Search Filters

Search Results

Found 39 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gammex PI Hybrid Micro Surgical Gloves are sterile, non-pyrogenic, are intended to be worn by operating room personnel to protect a surgical wound from contamination and have a shelf life of 3 years. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Tested chemotherapy drugs are as follows: Test chemotherapy drug and concentration Minimum breakthrough detection time in minutes Bleomycin (15.0 mg/ml) >240 Busulfan (6.0 mg/ml) >240 Carboplatin (10.0 mg/ml) >240 Carmustine (3.3 mg/ml) 24.1 Cisplatin (1.0 mg/ml) >240 Cyclophosphamide (20.0 mg/ml) >240 Cytarabine HCl (100.0 mg/ml) >240 Dacarbazine (10.0 mg/ml) >240 Daunorubicin HCl (5.0 mg/ml) >240 Docetaxel (10.0 mg/ml) >240 Doxorubicin HCl (2.0 mg/ml) >240 Epirubicin (2.0 mg/ml) >240 Etoposide (20.0 mg/ml) >240 Fludarabine (25.0 mg/ml) >240 Fluorouracil (50.0 mg/ml) >240 Gemcitabine (38.0 mg/ml) >240 Idarubicin (1.0 mg/ml) >240 Ifosfamide (50.0 mg/ml) >240 Irinotecan (20.0 mg/ml) >240 Mechlorethamine HCI (1.0 mg/ml) >240 Melphalan (5.0mg/ml) >240 Methotrexate (25.0 mg/ml) >240 Mitomycin C (0.5 mg/ml) >240 Mitoxantrone (2.0mg/ml) >240 Oxaliplatin (2.0 mg/ml) >240 Paclitaxel (6.0 mg/ml) >240 Rituximab (10.0 mg/ml) >240 ThioTEPA (10.0 mg/ml) 26.0 Vincristine Sulfate (1.0 mg/ml) >240 Warning: Do not use with Carmustine or ThioTEPA

    Device Description

    Gammex PI Hybrid Micro Surgical Gloves are sterile and disposable devices. The gloves are made of a synthetic rubber polyisoprene and neoprene blend. Ansell has performed independent laboratory testing to support use with Chemotherapy Drugs and Non-Pyrogenic claims. The glove has a shelf life of 3 years, is brown in color and comes in the following sizes: 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0. Each size is assigned with a model number.

    AI/ML Overview

    The document provides information on the non-clinical testing conducted for the GAMMEX PI Hybrid Micro surgical gloves. Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Dimensions (ASTM D3767-03)In accordance with ASTM D3577-19: Standard Specification for Rubber Surgical Gloves.PASS
    Physical Properties (ASTM D3577-19)Acceptance criteria for tensile strength, ultimate elongation and stress at 500% elongation before and after accelerated aging for synthetic surgical gloves per ASTM D3577-19.PASS
    Freedom from holes (ASTM D5151-19)In accordance with ASTM D3577-19: Standard Specification for Rubber Surgical Gloves with AQL requirements of 1.5.PASS
    Powder-Free (ASTM D6124)Meets applicable acceptance criteria for powder free ≤ 2mg per glove per ASTM D3577-19.PASS
    Sterility (ANSI/AAMI/ISO 11137-1:2006)Meets acceptance criteria requirement of 10-6 SAL per ISO 11137-1: Sterilization for Health Care Products - Radiation - Part 1: Requirements for Development, Validation and Routine Control of A Sterilization Process For Medical Devices.PASS
    Chemotherapy Drug Permeation Test (ASTM D6978-05(2019))Acceptance Criteria in accordance with ASTM D6978-05(2019): Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. (Specific breakthrough times for each drug are listed in the "Indications for Use" section).PASS (with specific breakthrough times. Noted that Carmustine (24.1 min) and Thiotepa (26.0 min) have extremely low permeation times, leading to a warning for these drugs.)
    ISO in vitro Cytotoxicity (ISO 10993-5)Acceptance criteria in accordance with ISO 10993-5: Biological Evaluation of Medical Devices, Part 5: Tests for In Vitro Cytotoxicity, 2009.FAIL (Surfaces at undiluted were found to be cytotoxic (grade 4) to L-929 cells and non-cytotoxic with grade 0 at 1:2, 1:4, 1:8, 1:16, 1:32 and 1;64 dilutions.)
    ISO Skin Irritation Study (ISO 10993-10:2010)Passes Primary Skin Irritation test per ISO 10993-10, Biological Evaluation Of Medical Devices, Part 10: Test for Irritation and Skin Sensitization. Under the conditions of the study, not an irritant.PASS
    ISO Maximization Sensitization Study (ISO 10993-10:2010)Passes Dermal Sensitization test per ISO 10993-10, Biological Evaluation of medical devices, Part 10: Test for Irritation and Skin Sensitization. Under the conditions of the study, not a sensitizer.PASS
    ISO Acute Systemic Toxicity Study (ISO 10993-11:2017)Passes Acute Systemic Toxicity Test per ISO 10993-11, Biological Evaluation of medical devices, Part 11: Test for Systemic Toxicity. Under the conditions of the study, there was no mortality or evidence of acute systemic toxicity.PASS
    Material Mediated Pyrogenicity Study (USP <151>)Passes Material Mediated Pyrogenicity Test per USP <151>. Under the conditions of the study, non-pyrogenic.PASS
    Endotoxin StudyLess than 20.0 EU/device.PASS

    Note regarding Cytotoxicity: Although the in vitro cytotoxicity test showed a "FAIL" at undiluted concentrations, the document states that the extracts were then evaluated by ISO 10993-11 for systemic toxicity, which passed, indicating no systemic response in the animal model. This suggests a different acceptance pathway for cytotoxicity, likely due to further evaluation in an in vivo model.

    2. Sample size used for the test set and the data provenance:

    • The document does not explicitly state the specific sample sizes for each non-clinical test.
    • The data provenance is from non-clinical tests performed by Ansell Healthcare at independent laboratories. There is no information regarding country of origin, but it is implied to be laboratory-based testing rather than human subject data. The studies are by nature prospective as they are conducted for the purpose of demonstrating safety and performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable. These are non-clinical tests based on standardized laboratory methods (e.g., ASTM, ISO, USP), not interpretation by human experts. The "ground truth" is defined by the objective measurement criteria of these standards.

    4. Adjudication method for the test set:

    • This information is not applicable, as the tests are objective measurements according to standardized protocols, not subjective assessments requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This information is not applicable. The device is a surgical glove, not an AI-assisted diagnostic or therapeutic device that would involve human readers or AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This information is not applicable. The device is a physical product (surgical glove), not an algorithm or software.

    7. The type of ground truth used:

    • The ground truth used for these tests is based on established international and national standards and methodologies (e.g., ASTM, ISO, USP) for evaluating the physical, chemical, and biological properties of medical devices like surgical gloves. These standards define the objective criteria and test methods to determine acceptable performance.

    8. The sample size for the training set:

    • This information is not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    • This information is not applicable, as it's not a machine learning driven device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230079

    Validate with FDA (Live)

    Date Cleared
    2023-08-23

    (225 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    0 - 5
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These surgical gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with chemotherapy drugs, have low dermatitis potential, and are nonpyrogenic. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential are sterile and disposable devices. Gloves are made of synthetic polyisoprene rubber, are white in color, and are available in sizes 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, and 9.0. A polyurethane polymer coating is applied to the inner surface of the glove to make donning easy.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device: "Polyisoprene Surgical Gloves Tested for Use with Chemotherapy Drugs and Low Dermatitis Potential" (K230079). It compares the new device to a predicate device (K190018) and provides evidence of performance through various tests.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document presents a comparison of the subject device (K230079) against a predicate device (K190018) and also summarizes non-clinical testing. Below is a compilation of the acceptance criteria (standards/guidance/test criteria) and the reported device performance.

    Test/CharacteristicAcceptance Criteria (Standard/Guidance/Criteria)Reported Device Performance (K230079)
    Chemotherapy PermeationASTM D6978-05 (2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy DrugsAll drugs with BDT >240 minutes met criteria.Carmustine (BCNU) (3.3 mg/ml): 12.6 minutesThiotepa (10.0 mg/ml): 26.6 minutes (Note: For marketing, the device labels a warning against using with Carmustine and Thiotepa due to "extremely low permeation times".)
    Dimensions (Length)ASTM D3577 requirements: Minimum 265mmPass
    Dimensions (Palm Width)ASTM D3577 requirements (specific to each size 5.5-9.0, e.g., 5.5: 70±6mm, 9.0: 114±6mm)Pass for all sizes
    Dimensions (Thickness - Finger)ASTM D3577 requirements: Minimum 0.10mmPass
    Dimensions (Thickness - Palm)ASTM D3577 requirements: Minimum 0.10mmPass
    Dimensions (Thickness - Cuff)ASTM D3577 requirements: Minimum 0.10mmPass
    Physical PropertiesASTM D3577-19 requirements for tensile strength, ultimate elongation, and stress at 500% elongation before and after accelerated aging for synthetic surgical glovesPass
    Freedom from holesASTM D3577-19 and ASTM D5151-06 requirements; AQL 1.5Pass
    Powder-FreeASTM D3577-19, ASTM D6124-06 requirements: ≤ 2 mg per glovePass
    SterilityANSI/AAMI/ISO 11137-1:2018 requirement of 10^-6 SALPass
    ISO In Vitro CytotoxicityISO 10993-5:2009: Non-cytotoxic (as the general acceptance criteria)Under the conditions of the study, the device was found to be cytotoxic. However, "the device extracts did not elicit acute systemic response in the animal model" in the follow-up Acute Systemic Toxicity test (ISO 10993-11). This is a notable finding, indicating the device failed the initial cytotoxicity but was deemed safe due to subsequent testing.
    ISO Skin Irritation StudyISO 10993-10:2010: Not an irritantPass
    ISO Maximization Sensitization StudyISO 10993-10:2010: Not a sensitizerPass
    ISO Acute Systemic Toxicity StudyISO 10993-11:2006: No mortality or evidence of acute systemic toxicityPass (addressed after initial cytotoxicity finding)
    Human Skin SensitizationModified Draize-95 Test (per FDA guidance): Not a sensitizer (demonstrates low dermatitis potential)Pass: "The results showed low dermatitis potential for the human subject tested."
    Endotoxin StudyLess than 20.0 EU/devicePass: "test results indicate low endotoxin level"
    Material Mediated Pyrogenicity StudyISO 10993-11:2017: Meets the material mediated pyrogenicity requirementPass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Chemotherapy Drug Permeation (ASTM D6978-05): The document specifies the standard used but does not provide the exact sample size for the test sets (number of gloves or replicates per drug). The data is generated from laboratory testing ("under the conditions of the study"). Data provenance is not explicitly stated but implies laboratory testing for regulatory submission.
    • Dimensions, Physical Properties, Freedom from Holes, Powder-Free, Sterility: These tests refer to ASTM and ISO standards (e.g., ASTM D3577, ASTM D5151, ASTM D6124, ISO 11137-1). These standards typically define the sampling plans (e.g., AQL 1.5 for freedom from holes), but the specific number of units tested for this submission is not directly stated. Data provenance implies manufacturer's testing for regulatory submission.
    • Biocompatibility (ISO 10993 series): These are biological evaluations. The document refers to "the conditions of the study" and "animal model" for acute systemic toxicity but does not provide specific sample sizes (e.g., number of animals used). Data provenance implies laboratory testing for regulatory submission.
    • Human Skin Sensitization (Modified Draize-95 Test): This test was performed on "human subject tested." The sample size (number of human subjects) is not explicitly given, it ambiguously states "the human subject tested." Data provenance is from a clinical test as per FDA guidance.
    • Endotoxin Study & Material Mediated Pyrogenicity Study: The document refers to "the conditions of the study" but does not provide sample sizes. Data provenance implies laboratory testing for regulatory submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This submission is for a physical device (surgical gloves) and its material properties, not an AI/algorithm-driven device requiring expert image interpretation or similar ground truth establishment. Therefore, the concept of "experts" establishing a ground truth in that context is not applicable here. The "ground truth" for the performance tests (e.g., breakthrough time, tensile strength) is established by the validated methodologies of the referenced ASTM and ISO standards.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of material and physical property testing of a device like surgical gloves. Test results are objective measurements based on defined standard protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This is not an AI device or a diagnostic imaging device where human reader performance is evaluated.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

    Not applicable as this is not an AI/algorithm device. The performance tests evaluate the physical gloves without any "human-in-the-loop" interaction in the testing methodology itself beyond conducting the tests.

    7. The Type of Ground Truth Used:

    • Objective Measurement Standards: For most physical and chemical properties (dimensions, strength, freedom from holes, powder, sterility, chemotherapy permeation, endotoxin, pyrogenicity), the "ground truth" is defined by the objective measurement criteria and methodologies specified in the referenced ASTM and ISO standards. The results are quantitative measurements against these predefined thresholds.
    • Biological Response in Models: For biocompatibility tests (cytotoxicity, skin irritation, sensitization, acute systemic toxicity), the "ground truth" is established by observing biological responses in in vitro or in vivo (animal) models as per ISO 10993 standards and the Modified Draize-95 Test for human skin sensitization.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/machine learning device that requires a training set. The device is a manufactured physical product.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213289

    Validate with FDA (Live)

    Date Cleared
    2023-05-26

    (602 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free surgeons' glove is intended to be worn by operating room personnel to protect a surgical wound from contamination. The glove system was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy testing was carried out on the system as a whole, including both the under-glove together. Please note that the following drugs have extremely low permeation times: 3.0 minutes and Thiotepa: 67.0 minutes. Warning: Do not use with Carmustine or Thiotepa.

    Device Description

    Gammex® Pl Plus Glove-in-Glove™ System tested for use with Chemotherapy Drugs is a sterile and disposable device. This glove system ismade of synthetic polyisoprene rubber. Gammex® Pl Plus Glove- in-Glove™ System Tested for Use with Chemotherapy Drugs is comprised of a single-use, sterile, white outer glove, which is a disposable, powder-free surgical glove made from synthetic polyisoprene and a single-use, sterile, green underglove which is a disposable, powder-free surgical glove made from syntheticpolyisoprene. The Gammex® Pl PlusGlove-in-Glove™ system Testedfor Use with Chemotherapy Drugs is comprised of the underglove being mechanically inserted into the outer glove prior to packing and sterilization. There is no adhesive present between the two gloves. This results in quick double gloving with only one donning event.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding the "Gammex® PI Plus Glove-in-Glove™ System Tested For Use with Chemotherapy Drugs." This document outlines the device's characteristics, its intended use, and the studies conducted to demonstrate its safety and effectiveness.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document includes a table of non-clinical tests performed on the Gammex® PI Plus Glove-in-Glove™ System.

    Test TitlePurpose of TestAcceptance CriteriaReported Device Performance
    ASTM D3767-03DimensionsIn accordance with ASTM D3577-19: Standard Specification for Rubber Surgical Gloves.PASS
    ASTM D3577-19Physical PropertiesFor tensile strength, ultimate elongation and stress at 500% elongation before and after accelerated aging for synthetic surgical gloves per ASTM D3577-19 : Standard Specification for Rubber Surgical Gloves.PASS
    ASTM D5151-19Freedom from holesIn accordance with ASTM D3577-19: Standard Specification for Rubber Surgical Gloves with AQL requirements of 1.5.PASS
    ASTM D6124Powder-FreeApplicable acceptance criteria for powder free ≤ 2mg per glove per ASTM D3577-19: Standard Specification for Rubber Surgical Gloves.PASS
    ANSI/AAMI/ISO 11137-1:2006SterilityRequirement of 10-6 SAL per ISO 11137-1: Sterilization for health care products – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.PASS
    ASTM D6978-05(2019)Chemotherapy Drug Permeation TestIn accordance with ASTM D6978-05(2019): Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Note: Carmustine (35.0 min) and Thiotepa (67.0 min) have extremely low permeation times, implying these are specific exceptions within the criteria or require special consideration/warnings.PASS
    ISO 10993-5 (Outer and Underglove)ISO in vitro Cytotoxicity Study (Undiluted, 1:2, 1:4, 1:8, 1:16, 1:32, 1:64 dilutions)Acceptance criteria in accordance with ISO 10993-5: Biological Evaluation of Medical Devices, Part 5: Tests for In Vitro Cytotoxicity, 2009. (Specifically, for the outer glove: non-cytotoxic at 1:4, 1:8, 1:16, 1:32, and 1:64 dilutions. For the underglove: non-cytotoxic at 1:16, 1:32, and 1:64 dilutions). Intermediate dilutions for both gloves showed varying levels of cytotoxicity.FAIL (at certain dilutions for both gloves, e.g., undiluted and 1:2 for outer; undiluted, 1:2, 1:4, 1:8 for under)
    ISO 10993-10:2010ISO Skin Irritation StudyPasses Primary Skin Irritation test per ISO 10993-10, Biological Evaluation of medical devices, Part 10: Test for irritation and skin sensitization. Not an irritant under study conditions.PASS
    ISO 10993-10:2010ISO Maximization Sensitization StudyPasses Dermal Sensitization test per ISO 10993-10, Biological Evaluation of medical devices, Part 10: Test for irritation and skin sensitization. Not a sensitizer under study conditions.PASS
    ISO 10993-11:2017ISO Acute Systemic Toxicity StudyPasses Acute Systemic Toxicity Test per ISO 10993-11, Biological Evaluation of medical devices, Part 11: Test for systemic toxicity. No mortality or evidence of acute systemic toxicity under study conditions.PASS

    Chemotherapy Drug Permeation Times (Detailed)

    Test Chemotherapy Drug & ConcentrationAverage Breakthrough Detection Time (Minutes)
    Bleomycin - 15.0 mg/ml>240
    Busulfan - 6.0 mg/ml>240
    Carboplatin - 10 mg/ml>240
    Carmustine - 3.3 mg/ml35.0
    Cisplatin - 1.0 mg/ml>240
    Cyclophosphamide - 20.0 mg/ml>240
    Cytarabine HCl - 100.0 mg/ml>240
    Dacarbazine - 10.0 mg/ml>240
    Daunorubicin HCl - 5.0 mg/ml>240
    Docetaxel - 10.0 mg/ml>240
    Doxorubicin HCl 2.0 mg/ml>240
    Epirubicin - 2.0 mg/ml>240
    Etoposide - 20.0 mg/ml>240
    Fludarabine - 25 mg/ml>240
    Fluorouracil - 50.0 mg/ml>240
    Gemcitabine - 38.0 mg/ml>240
    Idarubicin - 1.0 mg/ml>240
    Ifosfamide - 50.0 mg/ml>240
    Irinotecan - 20.0 mg/ml>240
    Mechlorethamine HCl - 1.0 mg/ml>240
    Melphalan - 5.0 mg/ml>240
    Methotrexate - 25.0 mg/ml>240
    Mitomycin C - 0.5 mg/ml>240
    Mitoxantrone - 2.0 mg/ml>240
    Oxaliplatin - 2.0 mg/ml>240
    Paclitaxel - 6.0 mg/ml>240
    Rituximab - 10.0 mg/ml>240
    Thiotepa - 10.0 mg/ml67.0
    Vincristine Sulfate - 1.0 mg/ml>240

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify sample sizes for individual tests like dimensions, physical properties, freedom from holes, powder-free, sterility, or biocompatibility studies. The general phrasing "Meets ASTM D..." or "Passes..." suggests that the sample sizes typically required by these standards were used.

    For the Chemotherapy Drug Permeation Test (ASTM D6978-05), while the results for each drug are provided (Average Breakthrough Detection Time), the document does not specify the number of samples (e.g., gloves or glove specimens) tested for each drug.

    The data provenance is prospective in the sense that these are tests conducted specifically for this 510(k) submission to demonstrate the device's performance against established standards. There is no mention of country of origin for the data; it is assumed to be from a testing facility compliant with the mentioned standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The "ground truth" for these tests is defined by the objective measurement methods and acceptance criteria specified in the referenced ASTM, ANSI, and ISO standards. These standards themselves are developed by expert consensus, but specific independent expert adjudication for this particular test set is not mentioned or typically required for this type of device and testing.

    4. Adjudication Method for the Test Set:

    An explicit adjudication method (like 2+1 or 3+1) is not applicable or mentioned for the non-clinical tests described. The tests are objective measurements against predefined criteria outlined in the consensus standards. The "Pass/Fail" results are determined directly by comparing the test outcomes to these established thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, a MRMC comparative effectiveness study was not done. The document explicitly states: "A clinical study was not conducted on the subject device." This type of study focuses on human reader performance, often with or without AI assistance, which is not relevant to the physical and chemical resistance properties of a surgical glove.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done:

    No, a standalone algorithm-only performance study was not done. This device is a physical medical device (surgical glove), not an algorithm or AI software. Therefore, studies evaluating algorithm performance are not relevant.

    7. The Type of Ground Truth Used:

    The ground truth used for the non-clinical tests is based on consensus standards and laboratory measurements.

    • For physical properties, dimensions, freedom from holes, powder residue, and sterility: The ground truth is objective measurement against the specified ASTM and ISO standards' criteria.
    • For chemotherapy drug permeation: The ground truth is the measured breakthrough time determined by the ASTM D6978-05 standard.
    • For biocompatibility (skin irritation, sensitization, acute systemic toxicity, cytotoxicity): The ground truth is the biological response observed in validated in vitro and in vivo models according to the ISO 10993 series of standards.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical product (surgical glove) and does not involve AI or machine learning algorithms that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212148

    Validate with FDA (Live)

    Date Cleared
    2022-03-10

    (244 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.

    Device Description

    Microflex 73-743 Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs is a non-sterile, single use only, disposable, powder free examination gloves. Characteristics: - . Ambidextrous with beaded cuff and straight fingers - Finger-textured . - . Blue colored - Five (5) sizes – extra-small, small, medium, large, and extra-large - . Tested against chemotherapy drugs High levels of ozone will degrade rubber material of the glove; therefore, the glove should be protected from ozone in particular. The glove is designed to meet the specifications of ASTM D6977-19, Standard Specification for Polychloroprene Examination Gloves for Medical Application.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device: "Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs".

    The information provided describes the product's acceptance criteria and the non-clinical testing performed to demonstrate that the device meets some of these criteria. However, it does not describe an AI/ML-driven medical device, nor does it detail a study proving that an AI/ML device meets acceptance criteria. Therefore, I am unable to answer the questions related to AI/ML device performance, ground truth, expert opinions, or MRMC studies.

    Instead, I will extract the acceptance criteria and study information related to the glove's performance as described in the document.


    Acceptance Criteria and Study for the Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs

    This submission is for a physical medical device (examination glove), not an AI/ML system. Therefore, the "acceptance criteria" relate to the physical and chemical performance characteristics of the glove. The "study" refers to non-clinical laboratory testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is assessed against recognized standards, primarily ASTM D6977-19, ASTM D5151, ASTM D6124, and ASTM D6978-05 for chemotherapy drug permeation, as well as ISO 10993 series for biocompatibility.

    Test Methodology / CharacteristicAcceptance Criteria (Standard)Reported Device Performance (Results)
    Physical Characteristics:
    Dimensions (ASTM D6977-19):Meets ASTM D6977-19 requirements for length, width and thickness
    - LengthMinimum 230mmMinimum 240mm
    - Palm Width (XS)$70 \pm 10$ mm$75 \pm 5$ mm
    - Palm Width (S)$80 \pm 10$ mm$85 \pm 5$ mm
    - Palm Width (M)$95 \pm 10$ mm$95 \pm 5$ mm
    - Palm Width (L)$110 \pm 10$ mm$105 \pm 5$ mm
    - Palm Width (XL)$120 \pm 10$ mm$115 \pm 5$ mm
    - Thickness (Finger)Minimum 0.05 mmFinger - min 0.09 mm
    - Thickness (Palm)Minimum 0.05 mmPalm - min 0.06 mm
    - Thickness (Cuff)- (Not specified in standard for comparison)Cuff - min 0.05 mm
    Physical Properties (ASTM D6977-19):Meets ASTM D6977-19 requirements for tensile strength and ultimate elongation before and after accelerated aging.Meets ASTM D6977-19 requirements for tensile strength and ultimate elongation before and after accelerated aging.
    - Tensile Strength (Before Aging)Minimum 14 MPaMinimum 16 MPa
    - Tensile Strength (After Aging)Minimum 14 MPaMinimum 14 MPa
    - Ultimate Elongation (Before Aging)Minimum 500%Minimum 500%
    - Ultimate Elongation (After Aging)Minimum 400%Minimum 400%
    Freedom from Holes (ASTM D6977-19, ASTM D5151-06):AQL 2.5Meets or exceeds ASTM D6977-19 and ASTM D5151-06 requirements of AQL 2.5
    Powder Residual (ASTM D6977-19, ASTM D6124-06):Not more than 2.0 mg/gloveMeets applicable requirement for powder free; ≤ 2 mg per glove
    Biocompatibility:
    - ISO Skin Irritation StudyISO10993-10:2010 (Not an irritant)Under the conditions of the study, not an irritant
    - ISO Maximization Sensitization StudyISO 10993-10:2010 (Not a sensitizer)Under the conditions of the study, not a sensitizer
    - ISO Acute Systemic ToxicityISO 10993-11: 2006 (No evidence of systemic toxicity)Under the conditions of the study, no evidence of systemic toxicity
    - ISO In Vitro CytotoxicityISO 10993-5:2009 (Non-cytotoxic at specified dilutions)Under the conditions of the study, undiluted, 1:2, and 1:4 dilutions were cytotoxic (grade 4). Dilutions of 1:8 (grade 2), 1:16 (grade 1), 1:32 and 1:64 (grade 0) were non-cytotoxic. (This indicates cytotoxicity at high concentrations, which is a common finding for glove materials, and the acceptance is based on non-cytotoxicity at lower, clinically relevant dilutions for biocompatibility).
    Chemotherapy Permeation (ASTM D6978-05):Times as per standard/previous testing (predicate device, if applicable)
    - Carmustine - 3.3 mg/ml- (Comparison to predicate: 47.9 minutes)45.4 minutes
    - Cisplatin - 1.0 mg/ml- (Not explicitly in predicate comparison table)>240 minutes
    - Cyclophosphamide - 20.0 mg/ml>240 minutes>240 minutes
    - Dacarbazine - 10.0 mg/ml- (Not explicitly in predicate comparison table)>240 minutes
    - Doxorubicin HCl - 2.0 mg/ml>240 minutes>240 minutes
    - Etoposide - 20.0 mg/ml>240 minutes>240 minutes
    - Fluorouracil - 50.0 mg/ml>240 minutes>240 minutes
    - Methotrexate - 25.0 mg/ml>240 minutes>240 minutes
    - Paclitaxel - 6.0 mg/ml>240 minutes>240 minutes
    - Thiotepa - 10.0 mg/ml>240 minutes (from predicate)23.5 minutes (Note: This is significantly lower than the predicate for the same drug, leading to a warning for its use with this glove.)
    - Vincristine Sulfate - 1.0 mg/ml>240 minutes>240 minutes

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify exact sample sizes for each test beyond indicating the tests were conducted. However, standards like ASTM and ISO have prescribed sample sizes and methodologies. For instance, for chemotherapy drug permeation testing as per ASTM D6978-05, a minimum of three specimens are tested per drug (indicated by "Average Minimum Breakthrough Detection Time (Sample 1,2,3)").

    The data provenance is not explicitly stated as retrospective or prospective or by country of origin, but it originates from non-clinical laboratory testing performed to meet recognized international standards. This type of testing is typically prospective, conducted under controlled laboratory conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This is not applicable as this is a non-clinical device that relies on standardized laboratory testing rather than human expert interpretation of data.

    4. Adjudication Method for the Test Set

    Not applicable. The testing is based on objective measurements and adherence to specified standard protocols, not expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used

    The "ground truth" for the performance claims of this medical glove consists of measurements obtained according to established and recognized international laboratory standards (e.g., ASTM, ISO). For example:

    • Physical dimensions and properties are measured directly.
    • Freedom from holes is determined by a water leak test.
    • Powder residual is measured gravimetrically.
    • Biocompatibility is assessed through in vitro and in vivo biological tests with specific endpoints (e.g., cell viability for cytotoxicity, skin reactions for irritation/sensitization).
    • Chemotherapy drug permeation is measured using highly sensitive analytical methods to detect breakthrough time.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device; there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is not an AI/ML device; there is no "training set."

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drug has an extremely low permeation time: Carmustine: 47.9 minutes (Blue), 49.6 minutes (Green), 27.6 minutes (Black). Warning: Do not use with Carmustine.

    Device Description

    Microflex® Nitrile Patient Examination Gloves Blue Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate, Microflex® Nitrile Patient Examination Gloves Green Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate, Microflex® Nitrile Patient Examination Gloves Black Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable, powder free examination gloves are made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the gloves to make donning easy.

    Characteristics:

    • . Ambidextrous with beaded cuff and straight fingers
    • . Finger-textured
    • Blue, green, or black colored .
    • . Four (4) sizes - small, medium, large, and extra-large
    • . Tested against chemotherapy drugs and fentanyl citrate

    High levels of ozone will degrade rubber material of the gloves; therefore, the gloves should be protected from ozone in particular.

    The gloves are designed to meet the specifications of ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    This is an analysis of a 510(k) Premarket Notification for medical gloves, which are considered Class I devices. The submission focuses on demonstrating substantial equivalence to a predicate device, primarily through non-clinical performance testing against established standards rather than clinical studies involving human patients or complex AI algorithms.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established by referring to ASTM and ISO standards for examination gloves. The reported device performance indicates that the gloves meet or exceed these standards.

    Microflex Nitrile Patient Examination Gloves (Blue, Green, Black)

    CharacteristicsStandard/Test/FDA GuidanceAcceptance Criteria (from Standard, implied)Reported Device Performance
    Physical Characteristics:
    Dimensions:ASTM D6319-10Meets ASTM D6319-10 requirementsMeets ASTM D6319-10 requirements for length, width, and thickness
    LengthMinimum 230mmMinimum 240mm (Exceeds)
    Palm width (S)80 ± 10 mm85 ± 5 mm (Meets)
    Palm width (M)95 ± 10 mm95 ± 5 mm (Meets)
    Palm width (L)110 ± 10 mm105 ± 5 mm (Meets)
    Palm width (XL)120 ± 10 mm115 ± 5 mm (Meets)
    Thickness (single-wall):
    Fingerminimum 0.05 mmFinger - min 0.09 mm (Exceeds)
    Palmminimum 0.05 mmPalm - min 0.06 mm (Exceeds)
    CuffNot specified (implied to meet standard)Cuff - min 0.05 mm
    Physical Properties:ASTM D6319-10Meets ASTM D6319-10 requirementsMeets ASTM D6319-10 requirements for tensile strength and ultimate elongation before and after accelerated aging
    Tensile Strength (Before Aging)minimum 14 MPaminimum 16 MPa (Exceeds)
    Tensile Strength (After Aging)minimum 14 MPaminimum 14 MPa (Meets)
    Ultimate Elongation (Before Aging)minimum 500%minimum 500% (Meets)
    Ultimate Elongation (After Aging)minimum 400%minimum 400% (Meets)
    Freedom from holes:ASTM D6319-10, ASTM D5151-06AQL 2.5Meets or exceeds ASTM D6319-10 and ASTM D5151-06 requirements of AQL 2.5
    Powder Residual:ASTM D6319-10, ASTM D6124-06≤ 2 mg per gloveMeets applicable requirement for powder free; ≤ 2 mg per glove
    Biocompatibility:
    ISO In Vitro CytotoxicityISO 10993-5:2009Non-cytotoxicUndiluted, 1:2, 1:4, 1:8, 1:16 dilution was cytotoxic. 1:32 and 1:64 are not cytotoxic. (Requires clarification on acceptable dilution for "non-cytotoxic")
    ISO Skin Irritation StudyISO10993-10:2010Not an irritantUnder the conditions of the study, not an irritant
    ISO Maximization Sensitization StudyISO 10993-10:2010Not a sensitizerUnder the conditions of the study, not a sensitizer
    ISO Acute Systemic ToxicityISO 10993-11:2006No evidence of systemic toxicityUnder the conditions of the study, no evidence of systemic toxicity
    Chemotherapy Drug Permeation (ASTM D6978-05):ASTM D6978-05No permeation within a specified time*Varies by drug and glove color. Carmustine has low permeation (27.6 to 49.6 minutes), others generally >240 minutes.

    Note: For chemotherapy drug permeation, the acceptance criterion isn't explicitly stated as a single value (e.g., ">X minutes"), but the comparison is made against the predicate device and the standard, indicating the longer the breakthrough time, the better the performance. The warning for Carmustine suggests that its specific permeation time is deemed too low for safe use with these gloves.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the total sample size for each specific test. However, for the Chemotherapy Drug Permeation testing, it indicates "Average Minimum Breakthrough Detection Time (Sample 1,2,3) (Minutes)", implying that at least 3 samples were tested for each drug/glove combination.

    The provenance of the data is not specified in terms of country of origin or whether it was retrospective or prospective. It is generally understood that such testing is conducted in a laboratory setting for regulatory submission.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. This submission pertains to physical and chemical performance testing of medical gloves, not a diagnostic device that requires expert interpretation of images or patient data to establish ground truth. The "ground truth" for these tests is defined by the objective measurements according to the specified ASTM and ISO standards (e.g., direct measurement of length, tensile strength testing, chemical permeation detection).

    4. Adjudication Method for the Test Set

    Not applicable. As there are no human interpretations or judgments required for establishing the "ground truth" (instead, it's objective measurement against standards), no adjudication method (like 2+1 or 3+1) is mentioned or relevant.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No MRMC study was done. This is a Class I medical device (gloves) and not an AI-assisted diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not applicable.

    6. Standalone Performance Study (Algorithm Only)

    Not applicable. This device does not involve an algorithm. The performance studies are laboratory tests of the physical, chemical, and biological properties of the gloves.

    7. Type of Ground Truth Used

    The ground truth for the performance claims is based on objective measurements obtained through standardized laboratory tests according to recognized national and international standards, specifically:

    • ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical Application (for dimensions and physical properties like tensile strength, elongation, freedom from holes).
    • ASTM D5151-06: Standard Test Method for Detection of Holes in Medical Gloves.
    • ASTM D6124-06: Standard Test Method for Residual Powder on Medical Gloves.
    • ASTM D6978-05: Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs.
    • ISO 10993-5:2009: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
    • ISO 10993-10:2010: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
    • ISO 10993-11:2006: Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

    8. Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device, so there is no "training set" in the computational sense. The product development likely involved iterative design and testing, but not a formally defined "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. (See point 8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K200671

    Validate with FDA (Live)

    Date Cleared
    2020-06-19

    (98 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    1 - 5
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable, powder free examination gloves. The glove is made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the glove to make donning easy.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves Ansell Healthcare Products LLC's Microflex® Nitrile Patient Examination Gloves (Blue, Green, and Pink colored) meet these criteria for use with chemotherapy drugs and, for the pink glove, Fentanyl Citrate.

    Here's the breakdown as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the gloves are based on various ASTM standards and ISO standards for biocompatibility, in addition to performance against specific chemotherapy drugs. The reported device performance is directly compared against these standards and against a legally marketed predicate device (Micro-Touch NitraTex Sterile Nitrile Powder-Free Blue Examination Gloves, K082457).

    Acceptance Criteria & Device Performance – Summary Table

    CharacteristicStandard/Acceptance CriteriaReported Device Performance (Microflex® Nitrile Patient Examination Gloves)
    Physical Properties:ASTM D6319-10 (Nitrile Examination Gloves for Medical Application)Meets ASTM D6319-10 requirements
    Dimensions:Length: Minimum 230mm; Palm width (XS-XL): 70±10 to 120±10 mm; Thickness (Finger, Palm): minimum 0.05 mmLength: Minimum 240mm; Palm width (XS-XL): 75±5 to 115±5 mm; Thickness (Finger, Palm, Cuff): Finger - min 0.09mm, Palm - min 0.06mm, Cuff - min 0.05mm.
    Tensile Strength:Before Aging: minimum 14 MPa; After Aging: minimum 14 MPaBefore Aging: minimum 16 MPa; After Aging: minimum 14 MPa
    Ultimate Elongation:Before Aging: minimum 500%; After Aging: minimum 400%Before Aging: minimum 500%; After Aging: minimum 400%
    Freedom from holes:ASTM D6319-10 & ASTM D5151-06 requirements of AQL 2.5Meets or exceeds ASTM D6319-10 and ASTM D5151-06 requirements of AQL 2.5
    Powder Residual:ASTM D6319-10 & ASTM D6124-06: ≤ 2 mg per gloveMeets applicable requirement for powder free; ≤ 2 mg per glove
    Biocompatibility:ISO 10993-5:2009 (Cytotoxicity), ISO 10993-10:2010 (Skin Irritation, Sensitization), ISO 10993-11: 2006 (Acute Systemic Toxicity)Cytotoxicity: Under the conditions of the study, undiluted, 1:2, 1:4, 1:8, 1:16 dilution was cytotoxic. 1:32 and 1:64 are not cytotoxic. Skin Irritation: Under the conditions of the study, not an irritant.Sensitization: Under the conditions of the study, not a sensitizer.Acute Systemic Toxicity: Under the conditions of the study, no evidence of systemic toxicity.
    Chemotherapy Drug Permeation (ASTM D6978-05):Predicate Device Permeation Times (for comparison)Blue Gloves:
    Carmustine (BCNU) - 3.3 mg/mlPredicate: 32.1 minBlue: 17.4 min; Green: 27.9 min; Pink: 23.4 min
    Cyclophosphamide - 20.0 mg/mlPredicate: >240 minAll: >240 min
    Doxorubicin HCl - 2.0 mg/mlPredicate: >240 minAll: >240 min
    Etoposide - 20.0 mg/mlPredicate: >240 minAll: >240 min
    Fentanyl Citrate - 100 mcg/2mLN/A (Tested for Pink only)Pink: >240 min
    Fluorouracil - 50.0 mg/mlPredicate: >240 minAll: >240 min
    Methotrexate - 25.0 mg/mlPredicate: >240 minAll: >240 min
    Paclitaxel - 6.0 mg/mlPredicate: >240 minAll: >240 min
    Thiotepa (THT) - 10.0 mg/mlPredicate: 140.7 minBlue: 67.1 min; Green: 48.6 min; Pink: 64.9 min
    Vincristine Sulfate - 1.0 mg/mlPredicate: >240 minAll: >240 min
    Warning: Carmustine and Thiotepa have low permeation times. The warning "Do not use with Carmustine" is consistently included for all glove types.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify the exact sample sizes (number of gloves or test replicates) used for each physical, chemical (chemotherapy permeation), and biocompatibility test. It only provides the summarized results (e.g., minimum values, average breakthrough times, or "meets requirements").
    • Data Provenance: The data comes from the testing performed by the manufacturer, Ansell Healthcare Products LLC, for their 510(k) premarket notification. The document itself does not specify the country of origin where the testing was physically conducted. The data is retrospective in the sense that it was generated for the purpose of this submission, rather than a prospective clinical trial.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • This submission is for medical gloves and relies on laboratory-based performance testing against established ASTM and ISO standards, not human expert interpretation of medical images or patient data. Therefore, the concept of "ground truth" established by a panel of medical experts (like radiologists for imaging devices) is not applicable here. The "ground truth" is defined by the objective measurement protocols outlined in the referenced standards (e.g., ASTM D6319-10 for physical characteristics, ASTM D6978-05 for chemotherapy drug permeation, ISO 10993 series for biocompatibility). The testing would have been conducted by trained laboratory personnel in accordance with these validated methods.

    4. Adjudication Method for the Test Set

    • As the "ground truth" is based on objective laboratory measurements against defined standards, there is no adjudication method by multiple human reviewers in the context of medical image interpretation. The testing results are quantitative and directly compared to the numerical criteria specified in the standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is a medical glove, not an AI-assisted diagnostic tool. Therefore, the concept of assessing human reader improvement with or without AI assistance is not relevant to this submission. The device's performance is measured by its physical properties, barrier protection, and biocompatibility.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done

    • No standalone algorithm performance study was done. This device is a physical product (a glove), not a software algorithm. Its performance is assessed through laboratory tests as described above.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • The "ground truth" for this device is based on objective, standardized laboratory test methods and their defined acceptance criteria. This includes:
      • Physical Properties: Measurements (dimensions, tensile strength, elongation) compared against ASTM D6319-10.
      • Barrier Integrity: Freedom from holes assessed against ASTM D6319-10 and ASTM D5151-06 (AQL 2.5).
      • Chemical Permeation: Breakthrough detection times for chemotherapy drugs and Fentanyl Citrate determined per ASTM D6978-05.
      • Biocompatibility: In vitro and in vivo testing for cytotoxicity, skin irritation, sensitization, and acute systemic toxicity evaluated against ISO 10993 series.

    8. The Sample Size for the Training Set

    • This submission describes a medical device (gloves), not an AI/ML algorithm. Therefore, the concept of a "training set" for model development is not applicable. The gloves were manufactured and then tested according to established quality control and regulatory requirements.

    9. How the Ground Truth for the Training Set was Established

    • As there is no training set for an AI/ML model, there is no ground truth established for a training set. The device's performance is demonstrated through direct testing against predefined physical, chemical, and biological standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190018

    Validate with FDA (Live)

    Date Cleared
    2019-05-23

    (139 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gammex Non Latex PI White Surgical Gloves Tested for Use with Chemotherapy Drugs are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Gammex Non Latex Pl White Surgical Gloves are sterile and disposable devices. Gloves are made of synthetic polyisoprene rubber. A polyurethane polymer coating is applied to the inner surface of the glove to make donning easy. Ansell has performed independent laboratory testing on the already cleared device to support a labeling modification for the subject device that the gloves are safe and ling chemotherapy agents

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA for a medical device: Gammex Non Latex Polyisoprene White Surgical Gloves Tested for Use with Chemotherapy Drugs. The document details the substantial equivalence determination for the device.

    Based on the provided text, the device in question is surgical gloves, not an AI/algorithm-based diagnostic or assistive device. Therefore, many of the requested elements of the prompt (ground truth establishment for training/test sets, expert adjudication, MRMC studies, AI effect size, etc.) are not applicable to this type of medical device submission.

    The "acceptance criteria" for a physical device like a glove are typically performance specifications based on established standards (e.g., ASTM standards for dimensions, physical properties, lack of holes, chemical permeation). The "study that proves the device meets the acceptance criteria" refers to non-clinical laboratory testing.

    Here's an attempt to answer the prompt using the available information, noting where AI-specific questions are not relevant:


    Device: Gammex Non Latex Polyisoprene White Surgical Gloves Tested for Use with Chemotherapy Drugs

    Type of Study: Non-clinical laboratory testing to assess the physical properties, chemical permeation resistance, and biocompatibility of the surgical gloves, demonstrating substantial equivalence to a predicate device.

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (from ASTM standards/device specifications)Reported Device Performance (Result Summary)
    Dimensions (ASTM D3577)Meets ASTM D3577 requirements for length, width, and thickness.
    - Length: Minimum 265mmAverage 301mm
    - Palm Width (per size): e.g., 5.5 (70±6mm), 6.0 (76±6mm) etc.Average values: 5.5 (74mm), 6.0 (80mm), 6.5 (86mm), 7.0 (92mm), 7.5 (97mm), 8.0 (103mm), 8.5 (111mm), 9.0 (119mm)
    - Thickness (Finger): Minimum 0.10mm0.233mm
    - Thickness (Palm): Minimum 0.10mm0.172mm
    - Thickness (Cuff): Minimum 0.10mm0.212mm
    Physical Properties (ASTM D3577-09)Meets ASTM D3577-09 requirements for tensile strength, ultimate elongation, and stress at 500% elongation before and after accelerated aging for synthetic surgical gloves.
    Freedom from holes (ASTM D3577-09, ASTM D5151-06)Meets ASTM D3577-09 and ASTM D5151-06 requirements of AQL 1.5. (Note: Comparison table implies a stricter AQL for subject device at 0.65 vs. predicate's 1.5, suggesting better performance).
    Powder-Free (ASTM D3577-09, ASTM D6124-06)Meets Applicable requirement for Powder Free; ≤ 2 mg per glove. Average residual powder content for the glove during process validation is 0.16mg per glove.
    Sterility (ANSI/AAMI/ISO 11137-1:2006)Meets ANSI/AAMI/ISO 11137-1:2006 requirement of 10-6 SAL.
    Biocompatibility:
    - ISO in vitro cytotoxicity (ISO 10993-5:2009)Under the conditions of the study, the device was found to be cytotoxic and therefore the device extracts were evaluated by ISO 10993-11 - Test for systemic toxicity. From Acute Systemic Toxicity device extracts, the device extracts did not elicit a systemic response in the animal model. (Note: Follow-up systemic toxicity test showed no systemic effect, addressing initial in vitro cytotox result)
    - ISO Skin Irritation Study (ISO 10993-10:2010)Under the conditions of the study, not an irritant.
    - ISO Maximization Sensitization Study (ISO 10993-10:2010)Under the conditions of the study, not a sensitizer.
    - ISO Systemic Toxicity Study (ISO 10993-11:2006)Under the conditions of the study, there was no mortality or evidence of systemic toxicity.
    Chemotherapy Permeation Standard (ASTM D6978-05(2013))Under the conditions of the study, the permeation is acceptable for most tested drugs (>240 minutes breakthrough time for most). Specific low breakthrough times noted for Carmustine (10.2 minutes) and Thiotepa (11.5 minutes), leading to a warning not to use with these drugs.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Testing: The document does not explicitly state the total number of gloves or samples tested for each criterion. It mentions "Minimum Breakthrough Detection Time (Sample 1,2,3)" in one table (page 5), implying at least 3 samples were used for chemotherapy permeation testing for each chemical. Standard testing protocols (like ASTM) would dictate specific sample sizes for each test.
    • Data Provenance: The testing was "independent laboratory testing" performed by Ansell. The country of origin of the data is not specified, but the submission is to the U.S. FDA by an LLC based in Reno, Nevada, USA. The data is retrospective in the sense that it was collected prior to this submission for regulatory review.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. For a physical device like surgical gloves, "ground truth" is established by adherence to recognized national and international testing standards (e.g., ASTM, ISO). These standards define the methodologies and acceptance limits for physical, chemical, and biological performance. Experts involved would be laboratory technicians and engineers specializing in materials testing and biocompatibility, but not in the sense of clinical "expert readers" for diagnostic images.

    4. Adjudication method for the test set:

    • Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used for establishing consensus "ground truth" in subjective assessments, such as radiology interpretations or clinical diagnoses. For physical device testing, results are typically quantitative measurementsagainst predefined criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a physical medical device (surgical gloves), not an AI-based diagnostic or assistive technology for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical device, not an algorithm. Standalone performance is assessed through the non-clinical laboratory tests showing the glove's inherent properties and resistance.

    7. The type of ground truth used:

    • Standard-based performance metrics and established laboratory methods. The "ground truth" for this device is based on meeting the performance requirements outlined in:
      • ASTM D3577 (Standard Specification for Rubber Surgical Gloves)
      • ASTM D5151 (Standard Test Method for Detection of Holes in Medical Gloves)
      • ASTM D6124 (Standard Test Method for Residual Powder on Medical Gloves)
      • ANSI/AAMI/ISO 11137-1:2006 (Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices)
      • ISO 10993 (Biological evaluation of medical devices), specifically Parts 5 (cytotoxicity), 10 (irritation and sensitization), and 11 (systemic toxicity).
      • ASTM D6978-05 (Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs).

    8. The sample size for the training set:

    • Not Applicable. There is no "training set" in the context of an AI algorithm for this type of device submission. The device itself is manufactured; it does not "learn" from data.

    9. How the ground truth for the training set was established:

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182948

    Validate with FDA (Live)

    Date Cleared
    2019-03-22

    (150 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gammex PI Hybrid Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Gammex Pl Hybrid Surgical Gloves are sterile and disposable devices. Gloves are made of synthetic rubber blend of polyisoprene and polychoroprene. A polyurethane polymer coating is applied to the inner surface of the glove to make donning eay. Ansell has performed independent laboratory testing on the already cleared device to support a labeling modification for the subject device that the gloves are safe and effective for handling chemotherapy agents.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of the Gammex PI Hybrid Surgical Gloves, specifically regarding their resistance to chemotherapy drug permeation.

    Here's the breakdown of the information requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for chemotherapy drug permeation are implicitly defined by the reported "Minimum Breakthrough Detection Time" for each drug when tested according to ASTM D6978-05. A longer breakthrough time indicates better performance. For most drugs, the performance (>240 minutes) suggests that the gloves resist permeation for at least 4 hours, which is likely the desired acceptance criterion for practical use. The specific acceptance criteria for most physical/biological aspects are meeting the referenced ASTM and ISO standards.

    Acceptance Criteria (Standard/Requirement)Reported Device Performance (Gammex PI Hybrid Surgical Gloves)
    Chemotherapy Permeation (ASTM D6978-05)
    Bleomycin sulfate (15.0 mg/ml)>240 minutes
    Busulfan (6.0 mg/ml)>240 minutes
    Carboplatin (10.0 mg/ml)>240 minutes
    Carmustine (3.3 mg/ml)23.6 minutes (Individual samples: 23.8, 26.5, 23.6)
    Cisplatin (1.0 mg/ml)>240 minutes
    Cyclophosphamide (20.0 mg/ml)>240 minutes
    Cytarabine (100.0 mg/ml)>240 minutes
    Dacarbazine (10 mg/ml)>240 minutes
    Daunorubicin (5.0 mg/ml)>240 minutes
    Docetaxel (10.0 mg/ml)>240 minutes
    Doxorubicin Hydrochloride (2.0 mg/ml)>240 minutes
    Epirubicin (2.0 mg/ml)>240 minutes
    Etoposide (20.0 mg/ml)>240 minutes
    Fludarabine (25.0 mg/ml)>240 minutes
    Fluorouracil (50.0 mg/ml)>240 minutes
    Gemcitabine (Gemzar) (38.0 mg/ml)>240 minutes
    Idarubicin (1.0 mg/ml)>240 minutes
    Ifosfamide (50.0 mg/ml)>240 minutes
    Irinotecan (20.0 mg/ml)>240 minutes
    Mechlorethamine HCl (1.0 mg/ml)>240 minutes
    Melphalan (5.0 mg/ml)>240 minutes
    Methotrexate (25.0 mg/ml)>240 minutes
    Mitomycin C (0.5 mg/ml)>240 minutes
    Mitoxantrone (2.0 mg/ml)>240 minutes
    Oxaliplatin (2.0 mg/ml)>240 minutes
    Paclitaxel (Taxol) (6.0 mg/ml)>240 minutes
    Rituximab (10.0 mg/ml)>240 minutes
    ThioTEPA (10.0 mg/ml)36.1 minutes (Individual samples: 36.1, 37.0, 36.7)
    Vincristine Sulfate (1.0 mg/ml)>240 minutes
    Dimensions (ASTM D3577-09)Meets ASTM D3577-09 requirements for length, width, and thickness.
    Length: Minimum 265mmAverage 303mm
    Palm Width (e.g., size 7: 89±6mm)Average values provided for sizes 5.5 to 9 (e.g., size 7: 91mm).
    Finger Thickness: Minimum 0.10mm0.22mm
    Palm Thickness: Minimum 0.10mm0.2mm
    Cuff Thickness: Minimum 0.10mm0.15mm
    Physical Properties (ASTM D3577-09)Meets ASTM D3577-09 requirements for tensile strength and elongation at break before and after accelerated aging.
    Freedom from holes (ASTM D3577-09 & ASTM D5151-06)Meets AQL 1.5 requirements.
    ASTM D3577-09 Inspection level/AQL: GI/AQL 0.65; Lot Size: 35,001 to 150,000 pieces; Sample size: 100 pieces; Accept number: 0Actual found: 0
    Powder-Free (ASTM D3577-09 & ASTM D6124-06)Meets applicable requirement for Powder Free; ≤ 2 mg per glove. Averaged residual powder content for the glove during process validation is 0.48mg per glove.
    Sterility (ANSI/AAMI/ISO 11137-1:2006)Meets requirement of 10-6 SAL.
    Biocompatibility:
    ISO in vitro cytotoxicity (ISO 10993-5:2009)Device extract was found to be cytotoxic. (However, ISO 10993-11 systemic toxicity test showed no systemic response in the animal model, which is typically the follow-up for cytotoxic findings).
    ISO Skin Irritation Study (ISO 10993-10:2010)Not an irritant.
    ISO Maximization Sensitization Study (ISO 10993-10:2010)Not a sensitizer.
    ISO Systemic Toxicity Study (ISO 10993-11:2006)No mortality or evidence of systemic toxicity.

    2. Sample Size Used for the Test Set and Data Provenance

    • Chemotherapy Permeation: For the permation testing, the table in section {5} (and {2}) indicates that the "Minimum Breakthrough Detection Time" is based on 3 samples (Sample 1, 2, 3), specifically shown for Carmustine and ThioTEPA. For other drugs, only the aggregate ">240" is given, implying consistent results across samples.
    • Freedom from holes: For the specific ASTM D3577-09 test mentioned in section {6}, the sample size for this attribute was 100 pieces for a lot size of 35,001 to 150,000 pieces.
    • Powder-Free: The average residual powder content (0.48mg per glove) was determined "during process validation," but the specific sample size for this validation is not detailed.
    • Other tests (Dimensions, Physical Properties, Biocompatibility, Sterility): The sample sizes for these tests are not explicitly stated in the provided documents, but they would have been conducted according to the requirements of the respective ASTM/ISO standards.
    • Data Provenance: The studies were conducted by Ansell Healthcare Products LLC, presumably at an independent laboratory, to support a labeling modification. The data is retrospective in the sense that it was performed on an already cleared device (K151694) to add a new claim (chemotherapy drug handling), rather than being part of the initial device development for the base glove. The country of origin of the data is not specified, but the applicant is based in the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the tests performed are objective, laboratory-based chemical and physical property tests, not interpretive studies requiring human expert assessment (like image analysis or clinical diagnosis). The "ground truth" is established by direct measurement of permeation time or other physical/chemical properties according to standardized methods.

    4. Adjudication Method for the Test Set

    This information is not applicable. As these are laboratory measurements, there is no adjudication method in the sense of reconciling different expert opinions. The results are quantitative measurements against a predefined standard. For instance, breakthrough time is measured directly using analytical techniques.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "A clinical study was not conducted on the subject or predicate devices." The evaluation focuses on the physical and chemical resistance properties of the gloves, not on human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This device (surgical gloves) is a physical product and does not involve any artificial intelligence algorithms or software. The evaluation is solely on the physical and chemical properties of the gloves themselves.

    7. The Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Direct Physical and Chemical Measurements: For chemotherapy permeation, dimensions, physical properties, freedom from holes, powder residue, and sterility.
    • Standardized Biological Assay Results: For biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity).
      These are objective, quantitative results obtained through testing methods specified in internationally recognized standards (ASTM, ISO).

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical product (surgical gloves) and does not involve any machine learning algorithms that require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as point 8. No training set or associated ground truth was required for the evaluation of this physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171375

    Validate with FDA (Live)

    Date Cleared
    2018-02-06

    (272 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gammex PI Breach Detect Powder Free Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. When used as an underglove, the glove is intended to promote visibility of a fluid that breaches/enters the glove via an increased spreading rate.

    Device Description

    Gammex Pl Breach Detect Powder Free Surgical Glove is a sterile, green in color, single use disposable powder free surgical glove. The glove is made of synthetic polyisoprene rubber. A polymer is applied to the inner surface of the glove to make donning easy. The glove's outer surface is coated with a surface-active agent and when used as an under-glove, this coating and the glove's green color can improve visibility of a glove breach by promoting wetting of the glove surface and spreading of bodily fluid penetrating the breach.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device called "Gammex PI Breach Detect Powder Free Surgical Glove" (K171375). However, the document does not contain the kind of detailed information about acceptance criteria and performance study results that would typically be found for an AI/ML-driven medical device, such as a diagnostic algorithm.

    The device in question is a surgical glove, which is a Class I device. The FDA's review for Class I devices generally focuses on demonstrating substantial equivalence to a predicate device based on defined physical, material, and biocompatibility characteristics, rather than extensive clinical performance studies or AI model validation.

    Therefore, many of the specific questions you've asked (e.g., sample sizes for test sets, number of experts for ground truth, MRMC studies, standalone performance of an algorithm, ground truth for training sets) are not applicable to this device and the information provided in this 510(k) summary.

    Here's a breakdown based on the provided text, addressing what is available and explaining why other aspects are absent:

    1. A table of acceptance criteria and the reported device performance

    The document provides a table outlining "Technological Characteristics" compared to ASTM or equivalent standards. These function as the acceptance criteria for a surgical glove.

    CharacteristicsStandard/Test/FDA GuidanceAcceptance Criteria (Implied by Standard)Reported Device Performance (Result Summary)
    Physical Characteristics:
    Dimensions: LengthASTM D3577-09Minimum 265mmAverage value: Minimum 297 mm (Meets)
    Dimensions: Palm Width (mm)ASTM D3577-09Size 5.5: $70 \pm 6$ Size 6.0: $76 \pm 6$ Size 6.5: $83 \pm 6$ Size 7.0: $89 \pm 6$ Size 7.5: $95 \pm 6$ Size 8.0: $102 \pm 6$ Size 8.5: $108 \pm 6$ Size 9.0: $114 \pm 6$Size 5.5: 76 Size 6.0: 82 Size 6.5: 86 Size 7.0: 91 Size 7.5: 98 Size 8.0: 100 Size 8.5: 112 Size 9.0: 120 (Meets)
    Dimensions: Thickness (mm) – single wall (Finger, Palm, Cuff)ASTM D3577-09Minimum 0.10 for Finger, Palm, CuffFinger: 0.25 Palm: 0.21 Cuff: 0.17 (Meets)
    Physical Properties:ASTM D3577-09Requirements for tensile strength, ultimate elongation, and stress at 500% elongation before and after accelerated aging for synthetic surgical gloves.Meets ASTM D3577-09 requirements (Meets)
    Freedom from holesASTM D3577-09, ASTM D5151-06AQL 1.5Meets ASTM D3577-09 and ASTM D5151-06 requirements of AQL 1.5 (Meets)
    Powder ResidualASTM D3577-09, ASTM D6124-06≤ 2 mg per gloveMeets applicable requirement for powder free; ≤ 2 mg per glove (Meets)
    SterilityANSI/AAMI/ISO 11137-1:2006$10^{-6}$ SAL (Sterility Assurance Level)Meets ANSI/AAMI/ISO 11137-1:2006 requirement of $10^{-6}$ SAL. (Meets)
    Biocompatibility:
    ISO In Vitro CytotoxicityISO 10993-5:2009Undiluted and 1:2 dilution not cytotoxic.Under the conditions of the study, undiluted and 1:2 dilution were cytotoxic. 1:4, 1:8, 1:16, 1:32 and 1:64 are not cytotoxic. (Note: This result is reported rather than a pass/fail against typical acceptance of "not cytotoxic at all concentrations". However, the overall conclusion for substantial equivalence indicates this was acceptable.)
    ISO Skin Irritation StudyISO10993-10:2010Not an irritantUnder the conditions of the study, not an irritant (Meets)
    ISO Maximization Sensitization StudyISO 10993-10:2010Not a sensitizerUnder the conditions of the study, not a sensitizer (Meets)
    ISO acute systemic toxicityISO 10993-11: 2006No evidence of systemic toxicityUnder the conditions of the study, no evidence of systemic toxicity (Meets)

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size: Not explicitly stated as a "test set" in the context of an AI/ML algorithm. The performance data refers to standard compliance tests (e.g., ASTM standards for glove characteristics). These tests involve sampling plans as defined within the respective standards, but the specific number of gloves tested for each characteristic is not provided in this summary.
    • Data Provenance: Not applicable. These are laboratory test results based on manufactured glove samples. The data is not patient-derived from specific countries or retrospective/prospective studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This device is a physical product (surgical glove), not an AI/ML diagnostic device requiring expert interpretation of medical images or data for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-assisted diagnostic device. The document explicitly states: "A clinical study was not conducted on the subject or predicate devices."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. There is no algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Ground truth for this device is based on established engineering standards and material science properties (e.g., ASTM standards for dimensions, tensile strength, freedom from holes, powder residue; ISO standards for biocompatibility and sterility). The "truth" is whether the glove meets the specified physical and biological performance characteristics.

    8. The sample size for the training set:

    • Not Applicable. There is no AI/ML model for which to define a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171737

    Validate with FDA (Live)

    Date Cleared
    2017-12-29

    (200 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Gloves are intended for medical purposes that are worn on the examiners hands to prevent contamination between patient and examiner.

    Device Description

    The MICRO-TOUCH® DENTA-GLOVE® Nitrile HydraSoft™ are non-sterile, single use only, disposable, powder free examination gloves. The glove is made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the glove to make donning easy. Hydrasoft coating (containing glycerine) is applied on top of the polymer coating on the glove inner surface..

    Characteristic:

    • Ambidextrous with beaded cuff and straight fingers
    • Finger-textured,
    • White colored
    • Featuring inner coating of polyacrylic polymer coating and HydraSoft™ coating.
    • . Five (5) sizes – extra-small, small, medium, large, and extra-large.
    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for a medical device, the "Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Gloves."

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicsAcceptance Criteria (Standard/Test/FDA Guidance)Reported Device Performance (Result Summary)
    Physical Characteristics:ASTM D6319-10Meets ASTM D6319-10 requirements for length, width and thickness
    LengthMinimum 230mmMinimum 240mm
    Palm width (XS)70 ± 10 mm75 ± 5 mm
    Palm width (S)80 ± 10 mm85 ± 5 mm
    Palm width (M)95 ± 10 mm95 ± 5 mm
    Palm width (L)110 ± 10 mm105 ± 5 mm
    Palm width (XL)120 ± 10 mm115 ± 5 mm
    Thickness (mm) - single-wallminimum 0.05 mm
    Fingerminimum 0.05 mm0.11 ± 0.03 mm
    Palmminimum 0.05 mm0.07 ± 0.02 mm
    CuffN/A (implied by minimum 0.05 for other areas)0.06 ± 0.02 mm
    Physical Properties:ASTM D6319-10Meets ASTM D6319-10 requirements for tensile strength and ultimate elongation before and after accelerated aging
    Tensile Strength
    Before Agingminimum 14 MPaminimum 17 MPa
    After Agingminimum 14 MPaminimum 17 MPa
    Ultimate Elongation
    Before Agingminimum 500%minimum 500%
    After Agingminimum 400%minimum 400%
    Freedom from holesASTM D6319-10, ASTM D5151-06 (AQL 2.5)Meets ASTM D6319-10 and ASTM D5151-06 requirements of AQL 2.5
    Powder ResidualASTM D6319-10, ASTM D6124-06 (≤ 2 mg per glove)Meets applicable requirement for powder free; ≤ 2 mg per glove
    Biocompatibility:
    ISO In Vitro CytotoxicityISO 10993-5:2009Under the conditions of the study, undiluted and 1:2 dilution was cytotoxic. 1:4, 1:8, 1:16, 1:32 and 1:64 are not cytotoxic
    ISO Skin Irritation StudyISO 10993-10:2010Under the conditions of the study, not an irritant
    ISO Maximization Sensitization StudyISO 10993-10:2010Under the conditions of the study, not a sensitizer
    ISO acute systemic toxicityISO 10993-11: 2006Under the conditions of the study, no evidence of systemic toxicity

    2. Sample Size Used for the Test Set and Data Provenance:

    The document focuses on non-clinical performance data and refers to established ASTM and ISO standards for testing. Specific sample sizes for each test are not explicitly stated in this summary but would be defined within the referenced ASTM and ISO standards. The data provenance is derived from these standard-compliant tests conducted on the device. No information on country of origin of the data is provided, nor whether it is retrospective or prospective, as these terms typically apply to clinical studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable to this document. The "ground truth" for evaluating this device's performance is based on established engineering and biocompatibility standards (ASTM and ISO), not expert consensus in a clinical context.

    4. Adjudication Method for the Test Set:

    This information is not applicable. Adjudication methods are typically used in clinical studies involving human readers/interpreters, not for performance testing against established engineering standards like those for examination gloves.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "A clinical study was not conducted on the subject or predicate devices." This type of study would involve human readers (e.g., medical professionals) evaluating cases, which is not relevant for this device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    This information is not applicable given the device is a patient examination glove, not an AI algorithm. The performance evaluation is based on direct physical and chemical testing of the glove itself.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance evaluation of the Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Gloves are established international and national standards for medical devices, specifically:

    • ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical Application.
    • ASTM D5151-06: Standard Test Method for Detection of Holes in Medical Gloves.
    • ASTM D6124-06: Standard Test Method for Residual Powder on Medical Gloves.
    • ISO 10993-5:2009: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.
    • ISO 10993-10:2010: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization.
    • ISO 10993-11:2006: Biological evaluation of medical devices – Part 11: Tests for systemic toxicity.

    8. The Sample Size for the Training Set:

    This information is not applicable. "Training set" refers to data used to train machine learning algorithms. This device is a physical product, not an AI algorithm, and therefore does not have a "training set" in that context.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4