510(k) Data Aggregation

    K Number
    K243343

    Validate with FDA (Live)

    Device Name
    BD CTGCTV2
    Date Cleared
    2025-04-22

    (179 days)

    Product Code
    Regulation Number
    866.3393
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD CTGCTV2 assay incorporates automated DNA extraction and real-time polymerase chain reaction (PCR) for the direct, qualitative detection of DNA from:

    • Chlamydia trachomatis (CT)
    • Neisseria gonorrhoeae (GC)
    • Trichomonas vaginalis (TV)

    The assay may be used for detection of CT, GC and/or TV DNA in patient- or clinician-collected vaginal swab specimens (in a clinical setting) and male and female urine specimens. The assay may also be used for the detection of CT and GC DNA in endocervical swab and Liquid-Based Cytology (LBC) specimens in ThinPrep® PreservCyt® Solution using an aliquot that is removed prior to processing for the ThinPrep® Pap test. The assay may also be used for the detection of CT and GC DNA in clinician-collected rectal and oropharyngeal swab specimens.

    The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial, gonococcal, and/or trichomoniasis urogenital disease and chlamydial and gonococcal extragenital infection.

    The BD CTGCTV2 assay is available for use with the BD MAX™ System (urogenital specimens) or the BD COR™ System (urogenital and extragenital specimens), as described above.

    Device Description

    The BD CTGCTV2 assay, performed on the BD COR™ System (hereafter referred to as BD CTGCTV2), is designed for use with the applicable BD Molecular specimen collection and transport devices for male and female urine, rectal swabs, oropharyngeal swabs, vaginal swabs, endocervical swabs, and LBC specimens (PreservCyt®). Specimens are collected and transported to the testing laboratory using their respective transport devices under conditions of time and temperature that have been determined to maintain the integrity of the target nucleic acids.

    The BD COR™ MX Instrument, when combined with the BD COR™ PX Instrument, is to be used for automated sample preparation, extraction, and purification of nucleic acids from multiple specimen types, as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based real-time PCR for simultaneous and differential detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis.

    The BD CTGCTV2 assay extraction reagents are dried in 96-well microtiter plates that contain binding magnetic affinity beads and Sample Processing Control (SPC). Each tube is capable of binding and eluting sample nucleic acids. The SPC monitors the integrity of the reagents and the process steps involved in DNA extraction, amplification and detection, as well as for the presence of potential assay inhibitors.

    The BD CTGCTV2 assay liquid reagent plate includes Wash, Elution and Neutralization buffers. The beads (described above), together with the bound nucleic acids, are washed and the nucleic acids are eluted by a combination of heat and pH. When performed on BD COR™ System, there is an additional buffer to rehydrate the dried extraction mix. Eluted DNA is neutralized and transferred to the Amplification reagent (described below) to rehydrate the PCR reagents. After reconstitution, the BD COR™ System dispenses a fixed volume of PCR-ready solution containing extracted nucleic acids into the BD PCR Cartridge. Microvalves in the BD PCR Cartridge are sealed by the system prior to initiating PCR in order to contain the amplification mixture and thus prevent evaporation and contamination.

    The BD CTGCTV2 assay is comprised of two targets for Chlamydia trachomatis (detected on the same optical channel), two targets for Neisseria gonorrhoeae (detected on two different optical channels) and one target for Trichomonas vaginalis (detected on one optical channel). Only one Chlamydia trachomatis target is required to be positive in order to report a positive result. Both Neisseria gonorrhoeae targets are required to be positive in order to report a positive result.

    The amplified DNA targets are detected using hydrolysis (TaqMan®) probes, labeled at one end with a fluorescent reporter dye (fluorophore), and at the other end, with a quencher moiety. Probes labeled with different fluorophores are used to detect the target analytes in different optical channels of the BD COR™ System. When the probes are in their native state, the fluorescence of the fluorophore is quenched due to its proximity to the quencher. However, in the presence of target DNA, the probes hybridize to their complementary sequences and are hydrolyzed by the 5'-3' exonuclease activity of the DNA polymerase as it synthesizes the nascent strand along the DNA template. As a result, the fluorophores are separated from the quencher molecules and fluorescence is emitted. The BD COR™ System monitors these signals at each cycle of the PCR and interprets the data at the end of the reaction to provide qualitative test results for each analyte (i.e., positive or negative).

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the BD CTGCTV2 assay detail its performance in detecting Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) in extragenital specimens (rectal and oropharyngeal swabs).

    Here's an analysis based on your request:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the BD CTGCTV2 assay are implicitly demonstrated through its clinical performance results, where the assay's sensitivity (Positive Percent Agreement - PPA) and specificity (Negative Percent Agreement - NPA) for extragenital specimens are compared against a Composite Comparator Algorithm (CCA). The FDA's clearance indicates that these performance metrics met the necessary standards for substantial equivalence.

    Table of Acceptance Criteria and Reported Device Performance:

    While explicit numerical acceptance criteria (e.g., "PPA must be >= X%") are not directly stated in the provided text, the reported performance measures are the ones that met the FDA's requirements for clearance.

    MetricTarget/Condition (Implicit Acceptance Criteria)Reported Device Performance (BD CTGCTV2)
    Chlamydia trachomatis (CT) - Oropharyngeal
    Sensitivity (PPA)Sufficiently high for diagnostic use100% (86.2–100% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.8% (99.5–99.9% CI)
    Neisseria gonorrhoeae (GC) - Oropharyngeal
    Sensitivity (PPA)Sufficiently high for diagnostic use92.8% (85.8–96.5% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.5% (99.1–99.7% CI)
    Chlamydia trachomatis (CT) - Rectal
    Sensitivity (PPA)Sufficiently high for diagnostic use97.7% (93.5–99.2% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.4% (99.0–99.7% CI)
    Neisseria gonorrhoeae (GC) - Rectal
    Sensitivity (PPA)Sufficiently high for diagnostic use95.8% (89.7–98.4% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.8% (99.5–99.9% CI)
    Non-Reportable Rate (Total CT and GC)Reasonably low for clinical utility (e.g., <5%)0.6% (0.4–0.9% CI) initial test

    Study that Proves the Device Meets the Acceptance Criteria:

    The key study proving the device meets the acceptance criteria is the Clinical Performance Study for Extragenital Specimens.

    1. Sample Size Used for the Test Set and Data Provenance:

    • Total Subjects Consented: 2,439
    • Eligible Subjects: 2,375
    • Total Eligible Specimens (test set): 4,652 (oropharyngeal and rectal swabs, considered for randomization)
    • Eligible Specimens for Testing: 4,579 (2,303 oropharyngeal swabs and 2,276 rectal swabs)
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the mention of "eight geographically diverse sites" suggests a multi-site clinical study, likely within the US, given FDA clearance.
      • Retrospective or Prospective: Prospective, as specimens were collected from subjects enrolled in the study. The study involved consenting subjects and collecting specimens specifically for this evaluation.

    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not provided in the document.
    • The ground truth was established by a Composite Comparator Algorithm (CCA), not direct human expert interpretation of images or other subjective data.

    3. Adjudication Method for the Test Set:

    • The adjudication method used for the ground truth (CCA) was based on the concurrence of at least 2 out of 3 commercially available NAAT assays. This means it was a "2 out of 3" agreement model for establishing the positive or negative status of a specimen.

    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a fully automated in vitro diagnostic (IVD) assay that detects nucleic acids, not an AI-assisted imaging device requiring human interpretation. Therefore, the concept of "human readers improving with AI assistance" does not apply.

    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone performance study. The BD CTGCTV2 assay is an automated system (BD COR™ System) that performs sample preparation, extraction, amplification, and detection. Its performance "without human-in-the-loop" is precisely what the clinical performance study measures, specifically in comparison to a CCA, which also represents an objective, algorithm-based comparator.

    6. The Type of Ground Truth Used:

    • The ground truth used was a Composite Comparator Algorithm (CCA), consisting of results from three commercially available Nucleic Acid Amplification Tests (NAATs).

    7. The Sample Size for the Training Set:

    • The document does not specify the sample size for the training set. The provided clinical performance data (4,579 specimens) pertains to the test set used for validation. Given that this is an IVD assay, the "training" analogous to machine learning would relate to the development and optimization of the assay's reagents, protocols, and analytical parameters, rather than a distinct "training set" of patient data in the typical sense of AI/ML model development.

    8. How the Ground Truth for the Training Set Was Established:

    • As the document does not explicitly discuss a "training set" in the context of machine learning, it also does not describe how a ground truth for such a set was established. For an IVD assay, the "ground truth" during development (analogous to training) would typically involve well-characterized reference strains, spiked samples with known concentrations, and analytical validation studies to optimize assay parameters, which are alluded to in the "Analytical Performance Evaluation" section (e.g., LoD, Inclusivity, Cross-Reactivity, Microbial Interference).
    Ask a Question

    Ask a specific question about this device

    K Number
    K230956

    Validate with FDA (Live)

    Date Cleared
    2023-07-31

    (118 days)

    Product Code
    Regulation Number
    866.3981
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD Respiratory Viral Panel for BD MAX™ System is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT- PCR) test intended for the simultaneous, qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) nucleic acid in nasopharyngeal swab (NPS) and anterior nasal swab (ANS) specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-COV-2, influenza, and RSV can be similar.

    BD Respiratory Viral Panel for BD MAX™ System is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and/or RSV infection if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV- 2, influenza A, influenza B, and RSV viral nucleic acid are generally detectable in NPS and ANS specimens during the acute phase of infection.

    Positive results do not rule out co-infection with other organisms. The agent(s) detected by the BD Respiratory Viral Panel for BD MAX™ System may not be the definitive cause of disease.

    Negative results do not preclude SARS-CoV-2, influenza B, and/or RSV infection.

    The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    BD Respiratory Viral Panel-SCV2 for BD MAX™ System:

    BD Respiratory Viral Panel-SCV2 for BD MAX™ System is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for the simultaneous, qualitative detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) viral nucleic acid in nasopharyngeal swab (NPS) and anterior nasal swab (ANS) specimens from individuals with signs and symptoms of respiratory tract infection. SARS-CoV-2 viral RNA is generally detectable in NPS and ANS specimens during the acute phase of infection.

    The BD Respiratory Viral Panel-SCV2 for BD MAX™ System is intended for use as an aid in the diagnosis of SARS-CoV-2 infection if used in conjunction with other clinical and epidemiological information, and laboratory findings.

    Positive results do not rule out co-infection with other organisms. The agent detected by the BD Respiratory Viral Panel-SCV2 for BD MAX™ System may not be the definitive cause of disease.

    Negative results do not preclude SARS-CoV-2 infection.

    The results of this test should not be used as the sole basis for diagnosis, treatment management decisions.

    Device Description

    The BD Respiratory Viral Panel (BD RVP) and BD Respiratory Viral Panel-SCV2 (BD RVP-SCV2) along with the BD MAXTM System are comprised of an instrument with associated hardware and accessories, disposable microfluidic cartridges, master mixes, unitized reagent strips, and extraction reagents. The instrument automates sample preparation including target lysis, Total Nucleic Acid (TNA) extraction and concentration, reagent rehydration, target nucleic acid amplification and detection using real-time PCR. The assay includes a Sample Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors RNA extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances. The BD MAX™ System software automatically interprets test results. For the BD Respiratory Viral Panel for BD MAX™ System and BD Respiratory Viral Panel-SCV2 for BD MAX™ System, a test result may be called as POS, NEG or UNR (Unresolved) based on the amplification status of the targets and of the Sample Processing Control. IND (Indeterminate) or INC (Incomplete) results are due to BD MAX™ System failure.

    AI/ML Overview

    The provided text describes the analytical and clinical performance evaluation of two molecular diagnostic devices: the BD Respiratory Viral Panel (BD RVP) for BD MAX System and the BD Respiratory Viral Panel-SCV2 (BD RVP-SCV2) for BD MAX System. Both are real-time RT-PCR tests for detecting respiratory viruses in nasopharyngeal and anterior nasal swab specimens.

    Here's an analysis of the acceptance criteria and study proving device performance, based on the provided text:

    Important Note: The provided text is a summary of a 510(k) premarket notification. As such, it reports performance data but does not explicitly state "acceptance criteria" in the format of a pre-defined table with specific percentage goals for PPA and NPA. Instead, the reported performance itself serves as the demonstration that the device is "substantially equivalent" to predicate devices, implying that the observed performance would be considered acceptable by the FDA for the given intended use. For this response, I will interpret "acceptance criteria" as the performance metrics and results deemed sufficient for clearance.


    1. Table of Acceptance Criteria (Interpreted) and Reported Device Performance

    Given that this is a 510(k) submission, the "acceptance criteria" are not explicitly listed with numeric targets in the document. Instead, the reported performance is presented to demonstrate substantial equivalence to legally marketed predicate devices. The implicit acceptance criteria are the high percentages of Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) achieved by the device when compared to a composite reference method (ground truth).

    The primary performance data comes from clinical evaluation studies for both the BD RVP and BD RVP-SCV2.

    BD Respiratory Viral Panel (BD RVP) for BD MAX System - Clinical Performance Summary

    Analyte (Specimen Type)Performance MetricReported Device PerformanceImplicit Acceptance Criteria (based on reported)
    SARS-CoV-2 (NP Swab)Positive Percent Agreement (PPA)Overall: 98.9% (517/523)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Overall: 97.7% (999/1022)High NPA (e.g., >95%)
    SARS-CoV-2 (ANS Swab)Positive Percent Agreement (PPA)Overall: 98.4% (478/486)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Overall: 97.7% (1050/1075)High NPA (e.g., >95%)
    Flu A (NP Swab)Positive Percent Agreement (PPA)Overall: 96.7% (59/61)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Overall: 99.5% (1494/1501)High NPA (e.g., >95%)
    Flu A (ANS Swab)Positive Percent Agreement (PPA)Overall: 96.9% (62/64)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Overall: 99.7% (1496/1500)High NPA (e.g., >95%)
    Flu B (NP Swab)Positive Percent Agreement (PPA)No data for PPA rate calculation (prospective); Retrospective: 100.0% (58/58)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Prospective: 99.9% (1561/1562); Retrospective: 98.9% (180/182)High NPA (e.g., >95%)
    Flu B (ANS Swab)Positive Percent Agreement (PPA)No data for PPA rate calculation (prospective); Retrospective: 100.0% (12/12)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Prospective: 100.0% (1564/1564); Retrospective: 98.9% (172/174)High NPA (e.g., >95%)
    RSV (NP Swab)Positive Percent Agreement (PPA)Overall: 100.0% (12/12)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Overall: 100.0% (1550/1550)High NPA (e.g., >95%)
    RSV (ANS Swab)Positive Percent Agreement (PPA)Overall: 91.7% (11/12)High PPA (e.g., >95%)
    Negative Percent Agreement (NPA)Overall: 99.9% (1551/1552)High NPA (e.g., >95%)

    BD Respiratory Viral Panel-SCV2 (BD RVP-SCV2) for BD MAX System - Clinical Performance Summary

    For BD RVP-SCV2 clinical performance, the text refers to the same clinical study results as BD RVP, but only for SARS-CoV-2.

    • SARS-CoV-2 (NP Swab): PPA 34.6% (541/1562), NPA not explicitly listed but implied from overall subject data. This seems to be a positivity rate, not an agreement rate. The clinical performance summary tables (Table 28 and 29) for SARS-CoV-2 are the relevant ones.
      • NP (Overall): PPA 98.9%, NPA 97.7%
      • ANS (Overall): PPA 98.4%, NPA 97.7%
        (These are the same as reported for the full BD RVP, which makes sense as SCV2 is a subset).

    2. Sample Size Used for the Test Set and Data Provenance

    BD Respiratory Viral Panel (BD RVP) for BD MAX System:

    • Prospective Clinical Evaluation:

      • Sample Size:
        • 2,005 subjects enrolled.
        • Test Set: 1,562 nasopharyngeal swabs (NPS) and 1,566 anterior nasal swabs (ANS).
        • For performance calculations, slightly fewer compliant specimens were used:
          • NPS: 1,545 for SARS-CoV-2, 1,562 for Flu A, Flu B, and RSV.
          • ANS: 1,561 for SARS-CoV-2, 1,564 for Flu A, Flu B, and RSV.
      • Data Provenance:
        • Country of Origin: United States (six geographically distinct U.S. study sites) and Europe (two geographically distinct sites).
        • Retrospective or Prospective: Primarily Prospective. Specimens were collected between January and August 2022.
          • January to early April 2022: Prospective archived/frozen (Category II).
          • Mid-April to August 2022: Prospective fresh (Category I).
    • Retrospective Clinical Evaluation:

      • Sample Size:
        • NPS: 240 frozen retrospective nasopharyngeal swabs.
        • Nasal Swabs: 187 frozen retrospective nasal swabs.
      • Data Provenance:
        • Country of Origin: Not explicitly stated within the retrospective section, but implied to be part of the broader US/Europe context from the prospective study.
        • Retrospective or Prospective: Retrospective. Specimens collected between December 2019 and January 2022 (for NPS) and February 2021 and February 2023 (for nasal swabs) from external sources. These were "archived, frozen specimens".

    BD Respiratory Viral Panel-SCV2 for BD MAX System:

    • The text states: "In the BD Respiratory Viral Panel-SCV2 for BD MAX™ System clinical study, reportable results from specimens compliant at the specimen and PCR levels were obtained from 8 geographically diverse sites. Nasopharyngeal and nasal specimens totaled 1.566 respectively." This refers to the same clinical study data as the broader BD RVP, specifically focusing on SARS-CoV-2 results from that dataset.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts or their qualifications for establishing ground truth.

    Instead, the ground truth for the clinical evaluation (test set) was established by comparison to highly sensitive molecular assays (NAATs).

    • For SARS-CoV-2: A composite method of "two out of three highly sensitive molecular assays (NAATS) that are FDA authorized under EUA for SARS-CoV-2."
    • For influenza B and RSV: "an FDA-cleared high sensitivity RT-PCR assay."

    This indicates an analytical or "reference method" based ground truth rather than a human expert consensus.


    4. Adjudication Method for the Test Set

    The adjudication method relies on a "two out of three" rule for SARS-CoV-2 and a single FDA-cleared RT-PCR assay for influenza B and RSV.

    • SARS-CoV-2: "Any specimen that tested positive by two EUA assays was considered positive for SARS-CoV-2, whereas any specimen that tested negative by two EUA assays was considered negative."
    • Influenza B and RSV: Comparison to a single "FDA-cleared high sensitivity RT-PCR assay."

    No explicit 2+1 or 3+1 human expert adjudication method is mentioned, as the ground truth derivation is assay-based.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a diagnostic assay (RT-PCR test), not an AI-assisted imaging product. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply. The performance is assessed on the device's ability to accurately detect specific viral nucleic acids.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    While the term "algorithm" is used to describe the software interpreting PCR data, the entire device (BD MAX System with the viral panel) is an automated system. The performance data presented (PPA, NPA) is the standalone performance of the device without human interpretation of the raw PCR signals for diagnosis. The BD MAX™ System software "automatically interprets test results" (page 7). Therefore, the clinical performance described is the standalone performance.


    7. The Type of Ground Truth Used

    The ground truth used was reference method comparison / composite molecular assay results:

    • SARS-CoV-2: Composite reference method using results from "two out of three highly sensitive molecular assays (NAATs) that are FDA authorized under EUA for SARS-CoV-2." (Page 42)
    • Influenza A, Influenza B, and RSV: An "FDA-cleared high sensitivity RT-PCR assay." (Page 42)

    This is a form of "analytical ground truth" established by other validated diagnostic tests, rather than pathology, expert consensus on images, or long-term clinical outcomes.


    8. The Sample Size for the Training Set

    The document does not explicitly state a separate "training set" sample size for the assay's development in the context of a statistical machine learning model.

    However, it mentions: "As the product undergoes product development, the data is supplemented, and the algorithm is adjusted ('trained') using viral cultures spiked into clinical background matrices at levels surrounding the limit of detection and expected clinical range." (Page 30, and repeated on Page 40 for SCV2). This refers to internal development and optimization using contrived samples, not a distinct, pre-defined "training set" of patient data for a typical AI/ML model for regulatory submission.

    Performance data for the submission is based on the "test set" (clinical evaluation studies described in section 2) and analytical studies.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, there isn't a "training set" ground truth in the sense of independently verified patient data used to directly train a submitted AI model. Instead, the "algorithm" (how the BD MAX software interprets PCR signals to call positive/negative) was "adjusted ('trained')" using:

    • Viral cultures spiked into clinical background matrices at various concentrations (e.g., around the Limit of Detection and expected clinical range).
    • These are contrived samples with a known (positive or negative) status and precise viral concentrations.

    The ground truth for these internal development/training stages would be the known concentration of spiked viral material.