Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K243788

    Validate with FDA (Live)

    Date Cleared
    2025-03-13

    (94 days)

    Product Code
    Regulation Number
    882.1320
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IceCaps (IceCap 2, IceCap 2 Small, IceCap Neonate (Sizes XS, S, & M)) are medical devices used as EEG electrodes. They are used by Healthcare Professionals on a patient in case of neurological disorders with a short or long-term EEG record (up to 72 hours).
    IceCap 2 shall be placed on patients weighing at least 10 kg (22.05 lbs) and having a head circumference above 43 cm (16.93 inches).
    IceCap Neonate shall be placed on the head of babies, newborns and premature babies.

    Device Description

    The IceCaps (IceCap 2, IceCap 2 Small, IceCap Neonate (Sizes XS, S, & M)) are medical devices used as EEG electrodes.
    The IceCaps are a single use cap which connects to the marketed EEG recorders using an IceAdapter or Touchproof adapter.
    The electrodes placement in IceCap Product line is done accordingly to the 10/20 system.
    The conductive tracks of the Flexible Printed Circuit are used to conduct EEG signals from the electrodes to the connectors.
    IceCap 2 shall be placed on patients weighing at least 10 kg (22.05 lbs) and having a head circumference above 43 cm (16.93 inches).
    IceCap Neonate (M,S,XS) shall be placed on the head of babies, newborns and premature babies.

    AI/ML Overview

    The IceCap product line does not perform comparative effectiveness studies with human readers or standalone algorithm performance studies. The device is a cutaneous electrode, and its evaluation focuses on safety and performance according to relevant standards, not on AI-driven diagnostic accuracy.

    Here's a breakdown of the acceptance criteria and supporting studies for the IceCap product line:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria TopicDescription/StandardDevice Performance (IceCap product line)
    Indications for UseFor use as EEG electrodes by Healthcare Professionals on patients with neurological disorders for short or long-term EEG record (up to 72 hours). Specific weight and head circumference ranges for IceCap 2 and IceCap Neonate.Meets stated indications for use, including up to 72 hours of use, matching predicate device (2).
    Safety StandardsCompliance with electrical safety and electromagnetic compatibility standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11).Conforms to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020, IEC 60601-1-2: 2014 + A1 (2020), IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020.
    BiocompatibilityMaterials in contact with the patient must be biocompatible (ISO 10993-1).Biocompatible and compliant with ISO 10993-1 Fifth edition 2018-08.
    Duration of UseUp to 72 hours of continuous use.Qualifies for 72 hours of use.
    Fit to Form and UsabilityAbility to accommodate different head sizes and proper installation.Qualified via fit to form test and usability test for installation.
    Signal Quality (Implied)The number of electrodes and material composition should not negatively impact the quality of EEG signal.Qualified via impedance test and general quality of signal.
    Material CompositionSpecific materials used for electrodes and adhesives.Materials listed (PET, Ag/AgCl inks, insulation inks, stiff PETG film, skin/silicone adhesive, graphical ink, protective polyolefin foam on acrylic adhesive) are biocompatible.
    Storage LifeExpected shelf life of the device.12 months. (Matches predicate 2, but shorter than predicate 1. This difference does not affect safety and effectiveness.)
    Single Use/SterilityNon-sterile, single-use device.Single use, non-sterile.
    Montage SystemConforms to the 10/20 System for electrode placement.10/20 System.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not explicitly state the sample size used for specific test sets (e.g., for fit-to-form, impedance, or usability tests). It mentions that the "clinical data were not necessary to determine substantial equivalence," indicating that animal or human subject testing for diagnostic or comparative effectiveness was not performed as a primary means of establishing substantial equivalence for this type of device.

    The document does not provide information on the country of origin of the data or whether the data was retrospective or prospective. The studies primarily involve non-clinical performance and safety testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    As this device is an EEG electrode, the primary "ground truth" for its performance is its ability to meet electrical and biocompatibility standards, and to effectively acquire EEG signals as confirmed by non-clinical tests. There is no mention of human experts being used to establish a ground truth for a diagnostic outcome, as the device itself does not provide diagnostic interpretations.

    4. Adjudication Method for the Test Set

    Not applicable. The evaluation performed is based on compliance with harmonized standards and engineering tests, not on human-based adjudication of diagnostic outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The IceCap product line is an EEG electrode, not an AI-powered diagnostic tool. Therefore, MRMC studies with human readers are not relevant to its clearance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a hardware component (EEG electrode) and does not involve a standalone algorithm for performance evaluation in a diagnostic context.

    7. The type of ground truth used

    The ground truth used for evaluating the IceCap product line is based on established engineering standards and regulatory requirements for medical devices, particularly for cutaneous electrodes. This includes:

    • Performance standards: e.g., electrical impedance, signal integrity (implied by "general quality of signal").
    • Safety standards: e.g., electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability for medical electrical equipment in the home healthcare environment (IEC 60601-1-11).
    • Biocompatibility standards: (ISO 10993-1) for materials in contact with the patient.
    • Functional tests: Fit-to-form, usability for installation.
      The "truth" is whether the device meets these specified, measurable criteria.

    8. The Sample Size for the Training Set

    Not applicable. This device is a hardware product (EEG electrode) and does not involve AI or machine learning algorithms that require a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This device is a hardware product (EEG electrode) and does not involve AI or machine learning algorithms that require a training set or ground truth for training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231366

    Validate with FDA (Live)

    Device Name
    Neuronaute Plus
    Manufacturer
    Date Cleared
    2023-11-09

    (182 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neuronaute Plus is a system intended to acquire, store, archive, and periodically transmit physiological signals from the brain using a full montage array to enable review at a physician's office, hospital, or other remote locations. It allows remote access by users via the BioSerenity Cloud.

    The Neuronaute Plus and its associated software are intended to assist in the diagnosis of neurological disorders. The Neuronaute Plus and its components do not provide any diagnostic conclusions or automated alerts of an adverse clinical event about a patient's clinical condition.

    The Neuronaute Plus is intended to be used by trained healthcare professionals, technicians or patients above 15 years old. In case of a patient below 15 years old, Neuronaute Plus is intended to be used by a care giver.

    Adequate training is recommended for proper use of the device.

    Neuronaute IceCap 2 electrodes are able to be used on patients weighing at least 10 kg and having a head circumference above 43 cm.

    Neuronaute is not intended to replace direct communication with healthcare providers. The system data should not be used alone, but should be used along with all other clinical data and exams to come to a diagnosis.

    The BioSerenity Cloud allows the display of signals in the diagnosis of physiological disorders through the data collected by recording.

    The BioSerenity Cloud should allow the analysis and filtering of data in order to aid doctors in diagnosing neurological disorders.

    Device Description

    The Neuronaute Plus is a non-invasive medical device which enables the acquisition, recording, storage and transmission of electrophysiological signals in order to analyze potential neurological disorders.

    The system is composed of the pieces of equipment listed below, and the Neuronaute Plus components are connected through Bluetooth and Wi-Fi.

    • -Neuronaute Plus (Core module): is a signal acquisition system: an electronical amplifier that records EEG, ECG, EMG, EOG and breathing signals via a connection to EEG caps which transmits the data to the Neuronaute mobile application and Cloud system.
    • -Neuronaute Plus IceCap extender : is a removable interface to connect the signal acquisition system to the IceCap and IceCap2 & 2 Small.
    • -Neuronaute Plus DB25 Extender : is a removable interface that can connect the signal acquisition system to any DB25-compatible EEG cap.
    • -Holding band : is a wearable textile band which allows the patient to place the Neuronaute Plus close around his chest over clothes. the use of the holding band is optional.
    • -Neuronaute IceCap: is a single use 21 electrode headset that connects to the Neuronaute Head Module via the IceAdapter, the DB25 cable and the Neuronaute BioAdapter, or Neuronaute Plus with its IceCap Extender. Recently, the Neuronaute Touchproof Adapter was developed to connect the IceCap electrode to Neuronaute Head module via a touchproof channel.
    • Neuronaute IceCap 2 & 2 Small: recently cleared version of the IceCap dedicated to adults and pediatric patients. It connects similarly to Neuronaute and Neuronaute Plus.
    • Neuronaute Mobile APP: is compatible with iOS systems. The mobile application enables the healthcare professionals to access and manage the prescribed recording sessions.
    • -Bioserenity CLOUD: is a web-based information system that receives the EEG signals from the recorder through a paired Wi-Fi connection. The cloud platform enables long-term storage and display of the recorded signals. The physician, who prescribes the use of this device, should monitor its use and confirm the proper functioning of signal recording through the Neuronaute Cloud.
    • -VEEG pack: it is composed of the Neuronaute N-DEO and Neuronaute N-WAY. The Neuronaute N- DEO camera enables the visualization of the patient during a recording. The IP camera provides HD resolution video sequence at 1080p, with advanced night mode and WDR technology. The Neuronaute N- WAY enables wireless connectivity (4G) for remote monitoring and data transmission to the cloud platform.
    AI/ML Overview

    The provided text does not contain the detailed information required to fill out a table of acceptance criteria and reported device performance for a diagnostic device, nor does it describe a study that proves the device meets specific performance criteria for diagnosis with quantitative metrics like sensitivity, specificity, or accuracy.

    The document is a 510(k) Premarket Notification for the Neuronaute Plus, which establishes its substantial equivalence to an existing predicate device (Neuronaute). This clearance pathway primarily focuses on demonstrating that the new device is as safe and effective as a legally marketed device, rather than proving its diagnostic performance against specific, quantitative acceptance criteria in a clinical study.

    Here's why the requested information is largely absent and what the document does provide:

    1. Acceptance Criteria and Reported Device Performance (Table):

      • The document lists several non-clinical tests (electrical safety, EMC, wireless safety, PEMS, cybersecurity, biocompatibility, usability, battery safety) that the Neuronaute Plus underwent.
      • For each of these, the "acceptance criterion" is implicitly "passed" or "fulfills the requirements of the standard."
      • The "reported device performance" is simply that "all tests... are in passed status" or "the product fulfills the requirements of the standard."
      • Crucially, there are no diagnostic performance metrics (e.g., sensitivity, specificity, AUC) or corresponding acceptance criteria because the device is an EEG acquisition and display system, not an AI diagnostic algorithm that provides conclusions or automated alerts. Its "performance" is primarily about its ability to accurately acquire and display physiological signals safely and effectively, comparable to its predicate.

      Therefore, a table of acceptance criteria and reported device performance related to diagnostic accuracy (as typically seen for AI/ML devices) cannot be constructed from this document. The performance testing described is focused on safety, electrical compatibility, and usability compared to a predicate device.

      Conceptual Table (based on safety/functional tests, not diagnostic accuracy):

      Area of TestAcceptance CriteriaReported Device Performance
      Electrical SafetyCompliance w/IEC 60601-1, -11, -26All tests passed; substantially equivalent to predicate.
      EMC TestsCompliance w/IEC 60601-1-2All tests passed; fulfills requirements of standard.
      Wireless SafetyRadio-emission safetyAll tests passed; substantially equivalent to predicate.
      PEMS (Firmware)Software safety & electrical perf.All tests passed.
      CybersecurityMeets FDA guidanceDesign process demonstrates safety; post-market plan.
      BiocompatibilityCompliance w/ISO 10993-1, FDA guid.No further testing required; classified as biocompatible.
      UsabilityNo critical/major use errorsValidation criteria met; no critical/major use errors.
      Battery SafetyCompliance w/IEC 62133, 60601-1All tests passed.
    2. Sample Size and Data Provenance for Test Set:

      • Diagnosis/AI Performance Test Set: Not applicable. The document explicitly states: "No clinical data was needed for demonstrating substantial equivalence of the Neuronaute Plus."
      • Usability Study Test Set:
        • Sample Size:
          • User group 1 (Healthcare professionals): n=9 (Physicians, nurses, technicians)
          • User group 2 (Patients): n=15 (aged 15-30, 31-50, >50 years)
        • Data Provenance: Not specified, but generally, human factors/usability studies are prospective and often conducted in a simulated environment by the manufacturer. Country of origin is not mentioned.
    3. Number of Experts and Qualifications for Ground Truth:

      • Not applicable for diagnostic ground truth as no clinical diagnostic performance study was conducted.
      • For the usability study, "trained healthcare professionals" were part of the user group, but their role was as study participants, not as ground truth experts for a diagnostic outcome.
    4. Adjudication Method for Test Set:

      • Not applicable for a diagnostic test set, as no such test was conducted.
      • For the usability study, the "validation criteria are met," and "no use errors leading to critical or major risks... occurred." This implies an assessment against predefined usability metrics, likely by study administrators, but no details on an adjudication panel for "ground truth" are provided.
    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

      • No. This study type is typically for assessing human reader performance with and without AI assistance for diagnostic interpretation. The Neuronaute Plus is an EEG acquisition and display system, not an AI interpretive algorithm. Its purpose is to assist in diagnosis by providing signals for a healthcare professional to review, not to provide diagnostic conclusions itself.
      • The Indications for Use clearly state: "The Neuronaute Plus and its components do not provide any diagnostic conclusions or automated alerts of an adverse clinical event about a patient's clinical condition."
    6. Standalone (Algorithm Only) Performance:

      • Not applicable. The Neuronaute Plus is a hardware system with associated software for signal acquisition, storage, and display. It does not contain a standalone diagnostic algorithm that produces a diagnostic output without human intervention.
    7. Type of Ground Truth Used:

      • For diagnostic purposes, none was used or required, as per the 510(k) clearance process for this type of device (EEG acquisition/display).
      • For non-clinical tests (electrical safety, EMC, etc.), the "ground truth" is adherence to established consensus standards (e.g., IEC standards).
    8. Sample Size for Training Set:

      • Not applicable. There's no mention of an AI/ML component that requires a training set for diagnostic classification or prediction. The device's software handles data acquisition, storage, transmission, and display, which are typically validated through verification and validation testing against functional requirements and standards, not through training data.
    9. How Ground Truth for Training Set was Established:

      • Not applicable, as no training set for an AI/ML diagnostic algorithm is mentioned.

    In summary: The provided document is a 510(k) clearance letter for an EEG acquisition and display system (Neuronaute Plus). Its "acceptance criteria" and "proof" primarily revolve around demonstrating safety, functional performance, and substantial equivalence to a predicate device through non-clinical testing and usability studies, rather than clinical diagnostic accuracy studies common for AI-powered diagnostic aids.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223644

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-12

    (37 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Neuronaute with IceCap 2 & IceCap 2 Small is a system intended to acquire, display, store, archive, and periodically transmit EEG signals from the brain using a full montage array to enable review at a physician's office, hospital, or other remote locations. It allows remote access by users via the Neuronaute N-CLOUD which receives EEG signals from Neuronaute Head Module which sends transmissions to the cloud.

    Neuronaute and its associated software are intended to assist in the diagnosis of neurological disorders. Neuronaute and its components do not provide any diagnostics conclusions or automated alerts of an adverse clinical event about a patient's clinical condition.

    The device is for use by trained medical professionals for patients under medical supervision. The device is intended for use on adults and pediatrics. Neuronaute is not intended to replace direct communication with healthcare providers. The system data should not be used alone but should be used along with all other clinical data and exams to come to a diagnosis.

    Device Description

    Neuronaute with IceCap 2 & IceCap 2 Small is a system intended to acquire, display, store, archive, and periodically transmit EEG signals from the brain using a full montage array. The system includes the Neuronaute Head Module, IceCap 2 & IceCap 2 Small electrodes, battery, Bioadapter with the IceAdapter, mobile app, and the Neuronaute N-CLOUD. The IceCap 2 and IceCap 2 Small are single-piece flexible printed circuit headsets with 21 EEG sensors (19 EEG channels, 1 reference and 1 ground), 4 skin adhesive areas and 2 connectors. The electrodes are made of Ag/AgCl sensors connected via conductive silver ink and insulated with dielectric ink. The system allows remote access by users via the Neuronaute N-CLOUD. An optional video system composed by N-way and N-Deo is available.

    AI/ML Overview

    The provided text describes the Neuronaute with IceCap 2 & IceCap 2 Small system, an electroencephalograph device. However, it does not contain a detailed study proving the device meets specific acceptance criteria for diagnostic performance. Instead, the document focuses on demonstrating substantial equivalence to a predicate device (Neuronaute, K202334) through a comparison of technical characteristics and non-clinical testing.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state quantitative acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, accuracy) or report such performance for the Neuronaute with IceCap 2 & IceCap 2 Small. The comparison table focuses on technical specifications and material changes relative to the predicate device.

    Acceptance Criterion (Implicitly compared to predicate)Reported Device Performance (Relative to Predicate)
    Ability to acquire, display, store, archive, and periodically transmit EEG signalsEquivalent: "The function of the system and the connection of the different elements remain the same."
    Target population (Adults and Pediatrics 5 years and older)Expanded: Subject device includes pediatric patients (5 years and older) due to two sizes of IceCap 2. Predicate was adults (18+).
    Electrode material (Ag/AgCl)Improved: Subject device uses Ag/AgCl for longer recording (up to 72h) compared to predicate's copper covered with silver (12.5h limitation due to copper migration).
    BiocompatibilityEquivalent: "IceCap 2 / 2 Small electrodes material has been assessed by consensus standard for biocompatibility testing, as previously done with its predicate Neuronaute."
    Electrical safetyEquivalent: Nonclinical tests demonstrated "safe" performance.
    Performance (General)Equivalent: Nonclinical tests demonstrated "as effective, and performs as well as the legally marketed predicate device."
    Number of electrodes (19 EEG, 1 reference, 1 ground)Equivalent: "Up to 21 electrodes" with the same distribution.
    Montage (10/20 System)Equivalent: Both use 10/20 System.
    Head size accommodationImproved: Subject device offers two sizes to accommodate head circumferences from 43 to 60 cm (including pediatric). Predicate had one size for "at least 90% of the population."
    Conductive electrolyte pasteEquivalent: Both use "Elefix" paste.
    EEG channels (Up to 24)Equivalent: Both have up to 24 channels.
    Data format (EDF)Equivalent: Both use EDF.
    Sampling Rate (250 or 500 Hz)Equivalent: Both support 250 or 500 Hz.
    Wireless Output (WiFi 2.4GHz, Bluetooth 2.4GHz)Equivalent: Same wireless specifications.
    Input dynamic range and differential offset voltage (±400mV)Equivalent: Same specifications.
    ADC Resolution (24 bits)Equivalent: Same specifications.
    ADC Common Mode Rejection Rate (CMRR) (> 105 dB)Equivalent: Same specifications.
    Input Impedance (> 1 Gohm)Equivalent: Same specifications.
    Input noise (< 6μVp-p over 0.1-50Hz)Equivalent: Same specifications.
    Electrode impedance checkEquivalent: Both perform electrode impedance check.
    Power source (Rechargeable LiPo 3.7V, 2.4 Ah battery)Equivalent: Same battery specifications.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a sample size for a test set related to diagnostic accuracy or clinical performance. It refers to "nonclinical tests" (biocompatibility, performance, electrical safety) but does not provide details on the test subjects or data used for these. Given that a clinical study was not required, there is no clinical "test set" and thus no data provenance for such a set.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. No clinical test set with ground truth established by experts is described in the document.

    4. Adjudication Method:

    Not applicable. As no clinical test set requiring ground truth establishment by experts is described, no adjudication method would have been used for that purpose.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. The document explicitly states: "N/A Clinical Testing was not required." Therefore, no MRMC study was conducted or reported.

    6. Standalone Performance Study:

    No. The document focuses on demonstrating substantial equivalence to a predicate device through technical comparisons and non-clinical testing. It does not describe a standalone performance study in terms of diagnostic accuracy or clinical outcomes. The device is intended to "assist in the diagnosis" and "should not be used alone."

    7. Type of Ground Truth Used:

    Not applicable for clinical performance. For the non-clinical tests (biocompatibility, performance, electrical safety), the ground truth would be based on established engineering and safety standards, but not clinical ground truth in the diagnostic sense.

    8. Sample Size for the Training Set:

    Not applicable. The document does not describe any machine learning or AI component that would require a "training set" in the context of diagnostic performance studies. The device is an EEG acquisition system and associated software, not an AI diagnostic tool.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as no training set for AI/ML is mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202334

    Validate with FDA (Live)

    Device Name
    Neuronaute
    Manufacturer
    Date Cleared
    2020-12-10

    (115 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Neuronaute is a system intended to acquire, display, store, archive, and periodically transmit EEG signals from the brain using a full montage array to enable review at a physician's office, hospital, or other remote locations. It allows remote access by users via the Neuronaute N-CLOUD which receives EEG signals from Neuronaute Head Module which sends transmissions to the cloud. Neuronaute and its associated software are intended to assist in the diagnosis of neurological disorders. Neuronaute and its components do not provide any diagnostics conclusions or automated alerts of an adverse clinical event about a patient's clinical condition.

    The device is for use by trained medical professionals for patients under medical supervision. The device is intended for use on adults (ages 18 and above). Neuronaute is not intended to replace direct communication with healthcare providers. The system data should not be used alone, but should be used along with all other clinical data and exams to come to a diagnosis.

    Device Description

    Neuronaute allows up to 24 channels EEG monitoring. It includes the following components: Neuronaute Head Module, Neuronaute High Capacity Battery Module, Neuronaute BioAdapter, Neuronaute Mobile App, Neuronaute N-CLOUD, Neuronaute N-DEO, Neuronaute N-WAY, Neuronaute IceCap.

    AI/ML Overview

    The provided text describes the Neuronaute device, an electroencephalograph (EEG) system, and its substantial equivalence to a predicate device (AE-120A EEG Head Set, K183529) for FDA clearance. However, it does not contain a specific study demonstrating that the device meets numerically defined acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, accuracy for a particular condition). Instead, the submission focuses on demonstrating substantial equivalence through non-clinical performance testing against recognized standards.

    Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable or cannot be extracted from this document, as a clinical performance study with such metrics was not submitted.

    Here's a breakdown of the information that can be extracted or inferred:

    1. Table of Acceptance Criteria and Reported Device Performance

    As a clinical performance study with specific diagnostic metrics (like sensitivity, specificity, or accuracy) was not submitted, there isn't a table of acceptance criteria for diagnostic performance against specific disease states. Instead, the acceptance is based on meeting technical and safety standards, and demonstrating signal quality, all of which are considered "performance" in this context.

    Performance CharacteristicAcceptance Criteria (Met by adherence to standards)Reported Device Performance (Demonstrated by testing)
    Safety & ElectricalAAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-26Met all listed standards requirements.
    Electromagnetic CompatibilityIEC 60601-1-2Met standard requirements.
    EEG Signal QualityIEC 60601-2-26 requirements for input noise; comparison to "EEG gold standards."Met IEC 60601-2-26 requirements. Signal quality demonstrated (implied to be acceptable through comparison).
    Software Verification & ValidationIEC 62304, FDA Guidance "General Principles for Software Validation"Met all listed standards and guidance requirements.
    UsabilityIEC 62366, FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices (2016)"Met all listed standards and guidance requirements. No new questions of safety or effectiveness identified.
    BiocompatibilityISO 10993-5, ISO 10993-10 (for IceCap and electrode gel paste)Conforms to ISO 10993-5 and ISO 10993-10.
    Input Dynamic Range & Differential Offset VoltageIEC 60601-2-26 requirements±400mV (device); conforms to IEC 60601-2-26.
    ADC ResolutionNot explicitly defined as an acceptance criterion for comparison, but described as "improved"24 bits
    ADC Common Mode Rejection Rate (CMRR)IEC 60601-2-26 requirements> 105 dB (device); conforms to IEC 60601-2-26.
    Input ImpedanceIEC 60601-2-26 requirements> 1 Gohm (device); conforms to IEC 60601-2-26.
    Input NoiseIEC 60601-2-26 requirements< 6µVp-p over 0.1-50Hz (device); conforms to IEC 60601-2-26.

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set): Not applicable. No clinical performance testing with a patient test set was submitted for diagnostic accuracy. The testing performed was non-clinical (bench, software V&V, usability).
    • Data Provenance: Not applicable. No clinical patient data was submitted for performance testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable. No clinical performance testing with expert-established ground truth was submitted.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method for the test set

    • Adjudication Method: Not applicable. No clinical performance testing requiring adjudication was submitted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. The device "Neuronaute" is an EEG data acquisition, display, storage, and transmission system. It is explicitly stated that "Neuronaute and its components do not provide any diagnostics conclusions or automated alerts of an adverse clinical event about a patient's clinical condition." Therefore, it does not include AI for diagnostic assistance, and an MRMC study comparing human readers with and without AI assistance is not relevant to this device's current claims.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: No. As noted above, the device does not provide diagnostic conclusions or automated alerts, implying no standalone diagnostic algorithm. Its primary function is to collect and transmit EEG signals for review by a physician.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: Not applicable for diagnostic performance. For the technical performance aspects (e.g., signal quality, electrical safety), the "ground truth" would be the specifications and requirements of the referenced industry standards (e.g., IEC 60601-2-26) or "EEG gold standards" referenced in the text.

    8. The sample size for the training set

    • Sample Size (Training Set): Not applicable. As the device does not provide diagnostic conclusions or automated alerts, it does not appear to involve a machine learning model that would require a patient-based training set for diagnostic classification.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable. See explanation for #8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173248

    Validate with FDA (Live)

    Device Name
    CARDIOSKIN
    Manufacturer
    Date Cleared
    2018-08-11

    (305 days)

    Product Code
    Regulation Number
    870.2920
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cardioskin is intended to aid the diagnostic evaluation for cardiac disorders by recording patient's ECG signals allowing to condition and monitor these signals. ECG data is stored, transferred and displayed for a review by a physician. Cardioskin is designed to be used by a patient to transmit multiple-lead ECG signal to enable review at a physician's office, hospital or other remote locations under medical supervision. Cardioskin target population is adults.

    Device Description

    Cardioskin is a wireless and wearable medical device.

    It enables acquisition, recording, storage, transmission and display of 3-12 leads electrocardiogram (ECG) to analyze potential cardiac pathological abnormalities from adult patients in healthcare facility, at home or clinical research environment. It is intended to be used by trained patients and healthcare professionals only.

    Cardioskin is composed of a garment, Cardioskin Textile, which is a T-Shirt where textilebased electrodes made of silver yarns are knitted and only functions on Cardioskin Battery and in association with a Cardioskin Recorder.

    Cardioskin functioning is coupled with a smartphone/tablet compatible with an iOS or Android system, and a mobile application, the Cardioskin App.

    A web-based information system, the Cardioskin Cloud, receives the ECG signals from the Cardioskin Recorder through a paired Wi-Fi connection for its long-term storage and display the signals in the web interface. The medical expert specialist, who prescribes the use of this device, should monitor its use and confirm the proper functioning of signal recording thanks to Cardioskin Cloud.

    AI/ML Overview

    The provided text describes the Cardioskin device, its indications for use, and a comparison to predicate devices, but it primarily focuses on regulatory approval (510(k) summary) and does not contain detailed acceptance criteria or a specific study demonstrating direct performance against those criteria.

    Instead, the document mentions that "Support for the substantial equivalence of Cardioskin was provided as a result of risk management and testing which included electrical and biological safety, performance and software tests." It also states, "Testing has provided reasonable assurance of safety and effectiveness for the intended use and supports the substantial equivalence demonstration." However, it does not elaborate on what these performance tests entailed, what the acceptance criteria were, or the specific results of those tests.

    Therefore, many of the requested items cannot be definitively answered from the provided text.

    Here is an attempt to answer the questions based only on the information available:

    1. A table of acceptance criteria and the reported device performance

      • Acceptance Criteria: Not explicitly stated in the document. The document refers to "requirements for safety, function and intended use" and compliance with specific IEC and ISO standards (IEC 60601-1, IEC 60601-1-2, ISO 14971, IEC 62304, IEC 60601-2-47). These standards imply certain performance criteria, but the specific thresholds or metrics for "acceptance" of the Cardioskin's ECG performance (e.g., accuracy of heart rate detection, signal-to-noise ratio) are not detailed.
      • Reported Device Performance: The document only makes a general statement: "The results of the verification and validation activities demonstrate that the software meets the requirements for safety, function and intended use." No specific performance metrics or quantitative results for ECG signal acquisition and analysis are provided.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • The document refers to "non-clinical and limited clinical testing" but does not specify the sample size for any test set or the data provenance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not specified.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not specified.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not specified, and it's unlikely given the nature of an ECG recording device where the primary output is the raw data for physician review, rather than an AI interpretation tool being evaluated for reader improvement.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • The device "records patient's ECG signals allowing to condition and monitor these signals. ECG data is stored, transferred and displayed for a review by a physician." This indicates a human-in-the-loop system, as the physician performs the review. There is no mention of a standalone algorithm for diagnosis.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not specified.
    8. The sample size for the training set

      • Not specified. This device is primarily for data acquisition and display, and while it has a "Cardioskin Cloud" that stores and displays signals, the document doesn't indicate extensive AI model training in the diagnostic sense that would require a large training set for a "ground truth" establishment.
    9. How the ground truth for the training set was established

      • Not specified.

    Summary of what can be inferred or directly stated from the text regarding performance and testing:

    • Type of tests conducted: Electrical and biological safety, performance (general statement), software verification and validation, electromagnetic compatibility, and compliance with specific medical device standards (IEC 60601-1, IEC 60601-1-2, ISO 14971, IEC 62304, IEC 60601-2-47).
    • Biocompatibility tests: Irritation and Sensitization, based on ISO 10993 for skin contact limited to 12 hours.
    • Software tests: Verification and validation according to IEC 62304:2006 and FDA Guidance "General Principles for Software Validation."
    • Outcome of tests: The tests provide "reasonable assurance of safety and effectiveness for the intended use" and demonstrate that "the software meets the requirements for safety, function and intended use." Compliance with recognized standards was demonstrated.

    The document primarily serves as a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices and meeting general regulatory safety and effectiveness requirements, rather than providing detailed performance study results against specific, quantified acceptance criteria for diagnostic performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170138

    Validate with FDA (Live)

    Device Name
    NEURONAUTE
    Manufacturer
    Date Cleared
    2017-12-26

    (343 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neuronaute is a medical, wireless and mobile equipment, which allows acquisition, record, storage, transmission and displaying of electroencephalogram (EEG) from adult patients.

    It can be used with patients in health care facility (data acquisition and reporting) or clinical research environment.

    The Neuronaute requires operation by a trained healthcare professional.

    The Neuronaute only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.

    Device Description

    The Neuronaute is a wearable and wireless equipment, which allows acquisition, recording, storage, transmission and displaying of electroencephalogram (EEG).

    The Neuronaute is composed of a textile cap where twenty-one titanium dry electrodes are attached. An electronic box is attached to the cap (EEG recording) associated to batteries. The system operates on battery power only.

    The system is connected to a device (i.e. smart phone or a digital tablet) loaded with a separate stand-alone software program ("app"). The app works with a cloud backend service, which allows the user management and authentication, the data storage for application recordings and access to this data via a web-based interface, the devices management and authentication, the recording scheduling, the download of the latest firmware version.

    The acquired signals are saved in a universal data format (European Data Format, EDF) that is intended to be visualized by a Physician through the web-based interface.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting study for the Neuronaute device, based on the provided FDA 510(k) summary:

    The Neuronaute device is an electroencephalograph (EEG) system intended for acquiring, recording, storing, transmitting, and displaying EEG from adult patients. The acceptance criteria and supporting studies focus on demonstrating that the Neuronaute performs similarly to or better than predicate devices, and meets relevant electrical, software, and biocompatibility standards.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list "acceptance criteria" with numerical targets and then "reported device performance" in a separate section. Instead, the "Performance Data" and "Technological Characteristics" sections describe testing conducted and the device's measured specifications, often compared directly to predicate devices to demonstrate substantial equivalence. I will synthesize the relevant information into a table, focusing on direct and inferable criteria for signal quality and electrical characteristics.

    Acceptance Criteria (Inferred from Predicate Equivalence & Standards)Neuronaute Reported Performance
    Electrical Performance:
    Sampling rate500 s/s (vs. predicate 256 s/s)
    Dynamic range+/- 600mV (vs. predicate +/- 1000 µV - note: units differ, but generally implies a range adequate for EEG)
    Noise< 6 µVpp (vs. predicate 3.7 µV typical - comparable or slightly higher, within acceptable range for EEG)
    CMRR (Common Mode Rejection Ratio)115dB (vs. predicate 110dB - better)
    Input Impedance1 Gohm (vs. predicate 100 Gohm - lower, but still high enough for good signal acquisition)
    Hardware filtering125Hz lowpass (vs. predicate 100 Hz Low Pass)
    BiocompatibilityCompliance with ISO 10993: Cytotoxicity, Irritation, Sensitization testing shown to be substantially equivalent to predicate. Materials (Ti6A14V ELI & OEKO-TEX® yarns) chosen for known biocompatibility.
    Electromagnetic compatibility (EMC) & Electrical safetyCompliance with IEC 60601-2-26:2002 and FDA Guidance demonstrated, substantially equivalent to primary predicate.
    Software ValidationVerification and validation according to IEC 62304:2006 and FDA Guidance "General Principles for Software Validation". Results demonstrated meeting requirements for safety, function, and intended use.
    Signal Quality:
    Acceptability of signal quality against gold standardTesting performed shows similar electrical and signal performance as the Rythmlink Disc Electrodes (K061148, considered gold standard).
    Alpha Rhythm presencePhysiological tests showed normal presence of alpha modulation in eyes opening/closing task.
    Impedance levelsMost impedances were below 1 MOhm.
    Data lossChirp and Ramp tests presented no data lost with the current version of software.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Physiological Tests (Impedance, Alpha Rhythm):
      • Impedance tests: 24 subjects.
      • Alpha Rhythm tests: 26 healthy adult subjects.
    • EEG signal testing for comparison with Rythmlink Disc Electrodes: The document doesn't explicitly state the number of subjects for this comparison, but it implies a test was conducted.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the mention of "healthy adult subjects," it implies a prospective clinical or research study was conducted internally or with a clinical partner.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    The document does not describe the use of experts to establish ground truth for the physiological or comparative EEG signal quality tests. The "gold standard" comparison for EEG signal quality refers to a predicate device (Rythmlink Disc Electrodes), not expert interpretation of data. The physiological tests (Alpha Rhythm, Impedance) likely relied on objective measurements and established physiological markers rather than expert consensus for ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable, as no expert-based ground truth establishment or adjudication process is described for the tests mentioned. The assessment of signal quality and electrical performance appears to be based on direct measurement and comparison to predefined technical specifications and predicate device performance.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The Neuronaute is described as a device that only acquires and displays physiological signals, with "no claims being made for analysis of the acquired signals with respect to the accuracy, precision and reliability." This means the device does not provide AI-assisted interpretations to be compared against human readers. Its function is data acquisition and display.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    A standalone algorithm performance study was not done, as the Neuronaute explicitly states it makes "no claims ... for analysis of the acquired signals." The "software" (firmware, mobile app, cloud) focuses on verification and validation of its functionality in acquiring, recording, storing, transmitting, and displaying data reliably and safely.

    7. Type of Ground Truth Used:

    • For EEG signal quality comparison: The "gold standard" was another cleared EEG electrode, Rythmlink Disc Electrodes (K061148). This serves as a technical ground truth for comparison of electrical and signal performance.
    • For physiological tests (Alpha Rhythm, Impedance): The ground truth is based on established physiological phenomena (e.g., normal presence of alpha modulation, expected impedance levels) and objective measurements.

    8. Sample Size for the Training Set:

    The document does not mention any "training set" as it is not an AI/ML device that performs analysis or interpretation. The software validation is for the functionality of data handling and display, not for learning from data.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1