Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K143396

    Validate with FDA (Live)

    Date Cleared
    2015-01-23

    (58 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to provide physicians with a tool for electromagnetic (EM) tracking of instruments with respect to pre-acquired or real-time data. The device is intended for use in Magnetic Resonance Imaging (MRI) and ultrasound (US) environments.

    Device Description

    The proposed device contains multiple fluid-filled cavities that serve as registration points (markers) in MR. Adhesive tape on the bottom of the proposed device is used to secure the device to the patient. Skin marking areas allow for repositioning in both MR and ultrasound environments. The device also provides a connection point for an electromagnetic sensor accessory. The proposed device is provided non-sterile, is intended for single patient use, and is manufactured of non-magnetic materials. The proposed device enables automatic or manual image fusion of real-time ultrasound to previously acquired MR data sets. The proposed device may also be used to aid in image fusion of real-time ultrasound to previously acquired ultrasound data sets.

    AI/ML Overview

    The provided document (K143396) describes the omniTRAX™ Active Patient Tracker, which is an active fiducial marker. The document focuses on demonstrating substantial equivalence to a predicate device, particularly regarding its safety and compatibility for use in Magnetic Resonance (MR) environments, as well as its intended use in ultrasound (US) environments for electromagnetic (EM) tracking.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based solely on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily focused on safety and compatibility in MR environments, rather than performance metrics like sensitivity, specificity, or accuracy, which are typical for diagnostic AI devices. The device's performance is reported in terms of passing these safety tests.

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    MR Safety & CompatibilityAdherence to specific ASTM standards for MR safety:- RF Heating (ASTM F2182-11a): The device must not pose an unacceptable risk of heating when exposed to Radiofrequency fields in an MRI scanner.- Magnetic Induced Torque (ASTM F2052-06): The device must not experience significant torque (rotational force) when subjected to the magnetic field of an MRI scanner, which could cause movement and potential injury.- Magnetically Induced Displacement Force (ASTM F2213-06): The device must not experience significant translational force (pulling or pushing) when subjected to the magnetic field gradient of an MRI scanner, which could cause it to move or become dislodged.- Magnetic Field Interactions (ASTM F2119-07 - modified): Although specifically for implants, modifications were made for external use, implying assessment of interactions with the magnetic field. The acceptance criterion would be that the device does not show adverse interactions.- RF Heating: "The devices passed the acceptance criteria for RF heating."- Magnetic Induced Torque: "The devices passed the acceptance criteria for... magnetic induced torque."- Magnetically Induced Displacement Force: "The devices passed the acceptance criteria for... magnetically induced displacement force."- Magnetic Field Interactions (modified F2119-07): "The devices passed the acceptance criteria for... [other aspects related to magnetic field interactions]." (Implied from the general statement, as F2119-07 covers this broader category).- Localized Image Artifact: While not an "acceptance criterion" in the sense of a pass/fail for safety, the document notes that "Localized image artifact was observed" and that "information regarding location and size of the artifacts has been included in the Instructions for Use," indicating this was an expected outcome that needed to be addressed through labeling rather than eliminated.
    BiocompatibilityCompliance with ISO 10993 standards for biological evaluation of medical devices (for skin contact):- Cytotoxicity (ISO 10993-5:2009/(R) 2014): Materials should not be toxic to cells.- Sensitization & Irritation (ISO 10993-10:2010): Materials should not cause allergic reactions or skin irritation. The device is intended for "limited contact duration (<24 hours) for surface devices (skin).""Biocompatibility testing was also completed for patient-contacting materials according to ISO 10993-5:2009/(R) 2014 and ISO 10993-10:2010." (Implied pass, as non-compliance would prevent substantial equivalence).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a separate "test set" in the context of a clinical study or performance evaluation with human data. The testing described is non-clinical testing of the device itself (materials, physical interactions).

    • Sample Size for Non-Clinical Testing: Not explicitly stated numerically. The text refers to "the devices" (plural), implying multiple units of the omniTRAX™ Active Patient Tracker were subjected to the tests.
    • Data Provenance: The data provenance is from laboratory testing (non-clinical) conducted against engineering standards (ASTM, ISO). There is no mention of human subject data, country of origin, or retrospective/prospective nature as would be relevant for clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to the type of testing described in the document. The ground truth for MR safety and biocompatibility testing is defined by the established ASTM and ISO standards themselves, which specify procedures and acceptable limits for physical and biological interactions. Experts are involved in conducting and interpreting these standardized engineering tests, but not in establishing a "ground truth" through consensus in the way clinical diagnostic studies do.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (like 2+1 or 3+1) is not applicable to the non-clinical engineering and biological safety testing performed. The results are typically quantitative measurements compared against predefined thresholds in the standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done and is not described in this document. This type of study is relevant for evaluating the impact of AI algorithms on human reader performance in diagnostic tasks, which is not the scope of this device (an active fiducial marker).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is a physical medical instrument (an active fiducial marker) for tracking, not an AI algorithm. Therefore, the concept of "standalone algorithm performance" is not applicable. The device's function is to serve as a physical marker for electromagnetic tracking in conjunction with imaging modalities and human intervention.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical testing is defined by adherence to established engineering standards (ASTM for MR safety, ISO for biocompatibility). These standards set the benchmarks and methodologies for determining safety and compatibility. For example, the ground truth for RF heating is that the temperature rise must be within acceptable limits as defined by ASTM F2182-11a.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a physical medical device, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As stated above, this device is not an AI algorithm requiring a training set or associated ground truth establishment for training.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1