Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K183593

    Validate with FDA (Live)

    Date Cleared
    2019-04-18

    (118 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

    Device Description

    The Lung Vision System (K163622) is designed to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

    The Lung Vision System is designed to assist the physician in guiding endobronchial tools towards the target area of interest inside the patient lungs. Prior to the endoscopic procedure the system allows planning the target location and the path to the target area on the CT scan. During the endoscopic procedure the system overlays planned data over fluoroscopic images to support endobronchial tool navigation towards the area of interest. The system does not include the Fluoroscope, Bronchoscope or the external monitor. Lung Vision system includes a main unit and a tablet vs. the previous PC based hardware platform. Image processing algorithms are executed on the main unit and the tablet is used as a primary method of interacting with the system. The Tablet is for planning but is not for diagnostic purposes. Both can perform the following functions: segment previously acquired DICOM 3D CT image data, register DICOM 3D CT image data with live fluoroscopic X-ray image, overlay the segmented 3D CT dataset over a live fluoroscopic X-ray image of the same anatomy, obtained on a Fluoroscopic system.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text.

    Important Note: The provided text is a 510(k) Summary, which is a premarket notification for new medical devices. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical trial results for efficacy or a comprehensive standalone AI performance study. Therefore, some information typically found in a clinical study report (like detailed MRMC results, specific effect sizes, or a large, prospectively collected, expert-adjudicated dataset for de novo AI training) is not present here. The performance testing described is primarily focused on demonstrating that the device functions as intended and safely, matching the predicate.

    Acceptance Criteria and Reported Device Performance

    The provided text doesn't explicitly state quantitative acceptance criteria in a table format with specific performance metrics (e.g., "accuracy > X%", "sensitivity > Y%"). Instead, the acceptance criteria are implicitly defined by demonstrating substantial equivalence to a predicate device and successful completion of bench testing to ensure the device meets its own "requirement specifications" and "hazards mitigations."

    The performance is reported in terms of functional equivalence and safety:

    Acceptance Criteria CategoryReported Device Performance (Summary of Findings)
    Functional EquivalenceThe Lung Vision System performs the same functions as the predicate device (LungVision - K163622).
    It includes new features (Virtual Bronchoscopy, C-Arm based Tomography, Multi-view set-up, Real-time compensation, 3D Guidance, Tablet use) which are similar in functionality or technological characteristics to either the predicate or reference devices (Covidien - superDimension™ Navigation System V7.2 - K173244).
    Performance & Accuracy"met all requirements specifications"
    "was found to be equivalent in comparison to the predicate"
    Accuracy testing in "deformable tissue" was performed using pig lungs. (No specific quantitative metric provided)
    SafetyComplies with ANSI/AAMI/ES 60601-1:2005(2012) and IEC 60601-1-2:2014 standards (electrical and electromagnetic compatibility).
    No patient-contacting parts (no biocompatibility testing needed).
    Clinical EfficacyNot directly assessed; presumed equivalent based on substantial equivalence to predicate.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: Not explicitly stated as a separate "test set" in the context of typical AI model validation. The "bench tests" included "verification testing of the requirements, testing of hazards mitigations and performance testing of the system." This suggests internal testing against pre-defined specifications rather than a distinct, large, clinical test dataset.
      • Data Provenance: Not specified. The document mentions "pig lungs" were used for some accuracy testing in "deformable tissue." This implies animal (in vitro/ex vivo) testing. There is no mention of human clinical data for the performance testing cited.
      • Retrospective or Prospective: Not explicitly stated for performance testing. Given the "bench tests" and no clinical testing, it's likely internal, controlled testing rather than a large-scale retrospective or prospective patient study.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

      • Not applicable/Not stated. The "ground truth" for the performance testing described here (bench tests, pig lungs) relates to the system's technical specifications and physical performance (e.g., registration accuracy on deformable tissue) rather than diagnostic interpretations by human experts.
    3. Adjudication Method for the Test Set:

      • Not applicable/Not stated. Since the "test set" described is for bench performance, there's no mention of human expert adjudication.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No. The document explicitly states: "Clinical: No clinical testing was performed." Therefore, no MRMC study comparing human readers with and without AI assistance was conducted or reported.
    5. Standalone (Algorithm Only Without Human-in-the-Loop) Performance Study:

      • A standalone performance study for the algorithm itself in terms of diagnostic accuracy (e.g., detecting lesions) was not the focus or purpose of this 510(k) submission. The device is a "system, image processing, radiological" intended to support catheter/device navigation by overlaying pre-acquired CT data onto live X-ray. Its performance is evaluated on its ability to accurately segment, register, and overlay images, and perform real-time compensation, not on its ability to autonomously diagnose.
      • The "Performance Testing" section states, "We have performed bench tests and found that the Lung Vision met all requirements specifications and was found to be equivalent in comparison to the predicate. Testing includes verification testing of the requirements, testing of hazards mitigations) and performance testing of the system." This implies internal measurements of system performance (e.g., accuracy of registration or compensation) but not a standalone diagnostic outcome.
    6. Type of Ground Truth Used:

      • For the bench testing, the ground truth would have been the engineering specifications of the device's functions (e.g., the expected accuracy of C-Arm based Tomography or real-time compensation).
      • For testing in "deformable tissue (pig lungs)," the ground truth would likely be physical measurements or imaging references to assess the system's ability to maintain accuracy in a dynamic environment, rather than a clinical diagnosis or pathology.
    7. Sample Size for the Training Set:

      • Not stated. The document doesn't discuss the details of the training data used for any algorithms (e.g., segmentation, registration, real-time compensation). As this is primarily a system modification and not a de novo AI diagnostic device, the specifics of algorithm training or AI model validation datasets are not a required part of this 510(k) summary. The "software upgrades" are described as "an algorithm improvement" or "a standard technology," implying refinements rather than entirely new AI models requiring extensive separate training data disclosure for this type of submission.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable/Not stated. As the training set details are not provided, the method for establishing its ground truth is also not mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172955

    Validate with FDA (Live)

    Device Name
    LungVision Tool
    Date Cleared
    2018-04-26

    (212 days)

    Product Code
    Regulation Number
    874.4680
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lung Vision Tool is an instrument designed as a working channel intended to be used with standard bronchoscopes, endotherapy accessories and ultrasound probe to guide the endotherapy accessories or ultrasound probe to the target area, specifically within the respiratory system.

    Device Description

    Lung Vision Tool is an instrument designed as a working channel intended to be used with standard bronchoscopes, endo-therapy accessories and ultrasound probe to guide the endo-therapy accessories or ultrasound probe to the target area, specifically within the respiratory system.

    LungVision tool is designed to be used by pulmonologists or thoracic surgeons during pulmonary procedures.

    The LungVision Tool is designed to be connected to a standard Bronchoscope Instrument Port.

    The Lung Vision Bronchoscope Adaptor is used to connect the LungVision Handle to the Bronchoscope instrument port / working channel entrance, and to allow the physician to release the tool from the bronchoscope. The current adaptor is for the Olympus Bronchoscope Model BF-1T160.

    More types of Bronchoscope Adaptors for other bronchoscopy brands and models.

    Lung Vision Tool Sheath can accommodate endo-therapy accessories and ultrasound probes with an outer diameter up to 1.9 mm.

    AI/ML Overview

    The provided text describes the LungVision Tool, a medical device intended to be used with standard bronchoscopes, endotherapy accessories, and ultrasound probes to guide them to a target area within the respiratory system. The submission demonstrates substantial equivalence to predicate devices K060243 (Olympus Guide Sheath) and K151315 (Boston Scientific Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle).

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a tabulated format with numerical targets for each test, but it does list the tests performed and implies that the device "met all requirements specifications." The comparison table on pages 4 and 5 highlights the features and specifications for the predicate, reference, and subject device.

    Feature / Test CategoryPredicate K060243 (Olympus Guide Sheath)Reference K151315 (Boston EPEUTAN)Subject Device K172955 (LungVision Tool)Reported Performance (LungVision Tool)Acceptance Criteria (Implied)
    Intended UseGuide endo-therapy accessories/ultrasound probeGuide endo-therapy accessories/ultrasound probeGuide endo-therapy accessories/ultrasound probeSubstantially equivalent to predicateTo be substantially equivalent to predicate
    Anatomical locationRespiratory organsRespiratory organsRespiratory organsSubstantially equivalent to predicateTo be substantially equivalent to predicate
    ComponentsHandle-like, Stopper, SheathHandle, Adaptor/Connector, Sheath, Wire/StyletHandle, Adaptor/Connector, Sheath, WireSimilar components to predicate/referenceTo be similar to predicate/reference
    Works with endo-therapy instrumentsYesYesYesYesYes
    ClassificationBronchoscope (flexible or rigid) and accessories. 21CFR 874.4680, Procut Code: EOQBronchoscope (flexible or rigid) and accessories. 21CFR 874.4680, Procut Code: EOQBronchoscope (flexible or rigid) and accessories. 21CFR 874.4680, Procut Code: EOQSubstantially equivalent to predicateTo be substantially equivalent to predicate
    Target anatomyRespiratory OrgansRespiratory OrgansRespiratory OrgansSubstantially equivalent to predicateTo be substantially equivalent to predicate
    Anatomy accessBronchial airwaysBronchial airwaysBronchial airwaysSubstantially equivalent to predicateTo be substantially equivalent to predicate
    Patient PopulationNot specified (undergoing endoscopic procedures)Not specified (undergoing endoscopic procedures)Patients undergoing endoscopic proceduresSubstantially equivalent to predicateTo be substantially equivalent to predicate
    Environment of useWhere endoscopic procedures performedWhere endoscopic procedures performedWhere endoscopic procedures performedSubstantially equivalent to predicateTo be substantially equivalent to predicate
    Single UseYesYesYesYesYes
    SterileYesYesYesYesYes
    Mechanism of actionManual attachmentManual attachmentManual attachmentSubstantially equivalent to predicateTo be substantially equivalent to predicate
    Number of device passes during a procedureMultiple passesMultiple passesMultiple passesSubstantially equivalent to predicateTo be substantially equivalent to predicate
    X-ray detectionYesNoYesYes (similar to predicate)Yes
    Sheath Max Outer Diameter2.7 mmN/A2.72 mm2.72 mm (comparable to predicate)< 2.8 mm (consistent with recommended channel size)
    Sheath Min Inner Diameter2.1 mmN/A2.08 mm2.08 mm (comparable to predicate)> 1.9 mm (to accommodate accessories)
    Working Length900 mmN/A1000 mm1000 mm (comparable to predicate)Sufficient for intended use
    BiocompatibilitySurface Contact Skin and Breached Limited duration (<24 hours)Surface Contact Skin and Breached Limited duration (<24 hours)Surface Contact Skin and Breached Limited duration (<24 hours)Non-cytotoxic, non-sensitizers, non-irritantsTo meet ISO 10993 requirements
    Bench TestingNot listedPassability, Tensile strength, Durability, Removal Force, RotationBronchoscope Compatibility, Simulated Use, Fluoroscopy, Tensile Tests, Repeatability Use, Sheath Insertion/Withdrawal, Twisting, Drop, Accelerated Aging, Package integrity, Package StabilityMet all requirements specificationsFunctional and safety requirements comparable to predicate
    SterilizationSterile (Yes)Sterile (Yes)Sterile (Yes)Met SAL of at least 10-6To achieve SAL of at least 10-6

    2. Sample Sizes used for the Test Set and Data Provenance:

    The document describes non-clinical/bench testing, biocompatibility testing, and sterilization validation. It does not refer to a "test set" in the context of clinical or performance data from human subjects.

    • Bench Testing: The sample sizes for each specific bench test (e.g., Bronchoscope Compatibility, Tensile Test) are not explicitly mentioned. The testing would involve multiple units of the LungVision Tool.
    • Biocompatibility: The specific number of samples for cytotoxicity, sensitization, irritation, and acute systemic toxicity tests are not provided, but these tests (ISO 10993) involve standardized sample sizes for in-vitro and animal testing.
    • Sterilization: "Single Lot Release validation" was performed, indicating testing on one lot of devices.
    • Data Provenance: The data is generated from non-clinical (bench) studies likely conducted in laboratories within the submitting company's or contractor's facilities. There is no mention of country of origin for the non-clinical data, nor whether it is retrospective or prospective, though bench testing is inherently prospective.

    3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as there was no test set involving human subjects or expert assessment to establish ground truth. The device is a medical instrument (tool), and its performance was evaluated through non-clinical testing against engineering specifications and comparison to predicate devices, not against a "ground truth" derived from human experts.

    4. Adjudication method for the test set:

    This information is not applicable as there was no test set involving human experts requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "No clinical testing was performed."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This device is a physical medical tool, not an algorithm or software. Therefore, the concept of "standalone (algorithm only)" performance is not applicable. Its performance is assessed through its physical and functional characteristics in bench testing.

    7. The type of ground truth used:

    The "ground truth" for this device's performance is established through:

    • Engineering specifications and design requirements: For bench testing (e.g., tensile strength, dimensions, compatibility).
    • International standards (e.g., ISO 10993 for biocompatibility) and regulatory requirements: For safety evaluations.
    • Comparison to legally marketed predicate devices: For demonstrating substantial equivalence in intended use, technology, and performance.

    8. The sample size for the training set:

    This information is not applicable. The device is a physical tool, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163622

    Validate with FDA (Live)

    Device Name
    Lung Vision
    Date Cleared
    2017-05-11

    (140 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lung Vision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

    Device Description

    The LungVision System is designed enabling users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

    The Lung Vision System is designed to assist the physician in guiding endobronchial tools towards the target area of interest inside the patient lungs. Prior to the endoscopic procedure the system allows planning the target location and the path to the target area on the CT scan. During the endoscopic procedure the system overlays planned data over fluoroscopic images to support endobronchial tool navigation towards the area of interest. The system does not include the Fluoroscope, Bronchoscope or the external monitor.

    Lung Vision image processing algorithms are executed on a PC based hardware platform, which can perform the following functions:

    • segment previously acquired DICOM 3D CT image data, .
    • register DICOM 3D CT image data with live fluoroscopic X-ray image
    • overlay the segmented 3D CT dataset over a live fluoroscopic X-ray image of the same anatomy, ● obtained on a Fluoroscopic system.
    AI/ML Overview

    The provided document describes the LungVision System, a device intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy to support catheter/device navigation during pulmonary procedures.

    The 510(k) summary (pages 3-6) focuses on demonstrating substantial equivalence to a predicate device (Philips EP-Navigator K062650) rather than providing detailed acceptance criteria and a study proving the device meets those specific criteria. The document states "No clinical testing was performed." and "No animal testing was performed."

    Therefore, based solely on the provided text, a comprehensive answer to your request is not possible. However, I can extract the available information regarding performance and testing:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it describes general claims of equivalence to a predicate device. The "Performance and Specifications" section states: "The performance and specifications demonstrate that the Lung Vision and predicate devices perform the same functions using the same technologies thus can be found substantially equivalent."

    The "Nonclinical / Bench" section states: "We have performed bench tests and found that the Body Vision met all requirements specifications and was found to be equivalent in comparison to the predicate. Testing includes verification testing of the requirements, testing of hazards mitigations, and performance testing of the system."

    Without the specific "requirements specifications" or direct comparative performance data, it's impossible to generate a table of acceptance criteria and reported device performance.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "bench tests" and "testing has also been performed with pig lungs to test accuracy in deformable tissue." However, it does not specify the sample size for these tests, the data provenance (e.g., country of origin), or whether the tests were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not describe the establishment of a ground truth for a test set using experts. Since no clinical or animal testing was performed to evaluate the diagnostic or navigational accuracy against defined ground truth, this information is not available.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical or animal studies with a test set requiring expert adjudication for ground truth were described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The document explicitly states "No clinical testing was performed."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document mentions "bench tests" and "verification testing of the requirements, testing of hazards mitigations, and performance testing of the system," as well as "testing has also been performed with pig lungs to test accuracy in deformable tissue." These appear to be standalone tests of the device's technical performance attributes. However, specific metrics of "algorithm only" performance (e.g., accuracy of segmentation or registration) are not provided, nor is the "performance" explicitly defined in terms of measurable outcomes. The application is for a navigation aid where a human is in the loop.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the "testing has also been performed with pig lungs to test accuracy in deformable tissue," the type of ground truth is not specified. Given it was in vitro testing on pig lungs, it would likely involve physical measurements or anatomical references, but this is not detailed. For other bench tests, "ground truth" would refer to the expected functional behavior or output as per the requirements specifications, rather than a clinical ground truth like pathology.

    8. The sample size for the training set

    The document describes the device as a "PC based software application" that utilizes algorithms to process existing 3D CT datasets and live fluoroscopic images. This suggests the system employs algorithms that might have been developed or trained. However, the document does not provide any information regarding a training set, its sample size, or how it was used in the development of the device's algorithms.

    9. How the ground truth for the training set was established

    Not applicable, as no information about a training set is provided.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1