Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K150932

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-22

    (473 days)

    Product Code
    Regulation Number
    870.1310
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bioteq Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures.

    Device Description

    Bioteq Sheath Introducer consists of a sheath introducer with side port extension and stopcock. The sheath hub includes a hemostasis valve and a suture ring. A 4-way stopcock is affixed to the proximal end of the side port extension. The Bioteq Sheath Introducer assembly includes a vessel dilator that rotates securely with the sheath introducer hub and guidewire which is cleared by the Food and Drug Administration (K920884).

    The sheath introducer is available in 5 French (F) to 7F sizes in length of 11cm or 23cm. The vessel dilator is tipped specifically to accept either a 0.038 inch or a 0.035 inch diameter guidewire with the length of 45cm or 80cm.

    AI/ML Overview

    The provided document describes the 510(k) premarket notification for a Sheath Introducer device, demonstrating its substantial equivalence to a legally marketed predicate device. This is not a study proving the device meets acceptance criteria for an AI/ML-based medical device.

    The document discusses:

    • The device: Sheath Introducer, used for percutaneous introduction of devices into veins/arteries.
    • The testing conducted: Sterilization validation, biocompatibility, and various performance tests (e.g., force at break, leakage, radiopacity, insertion force, guidewire compatibility, kink resistance).
    • The comparison: The proposed device is compared to the Merit Prelude™ Sheath Introducer (K050962).
    • The determination: The FDA found the device substantially equivalent based on intended use, design, technology/principles of operation, and performance.

    Therefore, the request for "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI/ML device testing (e.g., sample size for test/training sets, expert ground truth, MRMC studies, effect size, standalone performance) cannot be answered from this document.

    This document specifically states: "No clinical test data was used to support the decision of the substantial equivalence." and focuses on bench testing and material comparisons.

    However, if we interpret "acceptance criteria" and "study" in the context of this specific medical device (a physical sheath introducer) as a substantial equivalence determination, then the information would be as follows:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Bench Test Type)Reported Device Performance (Summary)
    Sterilization Validation (Bioburden, Endotoxin, Sterility, EO residual, Pyrogen, Batch release)Met related requirements.
    Packaging ValidationMet related requirements.
    Biocompatibility (Cytotoxicity, Intracutaneous Reactivity, Systemic Injection, Sensitization, Salmonella typhimurium Reverse Mutation, Hemolysis, Thromboresistance, Complement Activation, Material-Mediated Pyrogenicity)Met related requirements.
    Performance (Force at Break, Conical Fittings, Liquid Leakage, Chemical Characterization, Radiopacity, Insertion Force, Guidewire Compatibility, External Surface Inspection)Met related requirements and its intended use.
    Comparison Performance (Force at Break, Size Designation, Liquid Leakage, Insertion Force, Kink Resistance and Recovery, Minimum Tensile Force to Remove Dilator, Stiffness)Demonstrated substantial equivalence to predicate device; differences did not raise safety/effectiveness issues.

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not explicitly stated for each test, but standard test methods for medical devices typically involve specific sample sizes per ISO/ASTM standards.
    • Data provenance: Not specified beyond "A series of tests were performed." The tests are non-clinical, meaning they were conducted in a laboratory setting. No indication of country of origin for the data itself, but the manufacturer is in Taipei, Taiwan. The tests are "retrospective" in the sense that they are conducted on manufactured samples, not "prospective" clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a physical medical device undergoing bench testing, not an AI/ML algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for these tests comes from established engineering and material science standards and measurement techniques.

    4. Adjudication method for the test set:

    • Not applicable. Performance is measured against pre-defined engineering and material standards, not subjective expert judgment.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a physical, non-AI medical device. No human-in-the-loop performance or AI assistance involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical, non-AI medical device.

    7. The type of ground truth used:

    • For the non-clinical tests, the "ground truth" is derived from established engineering and material science standards (e.g., ISO, ASTM standards for medical devices), physical measurements, and chemical analyses. For instance, a "force at break" test's ground truth is the measured force value compared to an acceptable range.

    8. The sample size for the training set:

    • Not applicable. This is a physical, non-AI medical device. There is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no training set or machine learning "ground truth" for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K102633

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-09-09

    (361 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "BIOTEQ" Angiographic Catheter is indicated for the delivery of radiopaque contrast medium to selected sites in the vascular system. The device is for use in the coronary and peripheral vasculature, and not for use in the neurovasculature.

    Device Description

    The BIOTEO® ANGIOGRAPHIC CATHETER is with the braided proximal shaft. The distal shaft of the catheter has variety configuration and the tip is made soft to minimize trauma to the vessel wall. It consists of the following major components: Soft Tube, Braiding tube, Screw Cap, F.L.L. Adapter, Adaptor Butterfly.

    AI/ML Overview

    The provided document pertains to the 510(k) premarket notification for the BIOTEQ® ANGIOGRAPHIC CATHETER. This is a medical device and the document describes performance testing rather than a study typically associated with AI/ML devices that would involve acceptance criteria, sample sizes for test/training sets, ground truth establishment, or expert reviews for diagnostic accuracy.

    Therefore, the requested information regarding acceptance criteria and studies typical for AI/ML devices (e.g., sample sizes for test/training sets, ground truth, expert qualifications, adjudication methods, MRMC studies, standalone performance) cannot be extracted from this document, as it focuses on the physical and biological performance of a catheter.

    Here's a breakdown of the performance testing conducted for this device, which serves a similar purpose to acceptance criteria for traditional medical devices:

    1. Table of Acceptance Criteria and Reported Device Performance (based on performance testing, not AI/ML metrics):

    Performance Test CategorySpecific TestAcceptance Criteria (Implied by "according to protocols based on international standards and in-house requirements")Reported Device Performance
    Physical PerformanceSurface RequirementsMeets established standards for surface finish and integrity.Achieved
    Corrosion ResistanceNo significant corrosion observed.Achieved
    Strength TestWithstands specified forces without failure (ee.g., tensile, burst).Achieved
    Leakage Tests (positive pressure, vacuum)No leakage under specified pressure and vacuum conditions.Achieved
    Flow Rate TestAchieves specified flow rates for contrast media delivery.Achieved
    Torque Strength TestWithstands specified torque without failure.Achieved
    Flexibility and Kink TestMaintains integrity and patency under bending and kinking.Achieved
    Dimensional VerificationMeets specified dimensional tolerances.Achieved
    Particle TestWithin limits for particle shedding.Achieved
    Catheter Twist Transmitting TestTransmits torque effectively to the distal tip.Achieved
    Catheter Shape RetentionRetains intended shape after use/manipulation.Achieved
    Freedom from leakage and damage under high static pressure conditionsNo leakage or damage under specified high static pressure.Achieved
    Aged 3-year Shelf Life TestingMaintains all performance characteristics after simulated aging (3 years).Achieved
    Packaging TestsN/AMaintains sterility and device integrity.Achieved
    Biocompatibility TestsIntracutaneous Reactivity StudyNon-reactive or low reactivity within established limits.Achieved
    Guinea Pig Maximization Sensitization StudyNon-sensitizing.Achieved
    Cytotoxicity TestingNon-cytotoxic.Achieved
    Intramuscular Implant Study (1 week)Biocompatible; no adverse tissue reaction.Achieved
    Acute Systemic Injection StudyNo acute systemic toxicity.Achieved
    Hemolysis (direct and indirect contact)Non-hemolytic.Achieved
    Thrombogenicity Study ISONon-thrombogenic (within limits).Achieved
    Salmonella Typhimurium Reverse Mutation Assay (Ames Test)Non-mutagenic.Achieved
    Complement Activation TestMinimal or no complement activation.Achieved
    Material Mediated PyrogenicityNon-pyrogenic.Achieved
    Sterilization ValidationETO Sterilization Validation StudyAchieves sterility assurance level (SAL).Achieved

    The study supporting these criteria:

    The document states: "Performance testing was conducted on the BIOTEO® ANGIOGRAPHIC Catheter to establish substantial equivalence. Testing was conducted according to protocols based on international standards and in-house requirements and included dimensional and functional testing. Additionally, the catheters were subjected to biocompatibility testing per ISO 10993."

    Missing Information (as per AI/ML context):

    The following points are not applicable or not provided in the context of this traditional medical device submission for an angiographic catheter:

    1. Sample size used for the test set and the data provenance: Not applicable. Performance testing for physical and biological characteristics does not involve "test sets" in the AI/ML sense.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device is established through engineering and biological testing standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI/ML diagnostic tool for human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance is defined by established engineering standards, material science properties, biocompatibility guidelines (e.g., ISO 10993), and functional performance specifications.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K033862

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-02-14

    (430 days)

    Product Code
    Regulation Number
    876.5010
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOTEQ® Pigtail Drainage Catheter Set( One Step Type) is intended to be used for percutaneous drainage of abscesses, cysts, gall bladders and nephrostomies.

    Device Description

    The BIOTEQ® Pigtail Drainage Catheter Set ( One Step Type) is used for the drainage pathway of patients' fluids through the catheter out of body . It consist of the following major components: ① F.L.L. Adapter ② Screw Cap ③ Catheter ④ Curve Straightener (Sleeve) ⑤ Wire Cap --- (For With Safety String Lock model only) ⑥ Trocar Stylet ⑦ Trocar Needle ⑧ Sheath ⑨ Suture Wire --- (For With Safety String Lock model only)

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the BIOTEQ® Pigtail Drainage Catheter Set, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document describes performance in general terms rather than specific numerical acceptance criteria.

    Acceptance Criteria CategoryReported Device Performance
    Physical SpecificationConforms to applicable standards.
    Chemical SpecificationConforms to applicable standards.
    Biological SpecificationConforms to applicable standards (including ISO 10993 series, USP Pyrogenic standards).
    Sterilization SpecificationConforms to applicable standards (including ISO 11135, ISO 11607-1).
    Technological EquivalenceDifferences in technological characteristics do not raise new questions of safety or effectiveness compared to the predicate device.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated. The document indicates "bench testing" was performed, but the number of units tested is not specified.
    • Data Provenance: Not explicitly stated, but given the manufacturer (BIOTEQUE CORPORATION) is in Taiwan, it's likely the testing was conducted by or for the manufacturer. The testing is described as "bench testing," which refers to in-vitro laboratory tests, not human data. Therefore, the data would not have a country of origin in the context of patient data, nor is it retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not applicable. The "ground truth" for this type of device (a medical catheter set) is established through performance against industry standards and specifications during bench testing, not through expert interpretation of a test set.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method for the test set:

    • Adjudication Method: Not applicable. As the testing involves objective measurement against engineering and biological standards, there is no need for expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This is a medical device (catheter set), not an AI-powered diagnostic or interpretive tool. Therefore, a multi-reader multi-case comparative effectiveness study or AI assistance is not relevant or applicable.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    • Standalone Performance Study: Not applicable. This device does not involve an algorithm. The study described is bench testing of the physical, chemical, biological, and sterilization properties of the catheter set.

    7. The type of ground truth used:

    • Type of Ground Truth: The ground truth for this device's performance is derived from established international and national standards for medical devices, specifically for physical, chemical, biological compatibility, and sterility. Examples include ISO 10993 series, ISO 11607-1, ISO 11135, and USP Pyrogenic standards.

    8. The sample size for the training set:

    • Sample Size for Training Set: Not applicable. This device does not involve machine learning or AI, so there is no concept of a "training set."

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set Establishment: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K033210

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-12-30

    (454 days)

    Product Code
    Regulation Number
    876.4620
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOTEQ® DOUBLE PIGTAIL URETERAL STENT SET is intended to facilitate the temporary internal drainage of urine from the kidney to the bladder via placement endoscopically by a trained physician.

    Device Description

    The BIOTEQ® DOUBLE PIGTAIL URETERAL STENT SET consist of a ureter stent that is completely radiopaque, coiled cylindrical or open tip with drainage holes at intervals of 5-20 mm, centimeter graduations, continuous black positioning line to indicate the direction in which the catheter tip coils, and black marking at the catheter end. The Pusher Catheter (Introducer sheath) is approximately 45cm long. The Guide Wire (Spiral stylet) is made of stainless steel with PTFE coating and a flexible safety tip and a marked rigid tip and is approximately 120 cm long. The plastic dispenser is used to protect the Guide Wire (Spiral stylet). There are also plastic clips in the set.

    AI/ML Overview

    The provided text describes a 510(k) summary for the BIOTEQ® DOUBLE PIGTAIL URETERAL STENT SET. For medical devices, particularly those submitted under 510(k), the "acceptance criteria" and "study" typically refer to demonstrating substantial equivalence to a legally marketed predicate device through bench testing, performance testing, and adherence to relevant standards, rather than clinical studies with human participants that establish a device's performance in terms of diagnostic accuracy or effect sizes.

    Based on the provided text, here's an analysis of the requested information:

    1. Table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategorySpecific Criteria/Standard Adhered ToReported Device Performance / Compliance
    Physical SpecificationApplicable standards (unspecified detail)"conforms to applicable standards"
    Chemical SpecificationApplicable standards (unspecified detail)"conforms to applicable standards"
    Sterilization SpecificationISO 11135 (Sterilization of health care products - Ethylene oxide), ISO 11607-1 (Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems), USP Pyrogenic standards, & related standards"conforms to applicable standards"
    BiocompatibilityISO 10993 series (Biological evaluation of medical devices)"conforms to applicable standards"
    Substantial EquivalenceSame intended use and similar technological characteristics as predicate device (Rusch Ureter Stent Integral Set and Rusch Ureter DD Stent Sets - K982974). Any differences in technological characteristics do not raise new questions of safety or effectiveness."The BIOTEQ® DOUBLE PIGTAIL URETERAL STENT SET have the same intended use and similar technological characteristics as the Rusch Ureter Stent Integral Set and Rusch Ureter DD Stent Sets (K982974) marketed by RUSCH INTL.. Moreover, bench testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify a "sample size" in the context of a traditional test set with individual patient cases or data points. The assessment is based on bench testing and adherence to standards for the device itself and its material properties.
    • Data Provenance: Not applicable in the context of clinical data. The "data" refers to the results of the bench tests and compliance with international standards for manufacturing, materials, and sterilization.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable to this 510(k) submission. The "ground truth" for a medical device cleared via 510(k) is typically established by demonstrating that the device meets engineering specifications, material standards, biocompatibility requirements, and sterilization efficacy, which are evaluated by technical experts, engineers, and regulatory bodies against established standards, not clinical experts establishing diagnostic "ground truth" for a dataset.

    4. Adjudication method for the test set:

    • Not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies, particularly for diagnostic devices, to resolve discrepant expert opinions on case classifications. This 510(k) relies on bench testing and adherence to standards, which do not involve such adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was NOT done. This type of study is relevant for diagnostic imaging devices, especially those incorporating AI, where the performance of human readers with and without AI assistance is evaluated across multiple cases. The BIOTEQ® DOUBLE PIGTAIL URETERAL STENT SET is an implantable medical device, not a diagnostic imaging or AI-enabled device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was NOT done. This device is a physical medical instrument (ureteral stent) and does not involve an algorithm.

    7. The type of ground truth used:

    • The "ground truth" in this context is defined by engineering specifications, material safety and performance standards (e.g., ISO, USP), and biocompatibility requirements. It's about meeting pre-defined physical, chemical, and biological performance benchmarks, not clinical outcomes or pathology.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this device does not utilize machine learning or AI algorithms. The device undergoes manufacturing processes and quality controls to ensure each unit meets the required specifications.

    9. How the ground truth for the training set was established:

    • Not applicable. As there is no training set involving algorithms, there is no ground truth to be established for it. The "ground truth" for the device's design and manufacturing is established through adherence to recognized international and national standards and regulatory requirements.
    Ask a Question

    Ask a specific question about this device

    K Number
    K032292

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-03-22

    (242 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOTEQ® Arterial-Venous Fistula (A.V.F) Needle Set with Safety Flex is used for temporary cannulation for vascular access for extracorporeal blood treatment. The device is intended for single use only and is for temporary catheterization of less than 30 days. The Safety Feature(Safety Flex) aids in prevention of needle-sticks injuries when removing and discarding needles after dialysis session.

    Device Description

    The BIOTEQ® A. V. FISTULA NEDDLE SET WITH SAFETY FLEX consists of the following major components: ① Protection Cap ② Needle Cannula ③ Butterfly Wing ④ Needle Hub ⑤ Safety Flex ⑥ Small Pinch Clamp ⑦ PVC Tubing ⑧ Female Luer ⑨ Cap for Female Luer These components assembled together as BIOTEQ® Arterial-Venous Fistula (A.V.F) Needle Set with Safety Flex for use during hemodialysis procedures. The Safety Feature(Safety Flex) aids in prevention of needle-sticks injuries when removing and discarding needles after dialysis session. Various models of needle size manufactured such as 14 gauge, 15 gauge, 16 gauge, 17 gauge.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called the BIOTEQ® A. V. FISTULA NEEDLE SET WITH SAFETY FLEX. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than conducting a full clinical study with specific acceptance criteria related to device performance in a clinical setting.

    Therefore, the document does not contain detailed information about acceptance criteria or a study proving the device meets them in the way clinical performance studies typically do. Instead, it focuses on demonstrating that the device meets safety and effectiveness requirements by conforming to applicable standards and showing equivalence to a legally marketed predicate device.

    Here's a breakdown of the information that is and is not available based on your request:


    1. A table of acceptance criteria and the reported device performance

    • Not Applicable in the traditional sense. This 510(k) summary does not present specific clinical acceptance criteria (e.g., sensitivity, specificity, accuracy for an AI device) or reported performance metrics against such criteria.
    • What is reported: The document states, "In terms of Physical specification, Chemical specification, Biological specification & Sterilization Specification, the device conforms to applicable standards included ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards". This is the closest equivalent to "acceptance criteria" here, indicating that the device meets established industry standards for these categories.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Provided. This information is not typically included in a 510(k) summary focused on substantial equivalence through bench testing. The "test set" here refers to "bench testing" rather than a clinical dataset.
    • The document mentions "bench testing contained in this submission," but does not specify sample sizes or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. For a device like an A.V. Fistula Needle Set, "ground truth" is not established by human experts in the same way it would be for an AI diagnostic device evaluating images. Device performance is assessed against engineering and biological standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are relevant for studies where human interpretation or consensus is needed to establish ground truth, which is not the case for the type of testing described (physical, chemical, biological, sterilization specifications).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices, not for a physical medical device like a needle set.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a physical medical instrument, not an algorithm, so "standalone performance" in the context of AI is not relevant. The device itself is "standalone" in its function.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Ground truth for this device type is based on established engineering, biological, and sterilization standards. For example, "Physical specification" would refer to tensile strength, needle sharpness, flow rates; "Chemical specification" to material composition and biocompatibility; "Biological specification" to absence of certain toxins, and "Sterilization Specification" to sterility assurance levels.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" for this type of device. The document describes a manufactured product, not an AI/ML algorithm that requires training.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for this device type, the establishment of ground truth for it is irrelevant.

    Summary of the Study (as described in the 510(k)):

    The study described is bench testing to demonstrate substantial equivalence to a predicate device (JMS A.V. FISTULA NEEDLE WingEater, K 010406).

    • Purpose: To show that any differences in technological characteristics between the BIOTEQ® A. V. FISTULA NEDDLE SET WITH SAFETY FLEX and the predicate device "do not raise any new questions of safety or effectiveness."
    • Methods: The device was tested "in terms of Physical specification, Chemical specification, Biological specification & Sterilization Specification." It "conforms to applicable standards included ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards."
    • Conclusion: Based on this bench testing and comparison of intended use and technological characteristics, the device was deemed "substantially equivalent to the predicate devices."

    In essence, this 510(k) summary provides evidence of compliance with established standards and equivalence to a predicate, rather than reporting on a clinical trial with specific performance metrics against defined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013634

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-07-09

    (246 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs ( BT-102DA · BT-102DB · BT-102DC · BT-102DD) are sets of single-use disposable components intended to provide extracorporeal access to blood of patients suffering from end stage renal disease during Hemodialysis therapy. Each set will include an arterial/venous bloodline set, and any combination of two or three component of the following items: An intravenous administration set for (use in administration of intravenous fluids to a dialysis set in conjunction with hemodialysis procedures.) , AV fistula needle set to (apply on the access site of patient vessel to draw the blood flow adequate to pass through the dialyer and reinfusion of dialysed blood back to patient vessel via the A.V. Fistula Needle Set during hemodialysis procedures.) , and/or A transducer protector for (use as protective devices for pressure monitors and to help protect the sterility of the fluid pathway).

    Device Description

    The BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs (BT-102DA 、BT-102DB、BT-102DC、BT-102DD) are sets of single-use disposable components intended to provide extracorporeal access to blood of patients suffering from end stage renal disease during Hemodialysis therapy. Each set will include an arterial/venous bloodline set, and any combination of two or three component of the following items: An intravenous administration set for use in administration of intravenous fluids to a dialysis set in conjunction with hemodialysis procedures. , AV fistula needle set to apply on the access site of patient vessel to draw the blood flow adequate to pass through the dialyer and reinfusion of dialysed blood back to patient vessel via the A.V. Fistula Needle Set during hemodialysis procedures. , and/or A transducer protector for use as protective devices for pressure monitors and to help protect the sterility of the fluid pathway. BIOTEQ® 3 in 1 Hemodialysis Blood Tubing Packs includes 4 models (BT-102DA 、BT-102DB、BT-102DC、BT-102DD)

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device and, as such, focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and reported device performance in the way requested for AI/diagnostic devices. The information provided heavily emphasizes physical, chemical, biological, and sterilization specifications rather than clinical performance metrics often associated with "device performance" in the context of diagnostic or AI-driven tools.

    Therefore, the requested table of acceptance criteria and reported device performance, and several other points (2, 3, 4, 5, 6, 7, 8, 9), cannot be directly extracted or are not applicable from the provided text in the way they would be for a typical AI/diagnostic device study. The text primarily focuses on conformity to standards and substantial equivalence.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Applicable Standards/Specifications)Reported Device Performance (as stated in the document)
    Physical SpecificationConforms to applicable standards
    Chemical SpecificationConforms to applicable standards
    Biological SpecificationConforms to applicable standards (including ISO 10993 series, USP Pyrogenic standards)
    Sterilization SpecificationConforms to applicable standards (including ISO 11607-1, ISO 11135)
    Substantial Equivalence to Predicate DeviceBench testing demonstrates that differences in technological characteristics do not raise new questions of safety or effectiveness, making it substantially equivalent to Nipro Blood Tubing Set for Hemodialysis with Transducer Protectors and Priming set (K972206) & (K001465).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
    The document mentions "bench testing," but it does not specify the sample size used for these tests. Data provenance (country of origin, retrospective/prospective) is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This device is a blood tubing pack, and its evaluation relies on engineering, material, and sterility testing against established standards, not expert interpretation of outputs like an AI diagnostic tool.

    4. Adjudication method for the test set
    Not applicable. The evaluation is based on conformance to engineering and safety standards, not a diagnostic interpretation process requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI or diagnostic device that involves human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
    Not applicable, as this is not an AI algorithm. The performance evaluation focuses on the physical and biological characteristics of the blood tubing pack itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    The "ground truth" for this device is based on established international and national standards for medical device safety, material compatibility, and sterilization (e.g., ISO 10993 series for biological evaluation, ISO 11607-1 for packaging, ISO 11135 for sterilization, USP Pyrogenic standards).

    8. The sample size for the training set
    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K012526

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-09-14

    (39 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOTEQ® SCALP VEIN SET act as a pathway to administer parenteral fluid/ medication into the patient's vascular system.

    Device Description

    BIOTEQ® SCALP VEIN SET (BS-120 (22G); BS-121(23G) ; BS-122(24G))consist of the following 5 major components: Protective cap for needle, the stainless needle, the plastic butterfly wing , the PVC Tube, the PVC safety Cap. These 5 major components assembled together as SCALE VEIN SET for use as a pathway to administer medical fluid from extracorporeal system through the device into venous blood.

    AI/ML Overview

    The provided text is a 510(k) summary for the BIOTEQ® SCALP VEIN SET. It describes the device, its intended use, and its substantial equivalence to predicate devices. However, it does not include detailed information regarding a study that proves the device meets specific acceptance criteria in the way a clinical trial or performance study for a diagnostic AI device would.

    Instead, the submission relies on demonstrating substantial equivalence to already approved predicate devices through bench testing that confirms physical, chemical, biological, and sterilization specifications.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable or cannot be extracted from this document, as this is not the type of study described.

    Here's a breakdown of the information that is available based on your request:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance (as stated or implied)
    Physical SpecificationConforms to applicable standards (not explicitly detailed)
    Chemical SpecificationConforms to applicable standards (not explicitly detailed)
    Biological SpecificationConforms to applicable standards, including ISO 10993 series & USP Pyrogenic standards
    Sterilization SpecificationConforms to applicable standards, including ISO 11607-1, ISO 11135
    Intended UseServes as a pathway to administer parenteral fluid/medication into the patient's vascular system.
    Technological CharacteristicsSimilar to predicate devices; bench testing shows no new safety/effectiveness questions.

    Study Details (Based on available information)

    1. Sample size used for the test set and the data provenance: Not explicitly stated as a "test set" in the context of an AI/diagnostic device. The "bench testing" likely refers to laboratory-based evaluations of the device's components and assembled product against engineering and safety standards. The provenance (e.g., country of origin, retrospective/prospective) for this bench testing is not specified.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For a device like a scalp vein set, "ground truth" typically relates to objective measurements against established engineering and safety standards, rather than expert interpretation of data.

    3. Adjudication method for the test set: Not applicable. No human interpretation or adjudication process is described for this type of bench testing.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document pertains to a medical device (scalp vein set), not an AI diagnostic tool involving human readers.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance is based on established engineering standards, material specifications (e.g., ISO, USP), and functional performance requirements for medical devices of this type. It's objective, measurable criteria rather than subjective expert consensus or clinical outcomes data in the typical sense.

    7. The sample size for the training set: Not applicable. This is not a machine learning device, so there is no training set.

    8. How the ground truth for the training set was established: Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K002449

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-12-22

    (135 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K992930

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-02-09

    (162 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOTEQUE Hemodialysis Blood Tubing Set(Models BT-102A, BT-102B and BT-190) is intended to serve as a conduit through which blood is taken from the patient, delivered to the dialyzer, and returned to the patient during hemodialysis therapy. The device is intended to be used for patients suffering from renal disease only.

    Device Description

    Bioteque Hemodialysis Blood Tubing Set, consists of Arterial Line & Venous Line, is used as the conduit through which blood is taken from the patient, delivered to the dialyzer, and returned to the patient during dialysis therapy.

    AI/ML Overview

    The provided text for the BIOTEQUE HEMODIALYSIS BLOOD TUBING SET (K992930) discusses performance in terms of adherence to various standards rather than specific quantitative acceptance criteria and device performance values. It is a submission for a medical device that outlines its equivalence to a predicate device based on similar intended use and technological characteristics. The document does not describe a study comparing the device's performance against specific, pre-defined quantitative acceptance criteria in a clinical setting in the way an AI/ML device would be.

    Therefore, many of the requested categories for AI/ML device studies (such as sample size, data provenance, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable to this 510(k) summary for a physical medical device.

    However, I can extract the information that is present and indicate where the requested information is not provided or applicable.


    Acceptance Criteria and Device Performance for BIOTEQUE HEMODIALYSIS BLOOD TUBING SET (K992930)

    This 510(k) summary for the BIOTEQUE HEMODIALYSIS BLOOD TUBING SET does not present quantitative acceptance criteria or corresponding device performance results in the format typically seen for AI/ML diagnostic devices. Instead, the performance is demonstrated through conformance to established national and international standards for physical medical devices. The core argument for acceptance is substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Category (Standards Complied With)Reported Device Performance (Conformance)
    Physical SpecificationConforms to applicable standards
    Chemical SpecificationConforms to applicable standards
    Biological SpecificationConforms to applicable standards (e.g., ISO 10993 series, USP Pyrogenic standards)
    Sterilization SpecificationConforms to applicable standards (e.g., ISO 11607-1, ISO 11135)
    Intended UseSame as predicate device
    Technological CharacteristicsSimilar to predicate device

    Study Proving Device Meets Acceptance Criteria:

    The study described is primarily a bench testing and comparison to predicate device approach, rather than a clinical trial with quantitative endpoints in the context of AI/ML performance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document refers to "bench testing," which implies laboratory-based evaluations of the physical device components. Specific sample sizes for these tests are not provided in this summary.
    • Data Provenance: Not applicable in the context of patient data for an AI/ML device. The testing would have been conducted in a laboratory setting by the manufacturer, BIOTEQUE CORPORATION, likely in Taiwan (R.O.C.), given the submitter's address.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not Applicable: This type of information is relevant for studies involving human interpretation (e.g., radiologists reviewing images to establish ground truth). For a physical medical device like blood tubing, "ground truth" is typically established through validated laboratory tests against defined physical, chemical, and biological specifications, not through expert consensus on observational data.

    4. Adjudication Method for the Test Set

    • Not Applicable: Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human reviewers (experts) when establishing ground truth from observational data. This is not relevant for bench testing of a physical product.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is a physical medical device (blood tubing), not an AI/ML diagnostic algorithm. Therefore, an MRMC study and the concept of human reader improvement with/without AI assistance do not apply.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Standards Conformance: The "ground truth" for this device is based on conformance to established national and international standards for medical device physical, chemical, biological, and sterilization properties (e.g., ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards). The absence of safety and effectiveness concerns compared to a predicate device (K853851) also serves as a benchmark for acceptability.

    8. The Sample Size for the Training Set

    • Not Applicable: This is a physical medical device. There is no "training set" in the context of an AI/ML algorithm.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable: As there is no training set, the establishment of ground truth for it is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K993118

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-12-17

    (88 days)

    Product Code
    Regulation Number
    876.5540
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bioteque A.V. Fistula Needle is used during hemodialysis. It's a part of accessory of extracorporeal system for treatment of renal failure. A.V. Fistula Needle applied on the access site of patient's vessel to obtain blood flow adequate to pass through the dialyses, and the reinfusion of dialysed blood back to patient via the fistula needle during hemodialysis.

    Device Description

    Bioteque A.V. Fistula Needle Set consist of the following 6 major components: the stainless needle, protector cap for needle, the plastic butterfly wing, the PVC tubing, the female luer, the cap for female luer. These 6 major components assembled together as A.V. Fistula Needle Set for use during hemodialysis procedures. Various models of needle size manufactured such as 14 gauge, 15 gauge, 16 gauge.

    AI/ML Overview

    The provided document is a 510(k) summary for the BIOTEQUE A.V. FISTULA NEEDLE SET. This submission primarily focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a performance study with detailed acceptance criteria and testing results as would be typical for more complex devices or novel technologies.

    Here's an analysis based on the provided text, addressing the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly list quantitative acceptance criteria or specific performance metrics in a table format with numerical results. Instead, it states that the device "conforms to applicable standards." The acceptance criteria are implicitly tied to compliance with these standards and demonstrating similar technological characteristics to the predicate device.

    Acceptance Criteria (Implicit)Reported Device Performance (Summary)
    Physical specificationsConforms to applicable standards (including ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards). Acknowledged to have the same intended use and similar technological characteristics as the predicate device (TERUMO AV FISTULA NEEDLE SET K891062).
    Chemical specificationsConforms to applicable standards (including ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards). Acknowledged to have the same intended use and similar technological characteristics as the predicate device (TERUMO AV FISTULA NEEDLE SET K891062).
    Biological specificationsConforms to applicable standards (including ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards). Specifically mentions USP Pyrogenic standards. Acknowledged to have the same intended use and similar technological characteristics as the predicate device (TERUMO AV FISTULA NEEDLE SET K891062).
    Performance specificationsConforms to applicable standards (including ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards). Bench testing demonstrated that differences in technological characteristics do not raise new questions of safety or effectiveness. Acknowledged to have the same intended use and similar technological characteristics as the predicate device (TERUMO AV FISTULA NEEDLE SET K891062).
    Sterilization specificationsConforms to applicable standards (including ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards). Acknowledged to have the same intended use and similar technological characteristics as the predicate device (TERUMO AV FISTULA NEEDLE SET K891062).
    Substantial EquivalenceConcluded to be substantially equivalent to the TERUMO AV FISTULA NEEDLE SET (K891062).

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "bench testing contained in this submission" but does not specify the sample size for any test sets used. The nature of the device (a medical needle for hemodialysis) suggests bench testing primarily, not clinical trials with patient data. Therefore, questions of retrospective/prospective data or country of origin for a clinical test set are not applicable here. The testing would have been conducted in a laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable and not provided for this type of device and submission. The "ground truth" for mechanical and biological safety of an A.V. fistula needle is established by adherence to recognized international and national standards (e.g., ISO, USP) and engineering principles, not through expert consensus on diagnostic images or clinical outcomes in the same way as, for example, an AI diagnostic tool.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving subjective human interpretation (e.g., image reading) to establish a consensus ground truth. The tests mentioned (physical, chemical, biological, sterilization, performance) are objective laboratory/bench tests, not requiring human adjudication in this context.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic or AI-assisted devices where human readers interpret medical images or data. The BIOTEQUE A.V. FISTULA NEEDLE SET is a physical medical device (a needle), and its effectiveness is determined by its functional performance and safety according to engineering and biological standards, not by human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    This question is not applicable as the device is a physical medical needle, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance and safety is established through adherence to recognized industry standards and laboratory testing protocols. These standards (e.g., ISO 10993 series for biocompatibility, ISO 11607-1 for packaging, ISO 11135 for sterilization, USP Pyrogenic standards) define the acceptable limits and methods for evaluating the physical, chemical, biological, and performance characteristics of such devices. Therefore, the ground truth is based on objective measurements against established standards, not expert consensus, pathology, or outcomes data in the sense typically understood for diagnostic or interventional devices.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. The device is a physical medical device, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1