Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K162294

    Validate with FDA (Live)

    Date Cleared
    2017-04-21

    (249 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES 12", POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is a disposable device intended for medical purpose, to be worn on the examiners hand or finger to prevent contamination between patient and examiner.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    • Arsenic Trioxide (0.1mg/ml),(100ppm) >240 Minutes
    • Bleomycin Sulfate (15mg/ml),(15,000 ppm) >240 Minutes
    • Busulfan(6mg/ml),(6,000 ppm) >240 Minutes
    • Carmustine (BCNU) (3.3mg/ml),(3,300 ppm) 127.5 Minutes
    • Cisplatin (1.0mg/ml), (1,000 ppm) >240 Minutes
    • Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm) >240 Minutes
    • Cytarabine(100mg/ml), (100,000 ppm) >240 Minutes
    • Dacarbazine(DTIC) 10mg/ml, (10,000 ppm) >240 Minutes
    • Daunorubicin(5mg/ml), (5,000 ppm) >240 Minutes
    • Docetaxel(10mg/ml), (10,000 ppm) >240 Minutes
    • Doxorubicin HCL(2mg/ml),(2,000 ppm) >240 Minutes
    • Ellence (2mg/ml), (2,000 ppm) >240 Minutes
    • Etoposide /Toposar (20mg/ml), (20,000 ppm) >240 Minutes
    • Fludarabine (25mg/ml), (25,000 ppm) >240 Minutes
    • Fluorouracil(50mg/ml), (50,000 ppm) >240 Minutes
    • Gemcitabine/Gemzar (38mg/ml), (38,000 ppm) >240 Minutes
    • Idarubicin (1.0mg/ml), (1,000 ppm) >240 Minutes
    • Ifosfamide (50mg/ml), (50,000 ppm) >240 Minutes
    • Irinotecan(20mg/ml), (20,000 ppm) >240 Minutes
    • Mechlorethamine HCI(1.0mg/ml), (1,000 ppm) >240 Minutes
    • Melphalan(5mg/ml), (5,000 ppm) >240 Minutes
    • Methotrexate(25mg/ml),(25,000 ppm) >240 Minutes
    • Mitomycin C(0.5 mg/ml), (500 ppm) >240 Minutes
    • Mitoxantrone(2mg/ml),(2,000 ppm) >240 Minutes
    • Paclitaxel(6.0mg/ml),(6,000 ppm) >240 Minutes
    • Paraplatin (10mg/ml),(10,000 ppm) >240 Minutes
    • Rituximab(10mg/ml),(10,000 ppm) >240 Minutes
    • Thiotepa (10mg/ml),(10,000 ppm) >240 Minutes
    • Trisenox(0.1mg/ml), (100 ppm) >240 Minutes
    • Vincristine Sulfate(1.0mg/ml).(1.000 ppm) >240 Minutes

    Please note that the following drug noted permeation times below 240 mins: Carmustine (BCNU) 127.50 mins.

    Device Description

    The subject device in this 510(k) Notification is a Purple Nitrile Examination gloves, tested for use with Chemotherapy drugs.

    The subject device is a patient examination glove made from nitrile compound, Purple in color, powder free and non sterile (as per 21 CFR 880.6250, class I).

    The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    This device also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy drugs as per ASTM D6978-05

    AI/ML Overview

    This document describes the premarket notification (510(k)) for BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES 12", POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS, seeking to prove substantial equivalence to a predicate device. The "acceptance criteria" here refers to the performance standards and regulatory requirements that the device must meet to be considered substantially equivalent to a legally marketed predicate device, rather than specific AI/ML performance metrics. Similarly, the "study" proving the device meets these criteria refers to the non-clinical testing performed on the gloves.

    Given the nature of the device (examination gloves), the concepts of "AI assistance," "human observers," "training set," and "ground truth" (in the context of expert consensus or pathology for medical images/diagnoses) are not applicable. The "study" here is a series of non-clinical, laboratory-based tests to demonstrate the physical properties, biocompatibility, and chemical permeation resistance of the gloves.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established by recognized industry standards (ASTM, ISO) and FDA regulations. The reported device performance is the outcome of testing against these standards.

    Table of Acceptance Criteria and Reported Device Performance

    Characteristic / StandardAcceptance Criteria (from Standard)Reported Device Performance (Subject Device)Result of Comparison
    LabelingNo special labeling claims, no hypoanergic claimsNo special labeling claims, no hypoanergic claimsSame
    Device MaterialsNitrile CompoundNitrile CompoundSame
    ColourPurplePurpleSame
    Tensile strength (Before and after ageing) - ASTM 6319-10> 14 Mpa> 14 MpaSame
    Ultimate Elongation (Before and after ageing) - ASTM 6319-10> 500 %> 500 %Same
    Freedom from Pinholes (ASTM 6319-10, FDA 1000ml water leak test)PassPassSame
    Length (ASTM 6319-10)> 295 mm (for 12" glove)> 295 mm (295 mm minimum - 325 mm)Same
    Width (ASTM 6319-10)70±10 mm to 120±10 mm (sizes XS to XL)70±10 mm to 120±10 mm (sizes XS to XL)Same
    Thickness (ASTM 6319-10)> 0.05 mm (Palm 0.12mm Finger : 0.15mm)> 0.05 mm (Palm :0.14mm-0.18mm Finger : 0.20mm-0.24mm)Same
    Residual Powder (ASTM D6124)Less than 2 mg per glove; PASSLess than 2 mg per glove; PASSSame
    Primary Skin Irritation test (ISO 10993-10)Under conditions of the study, not an irritantUnder conditions of the study, not an irritantSame
    Dermal sensitization assay (ISO 10993-10)Under conditions of the study, not a sensitizerUnder conditions of the study, not a sensitizerSame
    Acute Systemic Toxicity (ISO 10993-11)Under conditions of the study, device does not induce acute systemic toxicityUnder conditions of the study, device does not induce acute systemic toxicitySame
    Chemotherapy Drug Permeation (ASTM D6978-05)
    - Arsenic Trioxide (0.1mg/ml)(No specific predicate value provided for comparison)> 240 MinutesN/A (Predicate not tested, Subject Device passed)
    - Bleomycin Sulfate (15mg/ml)> 240 Minutes> 240 MinutesSame
    - Busulfan (6mg/ml)> 240 Minutes> 240 MinutesSame
    - Carmustine (BCNU) (3.3mg/ml)30.7 Minutes (Predicate)127.5 MinutesSubject device showed significantly longer breakthrough time.
    - Cisplatin (1.0mg/ml)> 240 Minutes> 240 MinutesSame
    - Cyclophosphamide/Cytoxan (20mg/ml)> 240 Minutes> 240 MinutesSame
    - Cytarabine (100mg/ml)> 240 Minutes> 240 MinutesSame
    - Dacarbazine (DTIC) (10mg/ml)> 240 Minutes> 240 MinutesSame
    - Daunorubicin (5mg/ml)> 240 Minutes> 240 MinutesSame
    - Docetaxel (10mg/ml)> 240 Minutes> 240 MinutesSame
    - Doxorubicin HCL (2mg/ml)> 240 Minutes> 240 MinutesSame
    - Ellence (2mg/ml)> 240 Minutes> 240 MinutesSame
    - Etoposide /Toposar (20mg/ml)> 240 Minutes> 240 MinutesSame
    - Fludarabine (25mg/ml)> 240 Minutes> 240 MinutesSame
    - Fluorouracil (50mg/ml)> 240 Minutes> 240 MinutesSame
    - Gemcitabine/Gemzar (38mg/ml)> 240 Minutes> 240 MinutesSame
    - Idarubicin (1.0mg/ml)> 240 Minutes> 240 MinutesSame
    - Ifosfamide (50mg/ml)> 240 Minutes> 240 MinutesSame
    - Irinotecan (20mg/ml)> 240 Minutes> 240 MinutesSame
    - Mechlorethamine HCl (1.0mg/ml)> 240 Minutes> 240 MinutesSame
    - Melphalan (5mg/ml)> 240 Minutes> 240 MinutesSame
    - Methotrexate (25mg/ml)> 240 Minutes> 240 MinutesSame
    - Mitomycin C (0.5 mg/ml)> 240 Minutes> 240 MinutesSame
    - Mitoxantrone (2mg/ml)> 240 Minutes> 240 MinutesSame
    - Paclitaxel (6.0mg/ml)> 240 Minutes> 240 MinutesSame
    - Paraplatin (Carboplatin) (10mg/ml)> 240 Minutes> 240 MinutesSame
    - Rituximab (10mg/ml)> 240 Minutes> 240 MinutesSame
    - Thiotepa (10mg/ml)> 240 Minutes> 240 MinutesSame
    - Trisenox (0.1mg/ml)> 240 Minutes> 240 MinutesSame
    - Vincristine Sulfate (1.0mg/ml)> 240 Minutes> 240 MinutesSame
    Warning Statement (ASTM D6978-05)Acknowledge drugs with low permeation timesAcknowledge Carmustine (BCNU) 127.50 mins permeation timeSame (Both predicate and subject device have warning statements for specific drugs, though the specific drug and time differ due to varied performance)

    Study Information (Non-Clinical Testing for Gloves)

    Since this is a filing for medical examination gloves, the concepts below regarding AI/ML studies are not directly applicable. However, I will interpret them in the context of the testing performed on these physical devices.

    1. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: The document does not explicitly state the number of gloves tested for each characteristic (e.g., how many gloves were tested for pinholes, or how many samples for tensile strength). The standards referenced (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10:2010, ASTM D6978-05) would specify the required sample sizes for each test.
      • Data Provenance: The testing was performed according to international harmonized standards (ASTM, ISO). The manufacturer is Brightway Holdings Sdn Bhd. from Malaysia (Lot 1559, Jalan Istimewa, Batu Belah Klang, Selangor, 42100 MY). The data is implicitly prospective, as it was generated specifically for this 510(k) submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable in the AI/ML context. For physical device testing, the "ground truth" is established by the test methods themselves, defined by standards organizations (ASTM, ISO). The "experts" would be the certified laboratory technicians and engineers who perform the tests according to the standard operating procedures and interpret the results against the specified criteria. Their qualifications would be in laboratory science, materials testing, and adherence to quality systems.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable in the AI/ML context. For physical device testing, results are typically quantitative measurements or pass/fail determinations based on predefined criteria in the standards. Discrepancies would be resolved through re-testing or investigation into the test methodology, not through human consensus or adjudication of interpretations.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is a physical medical device (examination glove), not an AI/ML diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

      • The "ground truth" is based on standardized test methods and predefined quantitative/qualitative thresholds set by organizations like ASTM and ISO. For example, for "Freedom from Pinholes," the ground truth is a "Pass" or "Fail" based on a specified water leak test (FDA 1000ml water leak test) as incorporated into ASTM 6319-10. For chemotherapy permeation, the ground truth is the "breakthrough detection time" measured in minutes. Biocompatibility tests rely on established in-vitro and in-vivo assays with pass/fail criteria for irritation, sensitization, and toxicity.
    7. The sample size for the training set:

      • Not applicable. This is a physical device, not an AI/ML model that requires a training set. The manufacturing process is consistent, and quality control ensures consistency, but there isn't a "training set" in the AI sense.
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no training set for an AI/ML model, the concept of establishing ground truth for it does not apply.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162186

    Validate with FDA (Live)

    Date Cleared
    2017-03-31

    (239 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES 12", POWDER FREE, [STERLING/GREY] WITH CHEMOTHERAPY CLAIM is a disposable device intended for medical purpose, to be worn on the examiners hand or finger to prevent contamination between patient and examiner.

    The product has been tested with chemotherapy drugs in accordance with ASTM D6978. The breakthrough detection times are as follows:

    Test Chemotherapy Drug Name and Concentration Minimum Breakthrough Detection Time Carmustine (BCNU) (3.3mg/ml),(3,300 ppm) ● 4.3 Minutes >240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    240 Minutes

    10.9 Minutes

    240 Minutes

    • Cisplatin(1.0mg/ml), (1,000 ppm)
    • Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm)
    • Dacarbazine(DTIC) 10mg/ml, (10,000 ppm) ●
    • Doxorubicin HCL(2mg/ml),(2,000 ppm)
    • Etoposide /Toposar (20mg/ml), (20,000 ppm)
    • Fluorouracil(50mg/ml), (50,000 ppm)
    • Ifosfamide (50mg/ml), (50,000 ppm)
    • Mitoxantrone(2mg/ml),(2,000 ppm)
    • Paclitaxel(6.0mg/ml),(6,000 ppm)
    • Thiotepa (10mg/ml),(10,000 ppm)
    • . Vincristine Sulfate(1.0mg/ml),(1,000 ppm)

    Please note that Carmustine and Thiotepa have extremely low permeation times of 4.3 minutes and 10.9 minutes, respectively.

    Warning: Do not use with Carmustine and Thiotepa.

    Device Description

    The subject device in this 510(k) Notification is Sterling/Grey coloured Nitrile Examination gloves, with claims and tested for use with Chemotherapy drugs.

    The subject device is a patient examination glove made from nitrile compound, Sterling or Grey in colour, powder free and non sterile (as per 21 CFR 880.6250, class I).

    The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    This device also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy drugs as per ASTM D6978-05

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES 12", POWDER FREE, [STERLING/GREY] WITH CHEMOTHERAPY CLAIM meets these criteria.

    Here's the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsStandardAcceptance Criteria (Predicate Device K081089)Reported Device Performance (Subject Device)Result of Comparison
    LabelingN/ANo special labeling claims, no claim of hypoallergenic.No special labeling claims, no claim of hypoallergenic.Same
    Device MaterialsN/ANitrile CompoundNitrile CompoundSame
    ColourN/ASterlingSterling / GreySame
    Tensile strength (Before and after ageing)ASTM 6319-10> 14 MPa> 14 MPaSame
    Ultimate Elongation (Before and after ageing)ASTM 6319-10> 500 %> 500 %Same
    Freedom from PinholesASTM 6319-10 (FDA 1000 ml water leak test)PassPassSame
    Dimensions: LengthASTM 6319-10> 295 mm> 295 mm (295 mm minimum - 325 mm)Same
    Dimensions: WidthASTM 6319-1070±10 mm to 120±10 mm (sizes XS to XL)70±10 mm to 120±10 mm (sizes XS to XL)Same
    Dimensions: ThicknessASTM 6319-10> 0.05 mm (palm & finger)> 0.05 mm (Palm: 0.075mm-0.08mm; Finger: 0.085mm-0.10mm)Same
    Residual PowderASTM 6319-10Less than 2 mg per glove; PASSLess than 2 mg per glove; PASSSame
    Primary Skin Irritation testISO 10993-10Under conditions of the study, not an irritantUnder conditions of the study, not an irritantSame
    Dermal sensitization assayISO 10993-10Under conditions of the study, not a sensitizerUnder conditions of the study, not a sensitizerSame
    Systemic ToxicityISO 10993-11PassPassSame
    Chemical Permeation / Chemotherapy Testing (Breakthrough Detection Time)ASTM D6978-05Carmustine and Thiotepa have extremely low permeation times of less than 30 minutesCarmustine: 4.3 Minutes; Cisplatin: >240 Minutes; Cyclophosphamide/Cytoxan: >240 Minutes; Dacarbazine: >240 Minutes; Doxorubicin HCL: >240 Minutes; Etoposide /Toposar: >240 Minutes; Fluorouracil: >240 Minutes; Ifosfamide: >240 Minutes; Mitoxantrone: >240 Minutes; Paclitaxel: >240 Minutes; Thiotepa: 10.9 Minutes; Vincristine Sulfate: >240 MinutesSame (for Carmustine and Thiotepa, other drugs exceed the predicate's general criteria)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test mentioned (e.g., how many gloves were tested for pinholes, tensile strength, or chemotherapy permeation). It references standards like ASTM D6319-10 and ASTM D6978-05, which would have their own defined sample size requirements.

    The data provenance is from Malaysia, as the submitter, BRIGHTWAY HOLDINGS SDN. BHD., is located in Klang, Selangor Darul Ehsan, Malaysia. The data appears to be prospective as it involves direct testing of the device against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The acceptance criteria and performance are based on recognized industry standards (ASTM, ISO), meaning the "ground truth" is defined by these standards rather than expert consensus on specific cases. The testing was performed according to these established protocols.

    4. Adjudication method for the test set

    This information is not applicable/not provided as the evaluation is based on objective measurements against predefined standards (ASTM, ISO) rather than subjective assessment requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an examination glove, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is an examination glove, not an algorithm.

    7. The type of ground truth used

    The ground truth used is based on established industry standards and regulatory requirements. Specifically:

    • Physical properties (tensile strength, elongation, dimensions, freedom from pinholes, residual powder) are compared against specifications in ASTM D6319-10.
    • Biocompatibility (primary skin irritation, dermal sensitization, systemic toxicity) is evaluated against ISO 10993-10 and ISO 10993-11 requirements.
    • Resistance to chemotherapy drug permeation is tested according to ASTM D6978-05.
    • The overall criteria for patient examination gloves are also linked to FDA 21 CFR 880.6250.

    8. The sample size for the training set

    This is not applicable. The device is a physical product (gloves), not a machine learning model that requires a training set. The "training" in this context refers to the manufacturing process and quality control, not data training for an algorithm.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163267

    Validate with FDA (Live)

    Date Cleared
    2017-03-24

    (123 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [BLUE] is a disposable device intended for medical purpose to be worn on the examiners hand or finger to prevent contamination between the patient and the examiner.

    Device Description

    The subject device in this 510(k) Notification is a Blue Examination Glove. The subject device is a patient examination glove made from a Nitrile compound, Blue in colour, powder free and non sterile (as per 21 CFR 880.6250, class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device that meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10 ; Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the Brightway Nitrile Examination Glove, powder-free (blue).

    Here's the breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally based on the ASTM D6319-10 standard for Nitrile Examination Gloves. The device performance is reported as meeting these specifications.

    CharacteristicAcceptance Criteria (Standard / Description)Reported Device Performance
    Dimension MeasurementASTM D6319-10 (Specific ranges for weight, length, width, thickness by size)All samples passed all dimension measurements. No non-conformance reported.
    Water-tightness (Freedom from Pinholes)ASTM D5151-06; G2 AQL 1.0, Ac=5 Rej=61 total defective found across all sizes (in one sample's palm). Acceptance number is 5. Status: Pass.
    Tensile Strength (Before Aging)ASTM D6319-10: 14 MPa (Min)Average Result: 24.351 MPa. Status: Pass.
    Ultimate Elongation (Before Aging)ASTM D6319-10: 500% (Min)Average Result: 550.6%. Status: Pass.
    Tensile Strength (After Aging)ASTM D6319-10: 14 MPa (Min)Average Result: 22.134 MPa. Status: Pass.
    Ultimate Elongation (After Aging)ASTM D6319-10: 400% (Min)Average Result: 484.5%. Status: Pass.
    LabelingNo special labeling claims or claims of gloves as hypoallergenic.No special labeling claims or claims of gloves as hypoallergenic.
    Device MaterialsNitrile CompoundNitrile Compound
    Indications for UseIntended for medical purpose, to be worn on examiners hand or finger to prevent contamination between patient and examiner.Intended for medical purpose, to be worn on examiners hand or finger to prevent contamination between patient and examiner.
    ColorBlueBlue
    LengthASTM D6319-10: > 230 mm> 230 mm (230-258mm)
    WidthASTM D6319-10: 70±10 mm to 120±10 mm (sizes XS to XL)70±10 mm to 120±10 mm (sizes XS to XL)
    ThicknessASTM D6319-10: > 0.05 mm (palm & finger)> 0.05 mm (Palm: 0.08mm-0.16mm; Finger: 0.12mm-0.19mm)
    Residual PowderASTM D6319-10, ASTM D6124: Less than 2 mg per gloveLess than 2 mg per glove. Status: Pass.
    Primary Skin Irritation testISO 10993-10: Not an irritantUnder conditions of the study, not an irritant.
    Dermal Sensitization AssayISO 10993-10: Not a sensitizerUnder conditions of the study, not a sensitizer.
    Systemic ToxicityISO 10993-10: No systemic toxicity revealedUnder conditions of study, did not reveal any systemic toxicity.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Dimension Measurement: 20 pieces per batch. Random Sampling based on ISO2859-1:1999; S2 AQL 2.5, Ac=1 Rej=2. Data provided is for Batch No: 6H10 03.
    • Water-tight Test (Freedom from Holes): 200 pieces per batch. Random Sampling based on ISO2859-1:1999; G2 AQL 1.0, Ac=5 Rej=6.
    • Physical Properties (Tensile Strength, Ultimate Elongation): 20 samples (1 sample for each size XS, S, M, L, XL x 4 replicates per size, but the table shows 20 distinct samples with sizes varying, suggesting 4 per size covering XS-XL). Sampling Plan: Single Normal S2 AQL: 2.5 Acc/Rej: 1/2.
    • Biocompatibility (Primary Skin Irritation, Dermal Sensitization, Systemic Toxicity): Specific sample sizes are not explicitly stated for these tests within the provided document, but the tests were performed "under the conditions of the studies."
    • Data Provenance: The tests were conducted by Brightway Holdings Sdn Bhd, which is based in Malaysia. The testing certificates are dated August 22, 2016, and the overall submission is dated March 24, 2017. This indicates the studies were conducted prior to the submission, likely as part of the device development and verification, making them retrospective in the context of this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is a patient examination glove. The "ground truth" for its performance is established by adherence to recognized industry standards (like ASTM D6319-10, ASTM D5151-06, ISO 10993-10). These standards define objective, measurable physical, chemical, and biological properties. Therefore, no human experts (like radiologists) are typically involved in establishing a "ground truth" for the physical and biological characteristics of a glove. The "ground truth" is the quantitative measurement against the standard specification.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Since the assessment relies on objective measurements against predefined standards (e.g., tensile strength, dimensions, water leak test), there is no adjudication method described or required by human experts. The results are quantitative and directly compared to the specified limits.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a physical medical device (examination glove), not an AI-powered diagnostic tool, and therefore, an MRMC study is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical examination glove, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on objective measurements and laboratory test results compared against established international and national standards for medical gloves. This includes:

    • Physical properties (e.g., length, width, thickness, weight, tensile strength, ultimate elongation)
    • Barrier integrity (e.g., freedom from pinholes via water leak test)
    • Chemical properties (e.g., residual powder)
    • Biocompatibility (e.g., primary skin irritation, dermal sensitization, systemic toxicity).

    8. The sample size for the training set

    Not applicable. This device is a physical medical device (examination glove), not a machine learning model, and therefore does not have a "training set" in the context of AI/ML.

    9. How the ground truth for the training set was established

    Not applicable. As stated above, this device does not use a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162146

    Validate with FDA (Live)

    Date Cleared
    2017-01-05

    (157 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [LAVENDER] is a disposable device intended for medical purpose, to be worn on the examiners hand or finger to prevent contamination between patient and examiner.

    Device Description

    The subject device in this 510(k) Notification is a Lavender Examination Glove. The subject device is a patient examination glove made from a Nitrile compound, Lavender in colour, powder free and non sterile (as per 21 CFR 880.6250, class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device that meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10 ; Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Brightway Brand Nitrile Examination Gloves, Powder Free (Lavender), based on the provided document:

    Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Standard)Reported Device PerformanceResult
    Dimension and Thickness
    Weight (for each size XS-XL)$Range around target (e.g., XS: $2.7 \pm 0.3g$)All samples within specificationPass
    Length (Min-Max)Min - 230mm, Tar - 240mm, Max - 258mmAll samples within specificationPass
    Width (Min-Max for each size XS-XL)Range around targetAll samples within specificationPass
    Finger Tip Thickness (Min-Max)Min - 0.06mm, Tar - 0.08mm, Max - 0.11mmAll samples within specificationPass
    Palm Thickness (Min-Max)Min - 0.05mm, Tar - 0.07mm, Max - 0.09mmAll samples within specificationPass
    Cuff Thickness (Min-Max)Min - 0.05mm, Tar - 0.06mm, Max - 0.08mmAll samples within specificationPass
    Freedom from HolesASTM D5151; AQL 1.5, Ac=7 Rej=8 for 200 pcs1 defective foundPass
    Physical Properties (Before Aging)
    Tensile Strength (MPa)16 (Min) [ASTM D 6319-10]27.805 (Avg.)Pass
    Ultimate Elongation (%)500 (Min) [ASTM D 6319-10]588.8 (Avg.)Pass
    Physical Properties (After Aging)
    Tensile Strength (MPa)15 (Min) [ASTM D 6319-10]26.075 (Avg.)Pass
    Ultimate Elongation (%)450 (Min) [ASTM D 6319-10]523.5 (Avg.)Pass
    Biocompatibility
    Primary Skin Irritation testISO 10993-10: Not an irritantNot an irritantPass
    Dermal Sensitization assayISO 10993-10: Not a contact sensitizerNot a contact sensitizerPass
    Systemic ToxicityISO 10993-11: Non-systemically toxicNon-systemically toxicPass
    Residual PowderASTM D6124: Meets requirementsMeets requirementsPass

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Dimension Measurement: 20 pieces per batch. Random Sampling based on ISO2859-1:1999; S2 AQL 2.5, Ac=1 Rej=2.
      • Water-tight Test: 200 pieces per batch. Random Sampling based on ISO2859-1:1999; G2 AQL 1.5, Ac=7 Rej=8.
      • Physical Properties Analysis: Not explicitly stated but "Single Normal S2 AQL: 2.5 Acc/Rej : 1/2" for sampling plan. Given that 20 samples were tested and individual results are shown, it seems 20 samples were used for the physical properties test.
      • Biocompatibility Studies (Primary Skin Irritation, Dermal Sensitization, Systemic Toxicity): Sample sizes for these tests are not provided in the document.
      • Data Provenance: The tests were conducted by Brightway Holdings Sdn Bhd (the manufacturer) in Malaysia. The document does not specify if the data is retrospective or prospective, but as it's part of a 510(k) submission for a new device, it would typically involve prospective testing of newly manufactured devices.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This information is not applicable to the device described. The "ground truth" for examination gloves is established by standardized physical and chemical testing methods (ASTM, ISO), not by expert consensus or interpretations.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Performance is measured against objective standards, not subjective interpretations requiring adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is an examination glove, not an AI-powered diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is an examination glove, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for this device is based on objective, standardized measurements and tests outlined in recognized standards such as ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves), ASTM D5151 (Water Leak Test), ASTM D6124 (Residual Powder), and ISO 10993 (Biocompatibility).
    7. The sample size for the training set:

      • Not applicable. This device is not an AI/machine learning model that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. This device is not an AI/machine learning model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161215

    Validate with FDA (Live)

    Date Cleared
    2016-12-21

    (236 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [STERLING/GREY] TESTED FOR USE WITH CHEMOTHERAPHY DRUGS is a disposable device intended for medical purpose worn on the examiners hand or finger to prevent contamination between patient and examiner.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Test Chemotherapy Drug Name and Concentration

    • · Carmustine (BCNU) (3.3mg/ml),(3.300 ppm)
    • · Cisplatin(1.0mg/ml), (1,000 ppm)
    • · Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm)
    • · Dacarbazine(DTIC) 10mg/ml, (10,000 ppm)
    • · Doxorubicin HCL(2mg/ml),(2,000 ppm)
    • · Etoposide /Toposar (20mg/ml), (20,000 ppm)
    • · Fluorouracil(50mg/ml), (50,000 ppm)
    • · Ifosfamide (50mg/ml), (50,000 ppm)
    • · Mitoxantrone(2mg/ml),(2,000 ppm)
    • · Paclitaxel/Taxol(6.0mg/ml),(6.000 ppm)
    • · Thiotepa (10mg/ml),(10,000 ppm)
    • · Vincristine Sulfate(1.0mg/ml),(1,000 ppm)

    Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) 5.4 minutes and Thiotepa 40.4 minutes.

    Minimum Breakthrough Detection Time
    5.4 Minutes

    240 Minutes
    240 Minutes
    240 Minutes
    240 Minutes
    240 Minutes
    240 Minutes
    240 Minutes
    240 Minutes
    240 Minutes
    40.4 Minutes
    240 Minutes

    Device Description

    The subject device in this 510(k) Notification is a Sterling/Grey Nitrile Examination gloves, with claims and tested for use with Chemotherapy drugs.

    The subject device is a patient examination glove made from nitrile compound, Sterling or Grev in color, powder free and non sterile (as per 21 CFR 880.6250, class I).

    The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    This device is manufactured in facilities compliant to ISO 9001:2008 certified in Manufacture of Non Sterile Natural (Latex) and Synthetic Latex (Nitrile) Examination, Surgical and Industrial Gloves & Nitrile Sheath

    The device is manufactured to comply with ISO 13485:2003 / EN ISO 13485:2012 Manufacture of Non Sterile Natural (Latex) and Synthetic Latex (Nitrile) Examination & Sterile Surgical Gloves

    AI/ML Overview

    This document describes the premarket notification (510(k)) for "BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [STERLING/GREY] TESTED FOR USE WITH CHEMOTHERAPHY DRUGS". The acceptance criteria and the study proving the device meets these criteria are outlined through a comparison with a legally marketed predicate device (K151750).

    1. Table of Acceptance Criteria and Reported Device Performance

    Characteristic / TestAcceptance Criteria (Predicate K151750 and general standards)Reported Device Performance (Subject Device K161215)Comments / Proof of Meeting Criteria
    General
    Device TypePatient Examination Glove (21 CFR 880.6250)Nitrile Examination Gloves, Powder FreeSame
    Intended UsePrevent contamination between patient and examiner. Tested for use with chemotherapy drugs.Prevent contamination between patient and examiner. Tested for use with chemotherapy drugs.Same
    MaterialNitrileNitrileSame
    SterilityNon-SterileNon-SterileSame
    Single UseYesYesSame
    DesignMeet Requirements of ASTM D6319Meet Requirements of ASTM D6319Same (Implied by standard compliance)
    Physical PropertiesMeets Requirements of ASTM D6319Meets Requirements of ASTM D6319Same
    Dimensions
    Lengthmin 230 mm / ≥ 230mm, Meets Requirements of ASTM D6319minimum 230 mm, target 242 mm, maximum 258 mm, Meets Requirements of ASTM D6319Same
    Thickness (Cuff)0.032-0.050 mm0.05-0.07 mmMeets requirements of ASTM D6319. (While the ranges differ, both are stated to meet the standard, indicating the subject device's range is within acceptable limits of the standard.)
    Thickness (Palm)0.055-0.085 mm0.06-0.09 mmMeets requirements of ASTM D6319.
    Thickness (Finger)0.065-0.095 mm0.07-0.10 mmMeets requirements of ASTM D6319.
    Performance - Barrier
    Powder Free Residue≤ 2mg/glove≤ 2mg/gloveSame
    Freedom from pinholes testingTested in accordance with ASTM D5151 test method. Pass quality level at G1 AQL 1.5Tested in accordance with ASTM D5151 test method. Pass quality level at G2 AQL 1.0Subject device tested to a more stringent criteria of G2 AQL 1.0
    Performance - Chemotherapy Drug Permeation(Based on predicate performance and "above 30 minutes permeation time" criteria for some drugs)
    Carmustine (BCNU) (3.3mg/ml)10.1 minutes (Predicate) / "Below 30 minutes permeation time"5.4 minutesBelow 30 minutes permeation time, same as the predicate device. A warning statement is included.
    Cisplatin (1.0mg/ml)>240 minutes>240 minutesSame
    Cyclophosphamide/Cytoxan (20mg/ml)>240 minutes>240 minutesSame
    Dacarbazine (DTIC) 10mg/ml>240 minutes>240 minutesSame
    Doxorubicin HCL (2mg/ml)>240 minutes>240 minutesSame
    Etoposide /Toposar (20mg/ml)Not explicitly listed for predicate, assumed >240 as per other drugs>240 minutesImplicitly meets, with similar performance to most other listed drugs.
    Fluorouracil (50mg/ml)>240 minutes>240 minutesSame
    Ifosfamide (50mg/ml)>240 minutes>240 minutesSame
    Mitoxantrone (2mg/ml)>240 minutes>240 minutesSame
    Paclitaxel/Taxol (6.0mg/ml)>240 minutes>240 minutesSame
    Thiotepa (10mg/ml)30.2 minutes (Blue Predicate), 10.4 minutes (White Predicate)40.4 minutesSubject device had a permeation time of 40.4 mins which exceeded 30 mins, but less than 60 mins. A warning statement is included.
    Vincristine Sulfate (1.0mg/ml)>240 minutes>240 minutesSame
    Biocompatibility
    Dermal Sensitization (ISO 10993-10:2010)Not a contact sensitizer under the conditions of the studyNot a contact sensitizer under the conditions of the studySame
    Primary Skin Irritation Test (ISO 10993-10:2010)Not a primary skin irritant under the conditions of the studyNot a primary skin irritant under the conditions of the studySame
    Systemic Toxicity (ISO 10993-11)(Predicate not tested)No mortality or no evidence of Systemic Toxicity was observedPredicate device was not tested for Systemic Toxicity. Subject device was tested for Acute Systemic Toxicity and no mortality or evidence of Systemic Toxicity was observed.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes for the various tests (e.g., pinhole, chemotherapy permeation, physical properties, biocompatibility). It references standardized test methods (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10:2010, 16 CFR Part 1500.41) which typically define appropriate sample sizes. The data provenance is non-clinical performance data, likely generated prospectively by the manufacturer and/or a testing laboratory contracted by the manufacturer. The country of origin of the data is not specified beyond the manufacturer being based in Malaysia.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable. The device is a patient examination glove, and its performance is evaluated against objective, standardized physical, chemical, and biological tests, not against expert interpretation of medical images or other diagnostic data where ground truth would be established by human experts.

    4. Adjudication method for the test set

    This section is not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving human interpretation or subjective assessments, often in clinical trials or comparative imaging studies. The testing described here is based on objective, standardized laboratory methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. MRMC studies and the concept of AI assistance for human readers are relevant for diagnostic devices (e.g., AI for radiology interpretation). The device in question is a medical glove, for which such a study design is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This section is not applicable. The device is a physical medical glove, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance claims of this device is established through:

    • Standardized Test Methods and Specifications: Adherence to ASTM D6319-10 (Standard specification for Nitrile Examination Gloves), ASTM D5151-06 (standard for pinholes), ASTM D6124-06 (standard for powder residue), ISO 10993-10:2010 (biocompatibility - irritation and sensitization), ISO 10993-11 (biocompatibility - systemic toxicity), and 16 CFR Part 1500.41 (flammability).
    • Predicate Device Performance: The performance of the legally marketed predicate device (K151750) serves as a benchmark for substantial equivalence, particularly concerning chemotherapy drug permeation.

    8. The sample size for the training set

    This section is not applicable. A "training set" refers to data used to train machine learning models. This device is a physical product and does not involve AI or machine learning algorithms.

    9. How the ground truth for the training set was established

    This section is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1