Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K234012

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-03-14

    (451 days)

    Product Code
    Regulation Number
    866.1645
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VITEK® COMPACT PRO is intended for the automated quantitative and/or qualitative antimicrobial susceptibility testing of isolated colonies for most clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and yeast.

    The VITEK® COMPACT PRO is also intended for the automated identification of most clinically significant anaerobic organisms and Corynebacterium species, fermenting and nonfermenting Gram-negative bacilli, Gram-positive organisms, fastidious organisms, and yeasts and yeast-like organisms.

    Device Description

    The VITEK® COMPACT PRO instrument is an automated instrument designed for use in low-to medium-range applications in both Clinical and Industry laboratories. The instrument performs sample well filling, incubation, and optical readings. The VITEK® COMPACT PRO instrument is a two-step automated instrument for:

    • Hydrating reagents with sample inoculum
    • Pre-processing cards, incubating cards, and continuous reading for growth

    The VITEK® 2 Systems Software receives the instrument optical readings and performs analysis. The instrument then ejects the completed reagent card into the waste area for disposal.

    The system includes a VITEK® COMPACT PRO instrument with an internal computer, monitor, keyboard, mouse, handheld barcode scanner, and USB hub. The software provided with the internal computer includes analysis and limited data management programs. A bidirectional computer interface (BCI) may transfer results automatically to the user's laboratory information system (LIS).

    A Quality Control System is available to track the quality control results of the test cards. The Advanced Expert System™ (Clinical Use) is available to provide online, systematic validation of results and interpretation of resistant phenotypes found during susceptibility testing.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, the VITEK® COMPACT PRO, and its performance evaluation. However, it does not explicitly define "acceptance criteria" in a table format with specific numerical targets. Instead, it details the results of various performance tests as evidence that the device meets acceptable standards, primarily by demonstrating substantial equivalence to a predicate device.

    Here's an interpretation of the acceptance criteria and study that can be extracted from the provided information:

    Interpretation of Acceptance Criteria and Study Design:

    The VITEK® COMPACT PRO is an automated antimicrobial susceptibility testing system. The core of its acceptance criteria and the study proving it meets these criteria relies on demonstrating substantial equivalence to a previously cleared VITEK® 2 System (N50510 Supplement S082). This means the new device must perform comparably to the predicate device across various metrics.

    1. Table of Acceptance Criteria (Inferred) and Reported Device Performance:

    Since explicit acceptance criteria are not tabulated with thresholds, they are inferred from the reported performance, with the implicit criterion being "comparable to or better than the predicate device."

    Performance MetricInferred Acceptance Criterion (vs. Predicate VITEK® 2)Reported Device Performance (VITEK® COMPACT PRO)
    Quality Control (QC)Acceptable QC results within established rangesQC passing >99% for each of the AST card classes tested (GN, GP, YS, ST) for 39 antimicrobial agents and 13 QC organisms.
    Reproducibility>95% Reproducibility (within 1 doubling dilution)>97% reproducibility in the best case and >95% reproducibility in the worst case for all VITEK® 2 AST test cards/antimicrobials.
    Clinical Performance (Accuracy)High Essential Agreement (EA) with predicateAST GN cards: 98.8% EA
    (Essential Agreement - EA)High Categorical Agreement (CA) with predicateAST GP cards: 99.5% EA
    Clinical Performance (Accuracy)AST ST cards: 98.5% EA
    (Categorical Agreement - CA)AST YS cards: 100% EA
    AST GN cards: 97.3% CA
    AST GP cards: 97.4% CA
    AST ST cards: 98.7% CA
    AST YS cards: 100% CA
    Usability/Safety (Human Factors)Safe and effective for intended users/usesOverall favorable with respect to usability. Minor usability errors mostly due to workflow changes and risk of ignoring warnings, but with appropriate mitigations, found safe and effective.
    System Verification (Technical)Successful completion and passing of requirementsSuccessfully completed for instrument, software/firmware interfaces (VITEK® COMPACT PRO Instrument, VITEK® firmware, VITEK® Systems 10.0 software, VITEK® COMPACT PRO User Interface, VITEK Systems Communications protocols).

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • QC Testing: Each of the 13 quality control organisms was tested at least twenty times. This testing was conducted at "at least three of the clinical evaluation trial sites."
    • Reproducibility: Isolates were tested in triplicate for three trial sites. Individual test sets were selected for each antimicrobial, with each set including a minimum of 10 strains (at least 2 with on-scale results).
    • Clinical Performance Evaluation: "Representative, clinically relevant strains" were tested. Each strain was tested one time with both the VITEK® COMPACT PRO and the VITEK® 2.
    • Data Provenance: The document states testing was done at "at least three of the clinical evaluation trial sites" for QC and "three trial sites" for reproducibility. The country of origin for these sites is not specified, but bioMérieux Inc. is located in Hazelwood, Missouri, USA, implying the studies were likely conducted in the US. The studies appear to be prospective as they involve new testing of isolates for performance evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not explicitly stated in the provided text. The ground truth for the clinical performance evaluation was established by comparing the VITEK® COMPACT PRO's results against the VITEK® 2 System's results (N50510 Supplement S082), which implicitly serves as the "expert consensus" or "reference method" in this context. The document does not describe a separate human expert panel for adjudication or ground truth establishment the way it might for an AI imaging device.

    4. Adjudication Method for the Test Set:

    Not applicable in the traditional sense of human expert adjudication. The comparison was directly between the new device (VITEK® COMPACT PRO) and the predicate device (VITEK® 2 System) for MICs and interpretations (S/SDD/I/R). There's no mention of a 2+1 or 3+1 human expert adjudication process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC study was not done. This device is an automated in vitro diagnostic instrument, not an AI-assisted imaging device for human interpretation. The comparison is machine-to-machine (new instrument vs. predicate instrument).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the primary performance evaluation was a standalone "algorithm only" (instrument only) performance comparison. The VITEK® COMPACT PRO (device under test) generated MICs and interpretations, and these were directly compared to those generated by the predicate VITEK® 2 System. Human "in-the-loop" performance is not relevant for the core function of this automated susceptibility system in the context of its performance study.

    7. The type of ground truth used:

    The ground truth for the clinical performance evaluation was the MIC results and categorical interpretations (S/SDD/I/R) obtained from the legally marketed predicate device, the VITEK® 2 System (N50510 Supplement S082). This serves as the reference method against which the new device's performance is measured.

    8. The Sample Size for the Training Set:

    The document does not discuss a separate "training set" in the context of device development or performance evaluation. This is not an AI/ML device where a distinct training dataset is typically used. The development process likely involved internal verification and validation, but the reported studies (QC, Reproducibility, Clinical Performance) are essentially testing/validation datasets.

    9. How the ground truth for the training set was established:

    Not applicable, as no explicit training set for an AI/ML algorithm is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240279

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-01

    (90 days)

    Product Code
    Regulation Number
    866.5830
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VIDAS® TBI (GFAP, UCH-L1) test is composed of two automated assays - VIDAS® TBI (GFAP) and VIDAS® TBI (UCH-L1) - to be used on the VIDAS® 3 instrument for the quantitative measurement of Glial Fibrillary Acidic Protein (GFAP) and Ubiquitin C-terminal Hydrolase (UCH-L1) in human serum using the ELFA (Enzyme Linked Fluorescent Assay) technique. The results of both assays are requred to obtain an overall qualitative test interpretation.

    The overall qualitative VIDAS® TBI (GFAP, UCH-L1) test result is used, in conjunction with clinical information, to aid in the evaluation of patients (18 years of age or older), presenting within 12 hours of suspected mild traumatic brain injury (Glasgow Coma Scale score 13-15), to assist in determining the need for a Computed Tomography (CT) scan of the head. A negative interpretation of VIDAS® TBI (GFAP, UCH-L1) test is associated with the absence of acute intracranial lesions visualized on a head CT scan.

    Device Description

    The VIDAS® TBI (GFAP, UCH-L1) test is composed of two automated assays – VIDAS® TBI (GFAP) and VIDAS® TBI (UCH-L1) – to be used on the VIDAS® 3 instrument. Similar to other VIDAS assays, VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) test kits (specific to each biomarker) contain the solid phase receptacles (SPRs®), the reagent strips, Product Calibrator S1 and Product Control C1. These test kits will also contain the master lot entry (MLE) data i.e., a barcode printed on the outer label of the packaging, as well as the reference number of the package insert to download from the bioMérieux website.

    Whether it be for the GFAP or UCH-L1 quantification, the test combines a three-step enzyme immunoassay sandwich method with a final fluorescent detection step, also known as enzyme-linked fluorescent assay (ELFA).

    The Solid Phase Receptacle (SPR) serves as the solid phase as well as the pipetting device. The inner surface of the SPR is coated with antibodies aqainst the substance of interest i.e., anti-GFAP or anti-UCH-L1 antibodies. The reagent strip consists of 10 wells covered with a labeled foil seal. Well 1 is designated for the sample. Eight of the wells contain sample diluent, wash buffer, conjugate, and tracer. The last well contains the fluorescent substrate. All of the assay steps are performed automatically by the instrument.

    The intensity of the fluorescence is proportional to the concentration of the analyte the sample. At the end of the assay, the biomarker concentration is automatically calculated by the instrument in relation to the calibration curve and stored in the Master Lot Entry (MLE) data.

    VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) results are reported separately: the VIDAS 3 reports the calculated concentration and the qualitative interpretation for each. The final result i.e., the patient's status in relation to suspected mild traumatic brain injury, must be interpreted by the user according to the decision tree presented in the package insert.

    AI/ML Overview

    This document describes the validation of the VIDAS® TBI (GFAP, UCH-L1) test, an automated assay for diagnosing mild traumatic brain injury. The submission compares the device to a predicate device, the BANYAN BTI™, and summarizes non-clinical and clinical testing results. The following points address the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for each performance metric in a table format. However, it presents the results of various assays and often implies that the results "demonstrate" or "confirm" the required performance, indicating these are the achieved results compared to an internal standard or regulatory expectation. Below is a table summarizing various performance metrics and their reported results. Specific acceptance criteria values are not provided in this public summary.

    Performance MetricReported Device Performance
    Analytical Measuring Interval
    VIDAS TBI (GFAP)10.0 - 320.0 pg/mL
    VIDAS TBI (UCH-L1)80.0 - 2560.0 pg/mL
    Linearity
    VIDAS TBI (GFAP)Demonstrated on the range 6.7 - 354.5 pg/mL
    VIDAS TBI (UCH-L1)Demonstrated on the range 58.9 - 2769.1 pg/mL
    Detection Limits
    Limit of Blank (LoB) - GFAP4.4 pg/mL
    Limit of Detection (LoD) - GFAP5.4 pg/mL
    Limit of Quantitation (LoQ) - GFAP5.4 pg/mL
    Limit of Blank (LoB) - UCH-L141.8 pg/mL
    Limit of Detection (LoD) - UCH-L148.1 pg/mL
    Limit of Quantitation (LoQ) - UCH-L148.1 pg/mL
    Hook Effect
    VIDAS TBI (GFAP)No hook effect up to 200,000.0 pg/mL
    VIDAS TBI (UCH-L1)No hook effect up to 400,000.0 pg/mL
    Calibration FrequencyVerified for 56 days
    Sample StabilityVerified for specified storage conditions and freeze/thaw cycles
    Diagnostic Accuracy
    Diagnostic Sensitivity96.7%
    Diagnostic Specificity41.2%
    Positive Likelihood Ratio1.6
    Negative Likelihood Ratio0.1
    Positive Predictive Value9.9%
    Negative Predictive Value99.5%

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Test set sample size: For the diagnostic accuracy study, the sample size is not explicitly stated but refers to the "ALERT cohort." For the reference interval study, 513 apparently healthy US adult subjects were used.
    • Data provenance: The diagnostic accuracy study was performed using the "ALERT cohort." The reference interval study was conducted at three sites (one internal European site and two external US sites). It is not specified whether these studies were retrospective or prospective, though "ALERT cohort" could suggest a pre-existing dataset.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided in the document. The diagnostic accuracy study compares the device's results to the presence/absence of acute intracranial lesions visualized on a head CT scan, but the number or qualifications of experts interpreting these CT scans to establish ground truth are not mentioned.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in vitro diagnostic test for quantitative measurement of biomarkers, not an AI-assisted imaging device that impacts human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the diagnostic accuracy study presents the standalone performance of the VIDAS® TBI (GFAP, UCH-L1) assay. The results (sensitivity, specificity, etc.) are based on the device's output compared to the ground truth (CT scan findings). The device is used "in conjunction with clinical information," but the reported diagnostic accuracy figures are for the test itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the diagnostic accuracy study was "absence of acute intracranial lesions visualized on a head CT scan." This indicates that CT scan results were used as the reference standard for traumatic brain injury assessment.

    8. The sample size for the training set

    This document describes a diagnostic device and its validation. It does not explicitly mention a "training set" in the context of machine learning or AI models with distinct training and test phases. The "test set" for diagnostic accuracy is referred to as the "ALERT cohort." The reference interval was established using 513 apparently healthy subjects.

    9. How the ground truth for the training set was established

    As there is no explicitly defined "training set" for an AI model in this submission, the method for establishing ground truth for a training set is not applicable or described. The clinical performance data presented (Diagnostic Accuracy and Reference interval) seems to represent the evaluation of the final device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151817

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-02-18

    (227 days)

    Product Code
    Regulation Number
    866.1640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VITEK® 2 AST-YS Caspofungin is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Caspofungin is a quantitative test. Caspofungin has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifyingal.

    Active in vitro and clinical infections: Candida albicans Candida glabrata Candida guilliermondii Candida krusei Candida parapsilosis Candida tropicalis

    The VITEK @ 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast.

    Device Description

    The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

    The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Criteria CategoryAcceptance Criteria (typically from FDA guidance)Reported Device Performance (VITEK® 2 AST-YS Caspofungin)
    Overall Essential AgreementNot explicitly stated in the provided text as a numerical target, but implied to be high based on FDA Class II Special Controls Guidance Document.99.7%
    Overall Category AgreementNot explicitly stated in the provided text as a numerical target, but implied to be high based on FDA Class II Special Controls Guidance Document.99.8%
    ReproducibilityAcceptable performance (as per FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA)Acceptable results
    Quality ControlAcceptable performance (as per FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA)Acceptable results
    Substantial EquivalencePerformance comparable to a legally marketed predicate device and CLSI broth microdilution reference method.Demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method.
    Active MicroorganismsActive against specific Candida species according to FDA label for the antifungal.Confirmed for: Candida albicans, Candida glabrata, Candida guilliermondii, Candida krusei, Candida parapsilosis, Candida tropicalis.

    Note: The specific numerical acceptance criteria for Essential Agreement and Category Agreement are often found in FDA guidance documents for AST systems, which are referenced (FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, issued August 28, 2009). While the document states that the device demonstrated "acceptable performance," it does not explicitly list the numerical thresholds from that guidance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a total number. The study utilized "fresh and stock clinical isolates, as well as a set of challenge strains." The document does not specify the exact number of isolates for each category or the total.
    • Data Provenance: The document indicates the use of "clinical isolates" (both fresh and stock) and "challenge strains." The country of origin is not specified, but the context implies data collected for regulatory submission within the U.S.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • The document does not mention the use of experts to establish ground truth for the test set.
    • Instead, the ground truth was established by a reference method: the CLSI broth microdilution reference method.

    4. Adjudication Method for the Test Set

    • None. The ground truth was established by comparison to a standardized reference method (CLSI broth microdilution), not through expert adjudication of images or clinical cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. An MRMC comparative effectiveness study was not conducted as this device is an automated in vitro diagnostic for antimicrobial susceptibility testing, not a system involving human readers interpreting clinical cases with or without AI assistance. The performance is compared to a reference laboratory method.

    6. If a Standalone Performance Study Was Done

    • Yes. The study described is a standalone performance evaluation of the VITEK® 2 AST-YS Caspofungin system. It compares the results generated by the VITEK® 2 system (algorithm-driven interpretation of growth patterns) directly against the CLSI broth microdilution reference method. This is an "algorithm only" type of performance study for a diagnostic device.

    7. The Type of Ground Truth Used

    • Reference Method (CLSI Broth Microdilution): The ground truth was established by the CLSI broth microdilution reference method, incubated at 24 hours.

    8. The Sample Size for the Training Set

    • The document does not specify a separate "training set" or its sample size. For in vitro diagnostic devices like this, the development likely involves internal studies and database building, but the submission specifically focuses on the external evaluation data as the primary validation.

    9. How the Ground Truth for the Training Set Was Established

    • Since a training set is not explicitly mentioned or detailed in the document, how its ground truth was established is not provided. If a training set was used during the device's development, it would typically involve similar reference methods (like CLSI broth microdilution) to establish the correct susceptibility profiles the system is designed to learn from or emulate.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1