510(k) Data Aggregation

    K Number
    K981698

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-07-10

    (57 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chiron Diagnostics ACS:180 BR is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the Chiron Diagnostics ACS:180® Automated Chemiluminescence Systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.

    Device Description

    The Chiron Diagnostics ACS:180 BR assay is a fully automated, competitive, chemiluminescent assay. One reagent, designated Lite Reagent, is composed of a mouse monoclonal antibody specific for CA 27.29, labeled with acridinium ester. The antibody used in the assay, MAb B27.29, binds to a peptide epitope in the tandem repeat region of the MUC-1 gene product. The Solid Phase is composed of purified breast cancer antigen (CA 27.29) which is covalently coupled to paramagnetic particles (PMP). After onboard pretreatment, the patient serum sample is incubated with both reagents simultaneously for 7.5 minutes.

    The ACS: 180 system automatically performs the following steps:

    • dispenses sample and Pretreatment Reagent into a cuvette .
    • dispenses Lite Reagent and Solid Phase and incubates for 7.5 minutes u
    • separates, aspirates and washes the cuvettes
    • dispenses reagents which initiate the chemiluminescent reaction
    • = reports results

    An inverse relationship exists between the concentration of CA27.29 in a sample and the relative light units (RLU) detected by the system.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modified in vitro diagnostic device, the ACS:180 BR, which is an automated chemiluminescence system for detecting cancer antigen CA 27.29 in human serum.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" but rather focuses on demonstrating "substantial equivalence" to a predicate device. The primary performance metric reported is the correlation between the modified assay and the predicate device.

    Performance MetricAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device PerformanceComments
    Correlation ($r^2$)A high coefficient of determination, ideally close to 1, demonstrating a strong linear relationship with the predicate device.$>0.99$This indicates a very strong positive linear correlation between the modified and predicate assays.
    SlopeA slope close to 1, indicating that the modified assay measures values very similarly to the predicate device across the range.$1.0408$Close to 1, suggesting good agreement in the rate of change between the two methods.
    Y-interceptA Y-intercept close to 0, indicating minimal constant bias between the modified assay and the predicate device.$0.9907$ U/mLRelatively close to 0, suggesting a small positive bias but overall good agreement with the predicate.
    Assay RangeMust match the predicate device to ensure comparable applicability.$3.5$ U/mL to $450$ U/mLMatches the predicate device.
    Upper Limit of NormalMust match the predicate device, as it defines the clinical cutoff.$38.6$ U/mLMatches the predicate device, with a note that individual labs should determine their own reference ranges.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Over 875 specimens.
    • Data Provenance: The document does not specify the country of origin. It does not explicitly state whether the data was retrospective or prospective, but it implies a retrospective comparison of previously collected specimens, given the statement "These specimens had CA 27.29 levels that spanned the range from 3.68 U/mL to 420 U/mL."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The study described is a method comparison study between a modified device and a predicate device, not typically requiring expert-established "ground truth" in the same way an imaging or diagnostic AI device would. The "ground truth" here is the measurement provided by the existing, legally marketed predicate device. Therefore, no information is provided regarding expert qualifications or the number of experts used to establish ground truth in this context.

    4. Adjudication Method for the Test Set

    Not applicable. This was a direct comparison of measurements between two diagnostic assays, not a study requiring adjudication of interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This is a comparison between two automated diagnostic assays, not a study involving human readers or AI assistance for interpretation.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this was a standalone performance comparison study of the device (an automated chemiluminescence system) against its predicate device. The results reported are for the device (modified assay) performing on its own, without human intervention in the measurement process beyond standard laboratory procedures.

    7. The Type of Ground Truth Used

    The "ground truth" for this comparative study was the results obtained from the predicate device (the original ACS:180 BR assay). The objective was to show that the modified assay yielded results substantially equivalent to the established predicate.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI. This is a traditional in vitro diagnostic device modification, where performance is typically validated through comparison to a well-established method, not through machine learning training and testing paradigms. Therefore, information on a training set size is not applicable/provided.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set (in the machine learning sense) is described or implied for this type of device validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K980190

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-03-06

    (45 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chiron Diagnostics ACS:180 BR is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the Chiron Diagnostics ACS 180: Automated Chemiluminescence Systems. The test is intended for use as an aid in monitoring patients previously treated for Stage III breast cancer. Serial testing for CA 27,29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression of disease in response to treatment.

    Device Description

    The Chiron Diagnostics ACS:180 BR assay is a fully automated, competitive, chemiluminescent assay. One reagent, designated Lite Reagent, is composed of a mouse monoclonal antibody specific for CA 27.29, labeled with acridinium ester. The antibody used in the assay, MAb B27.29, binds to a peptide epitope in the tandem repeat region of the MUC-1 gene product. The Solid Phase is composed of purified breast cancer antigen (CA 27.29) which is covalently coupled to paramagnetic particles (PMP). The patient serum sample is incubated with both reagents simultaneously for 7.5 minutes. The ACS: 180 system automatically performs the following steps: dispenses sample into a cuvette, dispenses Lite Reagent and Solid Phase and incubates for 7.5 minutes, separates, aspirates and washes the cuvettes, dispenses reagents which initiate the chemiluminescent reaction, reports results. An inverse relationship exists between the concentration of CA 27.29 in a sample and the relative light units (RLU) detected by the system.

    AI/ML Overview

    Chiron Diagnostics ACS:180 BR - Acceptance Criteria and Study Details

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state formal "acceptance criteria" in a numerical or target-based format for the clinical performance of the ACS:180 BR. Instead, it presents the assay's performance characteristics and then concludes that these were "substantially equivalent" to the predicate device. Therefore, the "acceptance criteria" can be inferred as achieving substantial equivalence to the predicate device's performance.

    Performance MetricInferred Acceptance Criteria (Substantial Equivalence to Predicate Device)Reported Device Performance (ACS:180 BR)
    For monitoring recurrence (1 Value > ULN):Similar sensitivity, specificity, PPV, NPV to predicate.Sensitivity: 60% (39, 79), Specificity: 93% (87, 96), PPV: 60% (39, 79), NPV: 93% (87, 96)
    For monitoring recurrence (2 Consecutive Values > ULN):Similar sensitivity, specificity, PPV, NPV to predicate.Sensitivity: 48% (28, 69), Specificity: 98% (94, 100), PPV: 80% (52, 96), NPV: 91% (85, 95)
    For managing metastatic disease (Sensitivity to Change):Similar sensitivity to predicate.Sensitivity to Change in Disease Status: 73% (58, 85)
    For managing metastatic disease (Specificity to Change):Similar specificity to predicate.Specificity to Change in Disease Status: 71% (53, 85)
    For managing metastatic disease (PPV to Change):Similar PPV to predicate.Positive Predictive Value to Change in Disease Status: 77% (61, 88)
    For managing metastatic disease (NPV to Change):Similar NPV to predicate.Negative Predictive Value to Change in Disease Status: 67% (49, 81)
    Correlation with Predicate Device (203 specimens):Strong correlation (e.g., r > 0.90, slope ~ 1, y-intercept ~ 0).r = 0.96, slope = 1.05, y-intercept = 6 U/mL
    Correlation with Predicate Device (103 breast cancer specimens):Strong correlation (e.g., r > 0.90, slope ~ 1, y-intercept ~ 0).r = 0.96, slope = 1.04, y-intercept = 8 U/mL
    Analytical Sensitivity<3.5 U/mL (Inferred from reported value)<3.5 U/mL
    Assay Upper Limit of Normal (ULN)Established through testing healthy volunteers.38.6 U/mL (based on 314 healthy volunteers)

    2. Sample Size and Data Provenance for Test Set

    • Sample Size for Recurrence Monitoring Study: 162 patients for the primary analysis of clinical performance as an aid in monitoring recurrence. These patients provided 942 specimens.
    • Sample Size for Metastatic Disease Management Study: 97 breast cancer patients with metastatic disease. Calculations were based on the 79 patients who had assay levels above the ULN.
    • Sample Size for Correlation Study: 203 specimens overall, with a subset of 103 women with histologically confirmed breast cancer.
    • Data Provenance: The studies were described as "multicenter-prospective studies" (for recurrence monitoring and metastatic disease management). The specific countries of origin are not explicitly stated, but the submission is to the U.S. FDA by a U.S. company, suggesting North American or international sites. The "healthy volunteer donors" for ULN determination were tested at "three laboratories."

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the test set.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method like 2+1 or 3+1. For the recurrence monitoring study, "Breast cancer recurrence was established using clinical symptoms, and/or general or specific imaging techniques; e.g., x-ray, mammogram, magnetic resonance imaging, computerized tomography, ultrasound, bone or liver scan." This indicates a clinical diagnosis process, but no specific adjudication protocol is mentioned.

    For the metastatic disease management study, disease progression or regression was used as an endpoint, but the adjudication of these changes is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was described. The device is an in vitro diagnostic test (an automated assay), not an image reading device that would typically involve multiple human readers. The comparison was the performance of the assay against clinical outcome and against a predicate device, rather than human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies presented are for the standalone performance of the ACS:180 BR assay. It's an automated chemiluminescence system measuring a biomarker, so its performance is inherently "algorithm only" in terms of measurement and result reporting. The clinical utility is then evaluated by how well these results correlate with actual disease status.

    7. Type of Ground Truth Used

    • For Recurrence Monitoring: "Disease recurrence as defined by clinical outcome." This was established using "clinical symptoms, and/or general or specific imaging techniques; e.g., x-ray, mammogram, magnetic resonance imaging, computerized tomography, ultrasound, bone or liver scan." This indicates a combination of clinical follow-up and imaging.
    • For Metastatic Disease Management: "Changing disease status" (progression or regression) determined clinically, with a 20% increase in assay levels used as an indication of progressing disease.
    • For Specificity (Non-breast malignancies/Other conditions): Diagnosis of the specific malignancy or condition.
    • For Correlation with Predicate Device: The value obtained from the predicate device (Biomira TRUQUANT® BR™ RIA).
    • For ULN determination: Serum samples from "healthy volunteer donors."

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning. The ACS:180 BR is a quantitative immunoassay, not an AI/ML algorithm that is "trained" in the typical sense. The "training" in this context would be the assay development and optimization, which isn't described with a specific dataset size.

    9. How Ground Truth for Training Set Was Established

    As noted above, a distinct "training set" with established ground truth, as understood in machine learning, is not applicable or described for this type of in vitro diagnostic assay. Instead, the assay is developed based on chemical and biological principles to detect a specific antigen. The performance studies described (clinical performance, analytical sensitivity, specificity, etc.) serve as validation of the assay's accuracy and utility.