Search Results
Found 205 results
510(k) Data Aggregation
(56 days)
VITEK 2 AST-Streptococcus Cefuroxime (≤0.125 - ≥8) is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
VITEK 2 AST-Streptococcus Cefuroxime (≤0.125 - ≥8) is a quantitative test. Testing is indicated for Streptococcus pneumoniae as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).
VITEK 2 AST-Streptococcus Cefuroxime (≤0.125 - ≥8) has demonstrated acceptable performance with the following organism:
Streptococcus pneumoniae
The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus and Viridans Streptococcus to antimicrobial agents when used as instructed.
The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh (1) and Gerlach (2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (3).
Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 – 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK® 2 AST-Streptococcus Cefuroxime has the following concentrations in the card: 0.25, 0.5, 1, and 2 (equivalent standard method concentration by efficacy in µg/mL).
N/A
Ask a specific question about this device
(90 days)
VITEK 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) is a quantitative test. Testing is indicated for Enterobacterales (from infections other than uncomplicated UTI) as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).
VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) has demonstrated acceptable performance with the following organisms:
Enterobacterales (Escherichia coli, Proteus mirabilis, Klebsiella pneumoniae, Klebsiella oxytoca, Citrobacter koseri)
The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram- negative bacilli to antimicrobial agents when used as instructed.
The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh (1) and Gerlach (2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (3).
Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 – 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK® 2 AST-GN Cefazolin has the following concentrations in the card: 1, 2, and 8 (equivalent standard method concentration by efficacy in µg/mL).
The provided text describes the acceptance criteria and a study proving the device meets these criteria for the VITEK 2 AST-Gram Negative Cefazolin antimicrobial susceptibility testing system.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter | Acceptance Criteria (Generally Implied for AST Systems based on FDA Guidance) | Reported Device Performance (VITEK® 2 AST-GN Cefazolin) |
|---|---|---|
| Essential Agreement (EA) | Typically ≥ 90% (agreement between the MIC from the test device and the reference method, within a +/- 1 doubling dilution) | 97.5% (840/862) |
| Category Agreement (CA) | Typically ≥ 90% (agreement between the interpretive category generated by the test device and the reference method) | 86.8% (748/862) |
| Very Major Errors (VME) | As low as possible; specific limits usually apply (e.g., <1.5% and NME ≤30%) | 0.3% (1/299) |
| Major Errors (ME) | As low as possible; specific limits usually apply (e.g., <3.0% and NME ≤30%) | 0.7% (3/414) |
| Minor Errors (mE) | Not explicitly stated as an acceptance criterion percentage, but high mE often impacts CA. | 12.8% (110/862) |
| Reproducibility | Typically ≥ 95% | 100% |
Note on Acceptance Criteria: The document references the "FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)" for defining acceptable performance. While specific numerical acceptance criteria (e.g., for EA, CA, VME, ME) are not explicitly listed in the provided text, they are implied by the reported performance metrics and the statement that the device "demonstrated substantially equivalent performance." For AST systems, the FDA typically looks for high Essential and Category Agreement, with very low rates of Major and Very Major Errors. The low Category Agreement (86.8%) is acknowledged and attributed mainly to minor errors.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set:
- Total isolates tested: 862 (derived from the denominator for EA and CA calculations).
- The study included "fresh and stock clinical isolates" as well as "a set of challenge strains."
- Data Provenance:
- The study was an "external evaluation." This typically implies data collected from multiple sites outside of the manufacturer's primary lab.
- The document does not specify the country of origin for the data.
- The study included both "fresh" and "stock" clinical isolates, suggesting a mix of prospective (fresh isolates) and retrospective (stock isolates) data collection.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not specify the number or qualifications of experts used to establish the ground truth.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method for the test set. For AST devices, the "ground truth" is typically established by a reference method (broth microdilution in this case), not by expert consensus or adjudication in the way it might be for an imaging AI device.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Improvement
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed. This type of study (human readers assisting vs. not assisting) is typically relevant for interpretative diagnostic devices, especially in imaging, where human interpretation is a key part of the workflow. The VITEK 2 AST system is an automated device for determining antimicrobial susceptibility, not directly for human interpretation or image reading.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the primary performance study presented is a standalone (algorithm only) performance study. The device's performance (VITEK 2 AST-GN Cefazolin) was compared directly against the CLSI broth microdilution reference method. There is no mention of human-in-the-loop performance evaluation, as the VITEK 2 system is automated.
7. The Type of Ground Truth Used
The type of ground truth used was: Reference Method Comparison.
Specifically, the performance of the VITEK 2 AST-GN Cefazolin was compared to the CLSI broth microdilution reference method, incubated for 16-20 hours. This is the gold standard for antimicrobial susceptibility testing.
8. The Sample Size for the Training Set
The document does not specify the sample size for the training set. This submission is for a 510(k) clearance, which typically focuses on the performance of the final device rather than the specifics of its development (like training data for AI/ML models). The VITEK 2 system relies on growth pattern analysis algorithms rather than deep learning models that require large, labeled training datasets in the conventional sense. Its "training" would likely involve optimizing these algorithms based on extensive internal validation.
9. How the Ground Truth for the Training Set Was Established
The document does not describe how the ground truth for the training set was established, as details on the development of the device's analysis algorithms (likely proprietary) are not typically included in a 510(k) summary. Given the nature of AST systems, any "ground truth" used during development would also likely stem from comparisons to established reference methods like broth microdilution, similar to the test set's ground truth.
Ask a specific question about this device
(90 days)
ETEST® is a manual, quantitative technique for the determination of antimicrobial susceptibility of non fastidious Gram negative and Gram positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in μg/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.
Testing with ETEST Imipenem/Relebactam P. aeruginosa (IRPA) (0.008/4-128/4 µg/mL) is indicated for Pseudomonas aeruginosa, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).
The ETEST Imipenem/Relebactam P. aeruginosa (IRPA) (0.008/4-128/4 µg/mL) demonstrated acceptable performance with the following microorganism:
• Pseudomonas aeruginosa
ETEST® is a thin, inert and non-porous plastic strip carrying the MIC reading scale in μg/mL on one side and a predefined antibiotic gradient on the other side.
When the strip is applied to an inoculated agar surface, the preformed antibiotic gradient immediately transfers into the agar matrix, then forming a stable, continuous and exponential gradient of antibiotic concentrations directly underneath the strip. Bacterial growth becomes visible during incubation, and a symmetrical inhibition ellipse centered along the strip appears. The MIC value is read from the scale in terms of μg/mL at complete inhibition of bacterial growth, where the pointed end of the ellipse intersects the strip.
ETEST Imipenem/Relebactam P. aeruginosa (IRPA) with a concentration range of 0.008/4-128/4 µg/mL is specially designed and formulated for testing P. aeruginosa.
This document describes the regulatory clearance for the ETEST® Imipenem/Relebactam P. aeruginosa (IRPA) device, which is an antimicrobial susceptibility test. Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are defined by established performance metrics in antimicrobial susceptibility testing, specifically Essential Agreement (EA) and Category Agreement (CA), compared to a reference method. The reported performance demonstrates the device meets these criteria.
| Metric | Acceptance Criteria (Implicit) | Reported Device Performance (Pseudomonas aeruginosa) |
|---|---|---|
| Essential Agreement (EA) | Not explicitly stated, but generally expected to be high (e.g., typically >90% or 95% for AST devices). | 97.7% |
| Category Agreement (CA) | Not explicitly stated, but generally expected to be high (e.g., typically >90% or 95% for AST devices). | 91.1% |
| Reproducibility | Best case: >95% (as per predicate device) | Best case: 99.6% |
| Reproducibility | Worst case: >95% (as per predicate device) | Worst case: 99.6% |
| Quality Control | Results within range >95% (as per predicate device) | Results within range >95% of the times tested |
| Meets Guidance Document Performance Requirements | Yes (explicitly stated) | Yes |
Study Details Proving Device Meets Acceptance Criteria
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 437 strains of Pseudomonas aeruginosa.
- Data Provenance: The study involved external evaluations conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The specific country of origin is not mentioned, but "external evaluations" imply samples from clinical settings. The nature of "fresh" and "stock clinical isolates" suggests a mix of retrospective and prospective data, where fresh isolates would be prospective and stock isolates could be retrospective.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- This information is not explicitly provided in the document. Antimicrobial Susceptibility Testing (AST) generally relies on established laboratory methods rather than expert consensus for ground truth, but the personnel performing the reference method would be trained laboratory technicians.
-
Adjudication Method for the Test Set:
- None explicitly stated as it's a comparison to a reference laboratory method (CLSI M07-11th Ed broth microdilution). Discrepancy resolution (adjudication) is typically less relevant for quantitative laboratory assays where results are directly compared to a gold standard.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, a MRMC study was not done. This is an automated/manual laboratory test for determining Minimum Inhibitory Concentration (MIC), not an imaging device requiring human reader interpretation in a diagnostic workflow. The "human-in-the-loop" aspect for this device is the application of the strip and reading the MIC, which is part of the validation for the device's accuracy against a gold standard method.
-
If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This concept does not directly apply to this type of device. The ETEST device itself is a "standalone" test. The performance reported (EA and CA) is the performance of the ETEST system (strip + reading methodology) compared to the reference broth microdilution method. It inherently involves a human reading the result by interpreting the inhibition ellipse on the strip. The note "b) In the ETEST® Imipenem/Relebactam clinical studies, swabs were used for plate inoculation/streaking and forceps were used for ETEST® strip application. Testing with the optional Inoculator RETRO C80™, Vacuum Pen NEMA C88™ and Applicator SIMPLEX C76™ was not evaluated during the clinical studies" suggests that human manual steps were involved in the performance evaluation.
-
The Type of Ground Truth Used:
- Reference Method: The ground truth was established by the CLSI M07-11th Ed (January 2018) broth microdilution reference method, following specifications as defined in CLSI M100 34th Ed. (February 2024). This is considered the gold standard for antimicrobial susceptibility testing. Outcomes data or pathology are not relevant for this type of in vitro diagnostic device.
-
The Sample Size for the Training Set:
- The document does not mention a training set in the context of machine learning or AI models. This device is a manual, quantitative laboratory test (an ETEST strip system) and not an AI/ML-driven diagnostic. Therefore, there's no "training set" in the typical sense of AI development. The "strains" mentioned are for the performance evaluation against the reference method.
-
How the Ground Truth for the Training Set was Established:
- As there is no mention of a training set for an AI/ML model, this question is not applicable. The device's performance is validated against a well-established reference laboratory method using a test set of clinical isolates and challenge strains.
Ask a specific question about this device
(451 days)
The VITEK® COMPACT PRO is intended for the automated quantitative and/or qualitative antimicrobial susceptibility testing of isolated colonies for most clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and yeast.
The VITEK® COMPACT PRO is also intended for the automated identification of most clinically significant anaerobic organisms and Corynebacterium species, fermenting and nonfermenting Gram-negative bacilli, Gram-positive organisms, fastidious organisms, and yeasts and yeast-like organisms.
The VITEK® COMPACT PRO instrument is an automated instrument designed for use in low-to medium-range applications in both Clinical and Industry laboratories. The instrument performs sample well filling, incubation, and optical readings. The VITEK® COMPACT PRO instrument is a two-step automated instrument for:
- Hydrating reagents with sample inoculum
- Pre-processing cards, incubating cards, and continuous reading for growth
The VITEK® 2 Systems Software receives the instrument optical readings and performs analysis. The instrument then ejects the completed reagent card into the waste area for disposal.
The system includes a VITEK® COMPACT PRO instrument with an internal computer, monitor, keyboard, mouse, handheld barcode scanner, and USB hub. The software provided with the internal computer includes analysis and limited data management programs. A bidirectional computer interface (BCI) may transfer results automatically to the user's laboratory information system (LIS).
A Quality Control System is available to track the quality control results of the test cards. The Advanced Expert System™ (Clinical Use) is available to provide online, systematic validation of results and interpretation of resistant phenotypes found during susceptibility testing.
The provided text describes the regulatory clearance of a medical device, the VITEK® COMPACT PRO, and its performance evaluation. However, it does not explicitly define "acceptance criteria" in a table format with specific numerical targets. Instead, it details the results of various performance tests as evidence that the device meets acceptable standards, primarily by demonstrating substantial equivalence to a predicate device.
Here's an interpretation of the acceptance criteria and study that can be extracted from the provided information:
Interpretation of Acceptance Criteria and Study Design:
The VITEK® COMPACT PRO is an automated antimicrobial susceptibility testing system. The core of its acceptance criteria and the study proving it meets these criteria relies on demonstrating substantial equivalence to a previously cleared VITEK® 2 System (N50510 Supplement S082). This means the new device must perform comparably to the predicate device across various metrics.
1. Table of Acceptance Criteria (Inferred) and Reported Device Performance:
Since explicit acceptance criteria are not tabulated with thresholds, they are inferred from the reported performance, with the implicit criterion being "comparable to or better than the predicate device."
| Performance Metric | Inferred Acceptance Criterion (vs. Predicate VITEK® 2) | Reported Device Performance (VITEK® COMPACT PRO) |
|---|---|---|
| Quality Control (QC) | Acceptable QC results within established ranges | QC passing >99% for each of the AST card classes tested (GN, GP, YS, ST) for 39 antimicrobial agents and 13 QC organisms. |
| Reproducibility | >95% Reproducibility (within 1 doubling dilution) | >97% reproducibility in the best case and >95% reproducibility in the worst case for all VITEK® 2 AST test cards/antimicrobials. |
| Clinical Performance (Accuracy) | High Essential Agreement (EA) with predicate | AST GN cards: 98.8% EA |
| (Essential Agreement - EA) | High Categorical Agreement (CA) with predicate | AST GP cards: 99.5% EA |
| Clinical Performance (Accuracy) | AST ST cards: 98.5% EA | |
| (Categorical Agreement - CA) | AST YS cards: 100% EA | |
| AST GN cards: 97.3% CA | ||
| AST GP cards: 97.4% CA | ||
| AST ST cards: 98.7% CA | ||
| AST YS cards: 100% CA | ||
| Usability/Safety (Human Factors) | Safe and effective for intended users/uses | Overall favorable with respect to usability. Minor usability errors mostly due to workflow changes and risk of ignoring warnings, but with appropriate mitigations, found safe and effective. |
| System Verification (Technical) | Successful completion and passing of requirements | Successfully completed for instrument, software/firmware interfaces (VITEK® COMPACT PRO Instrument, VITEK® firmware, VITEK® Systems 10.0 software, VITEK® COMPACT PRO User Interface, VITEK Systems Communications protocols). |
2. Sample Sizes Used for the Test Set and Data Provenance:
- QC Testing: Each of the 13 quality control organisms was tested at least twenty times. This testing was conducted at "at least three of the clinical evaluation trial sites."
- Reproducibility: Isolates were tested in triplicate for three trial sites. Individual test sets were selected for each antimicrobial, with each set including a minimum of 10 strains (at least 2 with on-scale results).
- Clinical Performance Evaluation: "Representative, clinically relevant strains" were tested. Each strain was tested one time with both the VITEK® COMPACT PRO and the VITEK® 2.
- Data Provenance: The document states testing was done at "at least three of the clinical evaluation trial sites" for QC and "three trial sites" for reproducibility. The country of origin for these sites is not specified, but bioMérieux Inc. is located in Hazelwood, Missouri, USA, implying the studies were likely conducted in the US. The studies appear to be prospective as they involve new testing of isolates for performance evaluation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
This information is not explicitly stated in the provided text. The ground truth for the clinical performance evaluation was established by comparing the VITEK® COMPACT PRO's results against the VITEK® 2 System's results (N50510 Supplement S082), which implicitly serves as the "expert consensus" or "reference method" in this context. The document does not describe a separate human expert panel for adjudication or ground truth establishment the way it might for an AI imaging device.
4. Adjudication Method for the Test Set:
Not applicable in the traditional sense of human expert adjudication. The comparison was directly between the new device (VITEK® COMPACT PRO) and the predicate device (VITEK® 2 System) for MICs and interpretations (S/SDD/I/R). There's no mention of a 2+1 or 3+1 human expert adjudication process.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC study was not done. This device is an automated in vitro diagnostic instrument, not an AI-assisted imaging device for human interpretation. The comparison is machine-to-machine (new instrument vs. predicate instrument).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Yes, the primary performance evaluation was a standalone "algorithm only" (instrument only) performance comparison. The VITEK® COMPACT PRO (device under test) generated MICs and interpretations, and these were directly compared to those generated by the predicate VITEK® 2 System. Human "in-the-loop" performance is not relevant for the core function of this automated susceptibility system in the context of its performance study.
7. The type of ground truth used:
The ground truth for the clinical performance evaluation was the MIC results and categorical interpretations (S/SDD/I/R) obtained from the legally marketed predicate device, the VITEK® 2 System (N50510 Supplement S082). This serves as the reference method against which the new device's performance is measured.
8. The Sample Size for the Training Set:
The document does not discuss a separate "training set" in the context of device development or performance evaluation. This is not an AI/ML device where a distinct training dataset is typically used. The development process likely involved internal verification and validation, but the reported studies (QC, Reproducibility, Clinical Performance) are essentially testing/validation datasets.
9. How the ground truth for the training set was established:
Not applicable, as no explicit training set for an AI/ML algorithm is described.
Ask a specific question about this device
(265 days)
VITEK 2 AST-Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. VITEK 2 AST-Yeast Voriconazole is a quantitative test. Voriconazole has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antifungal.
Active in vitro and in clinical infections: Candida krusei Candida parapsilosis Candida tropicalis
The VITEK 2 Fungal Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.
The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique. The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique. Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45-0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK® 2 AST-YS Voriconazole has the following concentrations in the card: 0.03125, 0.125, 0.25, 1, and 2 (equivalent standard method concentration by efficacy in ug/mL).
The provided FDA 510(k) summary describes the VITEK® 2 AST-Yeast Voriconazole device, which is an antimicrobial susceptibility test system.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
The acceptance criteria are implicitly derived from the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems and are presented through "Essential Agreement" (EA) and "Category Agreement" (CA) performance metrics, along with error rates. The table below summarizes the reported performance for each microorganism. The specific acceptance thresholds for EA and CA are not explicitly stated as numerical percentages (e.g., >90% EA, >90% CA) with the exception of the individual error types (VME, ME, mE) having specified maximum allowable percentages. However, it is implied that the presented results were deemed acceptable by the FDA.
| Antimicrobial | Microorganism | Essential Agreement % | VME (%) | ME (%) | mE (%) | Category Agreement % | VME (%) (Cat Ag) | ME (%) (Cat Ag) | mE (%) (Cat Ag) | % Reproducibility |
|---|---|---|---|---|---|---|---|---|---|---|
| Voriconazole | C. albicans | 88.7 (228/257) | N/A | N/A | N/A | 86.0 (221/257) | 33.3 (4/12) | 1.3 (3/237) | 11.3 (29/257) | 100 |
| Voriconazole | C. krusei | 100.0 (76/76) | N/A | N/A | N/A | 98.7 (75/76) | 0.0 (0/0) | 0.0 (0/74) | 1.3 (1/76) | Not Specified |
| Voriconazole | C. parapsilosis | 94.6 (70/74) | N/A | N/A | N/A | 94.6 (70/74) | 0.0 (0/1) | 1.5 (1/68) | 4.1 (3/74) | Not Specified |
| Voriconazole | C. tropicalis | 97.7 (85/87) | N/A | N/A | N/A | 90.8 (79/87) | 0.0 (0/2) | 2.6 (2/77) | 6.9 (6/87) | Not Specified |
Key Definitions:
- Essential Agreement (EA): Agreement between the MIC results of the test device and the reference method within plus or minus one doubling dilution.
- Category Agreement (CA): Agreement between the interpretive categories (Susceptible, Intermediate, Resistant) of the test device and the reference method.
- Very Major Error (VME): The test device reports susceptible, but the reference method reports resistant.
- Major Error (ME): The test device reports resistant, but the reference method reports susceptible.
- Minor Error (mE): Any other disagreement in interpretive category (e.g., susceptible vs. intermediate, resistant vs. intermediate).
2. Sample Size Used for the Test Set and Data Provenance:
The sample sizes for the test set (external evaluation) are provided for each microorganism:
- C. albicans: 257 isolates
- C. krusei: 76 isolates
- C. parapsilosis: 74 isolates
- C. tropicalis: 87 isolates
Data Provenance: The study conducted an "external evaluation" with "fresh and stock clinical isolates, as well as a set of challenge strains." The document does not specify the country of origin of the data nor explicitly state if it was retrospective or prospective. However, "fresh clinical isolates" typically implies a prospective collection, while "stock clinical isolates" could be either.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
The ground truth was established by the "CLSI broth microdilution reference method." This is a standardized laboratory method, not reliant on human expert adjudication in the same way an imaging study would be. Therefore, the concept of "experts" and their qualifications as typically applied to visual diagnostic tasks (e.g., radiologists) is not directly applicable here. The ground truth method itself (CLSI broth microdilution) is the expert-defined standard.
4. Adjudication Method for the Test Set:
Not applicable in the human expert sense. The "adjudication method" for determining the ground truth in this context is the CLSI broth microdilution reference method, which serves as the gold standard for antimicrobial susceptibility testing. The device's results are compared directly against this established reference.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, often in conjunction with AI. The VITEK® 2 AST-Yeast Voriconazole is an automated in vitro diagnostic device for antimicrobial susceptibility testing; its performance is evaluated by comparing its automated output to a gold standard laboratory method, not by how it assists human readers.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
Yes, a standalone performance evaluation was done. The VITEK® 2 AST-Yeast Voriconazole system is an automated device designed to determine antimicrobial susceptibility without human interpretation of the primary data once the sample is loaded. The performance metrics (Essential Agreement, Category Agreement, error rates) directly reflect the device's standalone capability compared to the reference method.
7. The Type of Ground Truth Used:
The ground truth used was the CLSI broth microdilution reference method, incubated for 24 hours (up to 48 hours for slowly growing isolates). This is a well-established and scientifically accepted standard in microbiology for determining minimum inhibitory concentrations (MICs) of antifungal agents.
8. The Sample Size for the Training Set:
The document does not explicitly state the sample size for the training set. The descriptions focus on the "external evaluation" (test set). For IVD devices like this, the 'training set' often corresponds to historical data, internal studies, and method development efforts that lead to the final algorithm and concentration ranges. Without further information, the exact size of the training set used to develop the VITEK® 2 AST-Yeast Voriconazole algorithm is not provided in this 510(k) summary.
9. How the Ground Truth for the Training Set Was Established:
Similar to the answer for point 8, the document does not explicitly describe how the ground truth for any potential training set was established. However, given that the final performance is benchmarked against the CLSI broth microdilution reference method, it is highly probable that any internal development or training data would also have used this or a similar established reference method to determine the true susceptibility of isolates.
Ask a specific question about this device
(260 days)
VITEK 2 AST-Yeast Anidulafungin is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. VITEK 2 AST-Yeast Anidulafungin is a quantitative test. Anidulafungin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antifungal.
Active in vitro and in clinical infections: Candida albicans Candida glabrata Candida parapsilosis Candida tropicalis
In vitro data are available, but clinical significance is unknown: Candida guillermondii Candida krusei
The VITEK 2 Fungal Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.
The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh (1) and Gerlach(2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (3).
Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 – 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK® 2 AST-YS Anidulafungin has the following concentrations in the card: 0.0625, 0.125, 0.5, 2, and 8 (equivalent standard method concentration by efficacy in ug/mL).
Here's an analysis of the acceptance criteria and the study proving the device meets those criteria, based on the provided text.
Device: VITEK 2 AST-Yeast Anidulafungin
Indications for Use: Antifungal susceptibility testing of Candida species (C. albicans, C. glabrata, C. parapsilosis, C. tropicalis, C. guilliermondii, C. krusei) as a laboratory aid in determining in vitro susceptibility to antifungal agents.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document doesn't explicitly state the acceptance criteria in a separate table directly defining thresholds for Essential Agreement (EA), Category Agreement (CA), or error rates (VME, ME, mE) that the device must meet for approval. Instead, it presents the results of the performance study and implies that these results were deemed "acceptable" by the FDA. The performance is compared to the "CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)". This guidance document would contain the specific acceptance criteria.
However, based on the presented "Performance Overview" (Page 7) and the overall context of AST device approvals, typical acceptance criteria for Essential Agreement and Category Agreement are usually in the range of 90-95% or higher, with Very Major Error (VME) and Major Error (ME) rates usually being low (e.g., <3% or <1.5%).
Here's the reported device performance:
| Antimicrobial | Species | Essential Agreement (% EA) | Category Agreement (% CA) | VME (%) | ME (%) | mE (%) | Reproducibility (%) |
|---|---|---|---|---|---|---|---|
| Anidulafungin | C. albicans | 100% (206/206) | 100% (206/206) | 0.0% (0/0) | 0.0% (0/206) | 0.0% (0/206) | 98.15% |
| C. glabrata | 97.4% (113/116) | 98.3% (114/116) | 0.0% (0/4) | 0.9% (1/111) | 0.9% (1/116) | ||
| C. guilliermondii | 90.5% (19/21) | 90.5% (19/21) | 0.0% (0/0) | 0.0% (0/20) | 9.5% (2/21) | ||
| C. krusei | 98.6% (68/69) | 100% (69/69) | 0.0% (0/0) | 0.0% (0/69) | 0.0% (0/69) | ||
| C. parapsilosis | 100% (72/72) | 86.1% (62/72) | 0.0% (0/0) | 1.6% (1/62) | 12.5% (9/72) | ||
| C. tropicalis | 97.5% (79/81) | 96.3% (78/81) | 33.3% (1/3) | 0.0% (0/77) | 2.5% (2/81) |
Note on C. tropicalis VME: The document specifically highlights: "When evaluating VITEK 2 AST-YS Anidulafungin, there was a single very major error (VMJ) that resulted in an unacceptable VMJ rate of 33.3% (1/3) with C. tropicalis." Despite this, the overall conclusion states "VITEK® 2 AST-YS Anidulafungin demonstrated acceptable performance." This suggests that either the unacceptably high VME for C. tropicalis was mitigated by other factors (e.g., small sample size for VME calculation, or specific post-market surveillance requirements), or it points to a known limitation that the FDA found acceptable for clearance under the stated conditions.
2. Sample Size Used for the Test Set and Data Provenance
-
Sample Size for Test Set:
- C. albicans: 206 isolates
- C. glabrata: 116 isolates
- C. guilliermondii: 21 isolates
- C. krusei: 69 isolates
- C. parapsilosis: 72 isolates
- C. tropicalis: 81 isolates
- Total Isolates: 565 isolates across the specified Candida species.
-
Data Provenance: The study involved an "external evaluation" conducted with "fresh and stock clinical isolates, as well as a set of challenge strains." The document does not specify the country of origin of the data or whether it was retrospective or prospective. Typically, "clinical isolates" imply prospective collection from real patient samples, and "stock isolates" could be reference strains or previously collected clinical isolates. "External evaluation" implies data was collected from multiple sites, which is standard for clinical trials of this nature.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
This information is not provided in the document. For antimicrobial susceptibility testing (AST) devices, the ground truth is established by a reference method (here, the CLSI broth microdilution reference method), not by human expert interpretation of images or data in a consensus setting.
4. Adjudication Method for the Test Set
Not applicable. The ground truth for AST devices is established by a standardized laboratory reference method (CLSI broth microdilution), not through expert adjudication of ambiguous cases like in imaging studies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not performed. This type of study (comparing human readers with and without AI assistance on diagnostic tasks) is not relevant for an automated antimicrobial susceptibility testing device like the VITEK 2 AST-Yeast Anidulafungin. This device provides quantitative MIC values and interpretive categories, not an interpretation of a medical image or clinical data that would involve human readers.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the performance study effectively evaluates the standalone performance of the VITEK 2 AST-Yeast Anidulafungin system. The device automatically reads and interprets the growth patterns in the microdilution wells to generate MIC values and interpretive categories (Susceptible, Intermediate, Resistant). This automated result is then compared directly to the established ground truth (CLSI broth microdilution reference method). While human operators physically load the cards, the core performance being evaluated is the device's automated analysis.
7. The Type of Ground Truth Used
The type of ground truth used was the CLSI broth microdilution reference method (Clinical and Laboratory Standards Institute). This is a well-established, standardized, and validated laboratory method for determining the minimum inhibitory concentration (MIC) of an antimicrobial agent against a microorganism. The reference method results serve as the "gold standard" for accuracy in AST device performance studies.
8. The Sample Size for the Training Set
The document does not provide specific details about the training set size. The device uses "Discriminant Analysis" algorithms. Typically, for a device like this, the algorithms would be developed and refined using a large dataset of isolates with known reference MICs to optimize the mapping between the device's optical measurements and the true MICs. However, the exact size of this internal development/training set is not disclosed in this summary. The provided performance data (the "test set") is for validation purposes, distinct from the training set.
9. How the Ground Truth for the Training Set Was Established
While not explicitly stated for a "training set" in this document, it is highly probable that the ground truth for any isolates used during the development and training of the VITEK 2 system's algorithms would also have been established using the CLSI broth microdilution reference method or similar highly standardized and validated laboratory methods, consistent with industry best practices for AST device development. This ensures that the algorithm learns from accurate data.
Ask a specific question about this device
(90 days)
The VIDAS® TBI (GFAP, UCH-L1) test is composed of two automated assays - VIDAS® TBI (GFAP) and VIDAS® TBI (UCH-L1) - to be used on the VIDAS® 3 instrument for the quantitative measurement of Glial Fibrillary Acidic Protein (GFAP) and Ubiquitin C-terminal Hydrolase (UCH-L1) in human serum using the ELFA (Enzyme Linked Fluorescent Assay) technique. The results of both assays are requred to obtain an overall qualitative test interpretation.
The overall qualitative VIDAS® TBI (GFAP, UCH-L1) test result is used, in conjunction with clinical information, to aid in the evaluation of patients (18 years of age or older), presenting within 12 hours of suspected mild traumatic brain injury (Glasgow Coma Scale score 13-15), to assist in determining the need for a Computed Tomography (CT) scan of the head. A negative interpretation of VIDAS® TBI (GFAP, UCH-L1) test is associated with the absence of acute intracranial lesions visualized on a head CT scan.
The VIDAS® TBI (GFAP, UCH-L1) test is composed of two automated assays – VIDAS® TBI (GFAP) and VIDAS® TBI (UCH-L1) – to be used on the VIDAS® 3 instrument. Similar to other VIDAS assays, VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) test kits (specific to each biomarker) contain the solid phase receptacles (SPRs®), the reagent strips, Product Calibrator S1 and Product Control C1. These test kits will also contain the master lot entry (MLE) data i.e., a barcode printed on the outer label of the packaging, as well as the reference number of the package insert to download from the bioMérieux website.
Whether it be for the GFAP or UCH-L1 quantification, the test combines a three-step enzyme immunoassay sandwich method with a final fluorescent detection step, also known as enzyme-linked fluorescent assay (ELFA).
The Solid Phase Receptacle (SPR) serves as the solid phase as well as the pipetting device. The inner surface of the SPR is coated with antibodies aqainst the substance of interest i.e., anti-GFAP or anti-UCH-L1 antibodies. The reagent strip consists of 10 wells covered with a labeled foil seal. Well 1 is designated for the sample. Eight of the wells contain sample diluent, wash buffer, conjugate, and tracer. The last well contains the fluorescent substrate. All of the assay steps are performed automatically by the instrument.
The intensity of the fluorescence is proportional to the concentration of the analyte the sample. At the end of the assay, the biomarker concentration is automatically calculated by the instrument in relation to the calibration curve and stored in the Master Lot Entry (MLE) data.
VIDAS TBI (GFAP) and VIDAS TBI (UCH-L1) results are reported separately: the VIDAS 3 reports the calculated concentration and the qualitative interpretation for each. The final result i.e., the patient's status in relation to suspected mild traumatic brain injury, must be interpreted by the user according to the decision tree presented in the package insert.
This document describes the validation of the VIDAS® TBI (GFAP, UCH-L1) test, an automated assay for diagnosing mild traumatic brain injury. The submission compares the device to a predicate device, the BANYAN BTI™, and summarizes non-clinical and clinical testing results. The following points address the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for each performance metric in a table format. However, it presents the results of various assays and often implies that the results "demonstrate" or "confirm" the required performance, indicating these are the achieved results compared to an internal standard or regulatory expectation. Below is a table summarizing various performance metrics and their reported results. Specific acceptance criteria values are not provided in this public summary.
| Performance Metric | Reported Device Performance |
|---|---|
| Analytical Measuring Interval | |
| VIDAS TBI (GFAP) | 10.0 - 320.0 pg/mL |
| VIDAS TBI (UCH-L1) | 80.0 - 2560.0 pg/mL |
| Linearity | |
| VIDAS TBI (GFAP) | Demonstrated on the range 6.7 - 354.5 pg/mL |
| VIDAS TBI (UCH-L1) | Demonstrated on the range 58.9 - 2769.1 pg/mL |
| Detection Limits | |
| Limit of Blank (LoB) - GFAP | 4.4 pg/mL |
| Limit of Detection (LoD) - GFAP | 5.4 pg/mL |
| Limit of Quantitation (LoQ) - GFAP | 5.4 pg/mL |
| Limit of Blank (LoB) - UCH-L1 | 41.8 pg/mL |
| Limit of Detection (LoD) - UCH-L1 | 48.1 pg/mL |
| Limit of Quantitation (LoQ) - UCH-L1 | 48.1 pg/mL |
| Hook Effect | |
| VIDAS TBI (GFAP) | No hook effect up to 200,000.0 pg/mL |
| VIDAS TBI (UCH-L1) | No hook effect up to 400,000.0 pg/mL |
| Calibration Frequency | Verified for 56 days |
| Sample Stability | Verified for specified storage conditions and freeze/thaw cycles |
| Diagnostic Accuracy | |
| Diagnostic Sensitivity | 96.7% |
| Diagnostic Specificity | 41.2% |
| Positive Likelihood Ratio | 1.6 |
| Negative Likelihood Ratio | 0.1 |
| Positive Predictive Value | 9.9% |
| Negative Predictive Value | 99.5% |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Test set sample size: For the diagnostic accuracy study, the sample size is not explicitly stated but refers to the "ALERT cohort." For the reference interval study, 513 apparently healthy US adult subjects were used.
- Data provenance: The diagnostic accuracy study was performed using the "ALERT cohort." The reference interval study was conducted at three sites (one internal European site and two external US sites). It is not specified whether these studies were retrospective or prospective, though "ALERT cohort" could suggest a pre-existing dataset.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not provided in the document. The diagnostic accuracy study compares the device's results to the presence/absence of acute intracranial lesions visualized on a head CT scan, but the number or qualifications of experts interpreting these CT scans to establish ground truth are not mentioned.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in vitro diagnostic test for quantitative measurement of biomarkers, not an AI-assisted imaging device that impacts human reader performance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the diagnostic accuracy study presents the standalone performance of the VIDAS® TBI (GFAP, UCH-L1) assay. The results (sensitivity, specificity, etc.) are based on the device's output compared to the ground truth (CT scan findings). The device is used "in conjunction with clinical information," but the reported diagnostic accuracy figures are for the test itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the diagnostic accuracy study was "absence of acute intracranial lesions visualized on a head CT scan." This indicates that CT scan results were used as the reference standard for traumatic brain injury assessment.
8. The sample size for the training set
This document describes a diagnostic device and its validation. It does not explicitly mention a "training set" in the context of machine learning or AI models with distinct training and test phases. The "test set" for diagnostic accuracy is referred to as the "ALERT cohort." The reference interval was established using 513 apparently healthy subjects.
9. How the ground truth for the training set was established
As there is no explicitly defined "training set" for an AI model in this submission, the method for establishing ground truth for a training set is not applicable or described. The clinical performance data presented (Diagnostic Accuracy and Reference interval) seems to represent the evaluation of the final device.
Ask a specific question about this device
(84 days)
VITEK® 2 AST-Gram Positive Lefamulin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
VITEK® 2 AST-Gram Positive Lefamulin is a quantitative test. Lefamulin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active both in vitro and in clinical infections: Staphylococcus aureus (methicillin-susceptible isolates)
The VITEK® 2 Gram-positive Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., and S. agalactive to antimicrobial agents when used as instructed.
The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh (1) and Gerlach (2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique (0).
Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 - 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK® 2 AST-GP Lefamulin (≤ 0.03 –>4 µg/mL) has the following concentrations in the card: 0.125, 0.5, 1, and 2 (equivalent standard method concentration by efficacy in ug/mL).
The VITEK® 2 AST-Gram Positive Lefamulin (≤ 0.03 - ≥4 µg/mL) device is an antimicrobial susceptibility testing system designed for Gram-positive microorganisms. The acceptance criteria and performance of the device are detailed below.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria (from FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems) | Reported Device Performance (VITEK® 2 AST-GP Lefamulin) |
|---|---|---|
| Essential Agreement (EA) | Not explicitly stated in the provided text, but generally expected to be ≥90% for quantitative ASTs. | 90.8% (367/404) |
| Category Agreement (CA) | Not explicitly stated in the provided text, but generally expected to be ≥90% for categorical ASTs. | 99.8% (403/404) |
| Very Major Error (VME) | Not explicitly stated in the provided text, but generally expected to be ≤1.5% | 0.0% (0/3) |
| Major Error (ME) | Not explicitly stated in the provided text, but generally expected to be ≤3% | 0.2% (1/401) |
| Minor Error (mE) | Not explicitly stated in the provided text, but typically also considered in agreement measurements. | N/A (Not reported separately within the table for Lefamulin) |
| Reproducibility | Acceptable results | 100.0% |
2. Sample size used for the test set and the data provenance
The test set included:
- 404 isolates for Essential Agreement reporting and 404 isolates for Category Agreement reporting (derived from the numerators/denominators in Table 2).
- 3 resistant isolates were tested for VME (Very Major Error)
- 401 susceptible isolates were tested for ME (Major Error)
The data provenance is described as an "external evaluation" conducted with "fresh and stock clinical isolates, as well as a set of challenge strains." The document does not specify the country of origin of the data or explicitly state whether the study was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the given text.
4. Adjudication method for the test set
This information is not provided in the given text.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study with human readers assisted by AI is not applicable to this device. This device is an automated antimicrobial susceptibility test system, not an AI-assisted diagnostic tool interpreted by human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance evaluation was done. The VITEK® 2 AST-GP Lefamulin system's performance was compared directly to the CLSI broth microdilution reference method (the ground truth), without human intervention in the interpretation of the VITEK® 2 results. The system automatically generates MIC values and interpretive categories.
7. The type of ground truth used
The ground truth used was the CLSI broth microdilution reference method, incubated at 16-20 hours.
8. The sample size for the training set
The document does not explicitly mention a separate "training set" or its sample size. The description focuses on the external evaluation data used for performance assessment. As an AST system, the device's "training" for MIC determination is inherent in its design based on established microdilution principles and may not involve a distinct, large-scale machine learning training set in the way an AI algorithm might.
9. How the ground truth for the training set was established
Since a distinct training set is not explicitly mentioned as per the prompt's context (e.g., for an AI algorithm), details on how its ground truth was established are not provided. The device's operation is based on established microbiological principles, and its performance is validated against the CLSI broth microdilution reference method.
Ask a specific question about this device
(90 days)
VITEK® 2 Streptococcus Penicillin is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Penicillin is a quantitative test. Penicillin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active both in vitro and in clinical infections: Beta hemolytic Streptococci groups C and G Streptococcus pyogenes Streptococcus agalactiae Streptococcus viridans group Streptococcus pneumoniae
The VITEK® 2 Streptococcus Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Sireptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh(1) and Gerlach(2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique(3).
Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 - 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK® 2 Streptococcus Penicillin has the following concentrations in the card: 0.06, 0.12, 0.5, and 2ug/mL (equivalent standard method concentration by efficacy in ug/mL).
Here's a summary of the acceptance criteria and study details for the VITEK® 2 Streptococcus Penicillin device:
1. Acceptance Criteria and Reported Device Performance
The device performance is compared to the broth microdilution reference method. The key metrics are Essential Agreement (EA) and Category Agreement (CA), along with Very Major Errors (VME), Major Errors (ME), and Minor Errors (mE).
| Antimicrobial | Organism/Comment | Acceptance Criteria (Not explicitly stated as numerical, but implied to be "acceptable performance") | Reported Device Performance: Essential Agreement (EA) | Reported Device Performance: Category Agreement (CA) | Reported Device Performance: VME | Reported Device Performance: ME | Reported Device Performance: mE |
|---|---|---|---|---|---|---|---|
| Penicillin | Streptococcus pneumoniae (non-meningitis) | Acceptable performance | 97.7% (342/350) | 92.3% (323/350) | 0.0% (0/0) | 0.0% (0/304) | 7.7% (27/350) |
| Penicillin | Streptococcus pneumoniae (oral) | Acceptable performance | 97.7% (342/350) | 89.7% (314/350) | 0.0% (0/94) | 0.0% (0/189) | 10.3% (36/350) |
| Penicillin | Streptococcus pneumoniae (meningitis) | Acceptable performance | 97.7% (342/350) | 97.7% (342/350) | 1.2% (2/161) | 3.2% (6/189) | N/A (not applicable) |
| Penicillin | Streptococcus Viridans group (except Streptococcus pneumoniae) | Acceptable performance | 98.5% (385/391) | 94.1% (368/391) | 0.0% (0/15) | 0.0% (0/268) | 5.9% (23/391) |
| Penicillin | Beta-hemolytic Streptococcus | Acceptable performance | 100% (833/833) | 100% (833/833) | 0.0% (0/0) | 0.0% (0/833) | N/A (not applicable) |
The document states that the device "demonstrated substantially equivalent performance when compared with the broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)." This guidance document typically outlines specific numerical acceptance criteria for EA, CA, VME, ME, and mE, which are implicitly met by the reported percentages being considered "acceptable."
2. Sample Size Used for the Test Set and Data Provenance
The sample sizes for the test set are embedded within the "Reported Device Performance" column (e.g., 350 for Streptococcus pneumoniae (non-meningitis), 391 for Streptococcus Viridans group, 833 for Beta-hemolytic Streptococcus).
The data provenance is from "external evaluation... conducted with fresh and stock clinical isolates, as well as a set of challenge strains." This implies a combination of retrospective (stock clinical isolates) and prospective (fresh clinical isolates) data. The country of origin of the data is not specified in the provided text.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The ground truth is established by the CLSI broth microdilution reference method, but details on expert involvement in this process (e.g., reading/interpreting the reference method results) are not mentioned.
4. Adjudication Method for the Test Set
This information is not explicitly provided in the document. The study compares the VITEK® 2 results to the CLSI broth microdilution reference method. Adjudication might be part of the standard CLSI method, but it is not detailed here for discrepancies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study focuses on the performance of an automated diagnostic device against a reference method, not on human reader performance with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, a standalone performance study was done. The VITEK® 2 system is a "Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System," and its performance was evaluated against the CLSI broth microdilution reference method. This is an evaluation of the algorithm's output (MIC values and interpretive categories) without direct human intervention in the interpretation of the VITEK® 2 results.
7. The Type of Ground Truth Used
The type of ground truth used is the broth microdilution reference method, specifically the CLSI method incubated at 16-20 hours. This is typically considered the gold standard for antimicrobial susceptibility testing.
8. The Sample Size for the Training Set
The document does not explicitly state the sample size for the training set. The study describes "external evaluation" for performance, but it doesn't separate out a specific training set size for the VITEK® 2 algorithm itself.
9. How the Ground Truth for the Training Set Was Established
The document does not explicitly state how the ground truth for any training set was established, as it primarily focuses on the performance evaluation of the final device against a reference method. However, given that the device is an "Automated quantitative antimicrobial susceptibility test," it's highly probable that any internal training/development would have used an established reference method (like broth microdilution) to establish ground truth for algorithm development.
Ask a specific question about this device
(98 days)
VITEK® 2 AST-Gram Positive Daptomycin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Positive Daptomycin is a quantitative test. Daptomycin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active both in vitro and in clinical infections: Enterococcus faecalis (vancomycin-susceptible isolates only) Staphylococcus aureus (including methicillin-resistant isolates)
In vitro data are available, but their clinical significance is unknown: Enterococcus faecalis (vancomycin-resistant isolates)
The VITEK® 2 Gram-positive Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed.
The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh(1) and Gerlach(2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique(3).
Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 - 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
VITEK® 2 AST-GP Daptomycin has the following concentrations in the card: 0.5, 1, 2, 4, and 8 ug/mL (equivalent standard method concentration by efficacy in ug/mL).
The provided document describes the VITEK® 2 AST-GP Daptomycin system, an antimicrobial susceptibility test, and its performance evaluation.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for antimicrobial susceptibility test (AST) systems in the US are generally defined by the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems. While specific numerical targets for Essential Agreement (EA) and Category Agreement (CA) are usually outlined in this guidance, the summary states the device demonstrated substantially equivalent performance when compared with the broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document.
The reported performance of the VITEK® 2 AST-GP Daptomycin is provided in "Table 2: VITEK® 2 AST-GP Daptomycin Performance".
| Metric | Staphylococcus aureus | Enterococcus faecalis | Overall (implied from text) |
|---|---|---|---|
| Acceptance Criteria | (As per FDA Guidance, typical minimums for new ASTs are usually >90% for EA and >90% for CA with specific limits on VME/ME/mE) | ||
| Essential Agreement (%EA) | (183/194) 94.3% | (262/270) 97.0% | 95.9% |
| Very Major Errors (VME) | N/A | (1/8) 12.5% | |
| Major Errors (ME) | (2/185) 1.1% | (0/219) 0.0% | |
| Minor Errors (mE) | N/A | (41/270) 15.2% | |
| Category Agreement (%CA) | (192/194) 99.0% | (228/270) 84.4% | 90.5% |
Note on VME/ME/mE for Staphylococcus aureus: The document lists "N/A" for VME and mE for Staphylococcus aureus and "0.0%" for ME for Enterococcus faecalis. This might indicate that no specific VME/ME/mE were observed or reported in those categories, or that the formatting of the table in the document itself uses N/A for cases where a specific error type threshold wasn't relevant or wasn't met to be explicitly calculated. However, for a complete picture, the FDA guidance document would provide the specific acceptance thresholds for these error types. The (1/8) 12.5% VME for Enterococcus faecalis and (41/270) 15.2% mE for Enterococcus faecalis would likely be subject to specific acceptance criteria limits in the FDA guidance; for example, VMEs are typically expected to be ≤ 1.5% and MEs ≤ 3.0%, while mEs generally have a higher tolerance but still a limit. The document states a 90.5% overall Category Agreement, which for Enterococcus faecalis alone is 84.4%, falling below the typical 90% threshold for CA. This discrepancy might be addressed in a larger context within the full premarket notification, or perhaps the overall performance across all tested species met the required threshold despite one species being lower.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- Staphylococcus aureus: 194 isolates
- Enterococcus faecalis: 270 isolates
- Total Clinical Isolates in Performance Study: 194 + 270 = 464 isolates.
- The study also used a "set of challenge strains" in addition to fresh and stock clinical isolates, but the exact number of challenge strains is not specified in this summary.
- Data Provenance:
- "An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains." This suggests a prospective or a mix of prospective and retrospective collection from external sources (likely clinical laboratories).
- The document does not specify the country of origin for the data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- The ground truth was established by the CLSI broth microdilution reference method. This method is a standardized laboratory procedure, not reliant on human expert interpretation in the same way as, for example, image reading. Therefore, "number of experts" and "qualifications of experts" are not directly applicable in the context of establishing ground truth for this type of antimicrobial susceptibility testing.
4. Adjudication Method for the Test Set
- The reference method itself (CLSI broth microdilution) serves as the "ground truth" or "adjudication." No separate human adjudication process (e.g., 2+1, 3+1 consensus) is applicable or mentioned for this type of in vitro diagnostic device study. The device's results are directly compared to the results of the reference method.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study design is typically used for AI-powered diagnostic imaging devices where human readers interpret images with or without AI assistance. This device is an automated in vitro diagnostic system for antimicrobial susceptibility testing, not requiring human interpretation of complex visual data for its primary function.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, the performance study effectively evaluates the "standalone" performance of the VITEK® 2 AST-GP Daptomycin system. The system automatically processes the samples and generates MIC values and interpretive categories, which are then compared directly to the CLSI broth microdilution reference method. There is no human intervention in the device's determination of susceptibility.
7. The Type of Ground Truth Used
- The ground truth used was the CLSI broth microdilution reference method, which is a highly standardized and accepted laboratory method for determining antimicrobial susceptibility. This is essentially a "gold standard" laboratory method.
8. The Sample Size for the Training Set
- The document does not provide information on the training set size. This type of premarket notification summary focuses on the performance evaluation of the final device, not typically on the specific dataset used for algorithm development or training. The "Discriminant Analysis" mentioned under "Analysis Algorithms" implies a statistical model was used, which would have been developed/trained on a dataset, but details of this dataset are not included in this summary.
9. How the Ground Truth for the Training Set Was Established
- As the training set details are not provided, the method for establishing its ground truth is also not described in this summary. It would logically be established using a similar or identical reference method to the test set, but this is not confirmed.
Ask a specific question about this device
Page 1 of 21