Search Filters

Search Results

Found 20 results

510(k) Data Aggregation

    K Number
    K250309

    Validate with FDA (Live)

    Device Name
    BTL-199
    Date Cleared
    2025-08-07

    (185 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTL-199 is intended to stimulate peripheral nerves for relief of chronic intractable pain, post-traumatic pain, post-surgical pain and/or for relief of chronic painful diabetic peripheral neuropathy in the lower extremities. The BTL-199 is for use on patients 18 and older.

    Device Description

    The BTL-199 is an electromagnetic stimulator that delivers brief, focused electromagnetic pulses to non-invasively stimulate peripheral nerves, providing pain relief.

    The device consists of a main unit and specialized applicators, designed to deliver precise and effective therapy. The main unit is equipped with an intuitive color touch screen that provides step-by-step guidance for the operator throughout the therapy process. The device features a preset treatment protocol for ease of use, while also allowing operators to customize parameters to meet specific therapeutic needs. Adjustable parameters include frequency (up to 150 Hz), therapy time (up to 30 minutes), and intensity (0-100%). The interface displays detailed information about the therapy settings and session progress (i.e., treatment duration). The device uses a temperature sensor to monitor the temperature of the coil to ensure that the maximum permissible temperature of the applicator's contact areas (i.e., 43°C) are not exceeded. It also has an intensity predictor function that limits the maximum adjustable intensity of the device based on the therapy parameters set, so as not to overheat and damage the device.

    The device is equipped with three sets of applicators: AP-C-1, AP-C-2, and AP-C-4/5. Each applicator is intended for a different body area depending on the size and shape of the treated body part.

    • Applicator AP-C-1 is intended for treatment of larger body areas, such as back, waist, thighs, etc.
    • Applicator AP-C-2 is intended for treatment of smaller body areas, such as hands, arms, legs, etc.
    • Applicators AP-C-4/5 (AP-C-4 and AP-C-5 - for each side of the body respectively) is primarily intended for treatment of waist and lower back.
    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the BTL-199 device does not include any clinical study data or describe acceptance criteria related to device performance in a clinical setting.

    The submission is based on the device being "substantially equivalent" to predicate devices, which generally means that it has the same intended use and similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness. For devices like the BTL-199, classified as Class II, substantial equivalence can often be demonstrated through non-clinical testing (bench testing, electrical safety, biocompatibility, etc.) and a comparison of technological characteristics with predicate devices that have already been cleared.

    The "Clinical Testing" section explicitly states "Not applicable," which confirms that a clinical study proving device performance against specific acceptance criteria (like accuracy, sensitivity, or improvement in patient outcomes) was not required or submitted for this 510(k) clearance.

    Therefore, I cannot fulfill the request to describe acceptance criteria and a study proving the device meets them based on the provided text, as such information is not present. The document focuses on demonstrating substantial equivalence for regulatory clearance, not on presenting performance data from a clinical trial.

    If this were a device requiring clinical performance data (e.g., an AI-powered diagnostic device), the 510(k) summary would typically include a section detailing the clinical study, its methodology, and the results compared against pre-defined acceptance criteria. This is not the case for the BTL-199.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243290

    Validate with FDA (Live)

    Device Name
    BTL-785MJ
    Date Cleared
    2025-05-09

    (203 days)

    Product Code
    Regulation Number
    890.5850
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTL-785MJ device with the controller and single-use applicators has the following indications for use:

    1. To relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain
    2. Muscle re-education
    3. Increasing blood flow
    4. Maintain or increase mandibular range of motion
    Device Description

    The BTL-785MJ is a state-of-the-art electrostimulation and radiofrequency platform with accessories that enables the application of therapy by electrical stimulation.
    The control unit of the system is equipped with a large color touch screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.
    The generated electromagnetic field is intended to interact with the tissues of the human body in order to relieve symptoms associated with muscle spasm, to treat temporomandibular joint (TMJ) dysfunction and associated pain, muscle re-education, increase blood flow and maintain or increase mandibular range of motion.
    The BTL-785MJ device consists of the following main components:

    • Main unit
    • Handpiece
    • Holding arm
    • Connection cables
    • Single-use applicators
    • Therapy discomfort button
    AI/ML Overview

    The provided FDA clearance letter (K243290) for the BTL-785MJ device does not contain explicit acceptance criteria or a detailed clinical study demonstrating the device meets such criteria. Instead, it relies on substantial equivalence to a predicate device.

    Here's an analysis of the information available and what is missing relative to your requested points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in terms of specific performance metrics or thresholds for the BTL-785MJ device's effectiveness in treating the listed indications. The "device performance" reported is primarily compliance with various electrical safety, EMC, usability, software, risk management, and biological evaluation standards.

    Acceptance Criteria (Not explicitly stated for clinical performance)Reported Device Performance (Non-clinical)
    (Not stated for clinical efficacy)Complies with IEC 60601-1 (General safety & essential performance)
    (Not stated for clinical efficacy)Complies with IEC 60601-1-2 (EMC)
    (Not stated for clinical efficacy)Complies with IEC 60601-2-10 (Nerve & muscle stimulators)
    (Not stated for clinical efficacy)Complies with IEC 60601-1-6 (Usability)
    (Not stated for clinical efficacy)Complies with IEC 62304 (Software life cycle processes)
    (Not stated for clinical efficacy)Complies with ISO 14971 (Risk management)
    (Not stated for clinical efficacy)Complies with ISO 10993-1, -5, -10 (Biological evaluation: general, cytotoxicity, irritation/sensitization)

    2. Sample size used for the test set and the data provenance

    • Sample size: Not applicable. The document explicitly states: "A clinical study was not conducted." Therefore, there is no test set of patients for clinical efficacy.
    • Data provenance: Not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical study was conducted.

    4. Adjudication method for the test set

    Not applicable, as no clinical study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The device is a physical muscle stimulator, not an AI-powered diagnostic or interpretive tool that would involve "human readers" or AI assistance in that context. Therefore, an MRMC comparative effectiveness study is not relevant to this device, and none was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical electrostimulation device, not an algorithm.

    7. The type of ground truth used

    Not applicable for clinical efficacy. The substantial equivalence relies on comparing the BTL-785MJ's technical characteristics and indications for use to a legally marketed predicate device (D function, K203525) that presumably established its effectiveness. The ground truth for safety and performance is based on compliance with international standards (IEC, ISO) through non-clinical testing.

    8. The sample size for the training set

    Not applicable, as this is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.


    Summary of Device Acceptance and Study:

    The BTL-785MJ device received FDA clearance (K243290) based on substantial equivalence to an existing predicate device, the "D function" (K203525). This means its safety and effectiveness were demonstrated by showing that it has similar technological characteristics and the same intended use as the predicate device.

    Specifically, the "study" demonstrating this involves:

    • Non-clinical testing: Compliance with various electrical safety, electromagnetic compatibility (EMC), usability, software lifecycle processes, risk management, and biological evaluation standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-6, IEC 62304, ISO 14971, ISO 10993 series). These tests verify the physical and operational safety of the device.
    • Comparison to a predicate device: A detailed table in the 510(k) summary (page 8-10) outlines the similarities in indications for use, principle of action, clinical use, energy source, electrical safety, energy type, accessory attachment, therapy time, pulse width, output frequency, output current, max output voltage, max current density, pulse shape, channel numbers, electrical safety class, features (timer, buzzer, display), and electrode properties. The applicant argues that "Any differences between the predicate device and BTL-785MJ device have no significant influence on safety or effectiveness of the BTL-785MJ device."

    Crucially, no clinical study was conducted to directly prove the clinical efficacy of the BTL-785MJ device itself. Its acceptance is predicated on its similarity to a device that was previously cleared and is presumed safe and effective for its stated indications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241516

    Validate with FDA (Live)

    Device Name
    BTL-398
    Date Cleared
    2025-02-10

    (257 days)

    Product Code
    Regulation Number
    876.5320
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence.

    Device Description

    The BTL-398 is a non-invasive therapeutic device that delivers magnetic pulses to stimulate the pelvic floor muscles for the treatment of urinary incontinence in adult population. The BTL-398 device is intended for professional use in healthcare facilities as a prescription use only. The BTL-398 consists of a main unit and a Chair Applicator. The main unit consists of the electromagnetic field generators, the microprocessor-driven control unit, and the touch-screen control panel. The Chair Applicator is designed for non-invasive treatment of urinary incontinence. The magnetic coil is in the center of the seat, and its position is user-adjustable during therapy. Ongoing therapy is indicated by illuminated segments. The Chair Applicator is designed for precise patient positioning, ensuring the most appropriate posture before and during therapy. Prior to High-Intensity Focused Electromagnetic (HIFEM) therapy, the seat height and coil position can be adjusted using an attached remote controller. The Dynamic HFEM Technology (DHT) feature ensures the coil is optimally positioned at the right location beneath the patient before starting therapy. At the start of therapy, air is circulated through the seat cushion with the Pressure Wave Technology (PWT) feature (intensity of the PWT function is set based on patient feedback). The PWT feature helps patients maintain the correct therapeutic position during therapy. Both the PWT and DHT features can be readjusted, and the PWT feature can be turned off, at any time during the session. The touch screen display guides the user step-by-step through the entire therapy procedure. During therapy, the screen displays information about the applied therapy, remaining therapy time, and main therapy parameters. Therapy is administered through clothing.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for a device called "BTL-398". This document primarily focuses on demonstrating substantial equivalence to a predicate device (HPM-6000UF) for regulatory clearance. It does not include information about clinical studies with human participants, AI components, or detailed performance data against acceptance criteria for such clinical or AI-specific assessments.

    Therefore, many of the requested details about acceptance criteria, clinical study design, sample sizes, expert involvement, and ground truth establishment, particularly concerning AI or human-in-the-loop performance, cannot be extracted from this document. The document explicitly mentions "non-clinical testing" and "performance bench data," suggesting that equivalence was primarily established through engineering and electromagnetic performance tests, not clinical efficacy trials for the UI treatment itself.

    However, I can extract information related to what was used for acceptance and testing:


    Acceptance Criteria and Device Performance (Based on "Non-clinical Testing")

    Acceptance Criteria CategoryReported Device Performance (Summary)
    Electrical SafetyMet IEC 60601-1: 2020
    Mechanical StrengthMet IEC 60601-1: 2020
    Thermal SafetyMet IEC 60601-1: 2020
    Electromagnetic CompatibilityMet IEC 60601-1-2: 2020
    Software Verification & ValidationMet FDA's 2023 Guidance ("Content of Premarket Submissions for Device Software Functions
    Magnetic Field StrengthBench data submitted, all pre-determined acceptance criteria met. (Specific values not detailed)
    Coil Surface Temperature ControlsBench data submitted, all pre-determined acceptance criteria met. (Specific values not detailed)
    PWT featureBench data submitted, all pre-determined acceptance criteria met. (Specific functionality not detailed)

    Study Details (as per available information):

    1. A table of acceptance criteria and the reported device performance:
      (See table above) The document states, "All pre-determined acceptance criteria were met." However, it does not provide specific numerical targets or results for the bench performance tests (e.g., exact magnetic field strength measured vs. target range). It only confirms compliance with standards and successful completion of tests.

    2. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: Not applicable in the context of human subjects or AI test sets. The testing described is non-clinical bench testing of the device's electrical, mechanical, thermal, electromagnetic, and software functions.
      • Data Provenance: N/A for clinical data. The data is from internal bench testing (performance bench data).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable. Ground truth for clinical outcomes or diagnoses is not mentioned as this was primarily non-clinical testing.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      Not applicable. No human reader or expert adjudication process for clinical data or AI output is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No MRMC study was mentioned. The device is a non-invasive electromagnetic stimulator, not an imaging or diagnostic AI device that would typically involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable for an AI algorithm. The device, the BTL-398, is a physical medical device. Its software functions were verified and validated, but this does not imply an AI component in the typical sense of diagnostic or image analysis AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For the non-clinical tests, the "ground truth" was established based on engineering specifications and compliance with recognized standards (IEC 60601-1, IEC 60601-1-2) and FDA guidance for software verification and validation. For magnetic field strength, coil surface temperature, and PWT features, the ground truth would be the expected performance range derived from design specifications.

    8. The sample size for the training set:
      Not applicable. No machine learning training set is mentioned.

    9. How the ground truth for the training set was established:
      Not applicable. No machine learning training set is mentioned.


    Summary of what the document focuses on:

    The entire document pertains to establishing substantial equivalence of the BTL-398 to a predicate device (HPM-6000UF) through non-clinical testing. The key differences noted between the subject device (BTL-398) and the predicate are the inclusion of a "PWT feature" (Pressure Wave Technology for patient positioning) and a "DHT Feature" (Dynamic HFEM Technology for coil adjustment). The non-clinical tests were conducted to ensure these new features and the overall device design do not raise new questions of safety or effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241270

    Validate with FDA (Live)

    Device Name
    BTL-754
    Date Cleared
    2025-01-30

    (269 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTL-754 device with BTL-754-4 handpiece has the following indications for use:

    The 2940nm Fractional handpiece is intended for use in dermatological procedures requiring coagulation, resurfacing, and ablation of soft tissue. Procedures include skin resurfacing and treatment of wrinkles, furrows, fine lines, textural irregularities, pigmented lesions, and vascular dyschromia.

    The 1540nm Fractional handpiece is intended for use in coagulation of soft tissue, skin resurfacing procedures as well as treatment of melasma, striae, acne scars and surgical scars.

    The combined 1540mm and 2940mm Fractional handpiece is intended for dermatological procedures requiring coagulation, resurfacing, and ablation of soft tissue. Procedures include skin resurfacing and treatment of wrinkles, rhytides, furrows, fine lines, textural irregularities, dyschromia and pigmented lesions.

    Device Description

    The BTL-754 is a medical device containing two laser wavelengths. A laser is a device that can generate light in the visible or invisible part of the electromagnetic spectrum. The generation of radiation occurs by means of so-called stimulated emission, thanks to which its beam has specific properties - the laser beam is well directed and can propagate over long distances without much divergence. It can also be focused into a very small spot where a very high intensity is achieved.

    The BTL-754 consists of a main unit, articulated mirror arm and applicator. The main unit is equipped with a color touch screen with a wide that makes the device easy to use. The onscreen information informs the operator through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen. During therapy, the screen displays information about therapy parameters. The device is equipped with the Emergency laser stop button for immediate termination of the laser emission.

    AI/ML Overview

    The provided FDA 510(k) summary for the BTL-754 device does not describe a study involving an AI/Machine Learning component, nor does it specify acceptance criteria for such a system.

    Instead, the document details a medical device (a laser surgical instrument) and its substantial equivalence to a predicate device. The "Non-clinical Testing" section refers to compliance with applicable medical device safety standards and basic device performance verification, typical for hardware medical devices, not AI/ML performance. The "Clinical Testing" section explicitly states "Not applicable to this submission".

    Therefore, I cannot extract the requested information regarding acceptance criteria and the study proving the device meets them, as the provided text does not contain any reference to AI/ML device performance or associated studies.

    The information provided is geared towards demonstrating the BTL-754 is substantially equivalent to a legally marketed predicate device (Palomar Icon Aesthetic System, K142376) based on intended use, technological characteristics, and safety standards for laser surgical instruments.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242532

    Validate with FDA (Live)

    Device Name
    BTL-785BNF-E
    Date Cleared
    2024-12-13

    (109 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for aesthetic use including facial and neck or body skin stimulation.

    Device Description

    The BTL-785BNF-E is a state-of-the-art electrostimulation device that enables the application of therapy by electrical stimulation to patient's skin for aesthetic use. The main control unit is equipped with a large color touch screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy, the device displays information about the applied therapy type, the remaining therapy time and the main therapy parameters on the screen. The generated electrical stimulation is intenact with the tissues of the human body in order to achieve skin stimulation in the face. The device consists of the following main components: BTL-785 main unit, BTL-785-7 Handpiece, Single-use application electrodes, Holding arm, Connection cables, Therapy discomfort button.

    AI/ML Overview

    The provided document is a 510(k) summary for the BTL-785BNF-E device. It describes the device's indications for use, its substantial equivalence to a predicate device, and the non-clinical testing performed.

    Crucially, the document explicitly states under the "Clinical performance data" section: "A clinical study was not conducted."

    This means that the information typically required to answer your questions regarding acceptance criteria and performance studies (such as sample size, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment) is not available within this document because no clinical study was performed.

    The substantial equivalence determination for this device is based on its similarity to a legally marketed predicate device (BTL-785BNF Handpiece, K233604) and non-clinical testing (electrical safety, EMC, usability, risk management, and biocompatibility standards compliance), not on a new clinical performance study. The key difference in the subject device is the introduction of a new disposable applicator (BTL-785-7-11), which the manufacturer asserts does not significantly influence safety or effectiveness given its similar materials, manufacturing, and technology.

    Therefore, I cannot provide a table of acceptance criteria and proven performance based on a clinical study for this device, nor can I answer the specific questions about clinical study design parameters.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233849

    Validate with FDA (Live)

    Device Name
    BTL-499
    Date Cleared
    2024-04-23

    (140 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BTL-499 is indicated to be used for:
    Non-invasive lipolysis (breakdown of fat) of the abdomen.
    Reduction in circumference of the abdomen.
    Non-invasive lipolysis (breakdown of fat) of the thighs.
    Reduction in circumference of the thighs.
    BTL-499 is intended for use with skin types I - VI.
    Non-invasive lipolysis (breakdown of fat) of the flanks limited to skin types I - IV.
    Non-invasive lipolysis (breakdown of fat) of the upper arms limited to skin types II and III and BMI 30 and under.
    The BTL-499 is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The BTL-499 massage function is intended to provide a temporary reduction in the appearance of cellulite.

    Device Description

    The BTL-499 is a non-invasive therapeutic device is comprised of a main unit and applicators that deliver energy to the targeted tissue. The device's two outputs enable hands-free simultaneous treatment by two applicators. The BTL-499 is equipped with a large color touch-screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy, the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen. The system console consists of the RF, HV and massage generators, computer, and a touch-screen control panel. The system combines bipolar radiofrequency with electromagnetic stimulation and mechanical massage. The device is accompanied by the following applicators: 499-AP-MRA-1, 499-AP-MR-2, 499-AP-MRA-4 & 499-AP-MRA-5

    AI/ML Overview

    Given the provided document, the device in question is the BTL-499, an electrosurgical device. The document states "Clinical Testing: Not applicable to this submission" on page 5. This explicitly indicates that no clinical studies were performed for this submission, and therefore, no data exists to answer questions regarding acceptance criteria, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, or ground truth establishment for clinical performance.

    The document primarily focuses on non-clinical testing for electrical safety and compliance with medical device standards, and a comparison to predicate devices to demonstrate substantial equivalence based on technological characteristics and intended use, rather than direct comparative clinical performance.

    Therefore, most of the requested information cannot be extracted from this document, as it pertains to clinical performance evaluation, which was explicitly stated as "Not applicable to this submission."

    Here's a breakdown of what can be determined from the document:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (Non-Clinical/Safety Standards): The document lists several IEC and ISO standards that the BTL-499 device was found to comply with. These standards implicitly serve as the acceptance criteria for the non-clinical performance (safety, EMC, software lifecycle, risk management, biocompatibility).
      • IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
      • IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
      • IEC 62304: Medical device software - Software life cycle processes
      • ISO 14971: Medical devices – Application of risk management to medical devices
      • ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
      • ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
      • ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
    • Reported Device Performance (against standards): The document states: "The BTL-499 device has been thoroughly evaluated for electrical safety. The device has been found to comply with applicable medical device safety standards." This is the reported performance – that it complies with the listed standards. No specific quantitative performance metrics beyond compliance are provided in this summary.

    The following information is NOT available in the provided text because clinical testing was not applicable to this submission:

    • 2. Sample sized used for the test set and the data provenance: Not applicable. No clinical test set data is provided.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth was established from experts for this submission.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is not described as an AI-assisted device in this context, and no clinical comparative effectiveness study against human readers was performed.
    • 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. The device is an energy delivery system, not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. No clinical ground truth data was used for this submission.
    • 8. The sample size for the training set: Not applicable. No training set for clinical performance was used.
    • 9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233604

    Validate with FDA (Live)

    Device Name
    BTL-785S
    Date Cleared
    2024-03-28

    (140 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTL-785S device has the following indications for use:

    The BTL-785S with BTL-785-1 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The BTL-785-1-4 massage device is intended to provide a temporary reduction in the appearance of cellulite.

    The BTL-785S with BTL-785-2 applicator is indicated to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

    The BTL-785S with BTL-785-3 applicator is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

    The BTL-785-4 applicator used with tips BTL-785-4-1, BTL-785-4-2, BTL-785-4-5 and BTL-785-4-6 is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. The Applicator BTL-785-4 of BTL-785S device used with tips BTL-785-4-3, BTL-785-4-7 and BTL-785-4-8 is intended for dermatological procedures requiring fractional treatment of the skin. At higher energy levels greater than 62 mJ/pin, use of the BTL-785-4 applicator is limited to Skin Types I-IV.

    The BTL-785S with BTL-785-7 Handpiece used with hands-free applicators:

    BTL-785-7-1 & BTL-785-7-7, BTL-785-7-2 & BTL-785-7-8 single-use applicators are intended to provide:

    • Heating for the purpose of elevaing tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

    BTL-785-7-1 & BTL-785-7-2 single-use applicators are intended to provide:

    • Non-invasive temporary reduction of facial wrinkles.

    BTL-785-7-9 single-use applicator is intended to:

    • Affect the appearance of lax tissue in the submental area.
    Device Description

    The BTL-785S is a state-of-the-art radiofrequency device that enables the application of therapy by a high-frequency field.

    The control unit of the system is equipped with a large color touch screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen. For easier control, the hand-pieces are equipped with buttons, enabling operation of the device during therapy. The energy flow's quality is indicated by the illuminated treatment tip. The BTL-785S device comes with five different types of applicators.

    The BTL-785S device consists of the following main components:

    • microprocessor-driven control unit
    • radiofrequency generator
    • user interface with 15.6" color touch screen
    • applicators
    • exchangeable applicator tips
    AI/ML Overview

    The BTL-785S device, specifically with the BTL-785-7 handpiece and BTL-785-7-9 single-use applicator, was studied to support a new indication for use: "Affect the appearance of lax tissue in the submental area."

    Here's an analysis of the acceptance criteria and the study results:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Secondary Endpoints)Reported Device Performance
    A score improvement of at least 1.0 grade point in at least 75% of treated subjects.GAIS evaluation by three independent reviewers showed that a score of at least 1 point on the scale (referring to improved appearance) was documented in 51 of 56 subjects (91%). (Meets Criteria)
    Evaluation of effect on submental skin laxity: measurable reduction in the submental area and visible improvement by independent reviewers.At 3 months, 46 subjects (82%) exhibited a ≥ 20 mm² reduction in the submental area. Of those 46 subjects, 43 (93%) were identified by independent reviewers using GAIS as having visible improvement in the treated area at 3 months. (Meets Criteria)
    Overall satisfaction with the study treatment outcome.At 3 months, 56 (100%) subjects answered "satisfied" or "very satisfied" to the question "Overall how are you satisfied with the therapy?". 56 (100%) subjects answered "satisfied" or "very satisfied" to the question "Overall how satisfied are you with the chin-toning after the therapy". 55 subjects (98%) answered "satisfied" or "very satisfied" to the question "Overall how satisfied are you with the skin laxity of the treated area after the therapy". (Meets Criteria)
    Therapy comfort (assessed by subjects' agreement with "I found the treatment comfortable" and pain sensation).54 subjects (96%) "agreed" or "strongly agreed" with the statement "I found the treatment comfortable". Pain sensation assessed by 10-point analogue scale yielded an average score of 0.46 ± 0.81 points, with 55 (98%) subjects falling into the none to mild pain range (0-3). (Meets Criteria)
    Device safety (no adverse events).No adverse events occurred throughout the whole clinical investigation. (Meets Criteria)

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Sample Size: N=56 subjects.
    • Data Provenance: The study was a "single-center single-arm, open-label, interventional design" clinical investigation. The document does not explicitly state the country of origin, but given the FDA submission, it implicitly refers to data relevant for the U.S. market. It is a prospective clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Number of Experts: Three independent reviewers were used for the GAIS (Global Aesthetic Improvement Scale) evaluation.
    • Qualifications of Experts: The document does not specify the qualifications of these experts (e.g., radiologist with 10 years of experience). It only states they were "independent reviewers."

    4. Adjudication Method for the Test Set:

    • The document states "GAIS evaluation by three independent reviewers." This implies a form of consensus or independent grading by multiple experts, but the specific adjudication method (e.g., arithmetic mean, majority vote, or whether disagreements were arbitrated) is not explicitly detailed. It does not mention "2+1" or "3+1" specifically.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a Multi Reader Multi Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. This study focuses on the standalone performance of the device and its effect on patients. The device is a radiofrequency device, not an AI diagnostic tool for human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone clinical performance study was done for the device when used with the BTL-785-7 handpiece and BTL-785-7-9 single-use applicator. The study assessed the device's direct effect on patient outcomes (appearance of lax tissue, satisfaction, comfort, and safety) without human interpretation of device output in a diagnostic sense.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth was established through a combination of:
      • Expert Consensus: General Aesthetic Improvement Scale (GAIS) evaluation by three independent reviewers.
      • Objective Measurement/Outcomes Data: Measurable reduction in the submental area (≥ 20 mm²).
      • Patient-Reported Outcomes/Feedback: Subject satisfaction with therapy outcomes and comfort levels.

    8. The Sample Size for the Training Set:

    • The document does not provide information on a "training set" as it pertains to an AI/algorithm development. This was a clinical study designed to demonstrate the safety and efficacy of the physical device for a specific indication, not to train an AI model.

    9. How the Ground Truth for the Training Set Was Established:

    • As mentioned in point 8, there is no information about a training set or how ground truth for a training set was established, as this was not an AI/algorithm development study.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190456

    Validate with FDA (Live)

    Device Name
    BTL 799-2L
    Date Cleared
    2019-07-05

    (130 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BTL 799-2L is indicated to be used for:

    • · Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
    • · Strengthening, Toning and Firming of buttocks, thighs and calves.
    • · Improvement of muscle tone and firmness, for strengthening muscles in arms.
    Device Description

    The BTL 799-2L is a non-invasive therapeutic device. The device produces electromagnetic field that interacts with the tissues of the human body. The device has two outputs that enable simultaneous treatment by two applicators.
    The BTL 799-2L is equipped with a large color touch screen with wide view angle that significantly facilitates the use of the device. The on-screen information guides the user stepby-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device keeps information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

    AI/ML Overview

    This is a 510(k) premarket notification, which demonstrates substantial equivalence to a predicate device rather than conducting a de novo study to establish new performance criteria. Therefore, the information typically found in a study proving a device meets acceptance criteria, such as a table of acceptance criteria and reported device performance metrics like sensitivity, specificity, or AUC, a specific sample size for a test set, expert qualifications, or adjudication methods, is not present here.

    Instead, the submission focuses on non-clinical testing to ensure safety and essential performance, and a comparison of technological characteristics to predicate devices to argue for substantial equivalence.

    Here's a breakdown of the available information based on your request:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated as performance metrics in the typical sense (e.g., sensitivity, specificity). The acceptance criteria implicitly are compliance with various medical device safety and performance standards (listed below) and having technological characteristics and intended use similar enough to predicate devices to not raise new safety or effectiveness concerns.
    • Reported Device Performance: No specific performance metrics (e.g., accuracy, sensitivity) are reported as this is a comparison to a predicate device, not a de novo performance study. The device's "performance" is demonstrated by its adherence to safety standards and its functional similarity to legally marketed devices.

    2. Sample size used for the test set and the data provenance

    • This information is not provided as this submission relies on demonstrating substantial equivalence through non-clinical testing and comparison to predicate devices, not through a clinical performance study with a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided as there is no mention of a test set with expert-established ground truth.

    4. Adjudication method for the test set

    • This information is not provided as there is no mention of a test set or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This information is not provided as this device is a physical muscle stimulator, not an AI-assisted diagnostic or imaging device, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This information is not applicable/provided as this device is a physical muscle stimulator, not an algorithm.

    7. The type of ground truth used

    • Not applicable as no clinical study with a ground truth is described. The "ground truth" for this submission revolves around compliance with recognized medical device safety standards and the established safety and effectiveness of the predicate devices it is compared against.

    8. The sample size for the training set

    • This information is not provided as this is not an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    • This information is not provided as this is not an AI/machine learning device requiring a training set.

    Non-clinical Testing Summary:

    The BTL 799-2L device underwent thorough evaluation for electrical safety and compliance with the following applicable medical device safety standards:

    • IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
    • IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
    • IEC 60601-1-6: Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
    • IEC 60601-2-10: Medical Electrical Equipment – Part 2-10: Particular Requirements for the Basic Safety and Essential Performance of Nerve and Muscle Stimulators
    • IEC 62304: Medical device software - Software life cycle processes
    • ISO 14971: Medical devices – Application of risk management to medical devices
    • ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
    • ISO 10993-5: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
    • ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization

    The comparison table provided details the technological characteristics of the BTL 799-2L compared to its primary predicate BTL 799-2T (K182106) and secondary predicate HPM-6000 (K160992). The key argument is that any differences do not raise new types of safety or effectiveness questions, thus demonstrating substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190023

    Validate with FDA (Live)

    Device Name
    BTL-084
    Date Cleared
    2019-03-29

    (84 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BTL-084 RF device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The BTL-084 massage device is intended to provide a temporary reduction in the appearance of cellulite.

    Device Description

    The BTL-084 is a non-invasive therapeutic device enables the application of therapy by a high-frequency field. The control unit of the system is fitted with a color touch screen, to facilitate use of the device. The on-screen information guides the operator through the entire therapy. For easier control, the handpiece is equipped with buttons, enabling operation of the device during therapy. The energy flow is indicated by the illuminated treatment tip. The BTL-084 consists of the following main components: microprocessor-driven control unit, high-frequency electromagnetic energy generator, massage generator, user interface with 8.4" color touch screen, handpiece.

    AI/ML Overview

    There is no information about acceptance criteria or a study proving the device meets acceptance criteria in the provided text.

    The document is a 510(k) premarket notification clearance letter from the FDA for the BTL-084 device. It primarily discusses:

    • The FDA's determination of substantial equivalence to a legally marketed predicate device (TR-100 (K161551)).
    • The intended use of the device: "to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation" and "to provide a temporary reduction in the appearance of cellulite."
    • Non-clinical testing for electrical safety and compliance with various medical device safety and quality standards (e.g., IEC 60601-1, ISO 14971).
    • A brief mention of "Clinical testing" stating that "The substantial equivalence determination for the BTL-084 is based on clinical performance testing. The aim of the performance test was to prove that the device reaches the effective treatment temperature and maintain it for required time."

    However, the document does not provide any specific acceptance criteria (e.g., specific thresholds for temperature, pain reduction, or cellulite reduction), nor does it detail the results of the "clinical performance testing" that would demonstrate these criteria were met.

    Therefore, I cannot provide the requested table or detailed study information because it is not present in the given text.

    To address the specific points of your request:

    1. A table of acceptance criteria and the reported device performance: Not available.
    2. Sample size used for the test set and the data provenance: Not available. The document only mentions "clinical performance testing" but provides no details on participants, study design, or data origin.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/available. The context of "ground truth" (e.g., for AI model validation) is not present as this is a medical device clearance document, not an AI efficacy study report.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable/available.
    5. If a multi-reader multicase (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document describes a physical medical device, not an AI algorithm.
    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done: Not applicable. This document describes a physical medical device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable/available. The document mentions "clinical performance testing" to prove the device "reaches the effective treatment temperature and maintain it for required time," which implies physiological measurements, but no specific "ground truth" methodology is described.
    8. The sample size for the training set: Not applicable. This is not an AI model.
    9. How the ground truth for the training set was established: Not applicable. This is not an AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182363

    Validate with FDA (Live)

    Device Name
    BTL-703
    Date Cleared
    2019-01-24

    (147 days)

    Product Code
    Regulation Number
    890.5290
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BTL-703 is indicated to be used for applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as: 1. Relieving pain; 2. Reducing muscle spasm; 3. Increasing range of motion of contracted joints using heat and stretch techniques; and 4. Increasing blood flow to tissues in the treatment area.

    Device Description

    The BTL-703 is comprised of a device unit and applicator that deliver radiofrequency energy to the targeted tissue. The device is designed to enable hands-free treatment. Easyto-use color touch screen allows for maximum operator comfort. The main unit is placed in a specially designed cart, the shape of which provides maximum operator comfort and easy movement of the device in the office. The BTL-703 consists of the following main components: microprocessor-driven control unit, high-frequency energy generator, user interface with 15,6" HD touch screen, four-point applicator.

    AI/ML Overview

    The acceptance criteria and study information for the BTL-703 device, based on the provided FDA 510(k) summary, are described below.

    Please note: This document focuses on the regulatory submission information. For medical devices like the BTL-703 (Shortwave Diathermy), the "performance" in the context of a 510(k) submission primarily refers to meeting safety and fundamental operational requirements and demonstrating substantial equivalence to a predicate device. It typically does not involve the same kind of clinical trial efficacy metrics (sensitivity, specificity, etc.) as AI/software devices. The "acceptance criteria" here are therefore related to safety standards and the ability to achieve a therapeutic temperature range.


    1. Table of Acceptance Criteria and Reported Device Performance

    Criteria CategoryAcceptance CriteriaReported Device Performance
    Safety and Essential Performance StandardsCompliance with recognized medical device safety standards. Specifically:- IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance (Complied) - IEC 60601-1-2: Collateral standard: Electromagnetic Disturbance – Requirements and tests (Complied) - IEC 62304: Medical device software – Software life cycle processes (Complied) - ISO 14971: Medical devices - Application of risk management to medical devices (Complied) - ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process (Complied) - ISO 10993-5: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity (Complied) - ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization (Complied)
    Ablity to achieve therapeutic temperatureThe device should be able to maintain a therapeutic temperature range in vivo (in relevant body tissues). Specific target: approximately 40-45°C.The study confirmed the ability of the BTL-703 device to maintain the temperature in the range of approx. 40-45°C in vivo.
    Substantial EquivalenceDemonstration that the technological characteristics and indications for use are substantially equivalent to a legally marketed predicate device (K120093, BTL Elite), raising no new safety or effectiveness questions.Comparison table provided showing similar indications for use, principle of action, clinical use, electrical protection, user interface, firmware control, type of energy, working frequency (27.12 MHz), therapy time (1-30 min), hands-free application, stop remote control, and applicator contact monitor. Differences (e.g., no pulsed mode, negative pressure applicator fixation, thermal stabilization system) were determined not to raise new safety or effectiveness questions.

    2. Sample size used for the test set and the data provenance

    • Test Set (Animal Study): The document only mentions a "study in porcine model" for demonstrating temperature maintenance. It does not provide the specific sample size (number of animals or test points) used in this model.
    • Data Provenance: The animal study was conducted in a porcine model. The country of origin for the data is not specified. This was a prospective study designed to specifically evaluate the device's ability to maintain temperature in vivo.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for the animal study was objective temperature measurement, not a subjective assessment by human experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. The animal study involved direct physical measurements (temperature), not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The BTL-703 is a physical therapy device that applies heat, not an AI/imaging diagnostic device. Therefore, MRMC studies and AI assistance improvements are not relevant in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The BTL-703 is a physical medical device, not a software algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the animal study: The ground truth was based on objective temperature measurements within the porcine model.
    • For demonstrating substantial equivalence: The ground truth was based on adherence to established medical device safety standards (e.g., IEC, ISO) and comparison of technical specifications and indications for use with the predicate device.

    8. The sample size for the training set

    Not applicable. This device is not an AI/machine learning product and therefore does not have a "training set" in that sense. The non-clinical testing involved rigorous testing against established standards.

    9. How the ground truth for the training set was established

    Not applicable (as it's not an AI/ML device with a training set). The "ground truth" for the non-clinical testing was established by internationally recognized and harmonized medical device safety and performance standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2