510(k) Data Aggregation
(203 days)
The SDS Growing Rod is indicated for skeletally immature patients less than 10 years of age with the potential for additional spinal growth who require surgical treatment for correction and maintenance of the correction of severe, progressive, life-threatening early-onset deformities, including early-onset scoliosis, which are associated with or at risk of thoracic insufficiency syndrome.
The SDS Growing Rod is an implant designed to correct and actively lengthen the spine in children with early onset spinal deformities while allowing growth, without the need for re-operations or other interventions. This is achieved by a spinal construct with one or two springs and sliding bearings. The spring provides a distraction force to correct the spinal curve while the sliding bearing is designed to enable growth by allowing displacement of one rod with respect to the other rod.
The SDS Growing Rod is available with 5.5 mm rods. Two configurations of the SDS Growing Rod are possible: hybrid and bilateral. The SDS Growing Rod is delivered sterile and is for single use only. The intended lifetime of the SDS Growing Rod while allowing growth is up to 5 years. The SDS Growing Rod is removed or locked after the treatment is completed.
The SDS Growing Rod is composed out of the following components:
- Titanium alloy springs with UHMWPE bushings
- Titanium alloy Axial Sliding Bearing with UHMWPE bearings
- Titanium alloy Polyaxial Sliding Bearing with UHMWPE bearings
- Titanium alloy buttress
- CoCr rods
- Titanium alloy C-Clamp
The SDS Growing Rod is used in combination with:
- Dedicated instrument set
- Spring Compressor
- Spring Locker
- Torque Limiting Handle 9Nm
- C-Clamp Forceps
- SDS Tray
- Pedicle screws for 5.5 mm rods. The SDS Growing Rod is compatible with pedicle screws from the RESPONSE™ 5.5/6.0 mm System (Orthopediatrics) and their dedicated instrumentation for screw placement. Other components of the RESPONSE™ system, such as connectors, hooks, or ancillary fixation elements, may be used within the same spinal construct as clinically indicated. However, these components should not impede the sliding function of the growing rod or integrated spring mechanism.
N/A
Ask a specific question about this device
(341 days)
The PEEK SINEFIX is intended for soft tissue to bone reattachment in rotator cuff repairs for tendon ruptures up to 2 cm.
The SINEFIX is a PEEK implant which can be used to re-attach a ruptured rotator cuff tendon to the humerus of the shoulder. The implant can be used to re-attach complete or partially ruptured tendons up to 2 cm to facilitate the healing of the natural bone-tendon interface. The implant consists of a baseplate with a lateral and a medial anchor. The baseplate is placed over the tendon and attached to the humerus with the anchors. The medial anchor goes through the tendon, while the lateral anchor goes directly into the bone, lateral of the tendon.
The provided document is a 510(k) summary for a medical device called SINEFIX, a PEEK implant for soft tissue to bone reattachment in rotator cuff repairs. This document does not describe a study involving an AI/Machine Learning device or a multi-reader multi-case (MRMC) comparative effectiveness study. It focuses on the non-clinical testing of a physical implant.
Therefore, many of the requested categories for AI/ML device studies cannot be answered from this document. However, I can provide information based on the non-clinical testing for the physical SINEFIX device.
Here's a breakdown of the information that can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state numerical acceptance criteria for each test and then provide the exact performance results in a consolidated table. Instead, it lists the types of tests performed and concludes that the results demonstrate substantial equivalence.
Implied Acceptance Criteria (based on the type of tests performed and the conclusion of substantial equivalence):
- Mechanical Performance: The device must perform comparably to predicate devices in terms of mechanical strength, pullout strength, and response to cyclic loading. This would imply meeting or exceeding the mechanical properties of the predicate devices or relevant ASTM/ISO standards.
- Biocompatibility: The device must be biocompatible, meaning it does not cause adverse biological reactions. The acceptance criterion is compliance with ISO 10993-1.
- Sterilization: The device must be able to be effectively sterilized to a sterility assurance level (SAL) of 10^-6. Acceptance criterion is compliance with ISO 11137-1/2.
- Packaging: The packaging must maintain sterility and product integrity for the stated shelf life. Acceptance criterion is compliance with ISO 11607-1/2.
- Reprocessing (for reusable instruments): Reusable instruments (if any are associated with the implant) must be able to be reprocessed effectively. Acceptance criterion is compliance with the FDA guidance on reprocessing medical devices.
Reported Device Performance:
The document states: "Results of the performed tests demonstrate that the SINEFIX are substantially equivalent to legally marketed predicate devices." This is the overarching reported performance. Specific numerical results are not provided in this summary.
2. Sample size used for the test set and the data provenance
For the non-clinical tests (insertion, pullout, cyclic loading, static tensile, dynamic testing, pyrogenicity, endotoxins, packaging, biological evaluation, sterilization validation, reprocessing validation), the sample sizes are not specified in this 510(k) summary. These tests are performed in a laboratory setting (not on human data), so terms like "country of origin of the data" or "retrospective/prospective" are not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable as the tests are non-clinical (mechanical, biological, sterility, etc.) and do not involve "ground truth" derived from expert interpretation of medical images or patient data. The interpretation of results would be done by laboratory personnel according to established scientific and engineering standards and protocols.
4. Adjudication method for the test set
Not applicable, as there is no "test set" in the context of human data or expert review. Non-clinical tests follow laboratory protocols and standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is for a non-AI/ML physical implant, not a diagnostic or prognostic AI tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is for a non-AI/ML physical implant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable for a physical implant's mechanical or material performance. "Ground truth" for these types of tests refers to established physical properties, chemical composition, and performance against defined standards (e.g., maximum load at failure, sterility assurance level).
8. The sample size for the training set
Not applicable. This is for a non-AI/ML physical implant, so there is no "training set."
9. How the ground truth for the training set was established
Not applicable. This is for a non-AI/ML physical implant, so there is no "training set" or "ground truth" in that context.