Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K012520

    Validate with FDA (Live)

    Date Cleared
    2002-03-27

    (233 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gyro Tip Lariat Deflectable Loop Mapping Catheter is indicated for multi-electrode electrophysiological mapping, recording, and or temporary stimulation of cardiac structure.

    Device Description

    The Gyro-Tip Lariat Deflectable Loop Mapping Catheter of Cardiac Assist Devices, Inc. has been designed and developed as a diagnostic catheter for electrophysiological mapping of cardiac structures. The catheter has a braided outer tube that exhibits similar torsional stiffness, under a clockwise and counterclockwise-applied torque directions, to those of the named predicate devices. The distal portion of the subject catheter has up to 20 platinum electrodes that offer radiopacity under fluoroscopy for the required visibility during the cardiac electrophysiology mapping procedures. The accessory cables used to connect the subject device to a recorder/monitor comply with Section 12A of the Underwriters Laboratories UL 544 Standard for safety.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and the study that proves the device meets them:

    Disclaimer: This response is based solely on the provided text, which is a 510(k) summary, not a full submission. 510(k) summaries are intended to be concise and often reference more detailed information contained within the full submission (e.g., Appendices A, B, C, D, and E in this case). Therefore, some details you requested may not be explicitly present in this summary.


    This 510(k) summary (K012520) for the Gyro-Tip Lariat Deflectable Loop Mapping Catheter describes a device intending to demonstrate substantial equivalence to predicate devices, rather than meeting specific performance criteria against a pre-defined standard in a de novo study. The acceptance criteria and "proof" primarily stem from demonstrating similarity in performance to legally marketed predicate devices through non-clinical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance (Summary of Non-Clinical Tests)
    Safety and Effectiveness comparable to predicate devices."The comparison of the results indicates that the subject device is as safe, and as effective, as the named predicate devices.""The results of the non-clinical tests... indicates that the subject device... performs as well as the named predicate devices, and any minute differences in the test results are not significant to influence the safety and effectiveness of the subject device."
    Similar torsional stiffness (under clockwise and counterclockwise torque)."The catheter has a braided outer tube that exhibits similar torsional stiffness, under a clockwise and counterclockwise-applied torque directions, to those of the named predicate devices."
    Radiopacity under fluoroscopy for required visibility during cardiac electrophysiology mapping."The distal portion of the subject catheter has up to 20 platinum electrodes that offer radiopacity under fluoroscopy for the required visibility during the cardiac electrophysiology mapping procedures." (This is a feature description, implying it meets the need for visibility).
    Steering mechanism and actuation resulting in a deflectable loop/halo-shaped configuration, safe and effective."The steering mechanism of the subject device is exactly similar to the steering mechanism of the primary predicate device... The halo-shaped configuration... and the process of assuming this final halo-shaped configuration... has shown to be safe and effective through a series of non-clinical testing."
    Compliance with UL 544 Standard for Safety for accessory cables."The accessory cables used to connect the subject device to a recorder/monitor comply with Section 12A of the Underwriters Laboratories UL 544 Standard for safety."
    Non-clinical performance, as determined by FDA's "Electrode Recording Catheter preliminary Guidance," should compare favorably."These tests were performed according to FDA's 'Electrode Recoding Catheter preliminary Guidance'." The summary states the device performs "as well as" predicate devices with no significant differences.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text states that a "series of non-clinical testing" was performed, and the results are included in "Appendices A, B, and C" of the 510(k) submission. However, the specific sample size for each non-clinical test (e.g., number of catheters tested for torsional stiffness, radiopacity, or deflection) is not mentioned in this 510(k) summary.

    The data provenance is non-clinical testing, meaning it was performed in a laboratory or simulated environment, not on human subjects. This implies the data is retrospective in the sense that the tests were completed before submission. The country of origin of the data is not explicitly stated but would typically be where the manufacturer (Cardiac Assist Devices, Inc., Cleveland, Ohio) conducted their R&D and testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable / not provided in this context. The study is a non-clinical, benchtop, or simulated environment comparison to predicate devices, not a study involving human data requiring expert ground truth assessment (like image interpretation by radiologists). The "ground truth" here is the measured performance of both the subject device and the predicate devices against engineering and performance specifications.

    4. Adjudication Method for the Test Set

    This information is not applicable / not provided. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving human interpretation or subjective assessments, where multiple experts provide opinions that need to be resolved. For non-clinical tests, results are typically objective measurements compared against specifications or predicate device performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study involves human readers (e.g., clinicians) evaluating cases with and without AI assistance to measure performance improvement. The current submission is for a physical medical device (catheter) and focuses on its physical and functional performance, not on an AI algorithm's interpretative capabilities.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This submission is for a physical medical device, not a software algorithm or AI.

    7. The Type of Ground Truth Used

    The "ground truth" used for this non-clinical testing comprises objective measurements and specifications related to the device's physical properties and functionality. This would include:

    • Measurement of torsional stiffness.
    • Evaluation of radiopacity.
    • Assessment of deflection mechanism performance (loop formation, steerability, mechanical integrity).
    • Compliance with electrical safety standards (UL 544 for cables).

    The "ground truth" is established by the methods and standards used to measure these properties for both the subject device and the predicate devices.

    8. The Sample Size for the Training Set

    This information is not applicable / not provided. The device in question is a physical medical catheter, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable / not provided. As explained in point 8, there is no training set for a physical device. Ground truth for non-clinical testing is established through standard measurement techniques and engineering specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K960849

    Validate with FDA (Live)

    Device Name
    MAGNO ALARM
    Date Cleared
    1997-03-26

    (400 days)

    Product Code
    Regulation Number
    870.3690
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    useful, safe and effective for forewarning presence of environmental magnetic fields that might be potentially harmful for certain individuals with implanable devices such as pacemakers or ICD's.

    Device Description

    a device that in presence of magnetic fields produce sound and light signals to be noticed by the hearing and sight senses of an individual who is using it.

    AI/ML Overview

    The provided text is a K960849 510(k) Summary for the "Magno Alarm" device. Based on this document, here's an analysis of the acceptance criteria and study described:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state formal "acceptance criteria" in a quantitative, metrics-based format commonly seen in modern medical device submissions (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" are implied by the device's intended purpose and the qualitative claims made.

    Acceptance Criterion (Implied)Reported Device Performance
    1. Forewarning of Environmental Magnetic Fields: Device should detect and warn of potentially harmful environmental magnetic fields for individuals with implantable devices.The Magno Alarm has a "lower activation threshold from that of typical pacemakers and/or ICD's," providing a "safety factor of at least two (2) for the patient to get a warning." It produces "sound and light signals" in the presence of magnetic fields.
    2. Non-interference with Pacemakers/ICD's: Device should not disrupt the proper function of implantable cardiac devices."The Magno Alarm has been designed so that it does not interfere with the function of pacemakers/ICD's."
    3. Safety and Effectiveness: Device should be safe and effective for its stated purpose.The device is described as "useful, safe and effective for forewarning presence of environmental magnetic fields" and "a safe, effective and useful device."
    4. Substantial Equivalence to Predicate Device: Device performance should be comparable to a predicate device (test magnets).The Magno Alarm is "presented to be substantially equivalent to the test magnets employed as an external and noncontacting auxiliary tool for pacemakers/ICD's." The document then outlines three "bases of substantial equivalence" focusing on their respective interactions with magnetic fields and information they provide.
    5. Electromagnetic Compatibility (EMC): Device should operate without significant electromagnetic interference to or from other devices."The Magno Alarm has been tested for electromagnetic compatibility and the test results is included in the supplement to our 510(k) (K960849) of 08-12-96."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not provide any details regarding a specific "test set" for performance evaluation, nor does it mention sample sizes. It does not state the country of origin of data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not mention the use of experts to establish ground truth for a test set. The claims appear to be based on the device's design characteristics and a comparison to predicate devices, rather than a clinical study with expert-adjudicated outcomes.

    4. Adjudication Method for the Test Set:

    Since there is no mention of a test set or expert involvement, there is no adjudication method described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A MRMC comparative effectiveness study was not done or described. The device is a standalone alarm, not an interpretation aid for human readers. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" is not applicable.

    6. Standalone (Algorithm Only) Performance:

    Yes, a form of standalone performance is described. The Magno Alarm itself is the "algorithm only" in this context. Its performance is characterized by its "lower activation threshold" compared to pacemakers/ICDs, providing a "safety factor of at least two (2)". It produces "sound and light signals" in the presence of magnetic fields. The EMC testing also falls under standalone performance.

    7. Type of Ground Truth Used:

    The "ground truth" implicitly used for the device's claim of effectiveness appears to be:

    • Physical Principles and Known Thresholds: The understanding of how pacemakers/ICDs respond to magnetic fields and the design intention of the Magno Alarm to have a lower activation threshold.
    • Predicate Device Functionality: The working principle of test magnets and how the Magno Alarm's function relates to or complements this.

    There is no mention of expert consensus, pathology, or outcomes data being used for ground truth.

    8. Sample Size for the Training Set:

    The document does not provide any information about a "training set" or its size. This type of submission predates the common use of AI/ML models that require distinct training and testing sets. The device's design appears to be based on engineering principles and existing knowledge of medical devices, rather than machine learning.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no mention of a training set, there is no description of how its ground truth was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K955847

    Validate with FDA (Live)

    Date Cleared
    1997-01-29

    (400 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Monitor and record intra-cardiac electrical activities

    Device Description

    Not Found

    AI/ML Overview

    This document is a premarket notification (510(k)) for a medical device called the "Gyro Tip EP Catheter." It is for a device used to "monitor and record intra-cardiac electrical activities."

    Unfortunately, the provided text does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria. The document is primarily a summary for a 510(k) submission, focusing on:

    • Device Identification: Name, classification, intended use.
    • Regulatory Information: Manufacturer registration, classification.
    • Labeling: Mention of draft labeling, including directions for use, indications, cautions, warnings, and precautions.
    • Substantial Equivalence: A claim that the device is substantially equivalent to existing devices (Bard Dynamic Tip Catheter, Bard Tip Deflecting Catheter, Bard Tip Deflecting Electrode Catheter, and Steerocath).
    • General Information: Reference to a more comprehensive summary for material biocompatibility, physical characteristics, and results of electrical and mechanical tests.

    There is no mention of specific acceptance criteria, performance metrics, a study design, sample sizes, ground truth establishment, or expert involvement as requested. The "None Established Performance Standards" point further reinforces that no specific numerical performance standards are detailed in this summary.

    Therefore, I cannot populate the table or answer the specific questions about the study from the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1