510(k) Data Aggregation

    K Number
    K180805

    Validate with FDA (Live)

    Device Name
    Pure EP
    Date Cleared
    2018-08-08

    (133 days)

    Product Code
    Regulation Number
    870.1425
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PURE EP™ is a computerized system intended for acquiring, digitizing, filtering, measuring and calculating, displaying, recording and storing of electrocardiographic and intracardiac signals for patients undergoing electrophysiology (EP) procedures in an EP laboratory under the supervision of licensed healthcare practitioners who are responsible for interpreting the data.

    Device Description

    PURE EP™ is a system consists of two main parts: the amplifier, and the personal computer (PC), which are connected over a high-speed fiber optic cable. The patient is connected to the amplifier via a set of supplied ECG cables and commercially available intracardiac (IC) catheters. The amplifier includes hardware and embedded software necessary for acquiring ECG and IC signals from patients. The device is not intended for active patient monitoring. It consists of one ECG and seven IC modules designed to acquire 12-lead ECG and 56 IC signals and send the digitized data to the PC. The PC has preinstalled PURE-EP software and is connected to display monitors. The pre-installed software provides visualization of received data from the amplifier on real-time and review screens. It also includes a signal processing module that provides various filter options for real-time and review screens.

    AI/ML Overview

    The provided text describes the PURE EP™ system, a computerized system for acquiring, digitizing, filtering, measuring, calculating, displaying, recording, and storing electrocardiographic and intracardiac signals during electrophysiology (EP) procedures. The document does not contain specific acceptance criteria for device performance, nor does it detail a clinical study proving the device meets such criteria in terms of diagnostic accuracy or equivalent clinical outcomes.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (GE Healthcare's CardioLab) through engineering and safety testing, rather than a statistical study against a specific performance benchmark.

    Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is absent:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state specific acceptance criteria in terms of diagnostic performance metrics (e.g., sensitivity, specificity, accuracy). The "Performance Testing" section lists compliance with a variety of industry standards for software, electrical, and mechanical testing, which implies that the device achieved performance within the specified limits of those standards. However, these are general safety and engineering standards, not specific performance criteria for signal interpretation accuracy.

    Performance AspectAcceptance Criteria (from text)Reported Device Performance (from text)
    Functional/Performance RequirementsDevice design meets its functional and performance requirements as verified through software, electrical, and mechanical testing in accordance with industry standards.Representative samples underwent software, electrical, and mechanical testing in accordance with listed industry standards (e.g., ANSI/AAMI/IEC 62366-1, AAMI/ANSI/ISO 14971, IEC 60601-1-6, AAMI/ANSI HE75, IEC 62304, AAMI/ANSI ES60601-1, AAMI/ANSI IEC 60601-1-2). User validation was also performed.
    Electrical SafetyCompliance with AAMI/ANSI ES60601-1: 2012. Patient Source: <10 uA; Patient Sink: <10 uA; Patient Sink (single fault): <50 uA; Chassis Leakage: <100 uA.The device meets these electrical safety standards (listed with values identical to the predicate).
    Electromagnetic Compatibility (EMC)Compliance with AAMI/ANSI IEC 60601-1-2:2014.Device underwent testing to ensure electromagnetic compatibility.
    UsabilityCompliance with ANSI/AAMI/IEC 62366-1:2015 and IEC 60601-1-6: 2013 and AAMI/ANSI HE75: 2013.Device underwent usability engineering and human factors testing.
    Software Life CycleCompliance with IEC 62304: 2015.Software development followed life cycle processes specified by the standard.
    Risk ManagementCompliance with AAMI/ANSI/ISO 14971:2007/(R)2010.Risk management principles were applied during development.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states: "Similar to the predicate, the subject of this premarket submission, PURE EPIM, did not require clinical studies to support substantial equivalence." This indicates no clinical test set was used for a study proving device performance on patient data against specific performance criteria. The testing was focused on engineering, safety, and functional parameters.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Since no clinical studies were performed to establish performance on patient data, no experts were used to establish ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    As no clinical test set was used, there was no adjudication method applied.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, as the device's function is for acquiring, digitizing, filtering, measuring, calculating, displaying, recording, and storing signals, not for providing interpretative AI assistance to human readers. It's a data acquisition and processing system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is fundamentally a data acquisition and display system. It's not an "algorithm" in the sense of providing automated diagnostic insights independently. Its function is to present data to healthcare practitioners for their interpretation. Therefore, a "standalone algorithm only" performance study in a diagnostic context is not applicable to this device as described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the engineering and safety performance tests, the "ground truth" would be the expected output or behavior as defined by the engineering specifications and industry standards. For example, signal amplification accuracy would be judged against a known input signal and the device's specified gain. However, for a clinical diagnostic "ground truth," none was used as no clinical-diagnostic performance study was conducted.

    8. The sample size for the training set

    Since no clinical studies for diagnostic performance were conducted, and the device functions as a data acquisition and processing system rather than an AI-driven diagnostic algorithm, there is no concept of a "training set" in the context of diagnostic accuracy. Training here would refer to the development and internal testing of the system by the manufacturer to ensure it meets its engineering specifications.

    9. How the ground truth for the training set was established

    As there was no clinical "training set" for diagnostic performance, this question is not applicable. The "ground truth" during the device's development (analogous to a training phase) would have been established through design specifications, simulated signals, and controlled laboratory testing to ensure the electrical and software components performed as designed.