510(k) Data Aggregation

    K Number
    K230986

    Validate with FDA (Live)

    Device Name
    Bladder Scanner
    Date Cleared
    2023-12-29

    (267 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bladder Scanner is B-mode pulsed-echo ultrasound device adopts a 3D mechanical fan scanning probe for the ultrasonic scanning for bladder and measures the bladder volume from the abdominal surface of ultrasonic imaging. The Bladder Scanner is intended to be used only in hospital and by qualified medical professionals.

    Device Description

    The Bladder Scanner (Model: AS-2) is a hand-held battery-operated device, it provides non-invasive bladder volume measurement utilizing real-time ultrasound imaging. The proposed device consists of the main unit (include 3D probe), battery power adapter and USB charging cable.

    AI/ML Overview

    This FDA 510(k) summary provides information for a device called "Bladder Scanner" (Model: AS-2) by Avantsonic Technology Co., Ltd. The document primarily focuses on establishing substantial equivalence to a predicate device (K201316 Bladder Scanner by Suzhou Peaksonic Medical Technology Co., Ltd.) rather than detailing an independent study against acceptance criteria. However, it does state the acceptance criteria for volume measurement accuracy and reports whether the device met these criteria through non-clinical testing.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Bladder Volume Measurement Accuracy)Reported Device Performance
    ≤ ±10 ml (measured volume < 100ml)Met the requirements
    ≤ ±7.5% (measured volume ≥ 100ml)Met the requirements

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: The document states that the "Bladder Scanner volume measurement accuracy test" was conducted for the measurement range of 0 ml ~ 999 ml, implying a range of volumes were tested to verify accuracy across this spectrum. However, a specific number of samples (e.g., number of measurements, number of phantoms/models used) is not provided.
    • Data Provenance: The document does not specify the country of origin for the data or whether the tests were retrospective or prospective. Given that it's a non-clinical test, it likely refers to laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This information is not provided. The document mentions non-clinical tests were conducted, which typically involve measuring known volumes using standard metrology rather than expert judgment on images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is not provided. As it refers to non-clinical testing of volume measurement accuracy, an adjudication method in the context of expert review is not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical study is included in this submission." The device's primary function is to measure bladder volume using ultrasound, not to assist human readers in interpreting complex images or diagnoses.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in essence, a standalone evaluation was performed in terms of the device's ability to measure volume. The "Bladder Scanner volume measurement accuracy test" assesses the device's algorithmic and hardware performance in calculating volume against known standards, independent of human interpretation or assistance during the measurement process itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The ground truth for the "Bladder Scanner volume measurement accuracy test" would have been known, precisely measured volumes. These are typically established using phantoms or controlled liquid volumes where the true volume is known and verifiable through other metrological means.

    8. The sample size for the training set:

    • This information is not provided. The document describes a "Bladder Scanner" which is a hardware device that utilizes ultrasound principles. While it inherently relies on algorithms for volume calculation, the document does not detail any specific "training set" in the context of machine learning or AI. It's more likely based on established physics and signal processing principles rather than a trained AI model in the modern sense.

    9. How the ground truth for the training set was established:

    • This information is not provided and is likely not applicable in the context of the information given. As mentioned above, the device's mechanism is described in terms of "B-mode pulsed-echo ultrasound" and "3D mechanical fan scanning probe," suggesting a more traditional ultrasound technology rooted in physics rather than a deep learning model requiring a distinct training set with established ground truth labels.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171528

    Validate with FDA (Live)

    Date Cleared
    2018-01-24

    (244 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bladder Scanner (model: PadScan Z3, PadScan Z3, PadScan Z5) is B-mode pulsed-echo ultrasound device. It intended as a handheld battery-operated device. The Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder Volume noninvasively. The Bladder Scanner is intended to be used only by qualified medical professionals.

    Device Description

    The PadScan Series manufactured by AvantSonic Technology Co., Ltd. provides non-invasive volume bladder measurement utilizing real-time ultrasound imaging and measurement. The equipment consists of the main unit, 3D probe, battery and adapter. It features: Two Operation Modes: Expert Mode and Easy Mode, Non-invasive, comfortable, correct, reliable, fast and simple operation, Printouts with ultrasound images and various parameters through PC software, Touch screen keyboard operation, Urine volume setting and alarm setting, Multi-language selection, Combined power supply with AC adapter and a built-in battery. The difference between these models is the size of the LCD screen and enclosure structure. PadScan DS3 is provided 7-inch LCD screen. PadScan Z3 is provided 7-inch LCD screen and LCD screen stand. PadScan Z5 is provided 8-inch LCD screen and LCD screen which has a handle.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the AvantSonic Bladder Scanner, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Predicate Device K131227)Reported Device Performance (PadScan DS3, Z3, Z5)
    Volume Measurement Accuracy±15%, ±15ml±10%, ±10ml

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "Clinical testing is not required." This indicates that no clinical test set was used for the substantial equivalence determination. The performance criteria were based on comparisons to the predicate device's specifications.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable, as no clinical testing with a test set requiring expert ground truth was conducted. Performance was based on device specifications and non-clinical testing.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical testing with a test set requiring adjudication was conducted.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned or performed. The submission relies on demonstrating substantial equivalence to a predicate device through non-clinical testing and shared intended use/technological characteristics.

    6. Standalone (Algorithm Only) Performance Study

    While the device calculates bladder volume non-invasively, the provided information does not detail a standalone algorithm performance study. The performance is assessed based on the device's ability to measure bladder volume within specified accuracy against a reference.

    7. Type of Ground Truth Used

    For the reported device performance regarding volume measurement accuracy, the ground truth would have been established through a controlled measurement method (e.g., using known volumes of liquid in a phantom or other controlled environment) to calibrate and verify the device's calculations. This is implied by the accuracy specification but not explicitly described as "pathology" or "outcomes data."

    8. Sample Size for the Training Set

    Not applicable, as this is a medical device submission based on substantial equivalence, not a machine learning model requiring a training set in the typical sense. The device's "training" is inherent in its design and calibration processes to meet the specified accuracy.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8. The device's volume calculation mechanism is based on ultrasound physics and established algorithms for bladder volume estimation, not trained on a distinct dataset with "ground truth" labels in the context of machine learning. The accuracy is verified through metrological testing.