Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K082428

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-02-10

    (172 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The N-DEX® BWC™ Pink Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner. (21 CFR 880.6250)

    This device has been tested for use with specific chemotherapy drugs. This device is not intended to be used as a chemical barrier.

    Device Description

    The N-DEX® BWC™ Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner.

    AI/ML Overview

    This 510(k) submission describes the N-DEX® BWC™ Nitrile Powder-Free Medical Examination Glove (pink) and its substantial equivalence to a predicate device.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is a medical examination glove, and its performance is evaluated against established ASTM standards and biocompatibility tests.

    Acceptance Criteria (Standards/Tests)Predicate Device PerformanceProposed Device Performance
    Performance Standards (Physical Properties)
    ASTM D 6319-00a (Nitrile Examination Gloves)ConformsConforms (ASTM D 6319-00a3 listed, likely a minor revision)
    ASTM D 6124-01 (Powder Residue on Medical Gloves)ConformsConforms (ASTM D 6124-06 listed, likely a minor revision)
    ASTM D 5151-99 (Water penetration through finished surgical gloves)ConformsConforms (ASTM D 5151-06 listed, likely a minor revision)
    Biocompatibility
    ISO Skin Irritation Study (ISO 10993-10)PassesPasses
    ISO Closed Patch Sensitization Study (ISO 10993-10)PassesPasses
    Cytotoxicity Study (ISO 10993-5)Not tested (for predicate, but accepted for proposed as part of equivalence)Passes

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the sample sizes used for each ASTM standard test or biocompatibility study. However, these standards typically specify minimum sample sizes for testing to achieve statistically significant results.
    • Data Provenance: The studies were conducted by Best Glove, Inc. and are considered retrospective in the sense that the data is presented to demonstrate conformance to existing standards rather than generated from a new clinical trial. The country of origin for the data is not explicitly stated, but given the company's US address and compliance with ASTM and ISO standards, it is highly likely the testing was performed in the US or in accredited labs adhering to international standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This type of submission for a medical glove does not involve establishing ground truth from human expert consensus in the way an AI diagnostic device would. Instead, ground truth for material properties is established by adherence to validated ASTM and ISO test methods performed in laboratories. The "experts" involved would be qualified laboratory technicians, engineers, and scientists specialized in materials testing and biocompatibility. Their qualifications would be based on their expertise in performing and interpreting these specific tests according to regulatory and industry standards.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this device. The conformance to standards is determined by objective laboratory measurements and interpretations of those results against the criteria defined by the respective ASTM and ISO standards. There is no ambiguous result that requires human adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices that involve human interpretation of medical images or data. For a medical examination glove, the "effectiveness" is primarily determined by its physical properties and biocompatibility, as measured by standard laboratory tests, not by human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the performance evaluation in this submission is effectively a "standalone" evaluation of the glove's physical and biological properties. The tests (ASTM for physical properties, ISO for biocompatibility) are objective measurements of the device itself, without any human-in-the-loop diagnostic or interpretive component.

    7. The Type of Ground Truth Used

    The ground truth used is based on:

    • Standardized Test Methods and Specifications: Conformance to specific ASTM and ISO standards (e.g., ASTM D 6319-00a for glove properties, ISO 10993 for biocompatibility). These standards define objective, measurable criteria for acceptable performance.
    • Predicate Device Performance: The primary ground truth for demonstrating substantial equivalence is the performance of the legally marketed predicate device (Ultimate N-DEX® Nitrile Powder-Free Medical Examination Glove, Non-Sterile (blue), K001375). The proposed device must meet or exceed the performance of the predicate device for relevant parameters.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set. The "training" in the context of manufacturing would relate to process control and quality assurance to ensure consistent product attributes, not a data-driven training of a model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As noted above, there is no AI/ML training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082125

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-11-21

    (115 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner (21 CFR 880.6250).

    Device Description

    The N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove is a disposable device intended for medical purposes; that is worn on the examiner's hand, to prevent contamination between the patient and examiner.

    AI/ML Overview

    Here's an analysis of the provided 510(k) submission, focusing on the acceptance criteria and the study that proves the device meets those criteria:

    Device: N-DEX® NightHawk® Black Nitrile Powder-Free Medical Examination Glove (Non-Sterile)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard/Test)Predicate Device PerformanceProposed Device PerformanceType of Test
    ASTM D 6319-00a (Performance)ConformsConformsNon-Clinical
    ASTM D 6124-01 (Performance)ConformsN/A (Updated to 6124-06)Non-Clinical
    ASTM D 6124-06 (Performance)N/A (Older standard)ConformsNon-Clinical
    ASTM D 5151-99 (Water Tightness)ConformsN/A (Updated to 5151-06)Non-Clinical
    ASTM D 5151-06 (Water Tightness)N/A (Older standard)ConformsNon-Clinical
    ISO 10993-10 (Skin Irritation)PassesPassesBiocompatibility
    ISO 10993-10 (Sensitization)PassesPassesBiocompatibility
    ISO 10993-5 (Cytotoxicity)Not testedPassesBiocompatibility

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the non-clinical performance tests (ASTM standards) or the biocompatibility tests (ISO standards). The results are reported as "Conforms," "Passes," or "Not tested," implying that the testing was conducted according to the methodology outlined in the respective standards.

    For biocompatibility, the tests were conducted, but the specific number of animals or human subjects (if applicable for irritation/sensitization under specific protocols) is not provided.

    The data provenance is from testing conducted by the manufacturer, Best Glove, Inc. No country of origin for the data is specified beyond the company's location in the USA. The tests are prospective as they were conducted to demonstrate the safety and effectiveness of the new device for the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This submission does not involve an expert-reviewed test set in the way a medical image analysis or diagnostic device might. The "ground truth" for this device (medical examination glove) is established by its adherence to established performance standards (ASTM) and biocompatibility standards (ISO).

    Therefore, there were no "experts" in the traditional sense involved in establishing a ground truth for a test set for diagnostic accuracy. Instead, the "ground truth" is defined by the passing criteria of the specified international standards for glove performance and safety. The expertise lies in the certified laboratories and personnel who conduct these standardized tests according to the established protocols.

    4. Adjudication Method for the Test Set

    Not applicable. This device is not a diagnostic tool requiring expert adjudication of results. The performance and biocompatibility tests are based on objective measurements and established pass/fail criteria within the relevant ASTM and ISO standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret data (e.g., medical images) and AI might assist in that interpretation. This submission is for a medical examination glove, which does not involve human readers interpreting diagnostic information.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical medical glove and does not involve an algorithm or AI.

    7. The Type of Ground Truth Used

    The ground truth used is based on established objective performance criteria and biocompatibility thresholds defined by recognized international standards:

    • Performance: ASTM D 6319-00a/06, ASTM D 6124-01/06, ASTM D 5151-99/06. These standards define physical properties like tensile strength, elongation, puncture resistance, and watertightness, with specific acceptance limits.
    • Biocompatibility: ISO 10993-10 (for irritation and sensitization) and ISO 10993-5 (for cytotoxicity). These standards define acceptable biological responses when the device comes into contact with tissues.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical glove and does not involve a "training set" in the context of machine learning or AI algorithms. The manufacturing process and quality control are based on established engineering principles and quality management systems, not data training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1