Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K231513

    Validate with FDA (Live)

    Device Name
    Oral Matrix
    Date Cleared
    2024-02-16

    (267 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Oral Matrix is intended for use in extraction sockets only to contain or prevent migration of graft material. The device is supplied sterile and intended for one time use.

    Device Description

    The Oral Matrix consists of layered sheets of bioabsorbable extracellular collagen membrane matrix derived from porcine Small Intestinal Submucosa (SIS). These sheets are freeze-dried, packaged in a Tyvek pouch, and sterilized using ethylene oxide to achieve a SAL of 10-6.

    The Oral Matrix is available in eighteen different models, the differences between the models are shown in the table below.

    ModelSize (cm)ToleranceThickness
    SIS-ORP-4L-1×11×1
    SIS-ORP-4L-2×22×2
    SIS-ORP-4L-2×32×3
    SIS-ORP-4L-3×33×3
    SIS-ORP-4L-4×34×3
    SIS-ORP-4L-6×66×6
    SIS-ORP-4L-1.5×1.51.5×1.5±0.2cm0.05~0.13mm
    SIS-ORP-4L-3×53×5±0.2cm0.05~0.13mm
    SIS-ORP-4L-2.5×1.52.5×1.5±0.2cm0.05~0.13mm
    SIS-ORP-4L-5×45×4±0.2cm0.05~0.13mm
    SIS-ORP-4L-4×64×6±0.2cm0.05~0.13mm
    SIS-ORP-4L-7×77×7±0.2cm0.05~0.13mm
    SIS-ORP-4L-3×3.53×3.5±0.2cm0.05~0.13mm
    SIS-ORP-4L-3×73×7±0.2cm0.05~0.13mm
    SIS-ORP-4L-4×44×4
    SIS-ORP-4L-5×55×5
    SIS-ORP-4L-6×86×8
    SIS-ORP-4L-8×88×8
    AI/ML Overview

    The provided document describes the "Oral Matrix" device and its FDA 510(k) summary, which often includes information on acceptance criteria and supporting studies. However, this specific document focuses on demonstrating substantial equivalence to a predicate device rather than providing explicit acceptance criteria with numerical targets for clinical performance metrics or a detailed standalone study for human-in-the-loop performance. Instead, it relies on non-clinical and animal studies to justify equivalence based on product performance and composition.

    Here's an analysis of the provided information concerning acceptance criteria and studies:

    1. A table of acceptance criteria and the reported device performance:

    The document does not explicitly list quantitative acceptance criteria in a table format with specific performance targets for the device's intended clinical use (e.g., success rate of socket preservation, bone regeneration metrics). Instead, it presents various non-clinical and animal study findings which implicitly serve as evidence that the device performs similarly to the predicate device or meets regulatory standards.

    Implicit "Acceptance Criteria" through Equivalence and Standards Compliance:

    CategoryAcceptance Criterion (Implicit)Reported Device Performance
    Product Performance ComparisonBurst Strength: Similar to predicate device. Suture Retention Strength: Similar to predicate device. Tensile Strength: Similar to predicate device. Physical Performance: Similar to predicate device."The test results demonstrate the subject device has the similar product performance as those of the predicate device." (No specific numerical values provided for either subject or predicate)
    Product Composition ComparisonTotal Protein Content: Similar to predicate device. Total Sugar Content: Similar to predicate device. Lipid Content: Similar to predicate device. DNA Residue: Similar to predicate device. Peroxyacetic Acid Residue: Similar to predicate device."The test results demonstrate the subject device has the similar product composition as those of the predicate device." (No specific numerical values provided for either subject or predicate)
    Virus InactivationSum of log clearance of DNA and RNA virus from inactivation processes is at least 6 logs greater than anticipated virus concentration in unprocessed animal source (according to ISO 22442-3:2007)."The test results demonstrate that the virus inactivation processes of the subject device reduce more than 6 logs virus."
    Package Integrity TestingSeal Strength: No less than 1.5N/15mm (according to ASTM F88/F88M-15). Dye Penetration: No dye penetration across the package (according to ASTM F1929-15). ASTM F1886: Compliance with standard."The test results demonstrate the sterility maintenance package of the subject device meets the requirements of standards and thus provides sterility maintenance for the finished device." (No specific values provided, but compliance stated)
    Shelf Life TestingDevice safety and effectiveness maintained for the claimed shelf life (24 months) based on package integrity and product performance."The test results demonstrate the subject device can claim 24 month as shelf life."
    BiocompatibilityNo negative impacts from materials used (Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute systemic toxicity, Pyrogen, Implantation, Subchronic systemic toxicity, Bacterial reverse mutation, Mouse lymphoma TK assay, Hemolysis, Complement activation, Partial thromboplastin time, Chronic system toxicity, Chemical Characterization)."The test results demonstrate there are no negative impacts from the materials that are used in the subject device." Specific results include: No Cytotoxicity; No Sensitization; No Intracutaneous Reactivity; No Acute Systemic Toxicity; Non irritant after Implantation; No Subchronic Systemic Toxicity; No Bacterial Reverse Mutation; No mutagenic to mouse lymphoma cell; No Hemolysis; Pyrogen meet the acceptable USP limits; No Complement Activation; No Partial Thromboplastin; No Chronic System Toxicity; No unacceptable risks associated with the extract based on the chemical characterization evaluation. (All indicate successful meeting of acceptance criteria for benign biological response.)
    Animal Study (Performance/Safety)Ability to preserve the teeth extraction socket, comparable to predicate device and superior to blank control group."The results demonstrated that the subject device has ability to preserve the teeth extraction socket compared with predicate device." and "it was similar compared with the predicate device."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Test Sets:
      • Non-Clinical (Product Performance, Composition, Package Integrity, Shelf Life, Biocompatibility): The document doesn't specify sample sizes for these tests. The data provenance is not explicitly stated as country of origin, but the sponsor is Beijing Biosis Healing Biological Technology Co., Ltd. from China, and the testing was likely conducted in a setting compliant with international standards (ISO, ASTM). These are generally prospective tests conducted specifically for the submission.
      • Virus Inactivation: Not specified, but likely an in-vitro study designed to demonstrate log reduction, which is a prospective test.
      • Animal Study: The sample size for the canine model is not specified. The study was performed in an animal model, which is a prospective study. Data provenance in terms of country is not specified, but usually, animal studies are conducted by or for the submitting company.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable as the document describes non-clinical and animal studies, not studies involving human expert interpretation for ground truth. Ground truth for these types of tests is established through laboratory methods and histopathological analysis (for the animal study), not expert consensus on human data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable as the studies described (non-clinical, animal) do not involve human interpretation or adjudication processes typical of clinical diagnostic studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or mentioned. The device is a bone grafting material, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not Applicable. The device is not an algorithm, but a physical bone grafting material. There is no "algorithm only" performance to evaluate. The performance evaluations described are for the physical properties and biological effects of the material itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Non-Clinical (Product Performance, Composition, Package Integrity, Shelf Life): Ground truth is established by physical and chemical measurements against established standards (e.g., ASTM, ISO guidelines) and comparative data from the predicate device.
    • Virus Inactivation: Ground truth is established by virological assays to determine viral load reduction.
    • Biocompatibility: Ground truth is established by in vitro and in vivo biological assays (e.g., cytotoxicity, sensitization, implantation responses) adhering to ISO 10993 series standards.
    • Animal Study: Ground truth for socket preservation would be established by histopathological examination of the extracted socket sites, potentially including measurements of bone formation, reduction in ridge resorption, and inflammation, compared to control groups.

    8. The sample size for the training set

    Not Applicable. The device is a bioabsorbable extracellular collagen membrane matrix, not a machine learning or AI algorithm. Therefore, there is no "training set" in the context of an AI model.

    9. How the ground truth for the training set was established

    Not Applicable. As explained above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203622

    Validate with FDA (Live)

    Device Name
    Medical Gloves
    Date Cleared
    2021-08-20

    (252 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

    Device Description

    The proposed device is a powder free medical glove. The device is made from a nitrile latex compound, blue in color. The device meets the requirements of ASTM D6319-19: Standard specification for Nitrile Examination Gloves for Medical Application. The device is available in five sizes, which are XS, S, M, L, XL, it could be selected by the user depended on size of hand. The different between each size is just in the dimension. The device can be provided in non-sterile.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for "Medical Gloves" (K203622). It describes the device's characteristics, its intended use, and comparative testing against a predicate device (K180467) to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device Under Evaluation: Medical Gloves (powder-free patient examination glove, nitrile latex compound, blue color)

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    Freedom from Holes - ASTM D5151Detection of holes that allow water leakageDo not show droplet, stream or other type of water leakage (Meets ASTM D5151-19, AQL 2.5 Inspection, Level G-1, Accept at 5 failures, Reject at 6 failures).No leakage (Meets ASTM D5151-19, AQL 2.5 Inspection, Level G-1, Accept at 5 failures, Reject at 6 failures).
    Physical dimension - ASTM D6319Evaluate the glove physical dimensionLength: >220 mm (XS and S sizes), >230 mm (M, L and XL sizes) Width (±10mm): XS = 70mm, S = 80mm, M = 95mm, L = 110mm, XL = 120mm Thickness at Finger (mm): All Sizes ≥ 0.05 mm Thickness at Palm: All Sizes ≥ 0.05 mmLength: Larger than 220mm for XS and S size, Larger than 230mm for M, L and XL sizes Width: XS: within 70±10mm, S: within 80±10mm, M: within 95±10mm, L: within 110±10mm, XL: within 120±10mm Thickness: Larger than 0.05mm (at finger and palm)
    Physical requirement - ASTM D412Evaluate the physical requirement (Tensile Strength & Ultimate Elongation)Before Aging (Min): Tensile strength: 14Mpa, Ultimate elongation: 500% After Aging (Min): Tensile strength: 14Mpa, Ultimate elongation: 400%Before aging: Larger than 14Mpa and 500% After aging: Larger than 14Mpa and 400%
    Powder residue - ASTM D6124Evaluate the residue powderLess than 2.0mg per glove.Less than 2.0mg per glove.
    Skin sensitization - ISO 10993-10Evaluated for the potential to cause delayed dermal contact sensitizationMagnusson and Kligman grade shall be less than control group.No skin sensitization.
    Skin irritation - ISO 10993-10Evaluated for the potential to cause skin irritationNo significant reaction than the control group.No skin irritation.
    Acute toxicity - ISO 10993-11Evaluated for acute systemic toxicityNo significant reaction than the control group. Animal death or abnormal behavior no more than two animals. Body weight loss greater than 10% no more than 3 animals. (Note: Predicate device did not explicitly provide this data, but the proposed device showed no adverse effect.)No acute toxicity. (Implies meeting all specified criteria).

    Study that Proves the Device Meets the Acceptance Criteria (Non-Clinical Testing):

    The document explicitly states: "Non clinical tests were conducted to provide data that the proposed devices met the acceptance criteria or the specification for the test methodology or standard shown below. The test results demonstrated that the proposed devices comply with the following standards."

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: The document does not explicitly state the specific number of gloves or animals used for each test. However, it references specific ASTM and ISO standards (e.g., ASTM D5151-19, which specifies AQL levels and inspection levels for Freedom from Holes). These standards define appropriate sample sizes for testing compliance. For instance, ASTM D5151-19, AQL 2.5, Inspection Level G-1 means the sample size is determined by the lot size, but for the specifically mentioned acceptance/rejection criteria (Accept at 5 failures, Reject at 6 failures), it implies a certain sample size bracket within the standard.
    • Data Provenance: The document does not specify the country of origin where the non-clinical tests were performed. It states the sponsor is Beijing Biosis Healing Biological Technology Co., Ltd. from China, so it's reasonable to infer the tests were conducted in or overseen by this entity. The data is retrospective in the sense that the tests were performed, and the results are being reported as part of a premarket submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • This section talks about the establishment of ground truth for a test set, typically relevant for AI/ML device evaluations. This document describes the evaluation of a physical medical device (gloves) against established performance standards. Therefore, the "ground truth" is based on objective, standardized test methods and their defined acceptance criteria. It is not based on expert consensus interpreting subjective data.
    • The tests themselves are performed by trained personnel in a laboratory setting, but the document does not specify the number or qualifications of these individuals, as it's not a multi-reader study.

    4. Adjudication Method for the Test Set

    • Not applicable. This is not a study involving human readers or subjective interpretations requiring adjudication. The tests are objective and based on physical measurements and chemical analyses as per established standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • Not applicable. This is not an AI/ML device that assists human readers. It is a physical medical device (gloves).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This is not an algorithm or AI device. The "standalone" performance refers to the device's ability to meet the specified performance criteria as tested in isolation according to the listed ASTM and ISO standards. The results table clearly indicates that the device met these criteria.

    7. The Type of Ground Truth Used

    • The ground truth used for this device's evaluation is primarily objective, standardized test results based on established international (ISO) and national (ASTM) consensus standards. These standards define measurable physical, chemical, and biological properties, along with clear acceptance/rejection criteria. For biocompatibility tests (skin sensitization, irritation, acute toxicity), the ground truth is established by the biological responses observed in the test models (e.g., animal models) according to the protocols defined in ISO 10993.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device that requires a "training set." The device itself is a manufactured product, not a learned algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Circular Stapler has application throughout the alimentary tract to create end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic surgeries.

    The Disposable Hemorrhoidal Circular Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoid disease.

    The Disposable Linear Stapler and Reloads can be applied in abdominal, thoracic and pediatric surgical procedures for transection or resection of tissue.

    The Disposable Linear Cutter Stapler and Reloads can be applied in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.

    The Disposable Endoscopic Linear Cutter Stapler and Reloads has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transaction and creation of anastomosis.

    Device Description

    Disposable Circular Stapler places a double staggered, circular row of titanium staples upon activation, which was achieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resect by the circular knife, and then a circular anastomosis is created. The stapler are available in 25mm, 28mm, 29mm and 32mm five specifications. The staple is available in 4.5mm and 4.8mm two different heights.

    Disposable Endoscopic Linear Cutter Stapler and Unit places two, triple-staggered rows of titanium staples and simultaneously divides the tissue from central line. The devices are available in 30mm, 45mm and 60mm three specifications. Reloads are available in 2.5mm, 4.0mm and 4.8mm four staple sizes to accommodate various tissue thicknesses. The device may be reloaded and fired up to 25 times in a single procedure.

    Disposable Hemorrhoidal Stapler is a set of instruments that place a double staggered, circular row of titanium staples. Immediately after the formation of staples, the circular knife blade resects the excess of compressed mucosa. The stapler are available in 32mm, 33mm three specifications. The staple size is 3.8mm. It cannot be reloaded.

    Disposable Linear Cutter Staplers and Reload places two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The devices are available in 55mm, 60mm, 75mm, 80mm and 100mm five specifications. Reloads are available in two staple sizes to accommodate various tissue thicknesses: 3.8mm and 4.8mm. It may be reloaded and fired up to 8 times in a single procedure.

    The Disposable Linear Stapler and Reload places a double staggered row of titanium staples and is available in 30mm, 45mm, 60mm four specifications for use in various applications. Two staple sizes (3.5mm and 4.8mm) are available to accommodate various tissue thicknesses. It may be reloaded and fired up to 8 times in a single procedure.

    AI/ML Overview

    The provided document is a 510(k) Summary for various disposable staplers and reload devices. It describes non-clinical tests to verify performance but does not include any information about studies involving acceptance criteria related to AI/human reader performance, ground truth establishment by experts, or any MRMC comparative effectiveness studies.

    Therefore, most of the requested information cannot be extracted from this document, as it pertains to medical device regulatory submission for staplers, not AI medical devices.

    However, I can provide the following based on the Non-Clinical Test Conclusion section:

    1. A table of acceptance criteria and the reported device performance

    The document mentions several tests and standards the device complies with, which would imply passing these as acceptance criteria. It doesn't provide specific numerical acceptance criteria or performance metrics for each test, but generally states that the device "met all design specifications" and "complies with the following standards".

    Test/StandardReported Device Performance (Implied Acceptance)
    ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.Device complies with the standard.
    ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.Device complies with the standard.
    ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivityDevice complies with the standard.
    USP42-NF37 <85> Bacterial Endotoxins TestsDevice complies with the standard.
    ASTM F 88/F88M-15 Standard test method for seal strength of flexible barrier materials;Device complies with the standard.
    ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye PenetrationDevice complies with the standard.
    ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityDevice complies with the standard.
    Ex-vivo tissue test (porcine stomach and intestine tissue) - Pressure Rest, Closed Staple Dimension Test, Staple Formation Test, Force Required to Fire Stapler Test.Test results demonstrated the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device.
    Jugular vein test (porcine model) - Burst pressure, closed staple height and staple formation.Test results demonstrated the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any of the non-clinical tests.
    • Data Provenance: The ex-vivo and jugular vein tests were conducted on "porcine stomach and intestine tissue" and a "porcine model," respectively. This indicates animal tissue, not human data. The geographical origin of the tests is within "Beijing Biosis Healing Biological Technology Co., Ltd." in China (derived from company address). The tests are described as non-clinical "conducted to verify," implying prospective testing for the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. There is no mention of experts or ground truth establishment for a test set in the context of human interpretation or AI performance studies. The testing is for medical staplers, not an AI device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no mention of adjudication methods for a test set, as this document refers to physical device testing rather than expert-based assessments of data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document explicitly states: "No clinical study is included in this submission." Furthermore, the device is a surgical stapler, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The device is a surgical stapler, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the ex-vivo and jugular vein tests, the "ground truth" would be the direct measurements and observations of the physical performance characteristics (e.g., staple formation, burst pressure, seal strength) as defined by engineering specifications and relevant standards, demonstrated through the "device met all design specifications" claim. This is physical performance testing, not diagnostic accuracy evaluation.

    8. The sample size for the training set

    Not applicable. The document describes a physical medical device (stapler), not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, this document does not describe an AI algorithm or a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202844

    Validate with FDA (Live)

    Date Cleared
    2021-01-06

    (103 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Surgical Gowns is intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

    Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, the Medical Surgical Gowns met the requirements for Level 3 classification.

    Device Description

    The proposed device is a poly-reinforced surgical gown, the critical zone is front chest and sleeves. The critical zone is reinforced with PP/PE composite laminated material and the reinforced layer is attached with the gown by ultrasonic welding. The proposed device is available in five different sizes, include S-160, M-165, L-70, XL-175, XXL-180, and XXXL-185. Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, the proposed devices can meet the requirements for Level 3 classification. The proposed devices are disposable medical devices and provided in sterile.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for a medical device, specifically Medical Surgical Gowns. The information provided does not describe an AI medical device or a study involving human readers.

    Therefore, most of the requested points are not applicable. I can only extract information related to the performance of the Medical Surgical Gowns based on the provided text.

    Here's a summary of the information that is available:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally based on the ANSI/AAMI PB70:2012 Level 3 classification for liquid barrier performance, along with specific physical and biological properties.

    Acceptance Criteria / TestPerformance (Proposed Device K202844)
    Liquid Barrier Performance (ANSI/AAMI PB70:2012)Met Level 3 classification
    Hydrostatic pressure (AATCC 127)>50 cm
    Water impact (AATCC 42)≤1.0 g
    Physical Properties
    Break strength (ASTM D 5034-09)>20N
    Tear strength (ASTM D 5587-15)>20N
    Air permeability (ASTM D737-18)>15 cm3/s/cm2 (29 ft3/min/ft2)
    Lint (ISO 9073-10)Log10<4
    Flammability (16 CFR Part 1610)Class I
    Seal Strength (ASTM F88/F88M-15)Not explicitly stated in table, but tested
    Seal Leaks (ASTM F1929-15)Not explicitly stated in table, but tested
    Biocompatibility (ISO 10993)
    Cytotoxicity (ISO 10993-5)No Cytotoxicity
    Skin Irritation (ISO 10993-10)No Irritation
    Sensitization (ISO 10993-10)No Sensitization
    Sterilization
    Sterilization method & Sterility Assurance Level (SAL)Ethylene Oxide (EO), SAL=10-6

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes used for the non-clinical tests. Non-clinical tests were conducted to verify design specifications and similarity to the predicate device. The provenance of the data is not explicitly stated beyond the manufacturer being in Beijing, China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is for a physical medical device (surgical gown), not an AI diagnostic tool. "Ground truth" in this context would refer to the validated results of the physical and biological tests, not expert consensus on medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is for a physical medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is for a physical medical device, not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is for a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is established through adherence to recognized international and national standards for physical properties, barrier performance, and biocompatibility, such as ANSI/AAMI PB70:2012, ASTM standards, ISO standards, and CFR regulations. These standards define the acceptable limits for the device's characteristics.

    8. The sample size for the training set

    Not applicable. This is for a physical medical device, not an AI system that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202029

    Validate with FDA (Live)

    Date Cleared
    2020-12-16

    (147 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The medical surgical mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.

    Device Description

    The proposed device, Medical Surgical Mask is a three-layer, single-use, flat-pleated mask. The mask body is made of 25g/m2 PP non-woven cloth. The mask contains tie strings or ear loops to secure the mask over the users' mouth and face and includes a nosepiece to provide a firm fit over the nose. Ear loops are made of Nylon and PU, and tie strings are made of Nylon. The nose clip which is made of Iron strip covered by polypropylene covering. The device is provided in sterile.

    AI/ML Overview

    The provided document (K202029) describes the substantial equivalence determination for a Medical Surgical Mask. It is a 510(k) premarket notification, which means the device is being compared to a legally marketed predicate device to establish substantial equivalence, rather than proving novel safety and effectiveness.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are implied by the similarity comparison to the predicate device and the adherence to relevant ASTM and ISO standards for medical face masks. The "Performance" section within Table 1 provides the reported device performance, and the "Remark" column indicates how it compares to the predicate.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate and Standards)Reported Device Performance (Medical Surgical Mask)Remark (vs. Predicate)
    Fluid resistancePass at 120 mmHgPass at 120 mmHgSame
    Particulate efficiency levelAverage 99.74% at 0.1µm (Predicate's value)Average 98.87%Similar
    Bacterial filtration levelAverage 99.4% (Predicate's value)Average 99.46%Similar
    Differential pressureAverage 2.7 mmH2O/cm² (Predicate's value)Average 3.72 mmH2O/cm²Similar
    FlammabilityClass 1Class 1Same
    Cytotoxicity (Biocompatibility)No CytotoxicityNo CytotoxicitySame
    Sensitization (Biocompatibility)No SensitizationNo SensitizationSame
    Irritation (Biocompatibility)No IrritationNo IrritationSame

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each non-clinical test (e.g., how many masks were tested for particulate efficiency). It only states that "Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device."

    • Sample size: Not explicitly stated for each test.
    • Data provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The sponsor is based in Beijing, China, so it's likely the testing was conducted there or by a certified lab, but this is not confirmed in the text. The tests are "non-clinical," implying they were laboratory-based rather than involving human subjects or real-world data collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this submission. This is a 510(k) for a medical surgical mask, which relies on non-clinical performance testing against established standards and comparison to a predicate device, not on expert interpretation of medical images or clinical outcomes. Therefore, there are no "experts" establishing ground truth in the context of a typical AI/diagnostic device study.

    4. Adjudication method for the test set

    This information is not applicable. Since there are no human readers or expert interpretations involved in establishing ground truth for a medical mask's performance tests, there's no need for an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers (e.g., radiologists) interpret cases with and without AI assistance to assess the AI's impact on their performance. This 510(k) is for a physical medical device (surgical mask) and does not involve AI or human interpretation in its function.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm-only performance study was not done. This is also relevant for AI/diagnostic devices. The medical surgical mask is a physical barrier device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the device's performance is established by objective, standardized laboratory testing against specified ASTM and ISO standards for medical face masks. These standards define the methodologies and criteria for evaluating attributes like fluid resistance, particulate efficiency, bacterial filtration, differential pressure, flammability, and biocompatibility. The results are quantitative measurements or pass/fail determinations based on these established test procedures.

    8. The sample size for the training set

    Not Applicable. This device is a physical product (medical mask) and does not involve machine learning or AI models, so there is no "training set."

    9. How the ground truth for the training set was established

    Not Applicable. As there is no training set for an AI/ML model, there is no ground truth to be established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201641

    Validate with FDA (Live)

    Date Cleared
    2020-08-06

    (51 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Laparoscope Trocar has applications in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments. The trocar may be used with or without visualization for primary and secondary insertions.

    Device Description

    The proposed device is a basic equipment used during laparoscopic surgical, and the surgical device will enter into abdominal cavity via the passage established by the trocar. The proposed device is available in 5mm, 10mm, 12mm, 15mm four diameters to accommodate different sizes surgical instrument.

    AI/ML Overview

    The document provided is a 510(k) Summary for a medical device called "Disposable Laparoscope Trocar." This summary describes the device and its equivalence to predicate devices, focusing on non-clinical testing. It does not contain information about acceptance criteria for a device performance study in the context of diagnostic accuracy, AI, or human reader performance.

    Instead, the "acceptance criteria" discussed are related to standard engineering and biocompatibility tests for a surgical instrument. The study described is a series of non-clinical bench tests and an in-vivo animal study to demonstrate the safety and foundational performance characteristics of the trocar.

    Therefore, many of the requested categories (such as sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, etc.) are not applicable to this type of submission.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the benchmarks set by the predicate and reference devices, and by compliance with relevant ISO and ASTM standards. The reported device performance is mainly comparative to these existing devices and standards.

    Acceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (Subject Device K201641)
    Biocompatibility: Complied with ISO 10993 standards for Cytotoxicity, Sensitization, Intracutaneous, System Toxicity, Pyrogen.Comply with ISO 10993 standards (for Cytotoxicity, Sensitization, Intracutaneous, System Toxicity, Pyrogen)
    Insertion Force: Comparable to predicate/reference devices (e.g., BLTC12 Average 4.2 N, B5LT Average 2.35 N)RJTC-A5: Average 2.25 NRJTC-A15: Average 4.08 N
    Removal Force: Comparable to predicate/reference devices (e.g., BLTC12 Average 10.19 N, B5LT Average 6.27 N)RJTC-A5: Average 6.37 NRJTC-A15: Average 10.31 N
    Leak Resistance: No LeakageNo Leakage
    Snap Feature Retention Force: Comparable to predicate/reference devices (e.g., BLTC12 Average 11.98 N, B5LT Average 1.69 N)RJTC-A5: Average 1.8 NRJTC-A15: Average 11.8 N
    Sterilization Assurance Level (SAL): 10^-610^-6
    Endotoxin Limit: 20 EU per device20 EU per device
    Ethylene Oxide Sterilization Residuals: Complies with ISO 10993-7:2008(Tested, implied compliance)
    Bacterial Endotoxin Limit: Complies with USP <85>(Tested, implied compliance)
    Seal Strength: Complies with ASTM F88/F88M-15(Tested, implied compliance)
    Seal Leaks (Dye Penetration): Complies with ASTM F1929-15(Tested, implied compliance)
    Tip Integrity: Maintained after insertion (from in vivo study)(Evaluated, implied satisfactory)
    Visualization Performance (for optical type): Adequate (from in vivo study)(Evaluated, implied satisfactory)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated for each specific bench test, but implied to be sufficient for statistical validity in engineering tests. For the in vivo study, it mentions a "porcine model," implying a number of animal subjects were used, but the exact count is not given.
    • Data Provenance: Not specified where these tests were conducted, beyond the applicant being Beijing Biosis Healing Biological Technology Co., Ltd. (China). The tests are presumably retrospective data collected during device development.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This is not applicable as the evaluation involves physical and chemical property testing, not subjective expert assessment of clinical data or images.

    4. Adjudication Method for the Test Set

    • This is not applicable. Objective measurements and established testing protocols are used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not applicable as the device is a physical surgical tool (trocar), not an AI-powered diagnostic or assistive technology.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable as the device is a physical surgical tool and does not involve an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for these tests are the established definitions and methodologies within the cited ISO and ASTM standards, as well as the performance characteristics of legally marketed predicate devices. For biocompatibility, it's the biological response as per standardized tests. For mechanical properties, it's the quantifiable physical measurements.
    • For the in vivo study, the ground truth would be the direct observation of insertion force, fixation force, visualization, and tip integrity in a live biological model.

    8. The Sample Size for the Training Set

    • This is not applicable as there is no "training set" in the context of a physical medical device. This is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    • This is not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201243

    Validate with FDA (Live)

    Date Cleared
    2020-07-10

    (63 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use as a receptacle for the collection of tissue, organs and calculi during general and laparoscopic procedures.

    Device Description

    The proposed device, Specimen Retrieval Bag, is comprised of retracted bag, steel sheet, plug, inner sleeve, outer sleeve, pull ring, outer introducer handle, joint, tightening line, pick rod and inner introducer handle. The device is available in a series of models: RJRB-50, RJRB-100, RJRB-120, RJRB-150, RJRN-200, RJRB-250, RJRB-300, RJRB-320, RJRB-420. The difference between each model is the retracted bag size and volume. The device is provided sterile and single use

    AI/ML Overview

    This document is a 510(k) premarket notification for a "Disposable Specimen Retrieval Bag" and does not describe an AI/ML medical device. Therefore, it does not contain information about acceptance criteria for an AI device, a study proving an AI device meets acceptance criteria, sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth, or training set details.

    The document is for a physical medical device and is primarily focused on demonstrating "Substantial Equivalence" to a previously marketed predicate device based on non-clinical performance and biocompatibility tests.

    The device is a physical medical device, not an AI medical device. Therefore, the requested information regarding AI device acceptance criteria, studies, and related data is not applicable and not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1