510(k) Data Aggregation

    K Number
    K181737

    Validate with FDA (Live)

    Device Name
    Raycell MK1
    Date Cleared
    2019-01-02

    (184 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To irradiate cellular blood products to inactivate T-lymphocytes in order to prevent Graft Versus Host Disease.

    Device Description

    The Raycell Mk1 x-ray blood irradiator consists of one cabinet containing one x-ray tube, one highvoltage power supply, one radiation shielding chamber, control electronics, an internal cooling system, two removable shielding access panels to the irradiation chamber, a touch-based Graphical User Interface, and additional operator controls. The operator places the blood products to be irradiated in the canister or syringe holder, opens the door, places the canister or syringe holder on the turntable, closes the door and indicates the irradiation cvcle on the Graphical User Interface. The Operator then uses the Operator controls to start the irradiation cycle. The design of this device is substantially equivalent to the predicate device.

    AI/ML Overview

    This document describes the Raycell MK1, an x-ray blood irradiator. However, the provided text does not contain acceptance criteria or study details in the format requested, relating to AI/device performance metrics, sample sizes, expert ground truth, or MRMC studies.

    The document is a 510(k) summary for the Raycell MK1, which is a medical device for irradiating cellular blood products. The primary aim of a 510(k) submission is to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to prove specific performance metrics against pre-defined acceptance criteria in the manner often associated with AI/software performance studies.

    Here's a breakdown of what is and isn't in the provided text, in relation to your request:

    What is present (partial information):

    • Device Name: Raycell MK1
    • Intended Use: To irradiate cellular blood products to inactivate T-lymphocytes in order to prevent Graft Versus Host Disease.
    • Predicate Device: Raycell MK2 (K161324)
    • Safety & Effectiveness Statement: "The safety of the Raycell is equivalent or better than the predicate device." and "Validation testing demonstrated that the device is as safe and effective as the predicate device."
    • Compliance Standards: IEC 60601-1-2:2014-02 and IEC 60601-1:2012-08 (These are general medical electrical equipment safety and performance standards, not specific performance metrics for the irradiation itself).
    • Testing Method (general): "The performance of the device was tested against a set of functional specifications in an environment that simulated, as much as possible, the actual operating environment."

    What is NOT present in the provided text, and therefore cannot be answered from this document:

    • A table of acceptance criteria and reported device performance (in terms of specific metrics like sensitivity, specificity, accuracy, etc.): The document only generally states that "validation testing demonstrated that the device is as safe and effective as the predicate device." It does not provide quantitative performance metrics.
    • Sample size used for the test set and data provenance: No information on the number of samples (e.g., blood bags) tested, or where the data came from.
    • Number of experts used to establish the ground truth for the test set and their qualifications: This device is not an AI/diagnostic device that typically relies on expert interpretation for ground truth. Ground truth for an irradiator would be related to the actual dosage delivered and its biological effect.
    • Adjudication method: Not applicable for this type of device and study description.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, and its effect size: Not applicable/not mentioned. This is a physical device, not an AI assistance tool for human readers.
    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable/not mentioned.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): While the device's function is to inactivate T-lymphocytes, the document doesn't detail how the effectiveness of this inactivation was rigorously measured or what constituted "ground truth" for its performance validation.
    • The sample size for the training set: Not applicable, as this is not an AI/machine learning device requiring a training set in that context.
    • How the ground truth for the training set was established: Not applicable.

    Conclusion based on the provided text:

    The provided 510(k) summary focuses on demonstrating substantial equivalence to a predicate device by describing design changes and compliance with general safety standards. It does not provide the specific performance metrics, study designs (like MRMC), sample sizes, or ground truth details typically associated with the type of acceptance criteria and studies you've asked about for an AI-powered diagnostic/analysis device. The "performance of the device was tested against a set of functional specifications," but these specifications and their results are not detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161324

    Validate with FDA (Live)

    Device Name
    Raycell Mk2
    Date Cleared
    2016-08-12

    (93 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Raycell X-ray Blood Irradiator is intended for use in the irradiation of blood and blood products (packaged in transfusion bags) to inactivate T-lymphocytes for the prevention of Graft Versus Host Disease according to applicable FDA, AABB, Health Canada, and European guidelines.

    The Raycell X-ray Blood Irradiator is also intended for use in the irradiation of intra-operatively salvaged blood for cancer patients undergoing surgery to assist in the prevention of metastasis.

    Device Description

    The Raycell is a self-contained shielded cabinet X-Ray Blood Irradiator. The device consists of a lead shielded chamber containing two vertically opposed x-ray tubes with provision for a sample canister holder between them, dual power supplies and a operator control panel. The operator places the blood products to be irradiated in the canister, opens the drawer, places the canister in the holder, closes the drawer and starts the irradiation cycle at the Operator Panel. The design of this device is substantially equivalent to the predicate device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Raycell Mk2 X-Ray Blood Irradiator. This document focuses on demonstrating substantial equivalence to a legally marketed predicate device (Raycell, K051065) rather than presenting a clinical study to prove the device meets specific acceptance criteria based on patient outcomes or diagnostic performance.

    Therefore, many of the requested elements for describing an acceptance criteria study (like sample size for test sets, number of experts, adjudication methods, MRMC studies, standalone performance with ground truth) are not applicable or not present in this type of regulatory submission.

    However, I can extract the information related to the device's technical specifications and safety/effectiveness testing against functional specifications, as this is the "study" referred to in the document.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a direct table of quantitative acceptance criteria for clinical performance or diagnostic metrics. Instead, it focuses on demonstrating that the modified aspects of the Raycell Mk2 are equivalent or better than the predicate device's performance in terms of safety and technical functionality.

    Acceptance Criteria (General)Reported Device Performance
    Safety:Safety:
    Compliance with IEC 60601-1-2 (EMC)Complies with IEC 60601-1-2: Edition 3.0 2007-03
    Compliance with IEC 60601-1 (Basic Safety & Essential Performance)Complies with IEC 60601-1: Edition 3.0 2005 + CORR. 1 (2006) + CORR. 2 (2007)
    External radiation fields (compared to predicate)Decreased from the predicate design.
    Electrical interlocksMaintained.
    New electro-mechanical safety interlock (for sliding drawer)Added to prevent opening during x-ray generation.
    Effectiveness/Functionality:Effectiveness/Functionality:
    Irradiate blood and blood products (to inactivate T-lymphocytes)Uses the same x-ray tubes and high voltage generators as the predicate. Operates at a 25% higher power (4kW instead of 3.2kW).
    Software control systemNo changes to the control system software from the predicate design.
    Dose deliveredNot explicitly stated, but the purpose of the device (inactivation of T-lymphocytes) implies it delivers a sufficient dose. The optional digital traceability has no impact on radiation times or dose delivered.
    Ability to accommodate varying sample sizesAvailable in 3.5L or 2L sample canister configurations (predicate was 1.5L).
    Improved manufacturability and service accessShielded irradiation chamber modified to improve manufacturability and access for service and maintenance.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified as a number of patient cases or blood samples in the context of a clinical study. The "test set" refers to the device itself undergoing engineering and safety validation.
    • Data Provenance: The testing was likely conducted in a controlled laboratory or manufacturing environment by the manufacturer (Best Theratronics, Canada). It is not clinical data from patients or blood products in a real-world setting, but rather engineering verification and validation testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. As this is an engineering and safety validation of an irradiator, not a diagnostic or AI-based device, there were no clinical experts establishing "ground truth" for patient data. The "ground truth" here would be the physical and electrical parameters of the device meeting established engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. No clinical adjudication method was used. The device's performance was evaluated against technical specifications and regulatory standards by engineers and quality assurance personnel.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is an X-ray blood irradiator, not an AI-assisted diagnostic device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a standalone medical device (X-ray blood irradiator). The "standalone performance" refers to the device's technical operation as described in the "Safety & Effectiveness" section against functional specifications and regulatory standards. There is no "algorithm only" performance to evaluate in the context of AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the device's acceptable performance was based on engineering specifications, international safety standards (IEC 60601 series), and comparison to the predicate device's proven functionality and safety. The device's ability to irradiate blood effectively for its stated purpose is an established physical principle and was likely assumed to be met based on using the same core X-ray technology as the predicate, with increased power.

    8. The sample size for the training set:

    • Not Applicable. This device does not use machine learning or AI models that require a "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for an AI model, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K150191

    Validate with FDA (Live)

    Device Name
    GammaBeam 500
    Date Cleared
    2015-04-17

    (79 days)

    Product Code
    Regulation Number
    892.5750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.

    Device Description

    The GammaBeam 500 is a radiotherapy treatment unit with a cobalt-60 radiation delivery system that is designed to be a dedicated, standalone device for performing total body irradiation (TBI). The device consists of a source head, main frame, base, and controls. The design of this device is substantially equivalent to the predicate device Theratron Phoenix.

    The GammaBeam 500 is aimed to deliver total body irradiation to patients by generating a large fixed radiation treatment field at an extended Source - to - Skin Distance (SSD). This is a common radiotherapy technique performed by many clinics worldwide. A dedicated device improves the safety, effectiveness and clinical workflow of these treatments.

    The unit will be operated in a similar manner as the predicate and has the same redundant mechanisms for safety, auxiliary source drawer lock, secondary and manual source return, emergency stop switches on the unit and control console, source position indicator, and operational and inhibit state interlocks.

    AI/ML Overview

    The provided document is a 510(k) summary for the GammaBeam 500, a Cobalt Teletherapy unit. It primarily focuses on demonstrating substantial equivalence to a predicate device (Theratron Phoenix) rather than detailing specific clinical study results or acceptance criteria for a novel device's performance metrics.

    Based on the information provided, here's an analysis of the acceptance criteria and supporting studies as much as can be discerned:

    The document does not present specific acceptance criteria in the form of quantitative performance metrics (e.g., sensitivity, specificity, accuracy) for a device's diagnostic or predictive capabilities in the typical sense of a medical imaging or AI device. Instead, the acceptance criteria relate to regulatory compliance, safety, and functional equivalence to a predicate device.

    The "study that proves the device meets the acceptance criteria" is a collection of non-clinical tests and verification/validation activities.

    1. A table of acceptance criteria and the reported device performance

    The document doesn't provide a table of acceptance criteria for diagnostic/predictive performance. Instead, it lists standards for safety and EMC, and functional specifications. The "reported device performance" is the conclusion that the device complies with these standards and performs as intended.

    Acceptance Criteria (Category)Specific Standard/RequirementReported Device Performance
    Safety & EMCIEC 60601-1:2013 Edition 3.1 (Medical electrical equipment - General requirements for basic safety and essential performance)System complies with the standard for safety.
    IEC 60601-1-2:2007 Edition 3.0 (Medical electrical equipment Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests)System complies with the standard for EMC.
    EN 60601-2-11:2011 Edition 3.0 (Medical electrical equipment -- Part 2: Particular requirements for the safety of gamma beam therapy equipment)System complies with the standard for safety of gamma beam therapy equipment.
    EN 61217 (2014) (Radiotherapy equipment Coordinates, movements and scales)System complies with the standard.
    SoftwareCompliance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Software level of concern: "major" - failure could result in serious injury or death).Software verification and validation testing were conducted and documentation was provided, demonstrating that the software performs as intended.
    Mechanical PerformanceFunctional specifications (simulated actual operating environment)Mechanical validation of the main unit and patient couch, physics validation, and simulated use testing were performed. The device performs as intended in the specified use conditions.
    Substantial EquivalenceThe device is "as safe and effective as the predicate device that is currently marketed for the same intended use." This is the overarching "acceptance criterion" for a 510(k).The non-clinical data support the safety of the device and the hardware and software verification and validation support that the GammaBeam 500 performs as intended in the specified use conditions, demonstrating substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document refers to "non-clinical data" and "testing" rather than a clinical trial with a "test set" of patient data. The tests performed are:

    • Electrical safety and electromagnetic compatibility (EMC) testing.
    • Software Verification and Validation Testing.
    • Mechanical Testing (mechanical validation, physics validation, simulated use testing).

    These are engineering and software validation activities, not clinical studies on patient samples. Therefore, the concepts of "sample size for a test set" and "data provenance" in the context of patient data do not apply here. The testing was conducted on the GammaBeam 500 unit itself, and its components (main unit and patient couch).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the document does not describe a clinical study with a "test set" relying on expert interpretation for ground truth. The "ground truth" for these engineering tests would be derived from the specifications and expected physical/electrical performance defined by engineers and physicists.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there is no clinical test set requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for radiation therapy delivery, not an AI diagnostic or assistance tool. Therefore, MRMC studies and "human readers improve with AI" metrics are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a hardware device for radiation therapy. While it has a control system (software), its performance validation focuses on its safety and functional equivalence in delivering radiation, not on standalone interpretive algorithmic performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests would have been established by:

    • Standards for Safety and EMC: The defined requirements and limits specified in the referenced IEC and EN standards.
    • Software Verification and Validation: Software requirements specifications, design documents, and expected outputs/behaviors.
    • Mechanical and Physics Validation: Engineering specifications, design documents, and expected physical behavior/radiation output measurements based on established physics principles and industry best practices for teletherapy units.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a "training set" of data in the typical sense.

    9. How the ground truth for the training set was established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142219

    Validate with FDA (Live)

    Device Name
    GammaBeam 100
    Date Cleared
    2014-11-13

    (93 days)

    Product Code
    Regulation Number
    892.5750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.

    Device Description

    The GammaBeam 100 is a radiotherapy treatment unit with a cobalt-60 radiation delivery system. The device consists of a source head, collimator, gantry, main frame, base, controls and a pendulum or beamstopper style counterweight. The design of this device is substantially equivalent to predicate device Theratron Phoenix.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of device performance metrics like sensitivity, specificity, accuracy, or any statistical measures.

    The document is a 510(k) premarket notification letter from the FDA to Best Theratronics Limited regarding their "Gammabeam 100" device (a Radionuclide radiation therapy system). It primarily states that the device has been found substantially equivalent to a legally marketed predicate device (Theratron Phoenix).

    The document mentions:

    • Safety & Effectiveness: It states that "The safety of the GammaBeam 100 is equivalent or better than the predicate device" and that "Validation testing demonstrated that the device is as safe and effective as the predicate device."
    • Compliance with Standards: It lists several IEC and EN standards that the device is designed to comply with (e.g., IEC 60601-1-2, IEC 60601-1, EN 60601-2-11, EN 61217).
    • Functional Specifications: It vaguely mentions "The performance of the device was tested against a set of functional specifications in an environment that simulated, as much as possible, the actual operating environment."

    However, it does not provide:

    1. A table of acceptance criteria and reported device performance (with specific metrics).
    2. Sample sizes or data provenance for a test set.
    3. Information about experts, ground truth establishment, or adjudication methods for a test set.
    4. Details on a multi-reader multi-case (MRMC) comparative effectiveness study or related effect sizes.
    5. Details on a standalone algorithm performance study.
    6. The type of ground truth used (beyond implying functionality and safety against standards).
    7. Sample size for a training set or how its ground truth was established.

    This document is a regulatory approval letter based on substantial equivalence, not a detailed study report providing performance metrics against quantitative acceptance criteria.