510(k) Data Aggregation

    K Number
    K173760

    Validate with FDA (Live)

    Date Cleared
    2018-03-22

    (101 days)

    Product Code
    Regulation Number
    874.4680
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monarch Endoscopy Platform (Monarch Platform) and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

    Device Description

    The Monarch™ Endoscopy Platform (Monarch Platform) is intended to be used by qualified physicians to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Monarch Platform consists of three major components, (1) Monarch™ Endoscopy Cart, (2) Monarch™ Endoscopy Tower, and (3) Monarch™ Bronchoscope System, and working channel instruments.

    The Monarch Cart provides support for the effector arms. It includes two robotic arms and the electronic systems required to power and operate the robotic system.

    The Monarch Tower is the primary user (i.e. physician) procedural display interface. It contains a monitor for user viewing and computers running the system software. The tower provides connectivity for the bronchoscope camera and lighting, as well as the fluidics system. In addition, the tower includes an endoscopic controller that allows the user to control the system during a procedure.

    The user controls the system with an endoscopic controller which transmits user inputs through the electromechanical system to the end effectors. The Monarch Bronchoscope System comprising of a bronchoscope and sheath is attached at the end effector of a robotic arm with multiple degrees of freedom. The flexible bronchoscope has a working channel and a camera at the tip. The bronchoscope has an articulated tip that can bend in all directions. The working channel of the bronchoscope is used for irrigation, aspiration and to deliver the working channel instruments.

    The single-use manually controlled working channel instruments compatible with the Monarch Platform include the Aspirating Biopsy Needle, Biopsy Forceps, and Cytology Brush. All three have the same technological characteristics as the respective predicates and reference devices, that is, they are intended to operate in the same manner for collection of tissue or cells in the patient's lungs through a flexible endoscope or other working channel.

    The Aspirating Biopsy Needle is comprised of a polymeric shaft with a needle joined to the distal end. A Luer fitting at the proximal end of the shaft allows the user to provide suction to the needle via the supplied syringe. A handle mechanism interfaces with the shaft to provide extension and retraction of the needle, and an outer jacket provides protection when the needle is not in use.

    The Cytology Brush is comprised of an outer sheath and an inner catheter assembly. The inner catheter assembly consists of a thumb ring at the proximal end and a shaft to connect to the distal end which terminates in a brush. When the catheter is inserted into a working channel, the distal brush would be in a retracted position inside the outer sheath. When the catheter is in position, the brush can be extended into the tissue to obtain samples by advancing the proximal thumb ring.

    The Biopsy Forceps instrument is comprised of a coil pipe with a mechanical jaw assembly joined to the distal end. The jaws are opened by sliding the outer handle component distally relative to the inner, longer handle component. The jaws are closed by squeezing the two components together.

    AI/ML Overview

    The Monarch Endoscopy Platform (Monarch Platform) and its accessories were evaluated for safety and effectiveness through various performance tests, including biocompatibility, electrical safety, electromagnetic compatibility, software verification and validation, specific component performance testing, a navigational accuracy comparative study, animal and cadaver testing, and usability/human factors testing.

    Here's a breakdown of the acceptance criteria and the studies performed, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present acceptance criteria in a table format with specific quantitative targets for the Monarch Endoscopy Platform as a whole, or for each individual performance test. Instead, it states that the device "successfully tested for function, performance, and safety per FDA-recognized standards" and "complies with" or "conforms to" specific standards and design specifications. For the Navigational Accuracy Comparative Study, it mentions "substantially equivalent segmental, sub-segmental accuracy and on target accuracy, and thus equivalent Navigation Yield" compared to the predicate.

    Therefore, a direct table of "acceptance criteria" and "reported performance" like for a diagnostic accuracy study is not directly extractable for every aspect. However, we can synthesize the information for the Navigational Accuracy Comparative Study:

    Acceptance Criteria (Implied)Reported Device Performance
    Navigational Yield (Accuracy) substantially equivalent to predicate device (superDimension™ System)Substantially equivalent segmental, sub-segmental accuracy and on-target accuracy to the superDimension™ System, leading to equivalent Navigational Yield.
    Distance to Target accuracy substantially equivalent to predicate device (superDimension™ System)Substantially equivalent to the superDimension™ System.

    For all other performance testing categories (Biocompatibility, Electrical Safety, EMC, Software V&V, specific component testing, animal/cadaver, usability/human factors), the acceptance criterion was "compliance with recognized standards" and "meeting intended user requirements/design specifications," and the reported performance was that the device met these.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" and corresponding sample sizes for most of the performance evaluations in terms of human subjects or distinct data points in a clinical context. The studies involved:

    • Navigational Accuracy Comparative Study: The document doesn't explicitly state the sample size (number of cases or measurements) used in this comparative study.
      • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).
    • Animal and Cadaver Testing: Performed to evaluate the Monarch Platform under simulated use conditions.
      • Sample Size: Not specified (number of animals/cadavers).
      • Data Provenance: Prospective, simulated use.
    • Usability/Human Factors Testing:
      • Sample Size: Not specified (number of representative users).
      • Data Provenance: Prospective, simulated use.
    • Biocompatibility, Electrical Safety, EMC, Software V&V, and individual component performance testing: These are typically bench tests or engineering evaluations, not usually associated with a "test set" sample size in the clinical data sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document for any of the studies described. The studies appear to be primarily engineering, non-clinical (animal/cadaver), and simulated use evaluations rather than studies requiring expert adjudication of ground truth for diagnostic accuracy in patient data.

    4. Adjudication Method for the Test Set

    Since the document does not describe studies involving human medical experts establishing ground truth for a diagnostic test set, there is no mention of an adjudication method (e.g., 2+1, 3+1).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned in the provided text. The "Navigational Accuracy Comparative Study" compares the Monarch Platform's navigation metrics to the superDimension™ System, but it's not described as an MRMC study and therefore no effect size of human readers improving with AI vs. without AI assistance is provided. The Monarch Platform is described as a robotic endoscopy system that assists physicians, but the focus of the performance data section is on the device's inherent capabilities and comparison to existing technologies, not on augmented human performance in an MRMC setting.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The performance testing section primarily describes standalone performance aspects of the Monarch Platform and its accessories (e.g., biocompatibility, electrical safety, software V&V, mechanical performance of instruments, navigational accuracy of the system). The navigational accuracy study comparing Monarch and superDimension™ evaluates the system's navigation performance itself. The animal, cadaver, and usability studies involve human users but evaluate the system's safety and effectiveness with human interaction, not a standalone algorithm's diagnostic performance in the absence of human input.

    Therefore, while the individual component and software testing represent "standalone" evaluations of those specific parts, a standalone diagnostic algorithm performance study (without human-in-the-loop) for a clinical outcome is not explicitly detailed as described for typical AI/ML devices. The Monarch Platform is a surgical robotic system with navigation features, so its "standalone" performance relates to its ability to perform its function (e.g., navigate, manipulate instruments) rather than to interpret images or diagnose conditions like a standalone AI diagnostic software.

    7. The Type of Ground Truth Used

    For the specific tests mentioned:

    • Navigational Accuracy Comparative Study: The "ground truth" for navigation metrics (segmental, sub-segmental accuracy, on-target accuracy, distance to target) would likely be established through precise measurements against a known anatomical model or imaging data, but the document does not specify the exact method for this ground truth establishment.
    • Animal and Cadaver Testing: Ground truth established through observation of the system's performance and impact on the tissues/anatomy, likely assessed by skilled medical professionals involved in the study.
    • Usability/Human Factors Testing: Ground truth related to user needs, safety, and effectiveness is established through defined task completion rates, error rates, and user feedback against pre-defined success criteria.
    • Other performance tests (biocompatibility, electrical safety, etc.): Ground truth is established by adherence to recognized international standards and engineering specifications.

    8. The Sample Size for the Training Set

    The document does not mention a training set size. This is expected as the Monarch Platform is described as a robotic surgical system with navigation, not a machine learning or AI diagnostic device that typically requires a large training set of annotated data for model development. The software verification and validation are mentioned, but this refers to traditional software development and testing, not AI/ML model training.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned in the context of an AI/ML model, there is no information on how its ground truth was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152319

    Validate with FDA (Live)

    Date Cleared
    2016-05-26

    (283 days)

    Product Code
    Regulation Number
    874.4680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Auris Robotic Endoscopy System (ARES) is intended to provide bronchoscopic visualization of patient airways.

    Device Description

    The Auris Robotic Endoscopy System (ARES) is intended to be used by qualified physicians to provide visualization to the Bronchial Tree during Bronchoscopic procedures. The ARES consists of four major components, (1) the Patient Side System (PSS), (2) Controller Cart, (3) Surgeon Console (also known as the Master Device Workstation) and (4) the Bronchoscope and Accessories.

    The Patient Side System (PSS) includes the robot cart, two robot arms, both Endoscope and Sheath IDMs, IDS. servo drives box, Endoscope camera control box, power control box. illumination controller, and necessary cabling between the IDM/IDS and the robot cart.

    The Controller Cart houses the electronic systems required to power and operate the robotic systems. The Controller Cart is broken into two smaller carts, the system controller cart, and the arm controller cart.

    The Master Device Workstation is the workstation from which the surgeon drives the ARES. The console consists of a pendant that allows the surgeon to control various aspects of the system during a procedure.

    The system is based on a master - slave model, where the user (i.e. physician) is controlling the robots (slaves) using a pendant (master). The flexible bronchoscope is attached at the end effector of a robotic arm with multiple degrees of freedom. The flexible bronchoscope has a working channel and a camera at the tip. The Bronchoscope has an articulated tip that can bend in four directions. The working channel of the Bronchoscope is used for irrigation and aspiration.

    Each Slave includes a robotic arm with 6 degrees of freedom and an IDM (Instrument Drive Mechanism) with 4 actuated axes. The robotic arms are used to steer the Bronchoscope.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the Auris Robotic Endoscopy System (ARES). It details the device's indications for use, comparison to predicate devices, and testing completed. However, it does not contain specific acceptance criteria, reported device performance data, information about sample sizes for test sets, data provenance, number or qualifications of experts, adjudication methods, MRMC studies, standalone performance data, or details about the training set.

    The document focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and compliance with relevant medical device standards.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and study proving the device meets them with the level of detail requested, as that information is not present in the provided text.

    Here is what can be extracted or inferred from the text related to your request, with significant limitations:

    1. A table of acceptance criteria and the reported device performance:

    This information is not explicitly stated in the document. The document mentions "Preclinical testing included standard bench or in vitro testing confirming functionality and durability (e.g., tensile and other durability and functional evaluation)." and "Verification and validation testing was completed in compliance with the following standards: ... All clinical input requirements were validated." These statements imply that acceptance criteria were met, but the specific criteria and corresponding performance results are not provided.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not specified for bench or in vitro testing.
    • Data Provenance: The document states that a "Porcine animal model was used to validate system performance in vivo." This indicates some in vivo testing was performed in an animal model, but no details on the number of animals or the country of origin are provided. The broader preclinical testing would be considered prospective for the device's development.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided. Given the nature of the preclinical and animal model testing described, "ground truth" would likely be established by veterinary pathologists or engineers, but no specifics are given.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This device is not an AI/CAD-based diagnostic system. It is a robotic endoscopy system for visualization. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not be applicable to this type of device and is not mentioned or implied.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This device is a robotic system that is explicitly described as a "master - slave model, where the user (i.e. physician) is controlling the robots (slaves) using a pendant (master)." Therefore, it is inherently a "human-in-the-loop" device, and a standalone algorithm-only performance study would not be applicable. Its performance is tied directly to the physician's control.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the "in vivo" validation mentioned, the ground truth would likely be based on direct observation, procedural success (e.g., successful navigation, visualization), and potentially post-procedure pathology or findings in the porcine model. However, the document does not specify the exact nature of the ground truth. For the in vitro testing, ground truth would be based on engineering measurements and specifications.

    8. The sample size for the training set:

    The document mentions "Preclinical testing" and "Verification and validation testing." These types of studies are typically performed after development (which includes training/optimization) is complete. The document does not provide any information about a "training set" or how the device's control algorithms or visual processing might have been "trained." This is not an AI/machine learning device in the context of typical training sets.

    9. How the ground truth for the training set was established:

    As no training set is described, this information is not provided.

    In summary, the provided document details the regulatory submission (510(k)) for a robotic endoscopy system, focusing on its substantial equivalence to predicate devices and compliance with safety and performance standards. It broadly states that "clinical input requirements were validated" and "Preclinical testing included standard bench or in vitro testing confirming functionality and durability" and an "animal model was used to validate system performance in vivo." However, it does not provide the specific quantitative acceptance criteria or detailed study results for these tests that would address most of the points in your request.

    Ask a Question

    Ask a specific question about this device