510(k) Data Aggregation

    K Number
    K120126

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-10

    (237 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BaroSense Endogastric Tube (EGT) is indicated for use with an endoscope where multiple endoscopic intubations are anticipated.

    Device Description

    The EGT is a reusable overtube used in hospitals or surgery centers to provide a channel for the delivery and removal of endoscopic devices from the mouth to the stomach. It is intended for use when multiple endoscopic intubations, or endoscopic instrument/tool exchanges, are anticipated. A flexible Introducer Bougie has a tapered tip to assist in introduction of the EGT through the mouth and esophagus. The bougie has an internal lumen to accommodate a 5 mm endoscope. The EGT and Introducer Bougie are supplied non-sterile. The predicate device has identical technological characteristics. However, the predicate EGT is provided as a single-patient-use, disposable unit, with an internal bougie introducer included in the package. The modified device is provided as a reusable overtube and single-use bougie. The predicate device working tube is a PVC tube with an embedded stainless steel reinforcement coil, whereas the modified device is a Tecoflex tube with an embedded stainless steel reinforcement coil. The proximal handle of the modified device has an added feature to allow the endogastric tube to grip and seal to Introducer Bougie or other endoscopic devices. This gripping action prevents relative motion between the overtube and the other devices, e.g., the Introducer Bougie.

    AI/ML Overview

    The BaroSense Endogastric Tube (EGT) is a reusable overtube used with an endoscope where multiple endoscopic intubations are anticipated. The 510(k) summary indicates that the device's safety and effectiveness were established through a series of bench and animal tests. However, it does not provide acceptance criteria, reported performance values, or specific details for these studies.

    Here's a breakdown of the requested information based on the provided text, and where information is not available, it is noted:

    1. Table of Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance CriteriaReported Device Performance
    BiocompatibilityISO 10993 standards"Acceptable results"
    Safety and EffectivenessNot specified"Acceptable results"
    Comparison to predicate deviceSubstantial equivalence"Identical technological characteristics"

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text only mentions "bench and animal tests" but does not specify sample sizes for any test sets or the data provenance (e.g., country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned. The submission focuses on demonstrating substantial equivalence to a predicate device through bench and animal testing.

    6. Standalone (Algorithm Only) Performance Study

    This device is a medical instrument (Endogastric Tube), not an algorithm or AI system. Therefore, a standalone algorithm performance study is not applicable and not mentioned.

    7. Type of Ground Truth Used

    The ground truth for the "safety and effectiveness" appears to be based on the results of "bench and animal tests" and comparison to known biocompatibility standards (ISO 10993). However, the specific metrics for ground truth are not detailed.

    8. Sample Size for the Training Set

    As this is a physical medical device and not an AI or algorithm, there is no concept of a "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is not an AI/algorithm-based device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120147

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-02-17

    (30 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BaroSense ACE Stapler is indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract.

    Device Description

    The ACETM Stapler is a surgical stapler used in hospitals or surgery centers for staple closure on the wall of the stomach or gastrointestinal tract. The single-patient-use, disposable stapler head is supplied non-sterile and is fitted with a sterile, single-use staple cartridge. The stapler head is attached to a reusable, flexible stapler handle that controls the position and articulation of the stapler head. In use, the stapler is introduced into the patient through the mouth. A flexible endoscope passes through the stapler for gastric tissue visualization. The stapler works in conjunction with a vacuum pump to create a plication (tissue fold) in the GI tract, which is then compressed. The stapler then places a double ring of titanium staples (8 staples). A non-absorbable ring helps reinforce the staple placement in the tissue. The tissue compression and stapling functions are controlled by commercially available inflation syringes. An endogastric overtube may be used to protect the esophageal tissues during repeated insertions of the device. The overtube, flexible endoscope, vacuum pump and inflation syringes used with the stapler are all commercially available medical devices, not the subject of this 510(k), and are not supplied with the stapler.

    AI/ML Overview

    This document describes the BaroSense ACE™ Stapler and Cartridge, a surgical stapler used for staple closure on the wall of the stomach or gastrointestinal tract. However, it does not contain the specific information requested regarding acceptance criteria and a study proving the device meets those criteria. The provided text is a 510(k) summary for a medical device seeking substantial equivalence to a previously cleared predicate device.

    Here's a breakdown of why the requested information is absent and what is provided:

    Missing Information (and why it's missing from this type of document):

    • A table of acceptance criteria and the reported device performance: This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device, not on detailing specific performance acceptance criteria and results. Such criteria and results would typically be found in detailed test reports, not in this high-level summary.
    • Sample size used for the test set and the data provenance: Clinical study details are not typically provided in this type of 510(k) summary for a device seeking substantial equivalence based on bench and animal testing.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no human-read test set is described.
    • Adjudication method for the test set: Not applicable as no human-read test set is described.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a surgical stapler, not an AI-assisted diagnostic or imaging device.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a surgical stapler.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable as no human-read test set is described. Performance is based on physical and mechanical testing.
    • The sample size for the training set: Not applicable. This is a surgical stapler, not an AI/ML device requiring a training set.
    • How the ground truth for the training set was established: Not applicable.

    What the document does provide regarding "acceptance criteria" and "study":

    The document states:

    • "All patient-contacting components of the ACE Stapler are composed of materials of known biocompatibility and tested to the requirements of ISO 10993 for the predicate device." This implies that biocompatibility is an acceptance criterion, and testing to ISO 10993 is the study demonstrating it.
    • "The only new material for the modified device is a white orientation marking on the flexible portion of the stapler handle. The new material has also been tested to the requirements of ISO 10993." This reinforces biocompatibility testing as an acceptance criterion for new materials.
    • "The safety and effectiveness of the device was further established through a series of bench and animal tests. All testing yielded acceptable results." This is a general statement that safety and effectiveness were verified through bench and animal studies, and the results were acceptable. It does not provide specific performance metrics or acceptance criteria beyond "acceptable results."

    Summary of available information:

    Criteria/AspectInformation provided in the document
    Acceptance Criteria & Reported Performance- Biocompatibility: New and existing patient-contacting materials tested to ISO 10993 requirements. (Specific pass/fail values or reported performance metrics are not detailed in this summary, but the general statement is "tested to the requirements" and yielded "acceptable results".)- Safety and Effectiveness: Established through bench and animal tests. (Specific performance metrics and acceptance criteria for these tests are not detailed beyond "All testing yielded acceptable results.")- Substantial Equivalence: The primary "acceptance criteria" for K120147, as a 510(k), is substantial equivalence to the predicate device K110829 by demonstrating similar technological characteristics and indications for use, despite minor modifications (staple pattern, stapler head diameter, stapler handle diameter, longer articulation section). This is not a performance criterion in the typical sense, but rather a regulatory one.
    Sample size (test set) & Data ProvenanceNot specified for bench or animal tests. Data provenance is implied to be from BaroSense, Inc.'s internal testing.
    Number of experts & qualifications for ground truth (test set)Not applicable. No expert-based ground truth for a test set is described. Tests are physical/mechanical.
    Adjudication method for test setNot applicable.
    MRMC comparative effectiveness studyNot applicable. This is a surgical stapler, not an AI/imaging device.
    Standalone (algorithm only) performance studyNot applicable. This is a surgical stapler.
    Type of ground truth usedNot applicable in the context of human expert review. Performance is assessed against engineering specifications, biocompatibility standards (ISO 10993), and functional performance in bench and animal models (though specific metrics not detailed in this summary).
    Sample size for training setNot applicable. This is a surgical stapler, not an AI/ML device.
    How ground truth for training set was establishedNot applicable.

    In essence, the document serves as a regulatory summary to demonstrate substantial equivalence, not a detailed technical report on performance testing and acceptance criteria. It mentions "bench and animal tests" yielded "acceptable results" to establish safety and effectiveness, and new materials were "tested to the requirements of ISO 10993" for biocompatibility, but it does not elaborate on the specific acceptance criteria or the study details.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110829

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-06-03

    (71 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BaroSense ACE Stapler and Cartridge are indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract.

    Device Description

    The ACE™ Stapler is a surgical stapler used in hospitals or surgery centers for staple closure on the wall of the stomach or gastrointestinal tract. The single-patient-use, disposable stapler head is supplied non-sterile and is fitted with a sterile, single-use staple cartridge. The stapler head is attached to a reusable, flexible stapler handle that controls the position and articulation of the stapler head. In use, the stapler is introduced into the patient through the mouth. A flexible endoscope passes through the stapler for gastric tissue visualization. The stapler works in conjunction with a vacuum pump to create a plication [tissue fold] in the GI tract, which is then compressed. The stapler then places a double, circular row of titanium staples. A non-absorbable ring helps reinforce the staple placement in the tissue. The tissue compression and stapling functions are controlled by commercially available inflation syringes. The stapler may be introduced over a guidewire. If multiple plications are required, an endogastric overtube may be used to protect the esophageal tissues during repeated insertions of the device. The guidewire, overtube, flexible endoscope, vacuum pump and inflation syringes used with the stapler are all commercially available medical devices, not the subject of this 510(k), and are not supplied with the stapler.

    AI/ML Overview

    The BaroSense ACE Stapler and Cartridge underwent bench and animal testing to establish safety and effectiveness. The device's biological evaluation also conforms to ISO 10993-1:2009 for biocompatibility, ISO 11137-1:2006 for sterilization of healthcare products - radiation - Part 1, and ISO 11137-2:2006 for sterilization of healthcare products - radiation - Part 2. The report explicitly states that for each of these ISO standards, the test data demonstrated conformity and the standards themselves include acceptance criteria.

    Here's a breakdown of the specific aspects you requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Test/StandardAcceptance CriteriaReported Device Performance
    Material Biocompatibility (per ISO 10993-1:2009)Patient contacting components composed of materials of known biocompatibilityAll patient contacting components are composed of materials of known biocompatibility, meeting the requirements of ISO 10993. Test data demonstrated conformity to the requirements of this standard.
    Sterilization - Radiation Part 1 (per ISO 11137-1:2006)Standard defines acceptance criteria for radiation sterilization of healthcare products.Test data for the device demonstrated conformity to the requirements of this standard. This standard includes acceptance criteria.
    Sterilization - Radiation Part 2 (per ISO 11137-2:2006)Standard defines acceptance criteria for radiation sterilization of healthcare products.Test data for the device demonstrated conformity to the requirements of this standard. This standard includes acceptance criteria.
    Bench/Animal Testing (General Safety and Effectiveness)Acceptable results for safety and effectiveness through a series of bench and animal tests."All testing yielded acceptable results."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "a series of bench and animal tests" but does not specify the sample sizes for these tests. The country of origin for the data, whether it was retrospective or prospective, is also not provided in the given text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth. The evaluation mainly references compliance with recognized international standards (ISO) and general "bench and animal tests."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any adjudication method for test sets.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication of a multi-reader multi-case (MRMC) comparative effectiveness study being performed, nor is there any mention of AI assistance or human reader improvement with AI in the provided text. This device is a surgical stapler, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable. The BaroSense ACE Stapler is a medical device, not an algorithm or AI system. Its performance is inherent to its mechanical and material properties, and its functionality is directly operated by a human user (surgeon) with a flexible endoscope for visualization.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For biocompatibility and sterilization, the "ground truth" is adherence to the specifications and requirements defined within the referenced ISO standards (ISO 10993, ISO 11137-1, ISO 11137-2). For general safety and effectiveness, the document generally refers to "acceptable results" from bench and animal tests. No specific expert consensus, pathology, or outcomes data is detailed as the direct ground truth apart from the standards themselves.

    8. The sample size for the training set

    This question is not applicable as the device is a mechanical surgical stapler and does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082044

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-05-11

    (662 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BaroSense ACE Stapler is indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract.

    Device Description

    The ACE™ Stapler is a single use, disposable, surgical stapler used in hospitals or surgery centers to create a plication of the wall of the stomach or lower esophagus. The stapler is supplied non-sterile and is fitted with a sterile, single-use staple cartridge. In use, the stapler is introduced into the patient through the mouth using a disposable endogastric tube. A flexible endoscope passes through the stapler for gastric tissue visualization. The stapler works in conjunction with a vacuum pump to create a plication of the GI tract, which is then held in place by means of a double, circular row of titanium staples delivered by the stapler (along with a non- absorbable ring to help reinforce the staple placement in the tissue). The endogastric tube, flexible endoscope and vacuum pump used with the stapler are all standard medical devices and are not supplied with the stapler.

    AI/ML Overview

    The provided document is a 510(k) Summary for the BaroSense ACE™ Stapler and Cartridge. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than presenting detailed performance studies against specific acceptance criteria. Therefore, most of the requested information regarding acceptance criteria and study details for proving device performance cannot be extracted from this document.

    Here's an analysis of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria with corresponding device performance metrics. It generally states that "All testing yielded acceptable results" without detailing the specific criteria or the quantitative results.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "a series of bench and animal tests" but does not specify the sample sizes used for these tests, their provenance (country of origin), or whether they were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided as the submission focuses on substantial equivalence based on bench and animal testing, not human expert evaluations for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This type of study is not mentioned or discussed in the document.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is a surgical stapler, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document indicates that the safety and effectiveness were established through "bench and animal tests." This implies that the "ground truth" or evaluative standards would be based on engineering specifications, physiological measurements, and observations from animal models, rather than expert consensus, pathology, or human outcomes data in the context of clinical trials.

    8. The sample size for the training set

    The document does not discuss a "training set" as it is not an AI/algorithm-based device.

    9. How the ground truth for the training set was established

    This information is not applicable as the device is not an AI/algorithm-based device.


    Summary of available relevant information from the document:

    • Study Type: Bench and Animal Testing
    • Purpose of Study: To establish safety and effectiveness, and known biocompatibility of patient-contacting components.
    • Key Findings: "All testing yielded acceptable results."
    • Biocompatibility Testing: Patient-contacting components tested to requirements of ISO 10993.
    • Indications for Use: The BaroSense ACE Stapler is indicated for endoluminal trans-oral tissue approximation and ligation in the gastrointestinal tract.

    The document emphasizes substantial equivalence to legally marketed predicate devices (K062875, K011016, K073671) rather than detailing specific performance data against pre-defined acceptance criteria for a novel functionality.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082589

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-05-22

    (256 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    18 - 65
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BaroSense EGT is indicated for use with an endoscope where multiple endoscopic intubations are anticipated.

    Device Description

    The BaroSense EGT is a single use, disposable, overtube used in hospitals or surgery centers to provide a channel for the delivery and removal of endoscopic devices from the mouth to the stomach. It is intended for use when multiple endoscopic intubations, or endoscopic instrument/tool exchanges, are anticipated.

    The EGT is supplied non-sterile and is provided with a removable, flexible introducer with a tapered tip that assists in introduction of the EGT through the mouth and esophagus.

    AI/ML Overview

    The provided text is a 510(k) Summary for the BaroSense Endogastric Tube (EGT). This document primarily focuses on establishing substantial equivalence to existing predicate devices based on device description, indications for use, and biocompatibility.

    Crucially, the document does NOT contain information about acceptance criteria, device performance metrics (e.g., sensitivity, specificity, accuracy), or any study that directly reports on such performance data. It briefly mentions "bench tests" but does not detail their methodology, results, or specific acceptance criteria.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance or information about sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness studies, as these details are not present in the provided text.

    Based on the available information, here's what can be extracted:


    Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

    1. Table of Acceptance Criteria and the Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Not specified in the documentNot specified in the document
    However, based on the context of a 510(k) for an overtube, implicit performance criteria would likely include:Implicitly, the device successfully met these criteria, as substantial equivalence was determined.
    Biocompatibility requirements per ISO 10993"All patient contacting components of the EGT are composed of materials of known biocompatibility tested to the requirements of ISO 10993." (Stated as met)
    Functional performance characteristics (e.g., ease of insertion, ability to create a channel for endoscopic devices, durability during intended use, compatibility with endoscopes, prevention of kinking/collapse)"The safety and effectiveness of the device was further established through a series of bench tests. All testing yielded acceptable results." (Stated as met, but no specific results are provided)
    Dimensional specifications (length, diameter, compatibility with typical endoscopes)Not explicitly stated as an acceptance criterion, but would be assessed in bench testing.
    Sterilization (if applicable, though stated as non-sterile)Not applicable as the device is supplied non-sterile.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not specified. The document mentions "a series of bench tests," but provides no details on sample size or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable/Not specified. This device is an Endogastric Tube (overtube), not an AI/imaging diagnostic device that would typically involve expert ground truth for performance evaluation in the context of the questions asked. Its evaluation relied on engineering bench tests and biocompatibility testing.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable/Not specified. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI/imaging diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI/imaging diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For biocompatibility: Adherence to ISO 10993 standards and material conformance.
    • For functional performance: Engineering specifications and testing protocols for items like material strength, flexibility, dimensions, and ease of use in simulated environments. The "ground truth" would be whether the device met its pre-defined engineering specifications and performed as intended in bench simulations.

    8. The sample size for the training set:

    • Not applicable. This is a hardware medical device (an overtube), not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    Summary of Device and 510(k) Logic:

    The BaroSense Endogastric Tube (EGT) is a single-use, disposable overtube intended to provide a channel for endoscopic procedures requiring multiple intubations or instrument exchanges. Its 510(k) clearance (K082589) was based on demonstrating substantial equivalence to existing predicate devices (Guardus™ Disposable Overtube, NDO Surgical Overtube, CR Bard Endoscopic Overtube).

    The evidence for substantial equivalence primarily relied on:

    • Similar Indications for Use: "use with an endoscope where multiple endoscopic intubations are anticipated."
    • Similar Device Description: Single-use, disposable overtube with an introducer.
    • Biocompatibility: Confirmation that all patient-contacting components met ISO 10993 requirements.
    • Bench Testing: A series of unspecified "bench tests" were performed, with acceptable results, presumably demonstrating functional performance characteristics consistent with predicate devices and safety for intended use.

    The 510(k) process for this device focused on confirming that it did not raise new safety or effectiveness concerns compared to already marketed devices, rather than establishing novel performance metrics or conducting clinical trials as might be seen for a new drug or high-risk implant.