510(k) Data Aggregation

    K Number
    K172033

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-08-31

    (57 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pirouette High Pressure (HP) PTA Catheter is intended for balloon dilation of the iliac, femoral, popliteal, renal arteries, Arterio-venous Fistula (AVF) and Arterio-venous Graft (AVG).

    Device Description

    The Pirouette High Pressure (HP) Percutaneous Transluminal Angioplasty (PTA) Catheter Family are standard over the wire (OTW), semi-compliant, coaxial design catheters with a balloon mounted on the distal tip. The distal portion of the catheter has a hydrophilic coating. The manifold connector and shaft design consists of a guidewire lumen allowing the catheter to track over a guidewire and an inflation lumen, used to inflate and deflate the balloon. Radiopaque markers are positioned on the shaft within the balloon to enable visualisation of the catheter/balloon under fluoroscopy. The catheter is compatible with 0.018-inch (0.46 mm) wire guides. The Pirouette High Pressure (HP) PTA Catheter Family includes multiple balloon sizes ranging from 4 to 6 mm in diameter and 20 to 80mm in length. The nominal balloon diameter (mm) and the balloon length (mm) are inscribed on the guidewire hub of the manifold. The effective lengths of the balloon range from 45cm to 75cm.

    AI/ML Overview

    The provided document describes the ArraVasc Pirouette High Pressure (HP) PTA Catheter (K172033) and its substantial equivalence to predicate devices, rather than a study proving the device meets acceptance criteria in the context of an AI/ML medical device. The document focuses on regulatory approval based on comparison to existing devices and non-clinical performance testing.

    Therefore, many of the requested categories for AI/ML device studies (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of device submission.

    However, I can extract information related to the acceptance criteria and performance as presented for this conventional medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes "performance specifications" and "pre-determined acceptance criteria" that the device met during bench testing. While specific numerical acceptance criteria values are not explicitly listed in a table format in the provided text, the document states that the device met these criteria. The table below synthesizes the performance parameters and the reported outcome.

    Acceptance Criteria CategoryReported Device Performance
    Dimensional verificationMet pre-determined acceptance criteria.
    Balloon preparationMet pre-determined acceptance criteria.
    PushabilityMet pre-determined acceptance criteria.
    TrackabilityMet pre-determined acceptance criteria.
    DeploymentMet pre-determined acceptance criteria.
    WithdrawalMet pre-determined acceptance criteria.
    Balloon reinsertionMet pre-determined acceptance criteria.
    Balloon Rated Burst Pressure (RBP)Met pre-determined acceptance criteria. The comparable RBP is 22 atm (4.0-5.0mm diameter) and 20 atm (6.0mm diameter).
    Balloon Fatigue (repeated inflations)Met pre-determined acceptance criteria.
    Balloon Compliance (at nominal and RBP)Met pre-determined acceptance criteria.
    Balloon Inflation/Deflation TimeMet pre-determined acceptance criteria.
    Bond Strength (Tip-to-balloon, balloon-to-proximal shaft, shaft-to-manifold)Met pre-determined acceptance criteria.
    Flexibility and Kink TestMet pre-determined acceptance criteria.
    Torque StrengthMet pre-determined acceptance criteria.
    RadiopacityMet pre-determined acceptance criteria.
    Coating IntegrityMet pre-determined acceptance criteria.
    Particulate evaluationMet pre-determined acceptance criteria.
    Guidewire compatibilityMet pre-determined acceptance criteria. Compatible with 0.018-inch wire guides.
    Introducer sheath compatibilityMet pre-determined acceptance criteria. Compatible with 4-5 Fr introducer sheath.
    BiocompatibilityDetermined to be biocompatible by leveraging data from the reference device (Pirouette 018).
    Sterilization validationMet pre-determined acceptance criteria regarding sterility maintenance and acceptable EtO/ECH residues.
    Shelf life testingMet pre-determined acceptance criteria.
    Packaging validationMet pre-determined acceptance criteria.

    2. Sample size used for the test set and the data provenance

    This is a conventional medical device (PTA catheter), not an AI/ML device. Therefore, there is no "test set" in the context of AI/ML validation with patient data. The "testing" refers to bench testing of the physical device. The sample sizes for each specific bench test (e.g., how many balloons were tested for RBP) are not provided in this summary. Data provenance is not applicable in the AI/ML sense; physical specimens of the device were tested in a laboratory setting. All testing was non-clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As this is a physical medical device, not an AI/ML diagnostic tool, there is no "ground truth" derived from expert consensus on medical images or clinical data. Device performance was assessed against engineering specifications and industry standards.

    4. Adjudication method for the test set

    Not applicable. There was no clinical test set requiring adjudication of medical findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical interventional device, not an AI/ML diagnostic or assistive tool, so MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device and does not involve an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is based on engineering specifications, industry standards (e.g., ISO 10555-1, ISO 10555-4), and pre-determined acceptance criteria derived from these standards and the device's design requirements. For biocompatibility, it was established through studies on a reference device following ISO 10993-1:2009.

    8. The sample size for the training set

    Not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. There is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162316

    Validate with FDA (Live)

    Device Name
    Pirouette 014
    Manufacturer
    Date Cleared
    2016-11-22

    (96 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pirouette 014 PTA Catheter is intended for balloon dilation of the illac, femoral, popliteal, and renal arteries.

    Device Description

    The Pirouette 014 Percutaneous Transluminal Angioplasty (PTA) Catheter Family are standard over the wire (OTW), semi-compliant, coaxial design catheters with a balloon mounted on the distal tip. The distal portion of the catheter has a hydrophilic coating.

    The manifold connector and shaft design consists of a guidewire lumen allowing the catheter to track over a guidewire and an inflation lumen, used to inflate and deflate the balloon. Radiopaque markers are positioned on the shaft within the balloon to enable visualization of the catheter/balloon under fluoroscopy. The catheter is compatible with .014 inch (0.356 mm) wire guides.

    The Pirouette 014 PTA Catheter Family includes multiple balloon sizes ranging from 1.25 mm to 4.0 mm in diameter and 15 mm to 300 mm in length. The nominal balloon diameter (mm) and the balloon length (mm) are inscribed on the guidewire hub of the manifold. The effective lengths of the catheter are 90 cm, 130 cm and 150 cm.

    AI/ML Overview

    This document describes the Pirouette 014 PTA Catheter, a medical device for balloon dilation of arteries. However, the provided text does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML powered medical device. Instead, it describes a conventional medical device (a catheter) and its non-clinical performance testing.

    Therefore, I cannot fulfill the request as it pertains to an AI/ML powered device, an acceptance criteria table, or a study demonstrating AI performance.

    The information given is about:

    • Device Name: Pirouette 014
    • Device Type: Percutaneous Transluminal Angioplasty (PTA) Catheter
    • Intended Use: Balloon dilation of the iliac, femoral, popliteal, infra-popliteal, and renal arteries.
    • Predicate Device: Pirouette 018 OTW PTA Catheter (K151153) and Bantam α PTA Catheter (K093139)
    • Performance Data: Non-clinical bench tests, biocompatibility tests, sterilization, shelf-life, and packaging validation were conducted. No clinical studies were performed. The results from these tests "showed that the Pirouette 014 met the pre-determined acceptance criteria," but the specific acceptance criteria or the reported performance values are not detailed in a quantitative table.

    To answer your prompt for an AI/ML powered device, I would need a different type of document specifically detailing the performance of an AI/ML algorithm.