Found 8 results

510(k) Data Aggregation

    K Number
    K250971

    Validate with FDA (Live)

    Date Cleared
    2025-05-16

    (46 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    0 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Embosphere Microspheres are indicated for use in the embolization of:

    • Hypervascular tumors, including symptomatic uterine fibroids
    • Prostatic arteries for symptomatic Benign Prostatic Hyperplasia (BPH)
    • Arteriovenous malformations
    • Blood vessels to occlude blood flow to control bleeding/hemorrhaging in the peripheral vasculature
    Device Description

    Embosphere Microspheres are small, compressible, hydrophilic, biocompatible spheres made of acrylic polymer and porcine-derived gelatin. The microspheres are packaged in 0.9% saline and are provided sterile and non-pyrogenic in a vial or in a syringe.

    The product is provided in seven size ranges to allow physicians to choose the appropriate size necessary for the vessel being embolized. The size ranges available are:
    • 50-100 microns
    • 40-120 microns
    • 100-300 microns
    • 300-500 microns
    • 500-700 microns
    • 700-900 microns
    • 900-1200 microns

    The principles of operation for the subject device Embosphere Microspheres are the same as the predicate device Embosphere Microspheres (K181300). Embosphere Microspheres are permanent implantable devices and are designed for controlled, targeted embolization. All indications for Embosphere Microspheres; uterine arteries, arteriovenous malformations, hypervascular tumors and prostate arteries all involve arterial embolization. The procedure of arterial embolization is similar for all arteries. Appropriately sized microspheres for target vessel occlusion are chosen by the trained interventional radiologist. The delivery procedure involves arterial access through an artery, using a guidewire and microcatheter under fluoroscopic guidance. Once the catheter tip is placed in the artery(ies) supplying the targeted tissue, Embosphere Microspheres mixed with a non-ionic contrast agent are delivered in a controlled manner under visualization to occlude the feeding vessel(s) to interrupt artery blood flow to the targeted area. The device is intended for single use.

    AI/ML Overview

    I am sorry, but the provided text is an FDA 510(k) Clearance Letter for a medical device (Embosphere Microspheres). It details the regulatory clearance process, the device's intended use, and its equivalence to a predicate device.

    This document does NOT contain information about any AI/ML model, its acceptance criteria, or a study proving that an AI/ML device meets those criteria.

    Therefore, I cannot extract the information required to answer your request regarding the acceptance criteria and the study proving the device meets them, as it pertains to an AI model. The provided text describes a physical medical device and its regulatory review, not a software or AI-driven diagnostic tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192480

    Validate with FDA (Live)

    Date Cleared
    2019-11-21

    (72 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Torpedo Gelatin Foam is indicated for use in embolization of:
    · Hypervascular tumors
    · Blood vessels to occlude blood flow to control bleeding / hemorrhaging in the peripheral vasculature

    Device Description

    Torpedo Gelatin Foam is a hydrophilic medical device which consists of resorbable dry gelatin foam that is compressed into a cylindrical shape and preloaded into a cartridge with a standard female and male luer fittings. The device is available in two sizes (0.9 mm and 1.7 mm) and two lengths configurations (10 mm and 20 mm).
    Once rehydrated, the deformable torpedoes can be injected into the target vessel with an intravascular catheter or a micro-catheter (depending on the size range) to provide a mechanical barrier to blood flow. Contrast enhancement may be used to monitor the embolization procedure using fluoroscopy. The device is intended for single use and is provided sterile.

    AI/ML Overview

    This document is a 510(k) summary for the Torpedo Gelatin Foam device. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain the information requested in your prompt regarding acceptance criteria and the detailed study results that prove the device meets these criteria in the context of an AI-driven medical device.

    The provided text does not describe an AI medical device or its performance criteria. It is a traditional medical device 510(k) submission for an embolization device (gelatin foam).

    Therefore, I cannot provide the requested information about acceptance criteria and study details for an AI medical device based on the given text. The document refers to:

    • Bench testing: Leveraged from a previously cleared device (K183578) with no changes in material, design, or processing.
    • Animal testing: Leveraged from the predicate device (K183120) to support new indications for use. An animal study (7 test, 8 control sheep) evaluated vascular occlusion, local tissue effects, and in-vivo degradation over 4 weeks using angiography and histopathology. The outcome was successful embolization and no significant difference in artery patency between devices.

    None of this relates to AI performance metrics, sample sizes for AI model testing, expert labeling, MRMC studies, or training set details that would be pertinent to an AI medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183120

    Validate with FDA (Live)

    Date Cleared
    2019-08-06

    (270 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EmboCube Embolization Gelatin is indicated for use in embolization of:

    • Hypervascular tumors
    • Blood vessels to occlude blood flow to control bleeding / hemorrhaging in the peripheral vasculature
    Device Description

    EmboCube Embolization Gelatin is a hydrophilic, medical device which consists of pre-cut, dry cubes of resorbable gelatin sponge packaged in a syringe. The device is available in 2 cube sizes and 3 weight configurations. Once rehydrated, the deformable cubes can be injected into the target vessel with an intravascular catheter or a micro-catheter (depending on the size range) to provide a mechanical barrier to blood flow. Contrast enhancement may be used to monitor the embolization procedure using fluoroscopy. The device is intended for single use and is provided sterile.

    AI/ML Overview

    The provided text is a 510(k) summary for the EmboCube Embolization Gelatin and does not contain information about the acceptance criteria and study proving a device meets these criteria for an AI/ML-based medical device.

    This document is for a physical medical device, an embolization gelatin, and details its substantial equivalence to a predicate device. Therefore, it discusses performance attributes relevant to a physical medical implant (e.g., macroscopic appearance, digestibility, size range, density, moisture content, sterility, pyrogenicity, shelf-life, deliverability) and includes details about an animal study to evaluate its performance (immediate control of bleeding, complications, re-bleeding rate).

    The request, however, asks for information typically associated with the validation of an AI/ML medical device, such as:

    • Acceptance criteria and reported device performance (in an AI/ML context, this would involve metrics like sensitivity, specificity, AUC).
    • Sample sizes for test sets, data provenance (retrospective/prospective, country of origin).
    • Number and qualifications of experts for ground truth establishment.
    • Adjudication methods.
    • MRMC studies and human reader improvement with AI assistance.
    • Standalone algorithm performance.
    • Types of ground truth (expert consensus, pathology, outcomes data).
    • Training set sample size and ground truth establishment for training.

    Since the provided document is about a physical medical device and not an AI/ML device, it does not contain any of the information necessary to address the specific points in the prompt related to AI/ML device validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183578

    Validate with FDA (Live)

    Date Cleared
    2019-06-18

    (179 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use: Torpedo Gelatin Foam is indicated for use in embolization of hypervascular tumors.

    Device Description

    Torpedo Gelatin Foam is a hydrophilic medical device which consists of resorbable dry gelatin foam that is compressed into a cylindrical shape and preloaded into a cartridge with a standard female and male luer fittings. The device is available in two sizes (0.9 mm and 1.7 mm) and two lengths configurations (10 mm and 20 mm).

    Once rehydrated, the deformable torpedoes can be injected into the target vessel with an intravascular catheter or a micro-catheter (depending on the size range) to provide a mechanical barrier to blood flow. Contrast enhancement may be used to monitor the embolization procedure using fluoroscopy. The device is intended for single use and is provided sterile.

    AI/ML Overview

    The provided text is a 510(k) summary for the Torpedo Gelatin Foam, a vascular embolization device. It details the device's characteristics, intended use, and substantial equivalence to a predicate device (EmboCube Embolization Gelatin).

    However, the text does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria in the context of AI/ML device performance. The document is a traditional 510(k) submission for a physical medical device, not a software-as-a-medical-device (SaMD) or an AI-powered diagnostic/therapeutic device.

    Therefore, I cannot extract the following information from the provided text:

    1. A table of acceptance criteria and the reported device performance: This document focuses on demonstrating substantial equivalence through physical and chemical properties, sterilization, and animal studies for an embolization device, not on AI/ML performance metrics like sensitivity, specificity, or AUC.
    2. Sample size used for the test set and the data provenance: There is no "test set" in the context of AI/ML. The document mentions an animal study with 7 test and 8 control animals.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an embolization device involves successful occlusion of vessels and tissue necrosis, assessed through angiography and histopathology in animal models.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the animal study, ground truth was established through angiography (imaging of blood vessels) and histopathology (microscopic examination of tissue), confirming the embolic effect and tissue necrosis. Outcomes data (normal flow not recovered, necrotized renal tissue, clinically normal animals) also served as evidence.
    8. The sample size for the training set: Not applicable. There is no training set for an AI/ML model.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document describes the regulatory submission for a physical medical device, not an AI/ML driven device, and thus does not contain the specific information requested about acceptance criteria and study design for AI/ML performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181021

    Validate with FDA (Live)

    Date Cleared
    2018-09-27

    (163 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EmboCube Embolization Gelatin is indicated for use in embolization of hypervascular tumors.

    Device Description

    EmboCube Embolization Gelatin is a hydrophilic medical device which consists of pre-cut, dry cubes of resorbable gelatin sponge packaged in a syringe. The device is available in 2 cube sizes and 3 weight configurations. Once rehydrated, the deformable cubes can be injected into the target vessel with an intravascular catheter or a micro-catheter (depending on the size range) to provide a mechanical barrier to blood flow. Contrast enhancement may be used to monitor the embolization procedure using fluoroscopy. The device is intended for single use and is provided sterile.

    AI/ML Overview

    The provided text is a 510(k) summary for the EmboCube Embolization Gelatin and does not contain information about the acceptance criteria and study proving a device meets acceptance criteria related to AI/algorithm performance, multi-reader multi-case (MRMC) studies, or ground truth establishment for a test or training set.

    The document describes a medical device, EmboCube Embolization Gelatin, and its substantial equivalence to a predicate device based on various non-AI performance tests and an animal study. Therefore, I cannot address most of the points in your request based on the provided text.

    However, I can extract information related to the device itself and the general summary of testing:

    Device Information:

    • Trade/Device Name: EmboCube Embolization Gelatin
    • Regulation Name: Vascular Embolization Device
    • Regulatory Class: Class II
    • Product Code: KRD
    • Intended Use: For use in embolization of hypervascular tumors.
    • Device Description: Hydrophilic medical device consisting of pre-cut, dry cubes of resorbable gelatin sponge packaged in a syringe. Available in 2 cube sizes and 3 weight configurations. Once rehydrated, deformable cubes can be injected to provide a mechanical barrier to blood flow.

    Study Information (Non-AI related):

    1. A table of acceptance criteria and the reported device performance:
      The document lists the types of tests performed to demonstrate that the device meets "critical design specifications" and "clinical performance attributes," but it does not provide a table of
      specific quantitative acceptance criteria (e.g., "rehydration time must be < X minutes") nor the specific reported performance values for each test. Instead, it states that these tests were "successfully completed."

      Types of Performance Attributes Tested:

      • Macroscopic appearance (visual inspection)
      • Pepsin digestibility
      • Size range
      • Density
      • Moisture content (H2O analysis)
      • Hydration test
      • Effect on pH variation
      • Sterility
      • Pyrogenicity (Bacterial endotoxins)
      • Deliverability with the recommended catheters and radiopaque contrast agents
      • Flush test

      Biocompatibility Testing:

      • Cytotoxicity
      • Sensitization
      • Irritation
      • Acute Systemic Toxicity
      • Material-Mediated Pyrogenicity
      • Genotoxicity
      • Hemolysis
      • Complement Activation
      • Chemical characterization and toxicological risk assessment for Subacute/Subchronic Systemic Toxicity, Chronic Systemic Toxicity, and Carcinogenicity.
    2. Sample size used for the test set and the data provenance:

      • Animal Study: "An animal study was conducted on eight test (EmboCube Embolization Gelatin) and four control (Gel-Block 10x Embolization Pledgets, K143038) adult female sheep in the renal arteries over a period of 4 weeks."
      • Data Provenance: Conducted in sheep (animal study). No information on country of origin or retrospective/prospective nature. (Animal studies for device approval are typically prospective).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      Not applicable (N/A) as this is not an AI/algorithm study involving human expert review for ground truth. The animal study evaluated "vascular safety, local tissue effects, and in vivo degradation."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      N/A, as this is not an AI/algorithm study that would involve expert interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      N/A. This document describes a physical medical device, not an AI/algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      N/A. This document describes a physical medical device, not an AI/algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
      For the animal study, the "ground truth" was established by the observed "vascular safety, local tissue effects, and in vivo degradation" in the sheep. These are direct biological and physiological outcomes, not expert consensus on images or pathology as typically understood in AI studies.

    8. The sample size for the training set:
      N/A. This document describes a physical medical device, not an AI/algorithm with a training set.

    9. How the ground truth for the training set was established:
      N/A. This document describes a physical medical device, not an AI/algorithm with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181300

    Validate with FDA (Live)

    Date Cleared
    2018-07-16

    (60 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Embosphere Microspheres are indicated for use in the embolization of:
    • Hypervascular tumors, including symptomatic uterine fibroids
    • Prostatic arteries for symptomatic benign prostatic hyperplasia (BPH)
    • Arteriovenous malformations
    • Blood vessels to occlude blood flow to control bleeding/hemorrhaging in the peripheral vasculature (subject of this Traditional 510(k))

    Device Description

    Embosphere Microspheres are small, compressible, hydrophilic, biocompatible spheres made of acrylic polymer and porcine-derived gelatin. The microspheres are packaged in 0.9% saline and are provided sterile and non-pyrogenic in a vial or in a syringe.
    The product is provided in seven size ranges to allow physicians to choose the calibration necessary for the vessel being embolized. The size ranges available are:

    • 50-100 microns
    • 40-120 microns
    • 100-300 microns
    • 300-500 microns
    • 500-700 microns
    • 700-900 microns
    • 900-1200 microns .
      The principles of operation for the subject device Embosphere Microspheres are the same as the predicate device Embosphere Microspheres (K021397) and reference devices Embosphere Microspheres (K991549, DEN160040). Embosphere Microspheres are permanent implantable devices and are designed for controlled, targeted embolizations for Embosphere Microsphere Microspheres involve arterial embolization; embolization of uterine fibroids, arteriovenous malformations, hypervascular tumors and benign prostatic hyperplasia involve an embolization of the arteries supplying those areas. The procedure of arterial embolization is similar for all arteries. Appropriately sized microspheres for target vessel occlusion are chosen by the trained interventional radiologist. The delivery procedure involves arterial access through an artery, using a quidewire and microcatheter under fluoroscopic guidance. Once the catheter tip is placed in the artery(ies) supplying the targeted tissue Embosphere Microspheres mixed with a non-ionic contrast agent are delivered in a controlled manner under visualization to occlude the feeding vessel(s) to stop blood flow to the targeted area. The device is intended for single use.
    AI/ML Overview

    The provided document is a 510(k) Summary for Embosphere Microspheres, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed acceptance criteria and a study to prove a device meets those criteria directly.

    Given the context of a 510(k) submission, the "acceptance criteria" and "device performance" are framed in terms of substantial equivalence to a previously cleared predicate device. The primary "study" is a literature review of clinical data to support the safety and effectiveness of the existing device for a new indication.

    Here's an attempt to extract and synthesize the information based on your request, understanding that the document is not an efficacy study report in the traditional sense for a novel AI device:


    Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" for the subject device (Embosphere Microspheres with an expanded indication) are effectively demonstrating substantial equivalence to the predicate Embosphere Microspheres (K021397) and reference devices (K991549, DEN160040) in terms of intended use, design, technological characteristics, and safety/performance. The new indication itself needed to be supported by existing safety and effectiveness data.

    Table of Acceptance Criteria (Implied by Substantial Equivalence) and Reported Device Performance:

    Acceptance Criteria (Implied for Substantial Equivalence to Predicate)Reported Device Performance and Characteristics (Subject Device)
    Intended Use: Similar fundamental purpose as predicate.Intended Use: Embolization of: Hypervascular tumors (including symptomatic uterine fibroids), Prostatic arteries for symptomatic BPH, Arteriovenous malformations, AND Blood vessels to occlude blood flow to control bleeding/hemorrhaging in the peripheral vasculature (new Indication For Use). The function of both the subject and predicate device is to occlude blood vessel(s) to stop blood flow to a targeted area.
    Technological Characteristics: Identical to predicate device.Technological Characteristics: Identical to predicate. Includes: - Shelf Life (single use): Three years (36 months) - Material (spheres): Acrylic polymer and porcine-derived gelatin - Physical Characteristics: Biocompatible, hydrophilic, compressible, non-resorbable - Microspheres Size: Seven size ranges (50-100 to 900-1200 microns) - Sterilization: Steam sterilized - Pyrogenicity: Non-pyrogenic - Performance: Designed for controlled, targeted embolization at the desired level of vessel occlusion - Principle of Operation: Microspheres administered with contrast medium into artery via catheter - Volume of microspheres per container: 1ml or 2ml in 0.9% saline - Packaging: 8-mL glass vial or 20-mL plastic syringe
    Safety and Effectiveness: Demonstrate safety and effectiveness for the proposed indication, not raising new safety/effectiveness issues.Safety and Effectiveness: Extensive clinical data from 40 publications (8 prospective, 32 retrospective, 2000-2018) with 662+ patients reviewed. Embolization with Embosphere was found to be an effective method to control bleeding/hemorrhaging in the peripheral vasculature, with a low complication rate. Adverse events (catheterization complications, post-embolization syndrome, non-targeted embolization) are known and addressed in the predicate device's IFU. No new safety/effectiveness issues were identified with the expanded indication.

    Study Details: Clinical Literature Review (Not a direct device performance study)

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size (Patients): At least 662 patients in the reviewed literature.
      • Data Provenance: Clinical data from 40 publications (2000-2018). These included 8 prospective trials and 32 retrospective trials. The document does not specify the country of origin for the data, but it's reasonable to assume a mix of international and potentially US-based studies given the FDA review. The data consists of both retrospective and prospective studies.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • This section is not applicable in the traditional sense of an AI device's test set. The "ground truth" here is the clinical outcomes reported in published literature regarding the safety and effectiveness of Embosphere Microspheres for embolization. The ground truth was established by the clinical judgment and findings of the physicians and researchers who conducted the studies included in the literature review. The qualifications of these clinicians would implicitly be interventional radiologists or other specialists performing embolization procedures. The FDA, through its review process, acts as the ultimate body verifying the acceptability of the evidence presented.
    3. Adjudication Method for the Test Set:

      • Not applicable for a literature review. The clinical outcomes were observed and reported in the individual studies. The FDA's review process then evaluated the aggregated evidence.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This document is for a medical device (microspheres for embolization), not an AI/CADe device that assists human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a medical device, not an algorithm.
    6. The type of ground truth used:

      • Clinical Outcomes Data: The ground truth used was the reported clinical outcomes, safety profiles, and effectiveness data derived from published clinical studies (prospective and retrospective trials) on the use of Embosphere Microspheres for embolization procedures, specifically for controlling bleeding/hemorrhaging in the peripheral vasculature.
    7. The Sample Size for the Training Set:

      • Not applicable. This is not an AI/machine learning device. The "training data" in a conceptual sense would be the cumulative clinical experience and data that formed the basis for previous clearances of Embosphere Microspheres and the general understanding of embolization procedures.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable. As above, not an AI device with a distinct training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172372

    Validate with FDA (Live)

    Date Cleared
    2017-09-06

    (30 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    2 - 65
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QuadraSphere Microspheres are indicated for embolization of hypervascularized tumors including hepatoma, and peripheral arteriovenous malformations.

    Device Description

    QuadraSphere Microspheres are sterile, biocompatible, hydrophilic, nonresorbable, expandable, acrylic copolymer microspheres. QuadraSphere Microspheres are packaged in a vial containing 25 mg of dry microspheres. The expansion rate is dependent on ionic concentration. When in contact with blood, non-ionic contrast media, or normal saline (NaCl 0.9%), QuadraSphere Microspheres expand to approximately 4 times their dry state diameter; equal to 64x volume. They are available in a range of sizes.

    AI/ML Overview

    This document describes the QuadraSphere Microspheres, a vascular embolization device. The main change in this submission is the extension of the product line to include a 20-40 µm size, in addition to existing sizes.

    Here's the information regarding the acceptance criteria and study as requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a precise "acceptance criteria" table with numerical targets, but it lists performance parameters tested and states that the device met the pre-determined acceptance criteria for each.

    Parameter TestedReported Device Performance
    Macroscopic appearance of the powderMet pre-determined acceptance criteria
    Microscopic examination of the microspheresMet pre-determined acceptance criteria
    Swelling in Hexabrix of the microspheresMet pre-determined acceptance criteria
    Humidity of the microspheresMet pre-determined acceptance criteria
    Bacterial endotoxinsMet pre-determined acceptance criteria
    Catheter compatibilityMet pre-determined acceptance criteria
    Animal testingMet pre-determined acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes for each specific test (e.g., number of microsphere batches for appearance, number of animals for in-vivo testing).

    • Provenance: The studies were conducted by Biosphere Medical, S.A. No specific country of origin for the data is mentioned beyond the company's location in France. The document describes "design verification and validation testing," implying this was prospective testing directly supporting the device's clearance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This is not a study requiring expert readers to establish ground truth for image interpretation or diagnosis. The tests are focused on physical and biological properties of the microspheres themselves.

    4. Adjudication Method for the Test Set:

    Not applicable. This is a technical performance study, not one requiring adjudication of diagnostic interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No. This document does not describe an MRMC study. The device is a physical product (microspheres), not an AI algorithm for diagnostic assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    Not applicable. This device is not an AI algorithm.

    7. The Type of Ground Truth Used:

    The ground truth for the performance tests was based on established engineering and biological specifications for medical device manufacturing and safety. For example:

    • Physical properties (appearance, microscopic features, swelling) are typically compared against pre-defined specifications for the material and design.
    • Biocompatibility is assessed against ISO 10993-1 standards.
    • Bacterial endotoxin levels are measured against established safety limits.
    • Catheter compatibility is tested against functional requirements for delivery.
    • Animal testing would have pre-defined endpoints to evaluate safety and function in a biological system.

    8. The Sample Size for the Training Set:

    Not applicable. This device is not an AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. This device is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153102

    Validate with FDA (Live)

    Date Cleared
    2015-11-23

    (28 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QuadraSphere Microspheres are indicated for embolization of hypervascularized tumors including hepatoma, and peripheral arteriovenous malformations.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter from the FDA to Biosphere Medical, S.A. regarding their Quadrasphere Microspheres. This document largely focuses on regulatory approval and indications for use, and therefore does not contain information about acceptance criteria or a study proving the device meets said criteria.

    The typical content of such a study would include:

    1. Acceptance Criteria Table and Device Performance: A table outlining specific performance metrics (e.g., size distribution, mechanical strength, biocompatibility, embolization efficiency, clinical success rates, complication rates) with pre-defined acceptance limits and the observed performance of the Quadrasphere Microspheres.
    2. Sample Size and Data Provenance: Details about the number of patients or samples included in the test set, as well as the geographical origin of the data (e.g., USA, France, multi-national) and whether the data was collected retrospectively (from existing records) or prospectively (specifically for the study).
    3. Ground Truth Experts: Information on the number of medical experts (e.g., interventional radiologists, pathologists) who established the ground truth for the test set, along with their qualifications (e.g., years of experience, board certifications).
    4. Adjudication Method: If applicable, how discrepancies among experts were resolved to establish the ground truth (e.g., 2+1 means two experts agree, and a third is consulted if they disagree; 3+1 implies a similar consensus-based approach).
    5. MRMC Comparative Effectiveness Study: If a multi-reader, multi-case study was conducted, the document would describe its design and report the effect size quantifying how much human readers' performance (e.g., accuracy, diagnostic confidence) improved when using Quadrasphere Microspheres compared to a control or alternative treatment.
    6. Standalone Performance Study: Whether a study evaluating the device's performance in isolation (without human intervention, if applicable to the device type) was performed. For an embolic agent, this might relate to its physical properties and in-vitro performance.
    7. Ground Truth Type: The specific method used to establish the "true" clinical outcome or characteristic against which the device's performance was measured (e.g., confirmation by pathology after embolization, long-term patient outcomes, expert consensus on imaging results).
    8. Training Set Sample Size: If a machine learning component were involved (which is not the case for this physical medical device), the size of the dataset used to train the algorithm would be specified.
    9. Training Set Ground Truth: How the ground truth for any training set was established.

    Without a detailed study report, which is not part of this 510(k) letter, I cannot provide the requested information.