Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K212886

    Validate with FDA (Live)

    Date Cleared
    2022-02-16

    (159 days)

    Product Code
    Regulation Number
    874.4680
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Broncoflex® Agile and Broncoflex® Vortex

    This video-bronchoscope is intended to provide an optical display of the pulmonary tract using a monitor (Screeni®) and to be used with endotherapy accessories and instruments.

    This video-bronchoscope is designed for use exclusively in a hospital environment.

    Endotherapy accessories and instruments

    The effective length of an endoscopic instrument should be at least 30 cm greater than the effective length of the endoscope.

    ModelMinimum compatibleendotracheal tube sizeMinimum compatible duallumen endo-bronchial tube sizeMaximum size ofendotherapy instruments
    Broncoflex Agile$\ge 5.0$ mm$\ge 35$ Fr.$\le 1.2$ mm
    Broncoflex Vortex$\ge 6.0$ mm$\ge 41$ Fr.$\le 2.6$ mm

    Patient Population

    The Broncoflex can only be used on adult patients.

    Screeni®

    This device is designed for use exclusively in a hospital environment, in combination with an Axess Vision Broncoflex® video-bronchoscope.

    Device Description

    Broncoflex Agile and Broncoflex Vortex are a single-use video-bronchoscope which is part of a system made up of the endoscope (Broncoflex) and its reusable display monitor (Screeni). The videobronchoscope is intended to provide an optical display of the pulmonary tract using a monitor (Screeni) and to be used with endotherapy accessories and instruments. The applied part of the medical device is the whole of the flexible tubing containing the insertion tube, the articulated section and the distal tip (BF type applied part: BF). Broncoflex is sterile and supplied ready to use in its sealed packaging (sterilization method ethylene oxide). The Screeni is a reusable nonsterile medical device and is a video processor with integrated touch interface designed to display live imaging data captured by Broncoflex Agile and Broncoflex Vortex. The consumables (Broncoflex) are supplied in sets of 5 in a carton. Inside this box, each sterilized consumable is placed in a single sterile pouch. Screeni is supplied in a carton with all the accessories (quick mounting bracket, knurled screw, power supply, EU/US/UK/AU adaptors and video cable (HDMI/DVI)). Broncoflex Agile and Broncoflex Vortex are the same devices except for some dimensional aspects such as external diameter of insertion tube, distal tip outer diameter and working channel inner diameter. Indeed, Broncoflex Agile has an applied part with a smaller diameter than the Broncoflex Vortex.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria for a device's performance that would typically be described in a study. Instead, it describes a 510(k) submission for the Broncoflex Agile, Broncoflex Vortex, and Screeni devices, asserting their substantial equivalence to a predicate device.

    The document focuses on:

    • Device Description: Details the components of the Broncoflex system (single-use video-bronchoscope and reusable display monitor).
    • Indications for Use: Specifies the medical application and patient population.
    • Comparison to Predicate Device: Highlights and tabulates the technological characteristics of the subject device and the predicate device (which is a previously cleared version of the same product). It explicitly states that "The Broncoflex Vortex and Broncoflex Agile are the same as the predicate devices" in most listed areas, and the Screeni "is the same as the predicate device" in most areas, with one difference (HDMI port) that "has no impact on performance, safety nor effectiveness."
    • Non-Clinical Performance Data: Mentions that the system "meets all the requirements for overall design, sterilization, biocompatibility, sterile packaging integrity, shelf life, transportation, electromagnetic compatibility and electrical safety results confirming that the design output meets the design inputs and specifications for the device." It also lists the standards against which testing was performed (IEC 60601-1:2012, IEC 60601-2-18:2009, IEC 60601-1-2:2014/A1:2020 (Ed. 4.1), and IEC 62304:2006/A1 & FDA Guidance for software).

    The information provided does not include:

    • A specific table of acceptance criteria for performance metrics (such as sensitivity, specificity, accuracy) nor reported device performance against such criteria.
    • Details of a study involving a test set, including sample size, data provenance, number of experts for ground truth, or adjudication methods.
    • Any multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone algorithm performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Training set sample size or how its ground truth was established.

    The document primarily demonstrates substantial equivalence through technical comparison and compliance with relevant safety and performance standards for medical devices, rather than through a clinical performance study with specific acceptance criteria for diagnostic or algorithmic performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202180

    Validate with FDA (Live)

    Date Cleared
    2020-10-22

    (79 days)

    Product Code
    Regulation Number
    874.4680
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Broncoflex Agile and Vortex: This video-bronchoscope is intended to provide an optical display of the pulmonary tract using a monitor (Screeni®) and to be used with endotherapy accessories and instruments. This video-bronchoscope is designed for use exclusively in a hospital environment.
    Screeni: This device is designed for use exclusively in a hospital environment, in combination with an Axess Vision Broncoflex @ video-bronchoscope.

    Device Description

    Broncoflex® Agile and Broncoflex Vortex are a single-use video-bronchoscope which is part of a system made up of the endoscope (Broncoflex) and its reusable display monitor (Screeni). The applied part of the medical device is the whole of the flexible tubing containing the insertion tube, the articulated section and the distal tip (BF type applied part: BF). Broncoflex is sterile and supplied ready to use in its sealed packaging (sterilization method ethylene oxide). The Screeni is a reusable non-sterile medical device and is a video processor with interface touch interface designed to display live imaging data captured by Broncoflex Agile and Broncoflex Vortex.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study that proves the device meets the acceptance criteria in the format requested. The document is a 510(k) summary for a medical device (Broncoflex Agile, Broncoflex Vortex, and Screeni) and primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and non-clinical performance data.

    Here's an breakdown of the available information and what is missing:

    The document states:

    • "The Flexible Bronchoscope System passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device"
    • "Image quality comparison image quality was assessed same or better compared to predicate devices"
    • "Cleaning of display monitor – met performance requirements"
    • "Functionality tests (conditions in use) and climatic tests (transport and storage conditions) -. image present and number of deflections up/down specified during 5 defined cycles"

    However, it does not provide:

    • A table of specific acceptance criteria.
    • Reported device performance against those specific criteria (beyond general statements like "met performance requirements" or "same or better").
    • Details about the study design, such as sample sizes for test sets, data provenance, number or qualifications of experts, or adjudication methods for any image quality comparison.
    • Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies, effect sizes with or without AI assistance, or standalone algorithm performance.
    • The type of ground truth used (e.g., pathology, expert consensus, outcomes data) for any comparative studies.
    • Sample size for training sets or how ground truth for training sets was established, as this device does not appear to employ AI as a primary component.

    Therefore, I cannot fully answer your request based solely on the provided text, as much of the requested information is absent.


    Based on the provided document, here's what can be extracted, though it's incomplete for your specific questions:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Inferred from testing)Reported Device Performance Summary (from text)
    Overall design requirementsMet requirements
    SterilizationMet requirements (Ethylene oxide process validated per ISO 11135:2014)
    BiocompatibilityPassed testing per ISO 10993-1:2018
    Sterile packaging integrityValidated per ISO 11607:2019 and ASTM F 1980:2016 Standards
    Shelf lifeValidated per ISO 11607:2019 and ASTM F 1980:2016 Standards
    TransportationPassed testing per ISTA 3A:2018 and ISTA 3E:2017
    Electromagnetic Compatibility (EMC)Passed testing per IEC 60601-1-2:2014 (4th Ed.)
    Electrical safetyPassed testing per IEC 60601-1:2012 and IEC 60601-2-18:2009
    Software verification and validationPassed testing per IEC 62304:2006/A1 & FDA Guidance
    Compliance to Endoscope StandardsCompliant to ISO 8600-1:2015, ISO 8600-3:2019 and ISO 8600-4:2014
    Photobiological safetyCompliant per IEC 62471:2006
    Compatibility with endotherapy accessoriesDemonstrated passage of instrument through endoscope and vice-versa
    Image qualityAssessed same or better compared to predicate devices
    Cleaning of display monitorMet performance requirements
    Functionality (conditions in use)Image present and specified deflections up/down during 5 cycles
    Climatic tests (transport and storage)Met requirements

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the document for any specific clinical or comparative performance test. The non-clinical tests (e.g., biocompatibility, electrical safety) would have their own sample sizes, but these are not detailed.
    • Data Provenance: Not specified for any comparative or clinical data. The submitter is Axess Vision Technology, located in Joué-lès-Tours, France.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified. The document mentions "Image quality comparison image quality was assessed same or better compared to predicate devices" but does not detail how this assessment was done, by whom, or with what ground truth.

    4. Adjudication method for the test set:

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not mentioned. This device system (flexible bronchoscope and display monitor) is not described as having an AI component that would necessitate such a study. The focus is on physical and performance characteristics, and equivalence to existing predicate devices.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not mentioned. Again, an AI algorithm is not a described component of this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not specified for any comparative performance claims. For non-clinical tests, the "ground truth" would be the specified requirements and standards.

    8. The sample size for the training set:

    • Not applicable, as an AI component is not described.

    9. How the ground truth for the training set was established:

    • Not applicable, as an AI component is not described.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1