Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bead Block microspheres are intended to be used for the embolization of hypervascular tumours, including uterine fibroids (UFE) and arteriovenous malformations (AVMs). Bead Block microspheres are also intended for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH).

    Device Description

    Bead Block, a permanent intravascular implant, is made up of preformed soft, compressible, biocompatible, hydrophilic, non-resorbable and precisely calibrated microspheres that occlude vessels for the purpose of blocking the blood flow to a target tissue. Bead Block compressible microspheres consist of a macromer derived from polyvinyl alcohol (PVA). The fully polymerized microsphere is approximately 90-95% water and is compressible to approximately 30% by diameter. Bead Block microspheres are dyed blue to aid in the visualization of the microspheres in the delivery syringe. The microspheres can be suspended in contrast agents and delivered through microcatheters to the target location. Bead Block is available in bead sizes from 100 – 1200µm and supplied sterile in 20ml syringes which contain 1 or 2 ml of beads suspended in 6 or 5 ml of phosphate buffered saline, respectively. The different bead sizes of the product are differentiated by differently colored labels and syringe end caps. Bead Block is provided as a single use, non-pyrogenic, sterile (steam sterilized) device.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "Bead Block" device. It attests to the device's substantial equivalence to previously marketed predicate devices for the specified indications for use.

    Based on the provided text, the Acceptance Criteria and Device Performance for this medical device (Bead Block) are not defined in terms of typical AI/ML-based image analysis performance metrics (e.g., sensitivity, specificity, AUC). Instead, the document focuses on demonstrating substantial equivalence to existing predicate devices based on safety and effectiveness for its intended use, which is embolization.

    The "study" described is a retrospective data review of the clinical outcomes of patients treated with Bead Block for prostatic artery embolization (PAE) for symptomatic benign prostatic hyperplasia (BPH). This is a clinical effectiveness study, not a study of an AI/ML device's diagnostic or analytical performance.

    Therefore, many of the requested points related to AI/ML device study parameters (e.g., test set, ground truth experts, MRMC studies, standalone performance) are not applicable to this document.

    Here's an attempt to answer the prompt based on the provided text, reinterpreting the "acceptance criteria" and "study" in the context of a medical device submission for substantial equivalence:

    1. A table of acceptance criteria and the reported device performance

    The "acceptance criteria" for this 510(k) submission are not explicitly stated as quantitative thresholds for clinical performance but rather are implicitly tied to demonstrating safety and effectiveness comparable to predicate devices. The "reported device performance" refers to the clinical outcomes observed in the retrospective study.

    Acceptance Criteria (Implicit from 510(k) Process)Reported Device Performance (from Retrospective Study)
    Safety: Device is safe for intended use, with acceptable adverse event profile compared to predicate.Observed AEs: 149/232 patients (64.2%) reported at least one AE. Most AEs were non-serious and transient. Most common were renal/urinary disorders (15.2%), dysuria (41.4%), pollakiuria (45.3%). Conclusion: "well tolerated treatment."
    Effectiveness: Device effectively achieves intended clinical outcome, comparable to predicate.Clinical Improvements at 12 Months: - 85% of patients: decrease in total IPSS by at least 3 points. - 62% of patients: dropped at least 1 symptom category (severe to moderate to mild). - Statistically significant and clinically relevant improvements in total IPSS, QoL, PSA, and prostate volume (p<0.001). - Improvement tendency in Qmax (p=0.059) and PVR (p=0.1). - PAE did not impair sexual function (IIEF). Conclusion: "effective... with comparable outcomes to Embosphere Microspheres."
    Technological Characteristics: Similar to predicate devices, or differences do not raise new safety/effectiveness questions.Technological Comparison: - Similar size-calibrated spherical microspheres, delivered by microcatheters. - Similar size ranges, steam sterilized, single-use, biocompatible, non-resorbable polymer. - Delivery via microcatheter, visualization via radiographic imaging. - Minor material difference (PVA vs. acrylic/porcine gelatin) "do not raise different questions of safety and effectiveness."
    Biocompatibility: Device material is biocompatible for intended use.Previous Biocompatibility Testing: Genotoxicity, implantation, sensitization, intracutaneous reactivity, acute, subchronic toxicity. Studies showed material is biocompatible.
    Material/Device Performance: Physical and chemical properties suitable for intended use.Previous Bench Testing: Residuals, visual defects, color, solution clarity, catheter delivery (clogging, aggregates, injection ease), particle fiber shedding, pH, packaging integrity, shelf life, sterility, endotoxins.
    MR Safety: Demonstration of MR safety/compatibility.MR Safe: Based on scientific rationale (non-ferrous composition) in alignment with FDA guidance. Labeling updated to reflect MR safety.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 232 patients.
    • Data Provenance: Retrospective data review. The country of origin is not explicitly stated in this document but implied to be from a clinical setting where Bead Block was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was a retrospective clinical outcomes study, not a study involving expert readers establishing ground truth for image interpretation by an AI device. The "ground truth" here is the observed clinical outcomes in patients.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this was a clinical outcomes study, not an imaging study requiring reader adjudication. Clinical outcomes (IPSS, QoL, prostate volume, Qmax, PVR, IIEF, AEs) were measured directly.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a PMA (pre-market approval) or 510(k) submission for a physical medical device (embolization microspheres), not an AI/ML-based diagnostic or assistive device. Therefore, no MRMC study or AI assistance comparison was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the effectiveness portion of the study was clinical outcomes data (e.g., International Prostate Symptom Score (IPSS), erectile function (IIEF), Quality of Life (QoL), Prostate volume (PV), objective measures of urinary flow (Qmax, PVR)) and adverse events (AEs) reported by patients or clinicians over the course of the study.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set. The "study" described is a retrospective clinical review, which would be analogous to a "test set" in an AI/ML context, if at all.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/ML algorithm. The clinical data collected was used to demonstrate the performance of the device itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150876

    Validate with FDA (Live)

    Device Name
    Bead Block
    Date Cleared
    2016-03-07

    (341 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bead Block™ microspheres are intended to be used for the embolization of hypervascular tumors, including uterine fibroids and arteriovenous malformations (AVMs).

    Device Description

    Bead Block™ is made up of preformed soft, deformable microspheres that occlude arteries for the purpose of blocking the blood flow to a target tissue, such as a fibroid or a cancerous tumor. Bead Block™ compressible microspheres consist of a macromer derived from polyvinyl alcohol (PVA). The fully polymerized microsphere is approximately 90-95% water and is compressible to approximately 30% by diameter. Bead Block™ microspheres are dyed blue to aid in the visualization of the microspheres in the delivery syringe. The microspheres can be delivered through typical microcatheters in the 1.5-5Fr range.

    AI/ML Overview

    The provided document is a 510(k) summary for the Bead Block™ device, which is a vascular embolization device. The information requested pertains to the acceptance criteria and the study that proves the device meets the acceptance criteria. The document largely focuses on demonstrating substantial equivalence to a predicate device rather than performance against specific numeric acceptance criteria for the new device.

    Here's an analysis based on the available information:

    1. Table of acceptance criteria and the reported device performance

    The document does not present a table of explicit numeric acceptance criteria. Instead, it relies on demonstrating that Bead Block™ is "as safe and effective as the predicate device" and that its performance is "similar to the predicate device". The "performance" is described qualitatively through the results of non-clinical and animal testing.

    Acceptance Criteria CategoryReported Device Performance
    Material Properties:
    Residual starting materialSpecifications met (details not provided)
    Residual solvent materialSpecifications met (details not provided)
    Physical Appearance:
    Visual inspectionVisual defects, color, solution clarity evaluated (results not explicitly stated, but implied to be acceptable for substantial equivalence)
    Delivery & Functionality:
    Catheter deliveryEvaluated for catheter clogging, aggregation, ease of injection, and particle shape after injection (results not explicitly stated, but implied to be acceptable for substantial equivalence, and "similar to the predicate device")
    Particle size rangeConfirmation performed (results not explicitly stated, but implied to be within specified range as per product configurations table, e.g., 100-300μm, 300-500μm, etc.)
    Particle fiber sheddingEvaluated (results not explicitly stated, but implied to be acceptable for substantial equivalence)
    Biocompatibility:
    pHEvaluated (results not explicitly stated, but implied to be acceptable for substantial equivalence)
    Endotoxin<0.06 EU/ml
    BiocompatibilityEvaluated (results not explicitly stated, but implied to be acceptable for substantial equivalence and "minimal tissue necrosis and a common foreign body reaction; however, the intensity of inflammatory reaction is moderate, without humoral response or cytotoxicity" in animal studies)
    Sterility & Packaging:
    Packaging integrityEvaluated (results not explicitly stated, but implied to be acceptable for substantial equivalence)
    Shelf lifeEvaluated (results not explicitly stated, but implied to be acceptable for substantial equivalence)
    Sterilization validationPerformed (results not explicitly stated, but implied to be acceptable for substantial equivalence)
    Animal Study - UAE:
    Short term performance"similar to the predicate device, with both showing normal, thrombosis and some rupture of the internal elastic lamina"
    Side effects"minimal tissue necrosis and a common foreign body reaction; however, the intensity of inflammatory reaction is moderate, without humoral response or cytotoxicity"
    DistributionEvaluated (results not explicitly stated, but implied to be acceptable for substantial equivalence)
    Clinical - Adverse Events:"Review of the published and unpublished data regarding adverse events associated with Bead Block™ has not identified any unique safety concerns." (This implies a qualitative comparison to expected adverse events for similar devices).

    2. Sample size used for the test set and the data provenance

    • Non-clinical testing: No specific sample sizes are mentioned for the non-clinical tests (e.g., material specifications, visual inspection, catheter delivery). These typically involve lab-based testing.
    • Animal studies:
      • Study 1 (sheep penetration model): The sample size is not explicitly stated. The provenance is likely a research lab setting.
      • Study 2 (side effects in sheep uteri): The sample size is not explicitly stated. The provenance is likely a research lab setting.
    • Clinical information: This involved a "review of Bead Block™ published and unpublished data on the use of Bead Block™ for the treatment of leiomyoma uteri (uterine fibroids) [outside the United States] over the last ten years."
      • The exact sample size (number of patients/cases) is not provided.
      • Provenance: This data is retrospective, collected over "the last ten years," and originated from "outside the United States."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not detail the use of "experts" to establish a ground truth in the typical sense of a diagnostic device. The evaluation of non-clinical tests would be performed by qualified lab personnel. For animal studies, the assessment of tissue necrosis, inflammation, and vascular effects would be done by veterinary pathologists or other trained researchers, but no specific number or qualifications are given. For the clinical data review, it's not described as an adjudication process by experts to establish ground truth for a test set, but rather a review of existing clinical outcomes/adverse events.

    4. Adjudication method for the test set

    There is no mention of an adjudication method (like 2+1 or 3+1) for establishing ground truth from multiple readers/experts. This type of adjudication is typically relevant for diagnostic imaging studies where agreement on subtle findings is crucial. The studies described are primarily performance and safety assessments of an embolic device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. The device (Bead Block™) is an embolic agent, not an AI-powered diagnostic or assistive tool for human readers. Therefore, the question of human reader improvement with or without AI assistance is not applicable to this device submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. Bead Block™ is a physical medical device (microspheres for embolization), not an algorithm or AI system. Its performance is inherent to its physical properties and interaction with biological systems when used by a clinician, not a standalone software performance metric.

    7. The type of ground truth used

    • Non-clinical testing: Ground truth is established by standard laboratory test methods, specifications, and physical measurements (e.g., chemical analysis for residual materials, microscopy for particle size, functional testing for catheter delivery).
    • Animal studies: Ground truth is established by histological examination and pathological assessment of tissue samples from the embolized animals, comparing observed effects to normal physiology and effects seen with the predicate device.
    • Clinical data review: Ground truth is the documented adverse events and clinical outcomes collected in real-world use outside the US, reviewed descriptively (not as a systematically adjudicated "ground truth" for a specific test set, but rather a summary of real-world safety data).

    8. The sample size for the training set

    The concept of a "training set" is not applicable here. This is a medical device submission based on non-clinical, animal, and existing clinical data, not a machine learning model developed with training data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an algorithm/AI in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152157

    Validate with FDA (Live)

    Device Name
    LC Bead LUMI
    Date Cleared
    2015-12-11

    (130 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LC Bead LUMI are intended to be used for the embolization of hypervascular tumors and arteriovenous malformations (AVMs).

    Device Description

    LC Bead LUMI™ is an imageable spherical embolic product that can be visualised by X-ray based imaging (e.g. fluoroscopy and CT). The beads are non-resorbable microspheres with calibrated size ranges. LC Bead LUMI™ are intended to be used for the embolization of hypervascular tumors and arteriovenous malformations (AVM).

    LC Bead LUMI™ are supplied in a saline buffer solution within a sealed glass vial. A vial access device is also provided within the secondary packaging. LC Bead LUMI™ is provided sterile (moist heat sterilization).

    LC Bead LUMI™ consist of a macromer derived from a sulphonate modified polyvinyl alcohol (PVA) macromer which contains a radiopaque moiety that is covalently bound within the hydrogel structure. The incorporation of this radiopaque moiety into the co-polymer imparts Xray imageability by rendering the microspheres radiopaque.

    LC Bead LUMI™ has been designed as a radiopaque version of the LC Bead™. Currently embolization with LC Bead™ lacks post-procedural imaging feedback on exact bead location. The embolization process with LC Bead™ is monitored by detecting changes in antegrade flow of soluble iodinated contrast in which the beads are diluted. The embolization is continued until a desired embolization endpoint is reached. This process is completed without specific feedback on the bead location. In order to address this limitation, the manufacturer has developed an imageable spherical embolic bead that can be visualised by X-ray based imaging.

    At the time of use, LC Bead LUMI™ is mixed with non-ionic contrast agent in order to monitor the delivery of the product during the embolization procedure. LC Bead LUMI™ is delivered via typical microcatheters to physically block the target vessel. Once the product has been delivered the iodine moiety allows for visualization of the microspheres. LC Bead LUMI™ is a single use, tissue contacting, implantable device available only for prescription use.

    AI/ML Overview

    The provided document does not contain information regarding traditional acceptance criteria or a study design that outlines specific performance metrics for a diagnostic device. Instead, this document is a 510(k) premarket notification for a medical device called "LC Bead LUMI™," which is a vascular embolization device, not a diagnostic one.

    The document focuses on demonstrating substantial equivalence to a predicate device (LC Bead™) by comparing various characteristics and providing performance data relevant to safety and effectiveness for its intended use as an embolization agent.

    Therefore, I cannot fulfill the request to provide:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size for the test set and data provenance.
    3. Number and qualifications of experts for ground truth.
    4. Adjudication method for the test set.
    5. MRMC study details and effect size with AI assistance.
    6. Standalone performance details.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    These elements are typically found in submissions for diagnostic or AI-powered devices that assess a specific outcome or condition, often requiring metrics like sensitivity, specificity, accuracy, etc., against a defined ground truth.

    What the document does provide in terms of "performance data" for LC Bead LUMI™ is:

    • Bench testing data: To show that the device performs as expected in controlled laboratory settings.
    • Verification and validation studies: To confirm that the device meets its design specifications and user needs.
    • In vitro and in-vivo biocompatibility studies: To assess the safety of the material in biological systems.
    • In-vivo safety and performance studies: To evaluate the device's safety and how it functions within living organisms.
    • The conclusion is that these studies demonstrate equivalent safety and performance to the predicate device (LC Bead™), especially considering the key difference is the added radiopaque moiety for visualization.

    The studies were conducted in accordance with "FDA Guidance - Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices, December 29, 2004".

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1