The theraPORT Low-Profile Vascular Access System is a totally implantable venous access system intended for use with patients that require repeated venous access for injection or infusion therapy and/or venous blood sampling. Because of its low-profile design, it is indicated for either traditional chest placement or placement in the upper arm, and is appropriate for use in pediatric patients and small adults.
The theraPORT® Low-Profile Vascular Access System is a totally implantable venous access system consisting of a detached catheter and port.
The provided text describes the theraPORT® Low-Profile Vascular Access System, a medical device, and its regulatory submission. It does not contain information about acceptance criteria for an AI/CAD/software device, nor does it detail a study proving such a device meets acceptance criteria. The document is a 510(k) summary for a physical implanted medical device.
Therefore, I cannot provide the requested information. The document explicitly states: "Clinical studies were not conducted as they were determined to be not necessary due to the similarity in design, performance, materials, function and intended use of the theraPORT® Low-Profile Vascular Access System to other commercially available systems."
The "Performance Testing" section describes mechanical tests of the physical device rather than a study of an AI/CAD/software system. These tests are:
- Catheter-to-port connection strength tests
- Septum puncture durability tests
- Port leakage integrity tests
- Port/catheter clearance tests
These are not relevant to the performance metrics typically associated with AI/CAD systems (e.g., sensitivity, specificity, AUC).
Ask a specific question about this device
(81 days)
The theraPORT® is intended for use with patients that require repeated venous access for injection or infusion therapy and/or venous blood sampling.
The theraPORT® Vascular Access System is a totally implantable venous access system consisting of a detached catheter and port.
The provided text describes a medical device, the theraPORT® Vascular Access System, and outlines performance testing rather than acceptance criteria and a study to prove meeting those criteria in the context of diagnostic accuracy or clinical outcomes. The device is an implanted system, and the "studies" mentioned focus on engineering performance tests as opposed to clinical trials.
Here's an analysis based on the provided text, addressing your questions where possible:
1. Table of Acceptance Criteria and Reported Device Performance
The submission mentions that performance testing was conducted in accordance with the "Guidance on 510(k) Submissions for Implanted Infusion Ports, October 1990." While specific quantitative acceptance criteria are not explicitly stated, the types of tests performed indicate the areas of performance evaluated. The "reported device performance" is summarized as the device "functions properly" and is "substantially equivalent."
| Acceptance Criteria (Implied from tests performed) | Reported Device Performance |
|---|---|
| Catheter-to-port connection strength | Functions properly |
| Septum puncture durability | Functions properly |
| Port leakage integrity | Functions properly |
| Port/catheter clearance | Functions properly |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide the sample sizes for the performance tests conducted. It also does not specify the origin of any data, as the tests described are laboratory-based engineering performance tests, not clinical studies with human subjects or retrospective data.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This question is not applicable to the provided document. The ground truth for engineering performance tests is typically defined by engineering specifications and standards, not by expert consensus in a clinical diagnostic sense. No clinical experts were involved in what's described as "performance testing."
4. Adjudication Method for the Test Set
This is not applicable. The tests described are objective engineering performance tests, not subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Clinical studies were not conducted as they were determined to be not necessary due to the similarity in design, performance, materials, function and intended use of the theraPORT® Vascular Access System to other commercially available systems." Therefore, there is no effect size of human readers improving with or without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
This question is not applicable. The theraPORT® Vascular Access System is a physical medical device, not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for the performance tests would be defined by engineering standards and specifications outlined in the "Guidance on 510(k) Submissions for Implanted Infusion Ports, October 1990." For example, a "catheter-to-port connection strength" test would have a specified minimum force it must withstand, which serves as the ground truth for that particular test.
8. The Sample Size for the Training Set
This question is not applicable. The device is a physical product, not an AI or machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable for the reasons stated in point 8.