Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K251391

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-07

    (186 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.

    The BrainsWay Deep TMS™ System is also indicated as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (age 15-21 years).

    Device Description

    The BrainsWay Deep TMS™ System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The BrainsWay Deep TMS™ System is composed of the following main components:

    1. Cart
      a) TMS Neurostimulator
      b) Cooling System
      c) Positioning Device
    2. Helmet
      a) Aiming Apparatus (i.e., ruler/grid)
      b) Electromagnetic Coil (H Coil)
      c) Cap

    The BrainsWay Deep TMS™ System is identical to the previously cleared BrainsWay Deep TMS™ Systems.

    The purpose of this 510(k) submission is to enable modifications to the device software and for use as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (age 15-21 years).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251449

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-13

    (127 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.

    Device Description

    The BrainsWay Deep TMS™ System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The BrainsWay Deep TMS™ System is composed of the following main components:

    1. Cart
      a) TMS Neurostimulator
      b) Cooling System
      c) Positioning Device
    2. Helmet
      a) Aiming Apparatus (i.e., ruler/grid)
      b) Electromagnetic Coil (H Coil)
      c) Cap

    The BrainsWay Deep TMS™ System is identical to the previously cleared BrainsWay Deep TMS™ Systems.

    The purpose of this 510(k) submission is to enable modifications to the device software and expansion of the treatment stimulation protocols, to include the accelerated iTBS stimulation protocol.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K222196

    Validate with FDA (Live)

    Device Name
    Deep TMS System
    Manufacturer
    Date Cleared
    2024-05-31

    (679 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.

    Device Description

    The BrainsWay Deep TMS™ System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The BrainsWay Deep TMS™ System is composed of the following main components:

    1. Cart
      a) TMS Neurostimulator
      b) Cooling System
      c) Positioning Device
    2. Helmet
      a) Aiming Apparatus (i.e., ruler/grid)
      b) Electromagnetic Coil (H Coil)
      c) Cap
    AI/ML Overview

    The provided text is a 510(k) summary for the BrainsWay Deep TMS™ System, seeking to enable modifications to the device labeling based on clinical data. It is important to note that this document does not describe an AI medical device that has acceptance criteria based on diagnostic performance metrics (e.g., sensitivity, specificity, AUC) and a study proving those criteria were met via a test set adjudicated by experts. Instead, this document details a study for a therapeutic device (Transcranial Magnetic Stimulation System) and focuses on clinical outcomes (remission and response rates for Major Depressive Disorder (MDD)).

    Therefore, I cannot fulfill all parts of your request as it pertains to an AI medical device's performance study. However, I can extract the relevant information about the clinical studies conducted for the BrainsWay Deep TMS™ System's expanded indication for use.

    Here's an analysis based on the provided text, adapted to the context of a therapeutic device rather than a diagnostic AI:


    Acceptance Criteria and Study for BrainsWay Deep TMS™ System

    This 510(k) submission primarily focuses on expanding the indicated adult population for the BrainsWay Deep TMS™ System to include adult subjects >22 and <86 years of age suffering from MDD, who failed to achieve satisfactory improvement from previous antidepressant medication treatment. The "acceptance criteria" here are not diagnostic performance metrics, but rather clinical efficacy outcomes that demonstrate safety and effectiveness in the expanded patient population, comparable or superior to previously cleared indications and/or standard of care.

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a therapeutic device, the "acceptance criteria" are implied to be clinically meaningful improvements in depression symptoms (remission and response rates) and safety comparable to previous findings. The document measures performance against its own previous studies (MDD Multicenter Study) and sham treatment.

    Acceptance Criterion (Implicit)Reported Device Performance (Kaster 2018 Study - Active Deep TMS)Reported Device Performance (Kaster 2018 Study - Sham Deep TMS)Comparison to Prior Data (MDD Multicenter Study - Deep TMS)
    Remission Rate (HDRS-21 after 4 Weeks)64.0% (16/25) [42.5%; 82.0%]18.5% (5/27) [6.3%; 38.1%]Higher (vs. 30.4%)
    Response Rate (HDRS-21 after 4 Weeks)64.0% (16/25) [42.5%; 82.0%]22.2% (6/27) [8.6%; 42.3%]Higher (vs. 37.0%)
    NNT for Remission (Implied clinical benefit)2.2Not applicableBetter (vs. 6.85)
    NNT for Response (Implied clinical benefit)2.4Not applicableBetter (vs. 7.05)
    Change from baseline in HDRS-21 scores (Diff between active and sham)5.35 points (statistically significant, p=0.0025)Not directly applicable (this is a differential)Higher (vs. 2.23 points)
    Safety ProfileAdverse effects similar and typical of TMS treatment; no new or unanticipated adverse events.Adverse effects similar and typical of TMS treatment; no new or unanticipated adverse events.Comparable
    CGI-S Remission Rate (RWD)Comparable to MDD Multicenter Study CGI-S remission rateNot applicableComparable
    CGI-S Response Rate (RWD)Significantly greater than MDD Multicenter Study CGI-S response rateNot applicableSignificantly greater
    Change from baseline in CGI-S scores (RWD)Significantly better than MDD Multicenter Study CGI-S resultsNot applicableSignificantly better

    2. Sample Sizes and Data Provenance for Test Set (Clinical Studies)

    • Kaster 2018 Study (Double-blind, Randomized, Sham-controlled Trial):
      • Total Sample Size: 52 participants
      • Active Deep TMS Treatment Group: 25 participants
      • Sham Deep TMS Treatment Group: 27 participants
      • Provenance: Not explicitly stated, but the study is referenced as "Kaster 2018 Study", implying a published clinical trial. The context of a 510(k) submission generally implies prospective clinical trial data.
      • Population: Elderly MDD patient population (> 60 years old).
    • Real-World Data (RWD):
      • Total Patient Records: 1,198 patients
      • Source: Electronic health records from 20 sites over the US.
      • Time Span: 2012 to 2022 (10 years).
      • ITT Patient Population Analysis (met eligibility criteria): 152 subjects
      • Population: Elderly MDD patients (>69 years old: mean age 73.6, range 69-86).
      • Provenance: Retrospective, from the US.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This is a clinical efficacy study for a therapeutic device, not an AI diagnostic device. Therefore, the concept of "ground truth" established by experts for a test set (e.g., radiological reads) as would be the case for an AI diagnostic device, does not apply here.
    • Clinical outcomes (HDRS-21, CGI-S, PHQ-9 scores) are typically rated by trained clinicians (e.g., psychiatrists, psychologists). The document does not specify the number or qualifications of clinicians involved in assessing these scores for the Kaster 2018 Study or the RWD. It mentions that approximately 50% of RWD patients were treated by providers with >10 years or <10 years of experience, which relates to treatment providers, not necessarily independent assessors for ground truth.

    4. Adjudication Method for the Test Set:

    • Again, for a therapeutic device efficacy study measuring clinical scales, "adjudication" in the sense of reconciling reviewer disagreements on diagnostic labels is not directly applicable.
    • The Kaster 2018 Study was a double-blind, randomized, sham-controlled trial, which is a strong methodology for assessing treatment efficacy, inherently minimizing bias in outcome assessment. Blinding (neither patient nor assessor knows treatment assignment) often serves a similar purpose to prevent bias as adjudication attempts to for diagnostic interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size of human improvement with vs. without AI assistance:

    • No, an MRMC study was not done. This document is for a therapeutic device. MRMC studies are typically performed for diagnostic AI tools to evaluate their impact on human reader performance.
    • The study does compare the device's effect to a sham treatment and to previously documented performance (MDD Multicenter Study), showing improvement with the active device. However, this is not a human-AI collaboration study.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The BrainsWay Deep TMS™ System is a medical device, a transcranial magnetic stimulation system, that directly delivers therapy, not a diagnostic algorithm. Therefore, there is no "standalone" algorithm performance to evaluate. Its "performance" is its clinical efficacy when administered.

    7. The Type of Ground Truth Used:

    • The "ground truth" for efficacy was established through clinical outcome measures based on standardized psychiatric rating scales.
      • Remission: HDRS-21 (Hamilton Depression Rating Scale 21-item), PHQ-9 (Patient Health Questionnaire-9).
      • Response: HDRS-21, PHQ-9, CGI-S (Clinical Global Impression – Severity).
    • These are based on the assessment of clinical symptoms and patient status by trained clinicians, not pathology, imaging, or direct outcomes data in the sense of a definitive diagnostic gold standard.

    8. The Sample Size for the Training Set:

    • Not applicable. This device is a hardware therapeutic system, not an AI algorithm that undergoes a training phase with a dataset. The results presented are from clinical trials and real-world data specifically used to support the expanded indications for use, not to "train" the device.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable, for the same reasons as #8. There is no training set or corresponding ground truth for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220819

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-26

    (158 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.

    Device Description

    The BrainsWay Deep TMS™ System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure. The BrainsWay Deep TMS™ System is composed of the following main components: Cart (TMS Neurostimulator, Cooling System, Positioning Device), and Helmet (Aiming Apparatus, Electromagnetic Coil (H7 Coil), Cap).

    AI/ML Overview

    The provided text from the FDA 510(k) summary does not contain acceptance criteria or a study that directly proves the device meets those specific criteria in the way typically seen for AI/ML devices with quantitative performance metrics.

    Instead, the document details a non-inferiority study to demonstrate that a modified version of a device (BrainsWay Deep TMS™ System with H7 Coil) is as safe and effective as a previously cleared predicate device (BrainsWay Deep TMS™ System with H1 Coil). The "acceptance criteria" in this context are implicitly that the new device performs at least as well as the predicate device within a defined non-inferiority margin for key clinical outcomes.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Non-inferiority Limit)Reported Device Performance (H7 vs H1 Coil)
    Upper limit of one-sided 95% CI for the difference in change from baseline in HDRS-21 at 6 weeks is lower than 3 (non-inferiority limit).Upper limit of one-sided 95% CI was 1.9.
    No statistically significant difference in response rates and remission rates at week 6.Response rates and remission rates at week 6 were not statistically significantly different.
    No statistically significant difference in the incidence of adverse events.No statistically significant difference in the incidence of any adverse events reported.

    2. Sample size used for the test set and the data provenance

    • Sample size: 169 subjects total.
      • Specific breakdown by H7 and H1 coil groups is not explicitly stated, but the study was a "Randomized Controlled Trial," implying subjects were divided between these two groups.
    • Data provenance: Prospective, Multicenter. Country of origin not specified, but the study was titled "Multicenter H7 vs H1 Study."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The "ground truth" in this study relates to clinical outcomes over time (e.g., changes in HDRS-21 scores). These scores are typically assessed by trained clinicians, often psychiatrists or psychologists, but the exact number of experts or their specific qualifications for this study are not detailed in the provided text. Scores like HDRS-21 are standardized clinical assessments, not typically "established" by multiple experts for each patient like in imaging adjudication.

    4. Adjudication method for the test set

    Not explicitly stated. Clinical rating scales like HDRS-21 are usually administered by trained raters, potentially with inter-rater reliability checks, but a formal adjudication method like "2+1" or "3+1" is not mentioned as it would be for diagnostic imaging studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, this was not an MRMC comparative effectiveness study involving human readers with or without AI assistance. It was a clinical trial comparing two different device configurations (H7 coil vs. H1 coil) for direct patient treatment. The device itself is a treatment system, not a diagnostic AI tool. Therefore, an effect size of human reader improvement with AI assistance is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a deep transcranial magnetic stimulation (TMS) system for treating Major Depressive Disorder, not an algorithm or AI system used for diagnosis or image analysis in a standalone capacity.

    7. The type of ground truth used

    The ground truth used was clinical outcome data from patient assessments, specifically changes in the Hamilton Depression Rating Scale (HDRS-21) and Hamilton Anxiety Rating Scale (HARS) scores, as well as response and remission rates. These are standard measures of treatment efficacy in psychiatry.

    8. The sample size for the training set

    Not applicable. This device is a medical device for treatment, not an AI/ML algorithm that requires a "training set" in the conventional sense. The "study" described (CTP-0002-00) is a clinical trial to demonstrate safety and effectiveness for regulatory approval.

    9. How the ground truth for the training set was established

    Not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210201

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-08-17

    (204 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrainsWay Deep TMS™ System is indicated for the treatment of depressive episodes and for decreasing anxiety symptoms for those who may exhibit comorbid anxiety symptoms in adult patients suffering from Major Depressive Disorder (MDD) and who failed to achieve satisfactory improvement from previous antidepressant medication treatment in the current episode.

    Device Description

    The BrainsWay Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The BrainsWay Deep TMS System is composed of the following main components:

      1. Cart
      • a) TMS Neurostimulator
      • b) Cooling System
      • c) Positioning Device
      1. Helmet
      • a) Aiming Apparatus (i.e., ruler/grid)
      • b) Electromagnetic Coil (H1-Coil)
      • c) Cap
    AI/ML Overview

    The provided text describes the 510(k) summary for the BrainsWay Deep TMS System, primarily focusing on its expanded indication for decreasing anxiety symptoms in adult patients with Major Depressive Disorder (MDD). The document does not describe the acceptance criteria and the study that proves the device meets the acceptance criteria (except for performance standards from medical device norms and guidance mentioned on page 5). While it references clinical studies to demonstrate safety and effectiveness for the expanded indication, it doesn't detail the specific statistical acceptance criteria that the device had to meet based on these studies for clearance.

    Here's an analysis of the provided information relative to your request:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: The document does not explicitly state quantitative acceptance criteria for clinical performance. It mentions conformance to FDA recognized consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 10993-5, ISO 10993-10, ISO 14971) for non-clinical testing (Output Waveform, Electrical Field Spatial Distribution, Magnetic Field Strength Gradient Testing, electrical and mechanical safety, EMC, software validation). For clinical performance, the criterion seems to be demonstrating "safety and effectiveness" and "clinically significant decrease in anxiety symptoms" by showing "effect sizes" that are "consistent, robust and similar" to those reported for MDD.

    • Reported Device Performance: The document reports "effect sizes" (Cohen's d) from three Randomized Controlled Trials (RCTs) and meta-analyses, specifically related to the reduction of anxiety symptoms measured by the HDRS-21 Anxiety-Somatization Factor score.

      Study / AnalysisEffect Size (Cohen's d)95% Confidence Limits (Lower, Upper)
      Levkovitz 2015 (Deep TMS vs. Sham)0.34(0.009, 0.667)
      Kaster 2018 (Deep TMS vs. Sham)0.36(-0.23, 0.95)
      Filipcic 2019 (Deep TMS vs. Medication as SOC)0.90 (DH1 vs DSOC)(0.55, 1.26)
      Meta-analysis of three RCTs (Deep TMS vs. Sham/SOC)0.55 (weighted pooled)Not specified
      Kezior et al. Meta-analysis (6/11 studies)1.45 (pooled weighted)Not specified
      Hung et al. Meta-analysis (8/11 studies)-1.282 (Hedges' g)Not specified

      The document highlights that these effect sizes "may be compared to the effect size seen in the Multicenter MDD Study for all MDD subjects using the HDRS-21 score, where an effect size of 0.76... was reported."

    2. Sample sizes used for the test set and the data provenance:

    • Test Set Description: The "clinical performance data" section indicates that the evidence supporting the expanded indication came from "3 Randomized Controlled Trials (RCTs) and supportive data from meta-analyses from 11 clinical studies (including the 3 RCTs and another 8 open label studies) using the BrainsWay Deep TMS device in 573 patients."
    • Sample Sizes per RCT:
      • Levkovitz 2015: DTMS: 89, Sham: 92
      • Kaster 2018: DTMS: 25, Sham: 27
      • Filipcic 2019: DTMS: 65, Medication as SOC: 72
    • Data Provenance: The document implies these were clinical trials, which are typically prospective. The location of the studies (country of origin) is not explicitly stated, although the applicant and contact person are based in Israel, suggesting that some of these studies might have been conducted there or internationally. The term "Multicenter MDD Study" could imply international collaboration. The provenance is explicitly listed as a "prospective multicenter randomized controlled trial" for Levkovitz 2015, and "prospective randomized controlled trial" for Kaster 2018.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. Clinical trials of this nature would typically involve medical professionals (e.g., psychiatrists) for patient assessment and diagnosis (which forms the basis for "ground truth" for patient inclusion criteria and outcome measures), but the specific number and qualifications of experts for objective ground truth establishment (e.g., blinded rating panels) are not detailed.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This information is not provided in the document. For clinical outcome measures, some studies use blinded independent raters, but whether an adjudication process was used among multiple raters is not mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable to this device. The BrainsWay Deep TMS System is a therapeutic device (Repetitive Transcranial Magnetic Stimulation system), not an AI-assisted diagnostic or imaging interpretation tool that would typically involve human "readers" or MRMC studies. The clinical studies focus on direct outcomes from the device's therapeutic application.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable to this device. As a therapeutic device, its performance is evaluated based on its clinical effect on patients, not as an algorithm performing a standalone task. The device works with human operators (healthcare professionals) who administer the treatment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" in these studies is outcomes data based on clinical assessments.
      • Patient Diagnosis: Major Depressive Disorder (MDD) as per DSM-IV criteria, with patients who failed previous antidepressant medication treatment. This implies diagnosis by qualified clinicians.
      • Outcome Measure: Change in HDRS-21 Anxiety-Somatization Factor Score from baseline. The HDRS-21 (Hamilton Depression Rating Scale) is a widely used clinician-administered rating scale. While administered by humans, it yields quantitative data. The "ground truth" for the effectiveness of the treatment is the statistically significant reduction in these symptom scores compared to control groups (Sham or medication as standard of care).

    8. The sample size for the training set:

    • This information is not applicable as this is not an AI/ML device that requires a distinct "training set" in the traditional sense for algorithm development. The "training" for the device, if any, would refer to the pre-clinical characterization and validation. The clinical studies mentioned are for validation of the device's therapeutic effect.

    9. How the ground truth for the training set was established:

    • This is not applicable for the reasons stated above.

    In summary, the provided document details the non-clinical performance standards and clinical study results (effect sizes) that support the device's expanded indication. However, it does not provide explicit pre-defined quantitative acceptance criteria for the clinical studies, nor does it address aspects like expert consensus or MRMC studies, which are typically relevant for diagnostic AI/ML devices. The "acceptance" for this device appears to be based on demonstrating statistically and clinically meaningful improvement in anxiety symptoms, which the reported effect sizes aim to prove.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203735

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-23

    (123 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrainsWay Deep TMS System is indicated for the treatment of depressive episodes in adult patients suffering from Major Depressive Disorder who failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode.

    Device Description

    The BrainsWay Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.
    The BrainsWay Deep TMS System is composed of the following main components:

      1. Cart
      • a) TMS Neurostimulator
      • b) Cooling System
      • c) Positioning Device
      1. Helmet
      • a) Aiming Apparatus (i.e., ruler/grid)
      • b) Electromagnetic Coil (H1-Coil)
      • c) Cap
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BrainsWay Deep TMS System with iTBS Protocol, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this 510(k) submission are established through demonstrating non-inferiority of the new iTBS protocol compared to an already FDA-cleared High Frequency (HF) protocol for the same device. The primary endpoint for non-inferiority was the change in HDRS-21 score.

    Acceptance Criteria (Study Hypothesis - Null Hypothesis)Reported Device Performance (Summary of Results)
    The mean change in HDRS-21 score from baseline at 5 weeks in the iTBS Deep TMS group is inferior to the mean change in HDRS-21 score from baseline at 5 weeks in the HF Deep TMS group, by more than 3 points. (Non-inferiority margin: 3 points)The non-inferiority analysis of the change from baseline between the two treatment groups was assessed. Since the upper limit of the confidence interval was less than the non-inferiority margin of 3 points, the null hypothesis of inferiority was rejected, indicating non-inferiority between the iTBS Deep TMS and HF Deep TMS treatments, when adjusted for baseline HDRS-21 scores.
    Secondary Endpoints (No specific numerical acceptance criteria mentioned, but demonstrate similar performance)HDRS-21 Response Rate at 5 weeks, HDRS-21 Remission Rate at 5 weeks, Change in CGI-S from baseline to 6 weeks, CGI-S Response Rate at 6 weeks, CGI-S Remission Rate at 6 weeks. (The document states "the change from baseline was similar between the two treatment groups at 5 weeks," implying similar performance on these secondary endpoints as well.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 146 subjects
      • 66 subjects received iTBS stimulation protocol.
      • 80 subjects received HF stimulation protocol.
    • Data Provenance: Not explicitly stated as retrospective or prospective, nor country of origin. However, given it's a clinical study for FDA clearance, it's highly likely to be a prospective multi-center study, though the specific locations are not provided in this summary. The mention of "multicenter MDD study" for age eligibility criteria might hint at a multi-site study.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This particular study does not involve establishing ground truth through expert review of specific data like images. Instead, the "ground truth" for the effectiveness of the treatment is based on patient-reported outcomes and clinician-administered scales (HDRS-21, CGI-S), which inherently involve expert clinicians in their administration and interpretation during the study.

    4. Adjudication Method

    Not applicable in the typical sense of radiological or pathological adjudication. The primary and secondary endpoints are based on standardized clinical assessment scales (HDRS-21, CGI-S) administered by qualified clinical personnel as part of the study protocol.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This study is a clinical trial comparing two different stimulation protocols (iTBS vs. HF) on human patients for Major Depressive Disorder (MDD), not a reader study involving interpretation of cases by multiple human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a medical device for treatment, not an AI/algorithm for diagnosis or analysis where standalone performance would typically be evaluated. The "performance" being assessed is the clinical effectiveness of the device (iTBS protocol) in treating MDD. Therefore, a standalone algorithm-only performance study is not applicable.

    7. The Type of Ground Truth Used

    The ground truth used for determining the effectiveness of the device in this clinical study is:

    • Clinical Outcomes Data: Specifically, changes in patient depression levels as measured by established clinical rating scales (HDRS-21 and CGI-S scores) from baseline to follow-up, along with derived response and remission rates.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a clinical study to evaluate the effectiveness of a medical device's protocol, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set mentioned or implied for a machine learning algorithm. The "training" for this device refers to the clinical application of the stimulation protocol in the context of treating patients.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203616

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-16

    (127 days)

    Product Code
    Regulation Number
    882.5802
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brainsway Deep Transcranial Magnetic Stimulation System is indicated to be used as an aid in short-term smoking cessation for adults.

    Device Description

    The Brainsway Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The Brainsway Deep TMS System is composed of the following main components:

    1. Cart
      a) TMS Neurostimulator
      b) Cooling System
      c) Positioning Device
    2. Helmet
      a) Aiming Apparatus (i.e., ruler/grid)
      b) Electromagnetic Coil (H4-Coil)
      c) Cap
    AI/ML Overview

    Based on the provided text, the device in question is the Brainsway Deep TMS System, indicated as an aid in short-term smoking cessation for adults. The document outlines a 510(k) submission, classifying it as a Class II medical device.

    Crucially, the provided document explicitly states under the "Clinical Performance Data" section: "Not Applicable." This indicates that no direct clinical efficacy or performance study of the device itself (i.e., algorithm only or human-in-the-loop performance with the device) was conducted or provided for this specific 510(k) submission for the Brainsway Deep TMS System for smoking cessation.

    Instead, the submission focuses on demonstrating substantial equivalence to previously cleared predicate devices by Brainsway Ltd., specifically:

    • Main Predicate (K200957): Brainsway Deep TMS System for smoking cessation.
    • Reference Predicate (K183303): Brainsway Deep TMS System for OCD (used to demonstrate equivalence of device modifications like the new stimulator).

    The acceptance criteria and performance data provided are primarily non-clinical (bench) performance data showing that the modified Brainsway Deep TMS System performs similarly to already cleared versions. The "study that proves the device meets the acceptance criteria" in this context is the bench testing and compliance with recognized standards to demonstrate that the modifications made do not raise new safety or effectiveness concerns compared to the predicate devices.

    Therefore, many of the requested details about acceptance criteria, test set sample size, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth for training/test sets are not applicable or not provided in this specific 510(k) summary, as the clearance relies on substantial equivalence based on technical and performance similarities to existing cleared devices, rather than new clinical efficacy demonstrations.

    Given this, I will answer the questions based on the information available in the document, highlighting what is and is not present.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission based on non-clinical performance and a pre-existing indication, the "acceptance criteria" are not framed as specific clinical performance metrics (e.g., sensitivity, specificity for a diagnostic device) but rather as successful demonstration of technical equivalence and safety.

    Acceptance Criteria CategoryReported Device Performance
    Non-Clinical (Bench) Performance
    Output WaveformDemonstrated "substantially equivalent" to the previously cleared Brainsway DTMS System (K200957) through performance testing.
    Electrical Field Spatial DistributionDemonstrated "substantially equivalent" to the previously cleared Brainsway DTMS System (K200957) through performance testing.
    Magnetic Field Strength GradientDemonstrated "substantially equivalent" to the previously cleared Brainsway DTMS System (K200957) through performance testing.
    Electrical SafetyCompliance with IEC 60601-1 standard. Results of tests demonstrated substantial equivalence.
    Mechanical SafetyCompliance with IEC 60601-1 standard. Results of tests demonstrated substantial equivalence.
    Electromagnetic CompatibilityCompliance with IEC 60601-1-2 standard. Results of tests demonstrated substantial equivalence.
    Software ValidationCompliance with FDA guidelines for software validation and IEC 62304 standard requirements. The new stimulator software has been validated to ensure proper performance.
    BiocompatibilityMaterials are biocompatible.
    Functional Equivalence
    Intended Use/Indications for UseSame as the main predicate device (Brainsway Deep TMS System cleared in K200957) for aid in short-term smoking cessation for adults.
    Mechanism of ActionSimilar mechanism of action (electromechanical instrument producing pulsed magnetic fields to induce electrical currents in targeted brain regions for non-invasive stimulation of cortical neurons).
    System ComponentsBasic components are similar to cleared predicate devices. Minor modifications (new Brainsway stimulator, minor cart/cooling system/coil wire/helmet changes) introduced, but already present in reference predicate (K183303).
    Safety FeaturesIntroduces similar safety features and complies with the same relevant consensus standards as predicate devices. Potential hazards from modifications were mitigated through performance testing.
    Performance ParametersTreatment parameters (Magnetic Field Intensity, Frequency, Train duration, Inter-train interval, Number of trains, Magnetic Pulses per Session, Tx Session Duration, Session frequency) are identical to the main predicate (K200957).
    Environmental ParametersCoil's Operating Temperature, Storage Temperature, Atmospheric Pressure Range, and Relative Humidity Range are within the same or comparable bounds as the predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable / Not specified. The data provided focuses on non-clinical bench testing of the modified device's components and system performance against engineering and safety standards, rather than clinical performance on a patient test set.
    • Data Provenance: The document does not specify the country of origin for the non-clinical bench test data. It is inherently prospective in the sense that new tests were performed on the modified device, but these are engineering tests, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. As indicated in point 2, there was no clinical test set requiring expert ground truth for patient outcomes or conditions. The "ground truth" for the non-clinical tests would be the established engineering and safety standards (e.g., IEC standards, FDA guidelines for software validation).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set requiring adjudication was performed or reported in this 510(k) summary.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a medical device for treatment (Transcranial Magnetic Stimulation), not an AI diagnostic/imaging device. Therefore, a MRMC study for AI assistance to human readers is not relevant and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The device (Brainsway Deep TMS System) is a physical medical device. It does not operate as a standalone algorithm in the way a diagnostic AI would. Its performance is evaluated through its physical output (magnetic pulses) and electrical/mechanical characteristics as detailed in the non-clinical tests, and its substantial equivalence to a predicate device. Clinical performance of the therapy using the device relies on the operator (human-in-the-loop) and the patient, but the performance evaluation here is for the device's technical specifications.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For non-clinical performance: The "ground truth" is adherence to established engineering, electrical, mechanical, and software safety standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 62304) and comparison to the output/characteristics of the legally marketed predicate devices.
    • For the overall indication (smoking cessation): The acceptability of the indication itself and its efficacy as a therapy stems from the prior clearance of the predicate device (K200957), which would have established this efficacy through clinical trials (though not detailed in this particular 510(k) summary).

    8. The sample size for the training set

    • Not applicable. This device is not an AI/ML algorithm that requires a "training set" in the conventional sense for learning patterns from data. The "training" in this context would be the design and development processes for the hardware and software, and testing against specifications.

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no "training set" for an AI/ML algorithm in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200957

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-21

    (134 days)

    Product Code
    Regulation Number
    882.5802
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brainsway Deep Transcranial Magnetic Stimulation System is indicated to be used as an aid in short-term smoking cessation for adults.

    Device Description

    The Brainsway Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scap. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The FDA cleared Brainsway Deep TMS System is composed of the following main components:

      1. Electromagnetic Coil
      1. TMS Neurostimulator
      1. Cooling System
      1. Positioning System and Helmet
      1. Cart
    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the Brainsway Deep TMS System (with HADD-coil) for short-term smoking cessation were primarily based on the statistical significance and clinical meaningfulness of the 4-week Continuous Quit Rate (CQR), supported by secondary endpoints.

    Acceptance CriteriaReported Device Performance
    Primary Endpoint: Statistically significant and clinically meaningful higher 4-week Continuous Quit Rate (CQR) in the DTMS arm compared to the sham arm.Primary Endpoint: The 4-week CQR was statistically significantly higher (p-value = 0.0238) in the DTMS arm (17.1%) than in the sham arm (7.9%) for the ITT-Safety population (N=262) up to 4 months follow-up. This was considered clinically meaningful.
    Secondary Endpoint: Statistically significant higher 4-week CQR in subjects with at least 4 weeks of diary records.Secondary Endpoint: The 4-week CQR was statistically significantly higher (p-value = 0.0071) in the DTMS arm (27.3%) compared to the sham arm (11.3%) for subjects with at least 4 weeks of diary records.
    Secondary Endpoint: Statistically significant lower number of cigarettes smoked per day (per diary entry) in the DTMS treatment arm compared to the sham arm.Secondary Endpoint: The number of cigarettes smoked per day (per diary entry) was statistically significantly lower in the DTMS treatment arm compared to the sham arm (specific p-value not given for this point, but stated as statistically significant).
    Secondary Endpoint: Statistically significant higher 4-week CQR up to the 6th week visit.Secondary Endpoint: The 4-week CQR up to the 6th week visit was statistically significantly higher (p-value = 0.0022) in the DTMS arm (15.4%) than in the sham arm (4.3%).
    Safety Profile: No individual adverse event types with a significant difference between study groups, except for expected application site discomfort and muscle twitching.Safety Profile: No individual adverse event types showed a significant difference between groups, except for application site discomfort and muscle twitching. Heading reporting was not statistically significant. Application site discomfort (11.38% vs 2.16%, p=0.0043) and muscle twitching (5.69% vs 0%, p=0.0046) were higher in the DTMS group but were not considered to deter treatment.

    Study Details Proving Acceptance Criteria

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: N=262 (ITT-Safety population), with 123 subjects in the DTMS arm and 139 subjects in the sham arm.
      • Data Provenance: The study was a "prospective, double blind, randomized, sham controlled, multi-center trial." No specific countries of origin for the data are explicitly stated, but multi-center typically implies multiple sites, which could be in one or more countries. It is a prospective clinical trial.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not explicitly state the "ground truth" establishment by independent experts for the smoking cessation outcomes. The ground truth for effectiveness (smoking cessation) was determined by objective measures like Continuous Quit Rate (CQR) based on participant self-reporting and likely biochemical verification (though not explicitly stated for this 510(k) summary, it's common in smoking cessation trials). The study was clinically managed, but the role of external experts in adjudicating individual cases of "quit" status is not detailed.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • The document does not describe an adjudication method for the test set outcomes (i.e., whether participants met the smoking cessation criteria). The outcomes appear to be derived from direct data collection (e.g., self-reported smoking diary, likely backed by CO-oximetry or similar biochemical tests, though not mentioned in this excerpt) rather than subjective expert interpretation requiring adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, this was not an MRMC comparative effectiveness study. This device is a treatment device (Transcranial Magnetic Stimulation), not an AI diagnostic or assistance tool that would involve human readers interpreting cases.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This question is not applicable. The Brainsway Deep TMS System is a medical device for treatment, not an algorithm, and it is designed for use by a human operator (clinician) to administer the therapy. Its performance is as a therapeutic device, not an AI or algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth for device efficacy was outcomes data directly from the clinical trial, specifically the 4-week Continuous Quit Rate (CQR) for smoking cessation, along with other self-reported and indirect measures of smoking behavior (e.g., number of cigarettes per day).
    7. The sample size for the training set:

      • This question is not applicable. This is a medical device for treatment, not an AI or machine learning model that requires a "training set." The clinical trial data (N=262) served as the validation for its efficacy for the stated indication.
    8. How the ground truth for the training set was established:

      • This question is not applicable as there is no "training set" in the context of this traditional medical device validation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183303

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-08

    (101 days)

    Product Code
    Regulation Number
    882.5802
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brainsway Deep Transcranial Magnetic Stimulation System is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder.

    Device Description

    The proposed device is a modification to the Brainsway Deep TMS System that was cleared under DEN170078. Brainsway has implemented minor changes to the helmet configuration, improved the user interface, modified the measurement method, and made other minor component changes. None of these changes alter the technical specifications for the device, which maintains the same voltage and current, same frequencies, and same electrical mains compatibility. The same technological features have previously been cleared by FDA for use in major depressive disorder in K173540.

    Consistent with the predicate and reference device, the proposed Brainsway Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    AI/ML Overview

    The provided text describes modifications to the Brainsway Deep TMS System and asserts its substantial equivalence to a previously cleared predicate device. It does not provide detailed acceptance criteria or results of a clinical study to prove the device meets specific performance criteria. Instead, it focuses on demonstrating that the modifications do not introduce new questions of safety or effectiveness compared to the predicate device.

    Therefore, many of the requested details about acceptance criteria, clinical study methodology, and ground truth establishment cannot be found in the provided document. The document primarily confirms that the modified device functions as intended and meets the same acceptance criteria as the predicate and reference devices, based on software verification, electrical safety, EMC, and bench performance testing.

    Here's an analysis based on the provided text, highlighting what is available and what is not:


    Acceptance Criteria and Device Performance

    The document does not provide a specific table of acceptance criteria with numerical targets and corresponding reported device performance for a clinical study proving efficacy or diagnostic accuracy. Instead, it states:

    "In all instances, the subject device functions as intended and meets all the same acceptance criteria as the predicate and reference device."

    This implies that the acceptance criteria are generally related to the proper functioning of the device, electrical safety, electromagnetic compatibility, and bench performance, rather than clinical efficacy metrics. For an efficacy study demonstrating treatment outcomes for OCD, such details would typically be found in a separate clinical study report.

    The "Indications for Use" section states: "The Brainsway Deep Transcranial Magnetic Stimulation System is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder." This is the clinical claim, but the document doesn't present a study proving the device meets this claim with performance metrics. It relies on the substantial equivalence to the predicate device, which presumably had clinical data supporting its efficacy for this indication.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance (from text)
    Software Verification & Validation"the subject device functions as intended"
    Electrical Safety (IEC 60601-1)"meets all the same acceptance criteria as the predicate and reference device"
    EMC (IEC 60601-1-2)"meets all the same acceptance criteria as the predicate and reference device"
    Bench Performance Testing"the subject device functions as intended and meets all the same acceptance criteria as the predicate and reference device"
    Clinical Efficacy for OCD Treatment (Implied from Indications for Use)The document does not present a new clinical study to establish this for the modified device, but rather relies on substantial equivalence to the predicate, which previously established this.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not provided in the document. The document describes "software verification and validation," "electrical safety," "EMC," and "bench performance testing." These are engineering and performance validation tests, not typically clinical studies with "test sets" of patient data in the sense of a machine learning model.
    • If a prior clinical study for the predicate device existed for OCD treatment, that information is not detailed here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable / Not provided. Since this document describes engineering and bench testing, rather than a clinical trial or an AI/diagnostic device study requiring expert ground truth establishment for a patient dataset, this information is not relevant to the described testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. See explanation for #3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This document describes a medical device (Transcranial Magnetic Stimulation System) for treatment, not an AI-assisted diagnostic or imaging device used by "human readers." Therefore, an MRMC study is not relevant to the described device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Stand-alone performance was indirectly evaluated through bench testing and compliance. The device itself is a "human-in-the-loop" device as it delivers treatment under medical supervision. The "standalone" performance here refers to the device's technical specifications and electrical/mechanical performance, which was assessed through the described tests (software, electrical safety, EMC, bench performance). The document states: "In all instances, the subject device functions as intended and meets all the same acceptance criteria as the predicate and reference device." This implies standalone technical performance testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the technical and safety testing described (software, electrical safety, EMC, bench performance):
      • Ground Truth: Engineering specifications, regulatory standards (e.g., IEC 60601-1, IEC 60601-1-2), and comparison to the performance of the predicate and reference devices. The "ground truth" is that the device should operate within specified technical parameters and safety limits.
    • For the clinical efficacy of the device for OCD treatment (which is its intended use):
      • The document implies that the predicate device (DEN170078) already established clinical efficacy for OCD. This specific document for K183303 relies on "substantial equivalence" and does not describe a new clinical trial for efficacy. If such a trial existed for the predicate, the ground truth would likely have been clinical outcome measures for OCD (e.g., Yale-Brown Obsessive Compulsive Scale (YBOCS) scores or similar psychiatric outcome scales).

    8. The sample size for the training set

    • Not applicable / Not provided. The Brainsway Deep TMS System is a Transcranial Magnetic Stimulation device, not an AI/machine learning algorithm that requires a "training set" of data in the common sense. Its "training" is its engineering design and manufacturing.

    9. How the ground truth for the training set was established

    • Not applicable / Not provided. As explained in #8, there is no "training set" in the context of an AI model for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173540

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-05-03

    (169 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brainsway Deep TMS System is indicated for the treatment of depressive episodes in adult patients suffering from Major Depressive Disorder who failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode.

    Device Description

    The Brainsway Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure. The FDA cleared Brainsway Deep TMS System is composed of the following main components: 1. An Electromagnetic Coil 2. A TMS Neurostimulator 3. A Cooling System 4. A Positioning Device 5. A Cart

    AI/ML Overview

    This document is a 510(k) premarket notification for the Brainsway Deep TMS System. It focuses on demonstrating substantial equivalence to a previously cleared device (K122288), also from Brainsway Ltd., rather than presenting a de novo study with explicit acceptance criteria for a novel device performance claim. Therefore, the information requested in the prompt, which is typically found in clinical studies establishing the efficacy and safety of a new device, is largely not applicable or not provided in the context of this 510(k) submission.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria for diagnostic/AI performance, nor does it report specific device performance metrics in that context. The "acceptance criteria" here are implicitly related to demonstrating substantial equivalence for modifications to an existing device.

    From the text, the performance aspects evaluated are primarily related to safety and technical specifications:

    • Output Waveform
    • Electrical Field Spatial Distribution
    • Magnetic Field Strength Gradient Testing
    • Software validation testing (compliance with FDA guidelines and IEC 62304)
    • Compliance with electrical and mechanical safety, electromagnetic compatibility, and software validation standards (IEC 60601-1, IEC 60601-1-2, IEC 62304).

    The reported performance is that these characteristics are substantially equivalent to the predicate device and that the device complies with the mentioned standards.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not applicable as this submission is for modifications to an existing device, and the testing described is non-clinical (bench testing) and software validation. No new clinical test set (i.e. patients for diagnostic performance evaluation) was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. There was no clinical test set requiring expert-established ground truth for performance evaluation of a new diagnostic claim. The "ground truth" here is the established safety and performance of the predicate device and relevant international standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. No clinical test set requiring adjudication was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A MRMC comparative effectiveness study was not done. The device is not an AI-assisted diagnostic tool but a therapeutic device (Repetitive Transcranial Magnetic Stimulation System).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a therapeutic system, not an algorithm for standalone diagnostic performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this 510(k) submission, the "ground truth" for the device's technical performance and safety is its conformance with recognized international standards (IEC 60601-1, IEC 60601-1-2, IEC 62304) and demonstration of substantial equivalence to a previously cleared predicate device in terms of:

    • Output Waveform
    • Electrical Field Spatial Distribution
    • Magnetic Field Strength Gradient
    • Software validation

    Safety and effectiveness are established by showing that modifications do not introduce new hazards and that the device operates within established parameters, which are deemed safe and effective based on the predicate device's clearance.

    8. The sample size for the training set

    This information is not applicable. This is not an AI/ML device requiring a training set in the conventional sense. The "training" for the device's development and validation would refer to engineering design, testing iterations, and software development, not a data-driven machine learning training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reasons as point 8.