Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213775

    Validate with FDA (Live)

    Date Cleared
    2022-04-24

    (143 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glovatex™ White Latex Glove is a disposable device intended for Medical purpose that is worn on the examiner's hands or finger to prevent contamination patient and examiner.

    The Glovanil™ Blue Nittile Glove is a disposable device intended for Medical purpose that is worn on the examiner's hands or finger to prevent contamination patient and examiner.

    Device Description

    Glovatex™ White Latex Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D 3578-05 (Reapproved 2015), Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are in white color, non-sterile and powder free.

    Glovanil™ Blue Nitrile Gloves are manufactured to meet all the current specifications listed under the ASTM Specification ASTM D6319 - 19 Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from Nitrile compound. These gloves are in Blue color, non-sterile and powder free.

    AI/ML Overview

    This document is a 510(k) premarket notification for medical gloves, and as such, it does not describe an AI/ML device. Therefore, the questions related to AI/ML device performance (such as effect size, standalone performance, ground truth establishment, training set size, and expert qualifications for ground truth) are not applicable.

    The document focuses on demonstrating substantial equivalence to predicate devices through physical and biocompatibility testing.

    Here's the information extracted concerning the acceptance criteria and device performance for the two glove products described:

    1. Table of Acceptance Criteria and Reported Device Performance

    Glovatex™ White Latex Gloves (Predicate: K192329)

    CharacteristicsStandardsAcceptance Criteria (from Standard)Reported Device Performance (Subject Device)Comparison to Predicate
    Dimensions - Length (Small)ASTM D3578-19, Clause 8.4> 220 mm245 mmSimilar to Predicate
    Dimensions - Length (Medium)ASTM D3578-19, Clause 8.4> 220 mm245.6 mmSimilar to Predicate
    Dimensions - Length (Large)ASTM D3578-19, Clause 8.4> 220 mm247.5 mmSimilar to Predicate
    Dimensions - Length (X-Large)ASTM D3578-19, Clause 8.4> 220 mm244.3 mmSimilar to Predicate
    Dimensions - Width (Small)ASTM D3578-19, Clause 8.470-90mm84.4 mmSimilar to Predicate
    Dimensions - Width (Medium)ASTM D3578-19, Clause 8.485-105mm95.3 mmSimilar to Predicate
    Dimensions - Width (Large)ASTM D3578-19, Clause 8.4101-121mm103.7 mmSimilar to Predicate
    Dimensions - Width (X-Large)ASTM D3578-19, Clause 8.4110-130mm111.5 mmSimilar to Predicate
    Dimensions - Thickness - FingerASTM D3578-19, Clause 8.4> 0.08 mmSmall: 0.14, Medium: 0.14, Large: 0.15, X-Large: 0.15 mmSimilar to Predicate
    Dimensions - Thickness - PalmASTM D3578-19, Clause 8.4> 0.08 mmSmall: 0.12, Medium: 0.11, Large: 0.11, X-Large: 0.12 mmSimilar to Predicate
    Physical Properties - Tensile Strength (Before Ageing)ASTM D3578-19, Clause 8.5.1> 18 MpaSmall: 24.6, Medium: 23.1, Large: 23.7, X-Large: 22 MpaSimilar to Predicate
    Physical Properties - Tensile Strength (After Ageing)ASTM D3578-19, Clause 8.5.2> 14 MpaSmall: 23.6, Medium: 21.8, Large: 20.9, X-Large: 22.6 MpaSimilar to Predicate
    Physical Properties - Elongation at Break (Before Ageing)ASTM D3578-19, Clause 8.5.1> 650%Small: 1316, Medium: 1346, Large: 1330, X-Large: 1239 %Similar to Predicate
    Physical Properties - Elongation at Break (After Ageing)ASTM D3578-19, Clause 8.5.2> 500%Small: 1356, Medium: 1385, Large: 1371, X-Large: 1340 %Similar to Predicate
    Powder Free ResidueASTM D6124-06/2017≤ 2 mg/gloveSmall: 0.39, Medium: 0.35, Large: 0.37, X-Large: 0.42 mg/gloveSimilar to Predicate
    Biocompatibility - In vitro CytotoxicityISO10993-5:2009(E)Non-cytotoxicNon-cytotoxicSame as Predicate
    Biocompatibility - Primary Skin IrritationISO 10993-10:2010(E)Not-irritantNot-irritantSame as Predicate
    Biocompatibility - Dermal SensitizationISO 10993-10:2010(E)Not-sensitizerNot-sensitizerSame as Predicate
    Biocompatibility - Material mediated PyrogenicityISO 10993-11:2017(E)/USP 41<151>Non-pyrogenicNon-pyrogenicSame as Predicate
    Freedom From HolesASTM D5151-19PassesPassesSame as Predicate
    Intended Use-Prevent contamination between patient and examinerSameSame as Predicate
    Material-Natural LatexNatural LatexSame as Predicate
    Color-Blue (Predicate)White (Subject)Different from Predicate
    Texture-Finger TextureFinger TextureSame as Predicate
    SizeASTM D3578-19Small, Medium, Large, X-Large, XX-LargeSmall, Medium, Large, X-LargeSimilar (XX-Large not available for subject)
    Single UseMedical Glove Guidance Manual - LabelingSingle UseSingle UseSame as Predicate
    Sterile/non-sterile-Non sterileNon sterileSame as Predicate
    Powder/Powder free-Powder freePowder freeSame as Predicate
    Label and LabelingFDA Label requirementsMeets FDA's label and labeling requirementsMeets FDA's label and labeling requirementsSame as Predicate

    Glovanil™ Blue Nitrile Gloves (Predicate: K200326)

    CharacteristicsStandardsAcceptance Criteria (from Standard)Reported Device Performance (Subject Device)Comparison to Predicate
    Dimensions - Length (Small)ASTM D6319-19, Clause 7.4.2> 220 mm239.5 mmSimilar to Predicate
    Dimensions - Length (Medium)ASTM D6319-19, Clause 7.4.2> 230 mm239.7 mmSimilar to Predicate
    Dimensions - Length (Large)ASTM D6319-19, Clause 7.4.2> 230 mm238.3 mmSimilar to Predicate
    Dimensions - Length (X-Large)ASTM D6319-19, Clause 7.4.2> 230 mm239.3 mmSimilar to Predicate
    Dimensions - Width (Small)ASTM D6319-19, Clause 7.4.370-90mm81.7 mmSimilar to Predicate
    Dimensions - Width (Medium)ASTM D6319-19, Clause 7.4.385-105mm88.8 mmSimilar to Predicate
    Dimensions - Width (Large)ASTM D6319-19, Clause 7.4.3100-120mm104.3 mmSimilar to Predicate
    Dimensions - Width (X-Large)ASTM D6319-19, Clause 7.4.3110-130mm111.3 mmSimilar to Predicate
    Dimensions - Thickness - FingerASTM D6319-19, Clause 7.4.4> 0.05 mmSmall: 0.14, Medium: 0.14, Large: 0.14, X-Large: 0.14 mmSimilar to Predicate
    Dimensions - Thickness - PalmASTM D6319-19, Clause 7.4.4> 0.05 mmSmall: 0.11, Medium: 0.10, Large: 0.09, X-Large: 0.10 mmSimilar to Predicate
    Physical Properties - Tensile Strength (Before Ageing)ASTM D6319-19, Clause 7.5.1> 14 Mpa (Predicate met, Specific values not given for predicate.)Small: 22.8, Medium: 24.3, Large: 24.0, X-Large: 22.2 MpaSimilar to Predicate
    Physical Properties - Tensile Strength (After Ageing)ASTM D6319-19, Clause 7.5.1> 14 MpaSmall: 22.8, Medium: 24.3, Large: 24.0, X-Large: 22.2 MpaSimilar to Predicate
    Physical Properties - Elongation at Break (Before Ageing)ASTM D6319-19, Clause 7.5.1> 500% (Predicate met, Specific values not given for predicate.)Small: 791, Medium: 848, Large: 835, X-Large: 852 %Similar to Predicate
    Physical Properties - Elongation at Break (After Ageing)ASTM D6319-19, Clause 7.5.1> 400%Small: 791, Medium: 848, Large: 835, X-Large: 852 %Similar to Predicate
    Powder Free ResidueASTM D6124-06/2017≤ 2 mg/gloveSmall: 1.18, Medium: 1.91, Large: 0.81, X-Large: 1.17 mg/gloveSimilar to Predicate
    Biocompatibility - In vitro CytotoxicityISO10993-5:2009(E)Non-cytotoxic (Predicate: No information stated, Subject: Non-cytotoxic)Non-cytotoxicSimilar to Predicate
    Biocompatibility - Primary Skin IrritationISO 10993-10:2010(E)Not-irritantNot-irritantSimilar to Predicate
    Biocompatibility - Dermal SensitizationISO 10993-10:2010(E)Not-sensitizerNot-sensitizerSimilar to Predicate
    Biocompatibility - Acute Systemic ToxicityISO 10993-11:2017(E)Did not induce a systemic toxicityDid not induce a systemic toxicitySimilar to Predicate
    Biocompatibility - Material mediated PyrogenicityISO 10993-11:2017(E)/USP 41<151>Non-pyrogenic (Predicate: N/A, Subject: Did not produce a pyrogenic response)Did not produce a pyrogenic responseDifferent from Predicate (More testing for subject)
    Freedom From HolesASTM D5151-19PassesPassesSame as Predicate
    Intended Use-Prevent contamination between patient and examinerSameSame as Predicate
    Material-Nitrile compoundNitrile compoundSame as Predicate
    Color-Aqua Green (Predicate)Blue (Subject)Different from Predicate
    SizeASTM D3578-19Small, Medium, Large, X-Large, XX-LargeSmall, Medium, Large, X-LargeSimilar (XX-Large not available for subject)
    Single UseMedical Glove Guidance Manual - LabelingSingle UseSingle UseSame as Predicate
    Sterile/non-sterile-Non sterileNon sterileSame as Predicate
    Powder/Powder free-Powder freePowder freeSame as Predicate
    Label and LabelingFDA Label requirementsMeets FDA's label and labeling requirementsMeets FDA's label and labeling requirementsSame as Predicate

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each physical and chemical test. It only provides the performance results (e.g., average length, tensile strenth values) across different glove sizes. The provenance of the data is from laboratory testing performed on the "subject device" (the Glovatex White Latex Gloves and Glovanil Blue Nitrile Gloves) to demonstrate compliance with referenced ASTM and ISO standards. This is prospective testing for the purpose of the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as this is not an AI/ML device study. Performance is measured against established industry standards (ASTM, ISO).

    4. Adjudication method for the test set

    Not applicable, as this is not an AI/ML device study requiring expert adjudication of outputs. Compliance is determined by whether the test results meet the numerical and qualitative specifications of the referenced standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an AI/ML device.

    7. The type of ground truth used

    The "ground truth" for the device performance is the set of specifications and requirements outlined in the referenced consensus standards:

    • ASTM D3578-19 (Standard Specification for Rubber Examination Gloves)
    • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application)
    • ASTM D6124-06/2017 (Standard Test Method for Residual Powder on Medical Gloves)
    • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ISO 10993-5:2009(E) (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)
    • ISO 10993-10:2010(E) (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
    • ISO 10993-11:2017(E) (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)
    • USP 41<151> (Pyrogen Test)

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1