Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K954923

    Validate with FDA (Live)

    Date Cleared
    1996-11-21

    (392 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Qualitative culture and recovery of yeasts and fungi from blood

    Device Description

    BACTEC Fungal Medium is a growth medium providing an aerobic environment for the detection of yeast and fungi. It has been designed for blood volumes of four to 10 mL, and is used specifically with BACTEC non-radiometric (NR) series instruments in the monitoring of clinical blood specimens for the presence of yeasts and fungi. BACTEC Fungal medium has a higher weight/volume concentration of dextrose and sucrose then the predicate device to enhance the recovery and CO2 production of yeast. Saponin was added to the BACTEC Fungal Medium as a lysing agent, and ferric ammonium citrate was added as a source of iron to stimulate recovery of fungi other than yeast. Resin was removed from this medium. Principles for detection of CO2 produced by microorganisms metabolizing blood culture medium nutrients are identical to the predicate device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the BACTEC® Fungal Medium, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    The acceptance criteria are implicitly derived from the comparative study against predicate devices (BACTEC® PLUS 26 Culture Vials and ISOLATOR™ Microbial Tubes/ISOSTAT® Microbial System). The goal is to demonstrate that the BACTEC® Fungal Medium performs "as good or better than" or "equivalent to" these predicate devices for the recovery of yeasts and fungi from blood, particularly in terms of recovery rates and speed of detection.

    Acceptance Criteria CategorySpecific Criteria (Implicitly Derived)Reported Device Performance (BACTEC® Fungal Medium)
    Recovery of FungiVs. BACTEC® PLUS 26: Statistically significantly better recovery of all fungi combined and of Torulopsis glabrata. Vs. IS System: Equivalent recovery of fungi, with the exception of Histoplasma capsulatum. (i.e., performance should not be significantly worse for most fungi).Vs. BACTEC® PLUS 26: - 68 fungi isolates total. - 31 (45.6%) recovered in both media. - 11 (16.2%) recovered in BACTEC® PLUS 26 only. - 26 (38.2%) recovered in BACTEC® Fungal Medium only. - Recovery of all fungi combined and of Torulopsis glabrata was statistically significantly better (p < 0.025) in BACTEC® Fungal Medium. Vs. IS System: - 93 fungi isolates total. - 42 (45.2%) recovered in both media. - 34 (36.6%) recovered in IS System only. - 17 (18.3%) recovered in BACTEC® Fungal Medium only. - Excluding H. capsulatum (which was better in IS System), 17 fungi were isolated only in BACTEC® Fungal Medium and nine fungi were isolated only in the IS System; this difference is not statistically significant.
    Speed of Detection (Fungi)Vs. BACTEC® PLUS 26: No statistically significant difference in speed of detection for any or all fungi combined. Vs. IS System: No statistically significant difference in speed of detection for all fungi combined; potentially earlier detection for specific fungi (T. glabrata).Vs. BACTEC® PLUS 26: - No statistically significant difference in the speed of detection of any of the fungi in the two media, nor of all fungi combined. Vs. IS System: - T. glabrata was detected significantly earlier in BACTEC® Fungal Medium (p < 0.05). - The difference in speed of detection of all fungi in each of the two media was not significantly significant.
    Recovery of Bacteria(Implicitly, not the primary focus of the fungal medium, but generally should not be significantly worse if present in clinical specimens).Vs. BACTEC® PLUS 26: 161 bacteria isolates recovered (implicit understanding that the primary focus of the new medium is fungal recovery, not improved bacterial recovery compared to a general culture medium). Vs. IS System: 125 bacteria isolates recovered.

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • BACTEC® Fungal Medium vs. BACTEC® PLUS 26:
        • Sample Size: 4886 adequately filled paired vials.
        • Data Provenance: Clinical study conducted at four university-affiliated hospitals. This indicates a prospective collection of real-world clinical samples. The country of origin is not explicitly stated but is implied to be the US given the submission to the FDA.
      • BACTEC® Fungal Medium vs. ISOLATOR™ Microbial Tubes/ISOSTAT® Microbial System:
        • Sample Size: 4907 adequately filled paired vials and tubes.
        • Data Provenance: Clinical study conducted at four university-affiliated hospitals. This indicates a prospective collection of real-world clinical samples. The country of origin is not explicitly stated but is implied to be the US.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the summary. The "ground truth" for recovery is based on the actual growth and identification of microorganisms from the culture media, which would typically be performed by trained laboratory personnel (microbiologists/medical technologists). The summary does not specify the number or qualifications of these individuals.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not explicitly stated. For microbiology studies, "adjudication" in the traditional sense of multiple readers reviewing images isn't directly applicable. Instead, the results are likely determined by standard laboratory protocols for organism identification and confirmation.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, this was not an MRMC comparative effectiveness study involving human readers and AI assistance. This study evaluates the performance of different microbial culture media, an in vitro diagnostic device, not an AI-powered diagnostic imaging system.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in a sense, this was a "standalone" evaluation of the media. The BACTEC instrument automatically detects CO2 production. The study evaluates the medium's ability to support growth detectable by the instrument, without human intervention in the primary detection phase, although humans perform subsequent identification of the isolates.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth was established by microbiological culture and identification of isolates recovered from the blood samples. This is considered the gold standard for detecting and identifying microorganisms in blood cultures. "Clinically significant isolates" implies a determination by a clinician or microbiologist that the isolated organism is relevant to the patient's condition.
    7. The sample size for the training set:

      • This information is not explicitly provided. The study describes "clinical performance" testing, which typically refers to the final validation set for regulatory submission. Details of earlier development or optimization ("training") sample sizes for the medium formulation are not usually included in a 510(k) summary. The formulation of the medium itself is a scientific development process, not an AI training process.
    8. How the ground truth for the training set was established:

      • As with the training set, details on how ground truth was established during the development or "training" phase of the medium formulation are not in this summary. However, it would similarly have involved microbiological culture and identification methods to assess the growth characteristics of various target microorganisms in different medium formulations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K954932

    Validate with FDA (Live)

    Date Cleared
    1996-08-21

    (299 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BBL® MGIT™ Mycobacteria Growth Indicator Tube and BBL® MGIT™ supplemented with BBL® MGIT™ OADC and BBL® MGIT™ supplemented with BBL® MGIT™ PANTA antibiotic mixture, is intended for the detection and recovery of mycobacteria from decontaminated clinical specimens (except urine) and sterile body fluids (except blood). Acceptable clinical specimen types are digested and decontaminated.

    Device Description

    The BBL® MGIT™ Mycobacteria Growth Indicator Tube contains 4 mL of modified Middlebrook 7H9 Broth Base. The complete medium is supplemented with 0.5 mL OADC enrichment and 0.1 mL of PANTA antibiotic mixture, when necessary. The medium components are substances essential for the growth of mycobacteria. A fluorescent compound is embedded in the bottom of the BBL® MGIT™ tube which is sensitive to the presence of oxygen dissolved in the broth. Specimens are inoculated (0.5 mL) into the BBL® MGIT™ tube and incubated at the appropriate temperature. Tubes are read daily from the second day of inoculation. Actively respiring microorganisms consume the oxygen and allow the compound to fluoresce. Each inoculated BBL® MGIT™ tube is compared to the fluorescence Positive Control and Negative Control tube to assist in the interpretation of a positive signal from the BBL® MGIT™ tube. A BBL® MGIT™ tube which exhibits fluorescence comparable to the Positive Control tube is considered presumptively positive for growth.

    AI/ML Overview

    The provided text describes the BBL® MGIT™ Mycobacteria Growth Indicator Tube, intended for the detection and recovery of mycobacteria from clinical specimens. The device relies on an oxygen-sensitive fluorescent sensor that detects oxygen consumption by actively respiring microorganisms.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity, specificity, or false positive rates. However, the study aims to show comparable performance to predicate devices. Therefore, the "acceptance criteria" appear to be implicit: the BBL® MGIT™ tubes should perform similarly to or better than the BACTEC® 460TB System and the BBL® SEPTI-CHEK® AFB Mycobacteria Culture System in terms of recovery rate and time to detection, while maintaining an acceptable false positive rate.

    MetricAcceptance Criteria (Implicit: Comparable to Predicate)Reported BBL® MGIT™ Performance
    Mycobacterial Recovery RateComparable to BACTEC® 460TB System (79%) and BBL® SEPTI-CHEK® AFB (79%)77% (253 of 330 isolates)
    False Positive RateAcceptably low (no explicit numerical criterion, but likely aiming for minimal
                                   false positives relative to clinical impact) | 0.5%                            |
    

    | Time-to-Detection | Comparable to predicate devices (not explicitly stated, but implies being efficient | 14.1 days (average for all mycobacteria) |
    | Unrecovered Isolates | Minimally higher than predicate devices (with further recovery via recommended second medium) | 3.7% (isolates recovered by reference, but not MGIT) |
    | Breakthrough Contamination Rate | Acceptably low | 9.7% |

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: A combined total of 2801 clinical specimens were evaluated.
    • Data Provenance:
      • Country of origin: Not explicitly mentioned, but the submitting company is Becton Dickinson Microbiology Systems, located in Sparks, MD, USA. It's highly probable the studies were conducted in the USA.
      • Retrospective or Prospective: Not explicitly stated. However, external evaluations are typically conducted prospectively to compare a new device against existing methods in real-world clinical settings. The description of "specimens were evaluated" and "isolates recovered during the study" suggests a prospective collection and testing over a period.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not specify the number of experts or their qualifications for establishing the ground truth. The ground truth was established by culture results from reference methods:

    • BACTEC® 460TB System
    • BBL® SEPTI-CHEK® AFB Mycobacteria Culture System
    • Conventional solid agar media

    Presumably, trained laboratory personnel followed standard protocols for these reference systems to determine the presence or absence of mycobacteria and to identify isolates.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe an adjudication method involving human experts. The ground truth for mycobacterial presence was determined directly by the growth observed in the reference culture systems. If there was discordance between the BBL® MGIT™ and the reference, it was noted as an unrecovered isolate by the MGIT, not subjected to a separate expert adjudication process to determine a "true" positive or negative outside of the reference method's result.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, this was not a multi-reader, multi-case (MRMC) comparative effectiveness study.
    • AI (Artificial Intelligence): The BBL® MGIT™ device is a culture-based diagnostic system, not an AI-assisted diagnostic tool. Therefore, there's no discussion of AI assistance or its effect size on human readers. The growth detection is described as "Manual observation of fluorescence via excitation with longwave UV light."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The BBL® MGIT™ device itself is a standalone diagnostic tool in the sense that its detection mechanism (fluorescence) is independent of human interpretation for the initial signal. However, the final interpretation of a "presumptively positive" signal and subsequent identification of the mycobacteria would involve human intervention. The "manual observation of fluorescence" indicates human involvement in reading the fluorescence.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used was culture-based detection and recovery of mycobacteria using established reference methods:

    • BACTEC® 460TB System
    • BBL® SEPTI-CHEK® AFB Mycobacteria Culture System
    • Conventional solid agar media

    The presence of mycobacteria, and their specific isolates, as determined by growth in one or more of these reference systems, formed the ground truth against which the BBL® MGIT™ performance was compared.

    8. The sample size for the training set

    The document does not specify a separate "training set" in the context of machine learning or AI models. This device is a culture medium system, not a software algorithm that requires training data in that sense. The "internal testing" mentioned likely involved R&D and optimization of the medium components and detection method, rather than training a predictive model.

    9. How the ground truth for the training set was established

    As there was no explicit "training set" for an algorithm, the concept of establishing ground truth for it doesn't apply directly in this context. The "internal testing" would have relied on known mycobacterial strains and clinical samples where the presence/absence of mycobacteria was confirmed by standard laboratory methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962210

    Validate with FDA (Live)

    Date Cleared
    1996-08-08

    (59 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A non-radiometric automated blood culturing system which is designed for the monitoring and detection of clinical cultures of blood for the presence of microorganisms.

    Device Description

    BACTEC® 9050 System is a non-radiometric automated blood culturing system. It has a capacity of 50 specimens, uses Positivity Algorithms by Media Type for culture analysis, has agitation on during testing, uses a Florescence Detector (LED) for detection optics, incubates at 35°C ± 1.5°C, has a protocol length of 4 to 7 Days, uses Florescent Series Media, tests specimens every 10 Minutes, uses Low Voltage DC power supplies, and displays 1 rack (Rotor). Its dimensions are 24"W x 28.5"H x 25.5"D.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and study details:

    Acceptance Criteria and Device Performance

    The provided text focuses on demonstrating substantial equivalence to a predicate device rather than defining specific numerical acceptance criteria for a novel performance threshold. The core acceptance criterion appears to be that the BACTEC 9050 performs "equivalent" to the BACTEC 9240.

    Acceptance CriteriaReported Device Performance (BACTEC 9050 vs. BACTEC 9240)
    Recovery rate of microorganisms for clinically relevant organisms using available media and recommended blood volumes.Difference in recovery: 0.2%
    Mean time to detection of microorganisms.Mean time to detection: Equal (<0.01 hours difference)
    Overall performance compared to predicate device (BACTEC 9240).Performed "equivalent" to the BACTEC 9240.

    Study Details

    • Sample Size for Test Set: 485 paired seeded vials.
    • Data Provenance: Not explicitly stated, but given it's a non-clinical test using "seeded vials" (likely laboratory-prepared cultures), it's a controlled laboratory setting rather than patient-derived data. The country of origin is not mentioned. It is a prospective, non-clinical test.
    • Number of Experts Used to Establish Ground Truth for Test Set and Qualifications: Not applicable. The "ground truth" for these non-clinical tests would have been the known quantity and type of microorganisms seeded into the vials. There's no mention of expert consensus or interpretation required for this.
    • Adjudication Method: Not applicable. As the ground truth is based on known seeded organisms, no expert adjudication process is described or needed.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No. This study is a technical performance comparison between two automated blood culture instruments in a non-clinical setting, not an evaluation of human reader performance with or without AI assistance.
    • Standalone Performance (Algorithm Only): Yes, this study evaluates the standalone performance of the BACTEC 9050 system's ability to detect microbial growth compared to the BACTEC 9240. There isn't a human-in-the-loop component described.
    • Type of Ground Truth Used: "Known seeded organisms." The study involved using 23 specific organisms chosen for clinical relevance, implying that the presence and identity of these organisms in the seeded vials constituted the ground truth.
    • Sample Size for Training Set: Not applicable. This document describes a non-clinical validation study comparing a new device to a predicate, not the development or training of an AI algorithm.
    • How Ground Truth for Training Set was Established: Not applicable. (See above).
    Ask a Question

    Ask a specific question about this device

    K Number
    K946268

    Validate with FDA (Live)

    Date Cleared
    1996-05-21

    (515 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Detection of mycobacteria in clinical respiratory specimens

    Device Description

    BACTEC® 9000MB: 7H9 Middlebrook broth base with nutrient additives, Fluorescence sensor in silicon rubber base, Fluorescent detection of O₂ consumption by mycobacterial growth, On-board incubation at 37°C ± 1.5°C; internal instrument agitation every 10 minutes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BACTEC® 9000MB, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided text does not explicitly state "acceptance criteria" in a formal, quantifiable manner for overall device performance. Instead, it compares the BACTEC® 9000MB to other methods (BACTEC® 460TB and Conventional Media) across various technical specifications and time to detection.

    However, based on the comparative tables, we can infer some key performance aspects of the BACTEC® 9000MB and frame them as its reported performance relative to the benchmarks. The "acceptance criteria" for this device would likely be implicitly tied to demonstrating comparable or improved performance over existing methods.

    Feature / Performance MetricAcceptance Criteria (Inferred from comparison)Reported Device Performance (BACTEC® 9000MB)
    Primary Sample TypeRespiratory specimens (similar to comparators)Respiratory
    Growth Medium7H9 Middlebrook broth base withnutrient additives (a specific formulation)7H9 Middlebrook broth base withnutrient additives
    Growth Detection MethodFluorescence sensor for O₂ consumption (new method compared to comparators)Fluorescent detection of O₂consumption by mycobacterialgrowth
    Antimicrobial SupplementPANTA (similar to BACTEC® 460TB)Polymixin B, amphotericin B,nalidixic acid, trimethoprim &azlocillin (PANTA)
    Incubation TemperatureOn-board incubation at 37°C (with internal agitation)On-board incubation at 37°C ±1.5°C; internal instrument agitationevery 10 minutes.
    Time to Detection (TTD)Faster than Conventional Media (21-49 days); Comparable to BACTEC® 460TB (7-12 days) or slightly longer, but within a reasonable range.Eleven to fifteen days

    Important Note: The current document is a summary of safety and efficacy and focuses on describing the device and comparing its technical specifications and typical performance to other methods. It does not contain a dedicated section outlining pre-established, quantitative "acceptance criteria" that the device was then tested against in a formal study. The "acceptance" is implied by demonstrating a viable alternative with certain benefits (e.g., non-radiometric, faster than conventional).


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the given text.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the given text.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the given text.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The BACTEC® 9000MB is a diagnostic instrument for detecting mycobacteria using fluorescence, not an AI-assisted image interpretation tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable to this device.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, a standalone study of the BACTEC® 9000MB performance was implicitly done. The document describes the technical specifications and "Time to detection (TTD)" of the device itself, comparing it directly to other methods (BACTEC® 460TB and Conventional Media), not human performance. The TTD of 11-15 days is a direct measurement of the device's performance.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance comparisons (e.g., Time to Detection) would have been established by confirming the presence or absence of mycobacterial growth. Given the nature of microbiology diagnostics, this would typically involve:

    • Microbiological Culture Confirmation: Positive cultures on other standard media, possibly with further identification techniques.
    • Microscopy: Observation of acid-fast bacilli (AFB) in smears.
    • Clinical Outcomes/Patient Diagnosis (less likely for TTD, but relevant for overall efficacy): Correlation with the patient's clinical diagnosis of mycobacterial infection.

    The text does not explicitly state the specific ground truth method used, but it would undeniably rely on established microbiological techniques for identifying mycobacteria.


    8. The sample size for the training set

    This information is not applicable as the BACTEC® 9000MB is a laboratory diagnostic instrument, not a machine learning algorithm that requires a "training set" in the context of AI. Its "training" is in its engineering and calibration.


    9. How the ground truth for the training set was established

    This information is not applicable for the reasons stated above (not an AI algorithm).

    Ask a Question

    Ask a specific question about this device

    K Number
    K960827

    Validate with FDA (Live)

    Date Cleared
    1996-05-17

    (78 days)

    Product Code
    Regulation Number
    864.5220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QBC® AUTOREAD™ System provides a diagnostic hematology profile on venous or capillary blood: hematocrit, hemoglobin, MCHC, platelet count, white blood cell count, granulocyte count (% and number), and lymphocyte/monocyte count (% and number).

    Device Description

    The QBC® AUTOREAD™ System includes an analyzer, power pack, printer, centrifuge, and test accessories. The system uses specially designed tubes known as the QBC® AccuTube. AccuTubes are filled with blood and centrifuged. The spun tube is placed in the QBC® AUTOREAD™ analyzer which provides a report.

    AI/ML Overview

    The provided text describes a 510(k) summary for the QBC® AUTOREAD™ System, focusing on its equivalency to a predicate device. Here's an analysis of the acceptance criteria and the supporting study, based on the information given:

    Acceptance Criteria and Reported Device Performance

    The core of the acceptance criteria for this 510(k) submission is demonstrating equivalency to the predicate device and established methods for hematology parameters. The "reported device performance" is essentially the statistical measures (correlation coefficient, slope, intercept) derived from the equivalency studies.

    ParameterAcceptance Criteria (Implicit: Strong Correlation and Slope ~1, Intercept ~0 when compared to predicate/reference)Reported Device Performance (Correlation Coefficient, Slope, Intercept) - Against Reference or Predicate
    HematocritEquivalent to spun microhematocrit methodAgainst Spun Microhematocrit:Correlation Coefficient: 0.993Slope: 0.991Intercept: 0.270
    Equivalent to original QBC AccuTube/AUTOREAD SystemAgainst Original QBC AccuTube/AUTOREAD:Correlation Coefficient: 0.999Slope: 0.975Intercept: 1.015
    HemoglobinEquivalent to impedance cell counter methodAgainst Impedance System:Correlation Coefficient: 0.995Slope: 0.969Intercept: 0.410
    Equivalent to original QBC AccuTube/AUTOREAD SystemAgainst Original QBC AccuTube/AUTOREAD:Correlation Coefficient: 0.999Slope: 0.981Intercept: 0.239
    PlateletsEquivalent to original QBC AccuTube/AUTOREAD SystemAgainst Original QBC AccuTube/AUTOREAD:Correlation Coefficient: 0.989Slope: 0.932Intercept: 11.633
    White Blood CellsEquivalent to original QBC AccuTube/AUTOREAD SystemAgainst Original QBC AccuTube/AUTOREAD:Correlation Coefficient: 0.984Slope: 1.031Intercept: -0.116
    Granulocytes (#)Equivalent to original QBC AccuTube/AUTOREAD SystemAgainst Original QBC AccuTube/AUTOREAD:Correlation Coefficient: 0.985Slope: 1.045Intercept: 0.002
    Lymphocytes/Monocytes (#)Equivalent to original QBC AccuTube/AUTOREAD SystemAgainst Original QBC AccuTube/AUTOREAD:Correlation Coefficient: 0.967Slope: 0.921Intercept: 0.080

    Study Details

    Regarding the study that proves the device meets the acceptance criteria:

    1. Sample size used for the test set and the data provenance:

      • Hematocrit and Hemoglobin equivalency study: 160 fresh EDTA whole blood samples.
      • Data Provenance: The study was conducted at "two hospital sites". No specific country of origin is mentioned, but the company address is in Sparks, Maryland, USA, suggesting the data is likely from the United States. The samples were "fresh EDTA whole blood samples," indicating they were retrospectively collected from patients.
      • Other Parameters (Platelets, WBC, Granulocytes, Lymphocytes/Monocytes): The sample size for these parameters is not explicitly stated, but it's implied to be the same set of samples or a comparable set used for the hematocrit and hemoglobin equivalency against the original QBC AccuTube.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This type of device (automated differential cell counter) does not typically rely on human expert consensus for "ground truth" in the same way as image-based diagnostics. The ground truth here is established by reference methods (e.g., spun microhematocrit method, impedance cell counter, or the established predicate device).
      • Therefore, the concept of "number of experts" and "qualifications" doesn't directly apply here for establishing the ground truth, as it's a quantitative measurement device.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • No adjudication method is mentioned or relevant for this type of quantitative device study. Measurements are taken and compared directly to reference methods.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an MRMC comparative effectiveness study. It is a study to demonstrate the analytical equivalency of a diagnostic device to predicate devices and established reference methods. Human reader performance or AI assistance are not relevant to this type of submission.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, effectively. The "device" in question is an automated system (QBC® AUTOREAD™ System) which includes an analyzer and algorithms to generate results. The studies described evaluate the performance of this system (including its modified algorithm) in a standalone capacity against reference methods and the predicate device. There is no human interpretation component being evaluated in these specific studies; it's the machine's output.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Reference Methods:
        • For Hematocrit: Spun microhematocrit method.
        • For Hemoglobin: Impedance cell counter method.
        • For all parameters (Hematocrit, Hemoglobin, Platelets, White Blood Cells, Granulocytes, Lymphocytes/Monocytes): The original QBC AccuTube read with the original QBC AUTOREAD algorithm (the predicate device).
    7. The sample size for the training set:

      • The document does not explicitly mention a separate "training set" or its size. Since this is a 510(k) summary for a device with modified components (AccuTube and algorithms), the modifications were likely developed and refined using internal data, but the submission focuses on the validation or test data to demonstrate equivalency. It is possible that the algorithm adjustments were "trained" on some data, but this information is not provided in the summary.
    8. How the ground truth for the training set was established:

      • Given the lack of information on a "training set," there's no detail on how ground truth for such a set would have been established. However, for internal development, it would typically involve comparing preliminary device results against similar reference methods as used in the validation studies.

    Conclusion: The studies described successfully demonstrate the equivalency of the modified QBC® AUTOREAD™ System to established reference methods and its predicate device for the measured hematology parameters, addressing the implicit acceptance criteria for this 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K954925

    Validate with FDA (Live)

    Date Cleared
    1996-02-16

    (113 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K954921

    Validate with FDA (Live)

    Date Cleared
    1996-02-16

    (113 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K954927

    Validate with FDA (Live)

    Date Cleared
    1996-02-16

    (113 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K954930

    Validate with FDA (Live)

    Date Cleared
    1996-02-16

    (113 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K954924

    Validate with FDA (Live)

    Date Cleared
    1996-02-08

    (105 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A qualitative procedure for the culture and recovery of aerobic microorganisms (mainly bacteria and fungi) from pediatric and other blood specimens which are generally less than 3 mL in volume.

    Device Description

    BACTEC PEDS PLUS medium is a bacterial growth medium providing an aerobic environment for the detection of bacteria and fungi. It has been designed for the recovery of microorganisms from pediatric and other blood specimens which are generally less than 3 mL in volume and is used specifically with the BACTEC NR ( non-radiometric )blood culture series instruments. Each glass vial of medium contains 20 mL of enriched Soybean-Casein Digest Broth with resins and CO2. If microorganisms are present in the test sample inoculated into the BACTEC vial, CO2 will be produced and liberated into the atmosphere when the organisms metabolize the substrates present in the vial. The instrument analyzes the vial head space gas for CO2 and, If a threshold level is exceeded, indicates that the vial is positive, i.e., that the test sample contains viable organisms.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the BACTEC® PEDS PLUS™ Culture Vials, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit)Reported Device Performance (BACTEC PEDS PLUS)
    Detects clinically significant organismsOverall Recovery: 256 organisms detected (181 from both + 75 from PEDS PLUS only) out of 289 total clinically significant organisms (88.6%).
    Better or equivalent recovery of organisms compared to predicate device (BACTEC NR6A)Significantly better (p<0.001) with 75 organisms detected only in PEDS PLUS vs. 33 in NR6A.
    Shorter or equivalent time to detection compared to predicate deviceSignificantly better (p<0.001) with 39 organisms detected earlier in PEDS PLUS vs. 13 in NR6A.
    Improved recovery from patients receiving antimicrobial therapySignificantly better (p<0.001) with 45 organisms detected only in PEDS PLUS vs. 16 in NR6A.
    Qualitative procedure for culture and recovery of aerobic microorganisms (bacteria and fungi)Qualitatively detects CO2 produced by microorganisms, indicating presence of viable organisms.
    Suitable for pediatric and other blood specimens generally < 3 mLDesigned for this purpose as per intended use.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: 4,581 paired vials (meaning 4,581 BACTEC PEDS PLUS vials and 4,581 BACTEC NR6A vials were tested, for a total of 9,162 vials).
      • Data Provenance: Prospective, clinical studies conducted at three tertiary care pediatric hospitals. The country of origin is not specified, but given the submission is to the FDA, it's likely the US.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the text. The ground truth appears to be based on the isolation and identification of "clinically significant organisms," which implies laboratory culture and identification methods, rather than expert interpretation of results.
    3. Adjudication method for the test set:

      • This information is not provided in the text.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This was not an MRMC study and does not involve AI. It is a comparison of two culture media.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in essence. The "device" in this context is primarily the BACTEC PEDS PLUS medium and its performance in detecting growth, measured by the BACTEC NR series instruments. The detection process itself (sensing CO2) is automated by the instrument, and the results (positive/negative, time to detection) are generated without direct human interpretation of a visual image, for example. The "algorithm" here is the instrument's CO2 detection threshold.
    6. The type of ground truth used:

      • The ground truth was established by the isolation and identification of "clinically significant organisms" from the blood cultures. This implies standard microbiological culture and identification techniques, which is a form of laboratory-confirmed outcome/clinical outcome.
    7. The sample size for the training set:

      • Not applicable. This study is a clinical comparison of two culture media, not an AI model requiring a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. This study is a clinical comparison of two culture media, not an AI model requiring a ground truth for a training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2