510(k) Data Aggregation

    K Number
    K212364

    Validate with FDA (Live)

    Date Cleared
    2022-08-10

    (376 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BLT Dental Implant Systems are intended to be placed in the upper and lower jaw to support prosthetic devices and to restore a patient's chewing function. BLT Dental Implant systems are also indicated for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants through the corresponding components (abutments). In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.

    Device Description

    The proposed devices, BLT Dental Implant Systems, are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients. The proposed devices can also be used for immediate or early implantation following extraction or loss of natural teeth. The proposed device contains dental implant, healing cap and abutment. The dental implant system is available in two types, NC type and RC type.

    AI/ML Overview

    The request asks for acceptance criteria and details of the study proving these criteria are met. The provided text is a 510(k) Summary for a dental implant system. This type of submission focuses on demonstrating substantial equivalence to a predicate device, primarily through non-clinical testing and comparison of characteristics, rather than a clinical study with detailed acceptance criteria and performance metrics typically found for AI/ML devices. Therefore, much of the requested information (like effect size, ground truth, sample size for training/test sets, expert qualifications, and adjudication methods) is not applicable or unavailable from this document.

    However, I can extract information related to the device's performance based on non-clinical testing and the acceptance criteria implicitly defined by compliance with recognized standards and demonstrated equivalence to the predicate device.

    Here's the summary based on the provided document:

    Acceptance Criteria and Study Details for BLT Dental Implant System (K212364)

    The "acceptance criteria" for this device are implicitly defined by its compliance with various ISO and ASTM standards, and its demonstrated substantial equivalence to a legally marketed predicate device (K150388) and several reference devices. The "study" proving these criteria is a series of non-clinical tests and a comparison of technological characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicit via Standards/Equivalence)Reported Device PerformanceStudy that Proves this Performance
    Mechanical PerformanceNo significant difference in dynamic fatigue performance compared to predicate device. (Aligned with ISO 14801)"The test result does not show any significant difference." (between proposed device and predicate)Mechanical test performed according to FDA guidance and ISO 14801.
    BiocompatibilityMaterials are biocompatible and do not cause adverse effects. (Aligned with ISO 10993 standards)"The subject device is as safe... as the legally marketed predicate device..." as materials are identical to previously cleared device (K192274).Leveraged prior biocompatibility testing for K192274, as materials are identical in formulation, processing, and sterilization. Formal compliance with ISO 10993-1, -5, -10, -11 mentioned.
    SterilizationAchieves a Sterility Assurance Level (SAL) of $10^{-6}$. Maintain package integrity. (Aligned with ISO 11137-2)"Sterilized by irradiation to achieve a SAL of $10^{-6}$." "Vaccuum leak test and sterility test were provided to verify the package integrity."Sterilization method validated in accordance with ISO 11137-2. Package process is the same as K150388. Complies with ISO 11137-1, ISO 11607-1, -2, ASTM D3078-02(2013), ASTM F88/F88M-15, ASTM F1886/F1886M-16, ASTM F1929-15, ASTM F2096-11.
    Material CompositionComplies with relevant material specifications (Titanium alloys).Made of Pure Titanium (for implant), Titanium Alloy (for abutment/healing cap). Meets ASTM F136-13, ASTM F67-13.Chemical analysis leveraged from K150388 as surface modification is the same. Test reports for ASTM standards cited.
    Bacterial EndotoxinMeets bacterial endotoxin limits."Bacteria endotoxin limit were evaluated for the each lot device."USP <85> Bacterial Endotoxin Test.
    PackagingPackaging integrity for sterile barrier systems.Packaging system maintains sterility.Complies with ISO 11607-1:2019, ISO 11607-2:2019, ASTM D3078-02(2013), ASTM F88/F88M-15, ASTM F1886/F1886M-16, ASTM F1929-15, ASTM F2096-11 (Reapproved 2019), ASTM F1980-07(2011).
    Shelf Life5 years.5 years.(Details not explicitly stated in document, but assumed based on predicate equivalence and compliance with accelerated aging standards like ASTM F1980-07(2011) related to packaging.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for specific non-clinical tests. Non-clinical tests like mechanical testing (ISO 14801) typically involve a specific number of samples for destructive testing, but the exact count isn't provided in the summary.
    • Data Provenance: The tests were conducted by the manufacturer, Bio Concept Co., Ltd. and leverage data from previously cleared devices where materials or processes are identical. The document does not specify the country of origin of the raw test data (e.g., where the mechanical tests were physically performed) beyond the manufacturer's location in China. The data is retrospective in the sense that prior clearance data (K192274, K150388) is referenced for aspects like biocompatibility and surface analysis.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This is not applicable as this is a non-clinical submission for a physical medical device (dental implant system), not an AI/ML device relying on expert-annotated ground truth. The "ground truth" for non-clinical tests comes from physical measurements against established standards.

    4. Adjudication Method for the Test Set

    • Not applicable as this is a non-clinical submission for a physical medical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical dental implant system, not an AI/ML device, so no MRMC study was performed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical device, not an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for establishing device performance in this context is derived from:
      • Scientific Standards: Compliance with recognized international and national standards (ISO 14801 for mechanical properties, ISO 10993 for biocompatibility, ISO 11137-2 for sterilization, ASTM standards for materials and packaging).
      • Predicate Device Equivalence: Direct comparison of technological characteristics, materials, and intended use to an existing legally marketed device (K150388) and numerous reference devices.
      • Laboratory Testing: Results from mechanical tests, sterilization validation, bacterial endotoxin tests, and packaging integrity tests.

    8. The Sample Size for the Training Set

    • Not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192274

    Validate with FDA (Live)

    Date Cleared
    2020-06-18

    (301 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BV Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.

    Device Description

    The proposed devices, dental implant system consists of dental implants, abutments and healing cap, it is used in one or two-stage dental implant placement and restoration. The dental implants are available in Mini and Regular two types. The implant is made from titanium and the surface of the implants are treated with Sandblasting and Acid etching. The implants are provided in lengths from 7.0-15.0 and in diameter from 3.7 to 6.8.

    The abutments in this submission include transfer abutment, temporary abutment, multi abutment, multi angled abutment, equator abutment and Esthetic-low Temporary Cylinder. The abutment is also available in mini and regular two types to match with the corresponding implant. The abutment is made from titanium alloy.

    The healing cap in this submission include cover screw, healing abutment and equator healing abutment.

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter and summary for the BV Dental Implant System. It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a study to prove acceptance criteria for a novel AI/ML-driven device.

    Therefore, the document does not contain the specific information required to answer the questions regarding acceptance criteria and a study that proves a device meets those criteria for an AI/ML device.

    Specifically, the document does not include the following information relevant to AI/ML device evaluation:

    1. A table of acceptance criteria and the reported device performance: The tables provided are for comparison of the proposed dental implant system with predicate devices, outlining product codes, indications for use, and physical characteristics. There are no performance metrics related to an AI/ML component.
    2. Sample size used for the test set and data provenance: Not applicable as there is no AI/ML component or a test set described.
    3. Number of experts used to establish the ground truth for the test set and their qualifications: Not applicable.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of human readers improvement with AI vs. without AI assistance: Not applicable, as this is not an AI/ML device. The document explicitly states "No clinical study is included in this submission." (Page 6).
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used: Not applicable.
    8. The sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    Instead, the document details non-clinical tests conducted to demonstrate substantial equivalence for a dental implant system:

    • Biocompatibility test: Leveraged previous tests on K150388 (same applicant) for the implant. Cytotoxicity testing was performed for the abutment, showing no cytotoxicity.
    • Surface test: Leveraged previous tests on K150388.
    • Bacteria Endotoxin Test: Conducted per USP 85, with results showing endotoxin less than 20EU per device.
    • Mechanical test: Performed on the proposed and predicate device, showing no significant difference.
    • Sterilization: Radiation sterilization for implants (validated per ISO 11137-1 and -2) and steam sterilization for abutments and healing caps (validated per ISO 17665-1 and -2).

    The document is a traditional 510(k) submission for a physical medical device (dental implant system), not an AI/ML device, and therefore the requested information is absent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150388

    Validate with FDA (Live)

    Date Cleared
    2015-11-12

    (267 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dental implant systems are intended to be placed in the upper and lower jaw to support prosthetic devices and to restore a patient's chewing function. Dental implant systems are also indicated for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants through the corresponding components (abutments). In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.

    Device Description

    The proposed devices, Dental Implant Systems, are integrated systems of endosseous dental implants, which are intended to support prosthetic devices.

    The systems are mainly available in two models which are Tissue Level (TL) and Bone Level (BL), each model consists of the following components: (1) Dental Implants, (2) Abutments, (3) Occlusal Screws and (4) Healing Caps. Each component is available in various configuration. In addition, the systems also consist of various specific surgical instruments.

    Abutment models for TL dental implants include: Coping, Solid abutment, Retentive anchor abutment, Screw-retained abutment, Cementable abutment, Angled abutment.

    Abutment models for BL dental implants include: Healing Abutment, Anatomic abutment, Meso abutment, Cementable abutment, Multi-base abutment.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a Medical Device, specifically a "Dental Implant System." This document primarily focuses on demonstrating substantial equivalence to a predicate device for regulatory approval, rather than presenting a study of device performance against pre-defined acceptance criteria in the context of clinical or AI-based performance metrics.

    Therefore, the document does not contain the information requested for acceptance criteria and a study that proves the device meets those criteria, especially in relation to AI/algorithm performance. It is a regulatory submission demonstrating mechanical, material, and safety equivalence.

    Here's a breakdown of why the requested information cannot be extracted from this document:

    • No AI/Algorithm Component: There is no mention of any AI, machine learning, or algorithm being part of this "Dental Implant System." The device is a physical dental implant.
    • Focus on Substantial Equivalence: The entire document is geared towards proving that the proposed device is "substantially equivalent" to a legally marketed predicate device (STRAUMANN DENTAL IMPLANT SYSTEM). This is a regulatory pathway, not a clinical trial to demonstrate new performance metrics against acceptance criteria for a novel AI.
    • No Clinical Study: Section 7 explicitly states: "No clinical study is included in this submission." This means there is no data from human subjects or a clinical setting to evaluate performance.
    • Non-Clinical Test Conclusion: Section 6 details non-clinical tests. These tests are primarily for material properties, sterility, biocompatibility, and mechanical durability (e.g., dynamic fatigue test). These are engineering and material science tests, not performance studies for a diagnostic or AI-driven medical device.

    Therefore, it is impossible to complete the requested table or answer the specific questions about acceptance criteria, ground truth, expert adjudication, or MRMC studies, as these concepts are not applicable to the content of this 510(k) submission for a physical dental implant.