    K Number
    K140321

    Validate with FDA (Live)

    Date Cleared
    2014-04-17

    (58 days)

    Product Code
    Regulation Number
    870.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Better-Bladder™ is a device that isolates from blood contact when measurements of blood pressure in extracorporeal circuits are made during term procedures. The pressure signal can be used to control pump speed. It is also used as an inline reservoir to provide compliance in the circuit during short and long term procedures.

    Device Description

    The Better-Bladder™ (BB) is a length of standard perfusion tubing, a portion of which has been processed to form a sausage shaped balloon with a thin wall that is then sealed within a clear, rigid housing. This device can be used as an inline blood reservoir during extracorporeal circulation. It can also be used to monitor the blood pressure noninvasively: because the pressure of blood flowing inside the tubing is transmitted across the thin wall to the housing chamber, this pressure can be monitored via a pressure monitoring line to a pressure sensor. The BB thereby serves as a pressure transducer protector, isolating the blood from the pressure sensor and allowing noninvasive pressure measurements, which in turn can be used as input signals to control the pump speed.

    AI/ML Overview

    This 510(k) notification describes a medical device, the Better-Bladder™ BB14-72, and its substantial equivalence to a predicate device, focusing on its physical characteristics and performance in a mock circulatory system. The document does not describe the acceptance criteria and study design elements typically associated with AI/ML-based medical devices or diagnostics. Therefore, many of the requested fields cannot be directly answered from the provided text.

    Here's an analysis based on the provided text, addressing the points where information is available and noting when it is not:

    Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for the BB14-72 appears to be its ability to mitigate negative pressure spikes due to venous line obstruction, performing "as well or better than the predicate device," and providing "more time for a pump flow controller to respond than the predicate device."

    Acceptance Criteria (Implied)Reported Device Performance
    Lower negative pressure spike than predicate device."The negative pressure spike for the BB14-72 was lower than that for the predicate device."
    Longer time for pressure to reach steady state than predicate device."the time it took for the pressure to reach a steady state was longer with the BB14-72 than the predicate device."
    Substantially equivalent in design, intended use, technological characteristics, and performance characteristics to the predicate device.Concluded as "substantially equivalent" based on various comparisons and performance data.
    Materials identical to predicate device."The materials of the BB14-72 are identical to those of the predicate device."
    Method of manufacture same as predicate device."The method of manufacture is also the same for both the BB14-72 and the predicate device."

    Study Details

    The provided document describes a performance study to demonstrate the BB14-72 device's mechanical characteristics, not a clinical study involving human patients or complex AI/ML algorithms.

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: Not explicitly stated as a numerical sample size. The testing involved comparing the BB14-72 against the predicate device (BB14) and a 1/4" standard perfusion tubing in a "mock extracorporeal circuit." The graphs show results for three devices/configurations.
      • Data Provenance: The study was conducted in a "mock extracorporeal circuit," which implies a laboratory setting. There is no mention of country of origin of data or whether it was retrospective or prospective. It's a prospective engineering performance test.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. This was a physical device performance test using a mock circuit, not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" was mechanical measurements (pressure and time).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. See point 2.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device, and no MRMC study was performed.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI/ML algorithm; it is a physical medical device. The "standalone" performance shown is purely mechanical, not algorithmic.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" was direct physiological measurements (pressure and time) obtained from a simulated environment (mock extracorporeal circuit) under controlled conditions.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device, and therefore no training set was used.
    8. How the ground truth for the training set was established:

      • Not applicable. See point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K012942

    Validate with FDA (Live)

    Date Cleared
    2001-11-05

    (66 days)

    Product Code
    Regulation Number
    870.4400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V-Bag™ with the Vac-Box™ are designed for use in an extracorporeal circuit to provide a volume reservoir, compliance, and assist in removing air during the vacuum assisted venous drainage operating mode.

    Device Description

    Not Found

    AI/ML Overview
    1. Acceptance Criteria and Reported Device Performance:

    The provided document does not contain explicit acceptance criteria and corresponding reported device performance metrics in a tabular format. The document is a 510(k) clearance letter, which primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance study results against predefined acceptance criteria.

    1. Sample Size for Test Set and Data Provenance:

    The document does not specify the sample size used for any test set or the data provenance (e.g., country of origin, retrospective/prospective) because it is a 510(k) clearance letter, not a clinical study report. Such details would typically be found in a separate submission, often summarized in the "substantial equivalence" section or a specific performance testing report.

    1. Number of Experts and Qualifications for Ground Truth Establishment:

    The document does not mention the number of experts used or their qualifications for establishing ground truth for any test set. This information is not typically part of a 510(k) clearance letter for a device like this, which is a physical medical device (blood reservoir) rather than an AI/software device that requires human interpretation for ground truth.

    1. Adjudication Method for Test Set:

    The document does not specify any adjudication method for a test set. This is not applicable given the nature of the device and the content of the 510(k) clearance letter.

    1. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. This type of study is usually relevant for diagnostic imaging or AI devices where human reader performance is compared with and without AI assistance. This device (V-Bag and Vac-Box) is a cardiopulmonary bypass blood reservoir, not a diagnostic or AI-powered interpretative tool.

    1. Standalone (Algorithm Only) Performance Study:

    The document does not mention a standalone (algorithm only) performance study. This is not applicable as the device is a physical medical device, not an algorithm.

    1. Type of Ground Truth Used:

    The document does not explicitly state the type of ground truth used. For a physical device like a blood reservoir, ground truth would typically refer to objective measurements of device performance, such as volume capacity, air removal efficiency, biocompatibility, and structural integrity, against predefined engineering specifications and international standards. This information would be found in design verification and validation testing reports, not typically detailed in a 510(k) summary letter.

    1. Sample Size for Training Set:

    The document does not mention a sample size for a training set. This concept (training set) is relevant for machine learning or AI models, which is not applicable to the V-Bag and Vac-Box described.

    1. How Ground Truth for Training Set Was Established:

    The document does not mention how ground truth for a training set was established, as it is not applicable to this physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K001822

    Validate with FDA (Live)

    Device Name
    THE V-BAG
    Date Cleared
    2000-07-07

    (21 days)

    Product Code
    Regulation Number
    870.4400
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K981284

    Validate with FDA (Live)

    Date Cleared
    1998-08-12

    (126 days)

    Product Code
    Regulation Number
    870.4400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Better-Bladder™ is a device that isolates pressure transducers from blood contact when measurements of blood pressure in extracorporeal circuits are made during short and long term procedures. The pressure signal can be used to control pump speed. It is also used as an inline reservoir to provide compliance in the circuit during short and long term procedures.

    Device Description

    The BB consists of a unitary length of tubing, a section of which has been processed to form an enlarged diameter with a thinner wall. The enlarged section is very responsive to pressure differences across its thin wall. Because the entire device is made from a single length of smooth perfusion tubing, there are no connections between the thinker wall tubing portions. This unique design eliminates both physical and material (chemical) discontinuities. The thin walled section in its housing can withstand over 1500 mmHg (three times the maximum pressure expected under worst-case clinical conditions). The BB does not incorporate vent tubes for air aspiration or drug infusion. These can be added by placing a perfusion connector with a dual Luer fitting (e.g. Avecor Model # DL1414) at the inlet to the BB, see Fig. 7b.

    AI/ML Overview

    Here's an analysis of the provided text regarding the "Better-Bladder™" device's acceptance criteria and studies:

    Summary of Device Acceptance Criteria and Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device PerformanceStudy Type/Evidence
    Functional EquivalencePressure measurement noninvasively with BB vs. invasively.Equivalent to invasively measured pressure.Bench tests
    Performance ComparisonComparable or superior performance to predicate devices.Comparable or superior to predicate devices (Quest PMS-2, SciMed R30 Assist Reservoir).Bench tests
    Pressure Withstand (Thin-Walled Section)Withstand over 1500 mmHg (3x max clinical pressure).Withstood over 1500 mmHg.Bench tests
    Internal Pressure Test (Individual Unit)Withstand an internal pressure of 900 mmHg (1.8x max clinical pressure).Each BB individually tested to 900 mmHg.Individual unit testing during manufacturing
    Housing Pressure Test (Individual Unit)Withstand a housing pressure of 900 mmHg.Each BB individually tested to 900 mmHg.Individual unit testing during manufacturing
    Sterilization, Sterilant Residuals, PyrogenicityCompliance with established standards.See K964337 (implies previous FDA clearance or established protocol).Cross-reference to previous submission
    BiocompatibilityCompliance with ISO 10993 Part-1.Blood path material (polyvinyl chloride) passed biocompatibility tests in accordance with ISO 10993 Part-1.Material testing (implied by passing standard)
    General SafetyCompliance with established standards.See K964337.Cross-reference to previous submission
    Quality AssuranceCompliance with established standards.See K964337.Cross-reference to previous submission
    Design Integrity (Bond)Bond between tubing and housing external to blood path.Bond is external to blood path, no blood compatibility concerns.Design characteristic (verified by inspection/design review)

    Detailed Study Information:

    1. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • The document primarily refers to "bench tests" and "individual unit testing" for functional evaluations. Specific sample sizes for these tests are not provided.
      • The data provenance is not explicitly stated (e.g., country of origin). The testing appears to be prospective as it describes evaluations performed on the Better-Bladder™ to demonstrate its performance against predefined criteria and predicate devices.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • This device is a physical medical device (pressure monitoring/blood reservoir) and the "ground truth" for its functional performance (e.g., pressure measurement accuracy, pressure withstand, biocompatibility) is established through objective physical measurements and adherence to engineering/biological standards, not through expert review of data like imaging. Therefore, the concept of "experts establishing ground truth for a test set" in the context of human interpretation is not applicable here. The "experts" involved would be engineers, materials scientists, and quality control personnel following established protocols.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Again, this device is a physical, non-AI medical device. The "adjudication method" in the context of human interpretation of results (like in clinical trials or image analysis) is not applicable. Objective measurements and adherence to predetermined specifications are used.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This is a physical device, not an AI-assisted diagnostic tool involving human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in spirit, a standalone performance was done, as the device's functional integrity and performance were evaluated on its own (bench tests, individual unit tests) against established engineering and biological standards, without human interpretation in the loop as part of its primary function. The device's function is to objectively measure pressure and act as a reservoir, not to process data for human interpretation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for the device's performance is based on objective engineering measurements (e.g., pressure equivalence to invasive methods, maximum pressure withstand, material properties conforming to ISO 10993 Part-1). For comparison, the predicate devices themselves serve as a benchmark for performance.
    7. The sample size for the training set:

      • Not applicable. This device does not use machine learning or AI, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. This device does not use machine learning or AI, so there is no "training set" or ground truth for one.
    Ask a Question

    Ask a specific question about this device

    K Number
    K964205

    Validate with FDA (Live)

    Device Name
    BETTER-TUBING
    Date Cleared
    1998-04-08

    (537 days)

    Product Code
    Regulation Number
    870.4210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Better-Tubing™ is designed for use as a blood compatible conduit for pumping blood in a roller pump during cardiopulmonary bypass or other extracorporeal applications of up to 6 hours.

    Device Description

    The Better-Tubing™ (BT) is a 6 ft. length of ¼' ID tubing for use in an extracorporeal cardiopulmonary bypass circuit. It is biocompatible, has high tensile strength, and has many advantages over standard tubing made from polyvinyl chloride (PVC).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Better-Tubing™ device, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance ClaimBetter-Tubing™ Reported Performance
    BiocompatibilityPassed all ISO 10993 Part-1 Biocompatibility Tests (Intracutaneous Toxicity, Cytotoxicity, Sensitization, Systemic Toxicity, Genotoxicity, Implantation with Histopathology, Hemolysis, Whole Blood Coagulation Time, Pyrogenicity)
    Effect of Temperature on FlowMinimally affected by fluid temperature. Flow differed by no more than 3% at 22±2°C and 37±2°C.
    Hemolysis (compared to Tygon® tubing)Significantly less hemolytic than Tygon® tubing with standard occlusion (p<0.0046). No significant difference from Tygon® tubing set nonocclusively. IH (% STD-SO) of 30% ± 4% (BT-SO) and 35% ± 9% (BT-NO) vs. 100% (TY-SO).
    Pressure ResistanceAble to withstand 2500 mmHg (50 psi) for 60 min without leaks or bursting. Pressure decreased only 2% over 60 min.
    Tensile Strength (compared to predicate tubing)3 to 6 times greater than the predicate tubing.
    Pumping Life (compared to standard tubing)Equal to or greater than standard tubing. Decreases in flow after 6 hours were the same as Super-Tygon®.
    Abrasion Resistance / SpallationHigh abrasion resistance and thus lower spallation.
    Material FormulationDiffers from Tygon® in its material formulation. (Implicit acceptance: new material meets performance)
    Intended Use (Blood compatible conduit for pumping blood in a roller pump during cardiopulmonary bypass or other extracorporeal applications of up to 6 hours.)Met, as determined by the FDA's substantial equivalence finding.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Effect of Temperature on Flow: 3 BT tubes were tested.
    • Hemolysis Tests:
      • Number of circuits: Four identical circuits were tested simultaneously.
      • Tubing samples: Four tubings (Tygon-SO, Tygon-NO, Better-Tubing-SO, Better-Tubing-NO) were tested. This implies 'n=4' for each group as shown in Table 3.
      • Blood source: Fresh heparinized bovine blood obtained from a slaughterhouse.
    • Pressure Tests: A 6 ft. length of Better-Tubing™ was tested.

    Data Provenance: The studies appear to be primarily prospective, in-vitro (laboratory) tests conducted by the manufacturer, Circulatory Technology Inc. The blood for hemolysis tests was bovine and obtained from a slaughterhouse. The general provenance is implied to be within the US, corresponding to the FDA submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is a component (tubing) and the reported studies are performance tests (e.g., flow, pressure, hemolysis) against established benchmarks and predicate devices, not diagnostic assessments requiring expert interpretation of clinical images or data. Ground truth for these tests is based on objective measurements and established scientific protocols (e.g., ISO standards, spectrophotometric assays for hemolysis).

    4. Adjudication Method for the Test Set

    Not applicable. The reported studies are objective performance tests, not studies requiring expert adjudication of qualitative outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a medical device (tubing), not an AI/software device that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a medical device (tubing), not an algorithm. The reported tests are standalone performance tests of the tubing itself.

    7. The Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Objective Measurements: Flow rates (ml/min), pressure (mmHg), temperature (°C), time (min), percentage changes.
    • Established Test Standards: ISO 10993 Part-1 for biocompatibility.
    • Reference Devices/Materials: Comparison against Tygon® tubing (predicate device) and Super-Tygon® for pumping life.
    • Biochemical/Physical Assays: Spectrophotometric assay for plasma free hemoglobin to quantify hemolysis.

    8. The Sample Size for the Training Set

    Not applicable directly. This is a physical medical device. "Training set" typically refers to data used to train AI/machine learning models. For a physical device like this tubing, the "development" or "optimization" process would involve materials science, engineering, and manufacturing trials, which are not outlined in terms of a formal "training set" like in AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as explained in point 8. The "ground truth" for the development of such tubing would be the desired performance specifications (e.g., target tensile strength, desired flow characteristics, acceptable hemolysis levels) that the engineers aimed to achieve through material selection and design. These specifications would be derived from clinical needs, existing standards, and competitor analysis.

    Ask a Question

    Ask a specific question about this device

    K Number
    K964337

    Validate with FDA (Live)

    Date Cleared
    1997-05-23

    (204 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Better-Bladder™ is a device that isolates pressure transducers from blood contact when measurements of blood pressure in extracorporeal circuits are made. The noninvasive pressure obtained. with the BB can be used to control pump speed. It is also used as an inline reservoir to provide compliance in the circuit during routine bypass procedures.

    Device Description

    The Better-Bladder™ is a device that isolates pressure transducers from blood contact when measurements of blood pressure in extracorporeal circuits are made. The noninvasive pressure obtained. with the BB can be used to control pump speed. It is also used as an inline reservoir to provide compliance in the circuit during routine bypass procedures.

    AI/ML Overview

    I'm sorry, but the provided text does not contain the detailed information necessary to answer your request regarding acceptance criteria and study details for the "Better-Bladder™" device. The document is a 510(k) clearance letter from the FDA, which confirms that the device is substantially equivalent to a predicate device and can be marketed. It lists the device's indications for use but does not include:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size for the test set or data provenance. The letter mentions "premarket notification submission" but doesn't detail the study itself.
    3. Number or qualifications of experts used to establish ground truth.
    4. Adjudication method for the test set.
    5. Information on whether a multi-reader multi-case (MRMC) comparative effectiveness study was done or its effect size.
    6. Information on whether a standalone (algorithm only) performance study was done.
    7. The type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    The document primarily focuses on regulatory clearance based on substantial equivalence, not a detailed scientific study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K961594

    Validate with FDA (Live)

    Device Name
    BETTER-VENTER
    Date Cleared
    1996-09-05

    (134 days)

    Product Code
    Regulation Number
    870.4210
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1