510(k) Data Aggregation

    K Number
    K033679

    Validate with FDA (Live)

    Date Cleared
    2004-07-07

    (226 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CopiOs™ BVF, in combination with autologous bone marrow, is intended for use only for filling bone voids or gaps of the skeletal system (ie, extremities, pelvis, spine) that are not intrinsic to the stability of the bone structure. These voids may be a result of trauma or creation by surgeon. CopiOs™ BVF is intended to be gently packed into the void or gap and will resorb during the course of the healing process.

    Device Description

    CopiOs™ BVF is a synthetic bone graft material consisting of mineralized, lyophilized collagen that has been formed into three-dimensional pads of various sizes for surgical implantation. The CopiOs™ BVF product sponge consists of dehydrothermally (DHT) cross-linked bovine dermal collagen mixed with calcium phosphate salt. The pads are approximately 67% mineral by weight. The fibrous collagen mixed with the calcium phosphate salt is cast into sponges as described in this application. The average porosity of the product was determined to be 93.39%, which allows for favorable conditions for cell migration and adherence to the collagenous matrix. Further, the calcium phosphate provides the mineral required for osteogenesis and the specific CopiOs™ BVF formulation provides a pH environment conducive for osteoinduction.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the CopiOs™ Bone Void Filler (BVF). The study supporting its substantial equivalence is a comparative animal study.

    Here's an analysis of the acceptance criteria and study information, based on the provided document:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative "acceptance criteria" in the form of specific thresholds for success. Instead, it frames the performance in terms of "substantial equivalence" to a predicate device (HEALOS). The performance claims are qualitative comparisons.

    Acceptance Criteria (Implied)Reported Device Performance
    Composition: Similar collagen and calcium phosphate components.CopiOs™ BVF: 33% type 1 bovine collagen, 67% calcium phosphate salt. HEALOS (predicate): 70% collagen, 30% non-ceramic hydroxyapatite. (Note: While different percentages, the document asserts substantial equivalence.)
    Physical Structure: Similar physical form and porosity.CopiOs™ BVF: Mineralized, lyophilized collagen into 3D pads. Average porosity 93.39%. HEALOS (predicate): Supplied as lyophilized product. (Implied similar physical structure).
    Resorption Rate: Resorbed by the body at a comparable rate.CopiOs™ BVF & HEALOS: Resorption rates are "similar as evidenced in the comparative rabbit radial defect study."
    Bone Regeneration/Remodeling: Promotes new bone formation and is remodeled equivalently.CopiOs™ BVF & HEALOS: Both promote bone regeneration and are remodeled into new bone at equivalent rates "as demonstrated in the comparative rabbit radial defect study."
    Intended Use & Target Population: Same use case and patient group.CopiOs™ BVF & HEALOS: Intended for the same use (filling bone voids/gaps in skeletal system, combination with autologous bone marrow) and target populations.
    Sterility: Device must be sterile.CopiOs™ BVF & HEALOS: Both are sterile.
    Favorable Environment for Cell Migration/Adherence: Provides suitable environment for cellular activity.CopiOs™ BVF & HEALOS: Both provide a favorable environment for cell migration and adherence to the collagenous matrix.
    pH Environment for Osteoinduction: Conducive to bone formation.CopiOs™ BVF: Specific formulation provides a pH environment conducive for osteoinduction. (Not explicitly stated for HEALOS but implied through general equivalence).

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: The document mentions a "comparative rabbit radial defect study." The exact number of rabbits or defect sites used in this study (the "test set" for performance evaluation) is not specified in the provided text.
    • Data Provenance: The study was a "comparative rabbit radial defect study," indicating animal data. The country of origin is not explicitly stated, but the sponsor is based in Minneapolis, Minnesota. The study would be considered prospective in nature, as it's an experimental design to evaluate the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide information on the number or qualifications of experts involved in establishing the ground truth for the rabbit study. Histological analysis or imaging interpretation, which would likely involve experts, is implied for assessing bone regeneration and resorption, but details are absent.

    4. Adjudication method for the test set

    The document does not specify an adjudication method for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted, nor is this device an AI-based system. This question is not applicable to the provided information.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a medical implant (bone void filler), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the "comparative rabbit radial defect study," the ground truth for evaluating bone regeneration and resorption would likely be based on histopathology (microscopic examination of tissue samples) and potentially imaging analysis (e.g., radiography, micro-CT) of the radial defects. This would constitute a form of objective, scientific assessment.

    8. The sample size for the training set

    This device is not an AI/ML algorithm that requires a "training set" in the computational sense. Therefore, the concept of a training set sample size is not applicable. The "training" for this device's development would involve biological research and materials science, not data training.

    9. How the ground truth for the training set was established

    As in point 8, the concept of a "training set" and its ground truth in the context of AI/ML is not applicable to this bone void filler device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031511

    Validate with FDA (Live)

    Date Cleared
    2004-03-05

    (296 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a pedicle screw fixation system in skeletally mature patients, the DYNESYS Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, kyphosis, and failed previous fusion (pseudoarthrosis).

    In addition, when used as a pedicle screw fixation system, the DYNESYS Spinal System is indicated for use in patients:

    • Who are receiving fusions with autogenous graft only; a)
    • b) Who are having the device fixed or attached to the lumbar or sacral spine; and
    • c) Who are having the device removed after the development of a solid fusion mass.
    Device Description

    The DYNESYS® Spinal System consists of four pedicle screws, two cords and two spacers in a symmetric, bilateral arrangement. The pedicle screws are placed lateral to the facet joints with two screws in the cephalad position and two screws in the caudad position. The superior-inferior distance between the pedicle screws is maintained with the spacers. The cord passes through the eye of the pedicle screw heads and through the centers of the spacers.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for the DYNESYS® Spinal System. It is an application for regulatory clearance, claiming substantial equivalence to a predicate device, rather than a clinical study report describing acceptance criteria and device performance. Therefore, most of the requested information cannot be extracted from this document.

    Here's what can be gathered:

    1. Table of Acceptance Criteria and Reported Device Performance:
    This document does not specify quantitative acceptance criteria or report performance metrics from a study designed to meet such criteria. It states that "Establishment of equivalence is based on similarities of intended use, design, and performance characteristics" to the predicate device.

    2. Sample size used for the test set and the data provenance:
    Not applicable. This document does not describe a clinical study with a test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. This document does not describe a clinical study with a ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. This document does not describe a clinical study with adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This document is for a medical device (spinal system), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    Not applicable. This document is for a medical device (spinal system), not for an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    Not applicable. This document does not discuss a clinical study with ground truth. The basis for clearance is substantial equivalence to a predicate device.

    8. The sample size for the training set:
    Not applicable. This document describes a medical device, not a machine learning model that would have a training set.

    9. How the ground truth for the training set was established:
    Not applicable. This document describes a medical device, not a machine learning model.

    Additional relevant information from the document:

    • Device Name: DYNESYS® Spinal System
    • Intended Use: "When used as a pedicle screw fixation system in skeletally mature patients, the DYNESYS Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, kyphosis, and failed previous fusion (pseudoarthrosis)."
    • Predicate Device: Silhouette™ Spinal Fixation System (K980288)
    • FDA Limitation: The FDA added a limitation to the labeling: "The safety and effectiveness of this device for the indication of spinal stabilization without fusion have not been established." This indicates that while the device was found substantially equivalent for specified uses, potential broader uses were explicitly excluded without further proof.
    Ask a Question

    Ask a specific question about this device

    K Number
    K022615

    Validate with FDA (Live)

    Date Cleared
    2002-10-29

    (83 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spine-Tech™ Cement Restrictor is intended for use as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement.
    The Spine-Tech™ Cement Restrictor is NOT intended for any spinal indications.

    Device Description

    The Spine-Tech™ Cement Restrictor is a threaded, hollow cylinder that is manufactured from PEEK Optima LT1 with carbon fiber. This device is intended to be used in conjunction with PMMA cement.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device (Spine-Tech™ Cement Restrictor), focusing on its substantial equivalence to previously cleared predicate devices. It does not contain information about acceptance criteria, detailed study designs, or performance metrics typically found in clinical trials or validation studies for AI/algorithm-based devices.

    Therefore, most of the requested information cannot be extracted from this document.

    Here's an assessment based on the provided text, indicating what information is not available:

    1. Table of acceptance criteria and the reported device performance:

    • Not available. This document is a 510(k) summary and FDA clearance letter, which focuses on demonstrating substantial equivalence to predicate devices rather than setting or meeting specific performance acceptance criteria from a study.

    2. Sample size used for the test set and the data provenance:

    • Not available. No test set or data provenance from a study is mentioned. The clearance is based on comparison to predicate devices, not specific performance data from a clinical or analytical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not available. No ground truth establishment for a test set is mentioned.

    4. Adjudication method for the test set:

    • Not available. No test set or adjudication method is mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable/Not available. This device is a physical medical implant (cement restrictor), not an AI or imaging diagnostic tool. Therefore, an MRMC study related to AI assistance for human readers is irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable/Not available. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    • Not available. No ground truth for a study is mentioned. The "ground truth" for 510(k) clearance is often the historical performance (safety and efficacy) of the predicate device(s).

    8. The sample size for the training set:

    • Not available. No training set is mentioned as this is not an AI/algorithm device.

    9. How the ground truth for the training set was established:

    • Not available. No ground truth for a training set is mentioned.

    Summary based on the document:

    The Spine-Tech™ Cement Restrictor received 510(k) clearance based on demonstrating substantial equivalence to two predicate devices:

    1. RPX™ Titanium Cement Restrictor Device (K020344)
    2. Medtronic Sofamor Danek Cement Restrictor PEEK Optima™ LT (K010528)

    Equivalence was established based on:

    • Similarities of intended use: Both the subject device and predicates are intended for use as cement restrictors in orthopedic surgeries (e.g., femoral canal and tibial plateau in hip stem and total knee replacement).
    • Design similarities: All are cement restrictors. The subject device is a threaded, hollow cylinder manufactured from PEEK Optima LT1 with carbon fiber.
    • Physical characteristics similarities: Implied by the design and material, without specific physical characteristic data provided in this summary.

    The FDA added a warning that the device is not intended for any spinal indications, and its safety and effectiveness when implanted in the spine have not been established. This indicates an assessment of where the device could potentially be used but its safety and efficacy are not proven for that specific use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022344

    Validate with FDA (Live)

    Date Cleared
    2002-09-24

    (67 days)

    Product Code
    Regulation Number
    888.3060
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trinica Select™ Anterior Cervical Plate System is intended for anterior interbody screw fixation of the cervical spine at levels C2-T1. The System is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, and/or failed previous fusions.

    WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

    Device Description

    The Trinica Select Anterior Cervical Plate System is a fixation device consisting of cervical plates, locking caps, fixed bone screws and variable angle bone screws made from titanium alloy in conformance with ASTM F136. The locking cap is preassembled onto the plate and is designed with tabs that prevent bone screws from backing out. The plates and locking caps are treated with titanium anodization per AMS (Aerospace Material Specification) 2488 Type II. Bone screws are subjected to a color anodizing process to differentiate the screw type and diameter

    Plates are offered in one-level, two-level, and four-level fusion configurations (22 mm through 92 mm). Bone screws are available in lengths from 10mm through 18mm in 1mm increments. The screws have either a 4.2 inch 4.6 inch diameter. Fixed angle and variable angle screws are available.

    AI/ML Overview

    The provided document is a 510(k) summary for the Trinica Select Anterior Cervical Plate System. It describes a medical device and its substantial equivalence to predicate devices, rather than a study evaluating device performance against acceptance criteria for an AI/ML powered device. Therefore, the requested information regarding acceptance criteria, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies is not available in this document.

    The document states:

    • Summary of Non-Clinical Tests: "Based on risk analysis, appropriate testing was conducted to evaluate the impact of the changes to ensure that the modified device meets established criteria and that identified potential risks were mitigated. Results of the testing demonstrated that the modified device meets established criteria."
    • Conclusions Drawn From Testing: "Testing of the Trinica Select Anterior Cervical Plate System demonstrates that the device is substantially equivalent to the devices and that the design modifications do not affect device safety and effectiveness."

    However, it does not specify what these established criteria are, what the reported device performance was against these criteria, or provide any details about the study design that would allow me to populate the requested table and answer the specific questions about sample sizes, ground truth, experts, or comparative effectiveness.

    This document focuses on demonstrating substantial equivalence for a physical medical device (an anterior cervical plate system) based on material, design, and non-clinical performance testing (likely mechanical testing under loads), rather than clinical performance or the performance of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device