510(k) Data Aggregation

    K Number
    K032463

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-02-10

    (183 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CURAD® Silver Bandage is indicated for first aid to help in minor abrasions, cuts, lacerations, scrapes, and scalds.

    Device Description

    The CURAD® Silver Bandage is an adhesive bandage that includes silver in the wound pad. The bandage covers the wound to create a protected environment. At the same time, silver ions reduce bacterial growth in the wound pad.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the CURAD® Silver Bandage, which is an adhesive bandage containing silver in its wound pad. The purpose of the submission is to demonstrate substantial equivalence to legally marketed predicate devices.

    However, the provided document does not contain a detailed study proving the device meets specific acceptance criteria in the format requested. Instead, it refers to general performance testing.

    Here's a breakdown of the information presented, highlighting why a full acceptance criteria study cannot be extracted as requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states:

    Criteria CategoryDevice Performance
    Silver releaseAcceptable results
    Antimicrobial effectAcceptable results
    BiocompatibilitySatisfactory results
    • Missing Details: The document does not specify the quantitative or qualitative "acceptance criteria" for "acceptable" or "satisfactory" in terms of silver release, antimicrobial effect, or biocompatibility. It also does not provide the specific numerical results obtained during testing.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not provide information on the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective) for any of its performance testing.

    3. Number of Experts and Qualifications:

    The document does not provide information on the number of experts used to establish ground truth or their qualifications. This is not typically part of a 510(k) submission for this type of device, which focuses on device performance rather than interpretation of complex medical images or data by experts.

    4. Adjudication Method for the Test Set:

    The document does not provide information on any adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. This type of study is more relevant for diagnostic imaging devices where human interpretation is a key factor.

    6. Standalone (Algorithm Only) Performance:

    The document does not describe any standalone (algorithm only) performance. This device is a physical product, not an algorithm.

    7. Type of Ground Truth Used:

    For the performance testing mentioned (silver release, antimicrobial effect, biocompatibility), the "ground truth" would be established through laboratory testing and established scientific standards/methods for measuring these properties. The specific details of these methods are not provided in this 510(k) summary.

    8. Sample Size for the Training Set:

    The concept of a "training set" is not applicable to this physical medical device. This term is used for machine learning algorithms.

    9. How Ground Truth for Training Set Was Established:

    Not applicable, as there is no "training set" for this device.

    In summary, the provided 510(k) summary for the CURAD® Silver Bandage indicates that performance testing for silver release, antimicrobial effect, and biocompatibility was conducted and yielded "acceptable" or "satisfactory" results. However, it lacks the detailed quantitative acceptance criteria, specific study designs, sample sizes, and expert involvement that would be present in a comprehensive study report for certain types of medical devices, particularly those involving diagnostics or complex algorithms. The submission aims to demonstrate substantial equivalence to existing predicate devices based on technological characteristics and general performance, rather than providing detailed clinical trial data with expert agreement and specific statistical metrics often seen for higher-risk devices or software as a medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032462

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-02-10

    (183 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CURAD® Silver Gel Bandage is indicated for first aid to help in minor abrasions, cuts, lacerations, scrapes, and scalds.

    Device Description

    The CURAD® Silver Active Gel Bandage is a polyurethane occlusive bandage that includes silver oxide in the polyurethane matrix. The polyurethane provides a moist healing environment. At the same time, silver ions reduce bacterial growth in the wound pad.

    AI/ML Overview

    The provided text describes the 510(k) submission for the CURAD® Silver Active Gel Bandage. Here's an analysis of the acceptance criteria and the study conducted based on the given information:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Silver ReleaseAcceptable Results
    Antimicrobial EffectAcceptable Results
    BiocompatibilitySatisfactory Results

    Note: The document states that the CURAD® Silver Active Gel Bandage was tested for "silver release and antimicrobial effect and was demonstrated to have acceptable results." Additionally, a "biocompatibility assessment was performed... with satisfactory results." Specific quantitative thresholds for "acceptable" and "satisfactory" are not provided in the summary.

    Study Details

    1. Sample Size used for the test set and the data provenance:
      The document does not specify the sample size used for the performance tests (silver release, antimicrobial effect, and biocompatibility). It also does not mention the data provenance in terms of country of origin or whether the studies were retrospective or prospective.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This information is not applicable and not provided in the context of the performance testing for a wound dressing. Performance tests for silver release, antimicrobial effect, and biocompatibility typically involve laboratory assays and standardized protocols, not expert interpretation of diagnostic images or clinical data.

    3. Adjudication method for the test set:
      Not applicable (see point 2).

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This device is a wound dressing, not an AI-powered diagnostic tool requiring human reader studies.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable. This is not an algorithm-based device. The performance testing was for the intrinsic properties of the wound dressing.

    6. The type of ground truth used:
      The "ground truth" for this type of device would be established through laboratory measurements and standardized testing protocols for silver release, antimicrobial efficacy, and biocompatibility, as per relevant ISO standards or FDA guidance for wound dressings. The summary does not provide specific details on these protocols but states that results were "acceptable" and "satisfactory."

    7. The sample size for the training set:
      Not applicable. This device is not an AI/ML product developed with a training set.

    8. How the ground truth for the training set was established:
      Not applicable. This device is not an AI/ML product developed with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022645

    Validate with FDA (Live)

    Device Name
    CURAD
    Manufacturer
    Date Cleared
    2002-11-05

    (89 days)

    Product Code
    Regulation Number
    880.5090
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CURAD® Spray Bandage is indicated for providing a covering over minor cuts and scrapes that are clean and dry.

    Device Description

    The product is packaged in a pressurized can. The can is held about 5 inches from affected area, and a light coat is sprayed onto the skin. CURAD® Spray Bandage typically dries in 2-3 minutes. Once dried, additional coats can be applied, if needed.

    CURAD® Spray Bandage will typically adhere for several days before losing its adhesive strength. To remove it before it has lost its adhesive strength, the user can apply isopropyl alcohol to the bandage to dissolve it.

    CURAD® Spray Bandage is composed of the following ingredients:

    poly(methylacrylate-isobutene-monoisopropylmaleate), ethyl acetate, npentane, carbon dioxide, menthol

    AI/ML Overview

    Here's an analysis of the provided text regarding the CURAD® Spray Bandage's acceptance criteria and studies, organized as requested:

    Acceptance Criteria and Study for CURAD® Spray Bandage

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria ClaimedReported Device Performance
    FlexibleFlexible
    TransparentTransparent
    Water-tightWater-tight
    Allows passage of water vaporAllows passage of water vapor
    Seals out germsSeals out germs
    BiocompatibleBiocompatibility assessment performed with satisfactory results
    Adheres for several daysTypically adheres for several days
    Dries in 2-3 minutesTypically dries in 2-3 minutes

    2. Sample Size Used for the Test Set and Data Provenance

    The provided 510(k) summary does not explicitly state the sample size used for the test set for any of the performance claims. It also does not specify the data provenance (e.g., country of origin, retrospective/prospective). The document states that "A summary of testing to support each of these claims has been established and verified," implying that tests were conducted, but the details are omitted.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    The provided 510(k) summary does not mention the use of experts to establish ground truth for the test set. The performance claims appear to be based on physical/material property testing rather than expert-derived diagnoses or assessments.

    4. Adjudication Method for the Test Set

    Since the ground truth appears to be based on objective physical/material property testing rather than expert consensus on subjective assessments, an adjudication method for a test set is not applicable and not mentioned in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described in the provided document. The device in question is a liquid bandage, and the performance claims relate to its physical properties, not diagnostic accuracy where human readers would typically be involved.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This concept is not applicable to this device. The CURAD® Spray Bandage is a physical product, not a software algorithm that performs diagnostic or analytical tasks. Therefore, a "standalone algorithm performance" study was not conducted.

    7. Type of Ground Truth Used

    The ground truth used for the performance claims appears to be based on objective physical and material property measurements/testing. Examples include evaluating flexibility, transparency, water-tightness, water vapor permeability, germ sealing capability, drying time, adhesion duration, and biocompatibility.

    8. Sample Size for the Training Set

    There is no mention of a "training set" in the context of this device. A training set is typically associated with machine learning algorithms, which are not relevant to the CURAD® Spray Bandage.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set for this device, the question of how its ground truth was established is not applicable.

    Ask a Question

    Ask a specific question about this device