Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K240164

    Validate with FDA (Live)

    Date Cleared
    2024-06-28

    (158 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovaLine SP-C35 Transducer Protector is a single-use, disposable, prescription device intended for use as a protective device for pressure monitors and to help prevent contamination of the fluid pathway in hemodialysis delivery systems.

    Device Description

    The NovaLine SP-C35 Transducer Protector consists of a plastic housing that contains a hydrophobic filter with a reference pore size of 0.2 µ m. The housing has male and female luer lock connectors that allow its attachment between a blood tubing set and the pressure monitor on a hemodialysis machine.

    The NovaLine SP-C35 Transducer Protector is to be used as protective device for pressure monitors as well as to help maintain the sterility of the blood tubing fluid pathway. The 0.2 ↓ m hydrophobic filter helps prevent cross-contamination by viruses, bacterial and particulate matter while preventing the flow of fluids to the hemodialysis machine pressure monitor.

    The devices are packaged sterile and labeled for single use only.

    The proposed devices are provided in sterile condition, it is subject to e-beam radiation sterilization prior to release to achieve a Sterility Assurance Level (SAL) of 10-6.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the NovaLine SP-C35 Transducer Protector based on the provided FDA 510(k) summary.

    Note: This document describes a medical device, not an AI or software-based medical device. Therefore, many of the requested fields related to AI performance, ground truth, and expert evaluation are not applicable.


    1. Table of Acceptance Criteria and Reported Device Performance

    Device: NovaLine SP-C35 Transducer Protector

    Acceptance Criteria CategorySpecific Test / CharacteristicAcceptance Criteria (Target)Reported Device Performance (Result)Relevant Standard (if applicable)
    Physical PerformanceAppearanceComplies with specificationsComplies with specifications---
    Structural IntegrityMaintainedMaintained---
    DimensionComplies with standardComplies with standardISO 80369-7:2021
    SterilitySAL of 10^-6Achieved SAL of 10^-6---
    Positive Pressure Limitation500 mmHg500 mmHg (matches predicate)---
    Negative Pressure Limitation-500 mmHg-500 mmHg (matches predicate)---
    Connections ConformanceConforms to standardConforms to standardISO 80369-7:2016
    BiocompatibilityCytotoxicityNon-toxicComplies with ISO 10993-5ISO 10993-5
    SensitizationNon-sensitizingComplies with ISO 10993-10ISO 10993-10
    Irritation/Intradermal ReactivityNon-irritatingComplies with ISO 10993-23ISO 10993-23
    Systemic Toxicity/Acute Systemic ToxicityNon-toxicComplies with ISO 10993-11ISO 10993-11
    Pyrogens/Material-mediated PyrogensNon-pyrogenicComplies with ISO 10993-11ISO 10993-11
    In Vitro HemolysisNon-hemolyticComplies with ISO 10993-4ISO 10993-4
    Complement ActivityNo adverse complement activityComplies with ISO 10993-4ISO 10993-4
    Partial Thromboplastin TimeNo adverse effectComplies with ISO 10993-4ISO 10993-4
    Platelet CountNo adverse effectComplies with ISO 10993-4ISO 10993-4
    HematologyNo adverse effectComplies with ISO 10993-4ISO 10993-4
    SterilizationSterility Assurance LevelSAL of 10^-6Achieved SAL of 10^-6---
    PackagingSeal StrengthComplies with standardComplies with standardASTM F88/F88-15
    Dye PenetrationComplies with standardComplies with standardASTM F1929-15
    Shelf LifePackaging Performance (aged samples)Complies with acceptance criteriaComplies with acceptance criteriaISO 80369-7 (for connections)
    Performance Tests (aged samples)Complies with acceptance criteriaComplies with acceptance criteria---
    MicrobiologicalBacterial Endotoxin<20 USP EUComplies with limitUSP <85>
    Microbial ChallengeSterile Barrier IntegrityMaintains sterile barrierComplies with standardsASTM F1671/F1671M-2022, YYT1497-2016, ASTM F838-2020
    Distribution SimulationWithstands distribution stressesComplies with standardComplies with standardASTM D4169-22
    Comparative PerformanceOverall performance vs. predicateEquivalentEquivalent to predicate device---

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each test within the finished product performance, packaging, or shelf-life evaluations. It mentions "proposed devices" were tested, implying a statistical sample was used according to the relevant standards, but the specific number is not provided.

    Data Provenance: The tests were conducted by Bain Medical Equipment (Guangzhou) Co., Ltd. in China. The data is retrospective in the sense that the testing was performed on manufactured devices to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not Applicable. This is a medical device (transducer protector) and not an AI/software device. The "ground truth" here refers to the physical and biological properties of the device, established through standardized laboratory testing, not human expert interpretation of data.

    4. Adjudication Method for the Test Set

    Not Applicable. As this is not an AI/software device involving human interpretation, there is no adjudication method in the traditional sense. Device performance is determined by objective, measurable laboratory tests against pre-defined specifications and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is relevant for imaging or diagnostic AI, comparing human readers with and without AI assistance. This device is a component of a hemodialysis system, and its performance is assessed through physical, chemical, and biological testing.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not Applicable. This device is hardware, not an algorithm or software.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Standardized Test Results: Objective measurements from various laboratory tests (e.g., dimensional measurements, sterile barrier integrity, biocompatibility assays).
    • Compliance with Recognized Standards: Adherence to international and national standards such as ISO 80369-7, ISO 10993 series, ASTM F88, ASTM F1929, ASTM D4169, USP <85>, ASTM F1671, etc.
    • Predicate Device Equivalence: Direct comparison to the performance characteristics of a legally marketed predicate device (Nipro Transducer Protector TP-SURE, K072988) where applicable.

    8. The Sample Size for the Training Set

    Not Applicable. This is a physical medical device, not an AI/machine learning model, so there is no training set.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. See response to #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203524

    Validate with FDA (Live)

    Device Name
    Surgical Masks
    Date Cleared
    2021-02-25

    (86 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

    Device Description

    The Surgical Masks are single use, three-layer, flat-pleated masks with ear loops and nose clip. The inner and outer layers of the mask are made of spunbond polypropylene, and the middle layer is made of meltblown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the mask by ultrasound. The ear loops are made of polyester and spandex. The nose clip is made of polyethylene and galvanized iron wire. Users can adjust the nose clip according to the shape of the bridge of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. The size of the proposed device is 17.5cm×9.5cm. The proposed device is a Level 2 mask based on ASTM F2100-2019. The Surgical Masks are blue and provided non-sterile.

    AI/ML Overview

    This document describes the performance testing for the Bain Medical Equipment (Guangzhou) Co., Ltd. Surgical Masks (K203524).

    1. Table of Acceptance Criteria and Reported Device Performance

    Name of the Test / Methodology / StandardPurposeAcceptance CriteriaResults
    Resistance to Penetration by Synthetic Blood (ASTM F1862/F1862M-17)To evaluate the effectiveness of the test sample from possible exposure to blood and other body fluids.No penetration at 120 mmHgPass at 120mmHg
    Bacterial Filtration Efficiency (ASTM F2101-2019)To determine the bacterial filtration efficiency (BFE) of medical face mask materials, using a biological aerosol of Staphylococcus aureus.≥98%Average 99.9%
    Particulate Filtration Efficiency (ASTM F2299/F2299M-03 (2017))To determine the Particle Filtration Efficiency of materials used in medical face masks to penetration by particulates using latex spheres.≥98%Average 98.8%
    Differential Pressure (EN 14683:2019)To determine the differential pressure.<6.0 mm H2O/cm²Average 4.5 mm H2O/cm²
    Flammability (16 CFR 1610)To evaluate the flammability of the test sample.Class IClass I

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample size used for each test. However, it indicates that the tests were performed in accordance with recognized international standards (ASTM, EN, CFR), which typically specify sample sizes for such evaluations. The data provenance is from non-clinical testing conducted by Bain Medical Equipment (Guangzhou) Co., Ltd., as part of their 510(k) submission to the FDA. The country of origin for the data is implicitly China, where Bain Medical Equipment (Guangzhou) Co., Ltd. is located. It is retrospective in the sense that the testing was performed and the results are being reported for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This report details non-clinical laboratory performance testing of a medical device (surgical masks), not a diagnostic or AI-based device requiring human expert consensus for ground truth. The "ground truth" for these tests is established by the specified methodology and measurement against the defined acceptance criteria in the respective standards.

    4. Adjudication Method for the Test Set

    Not applicable, as this is laboratory performance testing against established standards, not a clinical study involving human assessment or adjudication of outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This submission is for a medical device (surgical masks) which is evaluated through non-clinical performance testing against established standards, not through clinical comparative effectiveness studies with human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This is not an AI/algorithm-based device.

    7. The Type of Ground Truth Used

    The ground truth for this device's performance is based on objective measurements and defined acceptance criteria outlined in specific industry standards and regulations (e.g., ASTM F1862, ASTM F2101, ASTM F2299, EN 14683, 16 CFR 1610). These standards dictate the methodology and expected outcomes for each performance characteristic of a surgical mask.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORA Disposable A.V. Fistula Needle Sets (Safety Needle Series) are single-use sterile medical devices intended to be used as vein puncture for hemodialysis treatment. Protective Shield aids in the prevention of accidental needlesticks.

    The DORA Disposable A.V. Fistula Needle Sets are single-use sterile medical devices intended to be used as vein puncture for hemodialysis treatment.

    Device Description

    The proposed device, DORA Disposable A.V. Fistula Needle Sets, is a non-implantable blood access device, which mainly consists of an adaptor, flexible tube and sharp needle. It is available in two types, 1) Needle sets with safety feature and 2) Needle sets without safety feature. Both the two types of proposed device are provided sterile and are for single use only.

    The two types of proposed devices are offered in difference configurations with options that include needle gauge, needle length, wing types (fixed wing or rotatable wing).

    The proposed device and its package are designed to be provided in Gamma sterilization. The package could maintain the sterility of the device for three years.

    The Protective Shield for injury prevention requires physical action by the clinician to activate and is designed to cover the cannula after treatment. Correct use of this safety feature will eliminate accidental needlestick injuries.

    The proposed device is used in conjunction with following hemodialysis blood tubing sets during clincial use.

    AI/ML Overview

    The provided document, a 510(k) Summary for the DORA Disposable A.V. Fistula Needle Sets, details the premarket notification of intent to market the device and its substantial equivalence to a legally marketed predicate device. This type of submission focuses on demonstrating that a new device is as safe and effective as an already approved device, rather than proving novel effectiveness through extensive clinical trials.

    As such, the document does not contain the information requested regarding acceptance criteria and a study proving a device meets those criteria in the context of:

    • A table of acceptance criteria and reported device performance for an AI/ML or diagnostic device. This document is for a physical medical device (AV fistula needle sets).
    • Sample sizes for a test set (for AI/ML validation). The testing described is for physical characteristics and biocompatibility, not for a 'test set' in the AI/ML sense.
    • Data provenance for a test set (for AI/ML validation). Not applicable to a physical medical device.
    • Number of experts and their qualifications for establishing ground truth. Not applicable.
    • Adjudication method for a test set. Not applicable.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study. Not applicable.
    • Standalone (algorithm only) performance. Not applicable.
    • Type of ground truth used (expert consensus, pathology, outcomes data). Not applicable.
    • Sample size for training set. Not applicable.
    • How ground truth for training set was established. Not applicable.

    Instead, the document focuses on non-clinical testing and substantial equivalence to a predicate device.

    Here's what the document does provide regarding acceptance criteria and performance:

    1. Acceptance Criteria (via Standards Compliance and Performance Testing) and Reported Device Performance:

    The acceptance criteria for this device are demonstrated through adherence to various international and national standards, as well as the successful completion of specific non-clinical performance tests. The reported performance is that the device "met all design specifications" and "is similar as that of the predicate device" or "acceptable."

    Acceptance Criteria CategorySpecific Criterion/StandardReported Device Performance
    Material/Component StandardsISO/FDIS 9626:2016 (Stainless steel needle tubing)Complies with ISO/FDIS 9626:2016 (Needle Performance)
    Packaging IntegrityASTM F88/F88M - 15 (Seal Strength)Complies
    ASTM F1929 – 15 (Seal Leaks by Dye Penetration)Complies
    Biological Safety (Biocompatibility)USP <85> (Bacterial Endotoxin Limit)Complies
    ISO 10993-3:2014 (Genotoxicity, carcinogenicity, reproductive toxicity)Complies with ISO 10993 series standards; tests performed (Cytotoxicity;Sensitization; Irritation sensitivity; Systemic toxicity;Pyrogen; Hemolysis; Thromboresistance; Partial thromboplastin time;Complement activation; Bacterial reverse mutation;Mammalian chromosome aberration; Mouse bone marrow micronucleus;Activated clotting time of whole blood; Platelet adhesion;Muscle implantation; Subchronic systemic toxicity)
    ISO 10993-4:2002 A1:2006 (Interactions with blood)Complies
    ISO 10993-5:2009 (Vitro Cytotoxicity)Complies
    ISO 10993-6:2007 (Local effects after implantation)Complies
    ISO 10993-10:2010 (Irritation and skin sensitization)Complies
    ISO 10993-11:2006 (Systemic toxicity)Complies
    ASTM F756-13 (Hemolytic Properties)Complies
    Functional/Mechanical PerformanceISO 594-1:1986 (Luer Taper - General)Complies (Female Conical Fitting Testing)
    ISO 594-2:1998 (Luer Taper - Lock Fittings)Complies (Female Conical Fitting Testing)
    ISO 10555-1:2013 (Intravascular catheters – General requirements)Complies
    Shipping/Packaging PerformanceASTM D4169-16 (Performance Testing Of Shipping Containers)Complies
    Specific Device PerformanceSimulated Use Needlestick Prevention TestingTest results demonstrated similarity to predicate device
    Needle Performance TestingTest results demonstrated similarity to predicate device
    Mechanical Testing (Wing torque, final lock, needle pushback, mechanical hemolysis)Test results demonstrated similarity to predicate device
    Tensile Strength Testing (Tube to wing pull, tube to joint, needle to cover pull, cannula to hub)Test results demonstrated similarity to predicate device
    Tubing Kinking TestTest results demonstrated similarity to predicate device
    Leakage Testing (Liquid leakage, air leakage)Test results demonstrated similarity to predicate device
    Clamp Stop TestingTest results demonstrated similarity to predicate device
    Flow Rate Testing (Also listed as an AV Fistula Needle Set Performance Test)Test results demonstrated similarity to predicate device
    Particulate Contamination Testing (Also listed as an AV Fistula Needle Set Performance Test)Test results demonstrated similarity to predicate device
    Sterility AssuranceSAL: 10-6 (Sterility Assurance Level) via Gamma SterilizationAchieved 10-6 SAL

    2. Sample sizes used for the test set and the data provenance:

    • Sample Sizes: Not explicitly stated with specific numbers (e.g., "n=X needles were tested"). The document generally states "Non clinical tests were conducted" and "The test results demonstrated...". For a 510(k) submission for a physical device, detailed sample sizes for each specific test are typically found in the full technical documentation, not summarized in the 510(k) summary as it is here.
    • Data Provenance: The tests were conducted internally by Bain Medical Equipment (Guangzhou) Co., Ltd. or by test labs they contracted. The document implicitly indicates the tests originate from China (Guangzhou) as that's the manufacturer's location. The tests are "non-clinical" and likely conducted in laboratory settings, not on patient data or retrospective/prospective human studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a physical medical device. "Ground truth" in this context refers to compliance with established engineering and biocompatibility standards, not expert human interpretation for a diagnostic or AI device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a MRMC study was not done. This device is a physical medical instrument, not an AI/ML diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, not applicable.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • For physical devices, the "ground truth" is adherence to predefined engineering specifications, internationally recognized standards (ISO, ASTM, USP), and established safety and performance benchmarks (e.g., sterilization effectiveness, material strength, biocompatibility profiles). It's a truth based on physical and chemical properties and their interaction, rather than clinical outcomes or diagnostic interpretations.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that uses a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1