    K Number
    K210204

    Validate with FDA (Live)

    Date Cleared
    2021-10-19

    (266 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Capsule Vitals Plus Patient Monitoring System is intended for monitoring and alarming of physiologic parameters, including non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, functional arterial oxygen saturation (SpO2), end-tidal and fractional concentration of inspired CO2, respiration, and temperature, on adult, pediatric, and neonatal patients in health care facilities when used by clinical physicians or appropriate medical staff under the direction of physicians.

    Device Description

    The Capsule Vitals Plus Patient Monitoring System which is the subject of this submission is intended for monitoring and alarming of physiologic parameters, including non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, functional arterial oxygen saturation (SpO2), end-tidal and fractional concentration of inspired CO2, respiration, and temperature, on adult, pediatric, and neonatal patients in health care facilities when used by.

    It operates with the Capsule Medical Device Information Platform (CMDIP) a Medical Device Data System to present patient information to the clinical user for active monitoring purposes at the point of care.

    The proposed Capsule Vitals Plus Patient Monitoring System consists of the following components:

    • -Capsule Neuron Mobile Platforms
    • Capsule Vitals Plus Application |
    • Capsule Vitals Plus NIBP Monitoring Modules -
    • Masimo NomoLine Infrared Sidestream Gas Analyzer ISA C02 —
    • Capsule Capnography Interface Module -
    • Nellcor SpO2 Oximetry Module -
    • Masimo uSpO2 Pulse Oximetry Cable
    • -Exergen TAT-5000S Temperature Scanner
    • -Capsule Early Warning Scoring System
    AI/ML Overview

    The provided text is a 510(k) summary for the Capsule Vitals Plus Patient Monitoring System. It describes the device, its intended use, comparison to a predicate device, and performance testing. However, it explicitly states that clinical studies were not required to support substantial equivalence. Therefore, the document does not contain the information requested regarding acceptance criteria related to a study proving the device meets those criteria, sample sizes for test sets, expert ground truth establishment, MRMC studies, standalone performance, or training set details.

    The summary focuses on demonstrating substantial equivalence by showing that the proposed device has the same intended use and similar technical characteristics as the predicate device (SmartLinx Vitals Plus Patient Monitoring System, K200856), with minor differences that do not raise new questions of safety or effectiveness. The performance testing mentioned refers to hardware and software verification and validation, as well as conformance with FDA recognized consensus standards (listed in Table 4).

    Based on the provided text, it is not possible to answer the detailed questions about acceptance criteria and a study proving the device meets them because such a study involving clinical performance with specific metrics (like sensitivity, specificity, or human reader improvement) was not part of this 510(k) submission.

    Here's a breakdown of why each requested point cannot be answered from the provided text:

    1. A table of acceptance criteria and the reported device performance: Not present. The document focuses on showing equivalence to a predicate device and conformance to standards, not on demonstrating performance against specific clinical metrics.
    2. Sample sized used for the test set and the data provenance: Not present. The performance testing described is primarily engineering verification and validation and compliance with standards, not a clinical test set with patient data.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/Not present. No clinical ground truth establishment is described.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable/Not present.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not done. This device is a patient monitoring system, not an AI diagnostic tool requiring MRMC studies.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a monitoring system intended for use by clinical physicians or appropriate medical staff.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable/Not present.
    8. The sample size for the training set: Not applicable/Not present. The document describes a medical device, not a machine learning model that would typically have a training set.
    9. How the ground truth for the training set was established: Not applicable/Not present.

    Conclusion stated in the document: "The subject of this premarket submission, Capsule Vitals Plus Patient Monitoring System, did not require clinical studies to support substantial equivalence." (Page 15)

    Ask a Question

    Ask a specific question about this device

    K Number
    K200856

    Validate with FDA (Live)

    Date Cleared
    2020-07-10

    (100 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartLinx Vitals Plus Patient Monitoring System is intended for monitoring and alarming of physiologic parameters, including non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, functional arterial oxygen saturation (SpO2), end-tidal and fractional concentration of inspired CO2, respiration, and temperature, on adult, pediatric, and neonatal patients in health care facilities when used by clinical physicians or appropriate medical staff under the direction of physicians.

    Device Description

    The SmartLinx Vitals Plus Patient Monitoring System is intended for monitoring and alarming of physiologic parameters, including non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, functional arterial oxygen saturation (SpO2), end-tidal and fractional concentration of inspired CO2, respiration, and temperature, on adult, pediatric, and neonatal patients in health care facilities when used by clinical physicians or appropriate medical staff under the direction of physicians.

    The SmartLinx Vitals Plus Patient Monitoring System operates with the SmartLinx Medical Device Information Platform (MDIP) a Medical Device Data System to present patient information to the clinical user for active monitoring purposes at the point of care.

    The proposed SmartLinx Vitals Plus Patient Monitoring System consists of the following components:

    • SmartLinx Neuron Mobile Platform
    • SmartLinx Vitals Plus Application
    • SmartLinx Vitals Plus NIBP Monitoring Module
    • Masimo NomoLine Infrared Sidestream Gas Analyzer ISA C02
    • Nellcor SpO2 Oximetry Module
    • Masimo uSpO2 Pulse Oximetry Cable
    • Exergen TAT-5000S Temperature Scanner
    • SmartLinx Early Warning Scoring System
    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study that proves a device meets those criteria. Instead, it describes a "SmartLinx Vitals Plus Patient Monitoring System" which is a medical device data system that integrates various vital signs modules.

    The document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to existing legally marketed predicate devices, rather than proving performance against specific acceptance criteria through a clinical study for a new, novel device.

    The "Performance Testing" section mentions adherence to FDA recognized consensus standards (Table 4), which are often general safety and performance standards for similar types of medical devices (e.g., electrical safety, electromagnetic compatibility, usability, software lifecycle, and particular requirements for specific monitoring components like non-invasive blood pressure, pulse oximetry, and respiratory gas monitors). This is a different type of "acceptance criteria" than what might be established for a novel AI/ML-based diagnostic device.

    Crucially, the document explicitly states:

    • "The subject of this premarket submission, SmartLinx Vitals Plus Patient Monitoring System, did not require clinical studies to support substantial equivalence."

    Therefore, I cannot extract the specific information requested about acceptance criteria and a study proving the device meets them from this document, as there was no such clinical study performed or required for this 510(k) submission.

    To directly answer your numbered points based on the provided text:

    1. A table of acceptance criteria and the reported device performance: Not available. The document refers to conformance with recognized consensus standards (Table 4) rather than specific performance acceptance criteria for a novel functionality.
    2. Sample sized used for the test set and the data provenance: Not applicable, as no clinical study was required or performed. Performance testing involved hardware and software verification, validation, and testing against consensus standards, not a "test set" of patient data in the typical sense of an AI/ML study.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical study was required or performed that involved expert-established ground truth on patient data.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI/ML-based diagnostic or assistive technology that would typically undergo an MRMC study. It is a patient monitoring system integrating existing vital sign modules.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm-only device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The device relies on the established accuracy and ground truth of its component modules (e.g., NIBP, SpO2, CO2 sensors) which are themselves regulated and have their own performance specifications.
    8. The sample size for the training set: Not applicable. This device does not use a "training set" in the context of AI/ML.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151071

    Validate with FDA (Live)

    Date Cleared
    2015-09-17

    (149 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartLinx Vitals Plus Patient Monitoring System is intended for monitoring physiologic parameters, including non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, functional arterial oxygen saturation (SpO2), and temperature, on adult, pediatric, and neonatal patients in healthcare facilities when used by clinical physicians or appropriate medical staff under the direction of physicians.

    Device Description

    The SmartLinx Vitals Plus Patient Monitoring System consists of the SmartLinx Vitals Plus mobile medical application and externally integrated vital signs modules operating on the SmartLinx Neuron 2 mobile platform. The application controls the modules at the point of care through interfaces on the SmartLinx Neuron 2 mobile platform and presents this to the monitoring purposes. The supported physiological parameters are: non-invasive blood pressure (systolic, diastolic, mean arterial pressure (MAP)), pulse rate, functional arterial oxygen saturation (SpO2), and temperature.

    The SmartLinx Vitals Plus Patient Monitoring System is composed of multiple products:

    • The SmartLinx Vitals Plus mobile medical application
    • The SmartLinx Vitals Plus NIBP Module with SunTech Medical cuffs and hoses
    • The Masimo SET uSpO2 Pulse Oximetry Cable
    • The Exergen TAT-5000S infrared thermometer
    • The SmartLinx Neuron 2 Mobile Platform
    • The SmartLinx Vitals Plus Roll Stand
    AI/ML Overview

    The provided document details the 510(k) premarket notification for the SmartLinx Vitals Plus Patient Monitoring System. It describes the device, its intended use, and a comparison to a predicate device, focusing on functional equivalence. However, the document does not contain detailed acceptance criteria for specific performance metrics (like accuracy or precision for NIBP, SpO2, or temperature) nor does it provide a study report proving the device meets said criteria with specific quantitative results.

    Instead, the document generally states that the device "complies with its predetermined specification" based on various types of testing, often referencing compliance with established electrical safety, EMC, and software standards, as well as a clinical performance standard for NIBP (ISO 81060-2:2013).

    Therefore, I cannot populate the table or answer all the requested questions with specific data from the provided text. I will indicate where information is not available.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Parameter/Test CategoryAcceptance Criteria (from document)Reported Device Performance (from document)
    Software TestingDesigned and developed in accordance with Capsule Technologie software development processes; verified and validated."Test results indicated that the SmartLinx Vitals Plus Patient Monitoring System complies with its predetermined specification."
    Electrical SafetyCompliance with IEC 60601-1:2005, IEC 60601-1-6:2013, IEC 60601-1-8:2012, IEC 80601-2-30:2013, ISO 80601-2-56:2009, ISO 80601-2-61:2011, ISO 81060-2:2013, IEC 62304:2006, IEC 62366:2014, IEC 62133:2012."Test results indicated that the SmartLinx Vitals Plus Patient Monitoring System complies with its predetermined specification."
    EMC TestingCompliance with IEC 60601-1-2:2007."Test results indicated that the SmartLinx Vitals Plus Patient Monitoring System complies with its predetermined specification."
    Bench PerformanceIn accordance with internal Capsule Technologie requirements and procedures; covers performance, functional, reliability, environmental, and packaging testing."Test results indicated that the SmartLinx Vitals Plus Patient Monitoring System complies with its predetermined specification."
    Clinical PerformanceIn accordance with ISO 81060-2:2013 (likely for NIBP). (Specific criteria within this standard, e.g., mean difference and standard deviation, are not stated in the document). The predicate device's specific performance metrics are also not provided in a similar manner, making a direct numerical comparison difficult from this document."Test results indicated that the SmartLinx Vitals Plus Patient Monitoring System complies with its predetermined specification." (No specific numerical results for accuracy, precision for NIBP, SpO2, temp or pulse rate are provided in the document.)
    SterilizationNot designed to be sterilized. Patient-applied parts' sterilization addressed in predicate/reference device submissions."The sterilization issues related to these patient applied parts were addressed in the predicate or reference device submissions, and were cleared by FDA."
    Shelf LifePatient-applied parts' shelf life addressed in predicate/reference device submissions. Neuron battery shelf life: three months (not in use). Exergen TAT-5000S 9V battery shelf life: typically seven years."The shelf life issues related to these patient applied parts were addressed in the predicate or reference device submissions, and were cleared by FDA." (For batteries, stated shelf lives are provided).
    BiocompatibilityPatient-applied parts' biocompatibility addressed in predicate/reference device submissions."The biocompatibility issues related to these patient applied parts were addressed in the predicate or reference device submissions, and were cleared by FDA."

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not specified in the document. For clinical performance testing (referencing ISO 81060-2:2013), this standard typically requires a certain number of subjects (e.g., 85 subjects for NIBP accuracy) and measurements, but the document does not report the actual sample size used for this specific device's clinical performance trial.
      • Data Provenance: Not specified. The document does not indicate the country of origin, nor whether the data was retrospective or prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not specified. The document indicates clinical performance testing was done "in accordance with ISO 81060-2:2013". For NIBP validation studies, ground truth (reference measurements) are typically established by trained technicians using a reference device (e.g., auscultation with mercury sphygmomanometer), but the number and qualifications of these individuals are not detailed in this submission.
    3. Adjudication method for the test set:

      • Not specified. If ground truth involved multiple measurements or experts, an adjudication method would typically be used, but this information is absent.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This device is a vital signs monitor, not an AI-assisted diagnostic imaging device that involves "human readers" or "AI assistance" in that context. Therefore, an MRMC study and AI improvement effect size is not applicable to this type of device and was not conducted.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, to an extent. The performance testing described (Software Testing, Electrical Safety, EMC Testing, Bench Performance Testing, and Clinical Performance Testing conforming to standards like ISO 81060-2:2013) essentially evaluates the device's inherent performance. While the device is intended for use by healthcare providers, these tests assess the accuracy and reliability of the device's measurements and functions independent of a human's interpretation of those measurements in a diagnostic workflow (which is the typical context for "standalone AI performance"). The software and hardware perform their functions (measuring NIBP, SpO2, temp) without direct "human-in-the-loop" assistance for the measurement process itself.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For NIBP, the ground truth would typically be established by simultaneous auscultatory measurements by trained observers using a reference device, as per the methodology specified in ISO 81060-2.
      • For SpO2, ground truth is usually established by CO-oximetry of arterial blood samples.
      • For temperature, ground truth would be from a precisely calibrated reference thermometer or other established method.
      • The document does not explicitly state the ground truth methodology beyond referencing the ISO standards, but these standards define the accepted methods for ground truth establishment.
    7. The sample size for the training set:

      • Not applicable. This document describes a traditional medical device (a vital signs monitor) and its validation. It does not refer to an AI/Machine Learning model that requires a "training set" in the conventional sense. The software development process mentioned is likely for deterministic software, not a learned model.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" for an AI/ML model described in this document.