510(k) Data Aggregation

    K Number
    K210382

    Validate with FDA (Live)

    Date Cleared
    2021-07-23

    (164 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blustone Synergy Diamond SA Cervical System are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous bone graft when used as an adjunct to fusion in patients with cervical degenerative disc disease (DDD) at one level or two contiguous levels from C2 to T1. Degenerative Disc Disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six weeks of non-operative treatment. The Blustone Synergy Diamond SA Cervical System may be used with additional supplemental fixation.

    Device Description

    The BluStone Synergy Diamond Stand Alone (SA) Cervical System consists of the Diamond cervical interbody plate and screws to be used in conjunction with the Blustone Synergy Interbody Fusion SLATE cervical interbody fusion devices to form the Diamond Stand Alone Cervical System. The Diamond Stand Alone cervical system is designed to be used with allograft and/or autograft. Use of the Diamond SA Cervical System is intended to expedite the Anterior Cervical Device instrumentation procedure, while minimizing tissue disruption through a minimally invasive approach. The Diamond plate includes anterior nail spikes to resist rotation and two holes for insertion of the included bone screws as well as an integrated locking plate to resist bone screw backout. The Diamond cervical interbody plate and screws are manufactured from titanium alloy. Previously cleared SLATE cervical cages to be used with the Diamond plate and screws are manufactured from PEEK and include tantalum markers. All implant components are available in various sizes to accommodate varying patient anatomy.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study that proves a device meets such criteria in the context of AI/ML performance. The document is an FDA 510(k) clearance letter for a medical device called the "Blustone Synergy Diamond SA Cervical System," which is an intervertebral body fusion device.

    The "Performance Testing" section mentions bench performance testing conducted on the device, including:

    • Static and dynamic axial compression
    • Static and dynamic axial compression shear
    • Static and dynamic torsion per ASTM F2077-18
    • Subsidence per ASTM F2267-04 (2018)
    • Expulsion

    It states that "Testing shows that the subject Blustone Synergy Diamond SA Cervical System performs equivalent to or better than the 5th percentile of FDA benchmark values." This implies there are FDA benchmark values that serve as acceptance criteria for mechanical performance.

    However, the document does not discuss AI/ML device performance, acceptance criteria, or studies related to AI/ML device evaluation. It primarily focuses on the substantial equivalence of a physical surgical implant based on mechanical engineering tests.

    Therefore, I cannot provide the requested information regarding AI/ML device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203520

    Validate with FDA (Live)

    Date Cleared
    2021-04-07

    (127 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blustone Synergy cervical (Slate Lavaflow) implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of carcellous bone graft when used as an adjunct to fusion in patients with cervical disc disease (DDD) at one level or two contiguous levels from C2 to T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six weeks of non-operative treatment. The Blustone Synergy cervical implants are also to be used with supplemental fixation.

    The Blustone Synergy lumbar (Lavaflow) implants are intervertebral body fusion devices indicated for use with autograff and/or allogenic bone graft comprised of cancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of non- operative therapy. Additionally, the Blustone Synergy lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scolosis. The Blustone Synergy lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine.

    Device Description

    The Blustone Synergy Interbody Fusion System is composed of cervical and lumbar interbody fusion devices. The BluStone Synergy Slate Lavaflow System is a Titanium Plasma Coated cervical interbody fusion system comprised of parallel and 6° lordotic cages in two footprints with varying heights designed to accommodate patient anatomy, and may be implanted as a single device via an anterior approach. The Blustone Synergy Lumbar Interbody Lavaflow System is a Titanium Plasma Coated lumbar interbody fusion system comprised of various device configurations based on surgical approach and patient anatomy, and may be implanted via one of the following approaches: bi-laterally in pairs via a posterior (PLIF) approach; as a single device via a transverse (T-PLIF) approach; as a single device via a transforaminal (TLIF) approach; or as a single device via a lateral (LLIF) approach.

    All Blustone Synergy Interbody Fusion System implant components are made of polyether-ether-ketone (Zeniva ZA-500 PEEK) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device.

    This Traditional 510(k) submission seeks to expand the sizes offered for the MAGMA LLIF cages as well as add commercially pure (CP) titanium plasma coating per ASTM F1580 to the subject new MAGMA cages as well as all previously cleared cervical and lumber interbody fusion devices. Plasma-coated implant options will be denoted as the LAVAFLOW subfamily. Finally, this submission seeks to offer all implants as sterile devices via sterilization by ethylene oxide (EO) in addition to the previously cleared non-sterile, non-coated options.

    AI/ML Overview

    The provided text is a 510(k) summary for the Blustone Synergy Interbody Fusion System. It describes the device, its indications for use, and a comparison to predicate devices to establish substantial equivalence. However, it does not contain information about specific acceptance criteria or a detailed study proving the device meets those criteria with statistical data, sample sizes, ground truth establishment, or expert involvement as requested in the prompt.

    The document primarily focuses on demonstrating substantial equivalence to previously cleared devices through:

    • Comparison of Indications for Use, Materials, and Geometry: Stating that these aspects are "all-inclusive of the subject device" compared to predicates.
    • Performance Testing: Mentioning that mechanical testing (expulsion, dynamic compression per ASTM F2077, and wear debris analysis per ASTM F1877) has been performed and "the results have shown them to be substantially equivalent to the predicate interbody devices." It also notes an "engineering rationale" for new MAGMA small IBDs.

    Therefore, I cannot fulfill your request for a table of acceptance criteria, reported device performance, sample sizes, data provenance, expert ground truth details, adjudication methods, MRMC study information, or standalone algorithm performance, as this information is not present in the provided text.

    The document concludes that "Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices," which is the central finding of a 510(k) submission. However, the specifics of these "test results" in terms of statistical metrics against predefined acceptance criteria are not detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172330

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-21

    (141 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blustone Synergy Silica System of implants is indicated for use to replace a vertebral body that has been resected or excised (i.e. partial or total vertebrectomy) due to tumor or trauma/fracture. The System is intended for use in the thoracolumbar spine (from T1 to L5) and is intended for use with supplemental fixation. These devices are designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. These devices are intended to be used with autograft or allograft bone.

    Device Description

    The Blustone Synergy Silica System will be offered in parallel and lordotic device configurations of various heights to accommodate patient anatomy. The System of devices may be implanted in the thoracolumbar spine via an anterior approach.

    The Blustone Synergy Silica System implant components are made of polyether ether ketone (Zeniva ZA-500 PEEK) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device. The System is implanted using a combination of device specific and universal class I instruments manufactured from stainless steel materials that conform to ASTM F899.

    AI/ML Overview

    The provided text is a 510(k) summary for the Blustone Synergy Silica System, which is a spinal intervertebral body fixation orthosis. This type of document is for medical device clearance, not for an AI/ML powered device. Therefore, no information is present regarding acceptance criteria, study design for AI evaluation, sample sizes for test/training sets, expert qualifications, or ground truth establishment relevant to an AI-powered medical device.

    The document discusses non-clinical (mechanical) testing of the physical implant device to demonstrate its substantial equivalence to predicate devices, focusing on:

    • Static and dynamic compression testing (ASTM F2077-14)
    • Static and dynamic torsion testing (ASTM F2077-14)
    • Subsidence testing (ASTM F2267-04(2011))
    • Expulsion testing (ASTM Draft Standard F-04.25.02.02)

    Since the request is specifically about a study proving an AI-powered device meets acceptance criteria, and this document pertains to a physical implant, I cannot extract the requested information from the provided text.