Search Results
Found 1 results
510(k) Data Aggregation
(328 days)
The Synovasure Alpha Defensin Lateral Flow Test Kit is a qualitative visually read immunochromatographic assay for the detection of human host response proteins, Alpha Defensins 1-3, in the synovial fluid of adults with a total joint replacement who are being evaluated for revision surgery. The Synovasure Alpha Defensin Lateral Flow Test Kit results are intended to be used in conjunction with other clinical and diagnostic findings as an aid in the diagnosis of periprosthetic joint infection (PJI). The Synovasure Alpha Defensin Lateral Flow Test Kit is not intended to identify the etiology or severity of a PJI.
The Synovasure Alpha Defensin Control Kit is used in the Synovasure Alpha Defensin Lateral Flow Test Kit as assayed quality control samples to monitor performance and reliability of the Synovasure Alpha Defensin Lateral Flow Test Kit.
The Synovasure Alpha Defensin Lateral Flow Test Kit (Synovasure LFT) is an immunoassay for the detection of alpha defensin levels in the synovial fluid of patients with a potential PJI. Antibodies specific to alpha defensin bind host alpha defensin in the synovial fluid, become immobilized on the lateral flow test strip, and are detected as a colored line due to the use of a colloidal gold reporter.
Synovasure Alpha Defensin Lateral Flow Test Kit contains two sub components:
-
- Synovasure Alpha Defensin Lateral Flow Test Device
-
- Synovasure Lateral Flow Sample Prep Assembly
The Synovasure Lateral Flow Sample Prep Assembly further contains
-
- One Synovasure Dilution Buffer Bottle
-
- One Sample Cup
-
- Two Microsafe Tubes
The Synovasure Alpha Defensin Lateral Flow Test Device is a cassette that includes a reagent strip. Each cassette contains a reagent strip with all the critical components for the assay.
The Synovasure Alpha Defensin Lateral Flow Test Kit is accompanied by the Synovasure Alpha Defensin Control Kit. The Synovasure Alpha Defensin Control Kit further contains
-
- Synovasure Alpha Defensin Positive Control
-
- Synovasure Alpha Defensin Negative Control
-
- Synovasure Control Reconstitution Bottle
The positive control contains 0.25 mL of 16 µg/mL alpha defensin in synthetic synovial fluid and the negative control contains 0.25 mL of synthetic synovial fluid.
Additional materials required but not provided include
- Timer
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The core clinical acceptance criteria for the Synovasure Alpha Defensin Lateral Flow Test Kit's performance in diagnosing periprosthetic joint infection (PJI) are related to its sensitivity and specificity.
| Acceptance Criteria (Clinical Performance) | Reported Device Performance (Prospective Data) |
|---|---|
| Sensitivity: | Sensitivity: 89.5% (95% CI: 78.5-96.0%) |
| >90% with lower confidence bound >85% | |
| Specificity: | Specificity: 94.8% (95% CI: 91.2-97.2%) |
| >90% with lower confidence bound >90% |
Notes on Acceptance Criteria Met: The device met the specificity criterion in the prospective study (94.8% with 91.2% LCB, exceeding the >90% and >90% LCB requirement). The device did not fully meet the sensitivity criterion in the prospective study (89.5% with 78.5% LCB), as the lower confidence bound of 78.5% is below the required >85%. However, the sponsor appears to have supplemented this with retrospective data to support overall performance for the De Novo classification.
2. Sample Sizes and Data Provenance
-
Test Set Sample Size:
- Prospective Study: 305 synovial fluid samples.
- Retrospective Study: 65 fresh remnant synovial fluid samples.
- Total Clinical Test Set: 370 samples (305 prospective + 65 retrospective).
-
Data Provenance:
- Country of Origin: United States (Prospective study conducted at 3 US medical centers).
- Retrospective/Prospective: Both.
- The main clinical study was a prospective study.
- A retrospective collection of fresh remnant samples was used to supplement the prospective data and reach a target of at least 100 positive specimens.
3. Number of Experts and their Qualifications for Ground Truth
- Number of Experts: A two-physician panel was used to adjudicate the final MSIS status for prospective data. A third physician was consulted to resolve discrepant opinions.
- Qualifications of Experts: The document states "two-physician panel" and "third physician," implying medical doctors, but does not specify their specialties (e.g., orthopedic surgeons, infectious disease specialists) or years of experience.
4. Adjudication Method for the Test Set
- Adjudication Method: 2+1 adjudication.
- The "final status determination was adjudicated by a two-physician panel, with discrepant opinions being resolved by consultation of a third physician."
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
-
Was an MRMC study done?: No, an MRMC comparative effectiveness study was not mentioned or conducted as this is a qualitative visually-read lateral flow immunoassay device, not an AI or imaging-based diagnostic where human reader performance would be a primary variable. The output is a visual line or no line, reducing inter-reader variability compared to complex interpretations.
-
Effect Size of Human Readers with AI vs. without AI assistance: Not applicable, as no AI assistance is involved in the interpretation of this device.
6. Standalone Performance Study (Algorithm Only)
- Was a standalone performance study done?: Not applicable in the context of an algorithm. This device is a manual, visually-read lateral flow immunoassay. Its "standalone performance" is essentially its analytical and clinical performance as a device read directly by an operator, which is what the provided sensitivity and specificity figures represent. There is no separate "algorithm" for this product that would operate without human interaction (as it is a visually-read test).
7. Type of Ground Truth Used
- Type of Ground Truth: Expert Consensus using established clinical criteria.
- The ground truth for Periprosthetic Joint Infection (PJI) was established using the Musculoskeletal Infection Society (MSIS) criteria.
- For the prospective study, the final MSIS status was adjudicated by a two-physician panel with a third physician for conflicts.
- For the retrospective samples, the ground truth was based on a "status determination based positive confirmation of three minor MSIS criteria (neutrophil %, positive culture, and WBC count)."
8. Sample Size for the Training Set
- Training Set Sample Size: The document does not explicitly mention a separate "training set" for an algorithm or model. This is consistent with the nature of a lateral flow immunoassay, which does not typically involve machine learning or AI models that require distinct training sets. The studies described are performance validation studies.
9. How the Ground Truth for the Training Set was Established
- Ground Truth Establishment for Training Set: Not applicable, as no distinct training set for an algorithm or machine learning model is described for this device.
Ask a specific question about this device
Page 1 of 1