Found 18 results

510(k) Data Aggregation

    K Number
    K153695

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-01-21

    (29 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a lumbar intervertebral body fusion device, the Biomet Spine Fusion System is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement

    Device Description

    The Biomet Spine Fusion System implants are intervertebral body fusion/vertebral body replacement devices consisting of a rectangular or semi-rectangular shape and various heights and footprints. The devices have a hollowed out central area to accommodate bone graft. The upper and lower surfaces of the devices have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The implants are available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The implants are made of PEEK (OPTIMA LT1®) per ASTM F2026, titanium alloy (Ti-6A1-4V ELI) per ASTM F136, commercially pure titanium per ASTM F1580, and may include tantalum markers per ASTM F560. The expandable Biomet Spine Fusion System implants incorporate PEEK (OPTIMA LT1® per ASTM F2026) spacer components and titanium alloy (Ti-6Al-4V ELI per ASTM F136) endplate components that orient into predetermined lordotic angles of 0, 8 and 12 degrees. The expandable implants are available in 7 to 14mm heights, an 11mm width, and 25, 30 and 35mm lengths. The subject Biomet Spine Fusion System implant is provided sterile only.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Biomet Spine Fusion System. The submission is to add a tantalum marker to the cam locking mechanism of the existing device. This is a modification to an existing device, not a new AI/ML-enabled device, and therefore does not include the detailed performance study information typically associated with AI/ML device submissions.

    Here's a breakdown of the information provided, tailored to the requested categories where applicable:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Mechanical Performance: Demonstrate substantial equivalence to predicate device (K151483) as per FDA guidance and ASTM standards (ASTM F2077 for static/dynamic axial compression, static/dynamic shear compression, static torsion; ASTM F2267 for subsidence).An engineering analysis demonstrated that no new mechanical testing was required for the modified device (adding a tantalum marker). The analysis demonstrated substantial equivalence of the new subject devices to the predicate device. The modified Biomet Spine Fusion System met all specified criteria.Mechanical testing on the predicate device (K151483) was conducted and demonstrated substantial equivalence to its own predicate devices.
    Safety and Effectiveness: The modifications should not raise new safety or performance questions.The modified Biomet Spine Fusion System met all specified criteria and did not raise new safety or performance questions. The minor differences in the new components (tantalum marker) do not raise any new issues of safety or effectiveness.
    Intended Use & Indications for Use: Maintain the same intended use and indications for use as the predicate.The modified Biomet Spine Fusion System implants included in this submission have the same intended use, indications, technological characteristics, and principles of operation as the previously cleared Biomet Spine Fusion System (K151483).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not applicable. For this specific 510(k) submission, there isn't a "test set" in the context of clinical data for an AI/ML device. The evaluation was primarily engineering analysis and mechanical testing comparison to a predicate device.
    • Data Provenance: Not applicable. The "data" here refers to mechanical testing results and engineering analysis, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. This is not an AI/ML device requiring expert ground truth for image or clinical interpretation. The evaluation was based on engineering and mechanical testing standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. As above, there is no "test set" in the sense of clinical data requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the predicate device's mechanical testing, the "ground truth" would be the established ASTM standards (F2077, F2267) and the performance criteria outlined in FDA guidances (Spinal System 510(k)s, Class II Special Controls Guidance Document: Intervertebral Body Fusion Device). For the current submission, the ground truth is simply the established safety and performance of the predicate device (K151483) against which the minor modification is compared via engineering analysis.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device.

    In summary, this 510(k) is for a minor modification (adding a tantalum marker) to an existing, previously cleared spinal fusion device. The "study" proving the device meets acceptance criteria is an engineering analysis, which determined that the modification did not alter the mechanical performance or safety/effectiveness of the device and thus no new mechanical testing or clinical studies were required beyond what was done for the original predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151483

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-07-31

    (59 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a lumbar intervertebral body fusion device, the Biomet Spine Fusion System is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement.

    Device Description

    The Biomet Spine Fusion System implants are intervertebral body fusion/vertebral body replacement devices consisting of a rectangular or semi-rectangular shape and various heights and footprints. The devices have a hollowed out central area to accommodate bone graft. The upper and lower surfaces of the devices have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The implants are available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The implants are made of PEEK (OPTIMA LT1®) per ASTM F2026, titanium alloy (Ti-6A1-4V ELI) per ASTM F136, commercially pure titanium per ASTM F1580, and may include tantalum markers per ASTM F560. The expandable Biomet Spine Fusion System implants incorporate PEEK (OPTIMA LT1® per ASTM F2026) spacer components and titanium alloy (Ti-6Al-4V ELI per ASTM F136) endplate components that orient into predetermined lordotic angles of 0, 8 and 12 degrees. The expandable implants are available in 7 to 14mm heights, an 11mm width, and 25, 30 and 35mm lengths. The subject Biomet Spine Fusion System implant is provided sterile only.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the Biomet Spine Fusion System, indicating it is a medical device submission to the FDA. The document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific acceptance criteria through a clinical study. Therefore, much of the requested information regarding acceptance criteria, study design, ground truth, and expert involvement is not present in this type of regulatory submission. The performance assessment here is primarily based on mechanical testing, not human reader performance or clinical outcomes.

    Here's the information that can be extracted and a clear statement about what is not available based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a formal table of acceptance criteria with corresponding performance metrics in a clinical context. Instead, it states that:

    Acceptance Criteria (Implied)Reported Device Performance
    Meet or exceed predicate device performance in mechanical tests"In all instances, the modified device met or exceeded predicate device performance, functioned as intended..."
    Function as intended"...functioned as intended and therefore demonstrated substantial equivalence to the predicate device(s)."
    Comparable properties to previously cleared devices"Performance data presented also demonstrated comparable properties to the previously cleared devices."

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable in the context of this mechanical testing study. The performance data refers to results from physical tests of the device components.
    • Data Provenance: Not applicable. The data is generated from mechanical testing of the device components, not from patient data, and thus does not have a country of origin or a retrospective/prospective classification in the typical sense of a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth, in this context, would relate to clinical outcomes or expert diagnoses, which are not part of this mechanical testing. The "ground truth" for mechanical testing would be the established engineering specifications and performance limits of the predicate device.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods are relevant for clinical trials or studies involving expert assessment, which is not described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission is for a physical intervertebral body fusion device and does not involve AI or human image readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone (algorithm-only) performance evaluation was not done. This device is a physical implant, not a software algorithm.

    7. The type of ground truth used

    For the mechanical testing, the "ground truth" is established by engineering safety and performance standards (e.g., ASTM F2077, ASTM F2267) and the performance specifications of the legally marketed predicate devices. The goal is to demonstrate that the new device's mechanical properties are equivalent to or better than those of the predicate.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that would have a training set. The "training" in product development for such a device refers to engineering design and testing iterative processes, not data training.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set in the context of machine learning. The design and testing of the device are based on established engineering principles, material science, and regulatory guidance for spinal implants.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133518

    Validate with FDA (Live)

    Date Cleared
    2014-11-05

    (355 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MaxAn® Anterior Cervical Plate System is intended for anterior interbody fixation of the cervical spine. Indications for use include the temporary stabilization of the anterior spine during the development of cervical fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin confirmed by patient history and radiographic studies), trauma including fractures, tumors, deformity (defined as kyphosis, lordosis), pseudarthroses, and/or failed previous fusions. The intended levels for treatment range from C2-T1.

    Device Description

    The MaxAn® Anterior Cervical Plate System Cervical Plate System consists of titanium alloy plates and screws intended for use in anterior cervical discectomy and fusion (ACDF) procedures. Cervical plates are intended for use at 1-5 levels, and are available in lengths from 8mm-130mm. Screws are available in both fixed and variable versions in sizes 4.0mm and 4.5mm in various lengths. All components are available in both sterile and non-sterile configurations.

    The purpose of this submission is to add MR Conditional labeling for the components of the MaxAn system.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the MaxAn® Anterior Cervical Plate System, based on the provided text:

    Important Note: The provided document is a 510(k) summary for adding MR Conditional labeling to an already cleared device (MaxAn® Anterior Cervical Plate System). Therefore, the "study" described herein specifically pertains to MR compatibility, not the primary clinical effectiveness or safety of the spinal fixation itself, which would have been established in the original K080646 submission. The information below reflects what's available in this specific document.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (ASTM Standard)Reported Device Performance (MaxAn® Anterior Cervical Plate System)
    ASTM F2052-06 (Magnetically Induced Displacement Force on Medical Devices)Test results indicate that the device meets the ASTM recommendations for MR Conditional labeling in terms of magnetically induced displacement.
    ASTM F2119-07 (Evaluation and Labeling of MR Image Artifacts)Test results indicate that the device meets the ASTM recommendations for MR Conditional labeling in terms of MRI artifacts.
    ASTM F2182-11a (Heating of Medical Devices During MRI)Test results indicate that the device meets the ASTM recommendations for MR Conditional labeling in terms of device heating.
    ASTM F1717-04 (Static & Dynamic Testing of Spinal Constructs)Qualitative torque testing was presented. (The document states these tests were conducted, implying performance met standards, but specific quantified acceptance criteria and performance values are not detailed in this summary for these mechanical tests).
    Static compression bending(Implied to meet ASTM F1717-04 recommendations, but specific criteria/results not detailed.)
    Static torsion(Implied to meet ASTM F1717-04 recommendations, but specific criteria/results not detailed.)
    Dynamic compression(Implied to meet ASTM F1717-04 recommendations, but specific criteria/results not detailed.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified in this 510(k) summary. The document refers to "testing conducted" according to ASTM standards, which would imply a number of device samples were used for each test. Specific quantities are not provided.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The testing described is bench testing (mechanical and MR compatibility), not human or animal studies, so "retrospective or prospective" is not applicable in the typical clinical sense. The tests were performed to evaluate the physical properties of the device components.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not applicable as the "ground truth" for these tests is the direct measurement of physical properties against established ASTM standards, not expert interpretation of clinical data or images. The "experts" involved would be engineers and technicians executing the tests and interpreting the raw data, but their number and specific qualifications are not detailed in this regulatory summary.

    4. Adjudication Method for the Test Set

    • This is not applicable. As these are bench tests measuring physical properties, there is no "adjudication method" in the sense of reconciling differing expert opinions. The outcome is determined by whether the measured physical property (e.g., heating, displacement) falls within the acceptable limits defined by the ASTM standard.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed as described in this document. This submission pertains to MR Conditional labeling for a device, which involves bench testing, not clinical performance comparison with human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • This is not applicable. The device is a physical spinal implant, not an algorithm or AI system. The testing performed is to assess the physical device's compatibility with an MRI environment.

    7. Type of Ground Truth Used

    • The "ground truth" used for these tests is the objective measurements of physical forces, temperatures, and image artifacts as defined and evaluated against the specified ASTM (American Society for Testing and Materials) standards. These standards provide the accepted methodology and limits for determining MR compatibility and mechanical performance.

    8. Sample Size for the Training Set

    • This is not applicable. The device is a physical implant, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    • This is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141804

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-09-25

    (84 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system for use with autograft. The Polaris Spinal System is indicated for the following conditions: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, Scheuerman's disease, and/or lordosis,), turnor, stenosis, pseudoarthrosis, or failed previous fusion.

    The Ballista and Cypher MIS instruments are intended to be used with Ballistal Cypher MIS Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/ Cypher MIS instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the above indications.

    For pediatric patients, the Polaris Spinal System may be used for posterior, non-cervical pedicle screw fixation as an adjunct to fusion to treat adolescent idiopathic scoliosis and is also indicated for treatment of the following conditions: spondylolisthesis/spondylolysis and fractures caused by tumor and/or trauma. Pedicle screw fixation is limited to a posterior approach.

    The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System, Lineum OCT Spine System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.

    Device Description

    The Polaris Spinal System is a non-cervical spinal fixation device made from titanium alloy (Ti-6Al-4V) per ASTM F136, unalloyed titanium per ASTM F67, stainless steel per ASTM F138 or ASTM F1314 and Cobalt Chrome Alloy (Co-28Cr-6Mo) per ASTM F1537. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos, various cross connectors and accessories. This submission is to clear modifications to the Polaris Spinal System to include Hydroxyapatite (HA) Coated Screws and a variation in shaft geometry on the multiaxial screws.

    AI/ML Overview

    This FDA 510(k) premarket notification describes the Biomet Spine, LLC's Polaris Spinal System. The provided document focuses on demonstrating substantial equivalence to predicate devices rather than providing a study proving performance against specific acceptance criteria in the manner typically seen for software or AI medical devices.

    Therefore, many of the requested sections (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set details) are not applicable to this type of submission. This document describes a medical device, specifically a spinal fixation system, not a software algorithm or AI. The "performance data" refers to mechanical testing against established standards for spinal implants.

    Here's the information extracted from the document based on the provided request format, with explanations where certain information is not present or not applicable:


    Acceptance Criteria and Study to Prove Device Meets Criteria: Polaris Spinal System (K141804)

    This 510(k) submission for the Polaris Spinal System focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than establishing performance against specific, quantifiable acceptance criteria for a novel functionality. The "acceptance criteria" here are implicitly meeting the performance and safety profiles of the predicate devices as demonstrated through mechanical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Category (Implicit via Substantial Equivalence to Predicates)Reported Device Performance
    Mechanical Strength & DurabilityStatic compression bending, static torsion, and dynamic compression bending fatigue per ASTM F1717 met requirements.
    Biocompatibility (HA Coated Screws)Followed FDA guidance document "Information Needed for Hydroxyapatite Coated Orthopedic Implants dated March 10, 1995"
    Material PropertiesMade from titanium alloy (Ti-6Al-4V) per ASTM F136, unalloyed titanium per ASTM F67, stainless steel per ASTM F138 or ASTM F1314, and Cobalt Chrome Alloy (Co-28Cr-6Mo) per ASTM F1537.
    Intended Use & Indications for UseSame as, or similar to, predicate devices (K140123, K090203, K122226).
    Fundamental TechnologySame as, or similar to, predicate devices (design, materials, manufacturing methods, sterility, operational principles).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable. For mechanical testing of medical implants, "test sets" typically refer to the number of constructs or samples subjected to each mechanical test. The document mentions "worst-case construct testing" was performed, but does not specify the number of samples per test type.
    • Data Provenance: Not applicable in the context of clinical data. The data provenance here refers to the mechanical testing conducted on the device components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. Ground truth for mechanical testing is established by adherence to recognized engineering standards (e.g., ASTM F1717) and internal protocols, not clinical expert consensus.
    • Qualifications of Experts: Not specified. Testing would be performed by engineers and technicians specializing in biomechanical testing.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Performance is judged against predefined criteria within the ASTM standards, not by expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No. This is not a software/AI device, but a physical implant. Therefore, MRMC studies are not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance Study: Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for this medical device submission is the adherence to established mechanical testing standards (specifically ASTM F1717) and FDA guidance documents for spinal systems and HA-coated implants. This ensures the device's physical properties meet recognized safety and performance benchmarks for spinal implants.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This refers to a physical device, not a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140710

    Validate with FDA (Live)

    Date Cleared
    2014-09-04

    (167 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biomet Graft Delivery Syringes are intended to deliver autograft, allograft or synthetic bone grafting materials to all orthopedic surgical sites.

    Device Description

    The Biomet Graft Delivery Syringes provide the surgeon with a method to manually deliver bone grafting material such as allograft, autograft and synthetic bone grafting materials to all orthopedic surgical sites. The sterile syringes are available in three different sizes to hold different volumes of grafting materials. The graft delivery syringe consist of a syringe barrel with a plunger, a removable female luer cap and removable end cap and will be provided empty or pre-filled with commercially available bone grafting materials. The syringes will be available with and without an internal stainless steel strainer.

    AI/ML Overview

    This is a 510(k) premarket notification for the Biomet Graft Delivery Syringes. The document indicates that the device is substantially equivalent to legally marketed predicate devices, and therefore does not provide the detailed acceptance criteria and study data typically found in a clinical trial report or a performance study summary for novel devices.

    However, based on the provided text, here's what can be extracted and what information is not available:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Functional Performance"device functions as intended to deliver bone graft materials to an orthopedic surgical site."
    Substantial Equivalence"The Biomet Graft Delivery System is substantially equivalent to the predicate syringes as a graft delivery syringe in regards to intended use, indications for use, fundamental technology including design, materials, manufacturing methods, sterility, and operational principles."
    Safety & Effectiveness"does not raise any new issues regarding safety and effectiveness."

    Missing Information: Specific quantitative acceptance criteria (e.g., maximum force to deliver graft, leakage rates, specific flow rates) and the precise quantitative performance results are not detailed in this summary. The performance is stated in qualitative terms (e.g., "functions as intended," "substantially equivalent").

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for the "simulated use testing."
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). The testing is described as "simulated use testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided as the study described is a "simulated use testing" focused on device function, not on clinical interpretation or an evaluation requiring expert ground truth for diagnostic accuracy.

    4. Adjudication method for the test set:

    • This information is not provided and is not applicable given the nature of the described "simulated use testing."

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a medical instrument (syringe) and not an AI-powered diagnostic tool, so such a study would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. This device is a physical instrument, not an algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the simulated use testing would likely be based on engineering specifications and functional requirements for the device (e.g., successful delivery of graft material without clogging, leakage, or structural failure). It is not based on expert consensus, pathology, or outcomes data in the clinical sense for this type of device.

    8. The sample size for the training set:

    • Not applicable. This device is hardware; there is no "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    Summary of Study:

    The document describes "Simulated use testing" which was conducted to demonstrate that the Biomet Graft Delivery Syringes function as intended to deliver bone graft materials. The purpose of this testing was to support the claim of substantial equivalence to predicate devices, ensuring that the new device does not raise new safety or effectiveness concerns. The details of this simulated use testing (e.g., protocols, specific pass/fail criteria, actual quantitative results, sample size) are not provided in this 510(k) summary, only the conclusion that it successfully demonstrated the device functions as intended.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132373

    Validate with FDA (Live)

    Date Cleared
    2014-03-27

    (240 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biomet probes/guidewires and dilators are intended for tissue dilation and stimulation of peripheral nerves including spinal nerve roots for location and identification during spinal surgery.

    Device Description

    The Biomet insulated probes/guidewires and dilators are intended to dilate tissue and be used as intraoperative neuromonitoring (IONM) stimulation accessories by connectially available IONM equipment for locating and identifying peripheral nerves including spinal nerve roots during spinal surgery. The Biomet insulated probes/guidewires have a length of 305mm, an exposed stimulation area of .167 cm², and an outer diameter of 2.6mm. The Biomet insulated dilators have a length ranging from 229mm to 254mm, an exposed stimulation are of .167 cm², an inner diameter ranging from 3.2mm to 9.8mm, and an outer diameter ranging from 5.8mm to 13.8mm.

    AI/ML Overview

    This Biomet submission (K132373) is for nerve stimulation accessories (probes/guidewires and dilators), not for a device that relies on AI/ML. Therefore, the requested information elements related to AI/ML device performance (such as sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment) are not applicable and are not present in this 510(k) summary.

    The acceptance criteria and supporting studies for this device are focused on electrical and mechanical performance, as well as biocompatibility, to demonstrate substantial equivalence to previously cleared predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/EvaluationAcceptance Criteria (Implicit from Predicate Equivalence)Reported Device Performance
    BiocompatibilityCytotoxicityNon-cytotoxic (ISO 10993-5)No evidence of cytotoxicity
    SensitizationNon-sensitizing (ISO 10993-10)No evidence of sensitization
    IrritationNon-irritant (ISO 10993-10)No evidence of irritation
    Systemic ToxicityNo systemically toxic (ISO 10993-11)No signs of toxicity
    Pyrogen TestNon-pyrogenicNo signs of pyrogens
    Electrical PerformanceTotal ResistanceEquivalent to predicate devicesResults demonstrate equivalence
    System ImpedanceEquivalent to predicate devicesResults demonstrate equivalence
    Current DensityEquivalent to predicate devicesResults demonstrate equivalence
    Mechanical PerformanceMechanical TestingEquivalent to predicate devicesResults demonstrate equivalence
    Material ConformanceWrought Stainless SteelsASTM F899:2012 complianceDevice conforms to ASTM F899:2012
    SterilizationRadiation SterilizationAAMI/ANSVISO 11137-2:2006 complianceDevice conforms to AAMI/ANSVISO 11137-2:2006
    Ethylene Oxide SterilizationANSI/AAMI/ISO 11135-1:2007 complianceDevice conforms to ANSI/AAMI/ISO 11135-1:2007
    PackagingPerformance TestingASTM D4169:2009 complianceDevice conforms to ASTM D4169:2009

    2. Sample size used for the test set and the data provenance
    The document does not specify the exact sample sizes for each bench test conducted (Total Resistance, System Impedance, Current Density, Mechanical/Electrical Testing) or for the biocompatibility tests. It states that "Biocompatibility testing was conducted" and "Bench testing was conducted," implying a sufficient number of samples were tested to gain confidence in the results, but the specific N is not provided. The data provenance is internal testing by Biomet Spine.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. The ground truth for this type of device (electrical stimulus and dilator) is based on objective measurements from bench testing and adherence to recognized standards, not subjective expert assessment of data like images or clinical outcomes.

    4. Adjudication method for the test set
    Not applicable. There was no expert adjudication process involved for the objective bench and biocompatibility testing. The outcomes are concrete measurements against defined standards or comparisons to predicate device performance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is not an AI/ML-driven diagnostic or assistive technology for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This device is a physical medical instrument (probes/guidewires and dilators) used with commercially available neuromonitoring equipment during surgery. It does not have an "algorithm-only" or "standalone" performance in the AI/ML sense.

    7. The type of ground truth used
    The "ground truth" for this device's performance is established through:

    • Objective Bench Test Measurements: Electrical properties (resistance, impedance, current density) and mechanical integrity were measured and compared to predicate devices' known performances and acceptable ranges.
    • Biocompatibility Standards: Adherence to international standards like ISO 10993 (Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Pyrogenicity) where "ground truth" is defined by the absence of adverse biological reactions as per the standard's methodologies.
    • Material and Sterilization Standards: Compliance with recognized ASTM and AAMI/ISO standards for materials and sterilization processes.

    8. The sample size for the training set
    Not applicable. This is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133746

    Validate with FDA (Live)

    Date Cleared
    2014-03-12

    (93 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system for use with autograft and/or allograft. The Polaris Spinal System is indicated for the following conditions: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, Scheuermann's disease, and/or lordosis,), tumor, stenosis, pseudoarthrosis, or failed previous fusion.

    The Ballista and Cypher MIS instruments are intended to be used with Ballistal Cypher MIS /Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/ Cypher MIS instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the above indications.

    For pediatric patients, the Polaris Spinal System may be used for posterior, non-cervical pedicle screw fixation as an adjunct to fusion to treat adolescent idiopathic scoliosis and is also indicated for treatment of the following conditions: spondylolisthesis/spondylolysis and fractures caused by turnor and/or trauma. Pedicle screw fixation is limited to a posterior approach.

    The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, Lineum OCT Spine System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.

    Device Description

    The Polaris Spinal System is a non-cervical spinal fixation device made from titanium alloy (Ti-6Al-4V) per ASTM F136, unalloyed titanium per ASTM F67, stainless steel per ASTM F138 or ASTM F1314 and Cobalt Chrome Alloy (Co-28Cr-6Mo) per ASTM F1537. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos, various cross connectors and accessories. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This line extension is to add downsized components and additional rod and connector styles to the Polaris Spinal System.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Polaris Spinal System:

    Acceptance Criteria and Device Performance Study

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from FDA Guidance)Reported Device Performance
    Static compression bendingMet all mechanical test requirements
    Static torsionMet all mechanical test requirements
    Dynamic compression bending fatigue (ASTM F1717)Met all mechanical test requirements
    Interconnection testing (ASTM F1798)Met all mechanical test requirements

    Note: The document explicitly states that the device "met all mechanical test requirements based on the worst-case construct testing." The specific numerical thresholds for these acceptance criteria are not provided in this summary but would be detailed in the full test reports referenced by the FDA's "Guidance for Industry and FDA Staff - Spinal System 510(k)s dated May 3, 2004."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated. The document mentions "worst-case construct testing," implying that specific configurations were tested, but the number of samples per test type is not detailed.
    • Data Provenance: This is a retrospective study in the sense that the testing was conducted on samples of the device components. The data is generated from laboratory mechanical testing, not from patient data or clinical trials. The country of origin of the data is implicitly the USA, where the testing was conducted to FDA guidance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This document describes a mechanical performance study, not a study requiring expert interpretation of medical images or patient data to establish ground truth. Therefore, this section is not applicable. The "ground truth" here is the physical performance of the device against established engineering standards.

    4. Adjudication Method for the Test Set

    Not applicable. As this is a mechanical performance study, there is no "adjudication method" in the clinical sense. The results are objective measurements against defined engineering standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for assessing the impact of AI on human reader performance, typically in diagnostic imaging. This document describes mechanical testing of a spinal implant system.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone performance study (in the context of an algorithm or AI) was not done. This document describes the mechanical performance of a physical medical device (spinal implant system).

    7. Type of Ground Truth Used

    The ground truth used for this study is based on established engineering standards and regulatory guidance for spinal implant testing, specifically:

    • ASTM F1717, Standard Test Methods for Static and Fatigue for Spinal Implant Constructs in a Vertebrectomy Model
    • ASTM F1798, Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants
    • FDA's Guidance for Industry and FDA Staff - Spinal System 510(k)s dated May 3, 2004

    8. Sample Size for the Training Set

    Not applicable. This is a study of a physical device's mechanical properties, not an AI or algorithm-based product that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131615

    Validate with FDA (Live)

    Date Cleared
    2013-09-04

    (93 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system for use with autograft and/or allograft. The Polaris Spinal System is indicated for the following conditions: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, Scheuermann's disease, and/or lordosis,), tumor, stenosis, pseudoarthrosis, or failed previous fusion.

    The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the above indications.

    For pediatric patients, the Polaris System may be used for posterior, non-cervical pedicle screw fixation as an adjunct to fusion to treat adolescent idiopathic scoliosis and is also indicated for treatment of the following conditions: spondylolisthesis/spondylolysis and fractures caused by tumor and/or trauma. Pedicle screw fixation is limited to a posterior approach.

    The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, Nextgen Altius OCT Spine System. The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.

    Device Description

    The Polaris Spinal System is a non-cervical spinal fixation device made from titanium alloy (Ti-6Al-4V) per ASTM F 136, unalloyed titanium per ASTM F 67, stainless steel per ASTM F 138 or ASTM F 1314 and Cobalt Chrome Alloy (Co-28Cr-6Mo) per ASTM 1537. The system includes screws, various types and sizes of rods, locking nuts, hooks. lateral connectors, plugs. fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This line extension is to expand the indications for use for Polaris Spinal System using previous cleared system components.

    AI/ML Overview

    This document describes a 510(k) submission for the Polaris Spinal System, a medical device. The information provided heavily focuses on the comparison to predicate devices to establish substantial equivalence rather than detailing specific device performance or clinical studies with acceptance criteria for a novel AI or diagnostic device.

    Therefore, many of the requested categories for a study proving device meets acceptance criteria are not directly applicable or are not explicitly stated in the provided text. The document is primarily a regulatory submission demonstrating equivalence, not a detailed performance study report.

    Here's an analysis based on the provided text, highlighting what is available and what is not:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance Metric / Acceptance CriteriaReported Device Performance
    Mechanical Strength (Static Compression Bending, Static Torsion, Dynamic Compression Bending per ASTM F1717)Implicit historical data: "The implants included in the submission were previously tested for static compression bending, static torsion and dynamic compression bending per ASTM F1717."
    Safety and Effectiveness for Pediatric Pedicle Screw Fixation (especially Adolescent Idiopathic Scoliosis)"The assessment concluded that pedicle screw fixation is safe and effective for use in pediatric patients for the treatment of adolescent idiopathic scoliosis." (Based on clinical literature assessment)
    Substantial Equivalence (Intended Use, Indications, Design, Materials, Manufacturing Methods, Sterility, Fundamental Technology, Operational Principles)The Polaris Spinal System is deemed "substantially equivalent" to multiple predicate devices (Revere" 4.5 Stabilization System, CD Horizon" Spinal System, TSRH* Spinal System, and previous Polaris Spinal System versions) based on these characteristics.

    Missing Information/Not Applicable: The document does not specify quantitative acceptance criteria (e.g., "must achieve X N-m of torque" or "failure rate must be below Y%"). The "reported device performance" for mechanical strength is a statement that it was "previously tested" and "already demonstrated," indicating compliance with prior standards or predicate device performance, but no specific values or criteria are provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a "test set" in the context of a clinical study with a specific sample size.

    • Mechanical Testing: While mechanical tests per ASTM F1717 were performed, the sample size for these tests is not mentioned.
    • Clinical Literature Assessment: This involved reviewing "published clinical data." The number of patients or studies included in this assessment is not specified, nor is the country of origin, or whether it was retrospective or prospective. It's a review of existing data, not a new data collection.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable. The document does not describe a study involving expert-established ground truth for a test set. The clinical literature assessment likely involved review by the regulatory affairs team, but not in the context of establishing "ground truth" for a device performance study.

    4. Adjudication Method for the Test Set

    This information is not applicable as no "test set" requiring adjudication is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was done or mentioned. This device is a surgical implant, not a diagnostic imaging AI, so such a study would not typically apply in this direct comparison context.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

    This information is not applicable. The Polaris Spinal System is a physical implant, not an algorithm. Standalone performance for mechanical devices relates to their physical integrity and function under simulated use, which is addressed by the ASTM F1717 testing.

    7. Type of Ground Truth Used

    • Mechanical Testing: The "ground truth" for mechanical testing would be the physical properties and performance metrics defined by the ASTM F1717 standard.
    • Clinical Efficacy (for pediatric pedicle screw fixation): The "ground truth" was derived from a "clinical literature assessment," meaning conclusions drawn from existing published clinical data and expert consensus in the medical community about the safety and effectiveness of pedicle screw fixation for the stated conditions.

    8. Sample Size for the Training Set

    This information is not applicable, as there is no "training set" in the context of a medical device like this. Training sets are relevant for machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123549

    Validate with FDA (Live)

    Date Cleared
    2013-02-25

    (98 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion.

    The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft.

    The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.

    Device Description

    The Polaris Spinal System is a non-cervical spinal fixation device. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This submission is a line extension to Polaris Spinal System to add other styles of multiaxial screw, including a screw with translating feature, and to provide the Polaris Spinal System sterile packed.

    AI/ML Overview

    The provided document describes a 510(k) summary for the Biomet Spine Polaris Spinal System - Translation Screw, which is a non-cervical spinal fixation device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Testing:
    Static compression bending per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Static torsion per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Dynamic compression bending fatigue per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Substantial Equivalence:Reported as substantially equivalent to predicate devices regarding material, design, operational principles, intended use, indications, and fundamental technology.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for algorithmic performance. The testing performed was mechanical testing of the device hardware.

    • Test Set Sample Size: Not applicable in the context of clinical data or algorithm performance.
    • Data Provenance: Not applicable. The "data" comes from physical mechanical tests conducted on the device components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth, in the context of algorithm performance, typically refers to expert-validated clinical diagnoses or outcomes. This submission focuses on mechanical performance data of a physical medical device.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are relevant for resolving discrepancies in expert opinions for clinical ground truth. Mechanical testing has well-defined pass/fail criteria per ASTM standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This device is a physical spinal fixation system, not an AI or imaging diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This submission is for a physical medical device.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the mechanical testing was the established pass/fail criteria defined by ASTM 1717-11a standard test methods for static and fatigue testing of spinal implant constructs in a vertebrectomy model.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical implant, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123451

    Validate with FDA (Live)

    Date Cleared
    2012-12-03

    (24 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion.

    The Ballista/Ballista II instruments are intended to be used with Ballista/Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion that warrant the use of a noncervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft.

    The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.

    Device Description

    The Polaris Spinal System is a non-cervical spinal fixation device. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This submission is a line extension to Polaris Spinal System to add another style of Ballista rods to the system.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Polaris Spinal System - Ballista II Rods," a non-cervical spinal fixation device. This submission is a line extension to add another style of Ballista rods to the existing system. The core assertion is that the new components are "substantially equivalent" to predicate devices, meaning they have the same or similar technological characteristics and do not raise new issues of safety or effectiveness.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Safety and Effectiveness (Substantial Equivalence)The new Ballista II rods are substantially equivalent to predicate rods (K121130 & K061441) with respect to intended use and indications, technological characteristics, and principles of operation. No new issues of safety or effectiveness are presented.
    Mechanical PerformanceAn engineering analysis concluded that mechanical testing was not required for this specific style of rod because the combination of components did not create a new worst-case construct.
    Compatibility with System InstrumentsValidation testing was conducted to verify that the modified rods can be used with the system's instrumentation.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not applicable in the context of this submission. The "study" here is primarily an engineering analysis and validation testing, not a clinical study on human subjects or a dataset in the typical sense for AI/machine learning.
    • Data Provenance: Not applicable. The validation testing would have been performed in a laboratory setting by the manufacturer, Biomet Spine. The data is internal to the company.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Number of Experts: Not applicable. This submission focuses on engineering analysis and mechanical/compatibility testing, not on establishing ground truth from expert interpretations of medical data.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable. There is no human interpretation of data requiring adjudication mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    • No, an MRMC comparative effectiveness study was not done. The submission focuses on device equivalence from a mechanical and functional perspective, not on comparing human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:

    • No, a standalone study (in the context of an algorithm's performance) was not done. This is a hardware device submission, not an AI or software device submission.

    7. The Type of Ground Truth Used:

    • Ground Truth: The "ground truth" in this context is defined by established engineering principles, mechanical testing standards for spinal implants, and the functional compatibility requirements with existing instrumentation. This is engineering/performance data and manufacturing specifications.

    8. The Sample Size for the Training Set:

    • Not applicable. This submission does not involve an AI system with a "training set."

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. This submission does not involve an AI system with a "training set."

    Summary of the Study Proving Device Meets Acceptance Criteria:

    The "study" in this context consists of an engineering analysis and validation testing.

    • Engineering Analysis: This analysis was conducted to determine if new mechanical testing was required for the different style of rod. The conclusion was that because the new component combination did not create a "new worst-case construct" compared to the predicate devices, additional general mechanical testing of the rod itself was not deemed necessary to demonstrate substantial equivalence for the main mechanical properties.
    • Validation Testing: This testing specifically verified the compatibility and functionality of the modified rods with the existing Polaris Spinal System instrumentation. This ensures that surgeons can safely and effectively use the new rods with the established toolset.

    The overall approach to proving the device meets acceptance criteria (which is primarily substantial equivalence) relies on demonstrating that the new components maintain the same intended use, indications, technological characteristics, and principles of operation as legally marketed predicate devices, and that the minor modifications do not introduce new safety or effectiveness concerns, supported by engineering analysis and specific compatibility testing.