Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K221036

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-01

    (85 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide Warming is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    Astroglide Warming is non-sterile, clear and water based personal lubricant. This product is not a spermicide or contraceptive. It is compatible with natural rubber latex and polyisoprene condoms only. It is not compatible with polyurethane condoms. The device is composed of water, glycerin, propylene glycol, polyquaternium-7 and citric acid.

    AI/ML Overview

    The provided document describes the K221036 submission for Astroglide Warming, a personal lubricant. It does not describe a study involving an AI/ML device and human readers. Therefore, I cannot fulfill all aspects of your request as it pertains to AI/ML device performance, ground truth, expert opinions, or MRMC studies.

    However, I can extract the acceptance criteria and the summary of non-clinical performance testing for the Astroglide Warming device itself.

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Test Method)Reported Device Performance
    Absence of particulate matterNo particlesMet specification (implied by "met the specifications at all time points")
    ColorColorlessMet specification (implied by "met the specifications at all time points")
    ClarityClearMet specification (implied by "met the specifications at all time points")
    OdorOdorlessMet specification (implied by "met the specifications at all time points")
    Viscosity175 - 330 cPMet specification (implied by "met the specifications at all time points")
    Osmolality2600 – 2800 mOsm/kg, dilution factor of 5Met specification (implied by "met the specifications at all time points")
    pH3.5-5.5Met specification (implied by "met the specifications at all time points")
    Total yeast/mold count (TYMC)<10 cfu/mL (USP <61>)Met specification (implied by "met the specifications at all time points")
    Total aerobic microbial count (TAMC)<100 cfu/mL (USP <61>)Met specification (implied by "met the specifications at all time points")
    Presence of pathogenic organisms (Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicans)Absent (USP<62>)Met specification (implied by "met the specifications at all time points")
    Antimicrobial effectivenessMeets USP<51> acceptance criteria for Category 2 products.Met specification (implied by "met the specifications at all time points")
    Biocompatibility: CytotoxicityNon-cytotoxic (per ISO 10993-5:2009/(R)2014)Non-cytotoxic
    Biocompatibility: Sensitization and IrritationNon-sensitizing, non-irritating (per ASTM D6355)Non-sensitizing, non-irritating
    Biocompatibility: Systemic ToxicityNot systemically toxic (per ISO 10993-11:2017)Not systemically toxic
    Shelf Life (8.5 months)Met all specifications in Table 1Met the specifications at all time points during accelerated aging study.
    Condom Compatibility: Natural Rubber Latex CondomsCompatible (per ASTM D7661-18)Compatible with natural rubber latex condoms.
    Condom Compatibility: Polyisoprene CondomsCompatible (per ASTM D7661-18)Compatible with polyisoprene condoms.
    Condom Compatibility: Polyurethane CondomsNot compatible (per ASTM D7661-18)Not compatible with polyurethane condoms.

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical performance testing for a personal lubricant, not an AI/ML device. Therefore, the concept of "test set" in the context of AI/ML is not applicable here. The testing involved samples of the product itself.

    • Biocompatibility testing: Not specified number of samples, but refers to ISO and ASTM standards.
    • Shelf Life Accelerated Aging Study: Not specified number of samples. The study duration was 3 months at 40°C.
    • Condom Compatibility: Not specified number of samples, but refers to ASTM D7661-18.
    • Data provenance: Not explicitly stated, but assumed to be from internal lab testing conducted by or for BioFilm, Inc. It is retrospective in the sense that the tests were completed before submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This is a physical product, not an AI/ML device requiring expert ground truth for image or data interpretation.

    4. Adjudication method for the test set:

    Not applicable. This is a physical product, not an AI/ML device requiring adjudication for ambiguous cases.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a physical product, not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical product, not an AI/ML algorithm. The "standalone" performance here refers to the device's inherent physical and chemical properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this product is based on established industry standards and validated test methods as defined by organizations like ISO, ASTM, and USP. For example:

    • Physical and Chemical Properties: The specifications (e.g., viscosity, pH, osmolality, color, odor, clarity, absence of particulate matter) are based on direct measurements against defined ranges.
    • Microbial Limits: Determined by specific USP <61> and <62> methods.
    • Antimicrobial Effectiveness: Determined by USP <51> methods.
    • Biocompatibility: Determined by ISO 10993-5, ISO 10993-11, and ASTM D6355 standards.
    • Condom Compatibility: Determined by ASTM D7661-18.

    8. The sample size for the training set:

    Not applicable. This is a physical product, not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220355

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-26

    (107 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide Ultra Gentle Gel is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    Astroglide Ultra Gentle Gel is a non-sterile, clear, odorless, and water based personal lubricant. This product is not a spermicide or contraceptive. It is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. The device is composed of water, propylene glycol, hydroxyethylcellulose, polyquaternium 7, potassium sorbate, sodium benzoate and citric acid. The product primary packaging is a white LDPE 3-ounce tube with a screw on cap. The tube and cap constitute the device final packaging.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the supporting study, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Specification)Reported Device Performance (Met/Not Met/Not Applicable)
    ColorClear to goldenMet (Implied, as testing "met specifications")
    ClarityClearMet (Implied, as testing "met specifications")
    OdorOdorlessMet (Implied, as testing "met specifications")
    Absence of particulate matterNo particlesMet (Implied, as testing "met specifications")
    Viscosity10,000-25,000 cPMet (Implied, as testing "met specifications")
    pH3.5-5.5Met (Implied, as testing "met specifications")
    Osmolality200-300 mOsm/kg, 1:5 dilution factorMet (Implied, as testing "met specifications")
    Total yeast/mold count (TYMC)<10 cfu/mL (USP <61>)Met (Implied, as testing "met specifications")
    Total aerobic microbial count (TAMC)<100 cfu/mL (USP <61>)Met (Implied, as testing "met specifications")
    Presence of Pathogenic Organisms (Staphylococcus aureus, Pseudomonas aeruginosa, Candida albicans)Absent (USP <62>)Met (Implied, as testing "met specifications")
    Antimicrobial effectivenessMeets USP <51> acceptance criteria for Category 2 productsMet (Implied, as testing "met specifications")
    CytotoxicityNon-cytotoxic (per ISO 10993-5:2009/(R)2014)Met
    Sensitization/IrritationNon-sensitizing, Non-irritating (per ASTM D6355)Met
    Systemic ToxicityNot systemically toxic (per ISO 10993-11:2017)Met
    Shelf Life (24 months)Maintain all specifications listed in Table 1Met (Confirmed by accelerated aging study)
    Compatibility with Natural Rubber Latex CondomsCompatible (per ASTM D7661-18)Compatible
    Compatibility with Polyisoprene CondomsCompatible (per ASTM D7661-18)Compatible
    Compatibility with Polyurethane CondomsNot Compatible (per ASTM D7661-18)Not Compatible (This is a reported finding, not a failure of a positive criterion)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test in the "Summary of Non-Clinical Performance Testing" section. For the Shelf Life study, it mentions an "accelerated aging study, conducted for 8 months at 40°C," but doesn't detail the number of samples or batches tested. The provenance of the data is internal to BioFilm, Inc., as these are non-clinical performance tests conducted for device clearance. There is no information regarding country of origin or whether it's retrospective or prospective beyond the description of the tests themselves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of device and study. The testing described (biocompatibility, shelf life, condom compatibility) are laboratory-based, objective performance tests conducted against established international or national standards (e.g., ISO, ASTM, USP). Ground truth is established by the test method's specified criteria, not by expert consensus in the way a clinical diagnostic device would require.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the tests are objective laboratory performance tests, not subjective interpretations requiring adjudication among experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI-driven diagnostic or screening devices involving human interpretation, not for a personal lubricant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a personal lubricant, not an AI algorithm. The performance tests are directly on the physical product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the acceptance criteria for most of the parameters (Color, Clarity, Odor, Absence of particulate matter, Viscosity, pH, Osmolality, Microbial Limits, Antimicrobial effectiveness) is based on pre-defined quantifiable specifications and standard laboratory test methods (USP <61>, USP <62>, USP <51>).

    For Biocompatibility, the ground truth is established by the pass/fail criteria defined in international standards like ISO 10993-1, ISO 10993-5, ISO 10993-11, and ASTM D6355.

    For Condom Compatibility, the ground truth is established by the pass/fail criteria defined in ASTM D7661-18.

    For Shelf Life, the ground truth is that the device must continue to meet all the physical, chemical, and microbiological specifications after accelerated aging, following ASTM F1980-16.

    8. The sample size for the training set

    This is not applicable as the device is not an AI/ML algorithm. There is no "training set" for a physical personal lubricant.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220653

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-06

    (60 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide® Glycerin & Paraben Free is a personal lubricant for penile, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    Astroglide® Glycerin & Paraben Free is a non-sterile, water-based liquid personal lubricant for penile, anal, and/or vaginal application. This product is not a spermicide or contraceptive. It is compatible with natural rubber latex and polyisoprene condoms. It is not compatible with polyurethane condoms. The subject device is sold as an overthe-counter (OTC) product in a 2.5 oz. clear PETE bottle with a screw on polypropylene flip-top cap. The bottle is labeled front and back and then packaged in a cardboard carton which constitutes the final packaging. This device is composed of water, butylene glycol, propylene glycol, xylitol, methyl gluceth-20, polyquarternium-7, hydroxyethyl cellulose, potassium sorbate, sodium benzoate, citric acid.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a personal lubricant called Astroglide® Glycerin & Paraben Free. It primarily focuses on demonstrating substantial equivalence to a previously marketed predicate device, rather than providing details on a study of a device's performance against specific clinical acceptance criteria for diagnostic or AI-driven devices.

    Therefore, much of the requested information (sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone AI performance, training set details) is not applicable to this type of regulatory submission for a personal lubricant.

    However, I can extract information related to the acceptance criteria for the product's physical and chemical properties, as well as its biocompatibility and condom compatibility.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Test Method)Reported Device Performance
    ColorClear to slight goldenNot explicitly stated as "reported performance," but the device is described as having this specification.
    ClarityClearNot explicitly stated as "reported performance," but the device is described as having this specification.
    OdorSlight sweet odorNot explicitly stated as "reported performance," but the device is described as having this specification.
    Absence of particulate matterAbsentNot explicitly stated as "reported performance," but the device is described as having this specification.
    Viscosity1000-2800 cPNot explicitly stated as "reported performance," but the device is described as having this specification.
    pH3.5-5.5Not explicitly stated as "reported performance," but the device is described as having this specification.
    Osmolality230-300 mOsm/kg, 1:5 dilution factorNot explicitly stated as "reported performance," but the device is described as having this specification.
    Total yeast/mold count (TYMC)<10 cfu/mL (USP <61>)Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Total aerobic microbial count (TAMC)<100 cfu/mL (USP <61>)Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Presence of Pathogenic OrganismsAbsent (USP <62>) (Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicans)Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Antimicrobial effectivenessMeets USP <51> acceptance criteria for Category 2 products.Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Biocompatibility (Cytotoxicity)Non-cytotoxic (per ISO 10993-5:2009/(R)2014)Demonstrated non-cytotoxic
    Biocompatibility (Sensitization)Non-sensitizing (per Human Repeat Insult Patch Testing (HRIPT) per ASTM D6355)Demonstrated non-sensitizing
    Biocompatibility (Irritation)Non-irritating (per Human Repeat Insult Patch Testing (HRIPT) per ASTM D6355)Demonstrated non-irritating
    Biocompatibility (Systemic Toxicity)Not systemically toxic (per ISO 10993-11:2017)Demonstrated not systemically toxic
    Shelf Life (Device Specifications)Met specifications in Table 1 across shelf-lifeMet specifications at all time points over 22 months
    Condom Compatibility (Natural rubber latex)Compatible (per ASTM D7661-18)Compatible
    Condom Compatibility (Polyisoprene)Compatible (per ASTM D7661-18)Compatible
    Condom Compatibility (Polyurethane)Not Compatible (per ASTM D7661-18)Not Compatible

    2. Sample size used for the test set and the data provenance

    • For physical/chemical specifications: The document refers to standard test methods (e.g., USP <61>, USP <62>) but does not specify the sample size for these tests. The data provenance is implied to be from BioFilm, Inc.'s internal testing, likely from the U.S.
    • For biocompatibility: No specific sample sizes for human subjects or animal models are provided for the HRIPT or systemic toxicity tests. For cytotoxicity, it's generally in vitro testing. Data provenance is implied to be from BioFilm, Inc.'s testing.
    • For condom compatibility: No specific sample size (i.e., number of condoms tested) is explicitly mentioned, but the testing was conducted according to ASTM D7661-18. Data provenance is implied to be from BioFilm, Inc.'s testing.
    • For shelf life: The study was an accelerated aging study conducted for 8 months at 40°C, per ASTM F1980-16. No sample sizes (number of devices) are given.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not a study requiring expert clinical review or ground truth establishment for diagnostic accuracy. The ground truth for the physical/chemical properties, biocompatibility, and condom compatibility are defined by the standardized test methods themselves.

    4. Adjudication method for the test set

    Not applicable. There is no subjective assessment requiring adjudication in this type of testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a personal lubricant device, not an AI-driven diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a personal lubricant device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the various acceptance criteria were established by:

    • Standardized laboratory methods/specifications: For color, clarity, odor, particulate matter, viscosity, pH, osmolality, microbial counts (TYMC, TAMC, pathogens), and antimicrobial effectiveness, the ground truth is defined by specified ranges, limits, or absence/presence criteria according to recognized pharmacopeial (USP) or internal standards.
    • International Standards (ISO) and ASTM Standards:
      • Biocompatibility: ISO 10993-5 (cytotoxicity), ASTM D6355 (sensitization/irritation), ISO 10993-11 (systemic toxicity).
      • Shelf Life: ASTM F1980-16 (accelerated aging for medical devices).
      • Condom Compatibility: ASTM D7661-18.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-driven device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213671

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-03-03

    (101 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide® Sensual Strawberry Personal Lubricant is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex condoms. This product is not compatible with polyisoprene or polyurethane condoms.

    Device Description

    Astroglide® Sensual Strawberry Personal Lubricant is a non-sterile, clear, strawberry-scented and flavored, water-based personal lubricant intended for penile, anal, or vaginal application. This product is not a spermicide or contraceptive. Astroglide® Sensual Strawberry Personal Lubricant is compatible with natural rubber latex condoms. It is not compatible with polyisoprene and polyurethane condoms. This product's primary packaging is a PETE clear bottle with a screw on polypropylene flip-top cap. The bottle is then packaged in a cardboard carton which constitutes the final packaging.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from BioFilm, Inc. to the FDA for their product, Astroglide® Sensual Strawberry Personal Lubricant. It aims to demonstrate substantial equivalence to a legally marketed predicate device (Glycerin & Paraben Free Astroglide, K072647).

    Therefore, this document does not describe:

    • Acceptance criteria and device performance for an AI/ML medical device.
    • A study proving the performance of an AI/ML medical device.
    • Data provenance, sample sizes for test/training sets, expert qualifications, or adjudication methods related to AI/ML model development or validation.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies or standalone AI performance.
    • Details about ground truth establishment for AI/ML models.

    Instead, the document focuses on:

    • Device Description: A water-based personal lubricant, its physical characteristics, and specifications.
    • Predicate Device Comparison: Highlighting similarities (indications for use, over-the-counter status, base type, pH) and differences (ingredients, osmolality, viscosity, packaging, shelf-life). The document asserts that these differences do not raise new questions of safety and effectiveness.
    • Non-Clinical Performance Testing: This is the equivalent of "proof" in this context, demonstrating the device's safety and effectiveness through:
      • Biocompatibility Testing: Cytotoxicity, sensitization, irritation, and acute systemic toxicity.
      • Condom Compatibility Testing: Specifically with natural rubber latex condoms using ASTM D7661-10.
      • Shelf-Life Testing: Using accelerated aging (ASTM F1980-16) to determine an 8.5-month shelf-life.

    To answer your specific questions based on the provided document, which is NOT about an AI/ML device:

    1. A table of acceptance criteria and the reported device performance:

      Physical Specification TestsAcceptance Criteria / Ranges/SpecificationsReported Device Performance (from text, implicitly met for approval)
      Particulate matterNo particlesMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study.")
      ColorClear to slight goldenMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study.")
      ClarityClearMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study.")
      OdorStrawberry scentMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study.")
      pH (per USP <791>)3.5-5.5Met (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study." Also stated in comparison table as "Same" to predicate.)
      Viscosity (per USP <912>)200 - 450 cpsMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study." The value 200-450 cps is the reported device performance for this specific product, which is different from the predicate, but deemed acceptable by the FDA.)
      Osmolality (per USP <785>)200 - 450 mOsm/kg, dilution factor of 5Met (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study." The value 200-450 mOsm/kg is the reported device performance for this specific product, which is different from the predicate, but deemed acceptable by the FDA.)
      Antimicrobial effectiveness (per USP <51>)Category 2: bacteria show ≥2.0 log reduction at 14 days, no increase from 14-day count at 28-day count; yeast/molds show no increase from initial calculated count at 14 and 28 days.Met (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study." Also stated in comparison table as "Same" to predicate.)
      Total yeast/mold count (per USP <61> & USP <62>)< 10 cfu/mLMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study." Also stated in comparison table as "Same" to predicate.)
      Total aerobic microbial count (per USP <61> & USP <62>)< 100 cfu/mLMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study." Also stated in comparison table as "Same" to predicate.)
      Absence of pathogenic organisms (Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicans - per USP <61> & USP <62>)AbsentMet (implied by FDA clearance and statement: "All specifications for the subject lubricant... were met throughout the shelf-life study." Also stated in comparison table as "Same" to predicate.)
      BiocompatibilityNon-cytotoxic, non-sensitizing, non-irritating, not systemically toxicDemonstrated to be non-cytotoxic, non-sensitizing, non-irritating, and not systemically toxic.
      Condom Compatibility (Natural Rubber Latex) (per ASTM D7661-10)CompatibleResults show that Astroglide® Sensual Strawberry Personal Lubricant is compatible with natural rubber latex condoms.
      Shelf-Life (per ASTM F1980-16)All specifications from Section IV met throughout the study.8.5 months shelf-life achieved; all specifications in Section IV were met throughout the shelf-life study.
    2. Sample size used for the test set and the data provenance: This document describes non-clinical performance testing for a personal lubricant, not an AI/ML device using a "test set" in the machine learning sense. The tests are laboratory-based.

      • Sample Size: Not explicitly stated for each test (e.g., number of lubricant samples tested for pH, viscosity, microbial counts). For biocompatibility, it's not specified how many samples/subjects were used for the in-vitro and in-vivo tests. For condom compatibility, ASTM D7661-10 would specify the number of condoms and lubricant samples, but this detail is not provided in the summary. For shelf-life, a sufficient number of product units would have been put on accelerated aging.
      • Data Provenance: Not applicable in the geopolitical sense. The data comes from the results of laboratory testing conducted for the purpose of this 510(k) submission. These are not "retrospective or prospective" patient data sets.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a lubricant is established by laboratory measurements (e.g., pH meter for pH, viscometer for viscosity, microbial culture for counts, established protocols for biocompatibility and condom compatibility). There are no human "experts" establishing a "ground truth" for classification or detection tasks as would be the case for an AI/ML device.

    4. Adjudication method: Not applicable. This is not a human-reader study where adjudication is needed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is not about an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This document is not about an AI/ML device.

    7. The type of ground truth used:

      • Physical Specifications: Instrumental measurements (pH meters, viscometers, osmolality readers) and laboratory assays (microbial cultures, particulate inspection).
      • Biocompatibility: In-vitro (cytotoxicity) and in-vivo (sensitization, irritation, acute systemic toxicity) testing according to ISO standards.
      • Condom Compatibility: Standardized mechanical and material property testing (ASTM D7661-10).
      • Shelf-Life: Stability testing over time under accelerated conditions according to ASTM F1980-16, measuring the physical specifications listed in the table.
    8. The sample size for the training set: Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210242

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-28

    (89 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide® X Silicone Personal Lubricant is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex condoms. This product is not compatible with polyisoprene and polyurethane condoms.

    Device Description

    Astroglide® X Silicone Personal Lubricant is non-sterile, clear, odorless, and siliconebased lubricant. This product is not a spermicide or contraceptive. Astroglide® X Silicone Personal Lubricant is compatible with natural rubber latex condoms. Astroglide® X Silicone Personal Lubricant is not compatible with polyurethane, or polyisoprene condoms. This product's primary packaging is a PETE clear bottle with a screw on polypropylene flip-top cap. The bottle is packaged in a cardboard carton which constitutes the final packaging.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study for the Astroglide® X Silicone Personal Lubricant.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Specification)Reported Device Performance
    AppearanceClearClear
    OdorCharacteristicOdorless
    Viscosity225 to 350 centipoise225 to 350 centipoise
    Total yeast/mold count (USP <61>)<10 cfu/mL<10 cfu/mL
    Total aerobic microbial count (USP <61>)<100 cfu/mL<100 cfu/mL
    Absence of Pathogenic Organisms (Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicans) (USP <62>)AbsentAbsent
    Water activity (USP<1112>)< 0.3 Aw< 0.3 Aw
    Condom CompatibilityCompatible with natural rubber latex condomsCompatible with natural rubber latex condoms only
    Shelf-Life3 years3 years
    Biocompatibility (Cytotoxicity)Non-cytotoxicNon-cytotoxic
    Biocompatibility (Sensitization)Non-sensitizingNon-sensitizing
    Biocompatibility (Irritation)Non-irritatingNon-irritating
    Biocompatibility (Systemic Toxicity)Not systemically toxicNot systemically toxic

    Note: The "Reported Device Performance" column reflects that the device met the stated specifications/criteria, as confirmed by the "Summary of Non-Clinical Performance Testing" section.

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" in the context of a clinical trial or a specific number of samples for each non-clinical test. The testing described is non-clinical performance testing.

    • Sample Size: Not explicitly stated for each test. The document refers to "the subject lubricant" being tested.
    • Data Provenance: Not explicitly stated. The tests are non-clinical (laboratory-based) and conducted according to international standards (ISO, ASTM, USP). There's no mention of country of origin for data or if it's retrospective/prospective in a clinical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable. The evaluations are based on internationally recognized laboratory test methods (ISO, ASTM, USP), not expert consensus on individual cases or ground truth derived from clinical experts.

    4. Adjudication method for the test set

    This section is not applicable. The evaluations are based on objective laboratory test results against defined specifications, not on human interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. This document pertains to the regulatory submission of a personal lubricant, not an AI-powered diagnostic device, so an MRMC comparative effectiveness study is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable. This document pertains to the regulatory submission of a personal lubricant, not an AI-powered algorithm.

    7. The type of ground truth used

    The "ground truth" for the device's performance is established by objective measurements against predetermined specifications and international standards. This includes:

    • Physical property measurements (viscosity, appearance, odor, water activity).
    • Microbiological testing (total yeast/mold count, aerobic microbial count, absence of specific pathogens).
    • Biocompatibility testing results (cytotoxicity, irritation, sensitization, systemic toxicity) as per ISO 10993.
    • Condom compatibility testing results as per ASTM D7661-10.
    • Shelf-life testing results as per ASTM F1980-16.

    8. The sample size for the training set

    This section is not applicable. There is no "training set" as this is a physical medical device (personal lubricant) undergoing non-clinical performance testing, not a machine learning model.

    9. How the ground truth for the training set was established

    This section is not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200239

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-13

    (287 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide® Organix® Gel is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. The product is not compatible with polyurethane condoms.

    Device Description

    Astroglide® Organix ® Gel is non-sterile, translucent, colorless to slight yellow, water-based personal lubricant intended for penile and vaginal application to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms and is not compatible with polyurethane condoms. This product is for over-the-counter (OTC) use. Astroglide® Organix ® Gel's formulation consists of water, glycerine, xanthan gum, gellan gum, aloe barbabensis powder, chamomilla recutita (matricaria) flower extract, calendula officinalis flower extract, vaccinium macrocarpon (cranberry) fruit extract, passiflora incarnata flower extract, sodium benzoate, potassium sorbate, and citric acid. It is packaged in white LDPE tube with a screw on flip top cap comprised of polypropylene. This product is also provided in 2.5 ml foil packs.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding a personal lubricant, Astroglide® Organix® Gel. It is a not a study of an AI/ML medical device. Therefore, I cannot extract the information required using the provided text.

    The request asks for specific details related to the validation of an AI/ML medical device, such as acceptance criteria based on AI performance metrics (e.g., sensitivity, specificity), details about test sets (sample size, provenance, ground truth establishment, expert qualifications, adjudication methods), information on multi-reader multi-case (MRMC) studies, and details about training sets. These concepts are entirely absent from the provided document, which focuses on the chemical, physical, and biocompatibility testing of a personal lubricant to demonstrate its substantial equivalence to a predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171985

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-08-30

    (58 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide® O Oil Personal Lubricant & Massage Oil is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with condoms.

    Device Description

    Astroglide® O Oil Personal Lubricant & Massage Oil is non-sterile, light yellow, lightly coconut and ylang-ylang scented, and oil-based lubricant. All oils used in the subject formulation are certified organic, and the subject lubricant is a USDA certified organic final product. This product is not a spermicide or contraceptive. It is NOT compatible with natural rubber latex, polyurethane, or polyisoprene condoms.

    This product primary packaging is a PETE amber bottle with a screw on polypropylene flip-top cap. The bottle and cap constitute the device are labeled front and back and then packaged in a cardboard carton which constitutes the final packaging.

    AI/ML Overview

    The provided text is a 510(k) summary for the Astroglide® O Oil Personal Lubricant & Massage Oil. It details the device, its intended use, and its comparison to a predicate device. However, it does not describe a study involving an AI-powered device or software. Therefore, I cannot extract the information requested about acceptance criteria, study details, ground truth, or MRMC studies for an AI device from this document.

    The document focuses on non-clinical performance testing for a personal lubricant to demonstrate its safety and effectiveness, which is a different type of evaluation than what would be conducted for an AI/ML medical device.

    If the request was intended to ask about the non-clinical performance testing detailed in the document, here is a summary of that information:

    1. A table of acceptance criteria and the reported device performance for Astroglide® O Oil Personal Lubricant & Massage Oil (non-AI device):

    ParameterSpecification (Acceptance Criteria)Reported Device Performance
    ColorLight yellowLight yellow (as described)
    ClarityClearClear (as described)
    OdorCoconut and ylang-ylangCoconut and ylang-ylang (as described)
    Absence of particulate matterAbsentAbsent (as described)
    Viscosity60-120 cPWithin the specified range of 60-120 cP
    Total yeast/mold count<10 cfu/mL (USP <61>)<10 cfu/mL
    Total aerobic microbial count<100 cfu/mL (USP <61>)<100 cfu/mL
    Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicansAbsent (USP <62>)Absent
    Biocompatibility - CytotoxicityNon-cytotoxic (ISO 10993-5: 2009/(R)2014)Found to be non-cytotoxic
    Biocompatibility - SensitizationNon-sensitizing (ISO 10993-10: 2010/(R)2014)Found to be non-sensitizing
    Biocompatibility - Vaginal IrritationNon-irritating (ISO 10993-10: 2010/(R)2014)Found to be non-irritating
    Biocompatibility - Systemic ToxicityNon-systematically toxic (ISO 10993-11: 2006/(R)2010)Found to be non-systematically toxic
    Condom Compatibility (Natural rubber latex, polyisoprene, polyurethane)Compatible (ASTM D7661-10 - implied predicate compatibility)NOT Compatible (This is a key difference and limitation)
    Shelf LifeMaintained specifications over 18-month shelf-life (ASTM F1980-07)Maintained specifications over the duration of its shelf life

    Regarding the other points (#2-9), this information is not applicable to an AI device and is not contained in the document for this non-AI product.

    • Sample size for test set and data provenance: N/A for an AI device. For the lubricant, samples were tested per the specified ASTM and USP methods.
    • Number of experts and qualifications: N/A for an AI device. Testing was conducted according to standardized laboratory methods.
    • Adjudication method: N/A for an AI device.
    • MRMC comparative effectiveness study: N/A as this is not an AI device.
    • Standalone performance: N/A as this is not an AI device. Its performance is evaluated through material and biological testing.
    • Type of ground truth used: N/A for an AI device. The ground truth for this product is based on established scientific standards and analytical test results for chemical and biological properties.
    • Sample size for training set: N/A as this is not an AI device.
    • How ground truth for training set was established: N/A as this is not an AI device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1