Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K252294

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-08

    (138 days)

    Product Code
    Regulation Number
    892.2060
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K251456

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-05

    (24 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrightHeart View Classifier device is intended to analyze fetal 2D ultrasound images and video clips using machine learning techniques to automatically detect standard views during fetal heart scanning.

    The BrightHeart View Classifier device is intended to be used as an adjunct to the acquisition and interpretation of fetal anatomic ultrasound examinations at the second or third trimester of pregnancy performed with transabdominal probes.

    Device Description

    BrightHeart View Classifier is a cloud-based software-only device which uses artificial intelligence (AI) to detect standard views during fetal heart scanning in fetal ultrasound images and video clips.

    BrightHeart View Classifier is intended to be used by qualified, trained healthcare professional personnel in a professional prenatal ultrasound (US) imaging environment (this includes sonographers, MFMs, OB/GYN, and Fetal surgeons), to help fetal ultrasound examination acquisition and interpretation of 2D grayscale ultrasound by providing automatic classification of video clips and images into standard views, by automatically extracting example frames of standard views from video clips, and by automatically assessing whether the documentation of each standard view in video clips and images satisfies an acquisition protocol defined by the center. Annotated DICOM files generated by the device cannot be modified by the user.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study proving the BrightHeart View Classifier device meets them, based on the provided FDA 510(k) clearance letter:


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state "acceptance criteria" as a set of predefined thresholds. However, it does present objective performance metrics derived from a validation study. For the purpose of this response, we will consider the reported performance metrics as demonstrative of meeting implicit acceptance criteria for clinical utility and safety, especially since the submission states the device "is as safe and effective as the predicate device and supports a determination of substantial equivalence."

    MetricAcceptance Criteria (Implicit)Reported Device Performance
    Mean Standard View Recognition SensitivityHigh sensitivity for detecting standard views, indicating a low rate of missed standard views. (Implicitly, the reported value was deemed sufficient for clearance given its comparison to the predicate, which shares the exact same algorithm).0.939 (95% CI: 0.917 – 0.960)
    Mean Standard View Recognition SpecificityHigh specificity for identifying standard views, indicating a low rate of incorrectly identified views. (Implicitly, the reported value was deemed sufficient for clearance).0.984 (95% CI: 0.973 – 0.996)
    Performance across subgroups (geographical region, US machine make, gestational age, mother's BMI, mother's age)Consistent performance across diverse subgroups."Performance was consistent across subgroups."
    Performance across mother's race (Asian and Black mothers)Consistent performance across mother's race."Specificity 95% CI lower bound was slightly lower for Asian and Black mothers, possibly due to large confidence intervals and small sample size." (This indicates a slight deviation but was seemingly acceptable given the context of small sample size in those subgroups).
    Supported Ultrasound Machine VendorsDevice performance should be validated for specific ultrasound machine makes.Validated with General Electric, Philips, Samsung, and Siemens ultrasound devices.
    Supported Gestational AgeDevice performance should be validated for specific gestational ages.Validated for pregnancies at 18 weeks of gestation or later.
    Supported Probe TypeDevice performance should be validated for specific probe types.Validated for transabdominal probes.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 2290 clinically acquired images and frames from video clips. These were derived from 579 fetal ultrasound examinations.
    • Data Provenance: The data was retrospective, consisting of clinically acquired images and frames. The country of origin for the data includes U.S.A. and France.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Two experts.
    • Qualifications of Experts: One sonographer and one MFM specialist (Maternal-Fetal Medicine) with experience in fetal echocardiography.

    4. Adjudication Method for the Test Set

    The adjudication method was described as a truthing process where the sonographer and MFM specialist independently determined the presence or absence of standard views. The document doesn't explicitly state a 2+1 or 3+1 method with a third tie-breaker, but it implies a consensus or agreement process between the two experts as they "determined" the presence or absence.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or reported in this document. The study described focuses on the standalone performance of the AI algorithm. Therefore, there is no reported effect size of how much human readers improve with AI vs. without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The reported sensitivity and specificity values directly relate to the BrightHeart View Classifier's (algorithm only) ability to identify standard views independently. The document states: "The performance testing demonstrated that BrightHeart View Classifier identifies standard views with a mean standard view recognition sensitivity of 0.939..." This confirms a standalone performance evaluation.

    7. The Type of Ground Truth Used

    The type of ground truth used was expert consensus / clinical expert interpretation. It was derived through a "truthing process" by a sonographer and an MFM specialist.

    8. The Sample Size for the Training Set

    The document explicitly states: "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in performance testing." However, the exact sample size for the training set is not provided in the given text.

    9. How the Ground Truth for the Training Set Was Established

    The document states: "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in performance testing." While it confirms distinct data, it does not explicitly describe how the ground truth for the training set was established. It can be inferred that it likely followed a similar expert review process as the test set, but this information is not detailed in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243684

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-07

    (159 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrightHeart View Classifier device is intended to analyze fetal 2D ultrasound images and video clips using machine learning techniques to automatically detect standard views during fetal heart scanning.

    The BrightHeart View Classifier device is intended to be used as an adjunct to the acquisition and interpretation of fetal anatomic ultrasound examinations at the second or third trimester of pregnancy performed with transabdominal probes.

    Device Description

    BrightHeart View Classifier is a cloud-based software-only device which uses artificial intelligence (AI) to detect standard views during fetal heart scanning in fetal ultrasound images and video clips.

    BrightHeart View Classifier is intended to be used by qualified, trained healthcare professional personnel in a professional prenatal ultrasound (US) imaging environment (this includes sonographers, MFMs, OB/GYN, and Fetal surgeons), to help fetal ultrasound examination acquisition and interpretation of 2D grayscale ultrasound by providing automatic classification of video clips and images into standard views, by automatically extracting example frames of standard views from video clips, and by automatically assessing whether the documentation of each standard view in video clips and images satisfies an acquisition protocol defined by the center. Annotated DICOM files generated by the device cannot be modified by the user.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the BrightHeart View Classifier, based on the provided FDA 510(k) Clearance Letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA letter does not explicitly state pre-defined acceptance criteria values that the device needed to meet. Instead, it reports the device's performance metrics directly from the validation study. However, based on the performance report, we can infer the achieved performance and understand that these values were deemed sufficient for clearance.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Mean Standard View Recognition SensitivityHigh (e.g., >0.90)0.939 (95% CI, 0.917 ; 0.960)
    Mean Standard View Recognition SpecificityHigh (e.g., >0.95)0.984 (95% CI, 0.973 ; 0.996)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 2290 clinically acquired images and frames from video clips.
    • Number of Fetal Ultrasound Examinations: 579
    • Country of Origin of Data: U.S.A. and France.
    • Retrospective or Prospective: The document implies retrospective data ("clinically acquired images and frames").

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Two experts: "a sonographer and an MFM specialist".
    • Qualifications: "with experience in fetal echocardiography". Specific years of experience are not mentioned.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Independence was maintained in the ground truth establishment. "The reference standard was derived from the dataset through a truthing process in which a sonographer and an MFM specialist with experience in fetal echocardiography determined the presence or absence of standard views on fetal ultrasound images. The truthing process was conducted independently of the BrightHeart View Classifier device." This indicates a consensus or independent review process, but not a specific 2+1 or 3+1 adjudication as those usually imply a tie-breaker. It seems like both experts independently determined the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described. The study evaluated the standalone performance of the AI device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    • Yes, a standalone performance study was conducted. The reported sensitivity and specificity values are for the BrightHeart View Classifier identifying standard views on its own.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus (from a sonographer and an MFM specialist with experience in fetal echocardiography).

    8. Sample Size for the Training Set

    • The sample size for the training set is not explicitly stated in the provided document. The document only mentions that "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in performance testing."

    9. How the Ground Truth for the Training Set Was Established

    • The document does not explicitly state how the ground truth for the training set was established. However, given the nature of the device and the ground truth method for the test set, it is highly probable that a similar expert review and annotation process was used for the training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K251071

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-02

    (25 days)

    Product Code
    Regulation Number
    892.2060
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fetal EchoScan is a machine learning-based computer-assisted diagnosis (CADx) software device indicated as an adjunct to fetal heart ultrasound examination in pregnant women aged 18 or older undergoing second-trimester anatomic ultrasound exams.

    When utilized by an interpreting physician, Fetal EchoScan provides information regarding the presence of any of the following suspicious radiographic findings:

    • overriding artery
    • septal defect at the cardiac crux
    • abnormal relationship of the outflow tracts
    • enlarged cardiothoracic ratio
    • right ventricular to left ventricular size discrepancy
    • tricuspid valve to mitral valve annular size discrepancy
    • pulmonary valve to aortic valve annular size discrepancy
    • cardiac axis deviation

    Fetal EchoScan is to be used with cardiac fetal ultrasound video clips containing interpretable 4-chamber, left ventricular outflow tract, right ventricular outflow tract standard views.

    Fetal EchoScan is intended for use as a concurrent reading aid for interpreting physicians (OB-GYN, MFM). It does not replace the role of the physician or of other diagnostic testing in the standard of care. When utilized by an interpreting physician, this device provides information that may be useful in rendering an accurate diagnosis regarding the potential presence of morphological abnormalities that might be suggestive of fetal congenital heart defects that may be useful in determining the need for additional exams.

    Fetal EchoScan is not intended for use in multiple pregnancies, cases of heterotaxy and postnatal ultrasound exams.

    Device Description

    Fetal EchoScan is a cloud-based software-only device which uses neural networks to detect suspicious cardiac radiographic findings for further review by trained and qualified physicians. Fetal EchoScan is intended to be used as an adjunct to the interpretation of the second-trimester fetal anatomic ultrasound exam performed between 18 and 24 weeks of gestation, for pregnant women aged 18 or more.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) Clearance Letter for Fetal EchoScan v1.1:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" but rather presents the performance metrics achieved by the device in both standalone and reader studies. The implication is that these performance levels were deemed acceptable for clearance.

    Table 1. Standalone Performance of Fetal EchoScan v1.1 for all suspicious radiographic findings Combined

    MetricAcceptance Criteria (Implied)Reported Device Performance (Worst-Case Sensitivity, Best-Case Specificity)Reported Device Performance (Best-Case Sensitivity, Worst-Case Specificity)
    Sensitivity for any suspicious findingsHigh (not numerically specified)0.977 (95% CI, 0.954 ; 0.989)0.987 (95% CI, 0.967 ; 0.995)
    Specificity for any suspicious findingsHigh (not numerically specified)0.977 (95% CI, 0.961 ; 0.987)0.963 (95% CI, 0.944 ; 0.976)
    Conclusive Output RateHigh (not numerically specified)98.8% (95% CI, 97.8 ; 99.3)98.8% (95% CI, 97.8 ; 99.3)

    Table 2. Reader Study Performance of Fetal EchoScan v1.1 for all suspicious radiographic findings Combined

    MetricAcceptance Criteria (Implied)Reported Device Performance (AI-Aided)Reported Device Performance (Unaided)Improvement (AI-Aided vs. Unaided)DBM-OR p-value
    ROC AUC for any suspicious findingsSignificantly higher with aid0.974 (95% CI 0.957-0.990)0.825 (95% CI 0.741-0.908)+0.149 (14.9%)0.002
    Mean Sensitivity for any suspicious findingsImproved with aid0.935 (95% CI 0.892-0.978)0.782 (95% CI 0.686-0.878)+0.153 (15.3%)Not explicitly stated for sensitivity/specificity
    Mean Specificity for any suspicious findingsImproved with aid0.970 (95% CI 0.949-0.991)0.759 (95% CI 0.630-0.887)+0.211 (21.1%)Not explicitly stated for sensitivity/specificity

    Note: The numerical acceptance criteria for "high sensitivity" and "high specificity" are not explicitly defined in the provided document, but the reported performance values surpassed what was considered acceptable by the FDA for substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size (Standalone Testing): 877 clinically acquired fetal ultrasound exams.
    • Test Set Sample Size (Reader Study): 200 exams.
    • Data Provenance:
      • Country of Origin: U.S.A. and France.
      • Retrospective or Prospective: The document doesn't explicitly state whether the data was retrospective or prospective, but it mentions "clinically acquired" exams, which often implies retrospective use of existing data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Three (3) pediatric cardiologists.
    • Qualifications of Experts: Pediatric cardiologists. No further details on years of experience or board certification are provided.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Majority voting among the three pediatric cardiologists.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes.
    • Effect Size of Human Readers' Improvement with AI vs. without AI assistance:
      • ROC AUC: Humans improved by +14.9% (from 0.825 unaided to 0.974 aided), with a p-value of 0.002.
      • Mean Sensitivity: Humans improved by +15.3% (from 0.782 unaided to 0.935 aided).
      • Mean Specificity: Humans improved by +21.1% (from 0.759 unaided to 0.970 aided).

    6. Standalone Performance Study

    • Was a standalone study done? Yes.
    • Performance Metrics: Refer to Table 1 above. The AI system had a conclusive output rate of 98.8%. Sensitivity ranged from 0.977 to 0.987, and Specificity ranged from 0.963 to 0.977 for the detection of any suspicious findings, depending on how inconclusive outputs were treated.

    7. Type of Ground Truth Used

    • Ground Truth Type: Expert consensus. Specifically, it was derived from a "truthing process in which three pediatric cardiologists assessed the presence or absence of each of the eight findings, and majority voting was used." This constitutes expert consensus.

    8. Sample Size for the Training Set

    • The document states: "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing." However, the specific sample size for the training set is not provided in the clearance letter. It only mentions that the data used for standalone testing (877 exams) and the reader study (200 exams) were distinct from the training and validation data.

    9. How the Ground Truth for the Training Set Was Established

    • The document states: "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing." However, the methodology for establishing ground truth for the training set is not explicitly detailed in the provided text. It can be inferred that a similar expert review process would have been used, but no specific details are given.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242342

    Validate with FDA (Live)

    Device Name
    Fetal EchoScan
    Manufacturer
    Date Cleared
    2024-11-14

    (99 days)

    Product Code
    Regulation Number
    892.2060
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fetal EchoScan is a machine learning-based computer-assisted diagnosis (CADx) software device indicated as an adjunct to fetal heart ultrasound examination in pregnant women aged 18 or older undergoing second-trimester anatomic ultrasound exams.

    When utilized by an interpreting physician, Fetal EchoScan provides information regarding the presence of any of the following suspicious radiographic findings:

    • overriding artery
    • septal defect at the cardiac crux
    • abnormal relationship of the outflow tracts
    • enlarged cardiothoracic ratio
    • right ventricular to left ventricular size discrepancy
    • tricuspid valve to mitral valve annular size discrepancy
    • pulmonary valve to aortic valve annular size discrepancy
    • cardiac axis deviation

    Fetal EchoScan is to be used with cardiac fetal ultrasound video clips containing interpretable 4-chamber, left ventricular outflow tract, right ventricular outflow tract standard views.

    Fetal EchoScan is intended for use as a concurrent reading aid for interpreting physicians (OB-GYN, MFM). It does not replace the role of the physician or of other diagnostic testing in the standard of care. When utilized by an interpreting physician, this device provides information that may be useful in rendering an accurate diagnosis regarding the potential presence of morphological abnormalities that might be suggestive of fetal congenital heart defects that may be useful in determining the need for additional exams.

    Fetal EchoScan is not intended for use in multiple pregnancies, cases of heterotaxy, and postnatal ultrasound exams.

    Device Description

    Fetal EchoScan is a cloud-based software-only device which uses neural networks to detect suspicious cardiac radiographic findings for further review by trained and qualified physicians. Fetal EchoScan is intended to be used as an adjunct to the interpretation of the second-trimester fetal anatomic ultrasound exam performed between 18 and 24 weeks of gestation, for pregnant women aged 18 or more.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Fetal EchoScan device, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as a set of predefined thresholds. Instead, it presents the performance of the device in various metrics and then concludes that these results demonstrate substantial equivalence. For the purpose of this request, I will infer the implied acceptance criteria from the reported performance and the conclusion of substantial equivalence.

    Inferred Acceptance Criteria & Reported Device Performance

    Metric / FindingInferred Acceptance Criteria (Implicit from conclusion of substantial equivalence)Fetal EchoScan Performance (Worst-Case Sensitivity / Best-Case Specificity)Fetal EchoScan Performance (Best-Case Sensitivity / Worst-Case Specificity)Aided Reader Performance (ROC AUC)
    Standalone Performance
    Any suspicious findingsHigh Sensitivity & High SpecificitySensitivity: 0.977 (0.954-0.989) Specificity: 0.977 (0.961-0.987)Sensitivity: 0.987 (0.967-0.995) Specificity: 0.963 (0.944-0.976)N/A
    Overriding arteryHigh Sensitivity & High SpecificitySensitivity: 0.894 (0.820-0.940) Specificity: 0.989 (0.977-0.995)Sensitivity: 0.942 (0.880-0.973) Specificity: 0.979 (0.963-0.988)0.953 (0.916-0.990)
    Cardiac crux septal defectHigh Sensitivity & High SpecificitySensitivity: 0.905 (0.823-0.951) Specificity: 0.995 (0.985-0.998)Sensitivity: 0.917 (0.838-0.959) Specificity: 0.989 (0.977-0.995)0.971 (0.943-0.999)
    Abnormal OT relationshipHigh Sensitivity & High SpecificitySensitivity: 0.869 (0.781-0.925) Specificity: 0.991 (0.979-0.996)Sensitivity: 0.952 (0.884-0.981) Specificity: 0.989 (0.977-0.995)0.972 (0.953-0.992)
    Enlarged CTRHigh Sensitivity & High SpecificitySensitivity: 0.955 (0.876-0.985) Specificity: 1.000 (0.993-1.000)Sensitivity: 0.955 (0.876-0.985) Specificity: 1.000 (0.993-1.000)0.960 (0.930-0.989)
    Cardiac axis deviationHigh Sensitivity & High SpecificitySensitivity: 0.945 (0.851-0.981) Specificity: 1.000 (0.993-1.000)Sensitivity: 0.945 (0.851-0.981) Specificity: 1.000 (0.993-1.000)0.967 (0.932-1.000)
    PV/AV size discrepancyHigh Sensitivity & High SpecificitySensitivity: 0.954 (0.914-0.975) Specificity: 0.989 (0.977-0.995)Sensitivity: 0.954 (0.914-0.975) Specificity: 0.989 (0.977-0.995)0.979 (0.962-0.997)
    RV/LV size discrepancyHigh Sensitivity & High SpecificitySensitivity: 0.950 (0.900-0.975) Specificity: 1.000 (0.993-1.000)Sensitivity: 0.950 (0.900-0.975) Specificity: 1.000 (0.993-1.000)0.991 (0.983-0.999)
    TV/MV size discrepancyHigh Sensitivity & High SpecificitySensitivity: 0.943 (0.896-0.970) Specificity: 1.000 (0.993-1.000)Sensitivity: 0.943 (0.896-0.970) Specificity: 1.000 (0.993-1.000)0.964 (0.938-0.990)
    MRMC Study Performance
    ROC AUC (any suspicious finding)Significantly higher with aid than unaidedN/AN/AAided: 0.974 (0.957-0.990) Unaided: 0.825 (0.741-0.908)
    Mean Sensitivity (Any finding)Increased with aidN/AN/AAided: 0.935 (0.892-0.978) Unaided: 0.782 (0.686-0.878)
    Mean Specificity (Any finding)Increased with aidN/AN/AAided: 0.970 (0.949-0.991) Unaided: 0.759 (0.630-0.887)
    Conclusive output rateHigh98.8% (95% CL, 97.8-99.3)N/AN/A

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size for Standalone Test Set: 877 clinically acquired fetal ultrasound exams.
    • Sample Size for MRMC Test Set: 200 exams.
    • Data Provenance: The data was collected from 11 centers in the U.S.A. and France. It was retrospectively collected as it refers to "clinically acquired fetal ultrasound exams".

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Three (3) pediatric cardiologists.
    • Qualifications of Experts: The document specifies "pediatric cardiologists" but does not provide details on their years of experience or other specific qualifications beyond their specialty.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Majority voting. This means that if at least two out of the three pediatric cardiologists agreed on the presence or absence of a finding, that was established as the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes.
    • Effect size of human readers improvement with AI vs. without AI assistance:
      • ROC AUC for any suspicious finding: +14.9% increase (from 0.825 unaided to 0.974 aided, p=0.002).
      • Mean Sensitivity for any suspicious finding: +15.3% increase (from 0.782 unaided to 0.935 aided).
      • Mean Specificity for any suspicious finding: +21.1% increase (from 0.759 unaided to 0.970 aided).

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes.
    • The results are presented in Table 1, showing sensitivity and specificity for "Any suspicious findings" and each individual finding, calculated under two scenarios for inconclusive outputs.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus. Specifically, it was derived from a truthing process by three pediatric cardiologists using majority voting.

    8. Sample Size for the Training Set

    • The document states that "The ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing," but it does not explicitly provide the sample size for the training set.

    9. How the Ground Truth for the Training Set Was Established

    • The document states that the "ultrasound examinations used for training and validation are entirely distinct from the examinations used in standalone testing." However, similar to the training set sample size, it does not explicitly describe how the ground truth for the training set was established. It only details the ground truth establishment for the test sets (standalone and MRMC).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1