510(k) Data Aggregation

    K Number
    K111609

    Validate with FDA (Live)

    Device Name
    BOSTON BAND
    Date Cleared
    2011-08-22

    (74 days)

    Product Code
    Regulation Number
    882.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Boston Band is intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial proportion and symmetry in infants from three to eighteen months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephic, brachycephalic and scaphocephalic patterned head shapes. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.

    Device Description

    The Boston Band is a cranial orthosis used to treat abnormally shaped craniums in infants three to eighteen months of age. This condition is clinically known as Positional or Deformational Plagiocephaly. The orthosis contains the protruding aspects of the cranium in a static equilibrium while guiding the growth of the flattened-areas of the skull into the created spaces to improve proportion and symmetry. The Boston Band is only available if prescribed by a physician. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.

    The Boston Band is essentially the same device as released by K072862. It is an orthosis designed for each patient from a cast or scan of the infant's head. The mold, either plaster from a cast, or foam, carved from the scan, is modified and prepared for fabrication according to the instructions provided by the treating practitioner using plaster/CAD modification techniques. Each orthosis is composed of an outer shell of thermoformable plastic, a single layer of ½ inch or 5-6 layers of 1/8 inch, or a combination of 1 - 2 1/4 inch layers with 1/8 inch layers of hypoallergenic polyethylene foam (Aliplast/Plastazote) and a strap for securing the orthosis. It has a top and side opening. Optimum fit and alignment is insured and monitored by the same clinical practitioner.

    AI/ML Overview

    The Boston Band is a cranial orthosis used to treat abnormally shaped craniums in infants between three and eighteen months of age. The document outlines a 510(k) submission for the Boston Band, focusing on proposed changes and comparing it to predicate devices.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of "acceptance criteria" in the typical sense of quantitative thresholds for clinical performance metrics (e.g., specific percentages of cranial symmetry improvement). Instead, the submission focuses on demonstrating substantial equivalence to predicate devices, particularly regarding safety and effectiveness with expanded indications and updated manufacturing processes.

    However, based on the non-clinical performance data section, we can infer some criteria and how the device met them:

    Acceptance Criterion (Inferred)Reported Device Performance
    Laser Scanner Safety (Class I)The Fastscan hands-free class I laser scanner was found to be "inherently safe for use without eye protection under all normal operating conditions." This aligns with Class I laser safety standards.
    Laser Scanner Effectiveness (Accuracy & Repeatability)"The effectiveness of this scanning device was evaluated through accuracy and repeatability testing. Positive foam models were scanned using the Fastscan hands free class I laser scanner. The scanned models were then carved and measured to compare to the original model. Each device met the predetermined acceptance criteria and was found acceptable." (Specific metrics for accuracy and repeatability are not provided, but the statement indicates meeting pre-defined thresholds.)
    CADCAM Software Effectiveness (Accuracy & Repeatability)"The effectiveness of this modification software was evaluated through accuracy and repeatability testing. A comparison was made between hand modified and Rodin4D modifications of the same models. Four models were selected. Hand modifications were done on carved models and CAD modifications were done on the electronic models. The electronic models were then carved. All models where then measured and compared. The results showed that both modification techniques produced an acceptable model for fabrication." (Specific metrics are not provided.)
    Manufacturing Process Equivalence (Carver Machine)The updated 5-axis carver machine maintains equivalence with the predicate device's manufacturing process, with the "only major change is the speed of carving." This implies that the output quality and structural integrity remain consistent.
    Clinical Equivalence to Predicate DevicesThe Boston Band is stated to be "essentially the same device as released by K072862." The comparisons tables (Tables 1, 3, 5) consistently show "Same" for contraindications, warnings, precautions, adverse effects, instructions for use, and expiration date, indicating no new safety or effectiveness concerns with the proposed changes or expanded indications.
    Expanded Indications Safety and EffectivenessThe core mechanism of the device (passively holding prominent cranial regions) remains the same. The expanded indications to include surgically corrected synostosis cases are considered safe and effective due to the device's fundamental function and similarity to predicate devices already approved for similar uses.

    2. Sample Size for Test Set and Data Provenance:

    • Laser Scanner Testing: "Positive foam models were scanned." The exact number of models is not specified, but the plural "models" indicates more than one.
    • CADCAM Software Testing: "Four models were selected."
    • Data Provenance: The document does not specify the country of origin for the data or whether it was retrospective or prospective. Given the nature of a 510(k) summary, which focuses on device characteristics and substantial equivalence, detailed clinical study data with provenance is often found in separate full study reports, not necessarily in this summary document. The context suggests these were internal engineering validation tests rather than large-scale clinical trials.

    3. Number of Experts and Qualifications for Ground Truth:

    • The document does not explicitly mention experts establishing ground truth for the test set for the scanner or CADCAM software testing. The "ground truth" for these tests seems to be the original physical models (for scanner accuracy) and the hand-modified carved models (for CADCAM modification equivalence).
    • For the overall device, the "predetermined acceptance criteria" mentioned for the scanner testing would likely have been established by Boston Brace International's internal engineering and quality control personnel. The "treating practitioner" and "clinical practitioner" are mentioned in the device description for prescribing, monitoring, and fitting, implying their expertise in clinical application but not in establishing ground truth for these specific non-clinical validation tests.

    4. Adjudication Method for the Test Set:

    • The document does not describe any formal adjudication method (e.g., 2+1, 3+1) for the non-clinical tests mentioned. The comparison was likely direct measurement against a reference or between two methods.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described in this 510(k) summary. This type of study is typically used for AI-powered diagnostic tools where human readers interpret medical images with and without AI assistance. The Boston Band is a physical cranial orthosis, not an imaging analysis tool. Therefore, a study of this nature would not be relevant.

    6. Standalone Performance Study (Algorithm Only):

    • No, a standalone (algorithm only without human-in-the-loop) performance study was not conducted in the context of an AI algorithm. The "Fastscan hands-free class I laser scanner" and "Rodin4D CADCAM software" are components of the manufacturing process, not standalone AI algorithms intended for clinical interpretation. Their performance was evaluated through accuracy and repeatability of scanning and modification, respectively.

    7. Type of Ground Truth Used:

    • For Laser Scanner Testing: The "original model" (presumably the physical foam model before scanning) served as the ground truth against which the carved models from scanned data were compared for accuracy.
    • For CADCAM Software Testing: The "hand modified" carved models served as the ground truth for comparison with the CAD-modified and subsequently carved electronic models.

    8. Sample Size for the Training Set:

    • The document does not mention any "training set." This device does not appear to involve machine learning or AI models that require training data in the conventional sense. The "modification software" is described as 3D rectification software that records modifications, implying a rule-based or guided modification process rather than an adaptive learning algorithm.

    9. How Ground Truth for Training Set Was Established:

    • As there is no mention of a training set, the establishment of ground truth for a training set is not applicable to this submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072862

    Validate with FDA (Live)

    Date Cleared
    2008-01-22

    (109 days)

    Product Code
    Regulation Number
    882.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Boston Band Cranial Molding Orthosis is intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial symmetry and/or shape in infants from three to eighteen months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephic, brachycephalic and scaphocephalic patterned head shapes.

    Device Description

    The Static Cranioplasty Orthosis is a cranial orthosis used to treat abnormally shaped craniums in infants three to 18 months of age. This condition is clinically known as positional or Deformational Plagiocephaly. The orthosis contains the protruding aspects of the cranium in a static equilibrium while guiding the growth of the flattened-areas of the skull into the created spaces. The Static Cranioplasty Orthosis is only available if prescribed by a physician.

    The orthosis is custom designed for each patient from a cast of the infant's head , a scan using the STARscanner from Orthomerica, or a scan from the Hand held FastScan Handheld Laser Scanner. The mold, either plaster from the cast, or foam from the scan, is modified and prepared for fabrication by the treating practitioner using mathernatical analyses and plaster modification techniques. The orthosis is then fabricated under the direction of the same practitioner. Each orthosis is composed of an outer shell of thermoformable plastic, 5-6 layers of hypoallergenic polyethylene foam and a strap for securing the orthosis. Optimum fit and alignment is insured and monitored by the same clinical practitioner.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text, structured as requested:

    1. Table of Acceptance Criteria & Reported Device Performance

    Acceptance Criteria (Stated or Implied)Reported Device Performance
    Effectiveness: Significantly effective in correcting abnormal head shape."The effects of treatment with cranial orthoses on infants have concluded that the devices are significantly effective in correcting abnormal head shape, without evidence of relapse following treatment."
    Safety: Device and materials are not expected to adversely affect infants; materials not reported to cause skin irritation or toxic effects; designed to avoid improper migration or harmful pressure."The safety of the cranial orthoses is established under standard biocompatibility assessments for each material used... These assessments reveal that the device and the materials used are not expected to adversely affect the infants under the intended conditions of use... The materials are not reported to cause skin irritation or any toxic effects. Further, the product is designed to avoid improper migration or harmful levels of pressure."
    Substantial Equivalence: Analogous to predicate device in production, instructions for use, materials, safety, and effectiveness."The Static Cranioplasty Orthosis and the predicate device are very similar with respect to production, instructions for use, materials, safety and effectiveness and special controls." "The proposed indications of use are analogous to those presented by the predicate device, and biocompaibility, function and effectiveness further parallel those of the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in the provided text. The text refers to "clinical trials" and "infants" in a general sense, but no specific number of patients or cases is mentioned for the effectiveness study.
    • Data Provenance: Retrospective or Prospective is not specified. The study cited, "Orthotic Treatment Protocols for Plagiocephaly, Jeff Larson, CO:JPO 2004, Vol 16, Num 4s," is a published clinical trial, suggesting prospective data collection, but this is an inference, not explicitly stated in the summary. The origin (country) is not specified, but the applicant is based in Avon, MA, USA, suggesting a US-based study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not explicitly stated. The document refers to "clinical trials" which implies that clinicians (likely orthopedic specialists, pediatricians, or orthotists) would be involved in patient assessment and outcome evaluation, but their specific roles in establishing ground truth (e.g., severity of plagiocephaly, correction) and their qualifications are not detailed.

    4. Adjudication Method for the Test Set

    Not explicitly stated. The nature of the study (clinical trial for a physical orthosis) is less likely to involve a formal adjudication method like 2+1 or 3+1 typically seen in diagnostic AI studies, as the "ground truth" would be the patient's physical head shape and its improvement over time, likely measured by practitioners or specialized equipment.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. The document describes a clinical trial evaluating the effectiveness of the device itself, not comparing human readers' performance with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) Was Done

    This question is not applicable to the device described. The Boston Band Cranial Remolding Orthosis is a physical medical device, not an AI algorithm. Therefore, there is no "standalone" algorithm performance to evaluate. Its "performance" is its physical effect on the infant's skull when worn.

    7. The Type of Ground Truth Used

    The ground truth appears to be based on clinical outcomes data related to the correction of abnormal head shape. The text states, "The effects of treatment with cranial orthoses on infants have concluded that the devices are significantly effective in correcting abnormal head shape..." This suggests that clinical assessments and measurements (though not detailed) of head shape before and after treatment served as the ground truth for effectiveness.

    8. The Sample Size for the Training Set

    This question is not applicable. The Boston Band Cranial Remolding Orthosis is a physical medical device, not an AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063098

    Validate with FDA (Live)

    Date Cleared
    2007-04-11

    (183 days)

    Product Code
    Regulation Number
    882.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infants. 3 - 18 months of age, with abnormally shaped craniums with no outstanding cranial abnormalities that have a RX from a physician who has assessed the child to rule out craniosynostosis.

    Device Description

    The Static Cranioplasty Orthosis is a cranial orthosis used to treat abnormally shaped craniums in infants three to 18 months of age. This condition is clinically known as positional or Deformational Plagiocephaly. The orthosis contains the protruding aspects of the cranium in a static equilibrium while guiding the growth of the flattenedareas of the skull into the created spaces. The Static Cranioplasty Orthosis is only available if prescribed by a physician.

    The orthosis is custom designed for each patient from a cast of the infant's head or a scan using the STARscanner from Orthomerica . The mold, either plaster from the cast, or foam from the scan, is modified and prepared for fabrication by the treating practitioner using mathematical analyses and plaster modification techniques. The orthosis is then fabricated under the direction of the same practitioner. Each orthosis is composed of an outer shell of thermoformable plastic, 5-6 layers of hypoallergenic polyethylene foam and a strap for securing the orthosis. Optimum fit and alignment is insured and monitored by the same clinical practitioner.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Boston Band Cranial Remolding Orthosis. This document is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device, not a study designed to prove the device meets specific acceptance criteria with reported performance metrics in the way a clinical trial for a novel AI or diagnostic device would.

    Therefore, many of the requested details (like sample sizes for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, or standalone performance) are not applicable or present in this type of regulatory document.

    However, I can extract the information related to the device's effectiveness as established by referencing a clinical trial and comparing it to a predicate device.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not present specific quantitative acceptance criteria or reported device performance in a dedicated table format. Instead, it claims substantial equivalence to a predicate device and references a clinical trial to establish "effectiveness." The "acceptance criteria" here are implicitly meeting the standard of effectiveness and safety demonstrated by the predicate and supported by existing clinical literature.

    AspectAcceptance Criteria (Implied)Reported Device Performance
    EffectivenessSignificant effectiveness in correcting abnormal head shape (e.g., positional plagiocephaly, brachycephaly, scaphocephaly) in infants aged 3-18 months, without evidence of relapse post-treatment. Ability to improve results of surgical correction in severe cases, potentially eliminating further surgical intervention."The effects of treatment with cranial orthoses on infants have concluded that the devices are significantly effective in correcting abnormal head shape, without evidence of relapse following treatment."
    SafetyBiocompatibility of materials, no skin irritation or toxic effects, designed to avoid improper migration or harmful pressure levels, smooth interior."The safety of the cranial orthoses is established under standard biocompatibility assessments for each material used... not expected to adversely affect the infants... materials are not reported to cause skin irritation or any toxic effects."
    Intended UseTo passively hold prominent cranial regions of an infant's skull to improve cranial symmetry and/or shape.The device is intended for this purpose, and its effectiveness (as described above) supports this intention.
    IndicationsInfants, 3 - 18 months of age, with abnormally shaped craniums with no outstanding cranial abnormalities, and a physician's RX ruling out craniosynostosis.The device performs according to these indications.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in this document for the Boston Band. The effectiveness claim relies on a clinical trial referenced for cranial orthoses in general, not specifically for the Boston Band.
    • Data Provenance: The document references a clinical trial: "Ottoto: Treatment Protocols for Plagiocephaly, Jeff Larson, CO:JPO 2004, Vol 16, Num 4s." The provenance of that study's data (country of origin, retrospective/prospective) is not provided in this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided because the document references an existing clinical trial for cranial orthoses to establish effectiveness. There is no "test set" in the context of validating a new algorithm, or a process described for establishing ground truth from experts for this specific 510(k) submission.

    4. Adjudication Method for the Test Set

    Not applicable. The document relies on a referenced clinical trial's methodology, not a separate adjudication process described here for a specific test set.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (cranial orthosis), not an AI or diagnostic imaging device that involves "human readers" or "AI assistance."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device.

    7. The Type of Ground Truth Used

    The "ground truth" for the effectiveness claim is based on the outcomes reported in the referenced clinical trial ("Ottoto: Treatment Protocols for Plagiocephaly, Jeff Larson, CO:JPO 2004, Vol 16, Num 4s"). This would typically involve clinical outcome data (e.g., changes in cranial measurements, resolution of deformities) rather than expert consensus, pathology, or other types of ground truth used for diagnostic devices.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device. There is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.