Found 2 results

510(k) Data Aggregation

    K Number
    K140703

    Validate with FDA (Live)

    Device Name
    OPA 30 SOLUTION
    Date Cleared
    2014-08-19

    (152 days)

    Product Code
    Regulation Number
    880.6885
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPA 30 Solution is a high level disinfectant for reprocessing heat sensitive semi-critical medical devices for which sterilization is not suitable, and when used according to the Directions for Use. OPA 30 Solution may be used of reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20°C (68°F), for a reuse period not to exceed 30 days. OPA 30 Solution may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25°C), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an immersion time of at least 5 minutes at a minimum of 25℃ (77ºF), for a reuse period not to exceed 14 days.

    Device Description

    The OPA 30 Solution is a clear, pale blue liquid with a pH of 7.5. It contains 0.60% ortho-Phthalaldehyde in an aqueous base containing buffers, chelating agents and a corrosion inhibitor. It is stable at 15-30° C (59-86° F) for its labeled shelf life of 24 months. OPA 30 Solution may be used or reused for manual (up to 30 days) or automated reprocessing (up to 14 days), according to the Directions for Use. OPA 30 Solution must be used at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with the immersion time and temperature specified for the disinfection method.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "OPA 30 Solution," a liquid chemical sterilant/high-level disinfectant. The document focuses on demonstrating substantial equivalence to a predicate device (Opaciden® Solution, K070627). Crucially, this document is for a chemical disinfectant, not an AI/ML-driven device, and therefore the concepts of "acceptance criteria," "device performance," "sample size for test set," "data provenance," "number of experts for ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "type of ground truth," and "sample size for training set" are not applicable in the context of an AI/ML device.

    The closest analogous information for a chemical disinfectant would be:

    1. A table of acceptance criteria and the reported device performance:

    The document implicitly defines "acceptance criteria" through comparison to the predicate device and adherence to FDA guidance for liquid chemical sterilants/high-level disinfectants. The "reported device performance" is largely established by stating that the chemical formulation is identical to the predicate and that previous testing (for the predicate) demonstrated acceptable performance. The key performance outcome tested explicitly for the new device is extended reuse periods.

    Acceptance Criteria (Implied by Predicate & Guidance)Reported Device Performance (OPA 30 Solution)
    Microbiological Efficacy:
    - Vegetative organisms effectivenessEffective against: Staphylococcus aureus, Salmonella choleraesuis, Pseudomonas aeruginosa, Mycobacterium bovis (Identical to predicate)
    - Fungi effectivenessEffective against: Trichophyton mentagrophytes (Identical to predicate)
    - Viruses, non-enveloped effectivenessEffective against: Poliovirus Type 1, Rhinovirus Type 42, Hepatitis A (AER) (Identical to predicate)
    - Viruses, enveloped effectivenessEffective against: Avian influenza, Hepatitis B, Influenza Type A (Hong Kong), HIV-1 (Identical to predicate)
    - Spores effectivenessEffective against: Herpes simplex Type 1, Clostridium sporogenes, Bacillus subtilis (Identical to predicate)
    Chemical Composition:
    - Active ingredient concentration0.60% ortho-Phthalaldehyde (Identical to predicate)
    - Inert ingredients percentage99.40% (Identical to predicate)
    - pH range7.5 (Identical to predicate)
    - Stability at specified temperatureStable at 15-30°C (59-86°F) for labeled shelf life (Identical to predicate)
    Minimum Effective Concentration (MEC):0.3% (Identical to predicate)
    Reuse Period:
    - Manual reprocessing reuse period30 Days Manual (Expanded from 14 days in predicate, confirmed by "re-use for manual processing" testing)
    - Automated reprocessing reuse period14 Days Automated (Identical to predicate)
    Toxicity:Positive for acute oral toxicity, moderate skin irritation. Negative for dermal and cytotoxicity. Probable eye irritant. (Identical to predicate)
    Device Compatibility:Olympus and Pentax endoscopes are compatible. (Identical to predicate)
    - Material compatibilityMaterials previously cleared in predicate and Acrylonitrile Butadiene Styrene (ABS) (Expanded from predicate, confirmed by "material compatibility" testing)
    Shelf Life:24 month shelf life for a closed container (Expanded from 15 months in predicate, confirmed by "stability" testing)
    Chemical Indicator:Opaciden OPA Reagent Strips (Identical to predicate)

    2. Sample sized used for the test set and the data provenance:

    This information is not provided in a way that defines "test set" as it would for an AI/ML device. The document states that performance testing was performed for "stability, re-use for manual processing and material compatibility." However, specific sample sizes (e.g., number of re-use cycles, number of material samples) and the provenance (country of origin, retrospective/prospective) are not detailed. The previous testing for microbiological efficacy was submitted and cleared as part of the predicate device (K070627). This previous testing would have been prospective studies following established disinfectant testing protocols.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable to a chemical disinfectant.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable to a chemical disinfectant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable, as this is not an AI/ML device. The "standalone" performance for a chemical disinfectant would be its efficacy in killing microorganisms under specific conditions, which was tested.

    7. The type of ground truth used:

    For a chemical disinfectant, "ground truth" would relate to the successful inactivation or killing of microorganisms. This is established through standardized laboratory testing (e.g., AOAC methods) where the presence or absence of viable microorganisms after exposure to the disinfectant is directly measured. This is akin to definitive pathology/microbiology laboratory results.

    8. The sample size for the training set:

    Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070627

    Validate with FDA (Live)

    Date Cleared
    2007-08-02

    (149 days)

    Product Code
    Regulation Number
    880.6885
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Opaciden Solution is a high level disinfectant for reprocessing heat-sensitive medical devices for which sterilization is not suitable, and when used according to the Directions for Use. Opaciden may be used or reused at or above its Minimum Recommended Concentration (MRC) of 0.3%, as determined by Opaciden OPA Reagent Strips, in manual reprocessing with an immersion time of at least 12 minutes at a minimum of 20° C (68° F), for a reuse period not to exceed 14 days. Opaciden may also be used or reused in a legally marketed automatic endoscope reprocessor (that can be set to a minimum of 25° C), at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with an immersion time of at least 5 minutes at a minimum of 25° C (77° F), for a reuse period not to exceed 14 days.

    Device Description

    Opaciden Solution is a clear, pale blue liquid with a pH of 7.5. It contains 0.60% ortho-Phthalaldehyde in an aqueous base containing buffers, chelating agents and a corrosion inhibitor. It is stable at 15-30°C (59-86°F) for its labeled shelf life. The product may be used or reused (up to 14 days) for manual or automated reprocessing, according to the Directions for Use. Opaciden Solution must be used at or above its Minimum Recommended Concentration (MRC), as determined by Opaciden OPA Reagent Strips, with the immersion time and temperature specified for the disinfection method.

    AI/ML Overview

    The provided text describes a 510(k) summary for a high-level disinfectant called Opaciden Solution. However, it does not contain the detailed information requested regarding acceptance criteria and a study proving the device meets those criteria, as typically found in a clinical study report or a more comprehensive performance study summary.

    The document primarily focuses on establishing substantial equivalence to a predicate device (CIDEX® OPA Solution) and describes the device's intended use and technological characteristics. It mentions "performance testing" but does not provide specific acceptance criteria or detailed results.

    Here's a breakdown of what can and cannot be extracted from the provided text, based on your request:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. While it states, "All test results were acceptable and demonstrated performance in accordance with the product labeling," it does not specify what those acceptance criteria were (e.g., specific log reduction values for microorganisms) or the quantitative performance results (e.g., "achieved X log reduction for Y organism").

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. The text mentions "tested, in both the manual and automated reprocessing modes," but does not give details about the samples (e.g., number of devices disinfected, types of microorganisms tested, or specific test conditions). It also does not specify data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. For a chemical disinfectant, "ground truth" typically refers to the confirmed microbial reduction or inactivation, which is determined through laboratory assays, not by expert consensus on images or clinical assessments. The document doesn't detail the methodology for establishing the efficacy ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation of medical images or clinical outcomes, where disagreements need to be resolved. For a disinfectant performance study, the outcome is usually based on quantitative laboratory measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. An MRMC study is relevant for medical imaging AI devices where human readers interpret cases. Opaciden Solution is a high-level disinfectant, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This question is for AI algorithms. Opaciden Solution is a chemical disinfectant. Its performance is inherent to the chemical formulation and its interaction with microorganisms, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for a chemical disinfectant's performance would be established through microbiological assays, demonstrating the reduction or inactivation of specific challenge microorganisms (e.g., bacteria, viruses, fungi, spores) to a certain log reduction level, as specified by regulatory standards for high-level disinfectants. However, the document does not explicitly state the type of ground truth used, only that "All test results were acceptable."

    8. The sample size for the training set

    This information is not applicable/provided. Training sets are relevant for machine learning or AI models. Opaciden Solution is a chemical product, not an AI system.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. Same as above, training sets and their ground truth establishment are for AI/ML models.

    In summary:

    The provided 510(k) summary for Opaciden Solution indicates that performance testing was conducted according to FDA guidance for high-level disinfectants. It claims that "All test results were acceptable and demonstrated performance in accordance with the product labeling, as well as substantial equivalence to the predicate device." However, it lacks the specific details of these performance tests, including quantitative acceptance criteria, actual performance metrics, sample sizes, and detailed methodology, which would be necessary to answer the questions comprehensively. The document is a summary focused on regulatory clearance based on substantial equivalence, not a detailed scientific study report.

    Ask a Question

    Ask a specific question about this device