510(k) Data Aggregation

    K Number
    K120443

    Validate with FDA (Live)

    Date Cleared
    2012-05-22

    (98 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Baxter Neutral Luer Activated Device is intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids. This device may be used with low pressure power injectors having a maximum pressure of 325 psi (2241 kPa) and a maximum flow rate of 10 mL/s.

    Device Description

    The Neutral Luer Activated Device (LAD) and Extension Sets with Neutral Luer Activated Device are single use disposable devices intended for use with a vascular access device for continuous or intermittent fluid administration or withdrawal of fluids. The device is an in-line access site and can be connected to male Luer adapters (e.g., syringes or sets) to allow needleless access to the fluid or vascular path and is designed to be easy to swab. A saline flush of 10 mL is able to clear the Neutral LAD of blood after sampling blood through the device. The Neutral LAD is clear and allows the clinician to view the fluid path of the device. The Neutral LAD has a low priming volume (0.08 mL) that allows medication to be flushed from the device with small amounts of fluid and is the maximum amount of medication that could potentially remain in the device when administered to a patient prior to flushing. A LAD with a low priming/residual volume is favored for pediatric patients or patients requiring fluid restriction. The device is a neutral fluid displacement LAD and does not require a specific clamping sequence in order to be used safely. This neutral displacement design has the added benefit of helping to reduce the occurrence of thrombotic catheter occlusions by limiting the reflux of blood into the catheter. The Neutral LAD may be used with low pressure power injectors having a maximum pressure of 325 psi (2241 kPa) and a maximum flow rate of 10 mL/s.

    AI/ML Overview

    This 510(k) premarket notification describes a medical device, specifically the "Neutral Luer Activated Device and Extension Sets with Neutral Luer Activated Device," not an AI/ML powered device. As such, the requested information regarding AI/ML study components (such as sample size for test/training sets, ground truth methodology, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance) is not applicable and cannot be extracted from the provided text.

    The document focuses on demonstrating the substantial equivalence of the new device to existing predicate devices through non-clinical testing.

    Here's the relevant information that can be extracted regarding acceptance criteria and performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance CriteriaReported Device Performance
    Intended UseAdministration of drugs/solutions without needles, eliminating needle-stick injuries. In-line injection site for continuous/intermittent fluid administration/withdrawal.The device is intended for these uses, aligning with the description. It functions as an in-line access site and allows needleless access.
    Power Injection CompatibilityMaximum pressure of 325 psi (2241 kPa) and maximum flow rate of 10 mL/s.The Neutral LAD may be used with low pressure power injectors having a maximum pressure of 325 psi (2241 kPa) and a maximum flow rate of 10 mL/s. Testing supports this use.
    Neutral Fluid DisplacementDoes not require a specific clamping sequence for safe use; helps reduce thrombotic catheter occlusions by limiting blood reflux.The device is described as "neutral fluid displacement LAD" which "does not require a specific clamping sequence in order to be used safely" and "has the added benefit of helping to reduce the occurrence of thrombotic catheter occlusions by limiting the reflux of blood into the catheter."
    SwababilityDesigned to be easy to swab.The Neutral LAD "is designed to be easy to swab."
    VisibilityClear to allow clinicians to view the fluid path.The Neutral LAD "is clear and allows the clinician to view the fluid path of the device."
    Priming VolumeLow priming volume for pediatric patients or those requiring fluid restriction.The Neutral LAD "has a low priming volume (0.08 mL)"
    Blood Clearance (Saline Flush)Capable of clearing blood after sampling with a 10 mL saline flush.A "saline flush of 10 mL is able to clear the Neutral LAD of blood after sampling blood through the device."
    Use Duration/ActuationsNot explicitly stated as acceptance criteria, but tested."Testing supports use of the device for 7 days and/or 200 actuations."
    Sterility & Non-pyrogenicSterile and non-pyrogenic."The product is sterile and non-pyrogenic."
    Functional, Biocompatibility, Microbial Ingress, Ease of UseMet all acceptance criteria."This device met all the acceptance criteria for all functional, biocompatibility, microbial ingress, sterility and ease of use requirements and the data support that the device is appropriately designed for its intended use."

    2. Sample size used for the test set and the data provenance:

    • Not applicable. This is a physical medical device, not an AI/ML algorithm requiring a data test set for performance evaluation in the context of an algorithm's output. The "testing" refers to benchtop or in-vitro tests on the devices themselves.
    • The provenance of data refers to the specific non-clinical tests performed (e.g., pressure testing, flow rate testing, biocompatibility testing, microbial ingress testing, etc.). These are typically laboratory-based tests and not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. As this is a physical device, ground truth is established through adherence to engineering specifications, validated test methods, and regulatory standards for device performance, sterility, and biocompatibility. There is no concept of "expert ground truth" in the way it is used for diagnostic AI.

    4. Adjudication method for the test set:

    • Not applicable. There is no "test set" in the context of human interpretation needing adjudication for an AI algorithm. Device performance is evaluated against predefined pass/fail criteria in laboratory settings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a physical medical device, not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an AI/ML algorithm.

    7. The type of ground truth used:

    • Engineering specifications and regulatory standards. For example, pressure testing would have a ground truth defined by a specific pressure range (e.g., 325 psi maximum), and the device either meets or fails this. Biocompatibility would be determined by established ISO standards. Microbial ingress would have a ground truth of preventing microbial contamination.

    8. The sample size for the training set:

    • Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is a physical medical device, not an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K093120

    Validate with FDA (Live)

    Date Cleared
    2010-02-17

    (138 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hemodialysis with Xenium XPM dialyzers is indicated for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate. It also may be indicated in the treatment of patients intoxicated with poisons or drugs.

    Device Description

    Xenium XPM dialyzers are Polyethersulfone fiber dialyzers and will be labeled for single use only. The dialyzers are available in six (6) sizes, which are differentiated by membrane surface area.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the Xenium XPM hemodialyzer. It focuses on demonstrating substantial equivalence to a predicate device and does not contain the detailed information typically found in studies for AI/ML-based devices regarding acceptance criteria, performance metrics, ground truth establishment, or sample sizes in the manner requested.

    Here's an analysis of the information that is available in the document, and where the requested details are missing:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: The document states, "All test results meet the acceptance criteria, and support that the devices are appropriately designed for their intended use." However, it does not specify what these quantitative acceptance criteria are (e.g., specific thresholds for clearance rates, ultrafiltration rates, etc.).
    • Reported Device Performance: No specific performance metrics or results are reported in this summary. The manufacturer conducted "design verification tests" but the outputs of these tests are not provided. The demonstration of substantial equivalence suggests the performance is similar to the predicate, but no direct performance data is given.

    Conclusion for Point 1: No specific acceptance criteria or quantitative device performance data are provided in this 510(k) summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document mentions "design verification tests" and "risk analyses per ISO 14971." These are non-clinical tests.
    • Sample Size: No sample size for any test set is mentioned.
    • Data Provenance: Not applicable, as this refers to non-clinical, in-vitro testing/design verification, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This device is a hemodialyzer, a physical medical device, not an AI/ML diagnostic or measurement tool. There is no concept of "ground truth" derived from expert consensus in the context of its performance verification as described here. Physical and chemical tests are likely used.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This term refers to the process of resolving discrepancies in expert interpretations, which is not relevant for the type of device and verification described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. The Xenium XPM is a hemodialyzer, not an AI-assisted diagnostic or imaging system. Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This device does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable in the AI/ML sense. For a hemodialyzer, "ground truth" for performance would be established by validated physical and chemical measurement methods (e.g., measuring solute clearance, ultrafiltration, pressure drops according to established standards).

    8. The sample size for the training set

    • Not applicable. This device does not involve an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This device does not involve an AI/ML algorithm.

    Summary of what the study provided in the document focuses on:

    The document describes a "Special 510(k) Premarket Notification" for a hemodialyzer, the Xenium XPM. The primary objective of this type of submission is to demonstrate substantial equivalence to an already legally marketed predicate device (in this case, the Baxter Xenium XPH dialyzer, K083778).

    The "study" or testing mentioned is non-clinical: "Baxter Healthcare Corporation conducts risk analyses, per ISO 14971, and design verification tests based on the result of these analyses. All test results meet the acceptance criteria, and support that the devices are appropriately designed for their intended use."

    This indicates that standard engineering and regulatory verification tests were performed to ensure the device met its design specifications and safety requirements, likely comparing its physical characteristics and functional performance (e.g., membrane surface area, materials, in-vitro clearance, and ultrafiltration rates) against the predicate device to demonstrate equivalence. The details of these tests, including specific acceptance criteria and detailed quantitative results, are not included in this 510(k) summary, which is typical for such submissions where the focus is on the conclusion of substantial equivalence rather than a full study report.