Found 1 results
510(k) Data Aggregation
(109 days)
The Medical Disposable Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided nonsterile.
The Medical Disposable Face Mask is blue color, single use, three-layer, flat-pleated mask with nose piece and ear loop. The blue colorant is polypropylene (PP) masterbatch. The Medical Face Mask is manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are made of polyester and spandex, not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable aluminum with polypropylene (PP) covering. Users can adjust the nose piece according to the shape of the bridge of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. The medical face masks are available in two size specifications, 17.5×9.5cm, 14.5×9.5cm. The medical face masks are sold non-sterile and are intended to be single use, disposable devices.
The provided document is a 510(k) summary for a medical device called "Medical Disposable Face Masks." It is a regulatory submission to the FDA demonstrating substantial equivalence to a legally marketed predicate device. This document does not contain information about an AI/ML-driven medical device, but rather a standard medical device (face masks).
Therefore, the specific information requested about acceptance criteria, study design for AI/ML performance, ground truth establishment, expert adjudication, MRMC studies, or training sets is not applicable to this document. The document describes non-clinical performance testing for the physical properties and biological compatibility of face masks.
However, I can extract the acceptance criteria and reported performance for the face mask product, as presented in the document:
1. Table of Acceptance Criteria and Reported Device Performance (for Face Masks)
| Test Standard | Performance Criteria (Acceptance Criteria) | Reported Device Performance |
|---|---|---|
| Fluid Resistance ASTM F1862 | 32 out of 32 per lot pass at 80 mmHg, 3 non-consecutive lots tested | 32 out of 32 per lot pass at 80 mmHg, 3 non-consecutive lots tested |
| Particulate Filtration Efficiency ASTM F2299 | ≥ 95% | ≥ 95% |
| Bacterial Filtration Efficiency ASTM F2101 | ≥ 95% | ≥ 95% |
| Differential Pressure (Delta P) EN14683 Annex C | < 5.0mmH2O/cm² | < 5.0mmH2O/cm² |
| Flammability 16 CFR 1610 | Class 1 | Class 1 |
| Biocompatibility ISO 10993-1 | Meets requirements of ISO 10993-1 for Cytotoxicity, Skin Irritation, Sensitization | Met all evaluation acceptance criteria for Cytotoxicity, Irritation, and Sensitization |
Information Not Applicable to this Document (as it's not an AI/ML device)
The remaining points of your request are specifically for an AI/ML medical device and are not present or applicable in this document for a physical medical product like face masks:
- Sample size used for the test set and the data provenance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
- Adjudication method for the test set.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, and the effect size.
- If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
- The sample size for the training set.
- How the ground truth for the training set was established.
This document clearly states: "No clinical study is included in this submission." (Page 7, Section 9). The testing performed is non-clinical and pertains to the physical and biological characteristics of the face mask material and construction.