Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K220756

    Validate with FDA (Live)

    Date Cleared
    2022-06-23

    (100 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Introcan Salety® 2 IV Catheter is inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure or administer fluids and blood intravascularly. The catheters may be used intravascularly with power injectors at a maximum pressure of 325 psi with a luer lock connection only.

    Device Description

    The Introcan Safety® 2 IV Catheter consists of an over-the-needle, peripheral catheter made of radiopaque polyurethane, an integrated septum, and a passive safety needleshielding mechanism. Introcan Safety® 2 is designed to protect clinicians and patients from blood exposure. During needle withdrawal through a septum that seals after the needle has been removed, blood is thus contained within the Introcan Safety® 2 device. The pressure exerted on the needle as it passes through the septum wipes blood from the needle further reducing potential blood exposure. The passive safety needle-shielding mechanism of the Introcan Safety® 2 is located inside the catheter hub. Upon withdrawal of the needle, the safety shield engages as the needle passes through the catheter hub and deploys automatically to shield the needle tip. The safety shield protects during disposal, aiding in the prevention of needlestick injuries. Once the safety shield engages and shields the needle tip, the user is unable to re-insert the needle which aids in the prevention of catheter shearing. This device may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The catheters may be used intravascularly with power injectors for which the maximum pressure setting is 325 psi with a luer lock connection only.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for the Introcan Safety® 2 IV Catheter. As such, it focuses on demonstrating substantial equivalence to a predicate device, rather than proving that the device meets novel acceptance criteria through a standalone study structured like a diagnostic AI performance evaluation. Therefore, many of the requested elements (like MRMC studies, number of experts, training set details) are not applicable or provided in the context of this regulatory submission for a physical medical device.

    However, based on the provided text, I can extract information related to the device's performance and the non-clinical testing conducted to support its substantial equivalence.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established based on international standards (e.g., ISO 10555-1, ISO 10555-5) and internal specifications. The reported performance is presented as the device successfully passing or meeting these criteria, particularly highlighting the changes from the predicate device.

    Performance Metric (Acceptance Criteria Implicit in Standards/Internal Specs)Reported Device Performance (Proposed Introcan Safety® 2 IV Catheter)Comparison/Notes
    Maximum Pressure for Power Injection325 psiMeets the new proposed maximum pressure. (Predicate was 300 psi). Confirmed through bench testing.
    Blood Control CapabilitiesMulti blood control (design changes to septum and septum opener)Meets the enhanced multi-access blood control capabilities. (Predicate had one-time blood control). Confirmed through bench testing.
    Gravity Flow Rate
    18ga x 32 mm105 mL/minSame as Predicate (105 mL/min)
    18ga x 45 mm90 mL/minMeets this flow rate. (Predicate was 100 mL/min). This specific flow rate for 18ga x 45mm changed and was confirmed through bench testing.
    20ga x 25 mm65 mL/minSame as Predicate (65 mL/min)
    20ga x 32 mm60 mL/minSame as Predicate (60 mL/min)
    20ga X 50 mm55 ml/minSame as Predicate (55 ml/min)
    22ga x 25 mm35 mL/minSame as Predicate (35 mL/min)
    24ga x 14 mm26 mL/minSame as Predicate (26 mL/min)
    24ga x 19 mm22 mL/minSame as Predicate (22 mL/min)
    Air tightness (ISO 10555-1)Testing successfully completed.Meets standard as demonstrated by successful completion of testing.
    High pressure (Burst Test) (ISO 10555-1)Testing successfully completed.Meets standard as demonstrated by successful completion of testing.
    Flow rate through capillary (ISO 10555-1)Testing successfully completed.Meets standard as demonstrated by successful completion of testing.
    Projecting length capillary tip (ISO 10555-1)Testing successfully completed.Meets standard as demonstrated by successful completion of testing.
    Catheter hub air aspiration (ISO 10555-1)Testing successfully completed.Meets standard as demonstrated by successful completion of testing.
    Power injection (ISO 10555-1)Testing successfully completed.Meets standard as demonstrated by successful completion of testing, confirming 325psi capability.
    Dynamic tensile load (ISO 10555-5)Testing successfully completed.Meets standard as demonstrated by successful completion of testing.
    Siliconization (Internal Requirement)Testing successfully completed.Meets internal specification.
    Flow rate through metal cannula (Internal Requirement)Testing successfully completed.Meets internal specification.
    Liquid tightness (Internal Requirement)Testing successfully completed.Meets internal specification.
    Force Testing (Internal Requirement)Testing successfully completed.Meets internal specification.
    Blood flashback (Internal Requirement)Testing successfully completed.Meets internal specification.
    Septum opener retention force (Internal Requirement)Testing successfully completed.Meets internal specification.
    Blood exposure (Internal Requirement)Testing successfully completed.Meets internal specification, demonstrating effective blood control.
    Flushing (Internal Requirement)Testing successfully completed.Meets internal specification.
    Cannula withdrawal angle test (Internal Requirement)Testing successfully completed.Meets internal specification.
    Biocompatibility ClassificationExternally communicating blood path, indirect prolonged contact (Same as predicate)Same as predicate, implying continued compliance with biocompatibility requirements.
    MRI labelingMRI Conditional (Same as predicate)Same as predicate, implying continued compliance with MRI safety requirements.
    SterilizationEthylene Oxide (Same as predicate)Same as predicate, implying continued compliance with sterilization requirements.
    Shelf life1 year (Same as predicate)Same as predicate, implying continued compliance with shelf-life requirements.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical bench testing rather than clinical studies with human data. Therefore, the concept of "test set" in the context of diagnostic AI is not directly applicable.

    • Sample Size: Not explicitly stated as a single number across all tests. The testing section lists various tests performed (e.g., ISO standards, internal specifications). For each test, an appropriate number of device samples would have been used to demonstrate compliance, but specific counts are not provided in this summary.
    • Data Provenance: N/A, as this is bench testing of physical devices, not clinical data provenance (country of origin, retrospective/prospective). The tests were conducted internally by B. Braun Medical Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. This is a physical device, and the "ground truth" for its performance is established through adherence to engineering standards and protocols, not through expert human interpretation of clinical data as would be done for a diagnostic AI.

    4. Adjudication method for the test set

    N/A. Adjudication methods (like 2+1, 3+1) are for resolving discrepancies in expert interpretations of data, typically in diagnostic studies. This is not relevant for bench testing of a physical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is a physical medical device, not an AI diagnostic tool. No MRMC study was performed, nor would it be relevant for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This is a physical medical device. It does not involve an algorithm, so "standalone performance" in the AI sense is not applicable.

    7. The type of ground truth used

    The "ground truth" for this device's performance is compliance with established international consensus standards (ISO 10555-1, ISO 10555-5) and internal engineering specifications and test protocols. These standards define measurable performance characteristics (e.g., flow rate, burst pressure, air tightness) that the device must meet.

    8. The sample size for the training set

    N/A. This device is not an AI algorithm and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    N/A. Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220626

    Validate with FDA (Live)

    Date Cleared
    2022-05-14

    (71 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Introcan Safety® IV Catheter is a passive anti-needle stick device to provide venous or arterial access for the infusion of fluids, drugs, and/or blood components, or to facilitate the placement of Vascular Access devices such as guidewires, indwelling central venous catheters, peripherally inserted central catheters, and midline catheters into the vascular system. The catheters may be used with power injectors at a maximum pressure of 325 psi with a luer lock connection only.

    Device Description

    The Introcan Safety® IV Catheter is a passive needle stick prevention device used for arterial and venous access for the infusion of fluids, drugs and/or blood components or to facilitate the placement of Vascular Access devices such as guidewires, indwelling central venous catheters, peripherally inserted central catheters, and midline catheters into the vascular system. The catheters may be used with power injectors for which the maximum pressure setting is 325 psi. The Introcan Safety IV Catheter is available in 14 - 24 gauge sizes, and both winged and non-winged versions and consists of an over-the-needle, peripheral intravascular catheter made of radiopaque polyurethane, and a passive safety needle-shielding mechanism.

    The passive safety needle-shielding mechanism of the Introcan Safety® IV Catheter is located inside the catheter hub. Upon withdrawal of the needle, the safety shield engages as the needle passes through the catheter hub and deploys automatically to shield the needle tip. The safety shield protects during disposal, aiding in the prevention of needlestick injuries. Once the safety shield engages and shields the needle tip, the user is unable to re-insert the needle which aids in the prevention of catheter shearing.

    This device may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The catheters may be used with power injectors with a maximum pressure setting is 325 psi with a luer lock connection only.

    AI/ML Overview

    This document is a 510(k) summary for the Introcan Safety® IV Catheter from B. Braun Medical Inc. It is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. This type of submission focuses on comparing the new device's characteristics and performance to an existing one, rather than presenting a study of the device's independent performance against predefined acceptance criteria for AI or a detailed clinical study with human readers.

    Therefore, the requested information cannot be fully extracted or is not applicable in this context. Here's why and what information can be provided:

    1. A table of acceptance criteria and the reported device performance:

    This document does not refer to "acceptance criteria" in the sense of performance metrics for an AI device (e.g., accuracy, sensitivity, specificity, AUC). Instead, it focuses on demonstrating that the proposed device performs comparably to its predicate device, particularly regarding "power injection capabilities" and "flow rate."

    The relevant performance information is for the power injection capabilities and gravity flow rate, presented as a comparison:

    Performance CharacteristicProposed Device (Introcan Safety® IV Catheter)Predicate Device (K020785) / Reference Device (K021094)Comparison/Difference
    Power InjectionMay be used with power injectors at a maximum pressure of 325 psi (for all gauge sizes)K020785: 14-22g may be used with power injectors at a maximum pressure of 300 psi. K021094: None claimed.Difference: All sizes to max pressure injection of 325psi. Bench testing performed to demonstrate this.
    Gravity Flow RateListed for all gauge sizes (e.g., 14ga x 32mm (1 1/4") 350 mL/min)K020785: Listed for some gauge sizes. K021094: Listed for all gauge sizes.Same (for comparable sizes), but the proposed device combines the full product offering's flow rates.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document refers to "bench testing" but does not specify the sample size (number of catheters tested). It mentions testing for "flowrate and device pressure per ISO 10555-1 Section 4.10 9 AnnexG" and "burst pressure under static conditions per ISO 10555-1 Section 4.10 Annex F." The number of units needed to demonstrate equivalence per these standards would be part of the full submission, but not typically detailed in the public 510(k) summary.
    • Data Provenance: This is not applicable in the typical sense of AI algorithm evaluation (country of origin, retrospective/prospective). The data comes from bench testing of the physical medical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable. This is a physical medical device (IV catheter), not an AI/imaging device requiring expert interpretation for ground truth establishment. The performance is assessed via engineering bench tests against industry standards (ISO).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This is not applicable. There is no human interpretation of data for ground truth establishment as it's a physical device being tested via engineering methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. This is not an AI-powered diagnostic device; it's an IV catheter. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This is not an algorithm, but a physical medical device. Performance is measured through bench testing of the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the device's performance here is defined by standard engineering test methods (ISO 10555-1), which dictate how concepts like "flow rate" and "burst pressure" are measured objectively and quantitatively.

    8. The sample size for the training set:

    This is not applicable. This is a manufactured physical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    This is not applicable for the same reason as point 8.


    Summary of what the document focuses on:

    The document focuses on demonstrating substantial equivalence of the Introcan Safety® IV Catheter to its predicate device (K020785 and K021094). The key "differences" and how they were addressed are:

    • Increased Power Injection Pressure: The proposed device is capable of handling a maximum pressure of 325 psi, an increase from the predicate's 300 psi for certain gauges. This extended capability was supported by bench testing (Power Injection test for flowrate and device pressure per ISO 10555-1).
    • Combined Indications for Use: The new 510(k) combines indications for use across the product family, which doesn't alter the device's fundamental performance but rather its labeling and marketing scope.
    • e-IFU: The Instructions for Use were revised to align with other Introcan products and to utilize an electronic format.

    The conclusion states that "Results of the testing conducted on the proposed devices demonstrate that the Introcan Safety® IV Catheters are substantially equivalent to the predicate device and are as safe and effective as the predicate device." This indicates that the bench testing results met the necessary criteria to show comparable safety and effectiveness, allowing for the 510(k) clearance despite the slight change in power injection capability.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213778

    Validate with FDA (Live)

    Date Cleared
    2022-03-03

    (90 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IV Administration Sets are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

    Device Description

    IV Administration Sets are gravity, single use, disposable, intravenous administration sets used to deliver fluids from a container into a patient's vascular system. These sets may be comprised of various components which are broadly used throughout industry including insertion spike, drip chamber, clamp, luer access device, stopcock, manifold, tubing, luer connections (connector, adaptor), filter, and hand pump. IV Administration sets are configured to ensure the intended use of the device is met.

    AI/ML Overview

    The provided text is a 510(k) summary for an IV Administration Set (K213778). It describes the device, its intended use, comparison to a predicate device, and nonclinical testing performed to demonstrate substantial equivalence.

    However, the request primarily asks about acceptance criteria and a study proving a device meets acceptance criteria, specifically in the context of an AI/Software as a Medical Device (SaMD). The document furnished details on a physical medical device (IV Administration Set) and its substantial equivalence to a predicate through bench testing and adherence to ISO standards.

    Therefore, the information required to answer the prompt regarding AI/SaMD acceptance criteria and proof of performance (e.g., sample size, expert ground truth, MRMC study, standalone performance) is not present in the provided text. The document clearly states: "No clinical testing was performed as these devices does not require clinical studies to demonstrate substantial equivalence with the predicate device."

    To reiterate, the provided document does not contain the information requested in your prompt regarding AI/SaMD performance studies and acceptance criteria. This device is a physical IV administration set, not an AI/SaMD.

    If you can provide a document related to an AI/SaMD, I would be happy to help answer your questions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213085

    Validate with FDA (Live)

    Date Cleared
    2021-10-25

    (31 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Introcan Safety® 3 Closed Intravascular Catheter is inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure or administer fluids and blood intravascularly. The catheters may be used with power injectors at a maximum pressure of 325 psi with luer lock connection only.

    Device Description

    The Introcan Safety® 3 Closed IV Catheter consists of an over-the-needle, peripheral intravascular catheter made of radiopaque polyurethane, an integrated bidirectional septum, a stabilization platform, and a passive safety needle-shielding mechanism. Introcan Safety® 3 design is a closed IV catheter since it protects clinicians and patients from blood exposure. Since the needle is withdrawn through a septum that seals after the needle has been removed, blood is thus contained within the Introcan Safety® 3 device. The pressure exerted on the needle as it passes through the septum wipes blood from the needle further reducing potential blood exposure. The Introcan Safety® 3 catheter has an integrated stabilization is designed to improve catheter stability while minimizing catheter movement within the vessel. The device controls the flow of blood, aiding in the prevention of blood exposure. The passive safety needle-shielding mechanism of the Introcan Safety® 3 is located inside the catheter hub. Upon withdrawal of the needle, the safety shield engages as the needle passes through the catheter hub and deploys automatically to shield the needle tip. The safety shield protects during disposal, aiding in the prevention of needlestick injuries. Once the safety shield engages and shields the needle tip, the user is unable to re-insert the needle which aids in the prevention of catheter shearing.

    AI/ML Overview

    The document is a 510(k) summary for the Introcan Safety® 3 Closed IV Catheter, indicating that the device has been found substantially equivalent to a previously cleared predicate device (K182870). The primary difference for the new submission is an extension of the power injection capabilities to higher pressure rates for all gauge sizes.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly laid out in a table format with specific pass/fail thresholds for each test in the provided document. However, the document lists "Nonclinical Testing" that was performed and states that the "Results of the testing conducted on the proposed devices demonstrate that the Introcan Safety® 3 Closed IV Catheters are substantially equivalent to the predicate device and are as safe and effective as the predicate devices."

    Based on the information provided, we can infer the tested parameters and their implied "acceptance" through comparison to the predicate. The "Comparison" column in the Technological Characteristics table effectively serves as the reported device performance relative to the predicate, with the implied acceptance criteria being "Same" or that the difference does not raise additional safety and effectiveness questions.

    Acceptance Criteria (Implied by Predicate & Standards)Reported Device Performance (Proposed Device)
    Indications for Use: Short term use to sample blood, monitor blood pressure, administer fluids/blood intravascularly. Power injectors at max pressure 300 psi for 18-24 gauge.Indications for Use: Short term use to sample blood, monitor blood pressure, administer fluids/blood intravascularly. Power injectors at max pressure 325 psi for 14-24 gauge.
    Configuration: Single Lumen, Tapered TipSingle Lumen, Tapered Tip
    Material Composition: (Catheter Tube, Hub, Needle, Safety Clip, Septum, Septum Opener, Septum Housing)Identical to predicate (Polyurethane, Polypropylene, Stainless steel, MABS, Stainless steel, Silicone or Polyisoprene Rubber, Polyoximethylene, Polypropylene)
    Catheter Gauge Sizes: 14ga-24ga14ga-24ga
    Catheter Length: 3/4" (19mm) - 2" (50mm)3/4" (19mm) - 2" (50mm)
    Gravity Flow Rate: (Specific values for each gauge x length)Identical to predicate for all listed gauge x length combinations (e.g., 14ga x 32mm: 325 mL/min; 24ga x 19mm: 22 mL/min)
    Sterilization: Ethylene OxideEthylene Oxide
    Shelf-Life: 5 year5 year
    Power Injection (Performance at specified pressure/flow): Max 300 psi for 18-24gBench testing performed to demonstrate acceptable performance at 325 psi for 14-24ga, concluding that differences do not raise additional questions of safety and effectiveness. (Specifically, "Power Injection test for flowrate and device pressure per ISO 10555-1 Section 4.10 Annex G" and "Test for burst pressure under static conditions per ISO 10555-1 Section 4.10 Annex F" were conducted.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of devices) used for each nonclinical test (Power Injection and Burst Pressure). It only states that "Bench testing performed on Introcan Safety® 3 Closed IV Catheters supports substantial equivalence of the proposed device."

    • Sample Size: Not explicitly stated for each test.
    • Data Provenance: The data is from "bench testing" performed by the manufacturer, B. Braun Medical Inc., located in Allentown, PA, USA. This is retrospective data collected for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not applicable as the study described is nonclinical bench testing of a physical medical device, not a diagnostic AI system requiring expert consensus for ground truth.

    4. Adjudication Method for the Test Set

    • This information is not applicable as it refers to clinical studies or studies using human interpretation, which is not the case here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    • No, an MRMC comparative effectiveness study was not done. The study was nonclinical bench testing.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This is not applicable as the device is a physical medical catheter, not an algorithm or AI system. The "standalone" performance here refers to the device's physical and mechanical properties. The bench tests conducted (Power Injection test, Burst Pressure test) represent the standalone performance of the device under simulated conditions.

    7. Type of Ground Truth Used

    The "ground truth" for the nonclinical tests is established by:

    • Engineering specifications and recognized international standards: The tests were performed "per ISO 10555-1 Section 4.10 Annex G" (Power Injection) and "per ISO 10555-1 Section 4.10 Annex F" (Burst Pressure). These standards define the methodology and expected performance characteristics for intravascular catheters.
    • Comparison to the predicate device: The new device's performance is gauged against the established and cleared performance of its predicate counterpart. The objective is to demonstrate that the proposed changes (higher pressure for power injection) do not compromise safety or effectiveness compared to the predicate, and meet the new specified performance.

    8. Sample Size for the Training Set

    • This is not applicable as there is no mention of a training set. This is a nonclinical bench test of a physical device, not an AI/ML algorithm that requires training.

    9. How the Ground Truth for the Training Set Was Established

    • This is not applicable as there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173361

    Validate with FDA (Live)

    Date Cleared
    2018-06-26

    (243 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IV Administration Sets are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

    Device Description

    IV Administration Sets are gravity, single use, disposable, intravenous administration sets used to deliver fluids from a container into a patient's vascular system. These sets may be comprised of various components which are broadly used throughout industry including insertion spike, drip chamber, clamp, needleless luer access device, injection site, check valve, tubing, clamp, and luer connection (connector, adaptor). IV Administration sets are configured to ensure the intended use of the device is met.

    AI/ML Overview

    The provided document is a 510(k) summary for an IV Administration Set, not an AI-powered medical device. Therefore, the information required to answer your query (acceptance criteria and study details for an AI/algorithm-based device) is not present in the given text.

    The document focuses on demonstrating substantial equivalence of a conventional medical device (IV administration set) to a predicate device, based on:

    • Similar indications for use and intended use.
    • Similar principle of operation and fundamental scientific technology.
    • Similar component types and materials (with specific testing for material differences).
    • Compliance with recognized performance standards (e.g., ISO 8536-4, ISO 10993-1).
    • Successful completion of non-clinical performance and biocompatibility testing.

    The concept of "acceptance criteria" in this context refers to meeting the performance specifications outlined in the relevant ISO standards and the functional and performance tests listed. There is no mention of an algorithm, AI, or comparative effectiveness studies involving human readers and AI assistance.

    To summarize why I cannot provide the requested information:

    1. Nature of the Device: The device is a physical medical administration set, not a software or AI-based diagnostic/assistive tool.
    2. Type of Submission: This is a 510(k) premarket notification for substantial equivalence, not a de novo or PMA submission that would typically involve extensive clinical performance data for novel AI devices.
    3. No AI/Algorithm Mentioned: The document does not describe any AI component, algorithm, or software functionality.
    4. No Human-in-the-Loop Study: Since there's no AI, there's no MRMC study, no human reader improvement data, and no standalone algorithm performance.
    5. No Ground Truth Establishment: The concept of "ground truth" for diagnostic accuracy is not applicable to a physical IV administration set.

    Therefore, I cannot populate the table or answer the specific questions related to AI device evaluation based on the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1