Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K233658

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-16

    (93 days)

    Product Code
    Regulation Number
    874.3880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biowy Tym Tube products are indicated for where chronic Eustachian tube dysfunction does not respond to conventional therapy.

    Device Description

    The tympanostomy tube is a silicone tube that is intended to provide ventilation and drainage to the middle ear.

    AI/ML Overview

    The provided text describes the Biowy Tym Tube, a tympanostomy tube, and its 510(k) summary for substantial equivalence to a predicate device. However, the document does not contain information about acceptance criteria or a study proving the device meets clinical acceptance criteria in the context of an AI/ML device.

    The provided information focuses on engineering and biocompatibility testing for a physical medical device, not a software or AI/ML product. Therefore, I cannot extract the requested details relevant to AI/ML device performance or clinical studies using AI models.

    To answer your request, here's an analysis based solely on the information provided, which is limited to the physical device's characteristics and safety testing:


    Based on the provided document, the "device" in question is the Biowy Tym Tube, a physical tympanostomy tube, not an AI/ML software device. The acceptance criteria and "study" described pertain to the physical properties, sterility, and biocompatibility of this physical device.

    Here's the information extracted from the document:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria / TestReported Device Performance
    Sterility: Appropriate sterilization and shelf-life validation.Met requirements (submission supported by appropriate validation).
    Pyrogenicity: Device labeled non-pyrogenic. Validated for LAL Endotoxin testing using the Kinetic-Chromogenic test method using three production lots.Met requirements (labeled non-pyrogenic, tested successfully).
    Bench Testing - Visual Inspection: No visual defects.Met requirements
    Bench Testing - Compression: Comparable to predicate.Met requirements
    Bench Testing - Tensile: Higher than that needed for placement and removal.Met requirements
    Biocompatibility: According to ISO 10993.Shown to be safe, non-toxic, and biocompatible.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set:
      • Pyrogenicity: "Three production lots of the test article identified as Biowy Tym Tube" were validated.
      • Bench Testing (Visual, Compression, Tensile): Not explicitly stated, but typically involves a sample size representative of production for such physical tests.
      • Biocompatibility: Not explicitly stated, but ISO 10993 testing requires specific sample sizes depending on the test.
    • Data Provenance: Not specified. This type of testing is typically performed by internal or contract labs and is generally considered part of the manufacturing quality control and design verification process, not clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not applicable and not provided in the document. The tests performed are engineering, chemical, and biological safety tests, not clinical evaluations requiring expert interpretation (like a radiologist establishing ground truth for image analysis).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable and not provided. Adjudication methods are typically used in clinical studies or expert reviews of complex data (e.g., medical images, clinical outcomes) to resolve discrepancies. The described tests are objective measurements against defined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and not provided. An MRMC study is relevant for evaluating the clinical performance of diagnostic or assistive AI/ML devices involving human readers. The Biowy Tym Tube is a physical medical device, not an AI/ML system.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and not provided. This concept applies to standalone AI/ML algorithms. The described device is a physical product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the tests mentioned are:

    • Sterility: Absence of microorganisms (verified through validated sterilization processes).
    • Pyrogenicity: Absence of pyrogens/endotoxins (quantified via LAL test).
    • Visual Inspection: Conformance to manufacturing specifications (e.g., no cracks, burrs, malformations).
    • Compression/Tensile: Objective physical measurements against defined engineering specifications and comparison to the predicate device.
    • Biocompatibility: Absence of adverse biological reactions (measured via ISO 10993 tests for cytotoxicity, sensitization, irritation, etc.).

    These are objective, measurable standards for a physical product, not clinical "ground truth" derived from patient data or expert interpretation.

    8. The sample size for the training set:

    This information is not applicable and not provided. A "training set" is relevant for AI/ML models. This document describes a physical medical device.

    9. How the ground truth for the training set was established:

    This information is not applicable and not provided. This concept is for AI/ML models.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192246

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-30

    (255 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biowy PICC catheter S Kit is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. The maximum pressure of power injectors used with the Biowy catheter may not exceed 300psi. The catheter is available in a kit configuration.

    Device Description

    The Biowy PICC catheters are opened-ended radiopaque polyurethane catheters. The catheters are sterile and for single use. They are available in 4F single lumen with 55cm usable length and 5F dual lumen with 60cm usable length. The catheters have a reverse taper design. Catheter shaft is marked with depth indicators, with "0" indicated to serve as a reference for the catheter taper point. Catheters are provided sterile in radiology and nursing configurations for single use. Colorants were added to the catheter shaft and hub to differentiate components. The extension leg, junction and clamp were printed with markings to identify the catheter and to include information to facilitate proper use of the device. The Catheter is provided in a kit with multiple components that includes a stylet, torque adapter and T-lock to facilitate catheter insertion and flushing which are subject to this submission. The stylet has a torque adapter at the end to facilitate handling and withdrawal of the stylet. The catheter and other components in the kit are unchanged from the predicate device.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for a medical device (Biowy PICC Catheter S Kit), not a study proving a device meets AI-related acceptance criteria. Therefore, most of the requested information regarding AI acceptance criteria and study design for AI performance evaluation is not present in the provided text.

    The document focuses on demonstrating substantial equivalence to a predicate device (Biowy PICC Catheter K173956) based on non-clinical bench testing for the physical catheter and its new accessories.

    Here's the information that can be extracted, and where the requested information is not applicable (N/A) because the document describes a traditional medical device clearance, not an AI/ML powered device:


    Device: Biowy PICC Catheter S Kit (K192246)
    Type of Device: Percutaneous, implanted, long-term intravascular catheter (PICC)


    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists "Tests" and "Results" which indicate whether the device "Passed" or "Met requirements" for various physical and material properties. These are not acceptance criteria related to AI performance metrics (e.g., sensitivity, specificity, AUC).

    TestStandardsAcceptance Criteria (Implied by "Passed/Met Requirements")Reported Device Performance
    BiocompatibilityISO 10993-1Meets ISO 10993-1 requirementsPassed / met requirements
    Sterility Assurance Level (SAL)ISO 11135-1:2014SAL of 10-610-6
    ParticulatesUSP <788>Meets USP <788> requirementsPassed / met requirements
    Stylet Inspections and Lock TestInternalMeets internal specificationsPassed / met requirements
    Catheter Burst test after stylet insertionISO 10555-1:2013Meets ISO 10555-1:2013 requirementsPassed / met requirements
    Stylet Adapter Tensile TestISO 10555-1:2013Meets ISO 10555-1:2013 requirementsPassed / met requirements
    Stylet-Adapter Separation ForceISO 10555-1:2013Meets ISO 10555-1:2013 requirementsPassed / met requirements
    Stylet performanceISO 10555-1:2013 / InternalMeets ISO 10555-1:2013 and internal specificationsPassed / met requirements
    Stylet - MaterialsASTM 138-13Meets ASTM 138-13 requirementsPassed

    Since this is not an AI-powered device, the following points are Not Applicable (N/A) to the information provided in the document:

    2. Sample size used for the test set and the data provenance: N/A (No AI test set described)

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (No AI ground truth establishment described)

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: N/A (No AI test set described)

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (Not an AI-assisted device)

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A (Not an AI algorithm)

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): N/A (Ground truth as understood for AI performance is not relevant here. "Ground truth" for this device's performance is based on physical property testing and material standards.)

    8. The sample size for the training set: N/A (No AI training set described)

    9. How the ground truth for the training set was established: N/A (No AI training set described)


    Summary for this document: The provided text is a 510(k) premarket notification for a physical medical device. It demonstrates substantial equivalence through non-clinical bench testing of the device's physical properties and materials against established medical device standards. It does not involve artificial intelligence, machine learning, or associated performance metrics and study designs that would typically address the questions posed about AI acceptance criteria and studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173956

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-05-21

    (144 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biowy PICC catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion or therapy, use 4 French or larger catheter. The maximum recommended infusion rate is 5ml/sec for power injection of contrast media. The maximum pressure of power injectors used with the Biowy catheter may not exceed 300psi.

    Device Description

    The Biowy PICC catheters are opened-ended radiopaque polyurethane catheters are sterile and for single use. They are available in 4F single lumen with 55cm usable length and 5F dual lumen with 60cm usable length. The catheters have a reverse taper design. Catheter shaft is marked with depth indicators, with "0" indicated to serve as a reference for the catheter taper point. Catheters are provided sterile in radiology and nursing configurations for single use. Colorants were added to the catheter shaft and hub to differentiate components. The extension leg, junction and clamp were printed with markings to identify the catheter and to include information to facilitate proper use of the device.

    AI/ML Overview

    The provided document is a 510(k) summary for the Biowy PICC Catheter, which is a medical device and not an AI/ML software device. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, standalone algorithm performance, number of experts for ground truth, adjudication methods, or training set details are not applicable in this context. The acceptance criteria and performance are based on physical and biological testing of the catheter.

    Here's the information pulled from the document regarding the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    TestStandards and Acceptance CriteriaReported Device Performance
    Critical DimensionsISO 9626:2015 (Met requirements)Passed and met requirement
    Priming VolumeInternal (Met requirements)Passed and met requirement
    Tensile ForceISO 10555-1:2014 (Met requirements)Passed and met requirement
    Elongation and StiffnessInternal (Met requirements)Passed and met requirement
    Flexural FatigueInternal (Met requirements)Passed and met requirement
    Aspiration Flow/Catheter collapseInternal (Met requirements)Passed and met requirement
    Air Leakage during AspirationISO 10555-1:2014 (Met requirements)Passed and met requirement
    Gravity FlowISO 10555-1:2014 (Met requirements)Passed and met requirement
    Pump FlowISO 10555-1:2014 (Met requirements)Passed and met requirement
    Power Injection and Leak TestingISO 10555-1:2014 (Met requirements)Passed and met requirement
    Burst PressureISO 10555-1:2014 (Met requirements)Passed and met requirement
    Power Injection/Burst PressureISO 10555-1:2014 (Met requirements)Passed and met requirement
    Power Injection (10x)/Burst PressureISO 10555-1:2014 (Met requirements)Passed and met requirement
    Printing MarkingISO 10555-3:2013 (Met requirements)Passed and met requirement
    6% Luer TaperISO 594-1:1986 (Met requirements)Passed and met requirement
    LuerISO 594-1:1986, ISO 594-2:1998 (Met requirements)Passed and met requirement
    Luer Color OrientationInternal (Met requirements)Passed and met requirement
    RadiopacityASTM F640-12 (Detectible)Passed and met requirement
    CytotoxicityISO 10993-5Non-cytotoxic
    SensitizationISO 10993-10Non-sensitizer
    IrritationISO 10993-10Non-irritant
    Acute Systemic InjectionISO 10993-11Non-toxic
    Material Mediated PyrogenUSP <151>Non-pyrogenic
    LAL (Bacterial Endotoxins)USP <85>, <161>Passed and met requirements
    Hemolysis (extract)ISO 10993-4Non-hemolytic
    Hemolysis (direct)ISO 10993-4Non-hemolytic
    Complement ActivationISO 10993-4 (Similar to predicate)Similar when compared to predicate
    Partial Thromboplastin Time (PTT)ISO 10993-4 (Minimal activator, same as predicate)Minimal activator (same as predicate)
    Dog ThrombogenicityISO 10993-4 (Similar to control)Similar to control
    ImplantationISO 10993-6Non-irritant
    Subacute/sub Chronic Toxicity, Genotoxicity, Chronic Toxicity, and CarcinogenicityISO 10993-17, ISO 10993-18Passed and met requirement
    EtO sterilizationISO 11135: 2014 (Met 10-6 SAL requirement)Met 10-6 SAL requirement
    Clinical EvaluationInternalPassed
    Accelerated Aging (1.5 year, 3.0 year)ISO 11607-1:2006, ASTM F88/F88M-15, ISO 10555-1:2014 (Met requirements)Passed and met requirement

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for each specific test. The data provenance is not mentioned, and tests appear to be laboratory-based design verification rather than clinical data from human subjects. The study is a non-clinical performance test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as this is a medical device clearance based on non-clinical performance testing and biocompatibility, not an AI/ML algorithm requiring expert ground truth for interpretation of images or other data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This type of adjudication is for studies involving human interpretation or assessments, such as in clinical trials or AI/ML evaluations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is for AI/ML software devices, not a physical medical device like a PICC catheter.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is for AI/ML software devices, not a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on established engineering and biological standards (e.g., ISO, ASTM, USP). The device's performance is compared against these predetermined specifications and, in some cases for biocompatibility, against a legally marketed predicate device (5Fr DL PowerPICC® Catheter, K051672).

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1