510(k) Data Aggregation

    K Number
    K162398
    Date Cleared
    2017-07-06

    (314 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The equipment is used for human tissue vaporization, coagulation in dermatology, plastic surgery, and general surgery. The fractional scanner is only for the treatment of wrinkles and skin resurfacing.

    Device Description

    The CO2 Diode Laser Therapy Systems mainly consist of console, articulated arm and a foot switch. CO2 laser wavelength is 10600nm, which is used to gasify the skin tissue based on the principle of the theory selective photothermolysis. The water in skin is the main target of the CO2 fractional laser, after absorbed the energy of the laser, the water will release heat, then be evaporated and gasified immediately, the cutaneous lesion will be eliminated instantly. By setting up the suitable energy and other parameters on skin tissue based on its specific thermal relaxation time, the healthy tissue can be prevented from being harmed.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the CO2 Laser Therapy Systems (K162398). The submission focuses on demonstrating substantial equivalence to a predicate device (CO2 Laser Therapy Machine, K161925) rather than providing a study where the device meets specific performance acceptance criteria from a clinical trial.

    Therefore, many of the requested elements for a study proving acceptance criteria cannot be extracted because such a study is not part of this 510(k) summary. Instead, the document provides non-clinical performance data (biocompatibility, electrical safety, EMC, and software verification/validation) to support safety and effectiveness and substantial equivalence.

    Here's a breakdown of the information that can be extracted, and where the requested information is not available:

    1. A table of acceptance criteria and the reported device performance

    This document does not present a table of specific acceptance criteria (e.g., specific clinical performance metrics like sensitivity, specificity, accuracy, or a reduction in wrinkles of a certain percentage) that the device was evaluated against in a clinical study. Instead, the performance data provided are non-clinical performance tests demonstrating compliance with recognized standards.

    Acceptance Criteria CategoryReported Device Performance (Non-Clinical)
    BiocompatibilityComplies with ISO 10993-1. Passed tests for Cytotoxicity, Sensitization, and Irritation for skin and subcutaneous tissue contact (< 24 hours).
    Electrical SafetyComplies with IEC 60601-1 (standard for safety).
    EMC (Electromagnetic Compatibility)Complies with IEC 60601-1-2 (standard for EMC).
    General PerformanceComplies with IEC 60825-1 (for laser product safety).
    Software V&VSoftware verification and validation conducted; documentation provided per FDA guidance. Software classified as "moderate" level of concern.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. No clinical test set or patient data (retrospective or prospective) was used for this 510(k) submission. The performance testing described is non-clinical (biocompatibility, electrical safety, EMC, software).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No clinical test set requiring expert-established ground truth was used.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a CO2 laser therapy system, not an AI diagnostic or assistance tool for human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or included.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a CO2 laser therapy system, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. No clinical ground truth (expert consensus, pathology, outcomes data) was used or required for this 510(k) submission as no clinical studies were performed. The "ground truth" for the non-clinical tests was compliance with the specified international standards.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set. This is a physical medical device (CO2 laser).

    9. How the ground truth for the training set was established

    Not applicable. As noted above, this is not an AI/machine learning device.


    Summary of what the document does provide regarding device performance:

    The document focuses on demonstrating that the CO2 Laser Therapy Systems is substantially equivalent to a predicate device (K161925) and meets all applicable non-clinical safety standards. This is the primary method for demonstrating safety and effectiveness in a 510(k) submission when clinical studies are not deemed necessary by the FDA.

    • Comparison of Technological Characteristics: A detailed table compares the proposed device with the predicate device across various specifications (Output power, Work mode, Laser medium/energy source, CO2 laser wavelength, Aiming Beam wavelength and power, Spot Size, Dot Quantity, Pulse Energy, Pulse Length, Cooling, Patient contact sites, Control System, Laser Operation, Power input). The conclusion is that differences do not raise new questions of safety and effectiveness.
    • Non-Clinical Performance Data:
      • Biocompatibility: The device components that contact the patient were tested and found to be biocompatible according to recognized standards (ISO 10993-1).
      • Electrical Safety & EMC: The device complies with IEC 60601-1 and IEC 60601-1-2 standards, ensuring electrical safety and electromagnetic compatibility.
      • General Performance Testing: The device was tested according to IEC 60825-1, a standard for the safety of laser products.
      • Software Verification and Validation: Software testing was conducted, and it was deemed a "moderate" level of concern, with documentation provided as per FDA guidance.
    • No Clinical Studies: The submission explicitly states, "The subject of this premarket submission, CO2 Laser Therapy Systems, does not require clinical studies to support substantial equivalence." This is why there is no information on clinical acceptance criteria or studies.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163123
    Date Cleared
    2017-02-02

    (86 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Q-Switched Nd: YAG Laser Therapy System (1064nm or 532nm) is indicated for use in tattoo removal, treatment of benign vascular lesions, treatment of benign pigmented lesions, incision, excision, ablation, vaporization of soft tissue for general dermatology as follows:

    1064nm: -Tattoo Removal Dark ink: blue and black. -Treatment of Benign Pigmented Lesions Nevus of ota. 532mm: -Tattoo Removal Light ink: red. Light ink: sky blue and green. - Treatment of Benign Vascular Lesions Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi. -Treatment of Benign Pigmented Lesions Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus.

    Device Description

    The Q-Switched Nd:YAG Laser Systems is based on the Q-Switched Nd: YAG (1064 nm) and frequency doubled KTP Nd: YAG (532nm) Laser Technology. There is one optical cavity containing the Nd: YAG Crystal. It is composed of laser module, articulated arm, laser power supply, cooling system and display and control system.

    The Q-Switched Nd: YAG Laser Therapy Systems works based on laser selective photothermy and blasting mechanism of Q-switched laser. Energy form particular wavelength with accurate dose will act on certain targeted color radicals: ink, carbon particles from derma and epidermis, exogenous pigment particles and endogenous melanophore from derma and epidermis. When suddenly being heated, pigment particles immediately blast into smaller pieces, which will be swallowed by macrophage phagocytosis and enters into lymph circulation system and finally be discharged out of body.

    The physician is able to select the desired wavelength and the related output energy, spot size and fluency via control panel.

    AI/ML Overview

    The provided document is a 510(k) summary for the Beijing Sincoheren Science And Technology Development Co.'s Q-Switched Nd:YAG Laser Therapy Systems. It aims to demonstrate substantial equivalence to a predicate device, not to establish new safety or effectiveness. Therefore, the document discusses performance data in the context of equivalence rather than proving acceptance criteria for a novel device.

    The document does not describe specific acceptance criteria in the typical sense (e.g., minimum sensitivity, specificity, or accuracy) for a new clinical indication or a new type of device. Instead, it relies on demonstrating that the performance characteristics of the proposed device are comparable to or better than those of the predicate device, and that it complies with relevant safety standards.

    Here's an analysis of the provided information based on your requested points, framed within the context of a 510(k) submission for substantial equivalence:

    1. A table of acceptance criteria and the reported device performance

    Since this is a 510(k) submission for substantial equivalence, the "acceptance criteria" are implicitly that the proposed device performs at least as well as, or is sufficiently similar to, the predicate device in terms of technical specifications and safety. The reported device performance is presented as a comparison table against the predicate device.

    SpecificationPredicate device performanceProposed device performanceComparison Discussion (Implicit Acceptance)
    K numberK133158--(Predicate ID)
    Product CodeGEXGEXIdentical
    ManufacturerBeijing Toplaser Technology Co., LtdBeijing Sincoheren Science and Technology Development Co., Ltd.(Different companies, but not a performance metric)
    Device nameNd: YAG Laser SystemQ-Switched Nd: YAG Laser Therapy SystemsSimilar
    Indications for use1064nm: -Tattoo Removal (Dark ink: blue and black) -Treatment of Benign Pigmented Lesions (Nevus of ota) 532nm: -Tattoo Removal (Light ink: red; Light ink: sky blue and green) - Treatment of Benign Vascular Lesions (Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi) -Treatment of Benign Pigmented Lesions (Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus.)1064nm: -Tattoo Removal (Dark ink: blue and black) -Treatment of Benign Pigmented Lesions (Nevus of ota) 532nm: -Tattoo Removal (Light ink: red; Light ink: sky blue and green) - Treatment of Benign Vascular Lesions (Port wine birthmarks; Telangiectasias; Spider angioma; Cherry angioma; Spider nevi) -Treatment of Benign Pigmented Lesions (Cafe-au-lait birthmarks; Solar lentiginos; Senile lentiginos; Becker's nevi; Freckles; Nevus spilus.)Identical (Demonstrates equivalence in intended use)
    TechnologyQ-Switched ND:YAG and KTP Nd: YAG LaserQ -Switched ND:YAG and KTP Nd: YAG LaserIdentical
    Energy SourceXenon LampXenon LampIdentical
    Wavelength (nm)1064nm and 532nm1064nm and 532nmIdentical
    Aiming beam wavelength635nm635nmIdentical
    Laser output modeQ-switched pulseQ-switched pulseIdentical
    Aiming laser output power<5mw0.1-5mwIdentical
    Maximum Pulse Energy@1064nm wavelength: 1-1000mJ @532nm wavelength: 50-300mJ@1064nm wavelength: 500mJ @532nm wavelength: 250mJLess than Predicate device, more safe (Acceptable as "safer" or at least equivalent for intended use)
    Pulse Duration6~8ns5ns±1ns or 5nsLess Pulse Duration, more peak power, more effective (Acceptable and potentially an improvement)
    Repetition Rate1~10Hz1-5HzLess than Predicate device, more safe (Acceptable as "safer" or at least equivalent for intended use)
    Nominal ocular hazard distanceNOHD 2.8kmNOHD 3.3kmSimilar (Acceptable)
    Spot SizeAdjustable Spot Size 3-9mm(Diameter)Adjustable Spot Size 2-10mm(Diameter)Wider adjustable range of spot size than predicate device, more useful (Acceptable and potentially an improvement)
    MaterialSteel, ABSSteel, ABSIdentical
    Beam deliveryArticulating Arm Light GuideArticulating Arm Light GuideIdentical
    CoolingInternal distilled water circulating coolingInternal distilled water circulating coolingIdentical
    Anatomical SitesSkin and subcutaneous tissueSkin and subcutaneous tissueIdentical

    In addition to these direct comparisons, the submission outlines compliance with various safety and performance standards as "performance data" to support the substantial equivalence claim:

    • Biocompatibility testing: Compliance with ISO 10993-1, including Cytotoxicity, Sensitization, and Irritation tests. (Implicit acceptance criteria: device materials must be biocompatible according to recognized standards for skin and subcutaneous tissue contact < 24 hours).
    • Electrical safety and electromagnetic compatibility (EMC): Compliance with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC). (Implicit acceptance criteria: device must meet electrical safety and EMC standards).
    • Performance testing: Compliance with IEC 60825-1 (laser safety). (Implicit acceptance criteria: laser performance must meet safety standards).
    • Software Verification and Validation Testing: Documentation provided as recommended by FDA guidance for "moderate" level of concern software. (Implicit acceptance criteria: software must be verified and validated according to FDA guidance for its risk level).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a "test set" in the context of clinical or image-based studies. The performance data presented are primarily technical specifications and compliance with international standards, not data from a clinical test set. Therefore, information on sample size and data provenance for a test set is not applicable to this submission as described.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as there is no clinical or image-based "test set" described with ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there is no clinical or image-based "test set" requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a laser therapy system, not an AI-assisted diagnostic or therapeutic tool that would involve human readers or image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical laser device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. For this type of device and submission, the "ground truth" for demonstrating substantial equivalence lies in meeting established engineering standards (e.g., IEC 60601-1, IEC 60825-1, ISO 10993-1) and demonstrating comparable basic technical performance to a legally marketed predicate device for the same indications for use. There is no biological "ground truth" involved in the sense of clinical validation of a diagnostic or treatment outcome for this specific 510(k) submission.

    8. The sample size for the training set

    Not applicable. This is a hardware medical device, not a machine learning model, so there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    In summary:

    This 510(k) submission primarily relies on bench testing, compliance with recognized international standards (IEC, ISO), and a direct comparison of technical specifications to a predicate device to demonstrate substantial equivalence. It does not involve experimental clinical studies, human reader studies, or AI algorithm validation data that would typically require "test sets," "ground truth," or "training sets" as you've defined them. The "acceptance criteria" are effectively met by demonstrating that the device is technically comparable or improved (e.g., safer, more effective according to physical principles) for the same intended use as a legally marketed predicate device, and that it adheres to relevant safety and performance standards for medical lasers.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152898
    Date Cleared
    2016-02-09

    (131 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy Systems are intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Therapy Systems mainly consist of console, treatment hand piece and a foot switch. The diode laser emits wavelength of 808nm from the treatment hand piece, the wavelength effectively penetrates deep into and absorbed by the target chromophore. The adequate pulse duration, energy density and epidermal cooling ensure an adequate thermal damage to the target tissue without damaging the surrounding tissue to achieve effective hair removal.

    AI/ML Overview

    Based on the provided document, here's an analysis of the acceptance criteria and the study that proves the device meets them:

    Note: This document is a 510(k) summary for a medical device (Diode Laser Therapy Systems, Model SDL-B) and focuses on establishing substantial equivalence to a predicate device, not necessarily on demonstrating clinical efficacy through a traditional comparative effectiveness study with pre-defined acceptance criteria for performance metrics like sensitivity or specificity. The "acceptance criteria" here refer to meeting regulatory standards and demonstrating equivalence in function and safety.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Standard MetReported Device Performance
    Intended Use EquivalenceTo be substantially equivalent to the predicate device for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), with a definition of permanent hair reduction as long-term, stable reduction measured at 6, 9, and 12 months post-treatment.The proposed device, Diode Laser Therapy Systems, has the same intended use as the predicate device (Emvera Diolux, K123257). Both are intended for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), with the same definition for permanent hair reduction.
    Technological EquivalenceDemonstrate similar technological characteristics or that differences do not raise new questions of safety or effectiveness.Wavelength: Same (808 nm) Fluence (Energy Density): Same (<120 J/cm²) Frequency: Same (1Hz-10Hz) Material: Same (Sapphire) Cooling: Same (Water cooling) Output Power: Proposed device (500W) is less than predicate (600W). Regarded as acceptable. Pulse Duration: Proposed device (5ms-400ms) has a smaller maximum pulse duration than predicate (5ms-625ms). Regarded as acceptable. Spot Size: Proposed device (12mm x 10mm) is smaller than predicate (12mm x 12mm). Regarded as acceptable. Anatomical Sites: Proposed device covers fewer anatomical sites (Axilla, Face, bikini, Upper limb and lower limb) than predicate (Axilla, Face, bikini, Upper limb and lower limb, brow, upper lip, neck, chest, back). Regarded as acceptable.
    BiocompatibilityCompliance with ISO 10993-1.Testing included Cytotoxicity, Sensitization, and Irritation for the laser output window. The evaluation was conducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1. Results are assumed to be compliant as the device was cleared.
    Electrical Safety & EMCCompliance with IEC 60601-1, IEC 60825-1, and IEC 60601-1-2.The system complies with IEC 60601-1 (safety), IEC 60825-1 (laser safety), and IEC 60601-1-2 (EMC).
    Software Verification & ValidationCompliance with FDA Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."Software V&V testing was conducted, and documentation was provided. The software was deemed a "moderate" level of concern. Results are assumed to be compliant as the device was cleared.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not report any clinical test set or data provenance for human subjects. The performance data presented are primarily non-clinical: biocompatibility testing, electrical safety and EMC testing, and software verification and validation. The claim of substantial equivalence is largely based on the device's technical specifications being similar to or safer than the predicate device.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as no clinical test set with human subject data requiring expert ground truth establishment is described in this 510(k) summary. The "ground truth" for this submission revolves around meeting engineering and biocompatibility standards.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as no clinical test set requiring adjudication is described.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This 510(k) is for a laser therapy system (hardware device), not an AI/software as a medical device (SaMD) that assists human readers.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study for an algorithm was not done. This device is a physical laser system for direct patient treatment, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context refers to established regulatory and engineering standards:

    • Biocompatibility: Adherence to ISO 10993-1 standards (e.g., negative results for cytotoxicity, sensitization, irritation).
    • Electrical Safety & EMC: Compliance with IEC 60601-1, IEC 60825-1, and IEC 60601-1-2 standards.
    • Software V&V: Adherence to FDA guidance for software in medical devices.
    • Substantial Equivalence: Comparison of technical specifications and intended use against a legally marketed predicate device.

    8. The sample size for the training set

    This information is not applicable as this submission does not describe an AI model with a training set. The "training" for this device would refer to its engineering design and validation against specified requirements.


    9. How the ground truth for the training set was established

    This information is not applicable as this submission does not describe an AI model with a training set. The "ground truth" for the device's development was established through engineering specifications, regulatory standards, and performance evaluation against those standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123435
    Date Cleared
    2013-02-11

    (96 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intense pulsed light therapy system is indicated for use in surgical, aesthetic and cosmetic application in the hair removal at a wavelength ranging from 690-1200nm.

    Device Description

    The IPL Therapy Systems is an Intense Pulsed Light Therapy System which delivers intense pulsed light at a wavelength ranging from 690nm -1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective photothermolysis. That is causing thermal damage to target chromophores by using light of appropriated wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin. IPL Therapy System is different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores. The Sincoheren IPL Therapy system controlled by computer is composed: - Main unit - LCD monitor - Treatment handle . The embedded microprocessor monitors can control the system continually. The operating Interface includes a full-LCD screen (8.0 inch), the key switch and emergency shut-off knob. LCD screen both shows the system work program, treatment parameters and number of light emitting. The treatment handle operated by hand comprises the intense emitter and transmission system.

    AI/ML Overview

    The provided document is a 510(k) summary for an Intense Pulsed Light Therapy System. It states that no clinical studies were required to support substantial equivalence. Therefore, it does not contain the information needed to answer many of the questions regarding acceptance criteria, study details, and performance.

    However, based on the information provided, here's what can be inferred or explicitly stated:

    1. Table of Acceptance Criteria and Reported Device Performance

    No specific quantitative acceptance criteria or reported device performance metrics are provided in this 510(k) summary. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting performance data from clinical trials.

    The "Indications for Use" section (page 6) does list "Density" (energy density in J/cm²) and "Program" settings for hair removal across different skin types (I-V, with N/A for VI). However, these are operational parameters, not performance metrics or acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. No clinical test set was used for this submission. The device did not require clinical studies.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. No clinical studies were performed, and thus no ground truth was established by experts for a test set.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. No clinical studies were done, so no MRMC study with or without AI assistance was performed.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    Not applicable. This device is a medical device (IPL system) operated by a human, not an AI algorithm. Its performance is tied to the operator and the device's physical parameters.

    7. Type of Ground Truth Used

    Not applicable. No clinical studies were performed. The "ground truth" for this 510(k) submission primarily relies on the equivalency of its non-clinical specifications and intended use to a legally marketed predicate device.

    8. Sample Size for the Training Set

    Not applicable. This is a hardware medical device; it does not involve machine learning or a "training set" in the computational sense. The "development" aspects mentioned (Risk Analysis, Requirements Reviews, Design Reviews, etc.) refer to engineering and quality assurance processes, not AI model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, there is no training set for this type of device.


    Summary of Device's Basis for Approval (from the document):

    The device's approval is based on:

    • Substantial Equivalence: The manufacturer asserts that the Intense Pulsed Light Therapy System is "as safe, as effective, and performance is substantially equivalent to the predicate device(s)" (K093627 IPLUSELIGHT IPL SYSTEM).
    • Non-Clinical Tests: Compliance with voluntary standards (IEC 60601-1, IEC 60601-1-2) and internal quality assurance measures (Risk Analysis, Requirements Reviews, Design Reviews, raw materials verification, unit/module verification, integration testing, final acceptance testing, performance testing, safety testing).
    • No Clinical Studies Required: "The subject of this premarket submission, Intense Pulsed Light Therapy System, did not require clinical studies to support substantial equivalence."