Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K091414

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-10-21

    (161 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioCare vTrust 701DH NIBP monitor is a compact, battery-powered monitor which measures noninvasive blood pressure (systolic and mean arterial pressure) and pulse rate based on the principle of plethysmography (oscillometric on inflation). This equipment is intended for use only by qualified medical providers in conjunction with established medical protocols. It is not intended for use in an MRU environment, or in an air transport environment.

    Device Description

    The BioCare vTrust 701DH NIBP monitor is a compact, battery-powered monitor which measures noninvasive blood pressure (systolic, diastolic and mean arterial pressure) and pulse rate based on the principle of plethysmography (oscillometric on inflation). The oscillometric method detects volume displacements within the artery and senses pressure variations within the blood pressure cuff during inflation. The measurement is more quickly and with less patient discomfort than with monitors which measure NIBP during cuff deflation. Cycle time function enables you to pre-set measurement intervals. User can set alarm limits by selectable for high or low range.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    ClausePerformance CharacteristicsMax. permissible error or acceptance criteriaReported Device Performance (Max. deviation)Passed/Failed
    4.2.1Storage conditions: [23°F(-5°C)] for 24 hrs; [122°F(50°C)] for 24 hrs; RH: 90% (noncondensing)≤± 3 mmHg (±0.4 kPa) ≤± 4 mmHg for in use-3.4 mmHgPassed
    4.2.2Operating conditions: Temp: 50°F (10°C)104°F(40℃); RH: 1590% (noncondensing); Bar. 105kPa80kPa (790mmHg600mmHg)≤± 3 mmHg (±0.4 kPa) ≤±4 mmHg for in use± 3 mmHgPassed
    4.2.3Vibration and shock: ISTA-standard drop and vibration test IEC 60601-1, sections 21.5 & 21.6ISTA-1A: 14,200 vibratory impacts; drop test-10 drops; before and after transportation test within ±5 mmHg of the average reading.no visible damage; SD ≤8 mmHgPassed
    4.2.4.1Voltage range: Battery-powered devices-indication of battery conditionRed light: low battery; Blue light: batteries are charging; Green light: fully charged<8.2VDC; 8.2~8.4VDCPassed
    4.2.4.2Life: Maintain the safety and performance characteristics for 10,000 full-scale cycles. Adult mode: 150 mmHg; Neo. mode: 75 mmHg≤ 3 mmHg (±0.4 kPa)2 mmHgPassed
    4.2.5Electromagnetic compatibilitycomplied with EC 60601-1-2 (EMC)N/APassed
    4.3.1.1Maximum cuff pressure: Adult: 300mmHg; neonate: 150 mmHgNIBP pressure calibration: <310 mmHg; Safety test: Release pressure before <320 mmHgN/APassed
    4.3.1.2Release rate: Rapid exhaust: 26015 mmHg ≤10s; 1505 mmHg ≤5sNIBP pressure calibration: deflate at 25 mmHg/sec; Speed test: Deflation >25 mmHg/secmmHg/secPassed
    4.3.2Electrical safety:IEC 60601-1, 1988; Amendment 1, 1991; & Amendment 2, 1995.N/APassed
    4.4.3.1.ARange: 0 mmHg to at least 260 mmHg by visual inspectionNIBP pressure calibration0~300 mmHgPassed
    4.4.3.2.AResolution: 1 mmHg by visual inspectionNIBP pressure calibration1 mmHgPassed
    4.4.3.3.AAccuracy of pressure measurement: Ambient temp 10°C~40°C; RH: 15%~90%; Pressure stages: 50, 100, 150, 200, 250, 300 mmHg≤±3 mmHg (±0.4 kPa) or 2% of the reading; ≤±4 mmHg for in use±3 mmHgPassed
    4.4.3.4.ARepeatability Cuff-Link simulator test for stability and reproduction of performance Adult: SYS120/DIA80/MAP90 mmHg, 80 BPM Neo: SYS80/DIA50/MAP62 mmHg, 120 BPMAverage difference ≤± 5 mmHg5 mmHgPassed
    4.4.3.6.ALeakage: Ambient temp 15°C25°C; RH: 2085%; Pressure stages: 50, 100, 150, 200, 250 mmHg, wait at least 60 s before reading the values.≤6 mmHg/min1.2 mmHg/minPassed
    4.4.4.BPressure transducer accuracy: Ambient temp 10℃~40℃; RH: 15%~90%; Pressure stages: 50, 100, 150, 200, 250, 300 mmHg≤±3 mmHg (±0.4 kPa) or 2% of the reading; ≤±4 mmHg for in use±3 mmHgPassed
    4.4.5.BOverall system efficacy-clinical testing (auscultatory method as the reference Std.) no fewer than 85 subjects with a minimum of 255(=3*85) observationsFor systolic and diastolic: mean≤±5 mmHg; SD ≤8 mmHgMean: Sys/Dia 3.4/-4.5; SD: Sys/Dia 5.23/4.64 mmHgPassed
    4.5.1Inflation source: pressure~300mmHg≤10sNIBP pressure calibration: inflate at 5 mmHg/sec; Speed test: Inflation > 5 mmHg/secN/APassed
    4.5.4.3.BAutomated valves-Release rate: rapid exhaust with fully-opened valve260mmHg → 15mmHg ≤10 sec or 150mmHg → 5mmHg ≤5 secdeflate at 25 mmHg/secPassed
    IEC 60601-2-30Automatic cycling function 1. In long term automatic mode (normal condition) Adult 120/80(90)[SYS/DIA(MAP)]--long term Neonate 80/50(62)[SYS/DIA(MAP)]--long term 2. In long term automatic mode (single fault condition) 3. In short term automatic mode (stat mode)1. deflated time ≥30 sec; mean≤± 5 mmHg; 2. deflated time ≤30 sec; release cuff pressure to below 15mmHg(adult)/ 5mmHg(neo.); 3. deflated time≥2sec. after each determination1. average for adults 34 sec / for neonates 42 sec; -4mmHg; 2. <15sec, safety activates; 3. average for adults 3 sec / for neonates 8 sec; --4mmHgPassed

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Clinical Testing (Test Set): No fewer than 85 subjects with a minimum of 255 observations (3 observations per subject).
    • Data Provenance: Not explicitly stated, but the submission is from BioCare Corporation (Taiwan), suggesting the study may have been conducted in Taiwan or a related region. The document does not specify if the data was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • The ground truth for the clinical testing was established using the auscultatory method as the reference standard. This method typically involves trained medical professionals (e.g., nurses, doctors) taking manual blood pressure measurements using a stethoscope and sphygmomanometer. The document does not specify the exact number or precise qualifications (e.g., years of experience) of the experts performing the auscultatory measurements for this study.

    4. Adjudication Method for the Test Set:

    • The document implies that the auscultatory measurements served as the direct reference standard. There is no mention of a separate adjudication process involving multiple experts for the clinical ground truth.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This study focused on assessing the standalone performance of the device against a known reference standard (auscultatory method), not on the improvement of human readers with AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Yes, a standalone performance study was done. The entire set of performance tests, including bench and clinical results, are for the BioCare vTrust 701DH Noninvasive Blood Pressure Monitor operating as a standalone device, without human-in-the-loop intervention beyond the initial application of the cuff and initiation of measurement. The clinical testing specifically compared the device's readings against the auscultatory method.

    7. Type of Ground Truth Used:

    • For the clinical study, the ground truth was established using the auscultatory method (a recognized clinical reference standard for blood pressure measurement). For other performance characteristics, the "acceptance criteria" themselves served as the ground truth, derived from the ANSI/AAMI SP-10: 2002 standard.

    8. Sample Size for the Training Set:

    • The document does not specify a separate training set or its sample size. Blood pressure monitors like this typically operate based on established oscillometric algorithms rather than being "trained" in the machine learning sense with a distinct training dataset that is then validated on a separate test set. The validation here refers to the device's overall performance against the ANSI/AAMI SP-10: 2002 standard and clinical reference.

    9. How the Ground Truth for the Training Set Was Established:

    • As a training set is not explicitly mentioned or implied in the context of the device's validation, the method for establishing its ground truth is not applicable or provided in this document. The device's underlying oscillometric principle is based on known physiological responses and engineering principles, not typically on machine learning model training.
    Ask a Question

    Ask a specific question about this device

    K Number
    K012829

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-11-02

    (71 days)

    Product Code
    Regulation Number
    862.2900
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stripmax 1210 Urine Chemistry Analyzer is a semi-automatic urinalysis instrument intended for reading UriMate Reagent Strips for Urinalysis. The tests provided on UriMate Reagent Strips are considered routine urinalysis.

    UriMate Reagent Strips for Urinalysis are dip-and-read test strips intended for use as an in vitro diagnostic aid using urine specimens. UriMate Reagent strips provide qualitative and semiquantitative tests for specific gravity, pH, Blood, leukocytes, nitrite, protein, bilirubin, ketone(acetoacetic acid), glucose, and urobilinogen by visual comparison to a color chart or by use of the Stripmax Urine Chemistry Analyzer. The tests provided are considered routine urinalysis.

    Device Description

    The Stripmax 1210 Urine Chemistry Analyzer is a semi-automatic urinalysis instrument intended for reading UriMate Reagent Strips for Urinalysis. UriMate Reagent Strips for Urinalysis are dip-and-read test strips intended for use as an in vitro diagnostic aid using urine specimens.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Stripmax 1210 Urine Chemistry Analyzer" and "UriMate Reagent Strip for Urinalysis." However, it does not contain any information about acceptance criteria, device performance, study designs, sample sizes, ground truth establishment, or expert qualifications.

    The document is a clearance letter from the FDA stating that the device is substantially equivalent to legally marketed predicate devices. It lists the device name, regulation numbers, classification, product codes, and indications for use.

    Therefore, I cannot fulfill your request for the detailed information about acceptance criteria and the study that proves the device meets them based on the provided text. The requested information is simply not present in this regulatory clearance document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K002855

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-10-12

    (29 days)

    Product Code
    Regulation Number
    862.1665
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1