Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K231406

    Validate with FDA (Live)

    Device Name
    ClearumTM HS
    Manufacturer
    Date Cleared
    2023-06-14

    (30 days)

    Product Code
    Regulation Number
    876.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Clearum™ HS 17 is intended for use in acute or chronic renal failure patients requiring hemodialysis

    Device Description

    The Clearum™ HS 17 dialyzer is consisting of a cylindrical polypropylene body containing a bundle of microporous hollow fibers made of polyethersulfone (PES) secured to the ends by means of hot-melt polyurethane resin.
    The modifed Clearum™ HS 17 dialyzer is substantially equivalent to the predicate Clearum™ HS 17 dialyzer manufactured by Bellco (K193542, cleared on October 8, 2020).
    The device is characterized by a membrane surface area equivalent to 1.7 m², and by an outer housing total height of 306 mm and a maximum outside diameter of 44 mm.
    ldentical to the predicate Clearum™ HS 17, blood and dialysate flow in a countercurrent in their respective compartments. In this process, toxins and fluid are transferred across the semipermeable membrane from the blood to the dialysate compartment.
    The dialyzers are sterilized using moist heat with saturated steam, have a non-pyrogenic fluid path, and are labeled for single use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called Clearum™ HS 17, a high permeability hemodialysis system. The submission seeks to demonstrate substantial equivalence to a predicate device (also Clearum™ HS 17, K193542) despite a minor modification in fiber undulation amplitude and effective fiber length.

    This document does not contain information about an AI/ML-based medical device or a study involving human readers and AI assistance. Therefore, it is impossible to extract the requested information regarding acceptance criteria and a study proving the device meets those criteria for an AI/ML context.

    The document focuses on the equivalence of a physical medical device (a dialyzer) based on in-vitro test results. The "study" referenced is a series of non-clinical, in-vitro tests comparing the modified device to its predicate.

    To directly address your request as if it were for an AI/ML device, I would need a different source document. However, based solely on the provided text, I can only report on the non-AI/ML aspects related to the Clearum™ HS 17 dialyzer.

    Here's a breakdown of what can be extracted, framed within the closest possible interpretation of your request for this non-AI/ML device:

    Acceptance Criteria and Device Performance (for a physical hemodialyzer)

    The document states that "The modified Clearum™ HS 17 has the same fundamental technological characteristics, principles of operation and control mechanisms as the predicate device." The acceptance criterion, implicitly, is that the modified device performs comparably to the predicate device in key in-vitro performance metrics, demonstrating substantial equivalence and not raising new safety or effectiveness concerns.

    Acceptance Criteria (Implicit - based on comparison to predicate)Reported Device Performance
    Maintain performance specifications as defined by relevant standards (e.g., ISO 8637-1:2017) and guidance (e.g., FDA guidance for hemodialyzers)."The results of the testing met the performance specifications demonstrating that the modified Clearum™ HS 17 performs as intended."
    Demonstrate substantial equivalence to the predicate Clearum™ HS 17 (K193542)."The predicate device substantial equivalence was also demonstrated."
    No new or increased biocompatibility concerns."The biocompatibility testing and/or chemical characterization followed by a toxicological risk assessment that was completed on the predicate Clearum™ HS 17 applies to the modified Clearum™ HS 17..."
    Maintain or improve performance across key metrics:In vitro testing conducted on both modified and predicate device.
    - Priming VolumeTest conducted.
    - Pressure Drop across Blood CompartmentTest conducted.
    - Pressure Drop across Dialysate CompartmentTest conducted.
    - Ultrafiltration CoefficientTest conducted.
    - Sieving Coefficient (Albumin, Myoglobin, Inulin)Test conducted.
    - Clearance (Urea, Creatinine, Phosphate, Vitamin B12)Test conducted.
    - Hemo-compatibility (mechanical hemolysis)Test conducted.

    Study Details (for a physical hemodialyzer)

    • 1. A table of acceptance criteria and the reported device performance: (See table above)

    • 2. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated for each test, but in-vitro testing was conducted on "both the modified and predicate Clearum™ HS 17 for comparative purpose." This implies multiple units were tested to generate the reported performance data. Specific numbers of dialyzers tested for each metric are not provided.
      • Data Provenance: The tests were conducted internally by Bellco S.r.l. or a contracted lab. The country of origin is Bellco S.r.l. in Mirandola (MO), Italy. The data is from in-vitro testing, not retrospective or prospective human clinical data.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable as the "ground truth" for a physical device's performance is established by standardized in-vitro measurements, not expert consensus or human reading. The "experts" would be the scientists and engineers conducting and interpreting the physical and chemical tests, presumably adhering to ISO standards.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. This concept applies to human interpretation or algorithm output, not to objective physical measurements of a device.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device. No MRMC study was performed.
    • 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable. This is not an AI/ML device, so there is no algorithm to evaluate in standalone mode.
    • 7. The type of ground truth used:

      • In-vitro performance metrics: The "ground truth" for this device's performance is derived from objective physical and chemical measurements (e.g., clearance rates, ultrafiltration coefficients, pressure drops) performed in controlled laboratory settings according to established scientific and engineering principles and international standards (ISO 8637-1:2017).
    • 8. The sample size for the training set:

      • Not applicable. This is not an AI/ML device, so there is no training set in the context of machine learning. The device design and manufacturing process are based on engineering principles and material science.
    • 9. How the ground truth for the training set was established:

      • Not applicable. As above, there's no training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193542

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-08

    (293 days)

    Product Code
    Regulation Number
    876.5860
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Clearum™ HS dialyzer family is intended for use in acute or chronic renal failure patients requiring hemodialysis.

    Device Description

    The Clearum™ HS family consists of dialyzers comprised by a cylindrical polypropylene body containing a bundle of microporous hollow fibers made of polyethersulfone (PES) secured to the ends by means of hot-melt polvurethane resin.

    The devices are available in different models which differentiate by membrane surface area, ranging from 1.3 to 2.2 m², and for the dimensions of the outer cylindrical body in terms of length and height (306 x 41 mm for the surface areas in the range 1.3-1.5 m², 306 x 44 mm for the surface area 1.7 m² and 366 x 44 mm for the surface areas in the range 2.0-2.2 m²).

    Similar to other commercially available hemodialyzers, blood and dialysate flow in a countercurrent in their respective compartments. In this process, toxins and fluid are transferred across the semipermeable membrane from the blood to the dialysate compartment.

    The dialyzers are sterilized using moist heat with saturated steam, have a non-pyrogenic fluid path, and are labeled for single use.

    AI/ML Overview

    The provided text describes the Clearum™ HS family of dialyzers and their comparison to a predicate device, the Polyflux H models, to demonstrate substantial equivalence for FDA 510(k) clearance. This report does not describe acceptance criteria for an AI/ML powered medical device, but rather for a physical medical device. It focuses on in-vitro test results and comparison to a predicate device.

    Here's an analysis of the provided information, noting where specific requested details (like those for AI/ML studies) are not available in this physical device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly list "acceptance criteria" in a table format for each performance metric, but rather states that "The results of the testing met the performance specifications demonstrating that the Clearum™ HS family performs as intended." It also aims to demonstrate substantial equivalence to the predicate device.

    The "performance specifications" are implicitly derived from the established performance characteristics of the predicate device and the industry standards (ISO 8637-1:2017).

    Performance MetricClearum™ HS Family Performance (Implicit)Predicate Device Performance (Polyflux H) (Implicit)Acceptance Criteria (Implicit)
    Priming VolumeMet performance specifications; demonstrated substantial equivalence to predicate.Polyflux 140H: 94 ml; Polyflux 170H: 115 ml; Polyflux 210H: 125 ml (These are the predicate's reported values, implying the Clearum™ HS values are comparable and within acceptable variation for substantial equivalence). Clearum™ HS: HS 13: 84ml, HS 15: 95ml, HS 17: 105ml, HS 20: 120ml, HS 22: 126mlPerformance must meet established design specifications and demonstrate substantial equivalence to the predicate device, consistent with ISO 8637-1:2017.
    Pressure Drop (Blood Compartment)Met performance specifications; demonstrated substantial equivalence to predicate.Not explicitly stated for predicate in summary, but comparative testing was done.Performance must meet established design specifications and demonstrate substantial equivalence to the predicate device, consistent with ISO 8637-1:2017.
    Pressure Drop (Dialysate Compartment)Met performance specifications; demonstrated substantial equivalence to predicate.Not explicitly stated for predicate in summary, but comparative testing was done.Performance must meet established design specifications and demonstrate substantial equivalence to the predicate device, consistent with ISO 8637-1:2017.
    Ultrafiltration CoefficientMet performance specifications; demonstrated substantial equivalence to predicate.Not explicitly stated for predicate in summary, but comparative testing was done.Performance must meet established design specifications and demonstrate substantial equivalence to the predicate device, consistent with ISO 8637-1:2017.
    Sieving Coefficient (Albumin, Myoglobin, Inulin)Met performance specifications; demonstrated substantial equivalence to predicate.Not explicitly stated for predicate in summary, but comparative testing was done.Performance must meet established design specifications and demonstrate substantial equivalence to the predicate device, consistent with ISO 8637-1:2017.
    Clearance (Urea, Creatinine, Phosphate, Vitamin B12)Met performance specifications; demonstrated substantial equivalence to predicate.Not explicitly stated for predicate in summary, but comparative testing was done.Performance must meet established design specifications and demonstrate substantial equivalence to the predicate device, consistent with ISO 8637-1:2017.
    Hemocompatibility (mechanical hemolysis)Met performance specifications; demonstrated substantial equivalence to predicate.Not explicitly stated for predicate in summary, but comparative testing was done.Performance must meet established design specifications and demonstrate substantial equivalence to the predicate device, consistent with ISO 8637-1:2017 and ISO 10993-1.
    SterilityEffectiveness of production techniques to assure sterility demonstrated.Moist heat with saturated steam.Must be sterile.
    Non-pyrogenicityEffectiveness of production techniques to assure non-pyrogenicity demonstrated. Device has a non-pyrogenic fluid path.Not explicitly stated for predicate in summary.Must be non-pyrogenic.
    BiocompatibilityIn accordance with ISO 10993-1 and FDA Guidance "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (limited to external communicating device, circulating blood, prolonged contact duration).Not explicitly stated for predicate in summary for comparison, but predicate would have met similar standards.Must comply with ISO 10993-1 for biological evaluation, specific to external communicating (circulating blood, prolonged contact).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document states "In vitro testing was conducted on the entire Clearum™ HS dialyzers family" and "For comparative purposes, the same testing, was also conducted on the Polyflux H predicate device, when applicable." However, the exact number of units tested (sample size) for each model or for the overall family is not specified in the provided text.
    • Data Provenance: The testing was "in-vitro," meaning it was conducted in a laboratory setting. There is no mention of human subject data, retrospective or prospective studies in the context of device performance. The manufacturer, Bellco S.r.l., is located in Mirandola MO, Italy, suggesting the tests were likely conducted there or at an affiliated lab.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device submission. For physical medical devices like dialyzers, "ground truth" is established through physical and chemical testing against recognized standards (e.g., ISO, FDA guidance) and comparison to the predicate device's established performance. It does not involve expert adjudication of clinical data in the way an AI/ML device would.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving discrepancies in expert interpretations of data, typically in AI/ML validation studies. For this physical device, performance is measured objectively through lab tests.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This is a submission for a physical medical device (hemodialyzer), not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-assisted human reader improvement metrics are relevant or discussed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    For this device, the "ground truth" is based on:

    • Industry Standards: Compliance with ISO 8637-1:2017 ("Cardiovascular implants and extracorporeal systems - Hemodialysers, hemodiafilters, hemofilters and hemoconcentrators") and ISO 10993-1 ("Biological evaluation of medical devices").
    • FDA Guidance: "Guidance for the Content of Premarket Notifications for Conventional and High Permeability Hemodialyzers" (1998) and "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing."
    • Predicate Device Performance: The established, legally marketed performance characteristics of the Polyflux H dialyzer family. The new device demonstrates "substantial equivalence" to the predicate.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" in the context of an AI/ML algorithm for this physical medical device.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as there is no training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN180055

    Validate with FDA (Live)

    Device Name
    CARPEDIEM System
    Date Cleared
    2020-04-29

    (568 days)

    Product Code
    Regulation Number
    876.5861
    Type
    Direct
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K160558

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-11-18

    (263 days)

    Product Code
    Regulation Number
    876.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 800 series RAPIDO BLS HOLLOW-FIBRE DIAL YZERS are single-use devices intended for hemodialysis (HD), hemofiltration (HF), and hemodiafiltration (HDF) to treat conditions of acute or chronic renal failure when conservative therapy is judged to be inadequate. They are also indicated in the treatment intoxicated with poisons or drugs.

    Device Description

    The above mentioned devices are dialyzers consisting of a cylindrical polycarbonate transparent body containing a bundle of microporous hollow fibers in highly-permeable polyethersulfone (Diapes® HF) secured to the ends by means of hot-melt polyurethane resin.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the RAPIDO BLS HOLLOW-FIBRE DIALYZER. It describes the device, its intended use, and comparative testing against predicate devices to demonstrate substantial equivalence, rather than a study proving the device meets specific acceptance criteria in the context of an AI/algorithm-based medical device.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set details) are not applicable to this type of regulatory submission and device.

    However, I can extract the relevant information regarding the acceptance criteria and the study type used for this medical device, which is an in-vitro performance study.

    Here's the information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "The results of these tests met established specifications." However, the specific numerical acceptance criteria for each performance parameter are not provided in this summary. The reported performance is implicitly that the device performs substantially equivalently to the predicate devices.

    Acceptance Criteria (General)Reported Device Performance (Implicit)
    Meets established specifications for mechanical integrityMet established specifications
    Meets established specifications for priming volumeMet established specifications
    Meets established specifications for pressure dropMet established specifications
    Meets established specifications for ultrafiltration rateMet established specifications
    Meets established specifications for sieving coefficientMet established specifications
    Meets established specifications for hemolysisMet established specifications
    Compliant with safety and effectiveness requirementsDemonstrated substantial equivalence and compliance
    Biocompatible for intended use (ISO 10993-1)Demonstrated biocompatibility
    Sterile and non-pyrogenicDemonstrated effectiveness of production techniques
    Package integrity testing met specificationsMet established specifications
    Shipping carton provides adequate protectionPassed basic testing and provided adequate protection

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in the document. The testing mentioned is "in vitro test results" and refers to "the device" (singular) implying testing on a sample, but the specific number of units tested is not given.
    • Data Provenance: In vitro test results. The device manufacturer is Bellco Srl, located in Mirandola (MO), Italy. The tests were conducted internally or by a contracted lab. The document describes these as "non-clinical" and "in vitro" tests, not involving human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is an in-vitro performance study for a physical medical device, not an AI/algorithm-based device relying on expert-annotated ground truth.

    4. Adjudication method for the test set

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an AI/algorithm. However, the performance described is of the device itself (standalone), not in conjunction with a human operator making clinical decisions based on its output (beyond operating the dialysis machine it connects to).

    7. The type of ground truth used

    • Established Specifications/Predicate Device Performance: The "ground truth" or reference for evaluating the RAPIDO BLS was its ability to meet established specifications for various performance parameters and to demonstrate substantial equivalence to legally marketed predicate devices (Xenium Dialyzer, Syntra, and Polyflux 6H). This relied on comparing the numerical values obtained from the in-vitro tests against these benchmarks.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device; there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The study was an in vitro and non-clinical performance evaluation comparing the RAPIDO BLS HOLLOW-FIBRE DIALYZER to existing predicate devices (Xenium Dialyzer, Syntra, and Polyflux 6H).

    The testing included:

    • Biocompatibility testing: Conducted in accordance with ISO 10993-1.
    • Sterility, pyrogenicity, EtO residual, and package integrity testing.
    • In vitro functional performance testing:
      • Mechanical integrity
      • Priming volume
      • Blood side pressure drop
      • Ultrafiltration rate
      • Sieving coefficient
      • Hemolysis

    For comparative purposes, the same in-vitro functional performance tests were also conducted on the predicate devices where applicable.

    Conclusion of the Study:
    The document concludes that the results from these in vitro studies demonstrate that the RAPIDO BLS performs in a manner substantially equivalent to the predicate devices and that it is biocompatible and safely manufactured (sterile, non-pyrogenic). All tests "met established specifications." This substantial equivalence argument is the basis for FDA clearance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1