in this submission.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided. The document does not specify quantitative acceptance criteria or report specific performance metrics from a study. It focuses on the LumaProbe's features being "virtually identical" to predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be provided. No test set or study data is described. The submission relies on "comparison of technological differences" and "overall performance characteristics" without detailing a specific study to generate this data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Cannot be provided. No ground truth establishment is described as there is no specific test set or clinical study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Cannot be provided. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Cannot be provided. This device is a light therapy device, not an AI-powered diagnostic tool, so an MRMC study comparing human readers with and without AI assistance is irrelevant and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Cannot be provided. This device is a physical light therapy device, not an algorithm. Therefore, a standalone algorithm performance study is not applicable and not mentioned.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Cannot be provided. No ground truth is established as no specific performance study is detailed.

    8. The sample size for the training set

    • Cannot be provided. No training set is mentioned as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    • Cannot be provided. No ground truth for a training set is relevant or mentioned.

    Summary of the K073600 Submission for LumaProbe:

    The K073600 submission for the LumaProbe device is a 510(k) Premarket Notification which asserts substantial equivalence to previously cleared predicate devices. It does not present novel performance data from de novo clinical studies with specific acceptance criteria that the device proves to meet. Instead, the core argument is that the LumaProbe is technologically similar to existing devices with established safety and effectiveness.

    The relevant section stating this is:

    "Comparison of Technological Differences: The intended use and technological characteristics of the LumaProbe system are virtually identical to the intended use and technological characteristics of the listed equivalent devices. Any differences between the LumaProbe and the equivalent devices have no significant influence on safety or effectiveness of the LumaProbe product."

    "Conclusions: Based upon an analysis of the overall performance characteristics for the LumaProbe, Clareblend, Inc. believes that no significant differences exist between this system and the predicate systems quoted, therefore, the LumaProbe Light Therapy device does not impose any new safety or effectiveness concerns."

    This means the "acceptance criteria" here are implicitly meeting the performance and safety profiles of the predicate devices based on a technological comparison, rather than direct, quantitative testing against specific numbers.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073022

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-10-01

    (341 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Clareblend Probes are intended to provide light to the body.

    Clareblend Probe (Red-631) is generally indicated for the treatment of superficial, benign vascular and pigmented lesions

    Clareblend Probe (Blue-415) is generally indicated for the treatment of dermatological conditions and specially indicated for the treatment of mild to moderate inflammatory acne vulgaris.

    Clareblend Probe (IR-830) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffiess; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

    Device Description

    The Clareblend Probes are hand held devices that utilizes Light Emitting Diodes to provide LED light to the body. The hand hold device contains the power supply. The probe will be simply activated by an alternote action ON/OFF pushbutton switch which is located on the top of the probe. The probe when activated anplies full power (100% duty cycle). The OFF button is provided for the user to immediately remove all power to the probes. The probe housing is 100% machined aluminum and has a recessed clear lexanowindow. The device delivers the light to the skin as it moves over the skin. The output is pre-tuned to one wavele. gth a narrow spectral bandwidth. The output wavelength of Red is 631 +/- 4nm, IR is 830 +/- 5mm and Blue is 415 +/- 5nm.

    AI/ML Overview

    The provided FDA 510(k) Summary for Clareblend LED Probes (K073022) does not contain information about a study with acceptance criteria and reported device performance in the typical sense of a clinical or analytical study.

    This document describes a substantial equivalence determination, meaning the device was deemed equivalent to already-marketed predicate devices. For this type of submission, extensive new clinical performance data is often not required if the new device's technological characteristics and intended use are very similar to established devices.

    Therefore, many of the requested details about acceptance criteria, study sizes, expert involvement, and ground truth are not applicable (N/A) for this particular document and submission type.

    Here's a breakdown based on the information available:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit)Reported Device Performance
    Technological Characteristics:
    - Wavelength output matching predicate devices.- Red: 631 +/- 4nm
    - IR: 830 +/- 5nm
    - Blue: 415 +/- 5nm
    - Device type: Hand-held, LED-based.- Hand-held device utilizing Light Emitting Diodes.
    - Power supply, ON/OFF switch, 100% duty cycle.- Contains power supply, "alternote action ON/OFF pushbutton switch," applies full power (100% duty cycle).
    - Housing material and design.- 100% machined aluminum, recessed clear lexan window.
    Intended Use:
    - Be equivalent to the indications for use of predicate devices:- Clareblend Probe (Red-631): Treatment of superficial, benign vascular and pigmented lesions. (Equivalent to Omnilux Revive in scope, although Omnilux Revive has more general indications).
    - Clareblend Probe (Blue-415): Treatment of dermatological conditions, specifically mild to moderate inflammatory acne vulgaris. (Equivalent to Omnilux Blue).
    - Clareblend Probe (IR-830): Temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness; promoting relaxation of muscle tissue; temporarily increasing local blood circulation. (Equivalent to Omnilux Plus).
    Safety and Effectiveness:
    - No significant differences in safety or effectiveness compared to predicate devices.- "The intended use and technological characteristics of the Clareblend Probes are virtually identical to the intended use and technological characteristics of the listed equivalent devices. Any differences between the Clareblend Probes and the equivalent devices have no significant influence on safety or effectiveness of the Clareblend Probes product." - "no significant differences exist between this system and the predicate systems quoted, therefore, the Clareblend Probes do not impose any new safety or effectiveness concerns."

    Study Details (Not Applicable/Information Not Provided for this type of Submission)

    1. Sample size used for the test set and the data provenance: N/A. This 510(k) relies on substantial equivalence and does not describe a clinical performance study with a test set of data. The "test" was a comparison to predicate devices, not a new data collection.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A. Ground truth establishment for a test set is not described as no new clinical study data evaluation is referenced.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A. No test set requiring adjudication is described.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A. This device is not an AI-assisted diagnostic tool; it's a therapeutic LED light probe. No MRMC study was conducted or mentioned.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A. This is not an algorithm or AI device. Its performance is inherent in its physical properties (wavelength, power) and intended biological effect.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): N/A. For a substantial equivalence claim, the "ground truth" implicitly refers to the established safety and effectiveness profiles of the identified predicate devices (Omnilux Blue K030883, Omnilux Revive K030426, Omnilux Plus K043317).
    7. The sample size for the training set: N/A. There is no mention of a training set as this is not an AI/ML device.
    8. How the ground truth for the training set was established: N/A. Not applicable as no training set is described.

    In summary: The provided document is a 510(k) summary for a device seeking clearance based on substantial equivalence to existing predicate devices. It focuses on demonstrating that the new device's technological characteristics and intended uses are "virtually identical" and pose "no new safety or effectiveness concerns" compared to the predicates. It does not detail new clinical trials or performance studies with specific acceptance criteria beyond the comparison to the predicate devices' established performance and safety profiles.